Endress+Hauser OUM980, OUSAF21 Operating Instructions Manual

Operating Instructions
Model OUM980 Photometric Analyzer/ Model OUSAF21 Dual Beam Inline Low Level Color Sensor
P/N 71105623 Rev 1 BA930C/07/en/10.09
Model OUM980 with OUSAF21 Inline Sensor
2
Model OUM980 with OUSAF21 Inline Sensor
Table of Contents
1. SAFETY INSTRUCTIONS........................................................................................................................................... 5
1.1 DESIGNATED USE ................................................................................................................................................................5
1.2 INSTALLATION, START-UP AND OPERATION .............................................................................................................................5
1.3 OPERATIONAL SAFETY ..........................................................................................................................................................5
1.4 RETURN .............................................................................................................................................................................6
2. GENERAL INFORMATION ....................................................................................................................................... 7
2.1 HOW THE MODEL 980 PHOTOMETRIC ANALYZER WORKS .........................................................................................................7
2.2 CONCENTRATION AND ABSORBANCE UNITS .............................................................................................................................7
2.3 EASY INTERFACING ..............................................................................................................................................................7
3. INSTALLATION......................................................................................................................................................... 9
3.1 MODEL OUM980 ANALYZER INSTALLATION...........................................................................................................................9
3.2 MODEL OUM980 CABLES AND WIRING ..............................................................................................................................10
3.3 DC INPUT POWER OPTION .................................................................................................................................................11
3.4 MODEL OUSAF21 SENSOR INSTALLATION............................................................................................................................ 12
3.5 CABLE OUK20 STRUCTURE AND TERMINATION OF MEASUREMENT CABLE................................................................................. 13
3.6 POST-CONNECTION CHECK ................................................................................................................................................14
4. OPERATION ............................................................................................................................................................ 15
4.1 QUICK OPERATION GUIDE ..................................................................................................................................................15
4.2 DISPLAY AND OPERATING ELEMENTS ....................................................................................................................................15
4.2.1 Operating Elements................................................................................................................................................15
4.2.2 Operation Legend...................................................................................................................................................16
4.3 LOCAL OPERATION ............................................................................................................................................................16
4.3.1 Measuring Mode ....................................................................................................................................................16
4.3.2 Configuration Mode ............................................................................................................................................... 18
4.4 REMOTE OPERATION..........................................................................................................................................................19
5. COMMISSIONING .................................................................................................................................................. 20
5.1 FUNCTION CHECK .............................................................................................................................................................20
5.2 SWITCHING ON.................................................................................................................................................................20
5.3 ANALYZER CONFIGURATION AND OPERATION ........................................................................................................................21
5.3.1 Factory Default Setting ........................................................................................................................................... 22
5.3.2 Sensor Calibration .................................................................................................................................................. 22
5.3.2.1 Liquid Calibration Procedure..............................................................................................................................22
5.3.2.2 Filter calibration Set up and Calibration .............................................................................................................. 24
5.3.3 Change Pathlength Setting......................................................................................................................................26
5.3.4 Setup Output Span ................................................................................................................................................. 27
5.3.5 Output Relay Set Point and Configuration...............................................................................................................28
5.3.6 Current Output Calibration ....................................................................................................................................30
5.3.7 Date and Time Setting............................................................................................................................................30
5.3.8 Configuration and Use of Digital Inputs...................................................................................................................31
5.3.9 Process Control ......................................................................................................................................................31
5.3.9.1 Access of Process Control...................................................................................................................................31
5.3.9.2 Baseline (AutoZero)............................................................................................................................................32
5.3.9.3 Optical Zero.......................................................................................................................................................32
5.3.9.4 Lamp Switch...................................................................................................................................................... 33
5.3.9.5 Optical Check ....................................................................................................................................................33
5.3.9.6 Password Protection........................................................................................................................................... 33
3
Model OUM980 with OUSAF21 Inline Sensor
5.3.9.7 Process Value Correlation...................................................................................................................................34
6. MAINTENANCE ...................................................................................................................................................... 37
6.1 MAINTENANCE OUTLINE ....................................................................................................................................................37
6.2 OUSAF21 MEASUREMENT DETECTOR/FILTER REPLACEMENT.................................................................................................40
6.3 SENSOR WAVELENGTH CHANGE ..........................................................................................................................................40
6.4 SENSOR WINDOW AND GASKET REPLACEMENT ......................................................................................................................41
7. REPLACEMENT PARTS LIST .................................................................................................................................. 42
7.1 MODEL 980 PHOTOMETRIC ANALYZER.................................................................................................................................42
7.2 MODEL OUSAF21 INLINE SENSOR ......................................................................................................................................42
8. TECHNICAL INFORMATION ................................................................................................................................. 43
8.1 PHOTOMETRIC ANALYZER OUM980....................................................................................................................................43
8.1.1 Input......................................................................................................................................................................43
8.1.2 Output...................................................................................................................................................................43
8.1.3 Power Supply.........................................................................................................................................................43
8.1.4 Performance...........................................................................................................................................................44
8.1.5 Environment ..........................................................................................................................................................44
8.1.6 Mechanical Construction........................................................................................................................................44
8.2 ABSORBANCE SENSOR OUSAF21......................................................................................................................................... 45
8.2.1 Sensor Dimensions .................................................................................................................................................45
8.2.2 Input and Output ...................................................................................................................................................45
8.2.3 Mechanic Construction ..........................................................................................................................................45
8.2.4 Environment ..........................................................................................................................................................46
8.2.5 Power Supply.........................................................................................................................................................46
8.2.6 Optical Specifications..............................................................................................................................................46
9. ORDERING INFORMATION ................................................................................................................................... 47
9.1 COLOR ANALYZER OUM980 ..............................................................................................................................................47
9.2 SENSOR OUSAF21............................................................................................................................................................48
9.3 OUK20 CABLE SET ...........................................................................................................................................................49
4
Model OUM980 with OUSAF21 Inline Sensor Safety Instructions
1. Safety Instructions
1.1 Designated Use
Integra Model OUM980 is an analyzer for determining the absorbance of a liquid medium at a specific wavelength
The analyzer is particularly suited for use in the following areas:
Bio-Technology industry Waste water processing Chemical processing industry Food industry Pharmaceutical industry
Any other use than the one described here compromises the safety of persons and the entire measuring system and is, therefore, not permitted. The manufacturer is not liable for damage caused by improper or non-designated use.
