Drager Konica Minolta JM-103 User manual

Page 1
Konica Minolta/Dräger Medical
WARNING:
For a full understanding of the performance characteristics of this equipment, the user should carefully read this manual before operating.
Model JM-103 Operating Instructions
Page 2
Page 3
PROPRIETARY AND CONFIDENTIAL DRAFT 9 Nov 04

Table of Contents

Section 1: Symbol Definition and Intended Use
Symbol Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Section 2: Introduction, Features, and Specifications
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Measuring Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Explanation of the Test . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 2
Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 5
Controls, Indicators, and Connections . . . . . . . . . . . . . . . .2 - 5
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 7
Standard Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 8
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 9
Standard Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 9
Regulations, Standards, and Codes . . . . . . . . . . . . . . . . . .2 - 10
Device Classification. . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 10
Electromagnetic Compatibility (EMC) Guidance and
Manufacturer’s Declarations . . . . . . . . . . . . . . . . . . . . . . .2 - 11
Section 3: Precautions and Safety Tips
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Electromagnetic Compatibility Precautions . . . . . . . . . . . .3 - 3
Safety Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 4
Warning and Caution Labels . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 5
Section 4: Installation and Assembly
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1
Charging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1
Selecting the Unit of Measurement. . . . . . . . . . . . . . . . . . .4 - 3
Operational Checkout of the Jaundice Meter . . . . . . . . . . .4 - 4
Section 5: Instructions for Use
Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
Taking Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
Jaundice Meter (Model JM-103) User Manual (usr070) Page i
Page 4
Setting the Number of Average Measurements . . . . . . . . . 5 - 5
Taking Average Measurements . . . . . . . . . . . . . . . . . . . . . 5 - 6
Section 6: Cleaning, Maintenance, Replacement Parts, and Storage and Handling
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Steam Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Cleaning Difficult to Access Areas . . . . . . . . . . . . . . . . . . 6 - 1
Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Storage and Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Section 7: Troubleshooting
Service Calls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
Appendix A: Clinical Performance Summary
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 1
Study Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 1
Performance Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 2
Appendix B: Doctors’ Office Data
Study Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B - 1
Performance Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B - 3
Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B - 8
Appendix C: Medical and Scientific References on Transcutaneous Bilirubinometry (1979 - 1997)
Page ii Jaundice Meter (Model JM-103) User Manual (usr070)
Page 5
DRAFT 18 May 2005
Section 1

Symbol Definition

and Intended Use
Symbol Definition
This manual contains different typefaces and icons designed to improve readability and increase understanding of its content. Note the following examples:
• Standard text—used for regular information.
Boldface text—emphasizes a word or phrase.
NOTE:—sets apart special information or important instruction clarification.
• The symbol below highlights a WARNING or CAUTION:
Warning and Caution
A WARNING identifies situations or actions that may affect
patient or user safety. Disregarding a warning could result in patient or user injury.
A CAUTION points out special procedures or precautions that
personnel must follow to avoid equipment damage.
• The symbol below highlights a type BF applied part:
Type BF Applied Part
The instrument provides a specified degree of protection
against electric shock, particularly the leakage current and reliability of the protective ground connection with an F-type applied part. An F-type applied part indicates an applied part isolated from all other parts of the instrument to such a degree that the patient leakage current allowable in a single-fault condition is not exceeded when a voltage equal to 1.1 times the highest-rated mains voltage is applied between the applied part and ground.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 1 - 1
Page 6
DRAFT 18 May 2005
• The symbol below highlights an ELECTRICAL SHOCK HAZARD WARNING:
Electrical Shock Hazard Warning
• The symbol below indicates INPUT RATING:
Input Rating Symbol
• The symbol below indicates that the product uses a RECHARGEABLE BATTERY:
Rechargeable Battery Symbol
• The symbol below indicates RESET:
RESET Button Symbol
• The symbol below, when applied to the device, indicates:
ATTENTION: Consult Accompanying Documents
Page 1 - 2 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 7
DRAFT 18 May 2005

Intended Use

CAUTION:
Magnetic Resonance Imaging (MRI) procedures interfere with Jaundice Meter operation. Inaccurate readings could occur.
CAUTION:
Do not use a mobile telephone when using the Jaundice Meter. A measurement error could occur.

Intended Use of the Jaundice Meter

The Jaundice Meter (JM-103) is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue in newborn infants.
The device is intended for use in hospitals or doctors’ offices under a physician’s supervision or at their direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical signs and laboratory measurements.
Newborn infants whose Jaundice Meter (JM-103) test results are indicative of hyperbilirubinemia should be evaluated by their physician(s) for appropriate patient management. Specific neonatal patient bilirubin levels should be confirmed by other methods, such as serum bilirubin, prior to treatment determinations.
The Jaundice Meter (JM-103) is not intended for home use.

Limitations (Doctors’ Office Use)

Use only on infants up to 14 days of age.
For doctors’ office application, use only the sternum location when taking measurements.
Please be aware, performance in doctors’ offices may vary from performance in hospitals.
Precocious Jaundice
Do not use this device on infants with precocious jaundice. If there is a possibility that the infant is suffering from precocious jaundice, as a result of an incompatible blood type or hemolytic jaundice, it is recommended that the total serum bilirubin be measured.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 1 - 3
Page 8
DRAFT 18 May 2005

Intended Use of the User Manual

This manual provides instructions for installation, use, operator maintenance, and troubleshooting of the Jaundice Meter (JM-103). Konica Minolta/Dräger Medical cannot be responsible for the performance of the Jaundice Meter if the user does not operate the unit in accordance with the instructions, fails to follow maintenance recommendations, or makes any repairs with unauthorized components. Only qualified service personnel should perform repair.
Page 1 - 4 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 9
DRAFT 20 June 2005
Section 2
Introduction, Features,
and Specifications

Introduction

To prevent kernicterus in newborn infants, it is very important to detect jaundice in its early stages. The Jaundice Meter (JM-103) is a non­invasive transcutaneous bilirubinometer. This hand-held device allows a quick, non-invasive estimate of bilirubin concentration, to be used as an aid for the management of jaundice in newborn infants. The measurements are taken automatically when you place the instrument’s measuring probe against the measuring site of the infant and press it gently; the measured value is then displayed.

Measuring Point

Measurements must be taken only on the infant’s forehead (at hospital sites only) or sternum (at hospital sites or physicians’ offices) where a sufficient amount of blood is circulated. A possibility exists that the bilirubin in the subcutaneous tissue may measure low for areas with minimal blood flow or areas in which the subcutaneous tissue is subject to keratinization.
Although correlation with serum bilirubin was observed for both forehead and sternum measurements, the clinical studies performed with the Jaundice Meter (JM-103) show consistently better results with measurements taken at the sternum versus the forehead. There is a possibility that this difference may be more pronounced for infants that have been exposed to sunlight, such as infants seen at doctors’ offices. Only sternum measurements were evaluated during the studies conducted at doctors’ offices; correlation of forehead measurements with serum bilirubin has not been evaluated, and the device is not intended for forehead measurements at doctors’ offices. Therefore, use
the sternum location when taking measurements at doctors’ offices.
SPECIFICATIONS
Phototherapy
WARNING:
Do not use the Jaundice Meter after initiation of phototherapy or after an exchange transfusion as results may be inaccurate under these conditions. Patient injury could occur.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 1
Page 10
DRAFT 20 June 2005

Explanation of the Test

Measuring Principle
The Jaundice Meter determines the yellowness of an infant’s subcutaneous tissue by measuring the difference in the optical densities for light in the blue (450 nm) and green (550 nm) wavelength regions. The measuring probe has two optical paths. This method allows for a more precise measurement of yellowness in an infant’s subcutaneous tissue by minimizing the influences of the melanin pigment and the skin maturity.
When the measuring probe is pressed against the forehead or sternum of the infant, the built-in xenon lamp flashes. The light from the xenon lamp passes through the glass fiber and illuminates the skin. The light scatters and is absorbed in the skin and subcutaneous tissue repeatedly, and then finally returns to the sensor side of the glass fiber. Of the light that returns, the part scattered from the shallow areas of the subcutaneous tissue passes through the inner core, or short-optical path, of the fiber. The part scattered from the deep areas of the subcutaneous tissue passes through the outer core, or long-optical path, and then reaches its corresponding photodiode.
Page 2 - 2 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 11
DRAFT 20 June 2005
By calculating the difference in the optical densities, the parts that are common to the epidermis and dermis are deducted, and as a result, the difference in the optical densities between the two wavelength regions can be obtained for the subcutaneous tissue only. Since the optical density difference shows a linear correlation with the total serum bilirubin concentration, it is converted to the estimated bilirubin concentration and is indicated digitally.
The Jaundice Meter (JM-103) device software uses a correlation coefficient to convert the measurement difference from the dual optical path to an estimated bilirubin concentration. The calculation formula used includes the correlation coefficients α and γ. These coefficients were determined in pre-clinical testing. The equation used is as follows:
J
= α(L-S) + γ
sample
Where L and S are the long and short optical path measurements.
Use of the Device
Patient Population
The Jaundice Meter (JM-103) is indicated for use in neonatal patients born >35 weeks gestation who have not undergone transfusion or phototherapy treatment.
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 3
Page 12
DRAFT 20 June 2005
Averaging of Measurements
Averaging measurements may allow for more precise results. Assess the advantages of using average measurements at your facility. There was no significant difference between the averaged measurements and the single measurement approaches in the largest study for sternum measurements. The mean of three measurements showed the highest degree of correlation (r=0.965), however, the difference was minimal with a single measurement (r=0.963).
Averaging was not evaluated in the physicians’ office study.
Action Levels
Each facility should determine their own action levels based on studies of performance of the device on their population. Appropriate action levels may vary depending on performance of the device, such as precision or correlation with serum bilirubin, in the hands of the user. Some factors that could affect performance of the device or appropriate action levels include skin color, age, or measurement site. The bias relative to serum bilirubin differs between hospital versus physicians’ office sites (see Appendixes A and B). Different action levels may be appropriate for hospital versus physicians’ offices. Careful selection of
action levels should be made so that false negatives do not prevent appropriate follow up measures.
Calibration
There is no user calibration of this device. The system does include a checker that measures the intensity of light from the device to ensure the light output is acceptable for proper use.
Processing of Measured Values
The Jaundice Meter (JM-103) determines the yellowness of the subcutaneous tissue by measuring the difference in the optical densities for light in the blue and green wavelength regions. The optical density difference has been shown to have a linear correlation with serum bilirubin concentration. The device computes an estimated bilirubin concentration based on this linear correlation and provides the value on the display.
Page 2 - 4 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 13
DRAFT 20 June 2005

