pH electrodes, analog or with digital Memosens
technology
For standard applications in process and
environmental technology, with dirt-repellent
PTFE diaphragm, built-in temperature sensor
(optional for analog sensor)
Application
• Long-term monitoring and limit value monitoring of processes with stable process
conditions
– Chemical industry: strong acids/bases, plastic, pulp and paper industry
– Power plants (e.g. flue gas cleaning), oil and gas
– Incinerator plants
• Water/wastewater treatment
– Boiler feedwater and cooling water
– Well water and drinking water
– All industrial and municipal treatment plants
With ATEX, IECEx, FM, CSA, TIIS and NEPSI approval for use in hazardous areas
Your benefits
• Low-maintenance and robust thanks to large PTFE ring junction
• Can be used at pressures up to 17 bar abs. (246 psi)
• Process glass also for very alkaline applications (BA and BT versions)
• Process glass for applications in media containing hydrofluoric acid (FA version)
• For media with low conductivity (AS version)
• Integrated NTC30K temperature sensor (Memosens) for effective temperature
compensation; Pt100 or Pt1000 for analog sensors
• Optional: Poison-resistant reference with ion trap
Other advantages of Memosens technology
• Maximum process safety
• Data security thanks to digital data transmission
• Very easy to use as sensor data saved in the sensor
• Recording of sensor load data in the sensor enables predictive maintenance with
the Memobase Plus CYZ71D
Page 2
Function and system design
3
4
1
2
Orbisint CPS11D and CPS11
Measuring principle
Measuring system
pH measurement
The pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The
membrane glass of the electrode supplies an electrochemical potential which is dependent upon the
pH value of the medium. This potential is generated by the selective penetration of H+ ions through
the outer layer of the membrane. An electrochemical boundary layer with an electric potential forms
at this point. An integrated Ag/AgCl reference system serves as the required reference electrode.
The transmitter converts the measured voltage into the corresponding pH value using the Nernst
equation.
A complete measuring system comprises at least the following components:
• pH electrode CPS11D or CPS11
• Transmitter, e.g. Liquiline CM42, CM44x, Mycom S CPM153, Liquisys M CPM2x3
• Memosens data cable CYK10 for Memosens sensors or CPK9 for analog sensors
• Immersion, flow or retractable assembly, e.g. Cleanfit CPA871
A0025757
1Example of a measuring system for pH measurement
1Retractable assembly Cleanfit CPA871
2pH electrode CPS11D
3Memosens data cable CYK10
4Liquiline M CM42 two-wire transmitter for hazardous area
2Endress+Hauser
Page 3
Orbisint CPS11D and CPS11
CPS11D communication and data processing
Communication with the transmitter
Always connect digital sensors with Memosens technology to a transmitter with Memosens
technology. Data transmission to a transmitter for analog sensors is not possible.
Digital sensors can store measuring system data in the sensor. These include the following:
• Manufacturer data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Slope at 25 °C (77 °F)
– Zero point at 25 °C (77 °F)
– Temperature offset
– Number of calibrations
– Serial number of the transmitter used to perform the last calibration
• Operating data
– Temperature application range
– pH application range
– Date of initial commissioning
– Maximum temperature value
– Hours of operation under extreme conditions
– Number of sterilizations
– Resistance of glass membrane
You can display the abovementioned data using the Liquiline CM44x, CM42 and Memobase Plus
CYZ71D.
Dependability
Reliability
IntegrityData security thanks to digital data transmission
Easy handling
Sensors with Memosens technology have an integrated electronics unit that stores calibration data
and other information (e.g. total operating hours and operating hours under extreme measuring
conditions). Once the sensor has been connected, the sensor data are transferred automatically to
the transmitter and used to calculate the current measured value. As the calibration data are stored
in the sensor, the sensor can be calibrated and adjusted independently of the measuring point. The
result:
• Easy calibration in the measuring lab under optimum external conditions increases the quality of
the calibration.
• Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the
availability of the measuring point .
• Maintenance intervals can be defined based on all stored sensor load and calibration data and
predictive maintenance is possible.
• The sensor history can be documented using external storage media and evaluation programs, e.g.
Memobase Plus CYZ71D. Thus, the current application of the sensors can be made to depend on
their previous history.
Memosens technology digitizes the measured values in the sensor and transmits the data to the
transmitter using a non-contact connection that is free from potential interference. The result:
• Automatic error message if sensor fails or connection between sensor and transmitter is
interrupted
• Immediate error detection increases measuring point availability
Endress+Hauser3
Page 4
Orbisint CPS11D and CPS11
Safety
Measured variables
Measuring range
Maximum process safety
With inductive transmission of the measured value using a non-contact connection, Memosens
guarantees maximum process safety and offers the following benefits:
• All problems caused by moisture are eliminated:
– Plug-in connection free from corrosion
– Measured values cannot be distorted by moisture.
– Can even be connected under water
• The transmitter is galvanically decoupled from the medium. Issues concerning "symmetrical highimpedance" or "asymmetry" or an impedance converter are a thing of the past.
• EMC safety is guaranteed by screening measures for the digital transmission of measured values.
