Endress+Hauser Orbisint CPS11D, Orbisint CPS11 Technical Information

Page 1
TI00028C/07/EN/14.15 71299911
Products Solutions Services
Technical Information
Orbisint CPS11D and CPS11
pH electrodes, analog or with digital Memosens technology
For standard applications in process and environmental technology, with dirt-repellent PTFE diaphragm, built-in temperature sensor (optional for analog sensor)
Application
• Long-term monitoring and limit value monitoring of processes with stable process conditions – Chemical industry: strong acids/bases, plastic, pulp and paper industry – Power plants (e.g. flue gas cleaning), oil and gas – Incinerator plants
• Water/wastewater treatment – Boiler feedwater and cooling water – Well water and drinking water – All industrial and municipal treatment plants
With ATEX, IECEx, FM, CSA, TIIS and NEPSI approval for use in hazardous areas
Your benefits
• Low-maintenance and robust thanks to large PTFE ring junction
• Can be used at pressures up to 17 bar abs. (246 psi)
• Process glass also for very alkaline applications (BA and BT versions)
• Process glass for applications in media containing hydrofluoric acid (FA version)
• For media with low conductivity (AS version)
• Integrated NTC30K temperature sensor (Memosens) for effective temperature compensation; Pt100 or Pt1000 for analog sensors
• Optional: Poison-resistant reference with ion trap
Other advantages of Memosens technology
• Maximum process safety
• Data security thanks to digital data transmission
• Very easy to use as sensor data saved in the sensor
• Recording of sensor load data in the sensor enables predictive maintenance with the Memobase Plus CYZ71D
Page 2

Function and system design

3
4
1
2
Orbisint CPS11D and CPS11

Measuring principle

Measuring system

pH measurement
The pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The membrane glass of the electrode supplies an electrochemical potential which is dependent upon the pH value of the medium. This potential is generated by the selective penetration of H+ ions through the outer layer of the membrane. An electrochemical boundary layer with an electric potential forms at this point. An integrated Ag/AgCl reference system serves as the required reference electrode. The transmitter converts the measured voltage into the corresponding pH value using the Nernst equation.
A complete measuring system comprises at least the following components:
• pH electrode CPS11D or CPS11
• Transmitter, e.g. Liquiline CM42, CM44x, Mycom S CPM153, Liquisys M CPM2x3
• Memosens data cable CYK10 for Memosens sensors or CPK9 for analog sensors
• Immersion, flow or retractable assembly, e.g. Cleanfit CPA871
A0025757
 1 Example of a measuring system for pH measurement
1 Retractable assembly Cleanfit CPA871 2 pH electrode CPS11D 3 Memosens data cable CYK10 4 Liquiline M CM42 two-wire transmitter for hazardous area
2 Endress+Hauser
Page 3
Orbisint CPS11D and CPS11

CPS11D communication and data processing

Communication with the transmitter
Always connect digital sensors with Memosens technology to a transmitter with Memosens technology. Data transmission to a transmitter for analog sensors is not possible.
Digital sensors can store measuring system data in the sensor. These include the following:
• Manufacturer data – Serial number – Order code – Date of manufacture
• Calibration data – Calibration date – Slope at 25 °C (77 °F) – Zero point at 25 °C (77 °F) – Temperature offset – Number of calibrations – Serial number of the transmitter used to perform the last calibration
• Operating data – Temperature application range – pH application range – Date of initial commissioning – Maximum temperature value – Hours of operation under extreme conditions – Number of sterilizations – Resistance of glass membrane
You can display the abovementioned data using the Liquiline CM44x, CM42 and Memobase Plus CYZ71D.

Dependability

Reliability

Integrity Data security thanks to digital data transmission

Easy handling
Sensors with Memosens technology have an integrated electronics unit that stores calibration data and other information (e.g. total operating hours and operating hours under extreme measuring conditions). Once the sensor has been connected, the sensor data are transferred automatically to the transmitter and used to calculate the current measured value. As the calibration data are stored in the sensor, the sensor can be calibrated and adjusted independently of the measuring point. The result:
• Easy calibration in the measuring lab under optimum external conditions increases the quality of the calibration.
• Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the availability of the measuring point .
• Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive maintenance is possible.
• The sensor history can be documented using external storage media and evaluation programs, e.g. Memobase Plus CYZ71D. Thus, the current application of the sensors can be made to depend on their previous history.
Memosens technology digitizes the measured values in the sensor and transmits the data to the transmitter using a non-contact connection that is free from potential interference. The result:
• Automatic error message if sensor fails or connection between sensor and transmitter is interrupted
• Immediate error detection increases measuring point availability
Endress+Hauser 3
Page 4
Orbisint CPS11D and CPS11