1.2 Installation, Start-Up and Operation
Please note the following items:
Installation, electrical connection, start-up, operation and maintenance of the measuring system must only be carried out
by trained technical personnel. The technical personnel must be authorized for the specified activities by the system operator.
Technical personnel must have read and understood these Operating Instructions and must adhere to them.
Before commissioning the entire measuring point, check all the connections for correctness. Ensure that electrical cables
and hose connections are not damaged.
Do not operate damaged products and secure them against unintentional commissioning. Mark the damaged product as
being defective.
Measuring point faults may only be rectified by authorized and specially trained personnel.
If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning.
Repairs not described in these Operating Instructions may only be carried out at the manufacturers or by the service
organization.
1.3 Operational Safety
Relevant regulations and standards have been met. As the user, you are responsible for complying with the following safety conditions:
Installation instructions Local prevailing standards and regulations.
EMC
This instrument has been tested for electromagnetic compatibility in industrial use according to applicable standards. Protection against interference as specified above is valid only for an instrument connected according to the instructions in these Operating Instructions.
5
Model OUM980 with OUSAF21 Inline Sensor Safety Instructions
1.4 Return
If the device requires repair, please contact your local sales and service representative. Refer to the website on the back page of this manual on where to find your local sales and service support office. Please use the original packaging, if possible.
Notes on safety icons and symbols
Safety icons
Warning!
This symbol alerts you to hazards. They can cause serious damage to the instrument or to persons if ignored.
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could cause damage to the instrument if ignored.
Note!
This symbol indicates important items of information.
Electrical symbols
Direct Current (DC)
A terminal at which DC is applied or through which DC flows.
Alternating Current (AC)
A terminal at which (sine-form) AC is applied or through which AC flows.
Ground connecting
A terminal is already grounded using a grounding system.
Protective earth terminal
A terminal which must be grounded before other connections may be set up.
Alarm relay
Input
Output
6
Model OUM980 with OUSAF21 Inline Sensor General Information
L
2. General Information
The Model OUM980 Photometric Analyzer is used in conjunction with a Model OUSAF21 sensor to measure the spectral absorbance of process liquid in the VIS/NIR region of the electromagnetic spectrum.
Dependent upon the optical pathlength of the connected sensor, the instrument can measure up to 50 Optical Density (OD) units.
The Model OUM980 Photometric Analyzer is manufactured from state of the art digital electronics. Plant interface is through quick disconnect screw terminals on the back of the module. The user interface is comprised of a 6-button tactile feel keypad and 4 x 20 character alphanumeric LCD display.
2.1 How the Model 980 Photometric Analyzer Works
The Model OUSAF21 sensor generates two photocurrent (nA) detector signals based upon the amount of energy present at its lamp source (reference) and measurement point internally. The Model OUM980 unit computes the logarithmic ratio of these two signals and determines the absorbance of the liquid passing through the sensor. The absorbance value is expressed in optical density units (OD) and is displayed on the front panel. Two galvanic isolated analog current outputs, proportional to the OD reading, are simultaneously transmitted for connection to other instrumentation and recording devices.
2.2 Concentration and Absorbance Units
The concentration of an optically absorbing material in a mixture can be determined since it is related to the amount of light absorbed from a beam of light passing through it. The absorbance of a substance is directly proportional to the concentration of the material that causes the absorption. The Lambert-Beer Law describes this relationship of absorbance (A) to concentration. Essentially, the amount of radiation transmitted through the absorbing material decreases logarithmically with its increasing concentration.
Where
A log
1
log
T
The above assumes that the optical pathlength remains constant. The optical density (OD) however, is defined as Absorption per unit length. Normalizing to an optical pathlength of 1cm, it follows that:
Io Ir
and
T
Ir Io
OD
Where OD = Optical Density
A = Absorbance
L = pathlength in cm
2.3 Easy Interfacing
The Model OUM980 Photometric Analyzer incorporates a front panel display and two analog current output signals. The display on the front panel is an alphanumeric 4 x 20 character LCD. The display may be configured for process variable or analog current output measurements. Optical Density (OD) readings are normalized to a 1cm pathlength.
)(1A
7
Model OUM980 with OUSAF21 Inline Sensor General Information
There are two analog current outputs from the unit. The range of each can be independently set to be any range within the measurement range of the instrument. One output is configured to operate with measurement baseline shift commands, while the other tracks full scale of the instrument.
The analog current outputs (4-20mA) will operate with loads up to 750 ohms. These outputs are galvanically isolated and conform to NAMURi Standard.
Model OUM980 is supplied as standard with four digital inputs and three digital output relays.
The digital inputs can be configured to perform a variety of functions such as baseline and sensor lamp off/on.
The outputs can be connected to a PLC for inclusion in a control scheme, for simple valve operation or for local indication (alarms) and annunciation. Alarm status indication is provided on the front panel display.
All digital inputs are optically isolated.
i
NAMUR standard limit detection when enabled is set to a BAD if the signal level is above 21 mA or below 3.6 mA for more than 4 seconds. The BAD status is cleared
when the signal returns to the normal 4-20 mA range.
8
Model OUM980 with OUSAF21 Inline Sensor Installation
3. Installation
3.1 Model OUM980 Analyzer Installation
Before starting installation, inspect the analyzer, sensor, and supplied cable set for any signs of shipping damage. Report any visual damage or discrepancies to the factory and the shipper immediately.
The Model OUM980 analyzer is a ¼ DIN enclosure which can be installed a variety of panel, wall and bench top housings. Refer to figure 3.1 for mounting dimensions. Mount or install the analyzer into an enclosure or area that is not subject to excessive vibration or shock and will protect the instrument from materials such as water and chemicals. Allow enough clearance behind it for cable access.
Figure 3.1: OUM980 Analyzer Mounting
9
Model OUM980 with OUSAF21 Inline Sensor Installation
3.2 Model OUM980 Cables and Wiring
All wiring terminals are located on the back panel of the Model OUM980. The analyzer/sensor interconnection cables supplied with the system have all been pre-terminated and labeled for ease of installation. The Model OUM980 has two (2) analog current outputs of 4 to 20mA. Both are capable of driving loads up to 750 ohms.
In addition to the analog outputs, three (3) galvanic isolated digital outputs (N.O.) are available. These outputs are rated for both AC and DC voltages up to 280 volts at 125mA. These outputs are intended for dry contact or pilot applications.