• Intrinsically safe electronics mean operation in hazardous areas is not a problem.
Input
pH value
Temperature
Electrode version AA (for water / wastewater), AS (for boiler feedwater):
pH:1 to 12
Temperature:-15 to 80 °C (5 to 176 °F)
Electrode version BA (for process):
pH:0 to 14
Temperature:0 to 135 °C (32 to 275 °F)
Electrode version FA (for hydrofluoric acid process):
pH:0 to 10
Temperature:0 to 70 °C (32 to 158 °F)
Electrode version BT with ion trap (for chemicals, scrubbers, pulp and paper):
pH:0 to 14
Temperature:0 to 135 °C (32 to 275 °F)
Please note the process operating conditions.
4Endress+Hauser
Page 5
Orbisint CPS11D and CPS11
15°
15°
A
B
Installation
Installation instructions
Do not install the electrodes upside down. The angle of inclination must be at least 15° from the
horizontal. A smaller inclination angle is not permitted as it could cause an air bubble to form in the
glass sphere and prevent the inner electrolyte from completely wetting the pH diaphragm.
NOTICE
Before screwing in the electrode, make sure the threaded connection of the assembly is clean
and runs smoothly.
Screw in the electrode finger-tight (3 Nm)! (Data apply only if installing with Endress+Hauser
‣
assemblies.)
Make sure to follow the installation instructions in the Operating Instructions of the used
‣
assembly.
A0024316
2Electrode installation; installation angle at least 15° from the horizontal
APermitted orientation
BForbidden orientation
Environment
Ambient temperatureNOTICE
Risk of damage due to frost
The sensor must not be used if the temperature drops below –15 °C (5 °F).
‣
Storage temperature
Degree of protection
0 to 50 °C (32 to 122 °F)
IP 68:
IP 68:TOP68 plug-in head, autoclavable up to 135 °C (275 °F), (1 m (3.3 ft) water column,
IP 67:GSA plug-in head (with closed connector system)
Memosens plug-in head, (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)
50 °C (122 °F), 168 h)
Endress+Hauser5
Page 6
Process
A
C
10
12
14
16
18
B
50100200
250
275
3005
0
40
80
120
160
200
246
[psi] p [bar]
T [°F]
400206080100
0
4
6
8
2
T [°C]
120
135
-15
7
1
15
17
101
D
32
176
158
70
Orbisint CPS11D and CPS11
Process temperature
Process pressure (absolute)
Pressure-temperature
ratings (absolute)
Version AA, AS:-15 to 80 °C (5 to 176 °F)
Version BA, BT:0 to 135 °C (32 to 275 °F)
Version FA:0 to 70 °C (32 to 158 °F)
Version AA, AS, FA:1 to 7 bar (15 to 101 psi)
Version BA, BT:1 to 17 bar (15 to 246 psi)
CAUTION
L
Pressurization of sensor due to prolonged use under increased process pressure
Risk of injury from glass breakage
Avoid excessive heating of such sensors if using them under reduced process pressure or under
‣
atmospheric pressure.
When handling such sensors, wear protective goggles and suitable gloves.
‣
3Pressure-temperature ratings
AVersion AA, AS
BVersion BA, BT
CVersion FA
DAtmospheric pressure
Minimum conductivity
Version AA, BA, BT, FA:
Min. 50 μS/cm (minimum flow; pressure and temperature must be
stable)
Version AS:Min. 0.1 μS/cm (stainless steel flow assembly with grounding;
stable and minimum flow; pressure and temperature must be
stable)
pH range
Version AA, AS:
1 to 12 pH
Version BA, BT:0 to 14 pH
Version FA:0 to 10 pH
NOTICE
6Endress+Hauser
Risk of damage to electrode
Never use the electrode outside of the listed specifications!
Glass to suit process
pH membrane glasses:Type A, B, F
Metal lead:Ag/AgCl
Diaphragm:Ring-shaped Teflon® diaphragm, sterilizable, not cytotoxic
Pg 13.5
CPS11D:
NTC30K
CPS11:Pt100, Pt1000
CPS11D:
Memosens plug-in head for digital, non-contact data transmission
A0025729
CPS11:
8Endress+Hauser
ESA:Threaded plug-in head Pg 13.5, TOP68 for electrodes with or without
temperature sensor, 17 bar abs. (246 psi) overpressure protection (threefold),
Ex
GSA:Threaded plug-in head Pg 13.5 for electrodes without temperature sensor
Page 9
Orbisint CPS11D and CPS11
Reference system
Ex approval for CPS11D
Ex approval for CPS11
(TOP68)
Biocompatibility
Version AA, BA, FA: Ag/AgCl reference lead with Advanced Gel 3M KCl, AgCl-free
Version AS:Ag/AgCl reference lead with Advanced Gel, saturated KCl (> 3M KCl) with salt
rings, AgCl-free
The following are indicative of used salt rings (fixed supply of KCl) under
constant process conditions (e.g. stable temperature and flow):
• a continuous upward trend in the pH value (to alkaline pH values)
• a continuous downward trend in the zero point (to acidic pH values)
following adjustment during calibration
BT version:Ag/AgCl reference lead with ion trap and Advanced Gel 3M KCl
Certificates and approvals
• ATEX II 1G Ex ia IIC T3/T4/T6 Ga
• FM / CSA Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters
Hazardous area versions of the digital sensors with Memosens technology are marked by a red-
orange ring in the plug-in head.
