Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical rep-
resentative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
•Install your equipment as specified in the Installation Instructions of the appropriate Instruc-
tion Manual and per applicable local and national codes. Connect all products to the proper
electrical and pressure sources.
•To ensure proper performance, use qualified personnel to install, operate, update, program,
and maintain the product.
•When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
•Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Ryton® is a registered trademark of Phillips Petroleum Co.
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
e-mail: gas.csc@EmersonProcess.com
652542 Outline and Mounting – Panel Mount
652543 Outline and Mounting – Wall Mount
652544 Outline and Mounting – Pipe Mount
652682 Pictorial Wiring Diagram
652683 Interconnect Diagram
652699 Installation Drawing
iv ContentsRosemount Analytical Inc. A Division of Emerson Process Management
Operator Manual
Model 7003M
PREFACE
The purpose of this manual is to provide information concerning the components,
functions, installation and maintenance of the Model 7003M Percent Oxygen Analyzer
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it. Read
this instruction manual completely.
DEFINITIONS
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout
this publication.
DANGER .
748223-K
June 2002
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial
property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure,
condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Operator Manual
748223-K
June 2002
Model 7003M
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems may be
impaired.
AUTHORIZED PERSONNEL
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property,
all personnel authorized to install, operate and service the this equipment should be thoroughly
familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can
cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded
three-wire source of power.
Relay contacts wired to separate power sources must be disconnected before servicing.
This instrument was shipped from the factory set up to operate on either 115 VAC 50/60 Hz, or 230
VAC 50/60 Hz, as specified by sales order. Verification that this voltage is proper for the installation
is the responsibility of the purchaser. See Section 2-5b on page 2-4.
WARNING .
OXYGEN LEVEL WILL NOT BE MONITORED
The instrument will not return automatically to the RUN mode from HOLD or DIAGNOSTICS modes.
Operator must manually return instrument to RUN mode to continue monitoring oxygen.
WARNING.
ERRONEOUS OXYGEN READINGS
If the pressure compensation function is not applied correctly, erroneous readings can be displayed. The display using span gas during calibration should be the same as the reading using the
same gas in RUN mode if the pressure values are correctly entered.
P-2 PrefaceRosemount Analytical Inc. A Division of Emerson Process Management
Operator Manual
Model 7003M
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair
WARNING.
SENSOR NOT INTRINSICALLY SAFE
The oxygen sensors used with this analyzer are not intrinsically safe. Use of the sensors in or near
flammable liquids, gases or vapors or in Class I, Division 1, Hazardous Locations should be carefully evaluated by qualified personnel at the site and is entirely the responsibility of the user.
The sensor is approved as non-incendive for use in Class I, Division 2 locations.
CAUTION .
748223-K
June 2002
HIGH PRESSURE GAS CYLINDERS
This module requires periodic use of pressurized gas. See General Precautions for Handling and
Storing High Pressure Gas Cylinders, page P-4
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Operator Manual
748223-K
June 2002
Model 7003M
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed
Gases" published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather
and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or
placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125
permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should never be
P-4 PrefaceRosemount Analytical Inc. A Division of Emerson Process Management
Operator Manual
Model 7003M
DOCUMENTATION
The following Model 7003M instruction materials are available. Contact Customer Service Center or the
local representative to order.
748223 Operator Manual (this document)
COMPLIANCES
The Model 7003M Percent Oxygen Analyzer and Models 4000 and 5000 series sensors are suitable for
use in Class I, Groups A, B, C, and D, Division 2 locations per North American electrical codes. Factory
Mutual Research (FM) approved under J.I. 1T5AO.AX. Canadian Standards Association (CSA) certified
under file LR 93812.
748223-K
June 2002
FM
APPROVED
®
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Operator Manual
748223-K
June 2002
Model 7003M
P-6 PrefaceRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 OVERVIEW
The Model 7003M Percent Oxygen Analyzer
continuously monitors the concentration of
oxygen in a gaseous sample. It consists of a
sensor and the analyzer. The sensor is
housed in a chamber assembly and is
connected to the analyzer by a shielded multiconductor cable.
The Model 7003M Percent Oxygen Analyzer
provides direct digital readout, isolated current
MODEFUNCTION
RUN (see Section 3-4 on page 3-10)Normal monitoring
SETUP (see Section 3-5 on page 3-11)Changing parameters
ALARM/CONTROL (see Section 3-6 on page 3-16)Set alarms or ON/OFF controllers
CALIBRATION (see Section 4-2 on page 4-1)Gas calibration/pressure inputs
HOLD (see Section 4-3 on page 4-5)Retain last output value
DIAGNOSTICS (see Section 4-4 on page 4-8)Electronics checks
The Model 7003M Percent Oxygen Analyzer
is housed in a NEMA 4X (CSA enclosure 4),
corrosion resistant housing suitable for panel,
pipe or wall mounting.
The analyzer has a microprocessor controlled
keyboard for operation. Signals from the
remote oxygen and temperature sensors are
processed to compute real time oxygen
concentration corrected for temperature,
constant pressure differences and residual
sensor current. The oxygen concentration is
compared with user-set alarm parameters,
and also generates an isolated 4 to 20 or 0 to
20mA signal. Data is stored in a memory
circuit that retains the data in case power is
removed.
The analyzer is shipped from the factory
configured for operation on either 115 VAC or
230 VAC, 50/60 Hz as specified by the
customer when ordered.
output, automatic temperature compensation
circuitry, and dual relay contact closures for
alarms or ON/OFF control devices as
standard features. The alarm ranges can be
selected independently from the output range,
eliminating the need to readjust alarm
setpoints if the range is changed.
The analyzer has six modes of operation:
1-2 PRINCIPLE OF MEASUREMENT
Measurement is based on the electrochemical reaction between oxygen and the
sensor cathode which produces a low level
electrical current. The sensor signal is
proportional to the partial pressure of oxygen
present at the membrane. This signal and a
thermistor signal are received by the analyzer
and processed by the electronics to output a
calibrated oxygen signal that is compensated
for temperature.
1-3 SENSORS
Rosemount Analytical offers rechargeable and
disposable oxygen sensors which can be
used with the Model 7003M. These sensors
are supplied in kits: Submersion, in-line flow,
fast response, and a cell separation kit.
Sensors are available constructed of
polypropylene or Ryton. See Section 74REPLACEMENT PARTS - SENSORS on
page 7-3, for additional information.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Operator Manual
748223-K
June 2002
Model 7003M
1-4 CONDITIONS AFFECTING MEASUREMENT
Oxygen measurement is affected by the
following:
•
Composition of the sampled gas
•
Sample and calibration gas pressures
•
Environment of sensor, cable and
analyzer
Each must be considered when calibrating
and operating the analyzer to obtain accurate
measurements.
a. Gas Composition
Any substance in the sample that
significantly interferes or "poisons" the
sensor oxygen reading should be
removed by a chemical and/or
mechanical sample handling system (see
Section 5-2 on page 5-1). If ambient air is
not used as the span gas, the
composition of calibration gases should
also be representative of the sample.
The sample and calibration gases must
meet certain specifications (see
Specifications in Preface) to ensure
accurate measurement. Physical
parameters that affect measurement
include:
b. Sample and Cal Gas Pressures
The sample and calibration gases may be
at different pressures as long as the
pressure of each remains constant. Any
variation in the pressure of either gas will
result in an incorrect output reading of the
same proportion.
Two methods of sample calibration are
recommended:
1. Calibrate at the sample pressure, or
2. Calibrated at a different pressure, and
enter the two pressure in the analyzer
(Section 4-2c on page 4-4)
Refer to Section 5-2 on page 5-1 for a
more detailed discussion of the
relationship between gas pressures and
apparent oxygen concentration.
c. Environmental Factors
If radio frequency interference (RFI)
normally occurs in the operating
environment, the sensor cable should be
electrically shielded during installation.
•
Pressure
•
Humidity
•
Temperature
•
Gas composition
Also refer to Section 5-2 on page 5-1 for
additional information about how gas
conditions affect the oxygen
measurement.
1-2 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
1-5 FEATURES
Suppressed Ranges
When setting the Model 7003M to a
suppressed output range (15 to 25%, etc.),
note that the absolute accuracy, drift, and
noise of the oxygen measurement will remain
the same for all ranges (see Specifications in
Preface).
ALARM Mode
ALARM Mode is not self-resetting. Operator
must press ACK to bring relays back to their
no-alarm state. Used when immediate
operator action is required.
NOTE:
Operator Can Bypass Alarm Relays
When relays are set to ALARM Mode (Section 3-6a on page 3-16) and an alarm condition has occurred, pressing the ACK key
will turn off the relay switch, whether the
oxygen level is out of range or not. If you
do not want an operator to be able to bypass the relay switches in this way, set the
instrument to ON/OFF -CONTROL Mode.
ON/OFF-Controller Mode
ON/OFF-CONTROLLER Mode is selfresetting. Pressing ACK has no effect on
relay coils. Used when automated action is
required.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Signal Output................................. Isolated 4 to 20 or 0 to 20 mA (selectable)
Alarm Relay Contacts.................... Two Form C Relays
Recorder Output............................ Isolated 4 to 20 mA
Operator Manual
748223-K
June 2002
600 ohms maximum load
SPDT, resistive load, independently adjustable
Deadband adjustable over full range
Alarm Relay Contacts Ratings
General Purpose (Ordinary) Locations:
115 VAC, 3.0 A (resistive), 230 VAC, 1.5 A resistive
Class I, Division 2 Locations:
CSA: 120 VAC, 3.0 A (non-inductive)
240 VAC, 1.5 A (non-inductive).
FM: 28 VDC (resistive) (Ci = 0, Li = 0)
Groups A&B: 150 mA
Group C: 400 mA
Group D: 540 mA
d. Sensors
Types............................................. Rechargeable and Disposable
Process Connections .................... Refer to flow chamber (if supplied)
Sample Pressure........................... 0 to 50 psig (0 to 345 kPa)
Sample Temperature..................... 32 to 110°F (0 to 44°C)
Ambient Temperature.................... 32 to 110°F (0 to 44°C)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Operator Manual
748223-K
June 2002
Model 7003M
1-6 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 2
INSTALLATION
2-1 UNPACKING
Carefully examine the shipping carton and
contents for signs of damage. Immediately
notify the shipping carrier if the carton or
contents is damaged. Retain the carton and
packing material until all components
associated with the Model 7003M Percent
Oxygen Analyzer are operational.