Safety

Measured variables

Measuring range

Maximum process safety
With inductive transmission of the measured value using a non-contact connection, Memosens guarantees maximum process safety and offers the following benefits:
• All problems caused by moisture are eliminated: – Plug-in connection free from corrosion – Measured values cannot be distorted by moisture. – Can even be connected under water
• The transmitter is galvanically decoupled from the medium. Issues concerning "symmetrical high­impedance" or "asymmetry" or an impedance converter are a thing of the past.
• EMC safety is guaranteed by screening measures for the digital transmission of measured values.
• Intrinsically safe electronics mean operation in hazardous areas is not a problem.

Input

pH value
Temperature
Electrode version AA (for water / wastewater), AS (for boiler feedwater): pH: 1 to 12 Temperature: -15 to 80 °C (5 to 176 °F) Electrode version BA (for process): pH: 0 to 14 Temperature: 0 to 135 °C (32 to 275 °F) Electrode version FA (for hydrofluoric acid process): pH: 0 to 10 Temperature: 0 to 70 °C (32 to 158 °F) Electrode version BT with ion trap (for chemicals, scrubbers, pulp and paper): pH: 0 to 14 Temperature: 0 to 135 °C (32 to 275 °F)
Please note the process operating conditions.
4 Endress+Hauser
Page 5
Orbisint CPS11D and CPS11
15°
15°
A
B

Installation

Installation instructions

Do not install the electrodes upside down. The angle of inclination must be at least 15° from the horizontal. A smaller inclination angle is not permitted as it could cause an air bubble to form in the glass sphere and prevent the inner electrolyte from completely wetting the pH diaphragm.
NOTICE
Before screwing in the electrode, make sure the threaded connection of the assembly is clean and runs smoothly.
Screw in the electrode finger-tight (3 Nm)! (Data apply only if installing with Endress+Hauser
assemblies.) Make sure to follow the installation instructions in the Operating Instructions of the used
assembly.
A0024316
 2 Electrode installation; installation angle at least 15° from the horizontal
A Permitted orientation B Forbidden orientation

Environment

Ambient temperature NOTICE

Risk of damage due to frost
The sensor must not be used if the temperature drops below –15 °C (5 °F).

Storage temperature

Degree of protection

0 to 50 °C (32 to 122 °F)
IP 68: IP 68: TOP68 plug-in head, autoclavable up to 135 °C (275 °F), (1 m (3.3 ft) water column,
IP 67: GSA plug-in head (with closed connector system)
Memosens plug-in head, (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)
50 °C (122 °F), 168 h)
Endress+Hauser 5
Page 6

Process

A
C
10
12
14
16
18
B
50 100 200
250
275
3005
0
40
80
120
160
200
246
[psi] p [bar]
T [°F]
400 20 60 80 100
0
4
6
8
2
T [°C]
120
135
-15
7
1
15
17
101
D
32
176
158
70
Orbisint CPS11D and CPS11

Process temperature

Process pressure (absolute)

Pressure-temperature ratings (absolute)

Version AA, AS: -15 to 80 °C (5 to 176 °F) Version BA, BT: 0 to 135 °C (32 to 275 °F) Version FA: 0 to 70 °C (32 to 158 °F)
Version AA, AS, FA: 1 to 7 bar (15 to 101 psi) Version BA, BT: 1 to 17 bar (15 to 246 psi)
CAUTION
L
Pressurization of sensor due to prolonged use under increased process pressure
Risk of injury from glass breakage
Avoid excessive heating of such sensors if using them under reduced process pressure or under
atmospheric pressure. When handling such sensors, wear protective goggles and suitable gloves.
 3 Pressure-temperature ratings
A Version AA, AS B Version BA, BT C Version FA D Atmospheric pressure

Minimum conductivity

Version AA, BA, BT, FA:
Min. 50 μS/cm (minimum flow; pressure and temperature must be stable)
Version AS: Min. 0.1 μS/cm (stainless steel flow assembly with grounding;
stable and minimum flow; pressure and temperature must be stable)

pH range

Version AA, AS:
1 to 12 pH Version BA, BT: 0 to 14 pH Version FA: 0 to 10 pH
NOTICE
6 Endress+Hauser
Risk of damage to electrode
Never use the electrode outside of the listed specifications!
A0025761
Page 7
Orbisint CPS11D and CPS11
3
4
5
6
1
2
7
8
12
(0.47)
120, 225, 360
(4.72, 8.86, 14.17)
50 (1.97)
3
4
5
6
1
2
7
8
9
12
(0.47)
120, 225, 360
(4.72, 8.86, 14.17)
50 (1.97)