Features

Controls, Indicators, and Connections

SPECIFICATIONS
Controls, Indicators, and Connections
Name Function
A Power switch Turns the Jaundice Meter on and off.
When used with the Reset button, the device switches to Check Mode and changes the unit of measurement.
B Ready lamp Illuminates to indicate that the Jaundice Meter
is ready for the next measurement.
C Display Displays the measured value.
D Measuring probe Takes the measurement when pressed against
the measuring point.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 5
Page 14
DRAFT 20 June 2005
Name Function
E Charger section Connect the charger unit to the charger sec-
tion.
F Reset button Deletes the currently displayed measured
value and prepares for the next measurement. When used with the Power switch, the device switches to Check Mode and changes the unit of measurement.
G Strap attachment area Is where the strap attaches.
H Checker cover Open this checker cover to check the Jaundice
Meter.
I Charger lamp Illuminates to indicate that the Jaundice Meter
is charging.
J DC jack Connect the AC adapter to this jack.
K Charger jack Connect the main body to this jack.
L Checker Checks for the intensity of light output by tak-
ing measurements in Check Mode.
M Standard checker
values
N DC plug Connect the charger’s DC jack to this.
O DC Plug (interna-
tional)
For reference.
Connect the charger’s DC jack to this.
Page 2 - 6 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 15
DRAFT 20 June 2005

Display

Name Function
A AVG Illuminates during averaging measurement.
B Optical path indicator
(•)
C Value Displays the measured value.
D Unit of measurement Displays the unit of measurement in either
E Battery indicator When the battery power is low, the battery
When verifying light output with the checker, (•)illuminates when the L-value appears and extinguishes when the S-value appears.
NOTE: When the measured value is greater than 20 mg/dl, the display shows “---” and the physician should be contacted.
milligrams per deciliter (mg/dL) or micro­moles of solute per liter (µmol/L)
indicator blinks. Charge the battery as soon as possible (see “Charging the Battery” on page 4-1). If only the battery indicator lights, the battery has run out. Go to “Charging the Battery” on page 4-1. If the power is on and the display is blank, the battery is completely exhausted. Go to “Charging the Battery” on page 4-1.
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 7
Page 16
DRAFT 20 June 2005

Standard Features

• Jaundice Meter (JM-103)
• Charger unit (Model JM-A30) with a checker
• AC adapter (Model JM-A31)
• Carrying case and wrist strap
Page 2 - 8 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 17
DRAFT 20 June 2005

Specifications

Standard Features

Feature Dimension
Model name JM-103
Measuring method Determines the yellowness of the
subcutaneous tissue by using two optical paths to measure the optical density difference at two wave­lengths
Measurement range 0.0 mg/dL to 20 mg/dL or 0 µmol/L
to 340 µmol/L
Clinical Data Standard Error of Estimate (SEE) *
Light source Pulse xenon arc lamp
Light source life 150000 measurements
Sensors Silicon photodiodes
Power source 2.4 V, Special Ni-MH battery
Protection type and level Internally-powered instrument, BF-
Minimum number of measurements when fully charged
Operating temperature range 10°C (50°F) to 40°C (104°F)
Operating relative humidity range 30% to 95% non-condensing
Storage temperature range -10°C (14°F) to 50°C (122°F)
Storage relative humidity range 30% to 95% non-condensing
Dimensions 48 mm (1.9") wide x 15.5 cm (6.1")
Weight, including battery 150 g (5.3 oz)
AC adapter input for North America 120V, 50/60 Hz, 10 W
AC adapter input for international use (old)
AC adapter input for international use (new)
AC adapter output 9V, 500 mA, 4.5W
± 1.5 mg/dL or ± 25.5 µmol/L
type
400 single measurements
high x 33 mm (1.3") deep
200V-240V, 50/60 Hz, 12.5VA
100V-240V, 50/60 Hz, 0.14-0.07A
SPECIFICATIONS
*The standard deviation shown above is based on the average of the
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 9
Page 18
DRAFT 20 June 2005
clinical data available. On average, 66% of results fall within this range, and the remainder fall outside this range. This value can be affected by variables such as age, skin color, and preformance of the device in the hands of the user. Refer to Appendixes A and B for a detailed description of results by clinical site, measurement location, and patient demographics. The SEE shown above are based on the clinical data available and can be affected by variables such as infant developmental age, ethnicity, etc. Therefore, we recommend that the JM-103 be used in conjunction with other clinical signs and laboratory measurements. "Specific Bilirubin Measurement" should be confirmed by other methods such as laboratory blood serum analysis.

Regulations, Standards, and Codes

In North America With respect to electrical shock, fire, and mechanical hazards only, this instrument complies with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
In Europe, this instrument complies with EN60601-1, EN60601-1-2, and EN ISO13485, and EN ISO14971.
Directive 2002/96/EC of the European Parliament and of the Council of
2003-01-27 on Waste Electrical and Electronic Equipment (WEEE) Annex IV, prEN 50419 applies.

Device Classification

The Jaundice Meter (JM-103) meets the requirements for the following classifications:
• Protection against electrical shock: Internally powered
• Type of applied part: BF
• Degree of protection against harmful ingress of water: Not applicable
• Not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
• Mode of operation of equipment: Continuous while in use (IEC 60601-1)
Page 2 - 10 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 19
DRAFT 20 June 2005

Electromagnetic Compatibility (EMC) Guidance and Manufacturer’s Declarations

Guidance and Manufacturer’s Declaration—Electromagnetic
Emissions
The Jaundice Meter (JM-103) is intended for use in the electromag­netic environment specified below. The customer or user of the unit should ensure that the unit is used in such an environment.
Emissions Test Compliance
Radio frequency
Group 1 The Jaundice Meter (JM-103) (RF) emissions —CISPR 11
Electromagnetic
Environment—Guidance
uses RF energy only for its inter­nal function. Therefore, its RF emissions are very low and are not likely to cause interference with nearby electronic equip­ment.
RF emissions— CISPR 11
Harmonic Emis­sions—IEC 61000-3-2
Voltage fluctua­tions/ flicker
Class B The Jaundice Meter (JM-103) is
suitable for use in all establish-
Class A
ments, including domestic and those directly connected to the public low-voltage power supply
Complies
network that supplies buildings used for domestic purposes.
emissions—IEC 61000-3-3
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 11
Page 20
DRAFT 20 June 2005
Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Jaundice Meter (JM-103) is intended for use in the electromag­netic environment specified below. The customer or user of the unit should ensure that the unit is used in such an environment.
Immunity
Test
Electrostatic discharge (ESD)— IEC 61000­4-2
Electrical fast tran­sient/burst —IEC 61000-4-4
Surge—IEC 61000-4-5
IEC 60601
Test L e v e l
± 6 kV con­tact ± 8 kV air
± 2 kV for power sup­ply lines
± 1 kV differential mode
Compliance
Level
± 6 kV con­tact ± 8 kV air
± 2 kV for power sup­ply lines
± 1 kV differential mode
Electromagnetic
Environment—
Guidance
The floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Page 2 - 12 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 21
DRAFT 20 June 2005
Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Jaundice Meter (JM-103) is intended for use in the electromag­netic environment specified below. The customer or user of the unit should ensure that the unit is used in such an environment.
Immunity
Test
Voltage dips, short interrup­tions, and voltage vari­ations on power sup­ply input lines—IEC 61000-4-11
Power fre­quency (50/60 Hz) magnetic field—IEC 61000-4-8
IEC 60601 Test Level
< 5% U
95% dip in
) for 0.5
U
T
cycles 40% U
(60% dip in
U
) for 5
T
cycles 70% U
(30% dip in
U
) for 25
T
cycles < 5% U
95% dip in
U
) for 5
T
seconds
(>
T
T
T
(>
T
Compliance
Level
< 5% U
T
(>
95% dip in
) for 0.5
U
T
cycles 40% U
T
(60% dip in
U
) for 5
T
cycles 70% U
T
(30% dip in
U
) for 25
T
cycles < 5% U
T
(>
95% dip in
U
) for 5
T
seconds
3 A/m 3 A/m The power frequency
Electromagnetic
Environment—
Guidance
Mains power quality should be that of a typical commercial or hospital environment. If the user of the unit requires con­tinued operation during power mains interrup­tions, it is recommended that the unit be powered from an uninterruptable power supply or battery.
magnetic fields should be at levels characteristic of a typical location in a typ­ical commercial or hospi­tal environment.
SPECIFICATIONS
NOTE:
U
is the AC mains voltage prior to the application of the test
T
level.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 13
Page 22
DRAFT 20 June 2005
Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Jaundice Meter (JM-103) is intended for use in the electromag­netic environment specified below. The customer or user of the unit should ensure that the unit is used in such an environment.
Immunity
Test
Con­ducted RF—IEC 61000-4-6
Radiated RF—IEC 61000-4-3
IEC
60601
Test
Level
3 Vrms 150 kHz to 80 MHz
3 V/m 3 V/m
80 MHz to 2.5 GHz
Compli
ance
Level
Electromagnetic Environment—
Guidance
Recommended Separation Distance
3 Vrms Portable and mobile RF communica-
tion equipment should be used no closer to any part of the Jaundice Meter (JM-103), including cables, than the recommended separation distance cal­culated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance
d 1.2 P=
d 2.3 P=
80 MHz to 800 MHz
800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufac­turer and d is the recommended separa­tion distance in meters (m). Field strengths from fixed RF transmit­ters, as determined by an electromag-
netic site survey the compliance level in each frequency
b
range.
a
, should be less than
Interference may occur in the vicinity of equipment marked with the follow­ing symbol:
Page 2 - 14 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 23
DRAFT 20 June 2005
Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Jaundice Meter (JM-103) is intended for use in the electromag­netic environment specified below. The customer or user of the unit should ensure that the unit is used in such an environment.
IEC
Immunity
Test
60601
Test
Level
NOTE:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
a. Field strengths from fixed transmitters, such as base stations for radio,
cellular/cordless telephones, land-mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed-RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the unit is used exceeds the applicable RF compliance level, observe the unit to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the unit.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be < 3
V/m.
Compli
ance
Level
Electromagnetic Environment—
Guidance
Recommended Separation Distance
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070) Page 2 - 15
Page 24
NOTES:
DRAFT 20 June 2005
Page 2 - 16 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 25
DRAFT 18 May 2005
Section 3
Precautions and Safety Tips