2-2 STORAGE
The sensors and analyzer unit may be
shipped together or separately. The sensors
are shipped assembly and charged, ready for
use.
If rechargeable sensors are stored longer than
six months, they may need to be recharged or
rejuvenated if performance is marginal.
Disposable sensors are not rechargeable.
Refer to documentation supplied with sensors
for more information.
2-3 ANALYZER INSTALLATION
The Model 7003M Percent Oxygen Analyzer
is designed to meet NEMA Type 4X (CSA
enclosure 4) requirements, provided that the
cutouts in the bottom of the enclosure are
fitted with approved conduit fittings or are
properly sealed by the user with Hoffman
plugs (PN A-S050) or equivalent.
The Model 7003M is supplied with the Panel
Mounting Kit PN 652527 as standard.
Optional Wall Mount Kit PN 652539 and Pipe
Mount Kit PN 652529 are available. Refer to
the Outline and Mounting drawings provided
in the rear of this manual.
2-4 SENSOR INSTALLATION
NOTE:
Refer to Figure 2-1 Sensor Ordering Matrix
on page 2-2, and verify that the sensor to
be used corresponds with the application.
If an incorrect sensor is used, calibration
may be impossible or incorrect. If a different type of sensor is to be used, the rocker
switch (SW1) on the Signal Board may
have to be changed (see Section 2-4a
below).
Install the sensor in an area where
temperature is relatively constant, taking care
to avoid changes in exposure to sunlight or
due to sudden drafts. A room-temperature
environment is preferable. The flow
chamber/sensor kit contains detailed
mounting and sample flow information. Add
distilled or deionized water to the sensor to
the maximum level indication on the sensor
reservoir. Let sensor stand for approximately
15 minutes and check for leaks around the
base of the reservoir, and at the seams and
corners. If a leak is found, contact the factory
before proceeding. Drain the sensor.
a. Analyzer/Sensor Jumper Configuration
The Model 7003M is configured to match
the type of sensor ordered with the
analyzer. If the Signal Board or Sensor is
repaired or replaced, verify rocker switch
SW1 on the Signal Board is properly set.
Rechargeable Sensor: Close
positions 1, 4, and 5
Disposable Sensor: Close positions
2, 4, and 6
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Operator Manual
748223-K
June 2002
Model 7003M
4000 Rechargeable Sensor
5000 Disposable Sensor
1 Polypropylene
2 Ryton (rechargeable only)
9 Special (consult factory)
1 Submersion
2 In-Line Flow
3 Fast Response (rechargeable only)
9 Special (consult factory)
4 0 0 03
Position →
Positions 1 through 4 = Sensor Type
Position 5 = Material of Construction (Sensor and Flow Assembly)
Position 6 = Flow Assembly Configuration
Example 400013 is a Rechargeable sensor, constructed of polypropylene, mounted in a
Fast-Response Flow Chamber Assembly.
1234
Figure 2-1. Sensor Ordering Matrix
1
5
6
2-2 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
2-5 ELECTRICAL CONNECTIONS
Observe all precautions stated in this manual
and in the drawings provided at the back of
this manual before installation.
All electrical connections to the analyzer are
made on the Power Supply Board connection
panel (see Figure 2-2 below and drawing
652683), which is accessed by removing the
rear cover.
WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without rear cover secured.
Servicing requires access to live parts
which can cause death or serious injury.
Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly
grounded three-wire source of power.
Electrical installation must be made in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any applicable national or local codes.
WARNING
ENCLOSURE INTEGRITY
Unused cable conduit entries must be securely sealed by non-flammable closures
to provide enclosure integrity in compliance with personnel safety and environmental protection requirements. For
watertight installation conforming to the
requirements of NEMA Type 4X, use
Hoffman plugs (PN A-S050) or equivalent
to seal any unused openings.
Provide three separate 1/2 inch conduits as
follows:
•
Power cable
•
Sensor cable
•
Alarm cables
1 +12V
2 -12V
1 ANODE
2 TEMP SENSE
3 NC
4 NC
5 EPOL
6 GND
7 Tx
8 Rx
9 RET
1 COM
2 NC
3 NO
4 COM
5 NC
6 NO
7 ISO GND
8 ISO CUR9 ISO CUR+
ALARM CONTACT RATING:
3A 120 VAC
1.5A 240 VAC
3A 30 VDC
AC POWER
AL 1
AL 2
HOT NEUT GND
L1 L2
FUSE RATING:
115V: 1/4A
250V
230V: 1/8A
250V
Figure 2-2. Model 7003M Rear Panel Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Operator Manual
748223-K
June 2002
a. Sensor Cable
The sensor cable used with the Model
7003M may be of any length up to a
maximum of 1000 feet (305 m). (See
Table 2-1 below.)
The signal cable should be routed through
a conduit (customer-supplied) in
permanent installations or in cases where
the sensor is located more than a few feet
from the control unit. (See Table 2-2
below.)
Model 7003M
Connect Sensor Cable
1. Ensure that power is disconnected
and remove rear cover.
2. Install the conduit or cable clamp in
the left opening in the bottom rear of
the case.
3. Install the sensor cable in the
instrument through the cable clamp
and tighten the nut on the plug to
create a seal around the cable.
NOTE:
In humid environments, temperature
changes can cause moisture to condense in the conduit, resulting in signal noise and corrosion. To minimize
the effect of condensation, put desiccant (available locally) inside the conduit before sealing it.
CABLE PART
NUMBER
19174810 Feet (3.1 m)
193265SPECL
Table 2-1.Sensor Cables
If conduit used:
1/2 inch NPT female connection on
bottom of control unit. (Conduit customer-supplied.
If no conduit used:
Use cable connector assembly (PN
856831) to provide watertight seal. Assembly includes: Aluminum inner sealing grommet for connector compression
nut.
Table 2-2.Analyzer Cable
Connections
LENGTH
Any specified length
up to 1000 Feet (305
m)
4. Refer to Figure 2-2 on page 2-3.
Using a flat-blade screwdriver,
connect the wires of the sensor cable
to the TB2-1, TB2-2, TB2-5 and TB26 on the rear panel of the analyzer.
5. Replace the rear cover if no other
connections are to be made. Do not
overtighten the cover retaining screws
to avoid stripping the threads in the
plastic case.
b. Power Connections
Electrical power is supplied to the
analyzer via a customer-supplied three
conductor cable, type SJT, minimum 18
AWG. Refer to the installation drawing in
the rear of this manual.
Jumpers and Fuses
To verify the jumper location for the
specified power (115 or 230 VAC) or if the
analyzer is to be connected to a different
power source, refer to Figure 2-3 on page
2-5and set jumpers and fuse as follows:
WARNING
ELECTRICAL SHOCK HAZARD
Disconnect power to analyzer before continuing.
2-4 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
WARNING
1. Remove the rear cover from the
analyzer.
2. Remove Power Supply Board.
3. Verify and/or set Jumpers E1, E2, E3,
and E4 per Figure 2-3 below.
4. Verify that fuse (F1) is correct for
power.
5. Re-install Power Supply Board.
6. Replace the rear cover if no other
connections are to be made. Do not
overtighten the cover retaining screws
to avoid stripping the threads in the
plastic case.
Conduit Connections
Route the power cable through conduit
and into the central opening in the bottom
of the case. The inlet hole is sized for
standard 1/2 inch conduit fittings.
ENCLOSURE NOT GROUNDED
The non-metallic enclosure does not provide grounding between conduit connections. Use grounding-type bushing and
jumper wires.
Connect Power Cable
Connect power wiring to the compression
type connectors on the rear panel (Figure
2-2 on page 2-3) as follows:
1. Loosen the connector screw.
2. Strip insulation back approximately
1/4 inch (6.35 mm), twist stranded
wires together. DO NOT tin braid
wire or the connection can work
loose.
3. Insert wire into connector about 1/4
inch (6.25 mm), ensuring that wires
do not touch any other wire or
connector.
4. Tighten screw.
115 VAC
Jumpers: E1 - E2
E3 - E4
Fuse: 1/4 Amp
115 VAC
Jumpers: E2 - E4
Fuse: 1/8 Amp
Power Supply Board
F1 E3 E4 E2 E1
1/4 AMP 115 VAC
F1 E3 E4 E2 E1
1/8 AMP 230 VAC
Figure 2-3. Power Jumpers and Fuses
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Operator Manual
748223-K
June 2002
Model 7003M
CAUTION
TOTAL OUTPUT RESISTANCE
Total resistance of all output devices and
associated interconnection cable must not
exceed 600 ohms.
c. Current Output Connections
Refer to Figure 2-4 on page 2-7. Isolated
current output (0 to 20 or 4 to 20 mA) is
standard on the Model 7003M. To
convert from current output to voltage
output, see Section 2-5d below.
1. Disconnect power to the analyzer.
Remove rear cover.
2. Connect the shielded recorder cable
leads to the rear panel of the analyzer
(Figure 2-2 on page 2-3):
TB3-7 has no connection
TB3-8 negative (-)
TB3-9 positive (+)
3. Connect the other end of the output
cable to terminals of the output
device. Ensure that polarity is
correct. Connect the shield to a
suitable ground point.
NOTE:
Current output terminals are not
grounded. Ground the current loop at
some point within the circuit to minimize noise or other undesirable interactions.
4. Replace the rear cover. Do not
overtighten cover retaining screws, to
avoid stripping the threads in the
plastic case.
