Drager Julian User Manual

Julian
Anaesthetic Workstation
Instructions for Use Software 2.n
1-36-96
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
How to use these Instructions for Use
In the header is the subject… of the main chapter.
Underneath is the title of the subchapter, to help you find your way around quickly.
On each page… the instructions for use
combining text with illustrations. The information is translated directly into actions to enable the user to learn „hands-on“ how to use the workstation.
Left-hand column: the text…
provides explanations and guides the user clearly and ergonomically with brief directions on the use of the product. Bullet points indicate the steps to be followed, and where there are several steps, numbers refer to the illustration and indicate the sequence.
Right-hand column: the illustration…
provides the link with the text and serves as a guide to the workstation. Points mentioned in the text are emphasised and non-essential information is omitted.
Screen displays guide the user and confirm the steps to be followed.
How to use these Instructions for Use
2
1 Press the »MAN/SPONT« key, 2 and confirm with the rotary knob.
Display (example):
O2 flush
For flushing and rapidly filling the breathing
system and breathing bag with O2 while bypassing the Vapor unit.
3 Press the »O2 +« button.
O2 flows into the breathing system without anaesthetic gas as long as the button is held down.
Julian
1
2
Operation Selecting the ventilation mode MAN/SPONT
Julian
D
3
MAN/SPONT
Pleth
40
0
CO
2
40 36
0.8 0.6 58 56
38
98 67
5.2 8
4.0030
Volumeter starten: Bestätigen!
O
2
SpO
2
N2O
Hal.
Fi
Fet
et CO
2
AMV
Freq
Grenzen CO
2
-Al.
aus/ein Alarm
Info Liste
Kurven
Konfig.
Frischgas O
2
+ N2O
O
2
%
L/min
VT
0.35
0
Volumeter
--s
--
05
10
0.5
1
Frischgas
intern extern
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
3
Contents
Index 142
For Your Safety and that of Your Patients 5
Intended Use 7
Operating Concept 11
Before Using for the First Time 19
Preparation 21
Starting Up 25
Operation 35
Monitoring 55
Configuring in Standby Mode 81
Care 95
Julian as Wall-mounted Unit 113
Maintenance Intervals 115
What's what 125
Technical Data 131
Fault – Cause – Remedy 119
Abbreviations / Symbols 139
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
4
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
For Your Safety and that of Your Patients
Strictly follow the Instructions for Use
Any use of the apparatus requires full understanding and strict observation of these instructions. The apparatus is only to be used for purposes specified here.
Maintenance
The apparatus must be inspected and serviced regularly by trained service personnel at six monthly intervals (and a record kept). Repair and general overhaul of the apparatus may only be carried out by trained service personnel. We recommend that a service contract be obtained with DrägerService and that all repairs also be carried out by them. Only authentic Dräger spare parts may be used for maintenance. Observe chapter "Maintenance Intervals".
Accessories
Do not use accessory parts other than those in the order list.
Not for use in areas of explosion hazard
This apparatus is neither approved nor certified for use in areas where combustible or explosive gas mixtures are likely to occur.
Safe connection with other electrical equipment
Electrical connections to equipment which is not listed in these Instructions for Use should only be made following consultations with the respective manufacturers or an expert.
Liability for proper function or damage
The liability for the proper function of the apparatus is irrevocably transferred to the owner or operator to the extent that the apparatus is serviced or repaired by personnel not employed or authorized by DrägerService or if the apparatus is used in a manner not conforming to its intended use.
Dräger cannot be held responsible for damage caused by non-compliance with the recommendations given above. The warranty and liability provisions of the terms of sale and delivery of Dräger are likewise not modified by the recommendations given above.
Dräger Medizintechnik GmbH
For Your Safety and that of Your Patients
5
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Intended Use............................................................................................. 9
7
Intended Use
Contents
Intended Use
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
8
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Intended Use
Julian Anaesthetic Workstation for patients with a body weight of 5 kg and over with the use of IPPV ventilation.