• ATEX II 1G Ex ia IIC T3/T4/T6 Ga
• FM Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters
Cytotoxicity verified in accordance with:
USP 2009, chapter <88> (USP Class VI) for diaphragm
TÜV certificate for
Memosens plug-in head
TÜV certificate for TOP68
plug-in head
CPS11D electromagnetic
compatibility
Product page
Product Configurator
Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
Interference emission and interference immunity as per EN 61326: 2012
Ordering information
www.endress.com/cps11d
www.endress.com/cps11
The navigation area is located on the right of the product page.
1.Under "Device support" click "Configure your selected product".
The Configurator opens in a separate window.
2.Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3.Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of
the screen.
Scope of delivery
The scope of delivery includes:
• Sensor in the version ordered
• Technical Information
Endress+Hauser9
Page 10
Orbisint CPS11D and CPS11
Accessories
The following are the most important accessories available at the time this documentation was
issued. For accessories not listed here, please contact your service or sales office.
Assemblies
Cleanfit CPA871
• Flexible process retractable assembly for water, wastewater and the chemical industry
• For applications with standard 12mm sensors
• Product Configurator on the product page: www.endress.com/cpa871
Technical Information TI01191C
Cleanfit CPA875
• Retractable process assembly for sterile and hygienic applications
• For in-line measurement with standard 12 mm sensors for parameters such as pH, ORP and
oxygen
• Product Configurator on the product page: www.endress.com/cpa875
Technical Information TI01168C
Cleanfit CPA472D
• Robust retractable assembly for pH, ORP and other industrial sensors
• Heavy-duty version made of durable materials
• For manual or pneumatic, remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa472d
Technical Information TI00403C
Cleanfit CPA450
• Manual retractable assembly for installing 120 mm sensors in tanks and pipes
• Product Configurator on the product page: www.endress.com/cpa450
Technical Information TI00183C
Cleanfit CPA471
• Compact stainless steel retractable assembly for installation in tanks and pipes, for manual or
pneumatically remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa471
Technical Information TI00217C
Cleanfit CPA472
• Compact plastic retractable assembly for installation in tanks and pipes
• For manual or pneumatic, remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa472
Technical Information TI00223C
Cleanfit CPA473
• Stainless steel process retractable assembly with ball valve shutoff for particularly reliable
separation of the medium from the environment
• Product Configurator on the product page: www.endress.com/cpa473
Technical Information TI00344C
Cleanfit CPA474
• Plastic process retractable assembly with ball valve shutoff for particularly reliable separation of
the medium from the environment
• Product Configurator on the product page: www.endress.com/cpa474
Technical Information TI00345C
Unifit CPA442
• Installation assembly for food, biotechnology and pharmaceutics
• With EHEDG and 3A certificate
• Product Configurator on the product page: www.endress.com/cpa442
Technical Information TI00306C
10Endress+Hauser
Page 11
Orbisint CPS11D and CPS11
Dipfit CPA111
• Immersion and installation assembly made of plastic for open and closed vessels
• Product Configurator on the product page: www.endress.com/cpa111
Technical Information TI00112C
Dipfit CPA140
• pH/ORP immersion assembly with flange connection for very demanding processes
• Product Configurator on the product page: www.endress.com/cpa140
Technical Information TI00178C
Flowfit CPA240
• pH/ORP flow assembly for processes with stringent requirements
• Product Configurator on the product page: www.endress.com/cpa240
Technical Information TI00179C
Flowfit CPA250
• Flow assembly for pH/ORP measurement
• Product Configurator on the product page: www.endress.com/cpa250
Technical Information TI00041C
Ecofit CPA640
• Set comprising adapter for 120 mm pH/ORP electrodes and sensor cable with TOP68 coupling
• Product Configurator on the product page: www.endress.com/cpa640
Technical Information TI00246C
Buffer solutions
Measuring cables
Flexdip CYA112
• Immersion assembly for water and wastewater
• Modular assembly system for sensors in open basins, channels and tanks
• Product Configurator on the product page: www.endress.com/cya112
Technical Information TI00432C
High-quality buffer solutions from Endress+Hauser - CPY20
The secondary buffer solutions have been referenced to primary reference material of the PTB
(German Federal Physico-technical Institute) and to standard reference material of NIST (National
Institute of Standards and Technology) according to DIN 19266 by a DKD (German Calibration
Service) accredited laboratory.
Product Configurator on the product page: www.endress.com/cpy20
CYK10 Memosens data cable
• For digital sensors with Memosens technology
• Product Configurator on the product page: www.endress.com/cyk10
Technical Information TI00118C
CPK9
• Terminated measuring cable for connecting analog sensors with TOP68 plug-in head
• Selection in accordance with product structure
For more information and to order, please contact your sales office.
CPK1
For pH/ORP electrodes with GSA plug-in head
Ordering information is available from your sales office or at www.endress.com.