5. To connect recorder(s), see Figure
2-4 on page 2-7.
d. Voltage Output Connections
Refer to Figure 2-4 on page 2-7. To
convert the voltage output to current
output see Section 0 above.
To monitor the signal output with a
voltage recorder or other voltage device,
the standard current output must be
converted into a voltage by connecting a
resistor in parallel with the output device
as follows:
1. Determine the resistor value (R)
required:
R
ohms = (fullscale voltage,
1
volts)/(0.02 amps)
2. Disconnect power; remove the rear
cover.
3. Connect resistor (1/4 watt or greater)
of value R between TB3-8 and TB3-9
(Figure 2-4 on page 2-7).
4. Connect leads of a shielded recorder
cable to the rear panel of the analyzer
(Figure 2-2 on page 2-3):
TB3-7 has no connection
TB3-8 negative (-)
TB3-9 positive (+)
5. Connect other end of output cable to
terminals of output device. Ensure
that polarity is correct.
6. Replace the rear cover. Do not
overtighten cover retaining screws, to
avoid stripping the threads in the
plastic case.
7. For the low end point to be 0 volts, set
the current output range to 0 to 20mA
when setting up the instrument
(Section 3-5b on page 3-13).
2-6 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
TB3
RECORDER
S - +
Operator Manual
748223-K
June 2002
ONE RECORDER
7
8
9
7
8
9
NEG (-)
POS (+)
NEG (-)
POS (+)
RECORDER
A
S - +
RECORDER
B
S - +
MULTIPLE
RECORDERS
Figure 2-4. Current Output Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Operator Manual
748223-K
June 2002
Model 7003M
2-6 RELAY CONTACTS FOR ALARMS OR
ON/OFF-CONTROLS
NOTE:
Whenever "Alarm" is used, it refers to
either alarm or ON/OFF Control, depending on how the relays are connected and configured by the
customer.
Ratings
Two sets of alarm relay contacts are
provided for actuation of customersupplied alarm and/or ON/OFF-Control
functions. (see Section 1-6 specifications
on page 1-4).
Power Failure
Removal of AC power from the analyzer,
as in a power failure, removes power from
both relay coils. The Model 7003M
should operate on a different AC power
source than the Alarm relay contacts.
Connecting Devices to Relay Contacts
WARNING
RELAY CONTACTS
Relay contacts wired to separate power
source must be disconnected before
servicing.
Connect alarm or ON/OFF-Control
devices to the relay contacts as follows:
1. Disconnect power; remove rear cover
of analyzer.
2. Connect leads from the external
alarm or ON/OFF Control system
(see Figure 2-2 on page 2-3).
TB3-1Alarm 1 Com
TB3-2Alarm 1 NC
TB3-3Alarm 1 NO
TB3-4Alarm 2 Com
TB3-5Alarm 2 NC
TB3-6Alarm 2 NO
Radio Frequency Interference (RFI)
If alarm contacts are connected to any
device that causes radio frequency
interference (RFI), an arc suppressor (PN
858728) should be used to minimize RFI.
3. Connect other end of output cable to
terminals of alarm device.
4. Replace the rear cover. Do not
overtighten cover retaining screws, to
avoid stripping the threads in the
plastic case.
NOTE:
For installation in a Class I, Division 2
location, refer to the Specifications in
the Preface section, and the approval
label on the analyzer for Factory Mutual limitations on contact ratings.
2-8 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 3
OPERATION
3-1 OVERVIEW
It is recommended that the operator read the
entire manual through before first operating
the instrument.
The operator can generally operate the Model
7003M using the flow chart in Figure 3-2 on
page 3-2 after the display symbols and modes
of operation are understood.
3-2 KEYPAD OPERATION
During each particular keystroke sequence,
the analyzer will recognize only certain keys.
If an unrecognized key is pressed, the
instrument will either not respond or will
display "SrY" ("Sorry") and return to RUN
Mode.
Generally, to escape from any routine, press
ENTER repeatedly until the instrument returns
to the RUN Mode and the oxygen
concentration is displayed.
Changing Parameters With The Keypad
All operation parameters are set with the
keypad. When the instrument is waiting for
operator input, the display will flash a digit. At
this point, if the number itself can be changed,
the operator can generally increase the
number (UP arrow) or decrease it (DOWN
arrow) at this time. An exception is when the
operator is restricted to pre-selected choices
(e.g., output current ranges, see Section 3-5b
on page 3-13).
When changing a value, an arrow key can be
pressed for a moment, and the display will
change by one digit. If it is held down, the
display will change, slowly at first, then
rapidly.
If the display has a digit flashing and will not
change, check the range of that parameter.
For example, an alarm setpoint cannot be set
higher than 25%.
When the number desired is displayed, press
ENTER. The Model 7003M will accept that
number and continue with the next step of that
routine.
Rosemount Analytical
ALARM1
ALARM2
%O2
°
C
MODEL 7003M
DISPLAYACK
PERCENT OXYGEN
Figure 3-1. Model 7003M Front Panel Controls and Indicators
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
ALARM1ALARM
ANALYZER
2
CALENTER
Operator Manual
748223-K
June 2002
Model 7003M
RUN MODE
Display Oxygen Temperature
SETUP MODE
CAL + ENTER
ENTER SECURITY
CODE
Set New
Security
Code
ENTERACK
Select
ACK
0-20 or 4-20mA
ENTER
Set Concentration
for Minimum
Output Current
ENTER
ENTER
ZERO GAS
ENTER
(0 or 4mA)
Set Concentration
for Maximum
Output Current
(20mA)
FLOW
Set Residual
Current Offset
ALARM MODE
ALARM1*
ENTER SECURITY
CODE
ENTER
ENTER
DISPLAYACK
DISPLAYACK
Disable
Alarm
ALARM 1*
Toggle
ON/OFF
Change "ON"
Setpoint
Change "OFF"
Setpoint
Acknowledge Alarm
CALIB. MODE
CAL
ENTER SECURITY
CODE
ENTER
Return to
RUN Mode
PRESSURE
COMPENSATION
ROUTINE
CAL
ENTER
ENTER
Reveal Security Code
HOLD MODE
ENTER SECURITY
Set Span
Concentration
DIAGNOSTICS
DISPLAY + ACK
Set Sample
Pressure
Set Span Gas
Pressure
ENTER
+
CODE
ENTER
Return to
RUN Mode
MODE
DISPLAY
Select Displayed
Parameter
ALARM 1*
Toggle
ON/OFF
Change
Current
Output
Return to
RUN Mode
Return to
RUN Mode
Return to
RUN Mode
Return to
RUN Mode
*Press ALARM 2 to set Alarm 2.
Figure 3-2. Model 7003M Operation Flow Diagram
3-2 OperationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
a. Single and Double Keystroke
Most procedures involve a sequence of
single keystrokes by which the operator
KEYFUNCTIONHOW TO USE IT
ACKAcknowledges an alarm conditionPress once to silence alarms
DISPLAY
ARROW keysIncrease/Decrease displayed value
CALAccess CALIBRATION mode
ENTEREnd command sequence
ALARM1Access ALARM 1 mode
ALARM2Access ALARM 2 mode
CAL + ENTERAccess SETUP mode
UP + DOWN ARROWAccess HOLD mode.
DISPLAY + ACKAccess DIAGNOSTICS mode
DISPLAY + ACKAccess security ID code
Switch between %O2 and °C
Press to display °C; press again to display O2.
Output signals are not affected.
Press once to change value by one digit. Hold key
down to change value quicker.
Press to calibrate. (If security is enabled, prompts
for ID code first)
Press to enter value or to end a command sequence
Press to check, set or disable Alarm 1 or ON/OFFControl relays
Press to check, set or disable Alarm 2 or ON/OFFControl relays
From RUN mode, press CAL and ENTER keys
simultaneously.
From RUN mode, press UP ARROW and DOWN
ARROW keys simultaneously.
From HOLD mode, press DISPLAY and ACK keys
simultaneously.
From RUN mode, press DISPLAY and ACK keys
simultaneously.
presses one key at a time. Some of the
more frequently used single-key
commands are listed in Table 3-1 below.
Table 3-1.Key Functions
INDICATORCARET/STATUSDESCRIPTION
%O2Caret ONOxygen percentage displayed
°C
ALARM1,2Caret OFFNormal condition
ALARM1,2Caret blinkingAlarm condition
-Caret ON
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 3-3
Caret ON
Sensor temperature displayed, °C
Displays current output in mA
(DIAGNOSTICS mode only)
Table 3-2.Display Indicators
Operator Manual
748223-K
June 2002
Model 7003M
b. Display Prompts
The Model 7003M prompts the operator
for inputs and also indicates the function
CODEDEFINITION
identer security code prompt
noinvalid security code entered - access denied
YESaccess granted or security code entered is correct
Table 3-3.Security Access Prompts and Responses
CODEDEFINITION
A-1setpoints for Alarm 1 - ALARM function
A-2setpoints for Alarm 2 - ALARM function
C-1setpoints for Alarm 1 - ON/OFF - Controller function
C-2setpoints for Alarm 2 - ON/OFF - Controller function
Onvalue in which to initiate an Alarm Action (On-setpoint)
OFFvalue in which to terminate an Alarm Action (Off-setpoint)
disAlarm is now disabled
Table 3-4.Alarm Settings Prompts and Responses
in progress. For example, when entering
the calibration mode, the prompt "CAL" is
displayed for one second after the CAL
key is pressed. Display acronyms are
listed in Table 3-3 through Table 3-9.