Precautions

WARNING:
Do not use the instrument in areas where flammable or combustible gases, such as anesthetic gases, are present. Doing so could result in a fire. Personal injury or equipment damage could occur.
WARNING:
If the instrument, the charger unit, or the AC adapter are damaged, or if smoke or an odd smell occurs, do not use the instrument, the charger unit, or the AC adapter. In such situations, immediately turn off the instrument, unplug the AC adapter from its power source, and contact the nearest authorized service facility. Failure to do so could result in fire, personal injury, or equipment damage.
WARNING:
Do not use the Jaundice Meter after initiation of phototherapy or after an exchange transfusion as results may be inaccurate under these conditions. Patient injury could occur.
SHOCK HAZARD:
Always plug the instrument into an AC outlet of the correctly rated voltage and frequency. Failure to do so could result in fire, personal injury, or equipment damage.
SHOCK HAZARD:
Do not disassemble or modify the instrument, the charger unit, or the AC adapter. Fire, personal injury, or equipment damage could occur.
CAUTION:
Do not place the instrument on an unstable or sloping surface. The instrument or charger unit could drop or overturn. Equipment damage could occur.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 3 - 1
Page 26
DRAFT 18 May 2005
CAUTION:
Do not use the instrument in direct sunlight. Equipment damage could occur.
CAUTION:
The Jaundice Meter is a precision instrument. When using it, do not drop it, expose it to shocks or strong vibrations, or place heavy objects on it. Equipment damage could occur.
CAUTION:
Do not allow blood or other liquids to come in contact with the instrument. Should blood or other liquids come in contact with the instrument, immediately clean the instrument (see “Cleaning” on page 6-1). Failure to do so could result in equipment damage.
CAUTION:
The instrument has a built-in, non-user-replaceable battery. Do not disassemble the instrument to replace the battery. To replace the battery, contact your dealer or authorized service center. Failure to do so could result in equipment damage.
Page 3 - 2 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 27
DRAFT 18 May 2005

Electromagnetic Compatibility Precautions

General information on electromagnetic compatibility (EMC) according to the international EMC standard IEC 60601-1­2: 2001
Pins of connectors identified with the ESD warning symbol shall not be touched and not be connected unless ESD precautionary procedures are used. Such precautionary procedures may include antistatic clothing and shoes, the touch of a ground stud before and during connecting the pins or the use of electrically isolating and antistatic gloves. All staff involved in the above shall receive instruction in these procedures.
NOTE:
Portable and mobile RF communications equipment can affect medical electrical equipment.
NOTE:
Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in the technical documentation available from Dräger Service upon request.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 3 - 3
Page 28
DRAFT 18 May 2005

Safety Tips

WARNING:
This instrument emits intense light to take its measurements. Take measurements only at the forehead or sternum (hospital application). Doctors’ office use should be performed at the sternum only. Do not press the measuring probe when it is directed toward the infant’s or caregiver’s eyes. Damage to the eyes could occur.
WARNING:
Before use, clean the measuring probe by wiping it with medicinal alcohol. Failure to do so could result in the spread of infection or infant injury.
WARNING:
The charger unit (JM-A30) and the AC adapter (JM-A31) are solely designed for use with the Jaundice Meter (JM-103). Use them only when charging the instrument. Using them to charge other equipment could result in personal injury or equipment damage.
WARNING:
Only facility-authorized personnel should troubleshoot the Jaundice Meter. Troubleshooting by unauthorized personnel could result in personal injury or equipment damage.
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could result in personal injury or equipment damage.
WARNING:
This product has been validated with the accessories and options listed in this manual and found to comply with all relevant safety and performance requirements applicable to the device. It is therefore the responsibility of that person or organization who makes an unauthorized modification, or incorporates an unapproved attachment to the device, to ensure that the system still complies with those requirements. [IHA036]
Page 3 - 4 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 29
DRAFT 18 May 2005
SHOCK HAZARD:
Do not plug or unplug the AC power cord’s plug with wet hands. Personal injury or equipment damage could occur.
SHOCK HAZARD:
Before cleaning, maintenance, or parts replacement, unplug the charger unit from its power source. Failure to do so could result in personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for liquid pooling. Personal injury or equipment damage could occur.
CAUTION:
Do not use harsh cleansers/detergents, such as scouring pads and heavy duty grease removers, or solvents, such as toluene, xylene, and acetone. Equipment damage could occur.

Warning and Caution Labels

Jaundice Meter (Model JM-103) User Manual (usr070) Page 3 - 5
Page 30
NOTES:
DRAFT 18 May 2005
Page 3 - 6 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 31
DRAFT 20 June 2005
Section 4
Installation and Assembly

Installation

Before using the instrument, charge and inspect the instrument.

Charging the Battery

When using the instrument for the first time, ensure that it is fully charged. To maintain a full charge at all times, place the instrument on the charger unit when it is not being used for measurements. When the battery power is low, the Battery display blinks.
If the Jaundice Meter is left uncharged for a long period of time, the power of the battery diminishes; ensure that it is charged prior to use. To charge the Jaundice Meter, perform the following:
WARNING:
The charger unit (Model JM-A30) and the AC adapter (Model JM­A31) are solely designed for use with the Jaundice Meter (JM-
103). Use them only when charging the instrument. Using them to charge other equipment could result in personal injury or equipment damage.
1. Plug the AC adapter into the DC jack of the charger unit. Use only the charger unit and AC adapter supplied with the Jaundice Meter.
NOTE:
The shape of the AC adapter varies according to region.
NOTE:
120V model shown.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 4 - 1
Page 32
DRAFT 20 June 2005
SHOCK HAZARD:
Do not plug or unplug the AC power cord’s plug with wet hands. Personal injury or equipment damage could occur.
2. Plug the AC adapter’s plug into an AC outlet. Never do so with wet hands.
CAUTION:
Unit must be placed in charger as shown, or damage to charging contacts could occur.
3. Place the Jaundice Meter on the charger unit so that its display faces you. When the Jaundice Meter is set
on the charger unit properly, the
Charger lamp lights up.
NOTE:
With a fully charged battery, approximately 400 measurements can be taken.
4. Allow approximately 32 hours for charging to complete.
CAUTION:
The instrument has a built-in, non-user-replaceable battery. Do not disassemble the instrument to replace the battery. To replace the battery, contact your dealer or authorized service center. Failure to do so could result in equipment damage.
5. To replace the battery, contact your dealer or an authorized service facility.
Page 4 - 2 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 33
DRAFT 20 June 2005

Selecting the Unit of Measurement

1. Hold down the Reset button, and turn on the Power switch.Do not release the Reset button.
2. While continuing to press the Reset button, allow approximately 15 seconds for the unit of measurement to switch from mg/dL to µmol/L, or vice versa.
3. Visually ensure that the unit’s display has changed.
4. Release the Reset button when unit displays an appropriate unit of measurement. The Ready lamp lights up, indicating that the instrument is ready to take a measurement.
5. To change the unit of measurement once more, turn off the Power switch, and repeat step 1.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 4 - 3
Page 34
DRAFT 20 June 2005

Operational Checkout of the Jaundice Meter

Do not touch the checker’s surface in the charging base with your fingers. If the checker gets dirty, wipe it with a soft cloth dampened with alcohol, and then wipe it with a dry cloth. Although the JM-103 device can be damamged as stated on pages 3-1 and 3-4, we recommend the unit always be verified to be operating properly by following the steps as stated in sections 4 and 5. Always remove the unit from service if there are concerns regarding the unit’s performance, and immediately contact your Dräger Medical representative.
For quality control purposes, periodically compare the JM-103 to serum bilirubin results. This checks that the instrument maintains consistent performance over time and that the operators are using the instrument properly.
Using the checker supplied with the charger unit, check the instrument to verify that the meter light output is within range for both long and short wavelengths of light. The labeling on the inside cover of the checker will state the acceptance ranges for both long and short wavelengths. The procedure to verify is as follows:
1. Hold the Reset button down, and set the Power switch to the On position.
NOTE:
If the Reset button is held down for longer than 15 seconds, the unit of measurement switches.
2. After CHE appears on the display window, immediately release the Reset button. If the Reset button is held down for longer than 15 seconds, switch the unit of measurement back to its previous setting (see “Selecting the Unit of Measurement” on page 4-3).
3. Visually confirm that CHE appears in the display and that the Ready lamp illuminates.
Page 4 - 4 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 35
DRAFT 20 June 2005
4. Open the cover of the checker. Use only the checker supplied with the Jaundice Meter (JM-103).
5. Place the measuring probe perpendicular to the checker, and push down gently until a click sounds.
6. If the measuring probe contacts the checker at an angle, place it perpendicular, and take the measurement again.
NOTE:
The display interchanges between the L­value, the measured value of the long­optical path, and the S-value, the measured value of the short-optical path. When the L­value is displayed, “•” appears in the upper left-hand corner of the display.
7. Confirm the measured value. If both the L-value and the S-value are within ± 1.0 of the reference values indicated on the checker cover, the unit is acceptable for use.
8. If the measured value exceeds ± 1.0 of the reference value, perform the following:
Jaundice Meter (Model JM-103) User Manual (usr070) Page 4 - 5
Page 36
DRAFT 20 June 2005
a. Clean both the checker and the measuring probe.
b. Place the measuring probe perpendicular to the checker, and
push down gently until a click sounds.
c. If the measured value still exceeds ± 1.0 of the reference value,
contact the nearest authorized service facility, and take the unit out of service.
9. Close the cover of the checker.
10. Set the Power switch to the Off position.
Page 4 - 6 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 37
DRAFT 20 June 2005
Section 5

