Emerson 7003M User Manual

Operator Manual
748223-K June 2002
Model 7003M
Percent Oxygen Analyzer
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical rep-
resentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruc- tion Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and per­sonal injury.
The information contained in this document is subject to change without notice.
Ryton® is a registered trademark of Phillips Petroleum Co.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model 7003M
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
Documentation....................................................................................................................................P-5
Compliances .......................................................................................................................................P-5
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview................................................................................................................................1-1
1-2 Principle of Measurement......................................................................................................1-1
1-3 Sensors..................................................................................................................................1-1
1-4 Conditions Affecting Measurement .......................................................................................1-2
a. Gas Composition.............................................................................................................1-2
b. Sample and Cal Gas Pressures......................................................................................1-2
c. Environmental Factors ....................................................................................................1-2
1-5 Features.................................................................................................................................1-3
1-6 Specifications ........................................................................................................................1-4
a. Performance....................................................................................................................1-4
b. Physical...........................................................................................................................1-4
c. Electrical..........................................................................................................................1-5
d. Sensors ...........................................................................................................................1-5
Operator Manual
748223-K
June 2002

TABLE OF CONTENTS

2-0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Storage ..................................................................................................................................2-1
2-3 Analyzer Installation ..............................................................................................................2-1
2-4 Sensor Installation .................................................................................................................2-1
a. Analyzer/Sensor Jumper Configuration ..........................................................................2-1
2-5 Electrical Connections ...........................................................................................................2-3
a. Sensor Cable ..................................................................................................................2-4
b. Power Connections.........................................................................................................2-4
c. Current Output Connections ...........................................................................................2-6
d. Voltage Output Connections ...........................................................................................2-6
2-6 Relay Contacts for Alarms or ON/OFF-Controls ...................................................................2-8
3-0 OPERATION .........................................................................................................................3-1
3-1 Overview................................................................................................................................3-1
3-2 Keypad Operation..................................................................................................................3-1
a. Single and Double Keystroke..........................................................................................3-3
b. Display Prompts ..............................................................................................................3-4
c. Error Messages...............................................................................................................3-6
d. Automatic Return to RUN Mode .....................................................................................3-7
e. Preventing Unauthorized Access....................................................................................3-7
f. Sensor Equilibrium..........................................................................................................3-7
3-3 Start-Up .................................................................................................................................3-8
a. Initial Power-Up...............................................................................................................3-8
b. System Restart................................................................................................................3-9
c. Automatic Restart - Due to Power Failure ......................................................................3-9
d. Manual Restart................................................................................................................3-9
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Operator Manual
748223-K June 2002
3-4 Run Mode ..............................................................................................................................3-10
a. Display Oxygen Concentration/Sensor Temperature .....................................................3-10
b. Acknowledge Alarms ......................................................................................................3-10
c. Access Other Modes.......................................................................................................3-10
3-5 Setup Mode ...........................................................................................................................3-11
a. Security Code..................................................................................................................3-13
b. Current Output Range - 0-20mA or 4-20mA................................................................3-13
c. Range Endpoint Adjustments .........................................................................................3-13
d. Zero Offset Adjustment ...................................................................................................3-15
3-6 Alarm Mode vs. ON/OFF-Control Mode ................................................................................3-16
a. Relay Configured as Alarm .............................................................................................3-16
b. Relay Configured as ON/OFF-Controller ........................................................................3-16
c. Relay Contacts................................................................................................................3-18
d. Fail Safe Operation .........................................................................................................3-18
e. Relay Reset Upon Fatal Error.........................................................................................3-18
f. Pressure Effects on Alarms ............................................................................................3-18
g. Setting Alarm and ON/OFF-Controller Relays................................................................3-18
h. Setting Alarm Setpoints ..................................................................................................3-18
i. Setting Alarm ON and OFF Levels .................................................................................3-19
Model 7003M
4-0 CALIBRATION AND ADJUSTMENTS.................................................................................4-1
4-1 Frequency of Calibration .......................................................................................................4-1
4-2 Calibration mode ...................................................................................................................4-1
a. Zero Calibration...............................................................................................................4-1
b. Span Calibration..............................................................................................................4-1
c. Pressure Compensation - Correction for Constant Pressure Differences......................4-4
4-3 Hold Mode .............................................................................................................................4-5
4-4 Diagnostics Mode ..................................................................................................................4-8
a. Raw Sensor Voltage .......................................................................................................4-8
b. Test Current Output ........................................................................................................4-8
c. Linearizing Current Output ..............................................................................................4-8
d. Test Alarm Relays...........................................................................................................4-10
4-5 Adjustments And Settings .....................................................................................................4-10
5-0 THEORY................................................................................................................................5-1
5-1 Principles Of Operation .........................................................................................................5-1
a. Electrochemical Theory ..................................................................................................5-1
5-2 Effects of Sample Conditions ................................................................................................5-1
a. Barometric Pressure .......................................................................................................5-1
b. Humidity ..........................................................................................................................5-1
c. Sample Temperature ......................................................................................................5-1
d. Interfering Gases.............................................................................................................5-2
5-3 Circuit Descriptions................................................................................................................5-2
a. Signal Board....................................................................................................................5-2
b. Power Supply Board .......................................................................................................5-2
c. Microprocessor Board .....................................................................................................5-2
d. Alarm Circuits..................................................................................................................5-2
6-0 ROUTINE SERVICING AND TROUBLESHOOTING...........................................................6-1
6-1 Routine Servicing ..................................................................................................................6-1
6-2 Troubleshooting.....................................................................................................................6-1
a. Symptoms .......................................................................................................................6-1
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
7-0 REPLACEMENT PARTS ......................................................................................................7-1
7-1 Circuit Board Replacement Policy .........................................................................................7-1
7-2 Matrix .....................................................................................................................................7-1
7-3 Replacement Parts - Model 7003M.......................................................................................7-2
7-4 Replacement Parts - Sensors................................................................................................7-3
a. Rechargeable Sensors ...................................................................................................7-3
b. Disposable Sensors ........................................................................................................7-4
8-0 RETURN OF MATERIAL ......................................................................................................8-1
8-1 Return Of Material .................................................................................................................8-1
8-2 Customer Service ..................................................................................................................8-1
8-3 Training..................................................................................................................................8-1
Operator Manual
748223-K
June 2002

LIST OF ILLUSTRATIONS

Figure 2-1. Sensor Ordering Matrix.......................................................................................... 2-2
Figure 2-2. Model 7003M Rear Panel Connections................................................................. 2-3
Figure 2-3. Power Jumpers and Fuses .................................................................................... 2-5
Figure 2-4. Current Output Connections .................................................................................. 2-7
Figure 3-1. Model 7003M Front Panel Controls and Indicators ............................................... 3-1
Figure 3-2. Model 7003M Operation Flow Diagram................................................................. 3-2
Figure 3-3. SETUP Mode Flowchart ...................................................................................... 3-12
Figure 3-4. Security Access Routine Flowchart ..................................................................... 3-14
Figure 3-5. ALARM Mode Flowchart...................................................................................... 3-17
Figure 3-6. Action of Alarm Relay ON and OFF Setpoints .................................................... 3-19
Figure 4-1. CALIBRATION Mode and PRESSURE COMPENSATION Flowchart.................. 4-3
Figure 4-2. HOLD and DIAGNOSTICS Modes Flowchart ....................................................... 4-7
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Operator Manual
748223-K June 2002
Table 2-1. Sensor Cables ....................................................................................................... 2-4
Table 2-2. Analyzer Cable Connections ................................................................................. 2-4
Table 3-1. Key Functions ........................................................................................................ 3-3
Table 3-2. Display Indicators .................................................................................................. 3-3
Table 3-3. Security Access Prompts and Responses............................................................. 3-4
Table 3-4. Alarm Settings Prompts and Responses............................................................... 3-4
Table 3-5. Calibration Prompts and Responses ..................................................................... 3-4
Table 3-6. System Setup Prompts and Responses................................................................ 3-4
Table 3-7. System Hold and Diagnostics Codes .................................................................... 3-5
Table 3-8. Data Displayed in DIAGNOSTICS Mode............................................................... 3-5
Table 3-9. Miscellaneous Prompts and Responses ............................................................... 3-5
Table 3-10. Power ON, Reset Diagnostics Errors (Fatal Errors ............................................... 3-6
Table 3-11. RUN Mode Errors (Non-Fatal Errors ..................................................................... 3-6
Table 3-12. Signal Board Errors ............................................................................................... 3-6
Table 3-13. Microprocessor Errors That Reset The System .................................................... 3-6
Table 3-14. Default Settings at Power-Up ................................................................................ 3-8
Table 3-15. RUN Mode Key Functions ................................................................................... 3-10
Table 3-16. SETUP Mode Key Functions .............................................................................. 3-11
Table 3-17. ALARM Mode Key Functions.............................................................................. 3-16
Table 3-18. Alarm Relay Contact Positions/Terminal Locations............................................. 3-18
Table 4-1. CALIBRATION Mode Key Functions........................................................................1
Table 4-2. Composition of Dry Atmospheric Air...................................................................... 4-2
Table 4-3. Composition of Humid Atmospheric Air................................................................. 4-2
Table 4-4. Barometric Pressure Effect on Apparent Oxygen Concentration ..............................4
Table 4-5. PRESSURE COMPENSATION Mode Key Functions..............................................4
Table 4-6. HOLD Mode Key Functions ......................................................................................5
Table 4-7. Caret Readings...................................................................................................... 4-6
Table 4-8. DIAGNOSTICS Mode Key Functions ................................................................... 4-8
Model 7003M

LIST OF TABLES

LIST OF DRAWINGS

652542 Outline and Mounting – Panel Mount 652543 Outline and Mounting – Wall Mount 652544 Outline and Mounting – Pipe Mount 652682 Pictorial Wiring Diagram 652683 Interconnect Diagram 652699 Installation Drawing
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Operator Manual
Model 7003M

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the Model 7003M Percent Oxygen Analyzer
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
DANGER .
748223-K
June 2002
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Operator Manual
748223-K June 2002
Model 7003M

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
Relay contacts wired to separate power sources must be disconnected before servicing.
This instrument was shipped from the factory set up to operate on either 115 VAC 50/60 Hz, or 230 VAC 50/60 Hz, as specified by sales order. Verification that this voltage is proper for the installation is the responsibility of the purchaser. See Section 2-5b on page 2-4.
WARNING .
OXYGEN LEVEL WILL NOT BE MONITORED
The instrument will not return automatically to the RUN mode from HOLD or DIAGNOSTICS modes. Operator must manually return instrument to RUN mode to continue monitoring oxygen.
WARNING.
ERRONEOUS OXYGEN READINGS
If the pressure compensation function is not applied correctly, erroneous readings can be dis­played. The display using span gas during calibration should be the same as the reading using the same gas in RUN mode if the pressure values are correctly entered.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Operator Manual
Model 7003M
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair
WARNING.
SENSOR NOT INTRINSICALLY SAFE
The oxygen sensors used with this analyzer are not intrinsically safe. Use of the sensors in or near flammable liquids, gases or vapors or in Class I, Division 1, Hazardous Locations should be care­fully evaluated by qualified personnel at the site and is entirely the responsibility of the user.
The sensor is approved as non-incendive for use in Class I, Division 2 locations.
CAUTION .
748223-K
June 2002
HIGH PRESSURE GAS CYLINDERS
This module requires periodic use of pressurized gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-4
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Operator Manual
748223-K June 2002
Model 7003M
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when lo­cated in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suit­able hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is at­tached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precau­tions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should never be
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Operator Manual
Model 7003M