CODEDEFINITION
CALinitiate a span calibration
PPrenter process pressure value
CPrenter calibration pressure value
GASInitiate a zero offset calibration (gas)
Table 3-5.Calibration Prompts and Responses
CODEDEFINITION
Setenter system setup mode
4-0define the current output as 4-20 or 0-20 mA
CLo% concentration that 0/4 mA represents (calibration low)
CHi% concentration that 20 mA represents (calibration high)
Table 3-6.System Setup Prompts and Responses
3-4 OperationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
CODEDEFINITION
A-1toggle ON/OFF check of relay coil for Alarm 1
A-2toggle ON/OFF check of relay coil for Alarm 2
D-2performing non-intrusive test Internal RAM (16#00#..16#FF#)
D-3performing non-intrusive test on Internal RAM (16#4000#..16#5FFF)
D-4performing non-intrusive test on EEPROM (16#B600#..16#B7FF#)
D-5performing non-intrusive test on PROM (16#C000#..16#FFFF#)
F-2failed D-2 test
F-3failed D-3 test
F-4failed D-4 test
F-5failed D-5 test
hldentering system Hold
OFFalarm relay is de-energized (normally-closed)
Onalarm relay is energized (normally-open)
tStentering on-line test diagnostics mode
CODEDEFINITION
caret litdisplayed
O2sensor membrane voltage from ADC
Tempsensor thermistor voltage from DC
minus signmilliamps sent to current output
Operator Manual
748223-K
June 2002
Table 3-7.System Hold and Diagnostics Codes
Table 3-8.Data Displayed in DIAGNOSTICS Mode
CODEDEFINITION
Sryinvalid keypad function (Sorry)
runcurrent function completed - entering RUN Mode
Table 3-9.Miscellaneous Prompts and Responses
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Operator Manual
748223-K
June 2002
Model 7003M
c. Error Messages
If an error condition occurs, E-# (non-fatal
error) or F-# (fatal error) will flash on the
Table 3-10. Power ON, Reset Diagnostics Errors (Fatal Errors
CODEDEFINITION
E-15 volts power supply above 5.5 V
E-25 volts power supply below 4.5 V
E-312 volts power supply above 13.2 V
E-412 volts power supply below 10.8 V
E-5sensor output too low for calibration
E-6sensor missing
E-8
E-9
sample temperature is below 1°C
sample temperature is above 44°C
display. Table 3-10 through Table 3-13
list error codes and corrective actions.
Pressing "ACK" or "ENTER" will
acknowledge any alarm and the error
message will disappear.
iotInvalid Op code Trap (internal computer command is invalid)
E-0power interruption occurred
sample temperature is above 50°C
sample temperature is below 0°C
Table 3-12. Signal Board Errors
Table 3-13. Microprocessor Errors That Reset The System
3-6 OperationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
d. Automatic Return to RUN Mode
If no keys are pressed for five minutes,
the instrument will automatically return to
the RUN Mode, except when in Zero
Offset, HOLD, or DIAGNOSTICS Modes.
Operator Manual
748223-K
June 2002
After the security code is entered, the
operator can change any instrument
setting and can go back and forth to the
RUN Mode without having to enter the
code again.
CAUTION
OXYGEN WILL NOT BE MONITORED
The instrument will remain in HOLD or DIAGNOSTICS Modes until the operator returns it to the RUN Mode or other
operation.
e. Preventing Unauthorized Access
The Model 7003M is equipped with a
security feature to prevent unauthorized
tampering with the instrument settings.
The instrument is shipped from the factory
with the security feature disabled. It can
be enabled or disabled at any time.
Security access is not required to:
• Display sensor data
• Acknowledge alarms
• Reveal security code
Security access is required to:
If security is enabled, and the instrument
is left in the RUN Mode for five minutes,
security will automatically be re-activated.
Within this five minute period, anyone can
access the functions without re-entering
the security code.
f. Sensor Equilibrium
When a new sensor is installed, power up
the instrument and let the sensor stabilize
for about 5 hours before calibrating it.
If the oxygen sensor is calibrated at a
relatively high oxygen concentration, such
as with ambient air (typically 20.95%) and
then is exposed to a low oxygen level, a
longer equilibrium time period may be
required for the sensor to equilibrate.
This time period depends on the
monitoring concentration level and how
long the sensor was operated at the
higher level.
• Calibrate with span gas
• Access options and diagnostics
• Change operation parameters
• Enable or disable alarms and
ON/OFF-Control
• Disable or change the security code
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Operator Manual
748223-K
June 2002
Model 7003M
3-3 START-UP
Before proceeding, the operator should be
familiar with the keypad and indicators
(Section 3-2 on page 3-1).
Operational checks and adjustments are not
normally required, provided that the sensor,
inter-connection cable and analyzer have
been properly installed. The analyzer
continually performs self-checking routines
during operation.
Calibration with ambient air is recommended
after start-up and at regular intervals (see
Section 4-2 on page 4-1)
The following sections provide instructions for
initial start-up and subsequent standardization
of the analyzer.
a. Initial Power-Up
1. Ensure that the proper sensor and
sensor cable are connected to the
analyzer.
2. Apply power to the analyzer.
The following internal tests are
performed:
ROM "Checksum" test
EEPROM Read/write test
3. If an error code flashes, refer to the
error code lists in Table 3-10 through
Table 3-13. Press ACK to continue
with the startup procedure.
After internal tests are passed, the
software version is displayed for a
few seconds, and the analyzer goes
directly into RUN mode. "RUN" is
briefly displayed, then current
oxygen level.
Error codes E-0 (power failure) and
E-5 (not calibrated) typically occur
upon power-up.
4. Monitor the oxygen output or display
until the level stabilizes (typically 5
hours). The alarms and current
output levels can be set immediately if
desired.
5. Once the sensor has stabilized, the
analyzer is operational and is ready
for subsequent operations. Refer to
Table 3-14 below for the default
settings present upon power-up.
RAM Read/write test
Mode:Alarm
Alarm 1
Range
Current Output Range4-20 mA
SecurityDisabled
Pressure
Residual Zero Offset0
Calibration StatusNot Calibrated
3-8 OperationRosemount Analytical Inc. A Division of Emerson Process Management
State:Disabled
Setpoints:On: 0.0 Off: 0.1%
Mode:Alarm
State:DisabledAlarm 2
Setpoints:On: 0.0 Off: 0.1%
0% of scale:0% Oxygen
100% of scale:25% Oxygen
Calibration:76.0 cm Hg
Process:76.0 cm Hg
Table 3-14. Default Settings at Power-Up
Model 7003M
Operator Manual
748223-K
June 2002
b. System Restart
The Model 7003M will attempt to restart
itself in the event of a power interruption,
power surge or other outside interference
to microprocessor operation (Section 3-3c
below).
The analyzer can also be manually
restarted if the RAM or EEPROM
becomes corrupted (Section 3-3d below).
c. Automatic Restart - Due to Power
Failure
If a power failure or other event occurs
that affects operation of the
microprocessor the system will attempt to
re-start itself. If it is successful, the
instrument will continue to operate
normally, except that the appropriate error
code will be displayed until the ACK
button in pressed. See Section 5-3c on
page 5-2 for circuit description.
d. Manual Restart
1. Turn power to analyzer OFF.
2. Wait 30 seconds, and turn the power
to the analyzer ON.
3. While the software revision is being
displayed and the self-tests are being
performed, press DISPLAY and ACK
simultaneously.
"dYn" is displayed.
4. Press ENTER to clear the RAM and
EEPROM.
The display will go blank, then continue
with the SETUP routine.
Continue with Section 3-5a on page 3-13,
step 2.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Operator Manual
748223-K
June 2002
Model 7003M
3-4 RUN MODE
Refer to Figure 3-2 on page 3-2 and Table
3-15 below. The RUN mode is the normal
mode of operation, from which the operator
can:
KEYDISPLAYWHAT HAPPENS
DISPLAYNumber changeToggles between O2 reading and temperature
ACKClears messageAcknowledges Alarms
ALARM 1"A-1" then ONTo Alarm 1 SETUP mode
ALARM 2"A-2" then ONTo Alarm 2 SETUP mode
UP ARROW"SrY" "run"Stays in RUN mode
DOWN ARROW"SrY" "run"Stays in RUN mode
CAL"CAL" then ONTo CALIBRATION mode
ENTER"SrY" "run"Stays in RUN mode
CAL + ENTER"SEt" then ONMoves to SETUP mode
UP + DOWN"hLd" then ONMoves to HOLD mode
DISPLAY + ACK"123"Shows the security code
• Display oxygen concentration and sensor
temperature
• Acknowledge alarms
• Display security code
• Access other modes
Table 3-15. RUN Mode Key Functions
a. Display Oxygen Concentration/Sensor
Temperature
Press DISPLAY once to display sensor
temperature. Press DISPLAY again to
display oxygen concentration.
b. Acknowledge Alarms
If an alarm setpoint has been reached, "A1" or "A-2" will flash with a concentration
value. Acknowledging the alarm will stop
the flashing display and turn off the relay
switch.
To Acknowledge Alarm - Press ACK
key. If relays are configured as ON/OFF
controllers, pressing the ACK key has no
effect.
3-10 OperationRosemount Analytical Inc. A Division of Emerson Process Management
c. Access Other Modes
The following lists which key(s) to press to
access various modes:
SETUP= CAL+ENTER
ALARM= ALARM1 (or ALARM2)
CALIBRATION = CAL
HOLD= UP and DOWN ARROWS
DIAGNOSTICS = DISPLAY+ACK (from HOLD)
Model 7003M
Operator Manual
748223-K
June 2002
3-5 SETUP MODE
Refer to Figure 3-3 on page 3-12 and Table
3-16 below. From the SET-UP mode the
operator can:
KEYDISPLAYWHAT HAPPENS
DISPLAYNothingNothing
ACK"<4>-0" / "4-<0>Selects 4-20 or 0-20 mA operation
ALARM 1NothingNothing
ALARM 2NothingNothing
UP ARROW
DOWN ARROW
10.8 →10.9
10.9 →10.8
CALNothingNothing
ENTERvarious to "run"Moves to the next step or goes to RUN mode
Table 3-16. SETUP Mode Key Functions
• Change security code
•
Set output current range (0-20mA or 420mA)
•
Set output concentration range endpoints
•
Adjust zero offset
Changes the number displayed
Changes the number displayed
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11
Operator Manual
748223-K
June 2002
SET-UP Mode
CALENTER
+
Model 7003M
Flashing
Flashing
May not be displayed
each time.