Insert Terminal
Here
1
2
3 4 5 6 7
Connector
8 9
Phoenix Cambion
10 11
12
Wire Pin Terminal
Figure 3.2: Wire Terminal Preparations
Cables installed for signal connection (i.e. analog outputs, lamp fail output) should be shielded twisted pairs.
Warning!
When routing the cables, separate signal cables from power wiring!
Caution!
Use dry contact only when connect digital input to analyzer!
Refer Figure 3.3 for wiring connection.
10
Model OUM980 with OUSAF21 Inline Sensor Installation
Figure 3.3: Model OUM980/OUSAF21 Dual Beam Wiring Diagram (AC Input Version)
3.3 DC Input Power Option
For instruments supplied for 24VDC operation, only the power input connection is changed. Figure 3.3 shows the connection detail for a 24VDC unit.
Input Power
+24VDC DC RTN
Ground
24 VDC
Figure 3.4: Integra DC Power Connection
11
Model OUM980 with OUSAF21 Inline Sensor Installation
3.4 Model OUSAF21 Sensor Installation
Sensors can be installed either directly in a process line or in a by-pass line. They can be mounted either horizontally or at an angle. If mounted horizontally, the sensor lamp and detector housings must be horizontal. This will insure that the optical window surfaces are in a vertical position, which will help prevent build up on the window surfaces. The sensor should be located upstream of pressure regulators. Operating sensors under pressure will help to avoid the possibility of air or gas bubble evolution, which can cause measurement noise and error.
When installing, allow adequate space for the connection of cables at the ends of the lamp and detector housings. Access to these areas is also important for connection/disconnection purposes. Sensor bodies should be supported when in line and care should be taken to ensure they are protected against damage caused by external forces such as carts on adjacent walkways.
Figure 3.5: Recommended Orientation for Sensor Installation
12
Model OUM980 with OUSAF21 Inline Sensor Installation
Figure 3.6: Typical OUSAF21 Sensor With Cables Connected.
3.5 Cable OUK20 Structure and Termination of Measurement Cable
Figure 3.7: OUSAF21 Cable at Lamp Side
13
Model OUM980 with OUSAF21 Inline Sensor Installation
Figure 3.8: OUSAF21 Cable at Detector Side
3.6 Post-Connection Check
After wiring up the electrical connection, carry out the following checks:
Device Status and Specifications
Is the analyzer or the cable externally damaged?
Electrical connection Remarks
Are the installed cables strain-relieved?
No loops or cross-overs in the cable run?
Are the signal cables correctly connected according to the wiring diagram?
Are all screw terminals tightened?
Are all cable entries installed, tightened and sealed?
Are the PE distributor rails grounded (if present)? Grounding at place of installation
Caution!
Improperly grounding will lead to unreliable measurement result!
Remark
Visual inspection
Remark
Grounding at place of installation
14
Model OUM980 with OUSAF21 Inline Sensor Operation
4. Operation
4.1 Quick Operation Guide
User has the following ways of operating the analyzer:
On site via the keypad Via the digital input interface to perform process control functions remotely.
4.2 Display and Operating Elements
4.2.1 Operating Elements
Figure 4.1: Operating Controls
1. LED for lamp warning functions and limit relay status.
2. Menu/ESC key serves two functions. Pressing Menu/ESC will always exit out of a screen and will not save any changes
Entered. Pressing Menu/ESC from the main PVii or Output screen will open the Sub-Menu screen.
3. Ack/ENT key accepts changes in a screen and continue forward through the screen
4. 4 keys navigate the user through the menus and screens.
5. LC display for displaying the measured values and configuration data LED.
ii
PV is the Process Value for PV1 and PV2 and corresponds to Output 1 and Output 2 respectively.
15
Model OUM980 with OUSAF21 Inline Sensor Operation
4.2.2 Operation Legend
In this manual, the following operation legend will be used when describing operation procedure.
Enter/Acknowledge key
Escape/Menu key
Left or Right key
Up or Down key
Figure 4.2: Operation Legend for Operation Instruction
4.3 Local Operation
4.3.1 Measuring Mode
The analyzer normally operates in measuring mode. Here, the relays and alarm LED are triggered by the analyzer according to process status. The analyzer will take configurable digital input to trigger certain function like BaseLine Zero
During the measuring mode, the LCD screen will display as figure 4.3. The process value, unit and analyzer label will be displayed permanently. BaseLine, Alarm and Lamp Status will be display when conditions are met.
iii
, Optical Zeroiv etc...
Figure 4.3: LCD display on measuring mode
iii
BaseLine Zero when enabled offsets the PV to zero. This often used to subtract background absorbance prior to initiating a process.
iv
Optical Zero is established during calibration and the set with a non-absorbing liquid in the flow cell, i.e. de-ionized water.
16
Model OUM980 with OUSAF21 Inline Sensor Operation
Table 4.1 shows the status display description and display condition.
Status Area Name Displayed symbol
Description and Display criteria
Analyzer Label “xxxxxxxxxxxx” Editable up to 12 characters for user
analyzer identification.
“OD” Display “OD” under normal condition, Process Unit
“xxxxxx” Displays user configured engineering unit
if PV correlation function is activated.
Over Range Symbol “OVER” Displays if measurable range is reached.
BaseLine Status
“BL OFF”
Displays if BaseLine is not activated.
“BL ON” Displays if BaseLine function is activated.
“BL SFT” Displays if BaseLine and BaseLine Shift
function are both activated.
Alarm “AL x” Displays if corresponding relays (1, 2, or
both) are triggered.
Lamp Status
“LAMP LOW” Flash if reference voltage is lower than
80% of initial value.
“LAMP FAIL” Displays if lamp reference signal is lost.
“LAMP OFF” Displays if lamp is turned off.
Cable Status “CABLE FAULT” Displays if lamp or signal cable has a faulty
connection.
Table 4.1: Operation screen under measuring mode
Under measuring mode, press control routine. The detail of process control function will be listed in Chapter 6.
Figure 4.4: Operations under Measuring Mode
button will let you enter the process control screen, where user can operate process
17
Model OUM980 with OUSAF21 Inline Sensor Operation
4.3.2 Configuration Mode
The configuration and calibration functions are arranged as function groups.
Use
Use
key to enter the function group menu from measurement screen.
arrow key to select the group from process set-up, maintenance, or system data.
Use to confirm the change and input the value.
If a modified setting is confirmed with , the old setting is retained.
