Endress+Hauser OUM980, OUSAF21 Operating Instructions Manual

Operating Instructions
Model OUM980 Photometric Analyzer/ Model OUSAF21 Dual Beam Inline Low Level Color Sensor
P/N 71105623 Rev 1 BA930C/07/en/10.09
Model OUM980 with OUSAF21 Inline Sensor
2
Model OUM980 with OUSAF21 Inline Sensor
Table of Contents
1. SAFETY INSTRUCTIONS........................................................................................................................................... 5
1.1 DESIGNATED USE ................................................................................................................................................................5
1.2 INSTALLATION, START-UP AND OPERATION .............................................................................................................................5
1.3 OPERATIONAL SAFETY ..........................................................................................................................................................5
1.4 RETURN .............................................................................................................................................................................6
2. GENERAL INFORMATION ....................................................................................................................................... 7
2.1 HOW THE MODEL 980 PHOTOMETRIC ANALYZER WORKS .........................................................................................................7
2.2 CONCENTRATION AND ABSORBANCE UNITS .............................................................................................................................7
2.3 EASY INTERFACING ..............................................................................................................................................................7
3. INSTALLATION......................................................................................................................................................... 9
3.1 MODEL OUM980 ANALYZER INSTALLATION...........................................................................................................................9
3.2 MODEL OUM980 CABLES AND WIRING ..............................................................................................................................10
3.3 DC INPUT POWER OPTION .................................................................................................................................................11
3.4 MODEL OUSAF21 SENSOR INSTALLATION............................................................................................................................ 12
3.5 CABLE OUK20 STRUCTURE AND TERMINATION OF MEASUREMENT CABLE................................................................................. 13
3.6 POST-CONNECTION CHECK ................................................................................................................................................14
4. OPERATION ............................................................................................................................................................ 15
4.1 QUICK OPERATION GUIDE ..................................................................................................................................................15
4.2 DISPLAY AND OPERATING ELEMENTS ....................................................................................................................................15
4.2.1 Operating Elements................................................................................................................................................15
4.2.2 Operation Legend...................................................................................................................................................16
4.3 LOCAL OPERATION ............................................................................................................................................................16
4.3.1 Measuring Mode ....................................................................................................................................................16
4.3.2 Configuration Mode ............................................................................................................................................... 18
4.4 REMOTE OPERATION..........................................................................................................................................................19
5. COMMISSIONING .................................................................................................................................................. 20
5.1 FUNCTION CHECK .............................................................................................................................................................20
5.2 SWITCHING ON.................................................................................................................................................................20
5.3 ANALYZER CONFIGURATION AND OPERATION ........................................................................................................................21
5.3.1 Factory Default Setting ........................................................................................................................................... 22
5.3.2 Sensor Calibration .................................................................................................................................................. 22
5.3.2.1 Liquid Calibration Procedure..............................................................................................................................22
5.3.2.2 Filter calibration Set up and Calibration .............................................................................................................. 24
5.3.3 Change Pathlength Setting......................................................................................................................................26
5.3.4 Setup Output Span ................................................................................................................................................. 27
5.3.5 Output Relay Set Point and Configuration...............................................................................................................28
5.3.6 Current Output Calibration ....................................................................................................................................30
5.3.7 Date and Time Setting............................................................................................................................................30
5.3.8 Configuration and Use of Digital Inputs...................................................................................................................31
5.3.9 Process Control ......................................................................................................................................................31
5.3.9.1 Access of Process Control...................................................................................................................................31
5.3.9.2 Baseline (AutoZero)............................................................................................................................................32
5.3.9.3 Optical Zero.......................................................................................................................................................32
5.3.9.4 Lamp Switch...................................................................................................................................................... 33
5.3.9.5 Optical Check ....................................................................................................................................................33