Instructions for Use

Instructions for Use

Taking Measurements

1. Remove the Jaundice Meter from the charger unit.
NOTE:
Check the light output of the device at least once each shift (refer to “Operational Checkout of the Jaundice Meter” on page 4-4). Proper light output is one factor that affects meter performance. Light output must be within the range shown on the inside cover of the checker to obtain reliable measurements.
CAUTION:
Unit must be placed in charger as shown, or damage to charging contacts could occur.
WARNING:
Before use, clean the measuring probe by wiping it with medicinal alcohol. Failure to do so could result in the spread of infection or infant injury.
2. Using medicinal alcohol, clean the measuring probe.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 5 - 1
Page 38
DRAFT 20 June 2005
3. Set the Power switch to the On position. The measured value for a single measurement, n-1, appears on the display.
NOTE: The JM-103 is a screening device for Bilirubin in the tissues and should never be used as a stand-alone measurement. Always contact the physician whenever the JM-103 measurements present a concern. The physician will make the determination regarding the laboratory blood serum TSB, as well as other clinical considerations, as these are appropriate measurement for Bilirubin concentration as stated in the indications for use.
4. Visually check that the Ready lamp illuminates.
5. If the battery indicator blinks, charge the battery (see “Charging the Battery” on page 4-1).
WARNING:
This instrument emits intense light to take its measurements. Take measurements only at the forehead or sternum (hospital application). Doctors’ office use should be performed at the sternum only. Do not press the measuring probe when it is directed toward the infant’s or caregiver’s eyes. Damage to the eyes could occur.
6. Perform the following:
a. Place the measuring probe
vertically against the infant’s forehead (at hospital sites only) or sternum (at hospital sites or physicians’ offices). Avoid any bruises or discolored areas of the skin.
Page 5 - 2 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 39
DRAFT 20 June 2005
NOTE: Only use the sternum location for JM-103
measurements when the device is used in a doctor’s office.
b. Push the measuring probe gently
until a click sounds. The instrument’s xenon lamp flashes momentarily, and the measured value appears on the display.
c. If the measured value is outside the
measurement range of 20 mg/dL or 340 µmol/L, the display shows “---” and user should contact the physician.
NOTE:
If the instrument is not operated for more than 60 seconds, the backlight on the display goes out.
7. To take another measurement, press the Reset button, and continue from step 4.
8. To stop measuring, perform the following:
a. Set the Power switch to the Off position.
b. Using medicinal alcohol, clean the measuring probe.
c. Place the Jaundice Meter on the charger unit. When the
Jaundice Meter is not in use, keep it on the charger unit.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 5 - 3
Page 40
DRAFT 20 June 2005
NOTE:
If the caregiver has any concerns regarding the values being presented by the JM 103 for Bilirubin concentration, the physician should be notified immediately for a determination as to whether supportive laboratory blood serum analysis is required or infant retest is appropriate.
Page 5 - 4 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 41
DRAFT 20 June 2005

Setting the Number of Average Measurements

1. Set the Power switch to On or press the Reset button to prepare the instrument for measurement.
n-1, n-2, and so on (up to n-5) will appear.
2. Press the Reset button for 5 s. The number of average measurements will switch as follows:
3. Release the Reset button when the required number of average measurements is displayed.
•If n-2 through n-5 is selected, AV G appears in the upper left corner of the display.
The selected number of average measurements will be recorded.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 5 - 5
Page 42
DRAFT 20 June 2005

Taking Average Measurements

1. Set the number of average measurements needed.
2. Ensure that the Ready lamp is illuminated once n-5 appears.
NOTE: n-5 is used in the steps below as an
example. The number of measurements you require and set is the number that should appear in the display.
NOTE:
Each measurement must be taken individually by the user. The first measurement is complete when the measuring probe is pressed against the patient and the unit clicks. The probe must then be lifted from the patient and reapplied for the total number of measurements selected (in this example 5 measurements).
WARNING:
This instrument emits intense light to take its measurements. Take measurements only at the forehead or sternum (hospital application). Doctors’ office use should be performed at the sternum only. Do not press the measuring probe when it is directed toward the infant’s or caregiver’s eyes. Damage to the eyes could occur.
NOTE:
It is recommended that all of the measurements taken for averaging be taken from the same measuring point, forehead (at hospital sites only) or sternum (at hospital sites or physicians’ offices).
3. Place the measuring probe vertically on the measuring point, and then apply gentle pressure until the probe clicks.
The measurement will be taken, and the number of remaining measurements will be displayed.
Page 5 - 6 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 43
DRAFT 20 June 2005
4. While ensuring that the Ready lamp is illuminated, repeat the measuring until the number of remaining measurements is 0.
When the remaining number of measurements is completed, the average of the measured values appears in the display.
If the instrument is left without the preset number of measurements taken, the setting will be canceled without the measurement value being displayed. To set the average measurements again, press the Reset button and reset the number of average measurements needed.
To change the number of average measurements, refer to “Setting the Number of Average Measurements” on page 5-5.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 5 - 7
Page 44
NOTES:
DRAFT 20 June 2005
Page 5 - 8 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 45
DRAFT 18 May 2005
Section 6
Cleaning, Maintenance,
Replacement Parts, and
Storage and Handling

Cleaning

WARNING:
Follow the product manufacturer’s instructions. Failure to do so could result in personal injury or equipment damage.
SHOCK HAZARD:
Before cleaning, maintenance, or parts replacement, unplug the charger unit from its power source. Failure to do so could result in personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for liquid pooling. Personal injury or equipment damage could occur.
CAUTION:
Do not use harsh cleansers/detergents, such as scouring pads and heavy duty grease removers, or solvents, such as toluene, xylene, and acetone. Equipment damage could occur.
If there is no visible soilage with possible body fluids, we recommend that you clean the unit with alcohol or a medical instrument detergent and warm wet cloth or gauze sponge. If disinfection is desired, you may use a disinfectant as explained in “Disinfecting” on page 6-2. Do not submerge unit in water or hold under running water to rinse.

Steam Cleaning

Do not use any steam cleaning device on the unit. Do not autoclave the unit. Excessive moisture can damage mechanisms or electronics in this unit.

Cleaning Difficult to Access Areas

Do not attempt to disassemble the reader or base for cleaning. Wipe exterior surfaces only. To remove difficult spots or stains, we
Jaundice Meter (Model JM-103) User Manual (usr070) Page 6 - 1
Page 46
DRAFT 18 May 2005
recommend that you use a soft bristle brush and alcohol or a medical instrument detergent. To loosen heavy, dried-on soil, you may first need to saturate the spot with a damp gauze sponge or cloth. Disinfecting is preferable in cases of contamination (visual).

Disinfecting

When there is visible soilage and between patients, we recommend that you disinfect the unit with a tuberculocidal disinfectant, such as Kleenaseptic®-b. For customers in the US, the disinfectant should be registered with the Environmental Protection Agency.
Dilute the disinfectant according to the manufacturer’s instructions.

Maintenance

There are no user serviceable parts and no maintenance or calibration required. Service is only required if the unit ceases to function as intended or fails the checker reading (see “Operational Checkout of the Jaundice Meter” on page 4-4).
NOTE:
For disposal of consummable materials,see “Disposal” on page 6-3.

Replacement Parts

CAUTION:
The instrument has a built-in, non-user-replaceable battery. Do not disassemble the instrument to replace the battery. To replace the battery, contact your dealer or authorized service center. Failure to do so could result in equipment damage.
There are no user-replaceable parts. To replace the battery, contact your dealer or an authorized service facility.
Page 6 - 2 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 47
DRAFT 18 May 2005

Storage and Handling

When storing the instrument, pay attention to the following conditions:
• Store the instrument at a temperature range of -10°C (14°F) to 50°C (122°F), and at a non-condensing relative humidity range of 30% to 95%.
• Keep the instrument dry.
• Do not store the instrument in locations that may have an adverse effect on its performance, such as:
Direct sunlight—do not store near windows.
Extreme dust—do not store in closets or bins where dust or lint
can gather.
Air having salinity or sulphur content.
Strong magnetic fields—do not store near MRI or other
imaging equipment, and do not store near operating rooms.
• Do not subject the instrument to severe vibration or impact.
• Do not store the instrument in locations where chemicals are stored or where solvent gases may be emitted.
• To ensure no problems will exist the next time the main body and charger are used, thoroughly clean the main body and charger with alcohol, and always store them together.

Disposal

This device is subject to EU Directive 2002/96/EC (WEEE). It is not registered for use in private households, and may not be disposed of at municipal collection points for waste electrical and electronic equipment.
Dräger Medical has authorized a firm to dispose of this device in the proper manner. For more detailed information, please contact your local Dräger Medical organization. (Alternatively: further information can be obtained from our national Dräger Medical organization.)
Jaundice Meter (Model JM-103) User Manual (usr070) Page 6 - 3
Page 48
NOTES:
DRAFT 18 May 2005
Page 6 - 4 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 49
DRAFT 18 May 2005
Section 7
Troubleshooting

Service Calls

When you call technical support about your unit, be prepared to give the serial number from the product identification label. When you give the serial number, the technical support representative can identify your unit and give you the information you need more quickly.