Design, dimensions of CPS11D

Mechanical construction

A0025726
 4 CPS11D with Memosens plug-in head, temperature
sensor
1 Memosens plug-in head, Pg 13.5 2 Viton O-ring with thrust collar 3 Ag/AgCl reference lead - reference 4 "Advanced Gel" electrolyte 5 Ag/AgCl reference lead - pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor NTC30K
A0025725
 5 CPS11D-7BTxx, with ion trap
1 Memosens plug-in head, Pg 13.5 2 Viton O-ring with thrust collar 3 Ag/AgCl reference lead - reference 4 "Advanced Gel" electrolyte 5 Ion trap 6 Ag/AgCl reference lead - pH 7 PTFE diaphragm 8 pH glass membrane 9 Temperature sensor NTC30K
Endress+Hauser 7
Page 8

Design, dimensions of CPS11

1
2
3
4
6
5
7
120, 225, 360, 425
(4.72, 8.86, 14.17, 16.73)
12 /
0.47
3
4
5
6
1
2
7
8
120, 225, 360, 425
(4.72, 8.86, 14.17, 16.73)
12 /
0.47
Orbisint CPS11D and CPS11

Weight

Materials

Process connection

Temperature sensor

Plug-in heads

A0025733
 6 CPS11 with GSA plug-in head
1 GSA plug-in head, Pg 13.5 2 Viton O-ring with thrust collar 3 Ag/AgCl reference lead - reference 4 "Advanced Gel" electrolyte 5 Ag/AgCl reference lead - pH 6 PTFE diaphragm 7 pH glass membrane
 7 CPS11 with TOP68 plug-in head, temperature sensor
1 TOP68 plug-in head, Pg 13.5 2 Viton O-ring with thrust collar 3 Ag/AgCl reference lead - reference 4 "Advanced Gel" electrolyte 5 Ag/AgCl reference lead - pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor Pt100
0.1 kg (0.2 lbs)
Electrode shaft:
Glass to suit process pH membrane glasses: Type A, B, F Metal lead: Ag/AgCl Diaphragm: Ring-shaped Teflon® diaphragm, sterilizable, not cytotoxic
Pg 13.5
CPS11D:
NTC30K
CPS11: Pt100, Pt1000
CPS11D:
Memosens plug-in head for digital, non-contact data transmission
A0025729
CPS11:
8 Endress+Hauser
ESA: Threaded plug-in head Pg 13.5, TOP68 for electrodes with or without
temperature sensor, 17 bar abs. (246 psi) overpressure protection (threefold), Ex
GSA: Threaded plug-in head Pg 13.5 for electrodes without temperature sensor
Page 9
Orbisint CPS11D and CPS11

Reference system

Ex approval for CPS11D

Ex approval for CPS11 (TOP68)

Biocompatibility

Version AA, BA, FA: Ag/AgCl reference lead with Advanced Gel 3M KCl, AgCl-free Version AS: Ag/AgCl reference lead with Advanced Gel, saturated KCl (> 3M KCl) with salt
rings, AgCl-free
The following are indicative of used salt rings (fixed supply of KCl) under constant process conditions (e.g. stable temperature and flow):
• a continuous upward trend in the pH value (to alkaline pH values)
• a continuous downward trend in the zero point (to acidic pH values) following adjustment during calibration
BT version: Ag/AgCl reference lead with ion trap and Advanced Gel 3M KCl

Certificates and approvals

• ATEX II 1G Ex ia IIC T3/T4/T6 Ga
• FM / CSA Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters Hazardous area versions of the digital sensors with Memosens technology are marked by a red-
orange ring in the plug-in head.
• ATEX II 1G Ex ia IIC T3/T4/T6 Ga
• FM Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters
Cytotoxicity verified in accordance with: USP 2009, chapter <88> (USP Class VI) for diaphragm

TÜV certificate for Memosens plug-in head

TÜV certificate for TOP68 plug-in head

CPS11D electromagnetic compatibility

Product page

Product Configurator

Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
Interference emission and interference immunity as per EN 61326: 2012

Ordering information

www.endress.com/cps11d
www.endress.com/cps11
The navigation area is located on the right of the product page.
1. Under "Device support" click "Configure your selected product".  The Configurator opens in a separate window.
2. Select all the options to configure the device in line with your requirements.  In this way, you receive a valid and complete order code for the device.
3. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of the screen.