DOCUMENTATION

The following Model 7003M instruction materials are available. Contact Customer Service Center or the local representative to order.
748223 Operator Manual (this document)

COMPLIANCES

The Model 7003M Percent Oxygen Analyzer and Models 4000 and 5000 series sensors are suitable for use in Class I, Groups A, B, C, and D, Division 2 locations per North American electrical codes. Factory Mutual Research (FM) approved under J.I. 1T5AO.AX. Canadian Standards Association (CSA) certified under file LR 93812.
748223-K
June 2002
FM
APPROVED
®
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Operator Manual
748223-K June 2002
Model 7003M
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 1

DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

The Model 7003M Percent Oxygen Analyzer continuously monitors the concentration of oxygen in a gaseous sample. It consists of a sensor and the analyzer. The sensor is housed in a chamber assembly and is connected to the analyzer by a shielded multi­conductor cable.
The Model 7003M Percent Oxygen Analyzer provides direct digital readout, isolated current
MODE FUNCTION
RUN (see Section 3-4 on page 3-10) Normal monitoring SETUP (see Section 3-5 on page 3-11) Changing parameters ALARM/CONTROL (see Section 3-6 on page 3-16) Set alarms or ON/OFF controllers CALIBRATION (see Section 4-2 on page 4-1) Gas calibration/pressure inputs HOLD (see Section 4-3 on page 4-5) Retain last output value DIAGNOSTICS (see Section 4-4 on page 4-8) Electronics checks
The Model 7003M Percent Oxygen Analyzer is housed in a NEMA 4X (CSA enclosure 4), corrosion resistant housing suitable for panel, pipe or wall mounting.
The analyzer has a microprocessor controlled keyboard for operation. Signals from the remote oxygen and temperature sensors are processed to compute real time oxygen concentration corrected for temperature, constant pressure differences and residual sensor current. The oxygen concentration is compared with user-set alarm parameters, and also generates an isolated 4 to 20 or 0 to 20mA signal. Data is stored in a memory circuit that retains the data in case power is removed.
The analyzer is shipped from the factory configured for operation on either 115 VAC or 230 VAC, 50/60 Hz as specified by the customer when ordered.
output, automatic temperature compensation circuitry, and dual relay contact closures for alarms or ON/OFF control devices as standard features. The alarm ranges can be selected independently from the output range, eliminating the need to readjust alarm setpoints if the range is changed.
The analyzer has six modes of operation:
1-2 PRINCIPLE OF MEASUREMENT
Measurement is based on the electro­chemical reaction between oxygen and the sensor cathode which produces a low level electrical current. The sensor signal is proportional to the partial pressure of oxygen present at the membrane. This signal and a thermistor signal are received by the analyzer and processed by the electronics to output a calibrated oxygen signal that is compensated for temperature.

1-3 SENSORS

Rosemount Analytical offers rechargeable and disposable oxygen sensors which can be used with the Model 7003M. These sensors are supplied in kits: Submersion, in-line flow, fast response, and a cell separation kit. Sensors are available constructed of polypropylene or Ryton. See Section 7­4REPLACEMENT PARTS - SENSORS on page 7-3, for additional information.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Operator Manual
748223-K June 2002
Model 7003M

1-4 CONDITIONS AFFECTING MEASUREMENT

Oxygen measurement is affected by the following:
Composition of the sampled gas
Sample and calibration gas pressures
Environment of sensor, cable and analyzer
Each must be considered when calibrating and operating the analyzer to obtain accurate measurements.

a. Gas Composition

Any substance in the sample that significantly interferes or "poisons" the sensor oxygen reading should be removed by a chemical and/or mechanical sample handling system (see Section 5-2 on page 5-1). If ambient air is not used as the span gas, the composition of calibration gases should also be representative of the sample.
The sample and calibration gases must meet certain specifications (see Specifications in Preface) to ensure accurate measurement. Physical parameters that affect measurement include:

b. Sample and Cal Gas Pressures

The sample and calibration gases may be at different pressures as long as the pressure of each remains constant. Any variation in the pressure of either gas will result in an incorrect output reading of the same proportion.
Two methods of sample calibration are recommended:
1. Calibrate at the sample pressure, or
2. Calibrated at a different pressure, and enter the two pressure in the analyzer (Section 4-2c on page 4-4)
Refer to Section 5-2 on page 5-1 for a more detailed discussion of the relationship between gas pressures and apparent oxygen concentration.

c. Environmental Factors

If radio frequency interference (RFI) normally occurs in the operating environment, the sensor cable should be electrically shielded during installation.
Pressure
Humidity
Temperature
Gas composition
Also refer to Section 5-2 on page 5-1 for additional information about how gas conditions affect the oxygen measurement.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002

1-5 FEATURES

Suppressed Ranges
When setting the Model 7003M to a suppressed output range (15 to 25%, etc.), note that the absolute accuracy, drift, and noise of the oxygen measurement will remain the same for all ranges (see Specifications in Preface).
ALARM Mode
ALARM Mode is not self-resetting. Operator must press ACK to bring relays back to their no-alarm state. Used when immediate operator action is required.
NOTE:
Operator Can Bypass Alarm Relays
When relays are set to ALARM Mode (Sec­tion 3-6a on page 3-16) and an alarm con­dition has occurred, pressing the ACK key will turn off the relay switch, whether the oxygen level is out of range or not. If you do not want an operator to be able to by­pass the relay switches in this way, set the instrument to ON/OFF -CONTROL Mode.
ON/OFF-Controller Mode
ON/OFF-CONTROLLER Mode is self­resetting. Pressing ACK has no effect on relay coils. Used when automated action is required.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Operator Manual
748223-K June 2002

1-6 SPECIFICATIONS

a. Performance

Accuracy........................................ ±0.25% oxygen
Operating Range ........................... 0 to 25% oxygen (endpoints set by user)
Linearity ......................................... ±0.25% oxygen
Zero Stability.................................. ±0.25% oxygen/week at 25°C
Span Stability................................. ±0.25% oxygen/24 hours at 25°C
Noise ............................................. ±0.25% oxygen
Response Time ............................. 90% of reading in 20 seconds at 25°C
Temperature Effects
32 to 110°F (0 to 44°C) ......... ±6% reading
60 to 90°F (15 to 32°C) ......... ±3% reading
other 30°F (16°C) ranges ...... +4% reading
Model 7003M
(with constant sample temperature after sensor zero offset correc­tion)

b. Physical

Mounting
Standard ................................ Panel Mount
Optional ................................. Wall Mount, Pipe Mount
Dimensions.................................... 5.7 x 5.7 x 7.6 inches (14 x 14 x 19 cm) HxWxD
Weight ........................................... Approximately 3.0 pounds (1.36 kg)
Enclosure....................................... ABS Plastic, Black, NEMA Type 4X, IP65
Ambient Humidity .......................... 95% relative humidity, non-condensing (maximum)
Ambient Temperature.................... 14 to 122°F (-10 to 50°C)
Sensor Cable................................. 2 twisted pairs, specified length up to 1000 feet (305 m) maximum
Power Cable.................................. 3 conductor, Type SJT, 18 AWG minimum
Connection Conduit
Power..................................... 1/2 inch
Sensor ................................... 3/4 inch
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M

c. Electrical

Power Requirements..................... 115/230 VAC ±10%, 50/60 Hz ±3 Hz
Display........................................... 3-1/2 digit liquid crystal (LCD)
Signal Output................................. Isolated 4 to 20 or 0 to 20 mA (selectable)
Alarm Relay Contacts.................... Two Form C Relays
Recorder Output............................ Isolated 4 to 20 mA
Operator Manual
748223-K
June 2002
600 ohms maximum load
SPDT, resistive load, independently adjustable Deadband adjustable over full range Alarm Relay Contacts Ratings General Purpose (Ordinary) Locations: 115 VAC, 3.0 A (resistive), 230 VAC, 1.5 A resistive Class I, Division 2 Locations: CSA: 120 VAC, 3.0 A (non-inductive) 240 VAC, 1.5 A (non-inductive). FM: 28 VDC (resistive) (Ci = 0, Li = 0) Groups A&B: 150 mA Group C: 400 mA Group D: 540 mA

d. Sensors

Types............................................. Rechargeable and Disposable
Process Connections .................... Refer to flow chamber (if supplied)
Sample Pressure........................... 0 to 50 psig (0 to 345 kPa)
Sample Temperature..................... 32 to 110°F (0 to 44°C)
Ambient Temperature.................... 32 to 110°F (0 to 44°C)
Ambient Humidity .......................... 95% relative humidity, non-condensing (maximum)
Also refer to instructions supplied with sensor.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Operator Manual
748223-K June 2002
Model 7003M
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 2

INSTALLATION

2-1 UNPACKING

Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or contents is damaged. Retain the carton and packing material until all components associated with the Model 7003M Percent Oxygen Analyzer are operational.

2-2 STORAGE

The sensors and analyzer unit may be shipped together or separately. The sensors are shipped assembly and charged, ready for use.
If rechargeable sensors are stored longer than six months, they may need to be recharged or rejuvenated if performance is marginal. Disposable sensors are not rechargeable.
Refer to documentation supplied with sensors for more information.

2-3 ANALYZER INSTALLATION

The Model 7003M Percent Oxygen Analyzer is designed to meet NEMA Type 4X (CSA enclosure 4) requirements, provided that the cutouts in the bottom of the enclosure are fitted with approved conduit fittings or are properly sealed by the user with Hoffman plugs (PN A-S050) or equivalent.
The Model 7003M is supplied with the Panel Mounting Kit PN 652527 as standard. Optional Wall Mount Kit PN 652539 and Pipe Mount Kit PN 652529 are available. Refer to the Outline and Mounting drawings provided in the rear of this manual.