Error Messages

For warnings that may appear on the display window, refer to the table below.
Error Messages
War ning Cause So lutio n
Er1 The measured value is
abnormal. In the case of an averaging measurement, the measurement fluctuation is excessively large.
Er2
through
Er6
A measurement error may have occurred during an aver­aging measurement, or the hardware is not functioning properly.
Place the measuring probe perpendicular to the infant’s forehead (at hospital sites only) or sternum (at hospital sites or physicians’ offices), and take the measurement again. If Er1 still appears, contact the nearest authorized service facility.
Set the Power switch to the Off position, and then return it to the On position. If the warning still appears, contact the nearest authorized service facility.
Jaundice Meter (Model JM-103) User Manual (usr070) Page 7 - 1
Page 50
DRAFT 18 May 2005

Troubleshooting

WARNING:
Only facility-authorized personnel should troubleshoot the Jaundice Meter. Troubleshooting by unauthorized personnel could result in personal injury or equipment damage.
If an abnormality occurs with the Jaundice Meter, perform the following:
1. Refer to the table below, and take the necessary action given.
2. If the abnormality still appears, set the Power switch to the Off position, and then return it to the On position.
3. If the abnormality still continues, contact the nearest authorized service facility, and take the unit out of service.
Troubleshooting
Symptom Possible Cause Action
The display is blank when the Power switch is in the On position.
The display suddenly goes blank during a measurement.
The Charger lamp does not illuminate when the Jaundice Meter is placed on the charger unit.
It is impossible to take measurements.
The batteries are exhausted.
The batteries are exhausted.
The charger unit and the AC adapter are not plugged into an AC outlet correctly.
The Jaundice Meter is not placed in the charger unit correctly.
The Ready lamp is not illuminated.
The batteries are exhausted.
Charge the battery (see “Charging the Battery” on page 4-1).
Charge the battery (see “Charging the Battery” on page 4-1).
Plug the charger unit and the AC adapter into an appropriate power source correctly.
Place the Jaundice Meter perpendicular to the measuring point in the charger unit.
Before taking a measurement, ensure that the Ready lamp is illuminated.
Charge the battery (see “Charging the Battery” on page 4-1).
Page 7 - 2 Jaundice Meter (Model JM-103) User Manual (usr070)
Page 51
DRAFT 20 June 2005
Appendix A
Clinical Performance Summary

Introduction

The Jaundice Meter (JM-103) has been the subject of clinical studies in Japan and the United States. The following is a summary review of two clinical studies in the US and a later study in the doctor’s office setting.
Because this device performs measurement through the use of light it is non-invasive and painless for the infant. The objective of the clinical studies was to confirm that the device measurement, displayed in units of estimated bilirubin concentration, does correlate with the serum bilirubin concentration sufficiently well to warrant its use as a screening tool.
The data in the following sections is provided to demonstrate results of the JM-103 clinical studies, in comparison to total serum bilirubin. The JM-103 reports values over a range of 0-20 mg/dL estimated bilirubin. Appendix A pages A2 - A18 illustrate results of hospital site studies. Appendix B pages B3 - B8 illustrate results of physicians’ office site studies.
The following is a summary of the study protocols:

Study Design

Selection Criteria

The patient selection criteria used for the studies included infants less than 30 days old and weighing greater than 1000 grams. Although the selection criteria was established to be “less then 30 days of age,” the infants in the hospital studies would be expected to be primarily NICU and newborn infants unless their medical condition required a longer duration of care. The test was performed on infants who were determined by their physician to require a serum bilirubin test. A tabulation of weight distribution is provided in graph 22 Infant Weight
Distribution. Error plots by weight are provided in graphs 23 Clinical Study Site A Birthweight Error Plot, Forehead and 24 Clinical Study Site A Birthweight Error Plot, Sternum.
Jaundice Meter (JM-103) User Manual (usr070) Page A - 1
Page 52
DRAFT 20 June 2005

Demographics of Patient Population

All patients meeting the above criteria were included in the study. There was significant effort to ensure that there was sufficient representation of all skin pigmentation to verify that the JM 103 could be used across all populations with consistent results. The demographics of the patient population included Caucasian, African-American, and other.

Sample Size

The total number of infants in the sample populations are shown on the graphs of the trials presented on pages A-3 through A-15. The hospital trials studies 613 patients. The data for the doctors’ office study is in Appendix B and encompassed 201 infants.

Measurement Selection

One study followed a protocol where three measurements were taken, each measurement was recorded and then the three measurements were averaged. The second study took one measurement only. Estimated bilirubin measurements taken during the studies ranged from 1.1 to 20.8 mg/dL. In the doctor’s office setting no measurement averaging was performed.

Body Sites Tested

In the hospital setting the measurements were taken on the forehead and sternum each time the measurements were taken for a given patient. In the doctor’s office setting all measurements were taken at the sternum.

Number of Hospital Sites

There were two hospital sites that participated in the hospital study. Hospital study site A with 513 patients studied. Clinical study site B with 100 patients studied.

Performance Data

The data in the following sections, as well as Appendix B, is provided to demonstrate the performance of the JM 103. This series of graphs show the correlation of the estimated bilirubin concentration taken non­invasively with the JM 103 to the actual serum bilirubin concentration measured from a blood sample taken from the patient (TSB) as explained in Sections 4 and 5. The device operates over a range of 0.0 –
Page A - 2 Jaundice Meter (JM-103) User Manual (usr070)
Page 53
DRAFT 20 June 2005
20.0 mg/dL (0 – 340 µmol/L). The data include graphs in the form of x­y plots where x is the total serum bilirubin concentration measured and y is the JM 103 estimated bilirubin measurement. Refer to the graphics results provided in this appendix for the results of studies at hospital sites. To address device performance in doctors’ office setting, a study of 201 infants was performed. Results showed similar performance at doctors’ offices as shown in Appendix B. The data for the doctor’s office setting contained no forehead measurements as all measurements were taken at the sternum. This data correlated well with prior sternum data (refer to page A-12). The serum bilirubin measurements were taken using direct spectrophotometry in Clinical Study A and specifically with the Beakman-Colter Synchron LX-20by in Hospital Study B. Both systems were used int he doctors’ office study as well.
Graph 1 – Hospital Study Site A All Patients, Forehead
Jaundice Meter (JM-103) User Manual (usr070) Page A - 3
Page 54
DRAFT 20 June 2005
Graph 2 – ospital Study Site A All Patients, Sternum
Graph 3 – Hospital Study Site A African-American Patients, Forehead
Page A - 4 Jaundice Meter (JM-103) User Manual (usr070)
Page 55
DRAFT 20 June 2005
Graph 4 – Hospital Study Site A African-American Patients, Sternum
Graph 5 – Hospital Study Site A Caucasian Patients, Forehead
Jaundice Meter (JM-103) User Manual (usr070) Page A - 5
Page 56
DRAFT 20 June 2005
Graph 6 – Hospital Study Site A Causasian Patients, Sternum
Graph 7 – Hospital Study Site A Other Patients, Forehead
Page A - 6 Jaundice Meter (JM-103) User Manual (usr070)
Page 57
DRAFT 20 June 2005
Graph 8 – Hospital Study Site A Other Patients, Sternum
Graph 9 – Hospital Study Site B All Patients, Forehead
Jaundice Meter (JM-103) User Manual (usr070) Page A - 7
Page 58
DRAFT 20 June 2005
Graph 10 – Hospital Study Site B All Patients, Sternum
Graph 11– Hospital Study Site B African-American Patients, Forehead
Page A - 8 Jaundice Meter (JM-103) User Manual (usr070)
Page 59
DRAFT 20 June 2005
Graph 12 – Hospital Study Site B African-American Patients, Sternum
Graph 13 – Hospital Study Site B Caucasian Patients, Forehead
Jaundice Meter (JM-103) User Manual (usr070) Page A - 9
Page 60
DRAFT 20 June 2005
Graph 14 – Hospital Study Site B Causasian Patients, Sternum
Graph 15 – Hospital Study Site B Other Patients, Forehead
Page A - 10 Jaundice Meter (JM-103) User Manual (usr070)
Page 61
DRAFT 20 June 2005
Graph 16 – Hospital Study Site B Other Patients, Sternum
Jaundice Meter (JM-103) User Manual (usr070) Page A - 11
Page 62
DRAFT 20 June 2005
Table 1: Summary Table of Slope, Intercept, Standard
Deviation, and Correlation Coefficients for Each Graph
Study Site / Patient
Population
SITE A Forehead All (n=513)
SITE A Sternum All (n=513)
SITE A Forehead Afri­can American (n=65)
SITE A Sternum African American (n=65)
SITE A Forehead Cau­casian (n=399)
SITE A Sternum Cauca­sian (n=399)
SITE A Forehead Other (n=49)
SITE A Sternum Other (n=49)
SITE B Forehead All (n=100)
SITE B Sternum All (n=100)
SITE B Forehead Afri­can American (n=48)
SITE B Sternum Afri­can American (n=48)
SITE B Forehead Cauca­sian (n=35)
SITE B Sternum Cauca­sian (n=35)
SITE B Forehead Other (17)
SITE B Sternum Other (17)
Correlation
Slope Intercept
1.05 -0.35 0.914 2.19
1.07 -0.74 0.946 1.02
1.15 -0.5 0.908 1.59
1.11 -0.4 0.908 1.55
1.01 -0.1 0.916 1.2
1.04 -0.6 0.956 0.88
1.06 -0.5 0.941 1.04
1.10 -1.0 0.977 0.65
1.07 -0.00 0.84 2.14
1.16 -0.43 0.89 1.85
1.40 +0.46 0.84 2.27
1.21 -0.17 0.89 1.9
1.10 -1.04 0.87 1.72
1.22 -1.69 0.88 1.81
1.03 -0.56 0.94 1.49
1.03 0.65 0.97 0.94
Coefficient
(r)
Standard
Deviation
(RMSE)
Page A - 12 Jaundice Meter (JM-103) User Manual (usr070)
Page 63
DRAFT 20 June 2005
Graph 17 – Hospital Study Site A Error Plot, Forehead
Graph 18 – Hospital Study Site A Error Plot, Sternum
Jaundice Meter (JM-103) User Manual (usr070) Page A - 13
Page 64
DRAFT 20 June 2005
Graph 19 – Hospital Study Site B Error Plot, Forehead
Graph 20 – Hospital Study Site B Error Plot, Sternum
Page A - 14 Jaundice Meter (JM-103) User Manual (usr070)
Page 65
DRAFT 20 June 2005
Graph 21 – Infant Weight Distribution
Infant Weight
(grams)
Hospital Study A Hospital Study B
Up to 999 6 2
1000 - 1499 15 0
1500 - 1999 23 13
2000 - 2499 37 23
2500 - 2999 68 19
3000 - and up 337 43
Unknown 27 0
Total 513 100
Graph 22 – Hospital Study Site A Birthweight Error Plot, Forehead
Jaundice Meter (JM-103) User Manual (usr070) Page A - 15
Page 66
DRAFT 20 June 2005
Graph 23 – Hospital Study Site A Birthweight Error Plot, Sternum