5.3.9.6 Password Protection........................................................................................................................................... 33
3
Model OUM980 with OUSAF21 Inline Sensor
5.3.9.7 Process Value Correlation...................................................................................................................................34
6. MAINTENANCE ...................................................................................................................................................... 37
6.1 MAINTENANCE OUTLINE ....................................................................................................................................................37
6.2 OUSAF21 MEASUREMENT DETECTOR/FILTER REPLACEMENT.................................................................................................40
6.3 SENSOR WAVELENGTH CHANGE ..........................................................................................................................................40
6.4 SENSOR WINDOW AND GASKET REPLACEMENT ......................................................................................................................41
7. REPLACEMENT PARTS LIST .................................................................................................................................. 42
7.1 MODEL 980 PHOTOMETRIC ANALYZER.................................................................................................................................42
7.2 MODEL OUSAF21 INLINE SENSOR ......................................................................................................................................42
8. TECHNICAL INFORMATION ................................................................................................................................. 43
8.1 PHOTOMETRIC ANALYZER OUM980....................................................................................................................................43
8.1.1 Input......................................................................................................................................................................43
8.1.2 Output...................................................................................................................................................................43
8.1.3 Power Supply.........................................................................................................................................................43
8.1.4 Performance...........................................................................................................................................................44
8.1.5 Environment ..........................................................................................................................................................44
8.1.6 Mechanical Construction........................................................................................................................................44
8.2 ABSORBANCE SENSOR OUSAF21......................................................................................................................................... 45
8.2.1 Sensor Dimensions .................................................................................................................................................45
8.2.2 Input and Output ...................................................................................................................................................45
8.2.3 Mechanic Construction ..........................................................................................................................................45
8.2.4 Environment ..........................................................................................................................................................46
8.2.5 Power Supply.........................................................................................................................................................46
8.2.6 Optical Specifications..............................................................................................................................................46
9. ORDERING INFORMATION ................................................................................................................................... 47
9.1 COLOR ANALYZER OUM980 ..............................................................................................................................................47
9.2 SENSOR OUSAF21............................................................................................................................................................48
9.3 OUK20 CABLE SET ...........................................................................................................................................................49
4
Model OUM980 with OUSAF21 Inline Sensor Safety Instructions
1. Safety Instructions
1.1 Designated Use
Integra Model OUM980 is an analyzer for determining the absorbance of a liquid medium at a specific wavelength
The analyzer is particularly suited for use in the following areas:
Bio-Technology industry Waste water processing Chemical processing industry Food industry Pharmaceutical industry
Any other use than the one described here compromises the safety of persons and the entire measuring system and is, therefore, not permitted. The manufacturer is not liable for damage caused by improper or non-designated use.
1.2 Installation, Start-Up and Operation
Please note the following items:
Installation, electrical connection, start-up, operation and maintenance of the measuring system must only be carried out
by trained technical personnel. The technical personnel must be authorized for the specified activities by the system operator.
Technical personnel must have read and understood these Operating Instructions and must adhere to them.
Before commissioning the entire measuring point, check all the connections for correctness. Ensure that electrical cables
and hose connections are not damaged.
Do not operate damaged products and secure them against unintentional commissioning. Mark the damaged product as
being defective.
Measuring point faults may only be rectified by authorized and specially trained personnel.
If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning.
Repairs not described in these Operating Instructions may only be carried out at the manufacturers or by the service
organization.
1.3 Operational Safety
Relevant regulations and standards have been met. As the user, you are responsible for complying with the following safety conditions:
Installation instructions Local prevailing standards and regulations.
EMC
This instrument has been tested for electromagnetic compatibility in industrial use according to applicable standards. Protection against interference as specified above is valid only for an instrument connected according to the instructions in these Operating Instructions.
5
Model OUM980 with OUSAF21 Inline Sensor Safety Instructions
1.4 Return
If the device requires repair, please contact your local sales and service representative. Refer to the website on the back page of this manual on where to find your local sales and service support office. Please use the original packaging, if possible.
Notes on safety icons and symbols
Safety icons
Warning!
This symbol alerts you to hazards. They can cause serious damage to the instrument or to persons if ignored.
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could cause damage to the instrument if ignored.
Note!
This symbol indicates important items of information.
Electrical symbols
Direct Current (DC)
A terminal at which DC is applied or through which DC flows.
Alternating Current (AC)
A terminal at which (sine-form) AC is applied or through which AC flows.
Ground connecting
A terminal is already grounded using a grounding system.
Protective earth terminal
A terminal which must be grounded before other connections may be set up.