2-4 SENSOR INSTALLATION

NOTE:
Refer to Figure 2-1 Sensor Ordering Matrix on page 2-2, and verify that the sensor to be used corresponds with the application. If an incorrect sensor is used, calibration may be impossible or incorrect. If a differ­ent type of sensor is to be used, the rocker switch (SW1) on the Signal Board may have to be changed (see Section 2-4a below).
Install the sensor in an area where temperature is relatively constant, taking care to avoid changes in exposure to sunlight or due to sudden drafts. A room-temperature environment is preferable. The flow chamber/sensor kit contains detailed mounting and sample flow information. Add distilled or deionized water to the sensor to the maximum level indication on the sensor reservoir. Let sensor stand for approximately 15 minutes and check for leaks around the base of the reservoir, and at the seams and corners. If a leak is found, contact the factory before proceeding. Drain the sensor.

a. Analyzer/Sensor Jumper Configuration

The Model 7003M is configured to match the type of sensor ordered with the analyzer. If the Signal Board or Sensor is repaired or replaced, verify rocker switch SW1 on the Signal Board is properly set.
Rechargeable Sensor: Close positions 1, 4, and 5
Disposable Sensor: Close positions 2, 4, and 6
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Operator Manual
748223-K June 2002
Model 7003M
4000 Rechargeable Sensor 5000 Disposable Sensor
1 Polypropylene 2 Ryton (rechargeable only) 9 Special (consult factory)
1 Submersion 2 In-Line Flow 3 Fast Response (rechargeable only) 9 Special (consult factory)
4 0 0 0 3
Position
Positions 1 through 4 = Sensor Type Position 5 = Material of Construction (Sensor and Flow Assembly) Position 6 = Flow Assembly Configuration
Example 400013 is a Rechargeable sensor, constructed of polypropylene, mounted in a Fast-Response Flow Chamber Assembly.
1234
Figure 2-1. Sensor Ordering Matrix
1
5
6
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002

2-5 ELECTRICAL CONNECTIONS

Observe all precautions stated in this manual and in the drawings provided at the back of this manual before installation.
All electrical connections to the analyzer are made on the Power Supply Board connection panel (see Figure 2-2 below and drawing
652683), which is accessed by removing the rear cover.
WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without rear cover secured. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this in­strument must be connected to a properly grounded three-wire source of power.
Electrical installation must be made in ac­cordance with the National Electrical Code (ANSI/NFPA-70) and/or any applicable na­tional or local codes.
WARNING
ENCLOSURE INTEGRITY
Unused cable conduit entries must be se­curely sealed by non-flammable closures to provide enclosure integrity in compli­ance with personnel safety and environ­mental protection requirements. For watertight installation conforming to the requirements of NEMA Type 4X, use Hoffman plugs (PN A-S050) or equivalent to seal any unused openings.
Provide three separate 1/2 inch conduits as follows:
Power cable
Sensor cable
Alarm cables
1 +12V 2 -12V 1 ANODE 2 TEMP SENSE 3 NC 4 NC 5 EPOL 6 GND 7 Tx 8 Rx 9 RET 1 COM 2 NC 3 NO 4 COM 5 NC 6 NO 7 ISO GND 8 ISO CUR­9 ISO CUR+
ALARM CONTACT RATING: 3A 120 VAC
1.5A 240 VAC 3A 30 VDC
AC POWER
AL 1
AL 2
HOT NEUT GND
L1 L2
FUSE RATING: 115V: 1/4A 250V 230V: 1/8A 250V
Figure 2-2. Model 7003M Rear Panel Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Operator Manual
748223-K June 2002

a. Sensor Cable

The sensor cable used with the Model 7003M may be of any length up to a maximum of 1000 feet (305 m). (See Table 2-1 below.)
The signal cable should be routed through a conduit (customer-supplied) in permanent installations or in cases where the sensor is located more than a few feet from the control unit. (See Table 2-2 below.)
Model 7003M
Connect Sensor Cable
1. Ensure that power is disconnected and remove rear cover.
2. Install the conduit or cable clamp in the left opening in the bottom rear of the case.
3. Install the sensor cable in the instrument through the cable clamp and tighten the nut on the plug to create a seal around the cable.
NOTE:
In humid environments, temperature changes can cause moisture to con­dense in the conduit, resulting in sig­nal noise and corrosion. To minimize the effect of condensation, put desic­cant (available locally) inside the con­duit before sealing it.
CABLE PART
NUMBER
191748 10 Feet (3.1 m) 193265­SPECL
Table 2-1. Sensor Cables
If conduit used:
1/2 inch NPT female connection on bottom of control unit. (Conduit cus­tomer-supplied.
If no conduit used:
Use cable connector assembly (PN
856831) to provide watertight seal. As­sembly includes: Aluminum inner seal­ing grommet for connector compression nut.
Table 2-2. Analyzer Cable
Connections
LENGTH
Any specified length up to 1000 Feet (305 m)
4. Refer to Figure 2-2 on page 2-3. Using a flat-blade screwdriver, connect the wires of the sensor cable to the TB2-1, TB2-2, TB2-5 and TB2­6 on the rear panel of the analyzer.
5. Replace the rear cover if no other connections are to be made. Do not overtighten the cover retaining screws to avoid stripping the threads in the plastic case.

b. Power Connections

Electrical power is supplied to the analyzer via a customer-supplied three conductor cable, type SJT, minimum 18 AWG. Refer to the installation drawing in the rear of this manual.
Jumpers and Fuses
To verify the jumper location for the specified power (115 or 230 VAC) or if the analyzer is to be connected to a different power source, refer to Figure 2-3 on page 2-5and set jumpers and fuse as follows:
WARNING
ELECTRICAL SHOCK HAZARD
Disconnect power to analyzer before con­tinuing.
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
WARNING
1. Remove the rear cover from the analyzer.
2. Remove Power Supply Board.
3. Verify and/or set Jumpers E1, E2, E3, and E4 per Figure 2-3 below.
4. Verify that fuse (F1) is correct for power.
5. Re-install Power Supply Board.
6. Replace the rear cover if no other connections are to be made. Do not overtighten the cover retaining screws to avoid stripping the threads in the plastic case.
Conduit Connections
Route the power cable through conduit and into the central opening in the bottom of the case. The inlet hole is sized for standard 1/2 inch conduit fittings.
ENCLOSURE NOT GROUNDED
The non-metallic enclosure does not pro­vide grounding between conduit connec­tions. Use grounding-type bushing and jumper wires.
Connect Power Cable
Connect power wiring to the compression type connectors on the rear panel (Figure 2-2 on page 2-3) as follows:
1. Loosen the connector screw.
2. Strip insulation back approximately 1/4 inch (6.35 mm), twist stranded wires together. DO NOT tin braid wire or the connection can work loose.
3. Insert wire into connector about 1/4 inch (6.25 mm), ensuring that wires do not touch any other wire or connector.
4. Tighten screw.
115 VAC
Jumpers: E1 - E2 E3 - E4 Fuse: 1/4 Amp
115 VAC
Jumpers: E2 - E4 Fuse: 1/8 Amp
Power Supply Board
F1 E3 E4 E2 E1
1/4 AMP 115 VAC
F1 E3 E4 E2 E1
1/8 AMP 230 VAC
Figure 2-3. Power Jumpers and Fuses
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Operator Manual
748223-K June 2002
Model 7003M
CAUTION
TOTAL OUTPUT RESISTANCE
Total resistance of all output devices and associated interconnection cable must not exceed 600 ohms.

c. Current Output Connections

Refer to Figure 2-4 on page 2-7. Isolated current output (0 to 20 or 4 to 20 mA) is standard on the Model 7003M. To convert from current output to voltage output, see Section 2-5d below.
1. Disconnect power to the analyzer. Remove rear cover.
2. Connect the shielded recorder cable leads to the rear panel of the analyzer (Figure 2-2 on page 2-3):
TB3-7 has no connection TB3-8 negative (-) TB3-9 positive (+)
3. Connect the other end of the output cable to terminals of the output device. Ensure that polarity is correct. Connect the shield to a suitable ground point.
NOTE:
Current output terminals are not grounded. Ground the current loop at some point within the circuit to mini­mize noise or other undesirable inter­actions.
4. Replace the rear cover. Do not overtighten cover retaining screws, to avoid stripping the threads in the plastic case.
5. To connect recorder(s), see Figure 2-4 on page 2-7.

d. Voltage Output Connections

Refer to Figure 2-4 on page 2-7. To convert the voltage output to current output see Section 0 above.
To monitor the signal output with a voltage recorder or other voltage device, the standard current output must be converted into a voltage by connecting a resistor in parallel with the output device as follows:
1. Determine the resistor value (R) required:
R
ohms = (fullscale voltage,
1
volts)/(0.02 amps)
2. Disconnect power; remove the rear cover.
3. Connect resistor (1/4 watt or greater) of value R between TB3-8 and TB3-9 (Figure 2-4 on page 2-7).
4. Connect leads of a shielded recorder cable to the rear panel of the analyzer (Figure 2-2 on page 2-3):
TB3-7 has no connection TB3-8 negative (-) TB3-9 positive (+)
5. Connect other end of output cable to terminals of output device. Ensure that polarity is correct.
6. Replace the rear cover. Do not overtighten cover retaining screws, to avoid stripping the threads in the plastic case.
7. For the low end point to be 0 volts, set the current output range to 0 to 20mA when setting up the instrument (Section 3-5b on page 3-13).
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
TB3
RECORDER
S - +
Operator Manual
748223-K
June 2002
ONE RECORDER
7
8
9
7
8
9
NEG (-)
POS (+)
NEG (-)
POS (+)
RECORDER
A
S - +
RECORDER
B
S - +
MULTIPLE
RECORDERS
Figure 2-4. Current Output Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Operator Manual
748223-K June 2002
Model 7003M
2-6 RELAY CONTACTS FOR ALARMS OR
ON/OFF-CONTROLS
NOTE:
Whenever "Alarm" is used, it refers to either alarm or ON/OFF Control, de­pending on how the relays are con­nected and configured by the customer.
Ratings
Two sets of alarm relay contacts are provided for actuation of customer­supplied alarm and/or ON/OFF-Control functions. (see Section 1-6 specifications on page 1-4).
Power Failure
Removal of AC power from the analyzer, as in a power failure, removes power from both relay coils. The Model 7003M should operate on a different AC power source than the Alarm relay contacts.
Connecting Devices to Relay Contacts
WARNING
RELAY CONTACTS
Relay contacts wired to separate power source must be disconnected before servicing.
Connect alarm or ON/OFF-Control devices to the relay contacts as follows:
1. Disconnect power; remove rear cover of analyzer.
2. Connect leads from the external alarm or ON/OFF Control system (see Figure 2-2 on page 2-3).
TB3-1 Alarm 1 Com TB3-2 Alarm 1 NC TB3-3 Alarm 1 NO TB3-4 Alarm 2 Com TB3-5 Alarm 2 NC TB3-6 Alarm 2 NO
Radio Frequency Interference (RFI)
If alarm contacts are connected to any device that causes radio frequency interference (RFI), an arc suppressor (PN
858728) should be used to minimize RFI.
3. Connect other end of output cable to terminals of alarm device.
4. Replace the rear cover. Do not overtighten cover retaining screws, to avoid stripping the threads in the plastic case.
NOTE:
For installation in a Class I, Division 2 location, refer to the Specifications in the Preface section, and the approval label on the analyzer for Factory Mu­tual limitations on contact ratings.
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 3

OPERATION

3-1 OVERVIEW

It is recommended that the operator read the entire manual through before first operating the instrument.
The operator can generally operate the Model 7003M using the flow chart in Figure 3-2 on page 3-2 after the display symbols and modes of operation are understood.