Reproducibility

Reproducibility of the light output of the device was tested daily using the checker. The checker determines the intensity of the light output of the device. The accuracy of the device is determined by how well the detectors in the unit measure the returning light. The results of the device testing using the checker show that the device produced output within the required range over the course of both hospital studies.
Reproducibility testing in the patient population can be derived from the data taken in Hospital study site A. The reproducibility data is based on 467 babies. Three independent measurements were taken at each site, forehead and sternum. Each measurement was recorded, the mean and standard deviation computed and recorded. The average standard deviation for forehead and sternum measurements was 0.3 for both.
Page A - 16 Jaundice Meter (JM-103) User Manual (usr070)
Page 67
DRAFT 20 June 2005
Table 2: Reproducibility Data
Mean
Estimated
Bilirubin
Measure
ment
Body Site
Deviation
Range
Deviation
Mean
Deviation
Median
Range
(mg/dL)
Sternum 0.0 - 2.2 0.3 0.3 0.8 - 18.5
Forehead 0.0 - 2.6 0.3 0.2 0.1 - 19.5

Conclusion

What the data presented shows is that the JM 103 estimated bilirubin concentration measurement correlates to the serum bilirubin measurements. This data supports the use of this non-invasive device along with other clinical indicators as an aid in the management of jaundice in the neonatal patient population.
Jaundice Meter (JM-103) User Manual (usr070) Page A - 17
Page 68
NOTES:
DRAFT 20 June 2005
Page A - 18 Jaundice Meter (JM-103) User Manual (usr070)
Page 69
DRAFT 20 June 2005
Appendix B
Doctors’ Office Data

Study Design

Studies were performed at 2 doctors’ office sites comparing JM-103 Total Calculated Bilirubin (TcB) versus laboratory measured total serum bilirubin (TSB).

Selection Criteria

The ages of the infants in the study ranged from approximately 24 hours to 7-10 days, with a mean of 3 days (at site 1) and 5 days (at site 2). The test was performed on infants who were determined by their physician to require a serum bilirubin test.

Demographics of Patient Population

All patients meeting the above criteria were included in the study. The majority of babies were Caucasian or no skin tone noted (n=167) with a small number of reported ethnicity defined infants (n=34). The demographics of the patient population included Caucasian, African­American, and other.
Jaundice Meter (JM-103) User Manual (usr070) Page B - 1
Page 70
DRAFT 20 June 2005
The doctor's office study was composed of the following ethnic groups:
TcB Range Qty Age
Caucasian 2.3-19.5 167
African American
Mid-Eastern 8.8 - 16.0 5
Indian 6.6- 17 5
Hispanic 7.6 1
Asian 4.8-12.6 10
NOTE:
Doctor’s Office #1: 83.5% Caucasian and 16.5% darker skin tone Doctor’s Office #2: 75% Caucasian and 25% darker skin tone

Exclusion Criteria

Infants requiring Exchange Transfusion or Phototherapy Initiated were not allowed in the study.

Sample Size

The doctor’s office use trial studied 201 patients.
4.9-17.3 13
97
72 hrs & 7072hrs
6
72 hrs & 772hrs
4
72 hrs & 172hrs
4
72 hrs & 172hrs
3
72 hrs & 772hrs

Measurement Selection

Estimated bilirubin measurements taken during the studies ranged from
2.3 to 19.5 mg/dL. No "Averaging of Readings" was used with the JM-
103.

Body Sites Tested

All infants were measured at the sternum location ONLY .

Number of Doctors’ Office Sites

For the doctor’s office study, two sites were chosen for application of the device. Sample size was 201 patients.
Page B - 2 Jaundice Meter (JM-103) User Manual (usr070)
Page 71
DRAFT 20 June 2005

Performance Data

The data in the following sections is provided to demonstrate results of the JM-103 clinical studies, in comparison to total serum bilirubin. The JM-103 reports values over a range of 0-20 mg/dL estimated bilirubin. Although the Doctor's Office Study showed correlation of JM-103 to the TSB laboratory values, the relationship (bias) of the JM-103 to the TSB values was somewhat different than previously reported for hospital sites (as previously shown in Appendix A). This may be related to the ages of the infants in the doctor's office study and the difference in individual development. Therefore, please be aware that results at Doctors offices may differ from results at hospitals and may have more variability.
The reported data shows that 27% of the JM-103 measurements were higher than the Laboratory TSB and 71% were lower (2% were matched sets). Of the readings falling below the TSB value, 87% were within 3 Mg/dL of the TSB value. Those readings that reported above the TSB value 94% were within 3 Mg/dL of the reported TSB value. The largest error was a single 5.2 Mg/dL reading. The serum bilirubin measurements were taken using direct spectrophotometry and the Beakman-Colter Synchron LX-20by.
See Regression Analysis and infant age graphs below for data at the individual sites. A regression analysis is also provided on the total of 201 infants.
Jaundice Meter (JM-103) User Manual (usr070) Page B - 3
Page 72
DRAFT 20 June 2005
Graph 1 – Infant Age, Doctors’ Office #1
Graph 2 – Infant Age, Doctors’ Office #2
Page B - 4 Jaundice Meter (JM-103) User Manual (usr070)
Page 73
DRAFT 20 June 2005
Graph 3 – Regression Analysis, Doctors’ Office #1
JM-103 Regression Analysis
30
25
20
15
10
5
JM-103 Reading TcB Mg/dl
0
0 5 10 15 20 25 30
Doctor's office No.1
-5
Serum Bilirubin TSB (mg/dl)
Regression Line Upper Limit Low er Limit Data Points
Regression Analysis Statistics Sample Size (n) = 133.
Y = -0.636 + 0.977 X
Root Mean Square Error (MSE) = 1.572
R-Square = 0.804
95% Confidence Interval (CI) for the Intercept = (-1.526, 0.254)
95% CI for the Slope = (0.894, 1.061)
Jaundice Meter (JM-103) User Manual (usr070) Page B - 5
Page 74
DRAFT 20 June 2005
Graph 4 – Regression Analysis, Doctors’ Office #2
JM-103 Regression Analysis
25
20
15
10
5
JM-103 Reading TcB Mg/dl
0
0 5 10 15 20 25 30
-5
Doctor's Office No.2
Serum Bilirubin TSB (mg/dl)
Regression Line Upper Limit Low er Limit Data Points
Regression Analysis Statistics Sample Size = 68
Y = 0.646 +0.859 X
Root MSE = 1.475
R-Square = 0.769
95% CI for the Intercept = (-0.846, 2.138)
95% CI for the Slope = (0.743, 0.974)
Page B - 6 Jaundice Meter (JM-103) User Manual (usr070)
Page 75
DRAFT 20 June 2005
Graph 5 – Regression Analysis, Doctors’ Office Setting, 2 Locations
JM-103 Regression Analysis
30
25
20
15
10
5
JM-103 Measurement Mg/dl
0
0 5 10 15 20 25 30
Doctor's office setting - 2 locations
-5
Serum Bilirubin TSB (mg/dl)
Regression Line Upper Limit Low er Limit Data Points
Regression Analysis Statistics Sample Size = 201
Y = 0.233 +0.934 X
Root MSE = 1.54
R-Square = 0.899
95% CI for the Intercept = (-0.233, 0.37)
95% CI for the Slope = (0.934, 0.032)
Jaundice Meter (JM-103) User Manual (usr070) Page B - 7
Page 76
DRAFT 20 June 2005