Scope of delivery

The scope of delivery includes:
• Sensor in the version ordered
• Technical Information
Endress+Hauser 9
Page 10
Orbisint CPS11D and CPS11

Accessories

The following are the most important accessories available at the time this documentation was issued. For accessories not listed here, please contact your service or sales office.

Assemblies

Cleanfit CPA871
• Flexible process retractable assembly for water, wastewater and the chemical industry
• For applications with standard 12mm sensors
• Product Configurator on the product page: www.endress.com/cpa871 Technical Information TI01191C
Cleanfit CPA875
• Retractable process assembly for sterile and hygienic applications
• For in-line measurement with standard 12 mm sensors for parameters such as pH, ORP and
oxygen
• Product Configurator on the product page: www.endress.com/cpa875 Technical Information TI01168C
Cleanfit CPA472D
• Robust retractable assembly for pH, ORP and other industrial sensors
• Heavy-duty version made of durable materials
• For manual or pneumatic, remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa472d Technical Information TI00403C
Cleanfit CPA450
• Manual retractable assembly for installing 120 mm sensors in tanks and pipes
• Product Configurator on the product page: www.endress.com/cpa450 Technical Information TI00183C
Cleanfit CPA471
• Compact stainless steel retractable assembly for installation in tanks and pipes, for manual or
pneumatically remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa471 Technical Information TI00217C
Cleanfit CPA472
• Compact plastic retractable assembly for installation in tanks and pipes
• For manual or pneumatic, remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa472 Technical Information TI00223C
Cleanfit CPA473
• Stainless steel process retractable assembly with ball valve shutoff for particularly reliable
separation of the medium from the environment
• Product Configurator on the product page: www.endress.com/cpa473 Technical Information TI00344C
Cleanfit CPA474
• Plastic process retractable assembly with ball valve shutoff for particularly reliable separation of
the medium from the environment
• Product Configurator on the product page: www.endress.com/cpa474 Technical Information TI00345C
Unifit CPA442
• Installation assembly for food, biotechnology and pharmaceutics
• With EHEDG and 3A certificate
• Product Configurator on the product page: www.endress.com/cpa442 Technical Information TI00306C
10 Endress+Hauser
Page 11
Orbisint CPS11D and CPS11
Dipfit CPA111
• Immersion and installation assembly made of plastic for open and closed vessels
• Product Configurator on the product page: www.endress.com/cpa111 Technical Information TI00112C
Dipfit CPA140
• pH/ORP immersion assembly with flange connection for very demanding processes
• Product Configurator on the product page: www.endress.com/cpa140 Technical Information TI00178C
Flowfit CPA240
• pH/ORP flow assembly for processes with stringent requirements
• Product Configurator on the product page: www.endress.com/cpa240 Technical Information TI00179C
Flowfit CPA250
• Flow assembly for pH/ORP measurement
• Product Configurator on the product page: www.endress.com/cpa250 Technical Information TI00041C
Ecofit CPA640
• Set comprising adapter for 120 mm pH/ORP electrodes and sensor cable with TOP68 coupling
• Product Configurator on the product page: www.endress.com/cpa640 Technical Information TI00246C

Buffer solutions

Measuring cables

Flexdip CYA112
• Immersion assembly for water and wastewater
• Modular assembly system for sensors in open basins, channels and tanks
• Product Configurator on the product page: www.endress.com/cya112 Technical Information TI00432C
High-quality buffer solutions from Endress+Hauser - CPY20
The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal Physico-technical Institute) and to standard reference material of NIST (National Institute of Standards and Technology) according to DIN 19266 by a DKD (German Calibration Service) accredited laboratory. Product Configurator on the product page: www.endress.com/cpy20
CYK10 Memosens data cable
• For digital sensors with Memosens technology
• Product Configurator on the product page: www.endress.com/cyk10 Technical Information TI00118C
CPK9
• Terminated measuring cable for connecting analog sensors with TOP68 plug-in head
• Selection in accordance with product structure For more information and to order, please contact your sales office.
CPK1
For pH/ORP electrodes with GSA plug-in head
Ordering information is available from your sales office or at www.endress.com.
Endress+Hauser 11
Page 12
www.addresses.endress.com
Loading...