Flashing
May not be displayed
each time.
Flashing
Flashing
ALARM1
ALARM2
%O2
C
°
SECURITY ACCESS ROUTINE
(see Figure 3-4)
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
Prompts for Security Code.
If left at 000, code will be disabled.
Use Arrow keys to set Security Code.
Select either 4-20mA or
ACK
0-20mA Output
ENTER
ENTER
ALARM1
ALARM2
%O2
C
°
Prompts for low end output.
(i.e., 0% O2 would be the low range
ENTER
representing O2 level of 0 or 4mA.
Flash for 2 seconds
(Hi range output selection)
Flash for 2 seconds
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
%O2
C
°
%O2
C
°
%O2
C
°
Prompts for zero offset
%O2
C
°
%O2
C
°
Set Span output level = 20mA
Set Span output level = 20mA
ENTER
ENTER
Figure 3-3. SETUP Mode Flowchart
3-12 OperationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
a. Security Code
Refer to Figure 3-4 on page 3-14.
Reveal Current Security Code
From RUN mode, hold down DISPLAY
and ACK keys simultaneously.
Change Current Security Code
1. From RUN mode, press CAL and
Enter simultaneously.
2. SET, id are briefly displayed. Then
"000" is displayed with the right digit
flashing.
NOTE
If RUN and CAL are not press exactly
at the same time, "SRY", or "RUN" will
appear, and the analyzer will return to
RUN mode.
3. If the security code has been
disabled, skip to step 4. Use the
ARROW keys to enter the current
security code, then press ENTER.
1. Press ACK until the desired range is
flashing, press ENTER.
2. Continue with the next SET-UP
routine, or hold ENTER key to return
instrument to RUN mode.
c. Range Endpoint Adjustments
The oxygen values corresponding to
minimum and maximum output current
must be set. Default is 0% for 0 or 4mA
and 25% for 20mA. Endpoints must be
within the range of the instrument and
must be at least 1% oxygen apart.
Set Oxygen Low Concentration Endpoint
Following the last procedure, "CLo" is
displayed briefly, followed by the oxygen
concentration last stored to correspond to
minimum output current.
NOTE
If "CLo" is set to a value greater than
"CHi", "CHi" will be automatically reset
to a value 1% greater than "CLo".
4. If the correct code is entered, YES,
SET, id will be displayed briefly, the
code will flash. If the incorrect code is
entered the analyzer will return to
RUN mode.
5. Use the ARROW keys to set the new
security code (or "000" to disable),
press enter.
6. Continue with the next set-up routine,
or hold down ENTER until "run"
appears (the instrument is back in
RUN mode).
b. Current Output Range - 0-20mA or 4-
20mA
Following the last procedure, "4-0" is
displayed with either digit flashing. "4"
indicated 4-20mA (default), and "0"
indicates 0-20mA.
Use the ARROW keys to set the low
oxygen concentration (0 to 24%), press
ENTER.
"CHi" is displayed briefly, followed by the
latest oxygen concentration stored
corresponding to maximum output
current.
Set Oxygen High Concentration Endpoint
"CHi" cannot be set to a value equal to or
less than "CLo".
Use the ARROW keys to set the high
oxygen concentration (1 to 25%), press
ENTER.
Continue with next SET-UP routine or
hold ENTER to return to RUN mode.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-13
Operator Manual
748223-K
June 2002
Model 7003M
RUN
SET-UP Mode
CALENTER
+
ALARMCAL
WAS SECURITY ID
ENTERED <5 MIN.
AGO
NO
ENTER ID
ENTER
ID
CORRECT
?
HOLD Mode
+
YES
OPERATOR GIVEN
ACCESS TO SE-
LECTED MODE
YES
NO
ACCESS DENIED
Returns to
RUN mode
Figure 3-4. Security Access Routine Flowchart
3-14 OperationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
d. Zero Offset Adjustment
Following the last procedure, the current
zero offset value will be flashing.
NOTE
When flowing zero gas over the sensor, ensure that there are no leaks and
the gas actually has zero oxygen concentration. Otherwise, significant errors will result. Zero offset adjustment
is generally not required. Errors are
more often due to incorrect zero gas
oxygen concentration than residual
current of the oxygen sensor.
The zero offset adjustment allows the
operator to compensate for any residual
sensor current. Adjustment range is 0-5%
equivalent oxygen.
Sensor Current = Current from
Oxygen + Residual Current
NOTE
Adjusting the zero offset affects the
output of the Model 7003M. Therefore,
if the zero offset is changed, the system must be re-calibrated with span
gas.
Zero Offset Is Not To Be Changed
Press ENTER to return analyzer to RUN
mode, and skip the remainder of this
procedure.
Zero Offset Is To Be Removed
1. Press the UP ARROW until the
maximum value is displayed.
2. Press ENTER and re-calibrate with
span gas per Section 4-2b on page 4-
1.
Zero Offset To Be Adjusted
1. Expose the sensor to 100% oxygenfree nitrogen or other inert gas and
allow the sensor signal to stabilize (25 hours depending on sample
conditions).
2. Use the DOWN ARROW to bring the
displayed value to zero (for maximum
offset).
3. Press ENTER to return to RUN mode.
4. Re-calibrate with span gas per
Section 4-2b on page 4-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-15
Operator Manual
748223-K
June 2002
Model 7003M
3-6 ALARM MODE VS. ON/OFF-CONTROL
MODE
Refer to Figure 3-5 on page 3-17 and Table
3-17 below. From the ALARM or ON/OFFControl Mode, the operator can:
KEYDISPLAYWHAT HAPPENS
DISPLAYNothingNothing
ACK"dIS"Disables alarms
ALARM 1"C-1" / "A-1"Toggles between alarm and controller mode
ALARM 2"C-2" / "A-2"Toggles between alarm and controller mode
UP ARROW
DOWN ARROW
CALNothingNothing
ENTER"OFF" / "run"Moves to the "OFF" setpoint step or goes back to RUN mode
a. Relay Configured as Alarm
Relays are configured as alarms (ALARM
Mode) when immediate operator action is
required.
When alarms are selected, "A-1" and "A2" are displayed in the ALARM mode setup procedure.
10.8 →10.9
10.9 →10.8
Table 3-17. ALARM Mode Key Functions
• Set alarm ON and OFF setpoints.
• Select alarm or ON/OFF-Control function
for relays
• Disable alarm(s)
Changes alarm setpoint
Changes alarm setpoint
b. Relay Configured as ON/OFF-
Controller
Relays are configured as ON/OFFControllers (ON/OFF-Controller mode)
when a non-critical, non-hazardous
automated action is desired, or when it is
important that the relay output not be
bypassed by the operator.
When an alarm condition occurs, the
caret blinks, the relay activates, and the
display flashes. Press the ACK key to
acknowledge the alarm, de-energize the
relay coil and stop the alarm actions.
3-16 OperationRosemount Analytical Inc. A Division of Emerson Process Management
When ON/OFF-Controllers are selected,
"C-1" and/or "C-2" is displayed in the
ALARM mode set-up procedure.
When an ON/OFF-Controller alarm
condition occurs, Alarm 1 or 2 caret blinks
and the relay coil is energized. However,
the display does not flash, and the ACK
key is inactive. Alarm conditions are
automatically reset when the sensor
indicates that an alarm condition no
longer exists.
Model 7003M
(
)
Operator Manual
748223-K
June 2002
ALARM/CONTROLLER MODE
To Disable Alarm
Indicates Alarm
is disabled
Flashes Alarm
Configuration
ALARM1
ALARM2
Flashing
ALARM1
ALARM2
Prompts "ON" Setpoint
Displays "ON" Setpoint
ACK
1
%O2
C
°
ALARM
ALARM1
ALARM2
Note: Alarms are disabled when instrument is received.
Alarm 2 uses identical sequence.
1
%O2
C
°
SECURITY ACCESS ROUTINE
see Figure 3-4
ALARM1
ALARM2
%O2
C
°
ALARM1
ALARM2
ALARM1
ALARM2
OR
%O2
C
°
ALARM1
ALARM2
%O2
C
°
%O2
C
°
STANDARD ALARM
Adjust Alarm ON Setpoint
ENTER
1
%O2
C
°
Flashes Control
Configuration
ALARM
1
Change Relay
configuration
ALARM1
ALARM2
%O2
C
°
ALARM1
ALARM2
ALARM1
ALARM2
%O2
Prompts for OFF Setpoint
C
°
%O2
C
°
Adjust Alarm OFF Setpoint
ENTER
ALARM1
ALARM2
New values/functions will not be stored until instrument returns to RUN mode with relay activated.
1
%O2
C
°
Figure 3-5. ALARM Mode Flowchart
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-17
Operator Manual
748223-K
June 2002
c. Relay Contacts
Relay contacts are listed in Table 3-18 below:
Model 7003M
RELAY POS
TERMINAL LOCATIONS
ALARM1ALARM2
NO (Normally open)TB3-3TB3-6
NC (Normally closed)TB3-2TB3-5
COM (Common)TB3-1TB3-4
The relays are setup to operated in "Fail
Safe" configuration. In this configuration
the relay coils are powered when the unit
is NOT in an alarm condition. When an
alarm condition occurs, power is removed
from the relay coils.
Wherever "Alarm" is referred to in this
section, the term "ON/OFF-Controller" can
be substituted, unless specifically called
out. Also, any reference to A1 or Alarm 1
also applies to Alarm 2, and to ON/OFF-
Controllers 1 and 2.
If a power failure occurs, power will be
removed from the relays and an alarm
h. Setting Alarm Setpoints
condition will be indicated to the device
connected to the relays.
This procedure can be used to:
Relays
e. Relay Reset Upon Fatal Error
If a fatal error is detected, the unit will
•
Verify alarm relay setpoints
•
Change setpoints
attempt to reset itself which will make
both relays go into alarm mode. If the
•
Enable alarms
microprocessor is able to continue its
operation, the relays will then be reset to
•
Disable alarms
their normal operating position.