For the use of:
Inhalation anaesthesia in rebreathing systems Inhalation anaesthesia in semi-closed to virtually
closed systems with "low flow" and "minimal flow" techniques (for minimal gas and anaesthetic agent consumption).
Inhalation anaesthesia in non-rebreathing systems,
with separate fresh gas outlet for the connection of e.g. the Bain system or Magill system
with a fresh gas flow of 0.5 to 12 L/min.
Ventilation modes:
Automatic ventilation (IPPV)
and
pressure-controlled ventilation (PCV). Manual ventilation (MAN). Spontaneous ventilation (SPONT).
The following measured values are displayed:
Peak pressure, mean pressure, plateau pressure and
PEEP Expiratory minute ventilation
Tidal volume VT
Breathing rate
Patient compliance Inspiratory and expiratory concentration of O2, N2O,
anaesthetic gas and CO2 optional: Functional oxygen saturation (SpO2) and pulse rate.
The following parameters are displayed as curves:
Airway pressure Expiratory flow Inspiratory and expiratory concentration of O2, CO2
and anaesthetic gas optional: Plethysmogram Trend curves and measured value lists are also available.
Monitoring
by means of adjustable alarm limits that are automatically adapted to the ventilation mode.
The workstation may only be used under the supervision of qualified medical personnel, so that assistance can be provided immediately in the event of any malfunctions.
Explosive anaesthetics, such as ether or cyclo- propane, must not be used due to the risk of fire.
Mobile radio telephones must not be used within 10 metres of the workstation!
Mobile telephones may interfere with the operation of electrical and electronic medical equipment.
Julian must not be used with nuclear spin tomography (MRT, NMR, NMI)!
Operation of the apparatus may be impaired.
Intended Use
9
Intended Use
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
10
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Screen ergonomics.................................................................................12
Selecting / setting ventilation parameters..................................................14
Selecting / setting monitoring functions.....................................................15
Screen layout..........................................................................................16
Three basic screens for monitoring...........................................................17
The standard screen.................................................................................17
The data screen....................................................................................... 17
The trend screen......................................................................................18
Operating Concept
Contents
11
Operating Concept
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operating Concept
Screen ergonomics
All the settings required for Fresh gas delivery Ventilation Monitoring are entered on the system screen using the appropriate
keys and the rotary knob. The keys are grouped in function fields:
Left-hand field Fresh gas delivery
Right-hand field Ventilation
Middle field Monitoring
The main functions for anaesthesia, e.g. selection of N2O or AIR, or selection of ventilation modes, can be selected directly by keys with permanently defined functions ("hardkeys"):
1 Left-hand block
The »N2O« or »AIR« keys are used to select the gas to be mixed with O2 for the fresh gas mix.
2 Right-hand block
The »MAN SPONT«, »IPPV« oder »PCV« keys select the ventilation mode.
These function keys are located in the bottom row on the control panel. Left-hand block for fresh gas setting Right-hand block for ventilation
Operating Concept Screen ergonomics
12
.
12...
.
NO
2
AIR
MAN
SPONT
IPPV PCV
Monitoring
Ventilation
Fresh gas delivery
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
.
12...
.
NO
2
AIR
MAN
SPONT
IPPV PCV
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
1
2
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Complementary "softkeys" with variable functions are provided at the bottom edge of the screen, above each group of hardkeys. These softkeys are used to set the fresh gas delivery parameters and ventilation parameters.
1 Left-hand block:
The keys for setting the O2 concentration and fresh gas flow.
2 Right-hand block:
The keys for setting the parameters for the relevant ventilation mode. The example shows the parameters for IPPV controlled ventilation.
These softkeys have different functions, depending on the operating status or ventilation mode.
The current parameter values are displayed in the softkey field.
In a prominent position at the bottom right-hand side: The "turn-and-push" rotary knob
is the main operating control of the apparatus and has the following functions in all setting operations:
3 select = turn 4 confirm = press
to confirm the selected carrier gas or a ventilation
mode,
to set and confirm the parameters for fresh gas and
ventilation modes,
to set and confirm the monitoring functions.