3-2 KEYPAD OPERATION

During each particular keystroke sequence, the analyzer will recognize only certain keys. If an unrecognized key is pressed, the instrument will either not respond or will display "SrY" ("Sorry") and return to RUN Mode.
Generally, to escape from any routine, press ENTER repeatedly until the instrument returns to the RUN Mode and the oxygen concentration is displayed.
Changing Parameters With The Keypad
All operation parameters are set with the keypad. When the instrument is waiting for operator input, the display will flash a digit. At this point, if the number itself can be changed, the operator can generally increase the number (UP arrow) or decrease it (DOWN arrow) at this time. An exception is when the operator is restricted to pre-selected choices (e.g., output current ranges, see Section 3-5b on page 3-13).
When changing a value, an arrow key can be pressed for a moment, and the display will change by one digit. If it is held down, the display will change, slowly at first, then rapidly.
If the display has a digit flashing and will not change, check the range of that parameter. For example, an alarm setpoint cannot be set higher than 25%.
When the number desired is displayed, press ENTER. The Model 7003M will accept that number and continue with the next step of that routine.
Rosemount Analytical
ALARM1
ALARM2
%O2
°
C
MODEL 7003M
DISPLAY ACK
PERCENT OXYGEN
Figure 3-1. Model 7003M Front Panel Controls and Indicators
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
ALARM1ALARM
ANALYZER
2
CAL ENTER
Operator Manual
748223-K June 2002
Model 7003M
RUN MODE
Display Oxygen Temperature
SETUP MODE
CAL + ENTER
ENTER SECURITY
CODE
Set New Security Code
ENTER ACK
Select
ACK
0-20 or 4-20mA
ENTER
Set Concentration for Minimum Output Current
ENTER
ENTER
ZERO GAS
ENTER
(0 or 4mA)
Set Concentration for Maximum Output Current
(20mA)
FLOW
Set Residual Current Offset
ALARM MODE
ALARM1*
ENTER SECURITY
CODE
ENTER
ENTER
DISPLAY ACK
DISPLAY ACK
Disable Alarm
ALARM 1*
Toggle
ON/OFF
Change "ON" Setpoint
Change "OFF" Setpoint
Acknowledge Alarm
CALIB. MODE
CAL
ENTER SECURITY
CODE
ENTER
Return to
RUN Mode
PRESSURE
COMPENSATION
ROUTINE
CAL
ENTER
ENTER
Reveal Security Code
HOLD MODE
ENTER SECURITY
Set Span Concentration
DIAGNOSTICS
DISPLAY + ACK
Set Sample Pressure
Set Span Gas Pressure
ENTER
+
CODE
ENTER
Return to
RUN Mode
MODE
DISPLAY
Select Displayed Parameter
ALARM 1*
Toggle
ON/OFF
Change Current Output
Return to
RUN Mode
Return to
RUN Mode
Return to
RUN Mode
Return to
RUN Mode
*Press ALARM 2 to set Alarm 2.
Figure 3-2. Model 7003M Operation Flow Diagram
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002

a. Single and Double Keystroke

Most procedures involve a sequence of single keystrokes by which the operator
KEY FUNCTION HOW TO USE IT
ACK Acknowledges an alarm condition Press once to silence alarms
DISPLAY
ARROW keys Increase/Decrease displayed value
CAL Access CALIBRATION mode
ENTER End command sequence
ALARM1 Access ALARM 1 mode
ALARM2 Access ALARM 2 mode
CAL + ENTER Access SETUP mode
UP + DOWN ARROW Access HOLD mode.
DISPLAY + ACK Access DIAGNOSTICS mode
DISPLAY + ACK Access security ID code
Switch between %O2 and °C
Press to display °C; press again to display O2. Output signals are not affected. Press once to change value by one digit. Hold key down to change value quicker. Press to calibrate. (If security is enabled, prompts for ID code first) Press to enter value or to end a command se­quence Press to check, set or disable Alarm 1 or ON/OFF­Control relays Press to check, set or disable Alarm 2 or ON/OFF­Control relays From RUN mode, press CAL and ENTER keys simultaneously. From RUN mode, press UP ARROW and DOWN ARROW keys simultaneously. From HOLD mode, press DISPLAY and ACK keys simultaneously. From RUN mode, press DISPLAY and ACK keys simultaneously.
presses one key at a time. Some of the more frequently used single-key commands are listed in Table 3-1 below.
Table 3-1. Key Functions
INDICATOR CARET/STATUS DESCRIPTION
%O2 Caret ON Oxygen percentage displayed
°C ALARM1,2 Caret OFF Normal condition ALARM1,2 Caret blinking Alarm condition
- Caret ON
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 3-3
Caret ON
Sensor temperature displayed, °C
Displays current output in mA (DIAGNOSTICS mode only)
Table 3-2. Display Indicators
Operator Manual
748223-K June 2002
Model 7003M

b. Display Prompts

The Model 7003M prompts the operator for inputs and also indicates the function
CODE DEFINITION
id enter security code prompt
no invalid security code entered - access denied
YES access granted or security code entered is correct
Table 3-3. Security Access Prompts and Responses
CODE DEFINITION
A-1 setpoints for Alarm 1 - ALARM function A-2 setpoints for Alarm 2 - ALARM function C-1 setpoints for Alarm 1 - ON/OFF - Controller function C-2 setpoints for Alarm 2 - ON/OFF - Controller function
On value in which to initiate an Alarm Action (On-setpoint)
OFF value in which to terminate an Alarm Action (Off-setpoint)
dis Alarm is now disabled
Table 3-4. Alarm Settings Prompts and Responses
in progress. For example, when entering the calibration mode, the prompt "CAL" is displayed for one second after the CAL key is pressed. Display acronyms are listed in Table 3-3 through Table 3-9.
CODE DEFINITION
CAL initiate a span calibration
PPr enter process pressure value
CPr enter calibration pressure value
GAS Initiate a zero offset calibration (gas)
Table 3-5. Calibration Prompts and Responses
CODE DEFINITION
Set enter system setup mode 4-0 define the current output as 4-20 or 0-20 mA
CLo % concentration that 0/4 mA represents (calibration low)
CHi % concentration that 20 mA represents (calibration high)
Table 3-6. System Setup Prompts and Responses
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
CODE DEFINITION
A-1 toggle ON/OFF check of relay coil for Alarm 1
A-2 toggle ON/OFF check of relay coil for Alarm 2 D-2 performing non-intrusive test Internal RAM (16#00#..16#FF#) D-3 performing non-intrusive test on Internal RAM (16#4000#..16#5FFF) D-4 performing non-intrusive test on EEPROM (16#B600#..16#B7FF#) D-5 performing non-intrusive test on PROM (16#C000#..16#FFFF#)
F-2 failed D-2 test
F-3 failed D-3 test
F-4 failed D-4 test
F-5 failed D-5 test
hld entering system Hold
OFF alarm relay is de-energized (normally-closed)
On alarm relay is energized (normally-open) tSt entering on-line test diagnostics mode
CODE DEFINITION
caret lit displayed
O2 sensor membrane voltage from ADC
Temp sensor thermistor voltage from DC
minus sign milliamps sent to current output
Operator Manual
748223-K
June 2002
Table 3-7. System Hold and Diagnostics Codes
Table 3-8. Data Displayed in DIAGNOSTICS Mode
CODE DEFINITION
Sry invalid keypad function (Sorry)
run current function completed - entering RUN Mode
Table 3-9. Miscellaneous Prompts and Responses
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Operator Manual
748223-K June 2002
Model 7003M

c. Error Messages

If an error condition occurs, E-# (non-fatal error) or F-# (fatal error) will flash on the
CODE DEFINITION
F-1 Invalid PROM (16#C000#..16#FFFF#) checksum detected F-2 Internal RAM (16#00#..16#FF#) failed testing F-3 External RAM (16#4000#..16#5FFF) failed testing F-4 EEPROM (16#B600#.. 16#B7FF#) bulk erase failure F-5 Signal Board not connected
F-6 thru F-9 EEPROM write failure
Table 3-10. Power ON, Reset Diagnostics Errors (Fatal Errors
CODE DEFINITION
E-1 5 volts power supply above 5.5 V E-2 5 volts power supply below 4.5 V E-3 12 volts power supply above 13.2 V E-4 12 volts power supply below 10.8 V E-5 sensor output too low for calibration E-6 sensor missing E-8 E-9
sample temperature is below 1°C sample temperature is above 44°C
display. Table 3-10 through Table 3-13 list error codes and corrective actions. Pressing "ACK" or "ENTER" will acknowledge any alarm and the error message will disappear.
Table 3-11. RUN Mode Errors (Non-Fatal Errors
CODE DEFINITION
HI
LO
Ahi excessive sensor signal
CODE DEFINITION
COP Computer Operating Properly error (error detected)
iot Invalid Op code Trap (internal computer command is invalid)
E-0 power interruption occurred
sample temperature is above 50°C sample temperature is below 0°C
Table 3-12. Signal Board Errors
Table 3-13. Microprocessor Errors That Reset The System
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M

d. Automatic Return to RUN Mode

If no keys are pressed for five minutes, the instrument will automatically return to the RUN Mode, except when in Zero Offset, HOLD, or DIAGNOSTICS Modes.
Operator Manual
748223-K
June 2002
After the security code is entered, the operator can change any instrument setting and can go back and forth to the RUN Mode without having to enter the code again.
CAUTION
OXYGEN WILL NOT BE MONITORED
The instrument will remain in HOLD or DI­AGNOSTICS Modes until the operator re­turns it to the RUN Mode or other operation.

e. Preventing Unauthorized Access

The Model 7003M is equipped with a security feature to prevent unauthorized tampering with the instrument settings. The instrument is shipped from the factory with the security feature disabled. It can be enabled or disabled at any time.
Security access is not required to:
Display sensor data
Acknowledge alarms
Reveal security code
Security access is required to:
If security is enabled, and the instrument is left in the RUN Mode for five minutes, security will automatically be re-activated. Within this five minute period, anyone can access the functions without re-entering the security code.

f. Sensor Equilibrium

When a new sensor is installed, power up the instrument and let the sensor stabilize for about 5 hours before calibrating it.
If the oxygen sensor is calibrated at a relatively high oxygen concentration, such as with ambient air (typically 20.95%) and then is exposed to a low oxygen level, a longer equilibrium time period may be required for the sensor to equilibrate. This time period depends on the monitoring concentration level and how long the sensor was operated at the higher level.
Calibrate with span gas
Access options and diagnostics
Change operation parameters
Enable or disable alarms and
ON/OFF-Control
Disable or change the security code
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Operator Manual
748223-K June 2002
Model 7003M

3-3 START-UP

Before proceeding, the operator should be familiar with the keypad and indicators (Section 3-2 on page 3-1).
Operational checks and adjustments are not normally required, provided that the sensor, inter-connection cable and analyzer have been properly installed. The analyzer continually performs self-checking routines during operation.
Calibration with ambient air is recommended after start-up and at regular intervals (see Section 4-2 on page 4-1)
The following sections provide instructions for initial start-up and subsequent standardization of the analyzer.

a. Initial Power-Up

1. Ensure that the proper sensor and sensor cable are connected to the analyzer.
2. Apply power to the analyzer.
The following internal tests are performed:
ROM "Checksum" test EEPROM Read/write test
3. If an error code flashes, refer to the error code lists in Table 3-10 through Table 3-13. Press ACK to continue with the startup procedure.
After internal tests are passed, the software version is displayed for a few seconds, and the analyzer goes directly into RUN mode. "RUN" is briefly displayed, then current oxygen level.
Error codes E-0 (power failure) and E-5 (not calibrated) typically occur upon power-up.
4. Monitor the oxygen output or display until the level stabilizes (typically 5 hours). The alarms and current output levels can be set immediately if desired.
5. Once the sensor has stabilized, the analyzer is operational and is ready for subsequent operations. Refer to Table 3-14 below for the default settings present upon power-up.
RAM Read/write test
Mode: Alarm
Alarm 1
Range
Current Output Range 4-20 mA Security Disabled
Pressure
Residual Zero Offset 0 Calibration Status Not Calibrated
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
State: Disabled Setpoints: On: 0.0 Off: 0.1% Mode: Alarm State: DisabledAlarm 2 Setpoints: On: 0.0 Off: 0.1% 0% of scale: 0% Oxygen 100% of scale: 25% Oxygen
Calibration: 76.0 cm Hg Process: 76.0 cm Hg
Table 3-14. Default Settings at Power-Up
Model 7003M
Operator Manual
748223-K
June 2002

b. System Restart

The Model 7003M will attempt to restart itself in the event of a power interruption, power surge or other outside interference to microprocessor operation (Section 3-3c below).
The analyzer can also be manually restarted if the RAM or EEPROM becomes corrupted (Section 3-3d below).
c. Automatic Restart - Due to Power
Failure
If a power failure or other event occurs that affects operation of the microprocessor the system will attempt to re-start itself. If it is successful, the instrument will continue to operate normally, except that the appropriate error code will be displayed until the ACK button in pressed. See Section 5-3c on page 5-2 for circuit description.