Menu
ESC
Ack
ENT
Menu
ESC
Ack
ENT
Menu
ESC
Ack
ENT
Figure 4.5: Diagram of configuration structure
Input data
During the system configuration and set up procedure, user will need to input the data into the screen. In the data input screen
Use
to move the cursor to select the digit to edit.
To change the decimal point of the input data, place the cursor on decimal point, use
point to the left or right.
Figure 4.6: Decimal Point Position Placement
to move the decimal
18
Model OUM980 with OUSAF21 Inline Sensor Operation
4.4 Remote Operation
Remote operation function will provide user the ability to use off site dry contact to perform the process control. Short input channel pin to COMMON pin will trigger the input the channel.
Refer section 5.3.8 to configure and use digital input function for remote control.
Digital Input #
1 Lamp On/Off
2 Enable BaseLine function
3 Update BaseLine value
1-
12
1+ 2-
2+
COM
1 2 3 4
Figure 4.7: Typical Digital Input Wiring Diagram
Description and Display criteria
Optical Zero 4
Hold current process measurement (both display and analog output)
Table 4.2: Digital Input Functions
13
C hannel 1 & 2
11 14
15
16 17
18 19
Analog Outputs (4-2 0 Ma)
D ry Contact or Panel Switch
19
Model OUM980 with OUSAF21 Inline Sensor Commissioning
5. Commissioning
5.1 Function Check
Warning!
Check all connections for correctness. Make sure that the supply voltage is identical to the voltage written on the technical information!
5.2 Switching On
Familiarize yourself with the operation of the analyzer before it is first switched on. Please refer in particular to the "Safety Instructions" and "Operation" sections. After power-up, the device performs a diagnosis for 30 seconds and then goes to the measuring mode.
Lamp Initial Set-Up
If the analyzer is first time switched on with a new lamp, a “LAMP LOW” symbol will be shown on the display screen. And the frond LED will be flashing and lamp alarm relay will be toggled. To remove this warning message, the lamp status has to be initialized from the software. The lamp low warning shows as follow
Figure 5.1: Lamp status during start-up
20
Model OUM980 with OUSAF21 Inline Sensor Commissioning
Reset New UV Lamp
Press , enter function group selection menu. Move the cursor
to MAINTENANCE and press to enter.
Press
.
The current lamp voltage will be displayed on the right upper
corner. The timer will start counting down once the is pressed to warm up the lamp.
Wait 30 minutes until the timer goes to 0:00
The current lamp voltage will be displayed in the first line and lamp low threshold voltage will be calculated and display on the third line.
to navigate to LAMP REPLACE, and press
Press to accept the setting and use to go back to measurement screen. The LED and lamp alarm relay will be reset.
5.3 Analyzer Configuration and Operation
After power-up and lamp setting, you must make some settings to configure the most important functions of the analyzer which are required for correct measurement.
The analyzer/sensor will be Plug& Play if been ordered together as complete solution!
If OUM980 Photometric Analyzer is ordered with OUSAF21 sensor and OUK20 cable, the factory will perform:
New lamp set up Pathlength adjustment in analyzer according to sensor order Calibrate analyzer
If application data has been given, factory will perform optical density to engineering unit correlation and deliver correlation data with analyzer. There will be an additional charge if sample preparation is required. Span and relay setting also can be performed in factory if requested. If analyzer ordered separately, the factory default setting will be loaded into the analyzer when analyzer is delivered.
21
Model OUM980 with OUSAF21 Inline Sensor Commissioning
5.3.1 Factory Default Setting
Parameters Factory Default Value
Measurement Unit
Measurement Pathlength
Output Span 1 5.0000 OD
Output Span 2 5.0000 OD
Relay 1 Set Point/Status 1.0000 OD/ Disabled
Relay 2 Set Point/Status 1.0000 OD/ Disabled
Process Value Correlation Off (Curve #0)
BaseLine Disabled
Password Disabled
Table 5.1: Factory Default Setting
OD
100mm to 250mm
5.3.2 Sensor Calibration
Model OUM980 analyzer supports 2 different calibration procedures: Liquid calibration and Filter Calibration.
5.3.2.1 Liquid Calibration Procedure
A liquid solution of a known optical density (at the wavelength of the sensor) must be used for instrument calibration. The following procedure is used in factory. Calibration liquid can be chosen by user depending on application.
22
Model OUM980 with OUSAF21 Inline Sensor Commissioning
The Liquid Calibration screen appears only if a Filter Calibration system is not installed and configured in the System Data section.
Liquid Calibration Procedure
Enter MAINTENANCE sub-menu to access
SENSOR CAL function. Press to ent.
Use to select LIQUID CAL, Press to enter liquid calibration process.
Fill the sensor with a zero fluid.
Press to record zero point.
Enter the value of the standard solution being used to calibrate the system. The solution should be approximately ½ of the full range of the system.
Press to continue.
Fill the sensor with the standard calibrating solution.
Wait until OD reading increase and stabilize. Press to record the data.
Remove the calibration solution, and rinse repeatedly with a zero solution. OD reading should decrease and
stabilize. Press
to proceed.
Remove the cover screw on the measurement detector
and install the reference rod. Press to proceed.
Remove the reference rod and install the cover screw on
the measurement detector. Press to proceed. The unit records the values of the calibration solution and ref rod value and displays these values. The standard solution is used for calibration, while the ref rod is for
checking calibration. Press
to accept these values.
23
Model OUM980 with OUSAF21 Inline Sensor Commissioning
5.3.2.2 Filter calibration Set up and Calibration
The EasycalTM unit contains an optical that was selected based on laboratory measurement of the customers submitted sample. This filter is placed into the optical measurement path of the instrument and displays an absorbance ratio process related. This filter has been scanned with a traceable source and actual absorbance at individual (measurement) wavelengths is ascertained.
The Easycal noted on the Calibration Certificate supplied with the unit. The absorbance value should be entered into the Model OUM980 as part of the configuration setup.
TM
allows the calibration without using liquid standards. Refer to the actual value of the EasycalTM optical filter as
Figure 5.2: EasycalTM Control for the OUSAF22 Dual Beam Sensor.
24
Model OUM980 with OUSAF21 Inline Sensor Commissioning
EasycalTM Installation
Enter SYSTEM DATA sub-menu to access CHANGE EASYCAL CNFG function.
Press to continue.
Use
to select [YES] or [NO] to decide if the
Easycalis installed. To install, select [YES] and confirm.