Beside the rotary knob: The standby key E for switching over to standby mode. 5 Press standby key E and confirm by pressing the
rotary knob.
13
Operating Concept Screen ergonomics
.
12...
.
NO
2
AIR
MAN
SPONT
IPPV PCV
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
1 2
_
CO
2
Ve
.
12...
.
NO
2
AIR
MAN
SPONT
IPPV PCV
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
3
5
4
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operating Concept Screen ergonomics
14
Selecting / setting ventilation parameters
Example: PEEP ventilation parameter
1 Press the softkey »PEEP«. 2 Select the PEEP value = turn the rotary knob. 3 Confirm the PEEP value = press the rotary knob.
The keys for the various monitoring functions are located on the right-hand side of the screen.
These keys also have different functions = softkeys, depending on the monitoring screen required.
Julian
1
2
3
.
12...
.
NO
2
AIR
MAN
SPONT
IPPV PCV
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
15
Operating Concept Screen ergonomics
Selecting / setting monitoring functions
For example, to change the lower alarm limit of the end- tidal CO2 concentration.
1 Press the »alarm limits« softkey. The alarm limits
menu is displayed on the screen.
Select the alarm limit value = turn the rotary knob.
Confirm the selection = press the rotary knob. The limit value is highlighted.
Example: lower alarm limit etCO2: <30
Set the alarm limit value = turn the rotary knob.
Confirm the new alarm limit = press the rotary knob.
The cursor returns to the zsymbol.
Julian
1
Pleth
IPPV
CO2
40
0
PAW
600 10
1:2
10 0251.0050
98 67 38
6.0 29
0.8
--
95
3.0
--
20
1.0
0.0 40
8
120
50
--
Select limit: Turn and confirm ! !
42 30
0
20
FiO
2
FiHal.
PAW
SpO
2
etCO
2
MV
alarm limits
auto set limits
alarm info
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
Pleth
IPPV
CO
2
40
0
600 10
1:2
10 0251.0050
98 67 38
6.0 29
0.8
--
95
3.0
--
20
1.0
0.0 40
8
120
50
--
Adjust limit value: Turn and confirm !
42
30
PAW
0
20
FiO
2
FiHal.
PAW
SpO
2
etCO
2
MV
alarm limits
auto set limits
alarm info
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O2 + N2O
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operating Concept Screen ergonomics Screen layout
16
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
➀ ➁ ➂ ➃
➄ ➅ ➆ ➇
Exit the alarm limits menu: 1 Press the rotary knob
or
2 Press the Q key.
The function keys for standard functions are located on the right-hand side of the control panel.
G Suppress the acoustic alarm for 2 minutes. S Select the screen page. Q Back to standard page.
Screen layout
Status field:
Displays information on the current operating mode
Graphics field:
For curves and bargraphs
Alarm field:
Displays information on the alarms, warnings, etc. and their priority
Numerical value field:
For numerical values
Lower softkeys – for fresh gas parameters Prompt field:
For user guidance
Lower softkeys – for ventilation parameters Right-hand softkeys – for monitoring:
For rapid selection of the desired monitoring function
.
12...
.
NO
2
AIR
MAN
SPONT
IPPV PCV
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
Julian
2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
17
Operating Concept
Screen layout
Three basic screens for monitoring To call up the standard screen, data screen and trend
screen in succession: 1 Briefly press the S key until the desired screen is
displayed.
Back to the standard screen: 2 Press the Q key.
The fresh gas and ventilation parameters can be set by means of the softkeys in each of the three basic screens.
The standard screen
with three selectable curves. The most important numerical values are displayed in groups on the right-hand side of the screen.
The data screen
contains all the numerical values with their units of measure.