Conclusion

What the data presented shows is that the JM 103 estimated bilirubin concentration measurement correlates to the serum bilirubin measurements. This data supports the use of this non-invasive device along with other clinical indicators as a aid in the management of jaundice in the neonatal patient population.
Page B - 8 Jaundice Meter (JM-103) User Manual (usr070)
Page 77
DRAFT 9 Nov 2004
Appendix C
Medical and Scientific References on
Transcutaneous Bilirubinometry (1979 - 1997)
1. Amato M, et al; Transcutaneous bilirubin determination: correlation in white prema­ture infants weighing less than 1500 gm. Schweiz Med Wochenschr, 1985 Jul 9.
2. Amato M, et al; Assessment of neonatal jaundice in low birth weight infants com­paring transcutaneous, capillary and arterial bilirubin levels. Eur J Pediatr, 1990 Nov.
3. Amato M
Transcutaneous, capillary, and arterial bilirubin levels. J. Pediatr. 125(2), 332, 1994
4. Amit Y, et al; Effect of skinfold thickness on transcutaneous bilirubin measure­ments. Biol Neonate, 1993.
5. Arimichi J, et al; Reliability and future application of a transcutaneous bilirubinome­ter. Josanpu Zasshi, 1984 Apr.
6. Berget M, et al; Transcutaneous bilirubinometry in newborn infants. Tidsskr Nor Laegeforen, 1984 May 20
7. Bhat V, et al; Correlation of transcutaneous bilirubinometry with serum bilirubin in south Indian neonates. Indian J Med Res, 1987 Jul.
8. Bhutta ZA, et al; Transcutaneous bilirubinometry in Pakistani newborns: a prelimi­nary report. JPMA J Pak Med Assoc, 1991 Jul.
Jaundice Meter (JM-103) User Manual (usr070) Page C - 1
Page 78
DRAFT 9 Nov 2004
9. Boo Nem Yun, et al; Transcutaneous bilirubinometry in Malay, Chinese and Indian term neonates. Med J Malaysia, 1984 Mar.
10. Bourchier D, et al; Transcutaneous bilirubinometry: 22 months experience at Waikato Women's Hospital. N Z Med J, 1987 Sep 23.
11. Brown A; et al; Transcutaneous bilirubinometry in infants: influence of race and phototherapy. Pediatr Res, 1981, 15:653
12. Brown LP, et al; Transcutaneous bilirubinometer: intermeter reliability. J Perinatol, 1990 Jun.
13. Brown LP, et al; Transcutaneous bilirubinometer: an instrument for clinical research. Nurs Res, 1990 Jul-Aug.
14. Brown LP, et al; Incidence and pattern of jaundice in healthy breast-fed infants dur­ing the first month of life. Nurs Res, 1993 Mar-Apr.
15. Brucker MC, et al; Neonatal jaundice in the home: assessment with a noninvasive device. J Obstet Gynecol Neonatal Nurs, 1987 Sep-Oct.
16. Cassady G
Transcutaneous monitoring in the newborn infant. J Pediatr, 1983 Dec.
17. Chen ZL, et al; Clinical use of transcutaneous bilirubinometry. Chung Hua I Hsueh Tsa Chih, 1986 Jan.
18. Christo GG, et al; Transcutaneous bilirubinometry in newborns. Indian Pediatr, 1988 Nov.
Page C - 2 Jaundice Meter (JM-103) User Manual (usr070)
Page 79
DRAFT 9 Nov 2004
19. Corchia C, et al; Comment on transcutaneous bilirubin device of Yamanouchi [letter] Pediatrics, 1981 Mar.
20. Dai J, et al; Clinical impact of transcutaneous bilirubinometry as an adjunctive screen for hyperbilirubinemia Clin Biochem 29 (6): 581-586, Dec 1996
21. Dai J, et al; Transcutaneous bilirubinometry: its role in the assessment of neo­natal jaundice Clin Biochem 30 (1): 1-9, Feb 1997
22. Derksen-Samsom JF, et al; The reliability of transcutaneous bilirubin measurement: a clinical study with statistical data and literature review. Tijdschr Kindergeneeskd, 1984 Oct.
23. Dominguez Ortega F, et al; Transcutaneous bilirubinometry: correlation of the reading site obtained with spectrophotometry and diazoreaction technique. An Esp Pediatr, 1993 Nov.
24. Douville P, et al; Diagnostic value of sequential readings with the Minolta transcuta­neous bilirubinometer in normal and low-birthweight infants [letter] Clin Chem, 1983 Apr, 29:4, 740-1.
25. Fabris C, et al; Evaluation of transcutaneous bilirubinometry in newborn infants. Minerva Pediatr, 1984 Jun 15.
26. Fisher B, et al; The cephalocaudal progression of jaundice in newborns in relation to the transfer of bilirubin from plasma to skin. Early Hum Dev, 1990 Apr, 22:1, 23-8
27. Foged N, et al; Transcutaneous bilirubinometry. A non-invasive method of measur­ing physiological jaundice. Ugeskr Laeger, 1983 Sep. 19.
Jaundice Meter (JM-103) User Manual (usr070) Page C - 3
Page 80
DRAFT 9 Nov 2004
28. Fok TF, et al; Transcutaneous bilirubinometer: its use in Chinese term infants and the effect of haematocrit and phototherapy on the TcB index. Aust Paediatr J, 1986 May.
29. Galletto P, et al; Supervision of neonatal jaundice by use of the transcutaneous bilirubinometer. Minerva Pediatr, 1983 Jan 31.
30. Goldman SL, et al; Jaundice meter: evaluation of new guidelines. J Pediatr, 1982 Aug.
31. Grande R, et al; Physiological variations in the pigmentation of newborn infants. Hum Biol, 1994 Jun.
32. Hanneman R; et al; Determination of serum bilirubin by skin reflectance: effect of pig­mentation Pediatr Res 13 (12): 1326-1329 (Dec 1979)
33. Hanneman R; et al; Evaluation of the Minolta bilirubin meter as a screening device in white and black infants. Pediatrics, 1982, 69, 107-109
34. Hegyi T, et al; Transcutaneous bilirubinometry. I. Correlations in term infants. J Pediatr, 1981 Mar.
35. Hegyi T, et al; Transcutaneous bilirubinometry. The cephalocaudal progression of dermal icterus. Am J Dis Child, 1981 Jun.
36. Hegyi T
Transcutaneous bilirubinometry: a new light on an old subject. Pediatrics, 1982 Jan.
37. Hegyi T, et al; Transcutaneous bilirubinometry II. Dermal bilirubin kinetics during phototherapy.
Pediatr Res, 1983 Nov.
Page C - 4 Jaundice Meter (JM-103) User Manual (usr070)
Page 81
DRAFT 9 Nov 2004
38. Hegyi T
Transcutaneous bilirubinometry in the newborn infant: state of the art. J Clin Monit, 1986 Jan.
39. Hegyi T, et al; Transcutaneous bilirubinometry. III. Dermal bilirubin kinetics under green and blue light phototherapy. Am J Dis Child, 1986 Oct.
40. Heick C, et al; Transcutaneous bilirubin determination in the newborn infant. Rec­ommendations of the Swiss Neonatology Group. Helv Paediatr Acta, 1982.
41. Hodr R, et al; Transcutaneous measurement of the severity of icterus and biliru­binemia in normal neonates. Cesk Pediatr, 1990 Nov.
42. Hung WT, et al; Diagnosis of atretic prolonged obstructive jaundice; technetium 99m hepatolite excretion study. J Pediatr Surg, 1990 Jul.
43. Karrar Z, et al; Transcutaneous bilirubin measurements in Saudi infants: the use of the jaundice meter to identify significant jaundice. Ann Trop Paediatr, 1989 Mar.
44. Kenny M, et al; Transcutaneous bilirubin monitoring of newborns. Ann N Y Acad Sci, 1984.
45. Keshishjan E.S., et al; Diagnostics and control of newborn hyperbilirubinemia with use noninvasive transcutaneous photometric analyzer "Bilitest" of a type AHP-02.
The methodical recommendations. Russia ministry of public health. 1992 Feb.
46. Keshishjan E.S., et al; Transcutaneous bilirubinometry method at newborn jaundice diag­nostics and control. The Russian bulletin perinatology and pediatrics, V38, 5, 1993
Jaundice Meter (JM-103) User Manual (usr070) Page C - 5
Page 82
DRAFT 9 Nov 2004
47. Keshishjan E.S., et al; Use of the method of transcutaneous bilirubinometry in hyperbi­lirubinemia in the newborn. Clinical laboratory diagnostics. 5, 1994 Moscow, Medicine
48. Kivlahan C, et al; The natural history of neonatal jaundice. Pediatrics, 1984 Sep.
49. Knudsen A. Prediction of the development of neonatal jaundice by increased umbilical cord blood bilirubin. Acta Pediatr Scand 1989;78:217-21.
50. Knudsen A, et al; Skin colour and bilirubin. Arch Dis Child 1989;64:605-9.
51. Knudsen A, et al; Maternal bilirubin, cord bilirubin, and placenta function at delivery and the development of jaundice in mature newborns. Acta Obstet Gynecol Scand 1989:68:719-24.
52. Knudsen A. The cephalocaudal progression of jaundice in newborns in relation to the transfer of bilirubin from plasma to skin. Early Human Development 1990;22:23-28.
53. Knudsen A. Measurement of the yellow colour of the skin as a test of hyperbi­lirubinemia in mature newborns. Acta Pediatr Scand 1990;79:1175-81.
54. Knudsen A. Maternal smoking and the serum bilirubin concentration in the first three days of life. Eur J Obstet Gynecol Reprod Bid 1991; 40:123-2 7.
55. Knudsen A. The influence of the reserve albumin concentration and pH on the cephalocaudal progression of jaundice in newborns. Early Human Development 1991;25:37-41.
56. Knudsen A. The influence of the basal yellow colour of the skin at birth on later jaundice meter readings in mature newborn infants. Acta Pediatr Scand 1992;81:494-97.
Page C - 6 Jaundice Meter (JM-103) User Manual (usr070)
Page 83
57. Knudsen A. Prediction of later hyperbilirubinemia by measurement of skin colour at first postnatal day and cord bilirubin. Dan Med Bull 1992;39:193-6.
58. Knudsen A
Predicting the need for phototherapy in healthy mature neonates using transcutaneous bilirubinometry on the first postnatal day. Biol Neonate, 1995.
59. Knudsen A, et al; Transcutaneous evaluation of icterus in healthy mature newborn infants. A substitution for blood tests. Ugeskr Laeger, 1995 Mar 13.
60. Knudsen A, et al; Transcutaneous bilirubinometry in neonatal intensive care units. Arch Dis Child Fetal Neonatal Ed, 1996 Jul.