Alarm relay
Input
Output
6
Model OUM980 with OUSAF21 Inline Sensor General Information
L
2. General Information
The Model OUM980 Photometric Analyzer is used in conjunction with a Model OUSAF21 sensor to measure the spectral absorbance of process liquid in the VIS/NIR region of the electromagnetic spectrum.
Dependent upon the optical pathlength of the connected sensor, the instrument can measure up to 50 Optical Density (OD) units.
The Model OUM980 Photometric Analyzer is manufactured from state of the art digital electronics. Plant interface is through quick disconnect screw terminals on the back of the module. The user interface is comprised of a 6-button tactile feel keypad and 4 x 20 character alphanumeric LCD display.
2.1 How the Model 980 Photometric Analyzer Works
The Model OUSAF21 sensor generates two photocurrent (nA) detector signals based upon the amount of energy present at its lamp source (reference) and measurement point internally. The Model OUM980 unit computes the logarithmic ratio of these two signals and determines the absorbance of the liquid passing through the sensor. The absorbance value is expressed in optical density units (OD) and is displayed on the front panel. Two galvanic isolated analog current outputs, proportional to the OD reading, are simultaneously transmitted for connection to other instrumentation and recording devices.
2.2 Concentration and Absorbance Units
The concentration of an optically absorbing material in a mixture can be determined since it is related to the amount of light absorbed from a beam of light passing through it. The absorbance of a substance is directly proportional to the concentration of the material that causes the absorption. The Lambert-Beer Law describes this relationship of absorbance (A) to concentration. Essentially, the amount of radiation transmitted through the absorbing material decreases logarithmically with its increasing concentration.
Where
A log
1
log
T
The above assumes that the optical pathlength remains constant. The optical density (OD) however, is defined as Absorption per unit length. Normalizing to an optical pathlength of 1cm, it follows that:
Io Ir
and
T
Ir Io
OD
Where OD = Optical Density
A = Absorbance
L = pathlength in cm
2.3 Easy Interfacing
The Model OUM980 Photometric Analyzer incorporates a front panel display and two analog current output signals. The display on the front panel is an alphanumeric 4 x 20 character LCD. The display may be configured for process variable or analog current output measurements. Optical Density (OD) readings are normalized to a 1cm pathlength.
)(1A
7
Model OUM980 with OUSAF21 Inline Sensor General Information
There are two analog current outputs from the unit. The range of each can be independently set to be any range within the measurement range of the instrument. One output is configured to operate with measurement baseline shift commands, while the other tracks full scale of the instrument.
The analog current outputs (4-20mA) will operate with loads up to 750 ohms. These outputs are galvanically isolated and conform to NAMURi Standard.
Model OUM980 is supplied as standard with four digital inputs and three digital output relays.
The digital inputs can be configured to perform a variety of functions such as baseline and sensor lamp off/on.
The outputs can be connected to a PLC for inclusion in a control scheme, for simple valve operation or for local indication (alarms) and annunciation. Alarm status indication is provided on the front panel display.
All digital inputs are optically isolated.
i
NAMUR standard limit detection when enabled is set to a BAD if the signal level is above 21 mA or below 3.6 mA for more than 4 seconds. The BAD status is cleared
when the signal returns to the normal 4-20 mA range.
8
Model OUM980 with OUSAF21 Inline Sensor Installation
3. Installation
3.1 Model OUM980 Analyzer Installation
Before starting installation, inspect the analyzer, sensor, and supplied cable set for any signs of shipping damage. Report any visual damage or discrepancies to the factory and the shipper immediately.
The Model OUM980 analyzer is a ¼ DIN enclosure which can be installed a variety of panel, wall and bench top housings. Refer to figure 3.1 for mounting dimensions. Mount or install the analyzer into an enclosure or area that is not subject to excessive vibration or shock and will protect the instrument from materials such as water and chemicals. Allow enough clearance behind it for cable access.
Figure 3.1: OUM980 Analyzer Mounting
9
Model OUM980 with OUSAF21 Inline Sensor Installation
3.2 Model OUM980 Cables and Wiring
All wiring terminals are located on the back panel of the Model OUM980. The analyzer/sensor interconnection cables supplied with the system have all been pre-terminated and labeled for ease of installation. The Model OUM980 has two (2) analog current outputs of 4 to 20mA. Both are capable of driving loads up to 750 ohms.