This screen displays the present NIST Filter Absorbance
values and the re-certification date. Press to change the date and filter values starting with the next screen.
Move cursor to edit the day, month and year. Save data and move to next step.
Refer NIST filter high absorption value listed on NIST filter certificate delivered with Easycal. The value should be
close to 1.0000 AU. Press to continue and use the same procedure to input the low filter absorption value, close to 0.5000 AU.
After put in both filter data use to confirm the setting. Easycal™ data installation is finished.
25
Model OUM980 with OUSAF21 Inline Sensor Commissioning
EasycalTM Calibration Procedure
Enter MAINTENANCE sub-menu to access
SENSOR CAL function. Press If Easycal has been installed. Analyzer will automatically enter Easycal
TM
procedure.
to change.
Both high and low NIST filters absorption value will be displayed in this screen. User set certificate due date is
displayed also. Press after review the data.
Assure both high and low filters are in OUT position.
Press
to record zero point.
Move HI filter adjustment screw to IN position.
Press to record high filter value.
Move LO filter adjustment screw to IN position.
Press to record HI+LO value.
Move HI filter adjustment screw to OUT position.
Press calibration calculation.
to record low filter value and perform
The screen then displays the low, high and high+low readings and % deviation from the certified value of each
value. Review the deviation level and press to accept the calibration result.
5.3.3 Change Pathlength Setting
If sensor pathlength need to be changed by user, the corresponding change will need to be performed to OUM980 analyzer also.
Refer to Chapter 6 Sensor and Analyzer Maintenance to perform sensor pathlength adjustment. This section only covers corresponding analyzer setting to match the measurement result.
26
Model OUM980 with OUSAF21 Inline Sensor Commissioning
Pathlength Setting
Enter SYSTEM DATA sub-menu to access
PATHLENGTH function. Press
The FROM value is the present pathlength Value. Use
position the cursor and each digit position. When new pathlength has been selected, press
to confirm the change.
The SYS DATA: PATHLENGTH is displayed together with the new values for pathlength and OD MIN/MAX.
to continue.
to select wanted digit. Repeat for
5.3.4 Setup Output Span
The 4~20mA span setting will provide user capability to achieve maximum resolution for preferred measurement range.
Change 4~20mA Output Span
Enter PROCESS SET-UP sub-menu to access OUTPUT SPANS
to
function. Use to move the cursor to O1 and press to select.
Use same procedure above to change Channel 2 output span.
This screen will display the current output span MIN/MAX span based
on pathlength or process value correlation setting. Press to change the span.
In this screen the current value for span is the FROM value. Position the cursor as done previously and select the number for that position. Repeat as needed for each digit position. When the new span has been
entered press
and confirm the sanity check to accept.
27
Model OUM980 with OUSAF21 Inline Sensor Commissioning
5.3.5 Output Relay Set Point and Configuration
Change Output Relay Set Point
Change Relay Set Point
Enter PROCESS SET-UP sub-menu to access OUTPUT SPANS function.
Use
Configure Output Relays
The Alarm Setup screen allows configuration of PV, Hi/Lo/Off, and NO/NC operation. is used to select the
parameter and is used to change the state. Press to continue.
Functions Description
Trigger Condition
Hysteresis Set relay hysteresis Delay Delay trigger timer (Sec)
HIGH Triggered if process value higher than set point. LOW Triggered if process value higher than set point. OFF Relay disabled PV1 Process value #1 Input Mapping PV2 Process value #2 NO Normally open Relay Operation Mode NC Normally close
In this screen the current value for span is the FROM value. Position the cursor as done previously and select the number for that position. Repeat as needed for each digit position. When the new relay set point
has been entered press and confirm the sanity check to accept.
Table 5.2: Output Relay Configuration
to move the cursor to ALx and press to select.
28
Model OUM980 with OUSAF21 Inline Sensor Commissioning
Configure Output Relay
Enter from MAITENANCE sub-menu. Scroll to the ALARM CONFIG
screen. Use and confirm.
to select the ALx line. Press to select
The ‘ALARM 1 SETUP’ screen will appear. Select the SOURCE option, PV1 or PV2.
Next with
select the CONFIG option. Use
to select OFF, HIGH and LOW.
Press
(N/C) alarm state of the relay contacts. Select the N/O option and
to scroll down to ALARM 1 SETUP STATE screen.
selects either normally open (N/O) or normally closed
press to accept and confirm.
The CHANGE ALARM 1 DELAY is next displayed. Set the needed
delay. Delay time unit is second. Press to accept and confirm.
The CHANGE ALARM 1 HYSTS option controls the alarm hysteresis. If the alarm set point is set 1.00 OD and the hysteresis is 0.100 OD, an alarm condition exists at 1.00 OD and will stay in alarm until the PV is
0.90 OD or less. Hysteresis can be called the deadband between an alarm on and off condition. Set the hysteresis to 0.100 OD and
press
to accept and confirm.
29
Model OUM980 with OUSAF21 Inline Sensor Commissioning
5.3.6 Current Output Calibration
This function calibrates 4~20mA current output to assure the measurement accuracy.
Calibrate 4~20mA Output
Enter MAINTENANCE Sub-Menu to access
5.3.7 Date and Time Setting
Set Clock
OUTPUT CAL function. Move cursor to OUTPUT 1 and press select this channel.
Move the cursor up and down to select the value need to be trimmed.
Use
The trimming resolution is 1% per step. Press
to trim 4mA or 20mA.
to accept and
confirm the trimming.
Follow same procedure to trim OUTPUT 2.
Enter MAINTENANCE sub-menu to access SET CLOCK function. Use
to select the CUR DATE line.
to
Press
Screen is displayed with the current date. Use
the cursor and
to select the DATE option and then the CHANGE DATE
to change the mm/dd/yy numbers. When
the wanted date is entered, press to accept and confirm.
Press to select the TIME option and then the CHANGE TIME
Screen is displayed with the current time. Use
the cursor and
the wanted time is entered, press
to change the hh/mm/ss numbers. When
to accept and confirm.
to position
to position
30
Model OUM980 with OUSAF21 Inline Sensor Commissioning
5.3.8 Configuration and Use of Digital Inputs
Configure Digital Input
Enter MAINTENANCE sub-menu to access DIGITAL IN function. Press
to continue.