Julian
2 1
Pleth
IPPV
alarm limits
auto set limits
alarm info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
40 36
0.8 0.6 58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
IPPV
PAW
40
20
0
600 10
1:2
10 0251.0050
O
2
%
N
2
O
%
Hal. %
036
40 33
0.8 0.6 58 56
Fi
Fet
CO
2
mmHg
98 67
6.0
0.60 10
27 20
5 15 65
MV V
T
freq
peak plat. PEEP mean compliance
mbar
mL/mbar
L/min
L
34
AW-temp.
°C
1/min
01. Nov. 98
sys. comp. 6.40 mL/mbar leckage 10 mL/min
01-10
8 : 00
12 : 00
SpO
2
%
1/min
alarm limits
auto se
t
limits
alarm info
list
config.
Freshgas O
2
+ N2O
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operating Concept Screen layout
18
The trend screen
displays the recorded progress of the numerical values over time since measurement started. The current measured values are displayed on the right-hand side.
Display (example): Trends for CO2 and minute ventilation MV
IPPV
40 36
0.8 0.6 58 56
38
98 67
6.0 10
600 10
1:2
10 0251.0050
PAW
40
20
0
11:00 12:00 13:00
60
30
0
15
10
5
0
Select zoom area:
Turn !
Zoom in: Press !
CO
2
MV
O
2
SpO
2
N2O
Iso.
Fi
Fet
etCO
2
MV
freq
alarm limits
AGas N
2
O
CO
2
MV
O
2
compl. SpO
2
pulse full
trend
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Charging the battery for emergency operation......................................20
When Julian is not in use........................................................................ 20
19
Before Using for the First Time
Contents
Before Using for the First Time
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Before Using for the First Time
Fit the enclosed O2 sensor, page 117 "Replacing O2 sensor" Fit the flow sensor, page 106
Charging the battery for emergency operation
Julian has a built-in uninterruptible power supply UPS which maintains the power supply for 30 minutes in the event of a mains failure, provided that the battery is charged. The switch-over to the battery-powered UPS is automatic and is indicated on the screen by the message: POWER FAIL!
The battery recharges automatically when the workstation is plugged into the mains.
The battery must be charged for 10 hours before using the Workstation for the first time.
Plug the mains plug of the Julian workstation into the
mains socket. The mains voltage must correspond to the voltage specified on the nameplate.
1 The green LED
NN
NN
lights up.
Leave Julian connected to the mains for 10 hours.
The workstation does not have to be switched on.
The devices connected to auxiliary power sockets will not be powered by the UPS in the event of a power failure.
When Julian is not in use
Charge battery at least every 4 weeks.
Allowing it to run down can lead to damage.
If Julian is out of use for an extended period:
Leave the workstation connected to the mains at all
times.
1 The green LED
NN
NN
is lit.
20
Before Using for the First Time Charging the battery for emergency operation When Julian is not in use
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Connecting the gas supply.....................................................................22
Connecting the backup gas cylinders for O2 and N2O .............................23
Caution when handling O2 cylinders.........................................................23
Connecting the anaesthetic gas scavenging system.............................23
Connecting the power supply.................................................................24
Connecting auxiliary systems....................................................................24
Equipotential bonding...............................................................................24
Connecting the power supply...................................................................24
21
Preparation
Contents
Preparation
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Preparation
Use only cleaned and disinfected parts!
Connecting the gas supply
1 Screw the central gas supply pressure hoses for O2,
AIR and N2O into the ports on the back of the gas supply block. The two ports at the front are reserved for the backup gas cylinders.
2 The vacuum supply block (VAC) is available as option
for secretion aspiration. Plug the other end of the pressure hoses into the wall
supply points.
Make sure the gas pressures of the central gas
supply are between 2.7 and 5.5 bar:
3 All three pressure gauges in the green zone.
22
Preparation Connecting the gas supply
N2O AIR O2 VAC
12
Julian
D
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Connecting the backup gas cylinders for O2 and N2O
Even if the workstation is connected to a central gas supply, the cylinders must remain on the apparatus as backup supply.