d. Manual Restart

1. Turn power to analyzer OFF.
2. Wait 30 seconds, and turn the power to the analyzer ON.
3. While the software revision is being displayed and the self-tests are being performed, press DISPLAY and ACK simultaneously.
"dYn" is displayed.
4. Press ENTER to clear the RAM and EEPROM.
The display will go blank, then continue with the SETUP routine.
Continue with Section 3-5a on page 3-13, step 2.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Operator Manual
748223-K June 2002
Model 7003M

3-4 RUN MODE

Refer to Figure 3-2 on page 3-2 and Table 3-15 below. The RUN mode is the normal mode of operation, from which the operator can:
KEY DISPLAY WHAT HAPPENS
DISPLAY Number change Toggles between O2 reading and temperature ACK Clears message Acknowledges Alarms ALARM 1 "A-1" then ON To Alarm 1 SETUP mode ALARM 2 "A-2" then ON To Alarm 2 SETUP mode UP ARROW "SrY" "run" Stays in RUN mode DOWN ARROW "SrY" "run" Stays in RUN mode CAL "CAL" then ON To CALIBRATION mode ENTER "SrY" "run" Stays in RUN mode CAL + ENTER "SEt" then ON Moves to SETUP mode UP + DOWN "hLd" then ON Moves to HOLD mode DISPLAY + ACK "123" Shows the security code
Display oxygen concentration and sensor temperature
Acknowledge alarms
Display security code
Access other modes
Table 3-15. RUN Mode Key Functions
a. Display Oxygen Concentration/Sensor
Temperature
Press DISPLAY once to display sensor temperature. Press DISPLAY again to display oxygen concentration.

b. Acknowledge Alarms

If an alarm setpoint has been reached, "A­1" or "A-2" will flash with a concentration value. Acknowledging the alarm will stop the flashing display and turn off the relay switch.
To Acknowledge Alarm - Press ACK key. If relays are configured as ON/OFF controllers, pressing the ACK key has no effect.
3-10 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

c. Access Other Modes

The following lists which key(s) to press to access various modes:
SETUP = CAL+ENTER ALARM = ALARM1 (or ALARM2) CALIBRATION = CAL HOLD = UP and DOWN ARROWS DIAGNOSTICS = DISPLAY+ACK (from HOLD)
Model 7003M
Operator Manual
748223-K
June 2002

3-5 SETUP MODE

Refer to Figure 3-3 on page 3-12 and Table 3-16 below. From the SET-UP mode the operator can:
KEY DISPLAY WHAT HAPPENS
DISPLAY Nothing Nothing ACK "<4>-0" / "4-<0> Selects 4-20 or 0-20 mA operation ALARM 1 Nothing Nothing ALARM 2 Nothing Nothing UP ARROW DOWN ARROW
10.8 10.9
10.9 10.8 CAL Nothing Nothing ENTER various to "run" Moves to the next step or goes to RUN mode
Table 3-16. SETUP Mode Key Functions
Change security code
Set output current range (0-20mA or 4­20mA)
Set output concentration range endpoints
Adjust zero offset
Changes the number displayed Changes the number displayed
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11
Operator Manual
748223-K June 2002
SET-UP Mode
CAL ENTER
+
Model 7003M
Flashing
Flashing May not be displayed each time.
Flashing May not be displayed each time.
Flashing
Flashing
ALARM1
ALARM2
%O2
C
°
SECURITY ACCESS ROUTINE
(see Figure 3-4)
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
Prompts for Security Code. If left at 000, code will be disabled. Use Arrow keys to set Security Code.
Select either 4-20mA or
ACK
0-20mA Output
ENTER
ENTER
ALARM1
ALARM2
%O2
C
°
Prompts for low end output. (i.e., 0% O2 would be the low range
ENTER
representing O2 level of 0 or 4mA.
Flash for 2 seconds
(Hi range output selection)
Flash for 2 seconds
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
%O2
C
°
%O2
C
°
%O2
C
°
Prompts for zero offset
%O2
C
°
%O2
C
°
Set Span output level = 20mA
Set Span output level = 20mA
ENTER
ENTER
Figure 3-3. SETUP Mode Flowchart
3-12 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002

a. Security Code

Refer to Figure 3-4 on page 3-14.
Reveal Current Security Code
From RUN mode, hold down DISPLAY and ACK keys simultaneously.
Change Current Security Code
1. From RUN mode, press CAL and Enter simultaneously.
2. SET, id are briefly displayed. Then "000" is displayed with the right digit flashing.
NOTE
If RUN and CAL are not press exactly at the same time, "SRY", or "RUN" will appear, and the analyzer will return to RUN mode.
3. If the security code has been disabled, skip to step 4. Use the ARROW keys to enter the current security code, then press ENTER.
1. Press ACK until the desired range is flashing, press ENTER.
2. Continue with the next SET-UP routine, or hold ENTER key to return instrument to RUN mode.

c. Range Endpoint Adjustments

The oxygen values corresponding to minimum and maximum output current must be set. Default is 0% for 0 or 4mA and 25% for 20mA. Endpoints must be within the range of the instrument and must be at least 1% oxygen apart.
Set Oxygen Low Concentration End­point
Following the last procedure, "CLo" is displayed briefly, followed by the oxygen concentration last stored to correspond to minimum output current.
NOTE
If "CLo" is set to a value greater than "CHi", "CHi" will be automatically reset to a value 1% greater than "CLo".
4. If the correct code is entered, YES, SET, id will be displayed briefly, the code will flash. If the incorrect code is entered the analyzer will return to RUN mode.
5. Use the ARROW keys to set the new security code (or "000" to disable), press enter.
6. Continue with the next set-up routine, or hold down ENTER until "run" appears (the instrument is back in RUN mode).
b. Current Output Range - 0-20mA or 4-
20mA
Following the last procedure, "4-0" is displayed with either digit flashing. "4" indicated 4-20mA (default), and "0" indicates 0-20mA.
Use the ARROW keys to set the low oxygen concentration (0 to 24%), press ENTER.
"CHi" is displayed briefly, followed by the latest oxygen concentration stored corresponding to maximum output current.
Set Oxygen High Concentration End­point
"CHi" cannot be set to a value equal to or less than "CLo".
Use the ARROW keys to set the high oxygen concentration (1 to 25%), press ENTER.
Continue with next SET-UP routine or hold ENTER to return to RUN mode.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-13
Operator Manual
748223-K June 2002
Model 7003M
RUN
SET-UP Mode
CAL ENTER
+
ALARM CAL
WAS SECURITY ID
ENTERED <5 MIN.
AGO
NO
ENTER ID
ENTER
ID
CORRECT
?
HOLD Mode
+
YES
OPERATOR GIVEN
ACCESS TO SE-
LECTED MODE
YES
NO
ACCESS DENIED
Returns to
RUN mode
Figure 3-4. Security Access Routine Flowchart
3-14 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002

d. Zero Offset Adjustment

Following the last procedure, the current zero offset value will be flashing.
NOTE
When flowing zero gas over the sen­sor, ensure that there are no leaks and the gas actually has zero oxygen con­centration. Otherwise, significant er­rors will result. Zero offset adjustment is generally not required. Errors are more often due to incorrect zero gas oxygen concentration than residual current of the oxygen sensor.
The zero offset adjustment allows the operator to compensate for any residual sensor current. Adjustment range is 0-5% equivalent oxygen.
Sensor Current = Current from Oxygen + Residual Current
NOTE
Adjusting the zero offset affects the output of the Model 7003M. Therefore, if the zero offset is changed, the sys­tem must be re-calibrated with span gas.
Zero Offset Is Not To Be Changed
Press ENTER to return analyzer to RUN mode, and skip the remainder of this procedure.
Zero Offset Is To Be Removed
1. Press the UP ARROW until the maximum value is displayed.
2. Press ENTER and re-calibrate with span gas per Section 4-2b on page 4-
1.
Zero Offset To Be Adjusted
1. Expose the sensor to 100% oxygen­free nitrogen or other inert gas and allow the sensor signal to stabilize (2­5 hours depending on sample conditions).
2. Use the DOWN ARROW to bring the displayed value to zero (for maximum offset).
3. Press ENTER to return to RUN mode.
4. Re-calibrate with span gas per Section 4-2b on page 4-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-15
Operator Manual
748223-K June 2002
Model 7003M
3-6 ALARM MODE VS. ON/OFF-CONTROL
MODE
Refer to Figure 3-5 on page 3-17 and Table 3-17 below. From the ALARM or ON/OFF­Control Mode, the operator can:
KEY DISPLAY WHAT HAPPENS
DISPLAY Nothing Nothing ACK "dIS" Disables alarms ALARM 1 "C-1" / "A-1" Toggles between alarm and controller mode ALARM 2 "C-2" / "A-2" Toggles between alarm and controller mode UP ARROW DOWN ARROW CAL Nothing Nothing ENTER "OFF" / "run" Moves to the "OFF" setpoint step or goes back to RUN mode

a. Relay Configured as Alarm

Relays are configured as alarms (ALARM Mode) when immediate operator action is required.
When alarms are selected, "A-1" and "A­2" are displayed in the ALARM mode set­up procedure.
10.8 10.9
10.9 10.8
Table 3-17. ALARM Mode Key Functions
Set alarm ON and OFF setpoints.
Select alarm or ON/OFF-Control function
for relays
Disable alarm(s)
Changes alarm setpoint Changes alarm setpoint
b. Relay Configured as ON/OFF-
Controller
Relays are configured as ON/OFF­Controllers (ON/OFF-Controller mode) when a non-critical, non-hazardous automated action is desired, or when it is important that the relay output not be bypassed by the operator.
When an alarm condition occurs, the caret blinks, the relay activates, and the display flashes. Press the ACK key to acknowledge the alarm, de-energize the relay coil and stop the alarm actions.
3-16 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
When ON/OFF-Controllers are selected, "C-1" and/or "C-2" is displayed in the ALARM mode set-up procedure.
When an ON/OFF-Controller alarm condition occurs, Alarm 1 or 2 caret blinks and the relay coil is energized. However, the display does not flash, and the ACK key is inactive. Alarm conditions are automatically reset when the sensor indicates that an alarm condition no longer exists.
Model 7003M
(
)
Operator Manual
748223-K
June 2002
ALARM/CONTROLLER MODE
To Disable Alarm
Indicates Alarm is disabled
Flashes Alarm Configuration
ALARM1
ALARM2
Flashing
ALARM1
ALARM2
Prompts "ON" Setpoint
Displays "ON" Setpoint
ACK
1
%O2
C
°
ALARM
ALARM1
ALARM2
Note: Alarms are disabled when instrument is received. Alarm 2 uses identical sequence.
1
%O2
C
°
SECURITY ACCESS ROUTINE
see Figure 3-4
ALARM1
ALARM2
%O2
C
°
ALARM1
ALARM2
ALARM1
ALARM2
OR
%O2
C
°
ALARM1
ALARM2
%O2
C
°
%O2
C
°
STANDARD ALARM
Adjust Alarm ON Setpoint
ENTER
1
%O2
C
°
Flashes Control Configuration
ALARM
1
Change Relay configuration
ALARM1
ALARM2
%O2
C
°
ALARM1
ALARM2
ALARM1
ALARM2
%O2
Prompts for OFF Setpoint
C
°
%O2
C
°
Adjust Alarm OFF Setpoint
ENTER
ALARM1
ALARM2
New values/functions will not be stored until instrument returns to RUN mode with relay activated.
1
%O2
C
°
Figure 3-5. ALARM Mode Flowchart
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-17
Operator Manual
748223-K June 2002