The following screen displays the current configuration of the external
digital inputs. Use the STANDARD CONFIG When disabled the external digital inputs are inoperative.
to scroll through the DISABLE and
When in STANDARD CONFIG, the digital inputs are configured as;
1 Lamp On-Off 2 Baseline On-Off 3 Baseline Update 4 Hold/Optical Zero
Digital Input #4 can be set as Optical Zero function if wanted. Press
to accept and confirm.
5.3.9 Process Control
From main measurement screen, press
Functions Status Description Remote Trigger
BaseLine
ON Enable baseline Digital Input 2 UPDATE Update baseline value Digital Input 3 OFF Disable baseline Digital Input 2 ON Turn on lamp Digital Input 1 Lamp Switch
OFF Turn off lam Digital Input 1 Optical Zero - Make current measurement at 0. Digital Input 4 Optical Check - Start sensor check -
5.3.9.1 Access of Process Control
The authority of access of process control function can be configured in OUM980 analyzer. The access control is password protected. See 5.3.9.6 for password protection.
to enter process control screen.
Table 5.3: Process Control Functions
31
Model OUM980 with OUSAF21 Inline Sensor Commissioning
Configure Process Control Access
Enter SYSTEM DATA sub-menu to access ACCESS function. Press
to continue.
Select the option to configure. Note: The Response is in brackets [ ]. With bracketed responses use
options to configure. When all changes are made, press and confirm the entries and return to top screen.
for changes. Select ‘YES’ for each option. There are 4
to accept
5.3.9.2 Baseline (AutoZero)
To operate the baseline function, the Process Control Screen must be assigned as an authority for the user. If assigned the authority,
use baseline use “On”, to remove an existing baseline use “Off”. To change an existing baseline setting to a new setting use “Update”. If the baseline was previously “On” when a power interruption occurred, the analyzer will retain its last saved baseline setting and return to that setting when power is restored, and the analyzer and the baseline control will remain “On”. Baseline function replaced the AutoZero function in the 800 series analyzer.
If a Baseline Shift value has been entered, activating the baseline shift will cause the 4~20mA output channel 2 to “Shift” to the entered % value. Both PV display would indicate “0” The PV screen will also indicate BL SFT in lieu of BL ON when a shift value has been entered (>0%).
Configure Process Control Access
to open the Process Control Screen. Use to select baseline and select as follows: To initiate a new
Enter SYSTEM DATA sub-menu to access ACCESS function. Press
5.3.9.3 Optical Zero
Optical Zero is a function to provide an over all zero to offset the initial background optical noise in OUSAF21 sensor. User can also use optical zero to setup the process zero (e.g. DI water optical density etc...).
to continue.
Select the option to configure. Note: The Response is in brackets [ ]. With bracketed responses use
options to configure. When all changes are made, press to accept and confirm the entries and return to top screen.
for changes. Select ‘YES’ for each option. There are 4
32
Model OUM980 with OUSAF21 Inline Sensor Commissioning
5.3.9.4 Lamp Switch
Use this function to switch lamp on and off.
Note!
Lamp alarm relay status will be affected.
5.3.9.5 Optical Check
When the Easycal secondary standard for routine system checking. Calibration can be traced to the controlled standard used to verify the original calibration solutions or process samples and can therefore satisfy validation procedures. During initial calibration of a sensor/analyzer pair, liquid standards should be used to calibrate the full-scale response of the system. After calibration, by filling the sensor with water and ensuring the analyzer reads zero, a value for the Easycal the ‘ON’ position (see Fig 5.2) and note the analyzer display reading. At any time in the future, setting the Easycal the instrument will display the the same value when the sample cell is filled with clear water.
Note!
All analyzers must be re-calibrated with liquid standards if there are any changes made to the optical chain, including replacement of seals or lamp.
TM
is used in conjunction with an initial liquid standard calibration procedure, the EasycalTM can be deemed a
TM
can be obtained by setting the EasycalTM in
TM
‘ON’ position
5.3.9.6 Password Protection
OUM980 support the password function to protect the analyzer from unauthorized operation.
Figure 5.3: Password protection level
33
Model OUM980 with OUSAF21 Inline Sensor Commissioning
Password Protection
Enter SYSTEM DATA sub-menu to access PASSWORD function. Press
to continue.
When in the ENTER OLD PASSWORD screen, the CURSOR highlights the
left digit position. Use the
to scroll thru 0-9 for each digit position. Move the cursor to
to wanted digit position. Use
the desired digit and enter a number. When the old password number has
been entered, press
to accept.
Note: When “0000” is set a password, the password protection mode is disabled.
When in the ER NEW PASSWORD screen, the CURSOR highlights
the left digit position. Use the
to wanted digit position. The
scrolls thru 0-9 for each digit position. Move the cursor to the desired digit and enter a number. Repeat as wanted for each cursor position. Any password number up to 4 digits can be entered. When the
password number has been selected, press twice to accept. The Password for MENU access is now the number entered.
5.3.9.7 Process Value Correlation
In OUM980 analyzer, user can define and store 3 linearization tables to correlate the optical density measurement to specified process parameter measurement. Each table supports up to 16 points of measurement value. User can switch correlation curves to fit different application.
Caution!
The overall course of the curve must be continuously rising. Horizontal or vertical segments or a zigzagging of the curve are not permitted.
Data points may be entered into the linearization table randomly; the software performs sorting automatically as needed. It is possible to increase accuracy in a certain segment of a linearization table later. To do so it is not necessary to enter all data points again, but to enter only additional data points.
34
Model OUM980 with OUSAF21 Inline Sensor Commissioning
Process Value Correlation Set Up
Enter PROCESS SET-UP sub-menu to access PV CORRELATE
function. Press
to continue.
The next screen asks which curve (1-3) you want to edit and store the
correlation table. Scroll with
Press to continue.
to select curve number.
If the original sample value is not put in yet. Select SAMPLE IN CELL?
By scrolling with to select. Press to continue.
The next screen asks if process sample is in the cell and displays measurement value. If this value is within the Min/Max acceptable
press to read and store.
The next screen asks if another sample is to be measured. Press
to read another sample, press if finished all samples.
Confirm the number of point to build correlation curve and press
to store original sample data.
SAMPLE EDIT is used to put in the target correlation value and unit.
Press
to continue to put in data.