61. Kumar A
Micro-invasive management of neonatal bilirubinemia. Indian Pediatr, 1992 Sep.
62. Kumar A, et al; Transcutaneous bilirubinometry in the management of bilirubine­mia in term neonates. Indian J Med Res, 1994 May.
63. Laeeq A, et al; Transcutaneous bilirubinometry: clinical application. JPMA J Pak Med Assoc, 1993 Feb.
64. Lebrun F, et al; Clinical evaluation of transcutaneous bilirubinometry. Ann Pediatr (Paris), 1982 Sep.
65. Lin YJ, et al; The clinical application of transcutaneous bilirubinometry in full­term Chinese infants. Acta Paediatr Sin, 1993 Mar-Apr.
66. Linder N, et al; Noninvasive determination of neonatal hyperbilirubinemia: stan­dardization for variation in skin color. Am J Perinatol, 1994 May.
DRAFT 9 Nov 2004
Jaundice Meter (JM-103) User Manual (usr070) Page C - 7
Page 84
DRAFT 9 Nov 2004
67. Loong EP, et al; Changes in neonatal transcutaneous bilirubinometer index follow­ing intravenous fluid and oxytocin infusion during labour. Asia Oceania J Obstet Gynaecol, 1988 Dec.
68. Maisels MJ, et al; Transcutaneous bilirubin measurements in full-term infants. Pediatrics, 1982 Sep.
69. Maisels MJ, et al; Transcutaneous bilirubin measurements: variation in meter response. Pediatrics, 1983 Mar.
70. Maisels MJ, et al; Transcutaneous bilirubinometry decreases the need for serum bilirubin measurements and saves money Pediatrics 99 (4): 599-601, Apr 1997
71. Moscicka A, et al; Usefulness of transcutaneous measurements of bilirubin in infants with jaundice. Ginekol Pol, 1994 Jun.
72. Myara A, et al; Early changes in cutaneous bilirubin and serum bilirubin isomers during intensive phototherapy of jaundiced neonates with blue and green light Biol Neonate 71 (2): 75-82, 1997
73. Onks D, et al; Effect of melanin, oxyhemoglobin and bilirubin on transcutaneous bilirubinometry. Acta Paediatr, 1993 Jan.
74. Pallas Alonso C, et al; Transcutaneous bilirubin measurement in neonates. An Esp Pediatr, 1993 Jan.
75. Palmer DC, et al; Transcutaneous bilirubinometry: use in Australia. Aust Paediatr J, 1982 Dec.
76. Palmer DC; et al; Jaundice: A 10 year review of 41,000 live born infants. Australian Pediatrric Journal, 1983, 19(2),86-89
Page C - 8 Jaundice Meter (JM-103) User Manual (usr070)
Page 85
DRAFT 9 Nov 2004
77. Peng DJ, et al; A study of the effect of pathologic factors on the TcBM readings. Hua Hsi I Ko Ta Hsueh Hsueh Pao, 1989 Mar.
78. Romo-Pinales J, et al; Preliminary study in Mexico on the use of the transcutaneous biliru­binometer. Bol Med Hosp Infant Mex, 1984 Oct.
79. Ruchala PL, et al; Validating assessment of neonatal jaundice with transcutaneous bilirubin measurement. Neonatal Netw, 1996 Jun.
80. Salmeron Escobar FJ
Differential diagnosis of jaundice. Rev Esp Enferm Dig, 1991 Jan.
81. Schubiger G, et al; Transcutaneous determination of bilirubin in the neonatal depart­ment: an analysis of its uses. Helv Paediatr Acta, 1986 Aug.
82. Schumacher RE, et al; Transcutaneous bilirubinometry: a comparison of old and new methods. Pediatrics, 1985 Jul.
83. Schumacher RE
Noninvasive measurements of bilirubin in the newborn. Clin Perinatol, 1990 Jun.
84. Serrao PA, et al; Significance of anti-A and anti-B isohemagglutinins in cord blood of ABO incompatible newborn infants: correlation with hyperbi­lirubinemia J Perinatol, 1989 Jun.
85. Sharma JN, et al; Transcutaneous bilirubinometry in newborns. Indian Pediatr, 1988 Aug.
86. Sheridan-Pereira M, et al; Transcutaneous bilirubinometry: an evaluation. Arch Dis Child, 1982 Sep.
Jaundice Meter (JM-103) User Manual (usr070) Page C - 9
Page 86
DRAFT 9 Nov 2004
87. Smith DW, et al; Use of noninvasive tests to predict significant jaundice in full-term infants: preliminary studies. Pediatrics, 1985 Feb.
88. Strange M, et al; Neonatal transcutaneous bilirubinometry. Clin erinatol, 1985 Feb.
89. Suckling RJ, et al; Transcutaneous bilirubinometry as a screening tool for neonatal jaundice. Scott Med J, 1995 Feb.
90. Taha SA, et al; Transcutaneous bilirubin measurement in evaluating neonatal jaun­dice among Saudi newborns. Ann Trop Paediatr, 1984 Dec.
91. Tan KL
Transcutaneous bilirubinometry in Chinese and Malay neonates. Ann Acad Med Singapore, 1985 Oct.
92. Tan KL, et al; Transcutaneous bilirubinometry in preterm very low birthweight infants. Acta Paediatr Scand, 1988 Nov.
93. Tan KL, et al; Efficacy of phototherapy for hyperbilirubinaemia in infants with the respiratory distress syndrome J Paediatr Child Health 31 (2): 127-129, Apr 1995
94. Tan KL, et al; Transcutaneous bilirubinometry in Chinese, Malay and Indian infants Acta Paediatr 85 (8): 986-990, Aug 1996
95. Tsai LT, et al; Clinical evaluation of transcutaneous jaundice meter in full-term newborns. Acta Paediatr Sin, 1988 Nov-Dec.
96. Tudehope DI, et al; Non-invasive method of measuring bilirubin levels in newborn infants. Med J Aust, 1982 Feb 20.
Page C - 10 Jaundice Meter (JM-103) User Manual (usr070)
Page 87
DRAFT 9 Nov 2004
97. Tudehope DI, et al; Multiple site readings from a transcutaneous bilirubinometer. Aust Paediatr J, 1982 Jun.
98. Uchida K, et al; Fundamental studies on transcutaneous bilirubinometry in newborn infants using an organ scanning spectrophotometer. Nippon Sanka Fujinka Gakkai Zasshi, 1988 Feb.
99. Utchajkin V.F., et al; Clinical importance of noninvasive bilirubinemia monitoring by children viral hepatitis.
2-nd Symposiums "NoninvasiveMethods of Diagnostics". The thesis of the reports. Moscow. 1995 Oct. 30 - Nov. 2.
100. Vaisman S, et al; Clinical trial of a transcutaneous bilirubinometer. Rev Chil Pediatr, 1983 Mar-Apr.
101. Veleminsky M
The significance of transcutaneous measurement of "bilirubin" in neonates. Cesk Pediatr, 1992 Nov.
102. Vocel J, et al; Transcutaneous bilirubinometry using the Minolta/Air-Shields in neonates. Cesk Pediatr, 1985 Nov.
103. Vocel J, et al; Transcutaneous bilirubinometry using the Minolta/Air-Shields apparatus in neonates with a low birthweight. Cesk Pediatr, 1987 Jan.
104. Vocel J, et al; Transcutaneous bilirubinometry in neonates with hyperbilirubine­mia treated with phototherapy. Cesk Pediatr, 1988 Dec.
105. Weiss WJ
A completely implanted left ventricular assist device. Chronic in vivo testing. ASAIO J 39(3), M427-M432, 1993
Jaundice Meter (JM-103) User Manual (usr070) Page C - 11
Page 88
DRAFT 9 Nov 2004
106. Yamanouchi I, et al; Transcutaneous bilirubinometry: preliminary studies of noninva­sive transcutaneous bilirubin meter in the Okayama National Hos­pital. Pediatrics, 1980 Feb.
107. Yamauchi Y, et al; Transcutaneous bilirubinometry. Effect of irradiation on the skin bilirubin index. Biol Neonate, 1988.
108. Yamauchi Y, et al; Transcutaneous bilirubinometry. Evaluation of accuracy and reli­ability in a large population. Acta Paediatr Scand, 1988 Nov.
109. Yamauchi Y, et al; Transcutaneous bilirubinometry: bilirubin kinetics of the skin and serum during and after phototherapy. Biol Neonate, 1989.
110. Yamauchi Y, et al; Transcutaneous bilirubinometry: serum bilirubin measurement using transcutaneous bilirubinometer (TcB). A preliminary study. Biol Neonate, 1989.
111. Yamauchi Y, et al; Transcutaneous bilirubinometry in normal Japanese infants. Acta Paediatr Jpn, 1989 Feb.
112. Yamauchi Y, et al; Transcutaneous bilirubinometry. Interinstrumental variability of TcB instruments. Acta Paediatr Scand, 1989 Nov.
113. Yamauchi Y, et al; Clinical application of transcutaneous bilirubin measurement. Early prediction of hyperbilirubinemia. Acta Paediatr Scand, 1990 Apr.
114. Yamauchi Y, et al; Transcutaneous bilirubinometry: effect of postnatal age. Acta Paediatr Jpn, 1991 Oct.
115. Yamauchi Y, et al; Transcutaneous bilirubinometry: variability of TcB measurements on the forehead with crying. Acta Paediatr Jpn, 1991 Oct.
Page C - 12 Jaundice Meter (JM-103) User Manual (usr070)
Page 89
DRAFT 9 Nov 2004
116. Yamauchi Y, et al; Factors affecting transcutaneous bilirubin measurement: effect of daylight. Acta Paediatr Jpn, 1991 Oct.
117. Yip WC, et al; Transcutaneous bilirubinometry [letter] Acta Paediatr Scand, 1983 Mar.
Jaundice Meter (JM-103) User Manual (usr070) Page C - 13
Page 90
NOTES:
DRAFT 9 Nov 2004
Page C - 14 Jaundice Meter (JM-103) User Manual (usr070)
Page 91
PROPRIETARY AND CONFIDENTIAL DRAFT 5/14/04
Page 92
Manufactured by: Distributed by: EC Representative:
Konica Minolta Sensing, Inc.
3-91, Daisennishimachi Sakai-ku, Sakai, Osaka
590-8551 Japan
Part number Revisions Date
usr070 (MU01380) rg October 2006
Draeger Medical Sys­tems, Inc.
3135 Quarry Road Telford, PA 18969 USA USA and Canada Cus-
tomer Support: (800) 437-2437 http://www.draegermedi­cal.com Technical Support:
(800) 437-2437 Fax: (215) 721-5782
Konica Minolta Sensing Europe B.V.
Edisonbaan 14-E 3439MN Nieuwegein,
the Netherlands Tel: +31 30 248 1166 Fax: +31 30 248 1282
Draeger Medical Systems, Inc. reserves the right to make changes without notice in design, specifications, and models. The only warranty Draeger Med­ical Systems, Inc. makes is the express written warranty extended on the sale or rental of its products.
The information in this manual is confidential and may not be disclosed to third parties without the prior written consent of Draeger Medical Systems, Inc.
© 2006 by Draeger Medical Systems, Inc. ALL RIGHTS RESERVED.
Loading...