c. Relay Contacts

Relay contacts are listed in Table 3-18 below:
Model 7003M
RELAY POS
TERMINAL LOCATIONS
ALARM1 ALARM2
NO (Normally open) TB3-3 TB3-6 NC (Normally closed) TB3-2 TB3-5 COM (Common) TB3-1 TB3-4
Relay positions are with power removed.
Table 3-18. Alarm Relay Contact Positions/Terminal Locations

d. Fail Safe Operation

g. Setting Alarm and ON/OFF-Controller
The relays are setup to operated in "Fail Safe" configuration. In this configuration the relay coils are powered when the unit is NOT in an alarm condition. When an alarm condition occurs, power is removed from the relay coils.
Wherever "Alarm" is referred to in this section, the term "ON/OFF-Controller" can be substituted, unless specifically called out. Also, any reference to A1 or Alarm 1 also applies to Alarm 2, and to ON/OFF-
Controllers 1 and 2. If a power failure occurs, power will be removed from the relays and an alarm

h. Setting Alarm Setpoints

condition will be indicated to the device connected to the relays.
This procedure can be used to:
Relays

e. Relay Reset Upon Fatal Error

If a fatal error is detected, the unit will
Verify alarm relay setpoints
Change setpoints attempt to reset itself which will make both relays go into alarm mode. If the
Enable alarms microprocessor is able to continue its
operation, the relays will then be reset to
Disable alarms their normal operating position.
Perform the following procedure for each

f. Pressure Effects on Alarms

Alarm setpoint, as required. Figure 3-6
illustrates how relays are energized and If the sample pressure changes, the partial pressure of oxygen, and thus the
de-energized in High and Low Alarm
configurations. reading, will change and the alarm relays will not actuate at the same oxygen levels
as they were set when the interment was last calibrated. See Section 5-2a on page 5-1 for a more detailed description of pressure effects.
3-18 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002

i. Setting Alarm ON and OFF Levels

Refer to flowchart Figure 3-5 on page 3-
17.
From RUN mode, press ALARM 1.
A-1, YES, A-1 (or C-1), ON flash briefly; the alarm ON setpoint is displayed with the right digit blinking.
A-1 indicates relay 1 acts as an ALARM; C-1 indicates relay one acts as an ON/OFF-Control.
To change the relay action from
ALARM to ON/OFF Control function (or
vise versa) press ALARM 1. C-1 or A-1
will flash indicating a new mode.
To disable ALARM 1 - Press ACK.
"DIS" will flash briefly and return to RUN
mode.
To set the ON setpoint for ALARM 1 -
Use the ARROW keys, press ENTER.
OFF is flashed briefly, then the alarm OFF
setpoint is displayed with the right digit
flashing.
To set the OFF setpoint for ALARM 1 -
Use the ARROW keys, press ENTER.
The instrument displays "run" and returns
to RUN mode.
HIGH Alarm
set for 8.24% O
LOW Alarm
set for 8.10% O
Figure 3-6. Action of Alarm Relay ON and OFF Setpoints
2
8.24%
8.10%
2
8.24%
8.10%
ACTIVATES DE-ACTIVATES
ALARM ON
ON
ACTIVATES DE-ACTIVATES
ON
SETPOINT
OFF SETPOINT
OFF
TIME
ALARM ON
OFF
OFF SETPOINT
SETPOINT
TIME
DEADBAND
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-19
Operator Manual
748223-K June 2002
Model 7003M
3-20 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
SECTION 4

CALIBRATION AND ADJUSTMENTS

Operator Manual
748223-K
June 2002

4-1 FREQUENCY OF CALIBRATION

When a new sensor is installed, adjust the zero offset current (Section 3-5d on page 3-
15) and calibrate the instrument with span gas (4-2b on page 4-1. Span the instrument weekly at first and note the stability. Use this information to determine the regular calibration frequency required by this particular application.
KEY DISPLAY WHAT HAPPENS
DISPLAY Nothing Nothing ACK Nothing Nothing ALARM 1 Nothing Nothing ALARM 2 Nothing Nothing UP ARROW DOWN ARROW
10.8 10.9
10.9 10.8 CAL "PPr" Moves to pressure compensation mode ENTER "run" Drops back to RUN mode
Table 4-1. CALIBRATION Mode Key Functions

a. Zero Calibration

Each oxygen sensor produces a constant residual electrical current in the absence of oxygen due to characteristics of the sensor. Typically this current is less that 0.1% apparent oxygen. It is different for each sensor, but will remain constant for the life of that sensor. The instrument can be zeroed by adjusting the residual current per Section 3-5d on page 3-15 After this procedure is performed once, the zero should not need adjusting again during the life of the sensor.

b. Span Calibration

4-2 CALIBRATION MODE

Refer to Figure 4-1 on page 4-3and Table 4-1 below. From the CALIBRATION Mode, the operator can:
Calibrate with span gas
Enter sample and span gas pressures
Changes calibration factor Changes calibration factor
2. Press CAL to enter the CALIBRATION mode. Enter the security code if required.
3. CAL flashes and oxygen concentration is displayed with the right digit flashing. Output is held at last reading until returned to RUN mode.
4. Use the ARROW keys to set the instrument to the concentration of the calibration gas. See Table 4-2 on page 4-2 and Table 4-3 on page 4-2 if using ambient air as span gas.
1. Expose the sensor to ambient air and let the oxygen reading and sensor temperature stabilize.
1
We recommend using ambient air in calibration, re-
gardless of the range. Only when monitoring at 0 to
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-1
1
1% range, can the accuracy of the instrument be slightly improved by using a gas of lower concentra­tion (e.g., 1%) for calibration. If bottled air is used, verify the oxygen concentration against National Bu­reau of Standards certified calibration gas. The true oxygen concentration in cylinders is often different from the concentration indicated on the cylinder label.
Operator Manual
748223-K June 2002
Model 7003M
5. Press ENTER to return to the RUN mode.
7. If the sample gas and the calibration gas are each at constant pressure, continue with Section 4-
6. If the ratio between the concentration value entered and
2c on page 4-4 and enter the
pressure of each. the sensor voltage is excessive, an error message will be flashed.
Oxygen 20.946% Nitrogen 78.084% Argon 0.934% CO2 0.033%
±0.002 ±0.004 ±0.004 ±0.001
Water Vapor 0.000% Other Gases Balance
760 mm Hg; 0% Relative Humidity
1

Table 4-2. Composition of Dry Atmospheric Air

GAS
15°C20°C25°C30°C
PERCENT COMPOSITION ( .002%) A@
Oxygen 20.641% 20.510% 20.338% 20.115% Water Vapor 1.683 2.307 3.126 4.187 Other Gases Balance Balance Balance Balance
760 mm Hg; 100% Relative Humidity
2

Table 4-3. Composition of Humid Atmospheric Air

1
CRC Handbook of Chemistry and Physics, R. Weast, ed., Chemical Rubber Co., Boca Raton, FL: CRC Press, 1987, p. F-
148.
2
Computed from data in CRC Handbook of Chemistry and Physics, p. D-190.
4-2 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
(
)
)
Operator Manual
748223-K
June 2002
CALIBRATION MODE
CAL
STANDARD CALIBRATION PROCEDURE
Example if using in Air
ALARM1
ALARM2
ALARM1
ALARM2
SECURITY ACCESS ROUTINE
see Figure 3-4
ALARM1
ALARM2
ALARM1
ALARM2
CURRENT %O
READING
%O2
C
°
ENTER
%O2
C
°
%O2
C
°
%O2
C
°
2
PRESSURE COMPENSATION SEQUENCE (i.e., CAL at atmospheric pressure, put sensor in different process pressure.
ENTER
ALARM1
ALARM2
ALARM1
ALARM2
%O2
Flash for 2 seconds
C
°
(Process Pressure)
%O2
C
°
ALARM1
ALARM2
%O2
C
°
Input Process Pressure in cm Hg
ENTER
ALARM1
ALARM2
ALARM1
ALARM2
%O2
Flash for 2 seconds
C
°
(Calibration Pressure)
%O2
C
°
Input Calibration Pressure in cm Hg
ENTER
ALARM1
ALARM2
%O2
C
°
Figure 4-1. CALIBRATION Mode and PRESSURE COMPENSATION Flowchart
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-3
Operator Manual
748223-K June 2002
Model 7003M
BAROMETRIC
PRESSURE
CM HG (IN. HG)
APPARENT O2 PARTIAL
PRESSURE
CM HG (DRY AIR)
O
CONCENTRATION WITH
2
NO PRESSURE CORRECTION
1
73 (28.7) 15.3 20.24% 74 (29.1) 15.5 20.52 75 (29.5) 15.7 20.80 76 (29.9) 15.8 20.95 77 (30.3) 16.1 21.36 78 (30.7) 16.3 21.63

Table 4-4. Barometric Pressure Effect on Apparent Oxygen Concentration

KEY DISPLAY WHAT HAPPENS
DISPLAY Nothing Nothing ACK Nothing Nothing ALARM 1 Nothing Nothing ALARM 2 Nothing Nothing UP ARROW DOWN ARROW
10.8 10.9
10.9 10.8
Changes pressure value
Changes pressure value CAL Nothing Nothing ENTER "CPr" / "run" Moves to calibration pressure input or goes back to RUN mode
Table 4-5. PRESSURE COMPENSATION Mode Key Functions
c. Pressure Compensation - Correction
for Constant Pressure Differences
Refer to Figure 4-1 on page 4-3 and Table 4-5 above. If the sample and calibration gases are each at constant pressure, the following procedure should be used to correct for the pressure difference between them. However, the instrument does not monitor pressure and will not correct the output if either sample or calibration gas pressure changes. (See Section 5-2 on page 5-1).
1. From RUN mode, press CAL.
2. CAL, YES, CAL flash briefly, then the oxygen concentration is displayed with the right digit flashing.
3. Press CAL again.
1
Computed using 20.946% oxygen at 760 m.
4. PPr flashes, then previous sample pressure setting is displayed with right digit flashing (76.0 cm Hg is default).
5. Use ARROW keys to enter sample pressure.
6. Press ENTER to accept the value.
7. CPr flashes, then previous span gas pressure setting is displayed with right digit flashing (76.0 cm Hg is default).
8. Use the ARROW keys to enter pressure of the calibration gas to be used. If the sample and calibration gas will always be at the same pressure, set PPr and CPr at the same value.
9. Press ENTER to accept the value. Instrument returns to RUN mode.
4-4 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M