The next screen asks if the user would like to change the original sample point OD value. If this value needs to be changed, use
continue.
and to change value. Press to
35
Model OUM980 with OUSAF21 Inline Sensor Commissioning
The next screen is for entry of target correlation values that can be assigned to the original correlated measured value. i.e. 1.32 measured units to 9999.9 units. The FROM value is the last entered value on the ‘ENTER USER VALUE’ screen. Enter the new value using
and to position the cursor and select the
number for each required position. If a different position is required for
the decimal point, do this first. Press
twice to change and
accept.
Repeat above procedure until all correlation points have been entered.
When all the sample points have been edited, the Change Units screen is next. The engineering unit can be entered is limited to 6 alpha-
numeric characters.( i.e. gr/l, %, mg/l, cel/cc, etc.). Press twice to change and accept entry.
The Max/Min values are calculated for entered values and units. These values are the new limits for output span and relay setting and the displayed Process Display PV value and units.
Note!
If the OD to correlation unit table is predefined and does not need liquid sampling, use MANUAL EDIT option to entry the table to analyzer. The procedure is similar to SAMPLE EDIT.
Switch Correlation Curves
Enter PROCESS SET-UP sub-menu to access RUN CAL CURVE
function. Press
to continue.
The next screen asks which curve (0-3) you want to execute. Curve #0 is the default OD measurement. Curves 1-3 are user defined. Scroll
with confirm.
to select curve number. Press to select and
NOTE: Entering a curve with Process Correlate will automatically set it as the current active curve.
36
Model OUM980 with OUSAF21 Inline Sensor Maintenance
6. Maintenance
6.1 Maintenance Outline
Once the unit is in operation, there is no requirement to access the interior of the Model OUM980 Photometric Analyzer housing for normal day-to-day operation and calibration.
Certain maintenance needs to be performed for OUSAF21 sensor. The service and maintenance interval time is application dependant.
Figure 6.1: Cross-Sectional View of a Typical OUSAF21 Sensor
Note!
Check list of maintenance!
Lamp Replacement O-Ring Replacement Filter Replacement Sensor window and gasket replacement
37
Model OUM980 with OUSAF21 Inline Sensor Maintenance
Caution! The procedures described in this section should only be carried out by qualified maintenance staff.
Model OUSAF21 sensors contain sensitive optical components and should be handled carefully. Particular care must be taken to prevent contamination of these components.
Warning!
Please refer to EXP-1 lamp instructions to replace lamp for hazardous area versions.
Caution!
Clean all optical components with a suitable lint free lens cleaning tissue and ethanol.
Both the High Luminescence and Gas-Filled Lamps are replaced identically.
38
Model OUM980 with OUSAF21 Inline Sensor Maintenance
Figure 6.2: Illustrative Standard Lamp Replacement Procedure
39
Model OUM980 with OUSAF21 Inline Sensor Maintenance
6.2 OUSAF21 Measurement Detector/Filter Replacement
Figure 6.6 – OUSAF21 Detector Assembly
1. The detector assembly is attached to the sensor with a threaded window ring. The detector assembly is removed
by unscrewing it off the window ring. Removal of the detector assembly does not break the liquid integrity of the sample cell.
2. After removal of the detector assembly, unscrew the detector endplate from the detector housing.
3. To replace either filter, remove the retaining screw on the detector, and carefully remove the detector. Remove the
o-rings in the cavity and gently tap the filter out of the detector block (note: the “mirror” side is directed towards the light source).
3. Replace filter/detector in reverse order.
5. After detector/filter replacement, the detector assembly may require alignment and the system to be re-calibrated.
6.3 Sensor Wavelength Change
Should it be required to change the operating wavelength of the sensor, both the measurement and reference filters must be changed. When changing filter wavelength, follow the same procedures as describe above.
40
Model OUM980 with OUSAF21 Inline Sensor Maintenance
6.4 Sensor Window and Gasket Replacement
Sensor optical pathlength is established by the window type(s) used. Each analyzer-sensor pair is configured for a given pathlength. Windows must be replaced with the same type to maintain pathlength. The illustration below is a typical flowcell assembly. Figure the available window types and possible pathlengths for sensors up to 2" line sizes. Replacement of windows or window seals, changing window spacing or other maintenance requiring the disassembly of the sensor uses the following procedure:
Note:
To replace and/or change the windows and seals, the sensor must be removed from the process line.
Figure 6.3: Typical Flowcell/Window Replacement
1. Remove the lamp and the detector housings from the sensor body.
2. Remove the 4 socket head screws from each window retaining ring and remove the rings. Be careful to loosen the screws
evenly and alternately around the window retaining ring. If the window is ‘stuck’, apply Acetone to the window seal area and let soak for several minutes. This may assist in freeing the windows from the seals.
3. Gently push/ease the windows out of the sensor.
4. Inspect the window area and clean as necessary. Inspect the windows for any signs of abrasive wear or chipping. If any is
apparent, replace the windows. Discard the 'O' rings and replace with new ones of the same material type. Re-assemble the sensor in the reverse order, taking care to cross-tighten the window retaining ring screws evenly to prevent uneven seating. If the sensor pathlength has been changed, the analyzer module must be configured to reflect the new pathlength. After every re­assembly of an OUSAF21 sensor, it is necessary to carry out a liquid or Easycal™ calibration with its associated analyzer.
Note!
Upon re-assembly, insure that the lamp assembly is mounted on to the side of the flow cell with the “shorter” length of the two windows.
5. Remove the lamp and the detector housings from the sensor body.
6. Remove the 4 socket head screws from each window retaining ring and remove the rings. Be careful to loosen the screws
evenly and alternately around the window retaining ring. If the window is ‘stuck’, apply Acetone to the window seal area and let soak for several minutes. This may assist in freeing the windows from the seals.
7. Gently push/ease the windows out of the sensor.
8. Inspect the window area and clean as necessary. Inspect the windows for any signs of abrasive wear or chipping. If any is
apparent, replace the windows. Discard the 'O' rings and replace with new ones of the same material type. Re-assemble the sensor in the reverse order, taking care to cross-tighten the window retaining ring screws evenly to prevent uneven seating. If the sensor pathlength has been changed, the analyzer module must be configured to reflect the new pathlength. After every re­assembly of an OUSAF21 sensor, it is necessary to carry out a liquid or Easycal
TM
calibration with its associated analyzer.
Note!
Some window faces may not be parallel to each other. This is normal, especially with fire-polished quartz windows. Take care to ensure pathlength gauge does not scratch window faces.