In addition to the analog outputs, three (3) galvanic isolated digital outputs (N.O.) are available. These outputs are rated for both AC and DC voltages up to 280 volts at 125mA. These outputs are intended for dry contact or pilot applications.
Insert Terminal
Here
1
2
3 4 5 6 7
Connector
8 9
Phoenix Cambion
10 11
12
Wire Pin Terminal
Figure 3.2: Wire Terminal Preparations
Cables installed for signal connection (i.e. analog outputs, lamp fail output) should be shielded twisted pairs.
Warning!
When routing the cables, separate signal cables from power wiring!
Caution!
Use dry contact only when connect digital input to analyzer!
Refer Figure 3.3 for wiring connection.
10
Model OUM980 with OUSAF21 Inline Sensor Installation
Figure 3.3: Model OUM980/OUSAF21 Dual Beam Wiring Diagram (AC Input Version)
3.3 DC Input Power Option
For instruments supplied for 24VDC operation, only the power input connection is changed. Figure 3.3 shows the connection detail for a 24VDC unit.
Input Power
+24VDC DC RTN
Ground
24 VDC
Figure 3.4: Integra DC Power Connection
11
Model OUM980 with OUSAF21 Inline Sensor Installation
3.4 Model OUSAF21 Sensor Installation
Sensors can be installed either directly in a process line or in a by-pass line. They can be mounted either horizontally or at an angle. If mounted horizontally, the sensor lamp and detector housings must be horizontal. This will insure that the optical window surfaces are in a vertical position, which will help prevent build up on the window surfaces. The sensor should be located upstream of pressure regulators. Operating sensors under pressure will help to avoid the possibility of air or gas bubble evolution, which can cause measurement noise and error.
When installing, allow adequate space for the connection of cables at the ends of the lamp and detector housings. Access to these areas is also important for connection/disconnection purposes. Sensor bodies should be supported when in line and care should be taken to ensure they are protected against damage caused by external forces such as carts on adjacent walkways.
Figure 3.5: Recommended Orientation for Sensor Installation
12
Model OUM980 with OUSAF21 Inline Sensor Installation
Figure 3.6: Typical OUSAF21 Sensor With Cables Connected.
3.5 Cable OUK20 Structure and Termination of Measurement Cable
Figure 3.7: OUSAF21 Cable at Lamp Side
13
Model OUM980 with OUSAF21 Inline Sensor Installation
Figure 3.8: OUSAF21 Cable at Detector Side
3.6 Post-Connection Check
After wiring up the electrical connection, carry out the following checks:
Device Status and Specifications
Is the analyzer or the cable externally damaged?
Electrical connection Remarks
Are the installed cables strain-relieved?
No loops or cross-overs in the cable run?
Are the signal cables correctly connected according to the wiring diagram?
Are all screw terminals tightened?
Are all cable entries installed, tightened and sealed?
Are the PE distributor rails grounded (if present)? Grounding at place of installation
Caution!
Improperly grounding will lead to unreliable measurement result!
Remark
Visual inspection
Remark
Grounding at place of installation
14
Model OUM980 with OUSAF21 Inline Sensor Operation
4. Operation
4.1 Quick Operation Guide
User has the following ways of operating the analyzer:
On site via the keypad Via the digital input interface to perform process control functions remotely.
4.2 Display and Operating Elements
4.2.1 Operating Elements
Figure 4.1: Operating Controls
1. LED for lamp warning functions and limit relay status.
2. Menu/ESC key serves two functions. Pressing Menu/ESC will always exit out of a screen and will not save any changes
Entered. Pressing Menu/ESC from the main PVii or Output screen will open the Sub-Menu screen.
3. Ack/ENT key accepts changes in a screen and continue forward through the screen
4. 4 keys navigate the user through the menus and screens.
5. LC display for displaying the measured values and configuration data LED.
ii
PV is the Process Value for PV1 and PV2 and corresponds to Output 1 and Output 2 respectively.
15
Loading...
+ 35 hidden pages