Perform the following procedure for each
f. Pressure Effects on Alarms
Alarm setpoint, as required. Figure 3-6
illustrates how relays are energized and
If the sample pressure changes, the
partial pressure of oxygen, and thus the
de-energized in High and Low Alarm
configurations.
reading, will change and the alarm relays
will not actuate at the same oxygen levels
as they were set when the interment was
last calibrated. See Section 5-2a on page
5-1 for a more detailed description of
pressure effects.
3-18 OperationRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
i. Setting Alarm ON and OFF Levels
Refer to flowchart Figure 3-5 on page 3-
17.
From RUN mode, press ALARM 1.
A-1, YES, A-1 (or C-1), ON flash briefly;
the alarm ON setpoint is displayed with
the right digit blinking.
A-1 indicates relay 1 acts as an ALARM;
C-1 indicates relay one acts as an
ON/OFF-Control.
To change the relay action from
ALARM to ON/OFF Control function (or
vise versa) press ALARM 1. C-1 or A-1
will flash indicating a new mode.
To disable ALARM 1 - Press ACK.
"DIS" will flash briefly and return to RUN
mode.
To set the ON setpoint for ALARM 1 -
Use the ARROW keys, press ENTER.
OFF is flashed briefly, then the alarm OFF
setpoint is displayed with the right digit
flashing.
To set the OFF setpoint for ALARM 1 -
Use the ARROW keys, press ENTER.
The instrument displays "run" and returns
to RUN mode.
HIGH Alarm
set for 8.24% O
LOW Alarm
set for 8.10% O
Figure 3-6. Action of Alarm Relay ON and OFF Setpoints
2
8.24%
8.10%
2
8.24%
8.10%
ACTIVATESDE-ACTIVATES
ALARM ON
ON
ACTIVATESDE-ACTIVATES
ON
SETPOINT
OFF SETPOINT
OFF
TIME
ALARM ON
OFF
OFF SETPOINT
SETPOINT
TIME
DEADBAND
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-19
Operator Manual
748223-K
June 2002
Model 7003M
3-20 OperationRosemount Analytical Inc. A Division of Emerson Process Management
SECTION 4
CALIBRATION AND ADJUSTMENTS
Operator Manual
748223-K
June 2002
4-1 FREQUENCY OF CALIBRATION
When a new sensor is installed, adjust the
zero offset current (Section 3-5d on page 3-
15) and calibrate the instrument with span
gas (4-2b on page 4-1. Span the
instrument weekly at first and note the
stability. Use this information to determine
the regular calibration frequency required
by this particular application.
KEYDISPLAYWHAT HAPPENS
DISPLAYNothingNothing
ACKNothingNothing
ALARM 1NothingNothing
ALARM 2NothingNothing
UP ARROW
DOWN ARROW
10.8 →10.9
10.9 →10.8
CAL"PPr"Moves to pressure compensation mode
ENTER"run"Drops back to RUN mode
Table 4-1.CALIBRATION Mode Key Functions
a. Zero Calibration
Each oxygen sensor produces a
constant residual electrical current in
the absence of oxygen due to
characteristics of the sensor. Typically
this current is less that 0.1% apparent
oxygen. It is different for each sensor,
but will remain constant for the life of
that sensor. The instrument can be
zeroed by adjusting the residual current
per Section 3-5d on page 3-15 After
this procedure is performed once, the
zero should not need adjusting again
during the life of the sensor.
b. Span Calibration
4-2 CALIBRATION MODE
Refer to Figure 4-1 on page 4-3and Table
4-1 below. From the CALIBRATION Mode,
the operator can:
2. Press CAL to enter the
CALIBRATION mode. Enter the
security code if required.
3. CAL flashes and oxygen
concentration is displayed with the
right digit flashing. Output is held
at last reading until returned to
RUN mode.
4. Use the ARROW keys to set the
instrument to the concentration of
the calibration gas. See Table 4-2
on page 4-2 and Table 4-3 on page
4-2 if using ambient air as span
gas.
1. Expose the sensor to ambient air
and let the oxygen reading and
sensor temperature stabilize.
1
We recommend using ambient air in calibration, re-
gardless of the range. Only when monitoring at 0 to
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-1
1
1% range, can the accuracy of the instrument be
slightly improved by using a gas of lower concentration (e.g., 1%) for calibration. If bottled air is used,
verify the oxygen concentration against National Bureau of Standards certified calibration gas. The true
oxygen concentration in cylinders is often different
from the concentration indicated on the cylinder label.
Operator Manual
748223-K
June 2002
Model 7003M
5. Press ENTER to return to the RUN
mode.
7. If the sample gas and the
calibration gas are each at constant
pressure, continue with Section 4-
6. If the ratio between the
concentration value entered and
2c on page 4-4 and enter the
pressure of each.
the sensor voltage is excessive, an
error message will be flashed.
Refer to Figure 4-1 on page 4-3 and
Table 4-5 above. If the sample and
calibration gases are each at constant
pressure, the following procedure should
be used to correct for the pressure
difference between them. However, the
instrument does not monitor pressure and
will not correct the output if either sample
or calibration gas pressure changes.
(See Section 5-2 on page 5-1).
1. From RUN mode, press CAL.
2. CAL, YES, CAL flash briefly, then the
oxygen concentration is displayed
with the right digit flashing.
3. Press CAL again.
1
Computed using 20.946% oxygen at 760 m.
4. PPr flashes, then previous sample
pressure setting is displayed with right
digit flashing (76.0 cm Hg is default).
5. Use ARROW keys to enter sample
pressure.
6. Press ENTER to accept the value.
7. CPr flashes, then previous span gas
pressure setting is displayed with right
digit flashing (76.0 cm Hg is default).
8. Use the ARROW keys to enter
pressure of the calibration gas to be
used. If the sample and calibration
gas will always be at the same
pressure, set PPr and CPr at the
same value.
9. Press ENTER to accept the value.
Instrument returns to RUN mode.
4-4 Calibration and AdjustmentsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
4-3 HOLD MODE
•
The last measured value is displayed
Refer to Figure 4-2 on page 4-7 and Table 4-6
•
below. In the HOLD mode:
•
All instrument functions are locked
•
All existing values are retained
•
The instrument stops responding to
sensor signals
KEYDISPLAYWHAT HAPPENS
DISPLAYNothingNothing
ACKNothingNothing
ALARM 1NothingNothing
ALARM 2NothingNothing
UP ARROWNothingNothing
DOWN ARROWNothingNothing
CALNothingNothing
ENTER"run"Drops back to RUN mode
DISPLAY + ACK"d-4" "tSt"Moves to DIAGNOSTIC mode
The last output value remains at the
isolated current output connections
•
Alarms and ON/OFF-Controllers are held
in their last state
•
No keys are recognized except entry to
the diagnostic routing.
Operator Manual
748223-K
June 2002
Table 4-6.HOLD Mode Key Functions
If the sensor cable must be disconnected, the
instrument can be put in the HOLD mode, in
which all existing signal levels and output
current levels and output current levels are
maintained. This may be useful to prevent
disturbing an ON/OFF-control loop connected
to the relay contacts.
1. Press the UP ARROW and DOWN
ARROW at the same time to put the
instrument on HOLD.
2. If the security is activated, enter the
security identification (id) code and press
ENTER to continue.
3. Hld, YES, flash briefly, then Hld flashes
alternately with the last oxygen reading.
The output signal will remain unchanged
until the instrument returns to the RUN
mode.
4. Press ENTER to return to the RUN mode,
or continue with Section 4-4 Diagnostics
Mode on page 4-8.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-5
Operator Manual
748223-K
June 2002
Model 7003M
CARET
LIT
VALUE DISPLAYED
%Raw Sensor Voltage
°C
-
3
Thermistor Voltage2.43No
Current Output17.43Yes
Use ARROW keys to make adjustments.
Table 4-7.Caret Readings
TYPICAL
VALUE
2
1
ADJUSTABLE
Yes
1
These typical values were obtained under the following conditions:
O2 Concentration:20.95%
Sensor Temperature:26.07°C
Current Output Range:4-20mA
Range Endpoint, Low:0%
Range Endpoint, High:25%
Zero Offset Signal:None
Pressure Compensation:None
Near the end of the life of the sensor, the sensor voltage is typically 0.53V (disposable) or 0.48V (rechargeable). At tempera-
tures (T) other than 25°C, multiply the voltages above by (1.025)exp(T-25).
3
"-" is a minus sign displayed on the left of the display between AL 1 and AL 2.
4-6 Calibration and AdjustmentsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
(
)
(
)
g
Operator Manual
748223-K
June 2002
HOLD MODE
+
SECURITY ACCESS ROUTINE
see Figure 3-4
ALARM1
ALARM2
Last displayed
value held
ENTER
Sensor Voltage
Display Parameters
ALARM1
ALARM2
%O2
C
°
DIAGNOSTICS MODE
+
%O2
C
°
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
CARET LIT
see Table 4-7
DISPLAY
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
Adjust Parameters (see Section 4-4)
WARNING
Testin
Relays will affect devices connected to them
Exit
ALARM1
ALARM2
ALARM1
ALARM2
ALARM
1
ALARM
1
%O2
°
%O2
°
C
C
Return to
HOLD
ENTER
DIAGNOSTIC
Mode
ALARM1
ALARM2
ALARM1
ALARM2
ALARM
2
ALARM
2
%O2
C
°
%O2
C
°
Figure 4-2. HOLD and DIAGNOSTICS Modes Flowchart
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-7
Operator Manual
748223-K
June 2002
Model 7003M
4-4 DIAGNOSTICS MODE
Refer to Figure 4-2 on page 4-7, Table 4-7 on
page 4-6 and Table 4-8 below. The
DIAGNOSTICS mode is used to:
• Check sensor circuit voltages
• Check relays
• Adjust the output current
Enter DIAGNOSTICS Mode
If the instrument is already in HOLD mode,
skip step 1 and proceed to step 2.