41
Model OUM980 with OUSAF21 Inline Sensor Replacement Parts
7. Replacement Parts List
7.1 Model 980 Photometric Analyzer
DESCRIPTION Reference No. Part Number
Fuse, AC Input 3.15A GDA Ceramic 1678-0017-00 63006917 Fuse, DC Input 1A MST 1678-1000-00 63006930 Fuse, Relays/Alarm 125mA MSF 1678-0125-00 63006925
7.2 Model OUSAF21 Inline Sensor
DESCRIPTION Reference No. Part Number
High Luminescence Lamp A011-0680-22 63009261 Gas-Filled Lamp A011-2200-00 63009277 Standard Lamp A011-0610-22 63009226
Interference Filter - 400nm 1410-0400-00 63006556 Interference Filter - 420nm 1410-0420-00 63006559 Interference Filter - 430nm 1410-0430-00 63006560 Interference Filter - 720nm 1410-0720-00 63006583
Type 'C' Pyrex Window 1420-0240-00 63006658 Type 'C' Quartz Window 1420-0240-03 63006659 Type 'C' Sapphire Window 1420-0240-04 63006660
Window Gasket Kit, Viton A000-0662-01 63009117 Window Gasket Kit, Kalrez A000-0662-03 71041399 Window Gasket Kit, EPR(EDPM) A000-0662-05 63009119 Window Gasket Kit, Silicone A000-0610-00 63009100
42
Model OUM980 with OUSAF21 Inline Sensor Technical Information
8. Technical Information
8.1 Photometric Analyzer OUM980
8.1.1 Input
Measured Variable Range Digital Input Keyboard
8.1.2 Output
Current Output
Dual Channel Current from Model OUSAF21 Inline Sensor 5 decades of absorbance signal from each detector 4 Channels, dry contact 6 push buttons keyboard
2 channel, 4~20mA NAMUR Compatible
Resolution
Isolation
Signal on Lamp Alarm
Actual Current Output Range Load
Front Panel Display
Alarm LED
General Relay Output
Relay Operation
Relay Delay
Lamp Alarm Relay
Lamp Power Supply
8.1.3 Power Supply
Supply Voltage Power Consumption Mains Fuse
12bit
O1 to PE 500 VDC
O2 to PE 500 VDC
O1 to O2 500 VDC
3.6mA or 21.5mA in case of error
3.8mA to 20.5mA
750 ohm Max.
4x20 Character LCD display
On front Panel, Red LED
2 Channel, Solid State Relays
Above/Below trigger configurable; NO/NC configurable
0~60 sec configurable
1 Channel, NC
3 to 5 vdc, 1.5A Max.
85~265 VAC, 24VDC as option 12 VA Max.
3.15 A
43
Model OUM980 with OUSAF21 Inline Sensor Technical Information
8.1.4 Performance
Sensor Photo Current Input
100pA to 700uA
Linearity
Repeatability
Accuracy
Initial Drifting
8.1.5 Environment
Temperature (Monitor) Temperature (Sensor) Relative Humidity
Altitude
Degree of Protection
8.1.6 Mechanical Construction
Dimensions
0.3% Max.
0.3% Max.
0.3% Max.
<0.5% for first 24 hours
0~55°C
0~55 °C
95%
2000 m
IP 65 (front panel), IP 30 (housing)
L x W x D: 102 x 102 x 177 mm
Weight
Material
Degree of Protection
1.213 kg
Aluminum (polyester coating on front panel)
IP 65 (front panel), IP 30 (housing)
44
Model OUM980 with OUSAF21 Inline Sensor Technical Information
8.2 Absorbance Sensor OUSAF21
8.2.1 Sensor Dimensions
FLOWCELL PATHLENGTH “A” DIMENSION
10cm Pathlength 5.9 in/150mm 15 cm Pathlength 7.9in/200mm 20 cm Pathlength 9.8in/250mm 25 cm Pathlength 11.8in/300mm 30 cm Pathlength 13.7in/350mm 35 cm Pathlength 15.7in/400mm
LAMP ASSY TYPE “B” DIMENSION
Standard Lamp 2.41in/61.2mm
EXP-1 Lamp 6.5in/165.0mm
DETECTOR ASSY “C” DIMENSION
OUSAF21 4.05in/102.8mm
8.2.2 Input and Output
Measured Input Variable
Process absorption and optical density
Output Signal
Photo detector current (100pA~ 700 μA)
8.2.3 Mechanic Construction
Weight
Housing Material
1.225 kg (flow cell not included)
Stainless steel 316L
45
Model OUM980 with OUSAF21 Inline Sensor Technical Information
8.2.4 Environment
Temperature (Process) Relative Humidity
0~90 °C Continuous up to130°C for 2 hours
95%
Altitude
Degree of Protection
Maximum Pressure in Flow Cell
8.2.5 Power Supply
Supply Voltage
Power Consumption
8.2.6 Optical Specifications
Wavelengths Available
Filters Detectors Lamp
2000m
IP 65
6.8 Bar/100 PSI
3-5 vdc
4 VA continuous 8 VA start up in-rush
400 nm / 720 nm 420 nm / 720 nm 430 nm / 720 nm 520 nm / 720 nm Interference filters, stray light .01% maximum.
Blue Enhanced Silicon, hermetically sealed. Lens End Halogen Lamp
46
Model OUM980 with OUSAF21 Inline Sensor Ordering Information
9. Ordering Information
9.1 Color Analyzer OUM980
Power Supply 1 100-250 VAC 424 VDC
Transmitter Unit A Optical Density Y Special version, TSP no.to be spec
Calibration 1 Single order/spare part 2 With sensor/cable order position 9 Special version, TSP no.to be spec
OUM980
Complete Order Code
47
Model OUM980 with OUSAF21 Inline Sensor Ordering Information
9.2 Sensor OUSAF21
48
Model OUM980 with OUSAF21 Inline Sensor Ordering Information
9.3 OUK20 Cable Set
Sensor 1OUSTF10 2 OUSAF21/OUSAF22 3OUSAF23
Transmitter A OUS900 Serie B OUS600 Serie C OUS700 Serie
Cable Length 10 10 ft / 3m 15 15 ft / 4,5 m 25 25 ft / 7,5 m 50 50 ft / 15 m
88 ..... ft; cable
89 ..... m; cable
Barriere A Standard (non-hazardous area) BFM, Busbar
CATEX, Busbar D FM, DIN Rail
E ATEX, DIN Rail
OUK10
Complete Order Code
49
www.endress.com/worldwide
Loading...