4-3 HOLD MODE

The last measured value is displayed
Refer to Figure 4-2 on page 4-7 and Table 4-6
below. In the HOLD mode:
All instrument functions are locked
All existing values are retained
The instrument stops responding to sensor signals
KEY DISPLAY WHAT HAPPENS
DISPLAY Nothing Nothing ACK Nothing Nothing ALARM 1 Nothing Nothing ALARM 2 Nothing Nothing UP ARROW Nothing Nothing DOWN ARROW Nothing Nothing CAL Nothing Nothing ENTER "run" Drops back to RUN mode DISPLAY + ACK "d-4" "tSt" Moves to DIAGNOSTIC mode
The last output value remains at the isolated current output connections
Alarms and ON/OFF-Controllers are held in their last state
No keys are recognized except entry to the diagnostic routing.
Operator Manual
748223-K
June 2002
Table 4-6. HOLD Mode Key Functions
If the sensor cable must be disconnected, the instrument can be put in the HOLD mode, in which all existing signal levels and output current levels and output current levels are maintained. This may be useful to prevent disturbing an ON/OFF-control loop connected to the relay contacts.
1. Press the UP ARROW and DOWN ARROW at the same time to put the instrument on HOLD.
2. If the security is activated, enter the security identification (id) code and press ENTER to continue.
3. Hld, YES, flash briefly, then Hld flashes alternately with the last oxygen reading. The output signal will remain unchanged until the instrument returns to the RUN mode.
4. Press ENTER to return to the RUN mode, or continue with Section 4-4 Diagnostics Mode on page 4-8.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-5
Operator Manual
748223-K June 2002
Model 7003M
CARET
LIT
VALUE DISPLAYED
% Raw Sensor Voltage
°C
-
3
Thermistor Voltage 2.43 No
Current Output 17.43 Yes
Use ARROW keys to make adjustments.
Table 4-7. Caret Readings
TYPICAL
VALUE
2
1
ADJUSTABLE
Yes
1
These typical values were obtained under the following conditions:
O2 Concentration: 20.95% Sensor Temperature: 26.07°C Current Output Range: 4-20mA Range Endpoint, Low: 0% Range Endpoint, High: 25% Zero Offset Signal: None Pressure Compensation: None
2
Typical voltages for new sensors at 25°C:
TYPE OF SENSOR MINIMUM NOMINAL MAXIMUM
Disposable 1.05V 1.39V 1.69V Rechargeable 0.97V 1.29V 1.76V
Near the end of the life of the sensor, the sensor voltage is typically 0.53V (disposable) or 0.48V (rechargeable). At tempera-
tures (T) other than 25°C, multiply the voltages above by (1.025)exp(T-25).
3
"-" is a minus sign displayed on the left of the display between AL 1 and AL 2.
4-6 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
(
)
(
)
g
Operator Manual
748223-K
June 2002
HOLD MODE
+
SECURITY ACCESS ROUTINE
see Figure 3-4
ALARM1
ALARM2
Last displayed value held
ENTER
Sensor Voltage
Display Parameters
ALARM1
ALARM2
%O2
C
°
DIAGNOSTICS MODE
+
%O2
C
°
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
ALARM1
ALARM2
CARET LIT
see Table 4-7
DISPLAY
%O2
C
°
%O2
C
°
%O2
C
°
%O2
C
°
Adjust Parameters (see Section 4-4)
WARNING
Testin
Relays will affect devices connected to them
Exit
ALARM1
ALARM2
ALARM1
ALARM2
ALARM
1
ALARM
1
%O2
°
%O2
°
C
C
Return to HOLD
ENTER
DIAGNOSTIC Mode
ALARM1
ALARM2
ALARM1
ALARM2
ALARM
2
ALARM
2
%O2
C
°
%O2
C
°
Figure 4-2. HOLD and DIAGNOSTICS Modes Flowchart
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-7
Operator Manual
748223-K June 2002
Model 7003M

4-4 DIAGNOSTICS MODE

Refer to Figure 4-2 on page 4-7, Table 4-7 on page 4-6 and Table 4-8 below. The DIAGNOSTICS mode is used to:
Check sensor circuit voltages
Check relays
Adjust the output current
Enter DIAGNOSTICS Mode
If the instrument is already in HOLD mode, skip step 1 and proceed to step 2.
1. From the RUN mode, press the UP ARROW and DOWN ARROW at the same time to enter the HOLD mode.
2. If the security is activated, enter the security identification (id) code and press ENTER to continue.
Hld, YES, flash briefly, then Hld flashes alternately with the last oxygen reading. The output signal will remain unchanged until the instrument returns to the RUN mode.
a. Raw Sensor Voltage
1. Press DISPLAY and ACK at the same
time to enter DIAGNOSTICS mode.
Tst, d-3, d-4 flash briefly, then the %O
caret lights and the raw sensor
2
voltage is displayed.
2. Press ENTER at any time to return to
HOLD mode. Press ENTER again to return to RUN mode.
KEY DISPLAY WHAT HAPPENS
DISPLAY "2.13" Scrolls among the DIAGNOSTIC displays ACK Nothing Nothing ALARM 1 "A-1" "On"/ "OFF" Toggles the Alarm 1 relay ON/OFF ALARM 2 "A-2" "On" / "OFF" Toggles the Alarm 2 relay ON/OFF UP ARROW DOWN ARROW CAL Nothing Nothing ENTER "hLd" Drops back to HOLD mode

b. Test Current Output

The current output can be tested (increased or decreased using the ARROW keys) and the actual output will change. When the instrument returns to the RUN mode, the output will return to the previous value.
1. Press DISPLAY until the minus sign (­) is displayed on left.
"2.13" "2.14" "2.14" "2.13"
Table 4-8. DIAGNOSTICS Mode Key Functions
Increases current output or span setting Decreases current output or span setting

c. Linearizing Current Output

For some samples, current output does not track the oxygen reading displayed. The operator can linearize output and store the compensating second-order polynomials in EEPROM. This data will remain and function there as long as the EEPROM is not erased. However, if a "cold boot" is performed, the operator must repeat this linearization routine.
2. Use the ARROW keys to increase or decrease the current output.
4-8 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
To determine if current output linearization is necessary, do the following:
1. Connect a separate current meter to the output terminals (any device that will display an accurate current measurement such as a computer input will suffice).
NOTE
Do not simply connect a current measurement meter in parallel with the usual load because this will lead to in­correct readings.
2. Place the instrument in DIAGNOSTIC mode by pressing the UP and DOWN ARROWS, followed by the DISPLAY and ACK keys. The display will show consecutively:
tSt, d-3, d-4, and oxygen input voltage
3. Press the DISPLAY key until the minus sign and a number are displayed. This reading is the current output (in milliamps) as determined by the Model 7003M.
8. Use the UP and DOWN ARROW keys to adjust the actual current output (as noted on the separate meter) to the correct amount. Note that the instrument display does not change.
9. Press ENTER, the instrument displays diA, returns to the DIAGNOSTIC mode, and continues to display the current output.
10. Use the UP and DOWN ARROW keys to adjust the current output to a lower value, representing the low range (the operator may prefer to use 4 mA to represent 0% oxygen).
11. Press CAL (the instrument displays CCL for lower current output setting), and repeat steps 3 and 4 of this procedure.
12. Use the UP and DOWN ARROW keys to adjust the current output to a value higher than the initial calibration value (step 1 of this procedure), representing the higher range (the operator may prefer to set it at
16.72mA to represent 20.9% oxygen.
4. Use the UP and DOWN ARROW keys to increase the display reading, and occasionally, compare the current displayed on both meters (the instrument display and the separate meter).
5. If the two readings are too disparate (different applications require lesser or greater tolerance), do the following 3-point calibration to linearize the current output:
6. Use the UP and DOWN ARROW keys to set the current output to a value halfway up its range. The operator may prefer to set it at a critical measurement point.
7. Press the CAL key. The instrument will flash CCA, illuminate the three right-hand carets, and continue displaying the current output.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-9
13. Press CAL (the instrument displays CCH for high current output setting), and repeat steps 3 and 4 of this procedure.
To return to the RUN mode, press ENTER twice.
At any time during this procedure, the operator may press the ACK key to reset all calibration coefficients to default values. Pressing DISPLAY at any time during this procedure will cause the instrument to display for a moment the relevant calibration efficient.
The operator may escape the current CALIBRATION and DIAGNOSTIC mode altogether by pressing ENTER twice. Note, all coefficients will return to the default values, and the current calibration procedure must be started over to linearize current output.
Operator Manual
748223-K June 2002
Model 7003M
NOTE
The low calibration point must be be­tween the initial calibration point (step 1 of previous procedure) and zero, and the high calibration point must be be­tween the initial calibration point and 20mA.
If the third point calibrated (step 8, previous procedure) is actually an­other low range (between the initial calibration point and CCL) it will over­ride the original CCL but won't affect the initial calibration.
NOTE
The current output linearizing proce­dure uses second order polynomials to adjust the output linearity. The correc­tion effect is maximized at the calibra­tion point, and is smoothly diminished towards the endpoints.

d. Test Alarm Relays

1. Disconnect any alarm or ON/OFF­Control devices that are connected to the analyzer, before testing.
2. Toggle the relays ON and OFF using the ALARM 1 and ALARM 2 keys.
A-1 ON or A-2 OFF is displayed. The alarm contacts can be heard opening and closing.
3. Connect any alarm or ON/OFF­Control devices that were disconnected in step 1.
4. Press ENTER to exit the DIAGNOSTICS mode.
"Hld" flashes alternately with the last oxygen reading and the analyzer returns to HOLD mode.
5. Press ENTER again to return to RUN mode.

4-5 ADJUSTMENTS AND SETTINGS

All operator settings and adjustments are made using the front keypad except if the power jumpers are to be changed. Refer to the following sections:
2-5b Power Connections on page 2-4
NOTE
When alarms are tested manually, any alarm or ON/OFF-Control devices con­nected to the relays will respond as they would in an alarm condition. The alarm devices may have to be discon­nected from the relays for this test.
4-10 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management
3-5d Zero Offset Adjustment on page 3-15
4-4a Raw Sensor Voltage on page 4-8
4-4c Linearizing Current Output on page 4-8
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 5

THEORY

5-1 PRINCIPLES OF OPERATION

The Model 7003M Percent Oxygen Analyzer consists of an amperometric sensor and analyzer interconnected by a multi-conductor shielded cable. The sensor responds to the partial pressure of oxygen and produces a current signal. The analyzer amplifies and processes the signal, providing readout in percent oxygen and current output.

a. Electrochemical Theory

With the sensor placed in the sample gas, a voltage is applied across the cathode and anode. Sample gas diffuses through the membrane and is reduced at the cathode resulting in a current proportional to the partial pressure of oxygen in the sample.
When no oxygen is present, a small residual sensor current, which can be offset during instrument setup.