1. From the RUN mode, press the UP
ARROW and DOWN ARROW at the
same time to enter the HOLD mode.
2. If the security is activated, enter the
security identification (id) code and press
ENTER to continue.
Hld, YES, flash briefly, then Hld flashes
alternately with the last oxygen reading.
The output signal will remain unchanged
until the instrument returns to the RUN
mode.
a. Raw Sensor Voltage
1. Press DISPLAY and ACK at the same
time to enter DIAGNOSTICS mode.
Tst, d-3, d-4 flash briefly, then the
%O
caret lights and the raw sensor
2
voltage is displayed.
2. Press ENTER at any time to return to
HOLD mode. Press ENTER again to
return to RUN mode.
KEYDISPLAYWHAT HAPPENS
DISPLAY"2.13"Scrolls among the DIAGNOSTIC displays
ACKNothingNothing
ALARM 1"A-1" "On"/ "OFF"Toggles the Alarm 1 relay ON/OFF
ALARM 2"A-2" "On" / "OFF"Toggles the Alarm 2 relay ON/OFF
UP ARROW
DOWN ARROW
CALNothingNothing
ENTER"hLd"Drops back to HOLD mode
b. Test Current Output
The current output can be tested
(increased or decreased using the
ARROW keys) and the actual output will
change. When the instrument returns to
the RUN mode, the output will return to
the previous value.
1. Press DISPLAY until the minus sign () is displayed on left.
"2.13" →"2.14"
"2.14" →"2.13"
Table 4-8.DIAGNOSTICS Mode Key Functions
Increases current output or span setting
Decreases current output or span setting
c. Linearizing Current Output
For some samples, current output does
not track the oxygen reading displayed.
The operator can linearize output and
store the compensating second-order
polynomials in EEPROM. This data will
remain and function there as long as the
EEPROM is not erased. However, if a
"cold boot" is performed, the operator
must repeat this linearization routine.
2. Use the ARROW keys to increase or
decrease the current output.
4-8 Calibration and AdjustmentsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
To determine if current output
linearization is necessary, do the
following:
1. Connect a separate current meter to
the output terminals (any device that
will display an accurate current
measurement such as a computer
input will suffice).
NOTE
Do not simply connect a current
measurement meter in parallel with the
usual load because this will lead to incorrect readings.
2. Place the instrument in DIAGNOSTIC
mode by pressing the UP and DOWN
ARROWS, followed by the DISPLAY
and ACK keys. The display will show
consecutively:
tSt, d-3, d-4, and oxygen input
voltage
3. Press the DISPLAY key until the
minus sign and a number are
displayed. This reading is the current
output (in milliamps) as determined by
the Model 7003M.
8. Use the UP and DOWN ARROW
keys to adjust the actual current
output (as noted on the separate
meter) to the correct amount. Note
that the instrument display does not
change.
9. Press ENTER, the instrument
displays diA, returns to the
DIAGNOSTIC mode, and continues to
display the current output.
10. Use the UP and DOWN ARROW
keys to adjust the current output to a
lower value, representing the low
range (the operator may prefer to use
4 mA to represent 0% oxygen).
11. Press CAL (the instrument displays
CCL for lower current output setting),
and repeat steps 3 and 4 of this
procedure.
12. Use the UP and DOWN ARROW
keys to adjust the current output to a
value higher than the initial calibration
value (step 1 of this procedure),
representing the higher range (the
operator may prefer to set it at
16.72mA to represent 20.9% oxygen.
4. Use the UP and DOWN ARROW
keys to increase the display reading,
and occasionally, compare the current
displayed on both meters (the
instrument display and the separate
meter).
5. If the two readings are too disparate
(different applications require lesser
or greater tolerance), do the following
3-point calibration to linearize the
current output:
6. Use the UP and DOWN ARROW
keys to set the current output to a
value halfway up its range. The
operator may prefer to set it at a
critical measurement point.
7. Press the CAL key. The instrument
will flash CCA, illuminate the three
right-hand carets, and continue
displaying the current output.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-9
13. Press CAL (the instrument displays
CCH for high current output setting),
and repeat steps 3 and 4 of this
procedure.
To return to the RUN mode, press
ENTER twice.
At any time during this procedure, the
operator may press the ACK key to reset
all calibration coefficients to default
values. Pressing DISPLAY at any time
during this procedure will cause the
instrument to display for a moment the
relevant calibration efficient.
The operator may escape the current
CALIBRATION and DIAGNOSTIC mode
altogether by pressing ENTER twice.
Note, all coefficients will return to the
default values, and the current calibration
procedure must be started over to
linearize current output.
Operator Manual
748223-K
June 2002
Model 7003M
NOTE
The low calibration point must be between the initial calibration point (step
1 of previous procedure) and zero, and
the high calibration point must be between the initial calibration point and
20mA.
If the third point calibrated (step 8,
previous procedure) is actually another low range (between the initial
calibration point and CCL) it will override the original CCL but won't affect
the initial calibration.
NOTE
The current output linearizing procedure uses second order polynomials to
adjust the output linearity. The correction effect is maximized at the calibration point, and is smoothly diminished
towards the endpoints.
d. Test Alarm Relays
1. Disconnect any alarm or ON/OFFControl devices that are connected to
the analyzer, before testing.
2. Toggle the relays ON and OFF using
the ALARM 1 and ALARM 2 keys.
A-1 ON or A-2 OFF is displayed.
The alarm contacts can be heard
opening and closing.
3. Connect any alarm or ON/OFFControl devices that were
disconnected in step 1.
4. Press ENTER to exit the
DIAGNOSTICS mode.
"Hld" flashes alternately with the last
oxygen reading and the analyzer
returns to HOLD mode.
5. Press ENTER again to return to RUN
mode.
4-5 ADJUSTMENTS AND SETTINGS
All operator settings and adjustments are
made using the front keypad except if the
power jumpers are to be changed. Refer to
the following sections:
2-5b Power Connections on page 2-4
NOTE
When alarms are tested manually, any
alarm or ON/OFF-Control devices connected to the relays will respond as
they would in an alarm condition. The
alarm devices may have to be disconnected from the relays for this test.
4-10 Calibration and AdjustmentsRosemount Analytical Inc. A Division of Emerson Process Management
3-5d Zero Offset Adjustment on page 3-15
4-4a Raw Sensor Voltage on page 4-8
4-4c Linearizing Current Output on page 4-8
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 5
THEORY
5-1 PRINCIPLES OF OPERATION
The Model 7003M Percent Oxygen Analyzer
consists of an amperometric sensor and
analyzer interconnected by a multi-conductor
shielded cable. The sensor responds to the
partial pressure of oxygen and produces a
current signal. The analyzer amplifies and
processes the signal, providing readout in
percent oxygen and current output.
a. Electrochemical Theory
With the sensor placed in the sample gas,
a voltage is applied across the cathode
and anode. Sample gas diffuses through
the membrane and is reduced at the
cathode resulting in a current proportional
to the partial pressure of oxygen in the
sample.
When no oxygen is present, a small
residual sensor current, which can be
offset during instrument setup.
5-2 EFFECTS OF SAMPLE CONDITIONS
a. Barometric Pressure
If ambient air is used to calibrate the
sensor, a change in barometric pressure
can be a significant factor. At 76 cm Hg,
the partial pressure of oxygen is 15.9 cm
Hg (20.9% of 76 cm Hg). As atmospheric
pressure deviates from 76 cm Hg, the
partial pressure of oxygen will vary
proportionally. The highest partial
pressure that can be read by the Model
7003M is 19 cm Hg (25% oxygen at 76
cm Hg).
oxygen without affecting the total
barometric pressure.
This relationship can be expressed as :
P atm = Po2 + PH20 = P other,
where:
P atm = total barometric pressure,
Po2 = partial pressure of oxygen,
PH20 = partial pressure of water
vapor,
P other = partial pressure of all
gases other than oxygen and water
vapor
c. Sample Temperature
The rate of oxygen diffusion through the
sensor membrane varies with
temperature, with a coefficient of about
+3% °C (5% °F), causing a corresponding
change in sensor current. The instrument
continuously monitors temperature with a
thermistor in the sensor body, and
compensates the output for both sample
temperature and membrane permeability.
Because of a slight time lag in the
response of the temperature sensor, a
rapid temperature change will cause a
temporary change in apparent oxygen
level, such as might be encountered if the
sensor is suddenly exposed to direct
sunlight from a shadow.
b. Humidity
If calibrating with humid ambient air, the
effect of the partial pressure due to water
vapor content should be taken into
consideration. High humidity has the
effect of reducing the partial pressure of
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 5-1
Extended use of the sensor at low
temperatures will shorten the life of the
sensor. As the sensor reaches the end of
its useful life, distinct step changes in the
oxygen reading will be evident as the
oxygen level changes.
Operator Manual
748223-K
June 2002
d. Interfering Gases
Gases that are reduced or oxidized in the
range of 0.5 to 0.8VDC contribute to
sensor current and can cause a readout
error. Only a few gases have this
characteristic. Common gases that
should be avoided include SO
oxides of nitrogen (NO
concentrations of hydrogen sulfide tend to
contaminate the sensor, but may not
seriously affect oxygen measurement.
If you suspect that any stream
components are affecting the operation of
the sensor, contact Rosemount Analytical
Technical Services.
If contaminated, the rechargeable sensor
must be rejuvenated and the disposable
sensor must be replaced. Refer to the
instructions supplied with the sensor.
5-3 CIRCUIT DESCRIPTIONS
a. Signal Board
The Signal Board is configured to match
the type of sensor ordered with the
analyzer. If the type of sensor is
changed, or the Signal Board is repaired
or replaced, verify the rocker switch
position (SW1):
Rechargeable Sensor: Close positions
1,4, and 5
2, Cl2, and
). Low-level
X
Model 7003M
•
+5V (isolated) for other functions
for that circuit
There are isolated and non-isolated
grounds on the Power Supply Board.
