EMERSON LPS60-M Service Manual

Page 1
Rev. 07.23.07
LPS60-M Series
1 of 3
Embedded Power for Business-Critical Continuity
LPS60-M Series
60 Watts Medical
• Medical safety approvals
• Universal input
• 3" x 5" footprint
•Remote sense
• Built in EMI filter
• Low output ripple
• Adjustable output voltage
• Overload protection
• 110 KHz switching frequency
• LPX60 enclosure kit available
Total Power: 60 - 80 Watts Input Voltage: 85 - 264 VAC
120 - 300VDC
# of Outputs: Single
Electrical Specifications
Safety
Input
Input ra nge 85 - 264 VAC (wide range)
120 - 300VDC
Frequency 47-440 Hz
Inrush current <18A peak @ 115VAC, <36A peak @230 VAC,
cold start @ 25°C
Input current 1.5A max. (RMS) @ 115 VAC
Efficiency 70% typical at full load
EMI filter FCC Class A conducted; CISPR22 Class A conducted;
EN55022 Class A conducted; VDE0878 PT3 Class A conducted
Safety ground leakage current
<75μA @ 50/60Hz; 264VAC input
Output
Maximum power 60W for convection; 80W with 30CFM forced air Adjustment range -5, +10% minimum
Hold-up time 20ms @ 60W load, 115VAC nominal line
Overload protection Short circuit protection on all outputs.
Case overload protected @ 110-145% above peak rating
Overvoltage protection 5V output; 5.7 - 6.7VDC. Other outputs 10% to 25%
above nominal output
Remote sense Compensates for 0.5V lead drop min. Will operate without
remote sense connected. Reverse connection protected.
VDE 0750/EN60950 (IEC601)
21310-3336-0015
UL UL2601
E182560
CSA CSA 22.2 No.601.1
LR109492
CE Mark (LVD)
RoHS
Page 2
Mechanical Drawing
Rev. 07.23.07
LPS60-M Series
2 of 3
Operating temperature: 0° to 50°C ambient derate each output as 2.5%
per degree from 50° to 70°C, -20°C startup
Storage temperature: -40°C to +85°C
Electromagnetic susceptibility:
designed to meet IEC801, -2, -3, -4, -5, -6, Level 3
Humidity: Operating; non-condensing 5% to 95% RH
Vibration: Three orthogonal axes, sweep at 1cot/min, 5 min.
dwell at four major resonances 0.75G peak 5Hz to 500Hz, operational
Temperature coefficient ±0.04% per °C
MTBF demonstrated >550,000 hours at full load and 25°C ambient
conditions
Environmental Specifications
0.22" (5.6)
GND
M1
3.00"
(76.2)
2.55"
(64.8)
SK 1
1
N
3
L
Output
Power
(Watts)
80W
60W
40W
20W
0W
SK201
Power Derating Curve
30 CFM Forced Air Cooling
Natural
Convection
Cooling
0°C
10°C 20°C 30°C 40°C 50°C 60°C 70°C
TA – Ambient Temperature – °C
M2
Remote Sense
+
_
Connector
Voltage Adj. Pot
SK 2
1 2 3 4 5 6
40W
30W
0.22" (5.6)
1.69"
(42.9)
0.12" (3)
4.56"
(115.8)
0.15" (4)
DIA (4 places)
5.00" (127)
Page 3
Rev. 07.23.07
LPS60-M Series
3 of 3
Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. ©2007 Emerson Electric Co.
Embedded Power for Business-Critical Continuity
Emerson Network Power.
The global leader in enabling business-critical continuity.
EmersonNetworkPower. com
AC Power
Connectivity
DC Power
Embedded Power
Inbound Power Integrated Cabinet Solutions Outside Plant
Precision Cooling
Site Monitoring and Services
Americas
5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698
Europe (UK)
Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355
Asia (HK)
16th - 17th Floors, Lu Plaza 2 Wing Yip Street, Kwun Tong Kowloon, Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888
For global contact, visit:
www.astecpower.com www.artesyn.com technicalsupport@astec.com technicalsupport@artesyn.com
While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions.
Ordering Information
Model
Number
Output
Voltage
Minimum
Load
Maximum Load
with Convection
Cooling
Maximum Load
with 30CFM
forced Air
Peak
Load
1
Regulation
2
Ripple
P/P (PARD)
3
LPS63-M 12V 0A 5A 6.7A 7.5A ±2% 120mV
LPS64-M 15V 0A 4A 5.3A 6A ±2% 150mV
LPS65-M 24V 0A 2.5A 3.3A 3.8A ±2% 240mV
1. Peak current lasting <30 seconds with a maximum 10% duty cycle.
2 . At 25°C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings.
3 . Peak-to-peak with 20 MHz bandwidth and 10μF (tantalum capacitor) in parallel with a 0. μF capacitor at rated
line voltage and load ranges.
Pin Assignments
Connector LPS62-M LPS63-M LPS65-M
SK1-1 Neutral Neutral Neutral SK1-3 Line Line Line
SK2-1 +12V +15V +24 SK2- 2 +12V +15V +24 SK2- 3 +12V +15V +24 SK2-4 Common Common Common SK2-5 Common Common Common SK2-6 Common Common Common SK201-1 +Sense +Sense +Sense SK201-2 -Sense -Sense -Sense
Mating Connectors
AC Input:
Molex 09-50-8031 (USA) Not required for (-T) option 09-91-0300 (UK) PINS: 08-58-0111
DC Outputs:
Molex 09-50-8061 (USA) Not required for (-T) option 09-91-0600 (UK) PINS: 08-52-0113
Remote Sense:
Molex 22-01-2025 PINS: 08-52-0113
Astec Connector Kit #70-841-006, includes all of the above
1. Specifications subject to change without notice.
2. All dimensions in inches (mm), tolerance is ±0.02" (±0.5mm)
3. Mounting holes M1 and M2 should be grounded for EMI purposes.
4. Mounting hole M1 is safety ground connection.
5. Specifications are for convection rating at factory settings at 115 VAC input, 25°C unless otherwise stated.
6. Warranty: 2 year
7. Weight: 0.75lbs/0.34kg