5-2 EFFECTS OF SAMPLE CONDITIONS

a. Barometric Pressure

If ambient air is used to calibrate the sensor, a change in barometric pressure can be a significant factor. At 76 cm Hg, the partial pressure of oxygen is 15.9 cm Hg (20.9% of 76 cm Hg). As atmospheric pressure deviates from 76 cm Hg, the partial pressure of oxygen will vary proportionally. The highest partial pressure that can be read by the Model 7003M is 19 cm Hg (25% oxygen at 76 cm Hg).
oxygen without affecting the total barometric pressure.
This relationship can be expressed as :
P atm = Po2 + PH20 = P other,
where:
P atm = total barometric pressure,
Po2 = partial pressure of oxygen,
PH20 = partial pressure of water vapor,
P other = partial pressure of all gases other than oxygen and water vapor

c. Sample Temperature

The rate of oxygen diffusion through the sensor membrane varies with temperature, with a coefficient of about +3% °C (5% °F), causing a corresponding change in sensor current. The instrument continuously monitors temperature with a thermistor in the sensor body, and compensates the output for both sample temperature and membrane permeability.
Because of a slight time lag in the response of the temperature sensor, a rapid temperature change will cause a temporary change in apparent oxygen level, such as might be encountered if the sensor is suddenly exposed to direct sunlight from a shadow.

b. Humidity

If calibrating with humid ambient air, the effect of the partial pressure due to water vapor content should be taken into consideration. High humidity has the effect of reducing the partial pressure of
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 5-1
Extended use of the sensor at low temperatures will shorten the life of the sensor. As the sensor reaches the end of its useful life, distinct step changes in the oxygen reading will be evident as the oxygen level changes.
Operator Manual
748223-K June 2002

d. Interfering Gases

Gases that are reduced or oxidized in the range of 0.5 to 0.8VDC contribute to sensor current and can cause a readout error. Only a few gases have this characteristic. Common gases that should be avoided include SO oxides of nitrogen (NO concentrations of hydrogen sulfide tend to contaminate the sensor, but may not seriously affect oxygen measurement.
If you suspect that any stream components are affecting the operation of the sensor, contact Rosemount Analytical Technical Services.
If contaminated, the rechargeable sensor must be rejuvenated and the disposable sensor must be replaced. Refer to the instructions supplied with the sensor.

5-3 CIRCUIT DESCRIPTIONS

a. Signal Board

The Signal Board is configured to match the type of sensor ordered with the analyzer. If the type of sensor is changed, or the Signal Board is repaired or replaced, verify the rocker switch position (SW1):
Rechargeable Sensor: Close positions 1,4, and 5
2, Cl2, and
). Low-level
X
Model 7003M
+5V (isolated) for other functions for that circuit
There are isolated and non-isolated grounds on the Power Supply Board. TP1 is the non-isolated ground and should be used with TP6 and TP7.

c. Microprocessor Board

The Microprocessor Board is the functional center of the instrument. With the exception of the program chip, no user serviceable functions are on this assembly. If the Microprocessor Board is to be repaired or replaced, note the positions of the configuration jumpers for setting jumpers on replacement board.
An instrument timer circuit monitors instrument power for a power failure, power surge or other outside condition interfering with the microprocessor operation. When conditions return to normal, the circuit will attempt a system restart. If the microprocessor is still capable of operating, both alarm relays will be put into alarm condition, the screen will clear, and the output current will drop to 0 mA. Then the instrument will go through a normal power-up routine and will eventually return to routine operation with alarms and output restored to their normal conditions. The appropriate error code (E-0, etc.) will be displayed until the ACK button is pressed.
Disposable Sensor: Close positions 2,4,

d. Alarm Circuits

and 6
Each alarm activation setpoint is designed

b. Power Supply Board

to be at a different level from the deactivation setpoint:
The Power Supply Board provides:
To prevent relay contacts from
±12V (non-isolated) for analog
signaling function
+5V (non-isolated) for microprocessor and associated logic
chattering if the oxygen level stays around the setpoint.
To indicate to the microprocessor whether alarm is HIGH or LOW.
To allow for a time lag between the time the alarm is activated and the
+24V (non-isolated) for the isolated current function
5-2 Routine Servicing and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
time it is deactivated for ON/OFF­Control purposes.
Model 7003M

ROUTINE SERVICING AND TROUBLESHOOTING

Operator Manual
748223-K
June 2002
SECTION 6
WARNING
ELECTRICAL SHOCK HAZARD
Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
Relay contacts wired to separate power sources must be disconnected before servicing.

6-1 ROUTINE SERVICING

Most service and maintenance involves cleaning and recharging the sensor. Refer to the instructions supplied with the sensor for recharging and maintenance procedures.
Service is rarely required on the Model 7003M Percent Oxygen Analyzer. Repair is limited to replacement of PCB boards, EPROM and fuses. See Section 7 Replacement Parts.

6-2 TROUBLESHOOTING

Troubleshooting should be referred to qualified service technicians. Refer to Section 5-3 Circuit Descriptions, on page 5­2, as an aide in isolating a problem.

a. Symptoms

Cannot adjust signal to calibration value
Possible cause
Sensor voltage error
Check/Test
Install know good sensor and verify voltage
Corrective Action
Recharge or replace sensor
Readings not consistent with calibra­tion curve
Possible cause
Incorrect sensor used for application
Check/Test
Verify sensor corresponds with application
Corrective Action
Install new sensor and calibrate with span gas
Possible cause
Sensor temperature incorrect
Check/Test
Sensor temperature should be 0 to 50°C
Calibration error code displayed
Possible cause
Exhausted Sensor
Check/Test
Install know good sensor and isolate
Corrective Action
Recharge or replace sensor
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing and Troubleshooting 6-1
Corrective Action
Install new sensor and calibrate with span gas
Possible cause
Incorrect span gas concentration
Check/Test
Verify span gas concentration
Corrective Action
Install new sensor and calibrate with span gas
Operator Manual
748223-K June 2002
Model 7003M
Error code E-0 displayed
Possible cause
Power interruption, power surge or other outside interference has occurred
Corrective Action
Press ACK. If this symptom occurs frequently, surge-protection is recommended.
Will not read low levels of oxygen
Possible cause
Ambient air leaking into system
Check/Test
Increase sample flow rate, block off flow of gas in and out of flow chamber, and not analyzer response
Corrective Action
If signal increases with test, isolate and seal leak. If signal does not increase, replace or recharge sensor.
Sudden change in output for no ap­parent reason
Possible cause
Sensor subjected to sudden change in conditions, as when a shadow moves and exposes the sensor to an increase in temperature
Corrective Action
Locate sensor in an area shaded from sunlight and away from cold or warm drafts.
Resolution degrading with time
Possible cause
Sensor exposed to low temperatures.
Check/Test
Per Section 4-4b on page 4-8, check current output of sensor.
If the sensor is degrading, the signal will change in steps as the oxygen level changes instead of changing continuously.
Corrective Action
Relocate sensor or replace more frequently.
6-2 Routine Servicing and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M
Operator Manual
748223-K
June 2002
SECTION 7

REPLACEMENT PARTS

7-1 CIRCUIT BOARD REPLACEMENT POLICY

In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replacement of an individual component. As standard policy, rebuilt boards are available on an exchange basis.

7-2 MATRIX

7003M MODEL 7003M OXYGEN MONITOR*
Code Power
01 115 VAC, 50/60 Hz 05 230 VAC, 50/60 Hz
Code Mounting Options
01 Panel Mount - Standard 07 Wall Mounting Plate w/ Junction Box 08 Two-Inch Pipe Mounting Bracket
Code Gain Setting
50 Signal: High Level (Used with Rechargeable Sensor % Gas) 51 Signal: Low Level (Used with Disposable Sensor % Gas) 99 Signal: Special
Because of the exchange policy covering circuit boards, the following sections on replacement parts do not include individual, circuit board mounted components. If circumstances necessitate replacement of an individual component, which can be identified by inspection or from the schematic diagrams, obtain the replacement component from a local source of supply.
7003M 01 01 50
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1
Operator Manual
748223-K June 2002

7-3 REPLACEMENT PARTS - MODEL 7003M

WARNING
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair
PART NUMBER DESCRIPTION
652630 Signal Board 652640 Power Supply Board 654605 Display/Keypad Board 655195 EPROM 655197 Microprocessor Board 777156 Fuse, 1/4 A, 120V (pkg of 5) 777360 Fuse, 1/8 A, 240 V (pkg of 5) 858728 Arc Suppressor
Model 7003M
When replacing Microprocessor Board PN 652650, replace with Microprocessor Board PN 655197 AND EPROM 655195.
7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M

7-4 REPLACEMENT PARTS - SENSORS

a. Rechargeable Sensors

Sensor:
MATERIAL PART NUMBER
Polypropylene 623371 Polypropylene 623370 (use with Fast Response Kit) Ryton 190408 Ryton 190409 (use with Fast Response Kit)
Sensor Kits (Sensor not included):
DESCRIPTION PART NUMBER USE WITH SENSOR
Submersion, Polypropylene 639904 623371 In-Line Flow, Polypropylene 639905 623371 Fast Response, Polypropylene 639906 623370 Submersion, Ryton 646628 190408 In-Line Flow, Ryton 646629 190408 Fast Response, Ryton 646630 190409
Operator Manual
748223-K
June 2002
Sensor Kits (Sensor included):
DESCRIPTION PART NUMBER Submersion, Polypropylene 400011
In-Line Flow, Polypropylene 400012 Fast Response, Polypropylene 400013 Submersion, Ryton 400021 In-Line Flow, Ryton 400022 Fast Response, Ryton 400023
Recharge Kit
Recharge Kit (10 recharges) 191755
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-3
Operator Manual
748223-K June 2002

b. Disposable Sensors

Sensor:
MATERIAL PART NUMBER
Polypropylene 623742
Sensor Kits (Sensor not included):
DESCRIPTION PART NUMBER USE WITH SENSOR
Submersion, Polypropylene 623715 623742 In-Line Flow, Polypropylene 623716 623742
Sensor Kits (Sensor included):
DESCRIPTION PART NUMBER
Submersion, Polypropylene 500011 In-Line Flow, Polypropylene 500012
Model 7003M
7-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003M

8-1 RETURN OF MATERIAL

If factory repair of defective equipment is required, proceed as follows:
1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
Rosemount CSC will provide the shipping address for your instrument.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
2. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
3. In a cover letter, describe completely:
The symptoms that determined the equipment is faulty.
The environment in which the equipment was operating (housing, weather, vibration, dust, etc.).
Site from where the equipment was removed.
Whether warranty or non-warranty service is expected.
Complete shipping instructions for the return of the equipment.
4. Enclose a cover letter and purchase order
and ship the defective equipment accord­ing to instructions provided in the Rose­mount Return Authorization, prepaid, to the address provided by Rosemount CSC.
Operator Manual
748223-K
June 2002
SECTION 8

RETURN OF MATERIAL

If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount’s option, and an operating unit will be returned to the customer in accordance with the shipping instructions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

8-2 CUSTOMER SERVICE

For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076

8-3 TRAINING

A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact the Technical Services Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 8-1
Operator Manual
748223-K June 2002
Model 7003M
8-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the com­mercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WAR­RANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DE­LAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WAR­RANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF AC­TION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDEN­TAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVE­NUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
Operator Manual
748223-K June 2002
Model 7003M
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
© Rosemount Analytical Inc. 2001
Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...