TP1 is the non-isolated ground and
should be used with TP6 and TP7.
c. Microprocessor Board
The Microprocessor Board is the
functional center of the instrument. With
the exception of the program chip, no
user serviceable functions are on this
assembly. If the Microprocessor Board is
to be repaired or replaced, note the
positions of the configuration jumpers for
setting jumpers on replacement board.
An instrument timer circuit monitors
instrument power for a power failure,
power surge or other outside condition
interfering with the microprocessor
operation. When conditions return to
normal, the circuit will attempt a system
restart. If the microprocessor is still
capable of operating, both alarm relays
will be put into alarm condition, the screen
will clear, and the output current will drop
to 0 mA. Then the instrument will go
through a normal power-up routine and
will eventually return to routine operation
with alarms and output restored to their
normal conditions. The appropriate error
code (E-0, etc.) will be displayed until the
ACK button is pressed.
Disposable Sensor: Close positions 2,4,
d. Alarm Circuits
and 6
Each alarm activation setpoint is designed
b. Power Supply Board
to be at a different level from the
deactivation setpoint:
The Power Supply Board provides:
•
To prevent relay contacts from
•
±12V (non-isolated) for analog
signaling function
•
+5V (non-isolated) for
microprocessor and associated
logic
chattering if the oxygen level stays
around the setpoint.
•
To indicate to the microprocessor
whether alarm is HIGH or LOW.
•
To allow for a time lag between the
time the alarm is activated and the
•
+24V (non-isolated) for the
isolated current function
5-2 Routine Servicing and TroubleshootingRosemount Analytical Inc. A Division of Emerson Process Management
time it is deactivated for ON/OFFControl purposes.
Model 7003M
ROUTINE SERVICING AND TROUBLESHOOTING
Operator Manual
748223-K
June 2002
SECTION 6
WARNING
ELECTRICAL SHOCK HAZARD
Servicing requires access to live parts
which can cause death or serious injury.
Refer servicing to qualified personnel.
Relay contacts wired to separate power
sources must be disconnected before
servicing.
6-1 ROUTINE SERVICING
Most service and maintenance involves
cleaning and recharging the sensor. Refer
to the instructions supplied with the sensor
for recharging and maintenance procedures.
Service is rarely required on the Model
7003M Percent Oxygen Analyzer. Repair is
limited to replacement of PCB boards,
EPROM and fuses. See Section 7
Replacement Parts.
6-2 TROUBLESHOOTING
Troubleshooting should be referred to
qualified service technicians. Refer to
Section 5-3 Circuit Descriptions, on page 52, as an aide in isolating a problem.
a. Symptoms
Cannot adjust signal to calibration
value
Possible cause
Sensor voltage error
Check/Test
Install know good sensor and verify
voltage
Corrective Action
Recharge or replace sensor
Readings not consistent with calibration curve
Possible cause
Incorrect sensor used for application
Check/Test
Verify sensor corresponds with
application
Corrective Action
Install new sensor and calibrate with
span gas
Possible cause
Sensor temperature incorrect
Check/Test
Sensor temperature should be 0 to 50°C
Calibration error code displayed
Possible cause
Exhausted Sensor
Check/Test
Install know good sensor and isolate
Corrective Action
Recharge or replace sensor
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing and Troubleshooting 6-1
Corrective Action
Install new sensor and calibrate with
span gas
Possible cause
Incorrect span gas concentration
Check/Test
Verify span gas concentration
Corrective Action
Install new sensor and calibrate with
span gas
Operator Manual
748223-K
June 2002
Model 7003M
Error code E-0 displayed
Possible cause
Power interruption, power surge or other
outside interference has occurred
Corrective Action
Press ACK. If this symptom occurs
frequently, surge-protection is
recommended.
Will not read low levels of oxygen
Possible cause
Ambient air leaking into system
Check/Test
Increase sample flow rate, block off
flow of gas in and out of flow
chamber, and not analyzer response
Corrective Action
If signal increases with test, isolate
and seal leak.
If signal does not increase, replace or
recharge sensor.
Sudden change in output for no apparent reason
Possible cause
Sensor subjected to sudden change in
conditions, as when a shadow moves
and exposes the sensor to an increase
in temperature
Corrective Action
Locate sensor in an area shaded from
sunlight and away from cold or warm
drafts.
Resolution degrading with time
Possible cause
Sensor exposed to low temperatures.
Check/Test
Per Section 4-4b on page 4-8, check
current output of sensor.
If the sensor is degrading, the signal will
change in steps as the oxygen level
changes instead of changing
continuously.
Corrective Action
Relocate sensor or replace more
frequently.
6-2 Routine Servicing and TroubleshootingRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 7
REPLACEMENT PARTS
7-1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a
circuit board, it is more practical to replace the
board than to attempt isolation and
replacement of an individual component. As
standard policy, rebuilt boards are available
on an exchange basis.
7-2 MATRIX
7003M MODEL 7003M OXYGEN MONITOR*
Code Power
01 115 VAC, 50/60 Hz
05 230 VAC, 50/60 Hz
Code Mounting Options
01 Panel Mount - Standard
07 Wall Mounting Plate w/ Junction Box
08 Two-Inch Pipe Mounting Bracket
Code Gain Setting
50 Signal: High Level (Used with Rechargeable Sensor % Gas)
51 Signal: Low Level (Used with Disposable Sensor % Gas)
99 Signal: Special
Because of the exchange policy covering
circuit boards, the following sections on
replacement parts do not include individual,
circuit board mounted components. If
circumstances necessitate replacement of an
individual component, which can be identified
by inspection or from the schematic diagrams,
obtain the replacement component from a
local source of supply.
7003M010150
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1
Operator Manual
748223-K
June 2002
7-3 REPLACEMENT PARTS - MODEL 7003M
WARNING
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair
PART NUMBERDESCRIPTION
652630Signal Board
652640Power Supply Board
654605Display/Keypad Board
655195EPROM
655197Microprocessor Board
777156Fuse, 1/4 A, 120V (pkg of 5)
777360Fuse, 1/8 A, 240 V (pkg of 5)
858728Arc Suppressor
Model 7003M
When replacing Microprocessor Board PN 652650, replace with Microprocessor Board PN 655197 AND
EPROM 655195.
7-2 Replacement PartsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
7-4 REPLACEMENT PARTS - SENSORS
a. Rechargeable Sensors
Sensor:
MATERIALPART NUMBER
Polypropylene623371
Polypropylene623370 (use with Fast Response Kit)
Ryton190408
Ryton190409 (use with Fast Response Kit)
Sensor Kits (Sensor not included):
DESCRIPTIONPART NUMBERUSE WITH SENSOR
Submersion, Polypropylene639904623371
In-Line Flow, Polypropylene639905623371
Fast Response, Polypropylene639906623370
Submersion, Ryton646628190408
In-Line Flow, Ryton646629190408
Fast Response, Ryton646630190409
Operator Manual
748223-K
June 2002
Sensor Kits (Sensor included):
DESCRIPTIONPART NUMBER
Submersion, Polypropylene400011
In-Line Flow, Polypropylene400012
Fast Response, Polypropylene400013
Submersion, Ryton400021
In-Line Flow, Ryton400022
Fast Response, Ryton400023
Recharge Kit
Recharge Kit (10 recharges)191755
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-3
7-4 Replacement PartsRosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
8-1 RETURN OF MATERIAL
If factory repair of defective equipment is
required, proceed as follows:
1. Secure a return authorization from a
Rosemount Analytical Inc. Sales Office or
Representative before returning the
equipment. Equipment must be returned
with complete identification in accordance
with Rosemount instructions or it will not
be accepted.
Rosemount CSC will provide the shipping
address for your instrument.
In no event will Rosemount be
responsible for equipment returned
without proper authorization and
identification.
2. Carefully pack the defective unit in a
sturdy box with sufficient shock absorbing
material to ensure no additional damage
occurs during shipping.
3. In a cover letter, describe completely:
•
The symptoms that determined the
equipment is faulty.
•
The environment in which the
equipment was operating (housing,
weather, vibration, dust, etc.).
•
Site from where the equipment was
removed.
•
Whether warranty or non-warranty
service is expected.
•
Complete shipping instructions for the
return of the equipment.
4. Enclose a cover letter and purchase order
and ship the defective equipment according to instructions provided in the Rosemount Return Authorization, prepaid, to
the address provided by Rosemount CSC.
Operator Manual
748223-K
June 2002
SECTION 8
RETURN OF MATERIAL
If warranty service is expected, the defective
unit will be carefully inspected and tested at
the factory. If the failure was due to the
conditions listed in the standard Rosemount
warranty, the defective unit will be repaired or
replaced at Rosemount’s option, and an
operating unit will be returned to the customer
in accordance with the shipping instructions
furnished in the cover letter.
For equipment no longer under warranty, the
equipment will be repaired at the factory and
returned as directed by the purchase order
and shipping instructions.
8-2 CUSTOMER SERVICE
For order administration, replacement Parts,
application assistance, on-site or factory
repair, service or maintenance contract
information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
8-3 TRAINING
A comprehensive Factory Training Program of
operator and service classes is available. For
a copy of the Current Operator and ServiceTraining Schedule contact the Technical
Services Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 8-1
Operator Manual
748223-K
June 2002
Model 7003M
8-2 Return of MaterialRosemount Analytical Inc. A Division of Emerson Process Management
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from
defects in workmanship and material under normal use and service for a period of twelve (12)
months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid
junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and
material under normal use and service for a period of ninety (90) days from date of shipment by
Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or
material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the
goods, part(s) or consumables are returned to Seller's designated factory, transportation charges
prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in
the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in
effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day
warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and
consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly
or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from
any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS,
SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED
WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD
WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC
GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER
AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT
ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT.
IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER
NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS
COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO
BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED
HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
Operator Manual
748223-K
June 2002
Model 7003M
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
E gas.csc@emersonprocess.com
ASIA - PACIFIC
Fisher-Rosemount
Singapore Private Ltd.
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T 65-777-8211
F 65-777-0947