On the back of the workstation:
place full cylinders in the cylinder holders and secure in position.
1 Screw the pressure reducing adapters onto the
cylinder valves.
2 Screw the gas hoses into the ports on the front of the
gas supply block.
Switch over immediately to cylinder supply if the central gas supply fails:
Open the cylinder valves.
Caution when handling O2 cylinders
Do not oil or grease the O2 cylinder valves or O2
pressure reducing adapters, and do not handle with greasy fingers.
Danger of explosion!
The cylinder valves must be opened / closed by hand.
Do not use tools.
If a cylinder valve is leaky or difficult to operate, it
must be repaired by an expert.
Connecting the anaesthetic gas scavenging system
3 Connect the transfer hose to the waste gas port and
to the port of the scavenging system.
4 Connect the scavenging hose to the port of the
scavenging system.
5 Connect the anaesthetic waste gas probe to the
aspirating hose.
6 Ensure that the second connection to the scavenging
system is sealed by a screw plug.
Follow the Instructions for Use of the anaesthetic gas
scavenging system.
23
Preparation
Connecting the gas supply
Connecting the anaesthetic gas scavenging system
N2O
1
1
O2
22
6
3
3
5
4
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Connecting the power supply
Connecting auxiliary systems 1 connect to auxiliary sockets on the back of the
workstation.
The auxiliary sockets are not powered by the uninterruptible power supply UPS in the event of a power failure!
Do not connect HF surgical devices to the auxiliary sockets!
Connecting equipment to the auxiliary sockets may cause the patient leakage current to rise above the permitted values in the event of failure of a protective earth conductor. The risk of electric shock cannot be excluded in such cases. Additional power adapter sockets must not be connected to the auxiliary sockets.
Equipotential bonding
e.g. for intracardial or intracranial operations. 2 Connect one end of the earth cable to one of the
connecting pins on the back of the workstation.
Connect the other end of the earth cable to the
specified equipotential bonding point, e.g. on the operating table or ceiling lamp.
Connecting the power supply
The mains voltage must be the same as specified on the nameplate on the back of the workstation. Voltage range: 90 to 265 V
Plug the mains plug into the wall socket.
24
Preparation Connecting the power supply
1
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Checking the workstation against the checklist.................................... 26
Switching on............................................................................................26
Gas pressures.........................................................................................29
Central gas supply................................................................................... 29
Backup gas cylinders............................................................................... 29
Suction system.........................................................................................30
O2 flush .................................................................................................. 30
O2 emergency metering........................................................................... 30
Preparing Julian for the self-test................................................................31
Self-test...................................................................................................31
Electronics...............................................................................................31
Fresh gas mixer........................................................................................31
Ventilator and breathing system ...............................................................31
System compliance................................................................................ 32
Leakage...................................................................................................32
Emergency start......................................................................................33
25
Starting Up
Contents
Starting Up
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Starting Up
Checking the workstation against the checklist
Preconditions: The workstation must be prepared as described on pages 95 to 105 and assembled ready for operation as described on pages 106 to 111. The gas supply and power supply must be connected.
Switching on 1 Switch on Julian: press the power switch » «.
Version without rotary knob for O2 emergency metering:
After switching on, Julian delivers an O2 flow of 4 L/min for manual ventilation. This O2 flow is switched off automatically by Julian as soon as the »cancel test« key appears after acknowledging the checklist.
The opening screen appears with the current software version after about 10 seconds. Julian now loads its software and tests its internal memory.
The checklist is displayed after about 40 seconds.
Check the components as instructed in the checklist
on the screen.
If the self-test has to be interrupted, e.g. for a quick start in an emergency:
Acknowledge the checklist:
Press the rotary knob. After a few seconds, the »cancel test« softkey appears on the right-hand edge of the screen.
Press the »cancel test« key. See "Emergency start"
on page 33 to continue.
Starting Up Checking the workstation against the checklist
26
Julian
D
1
Technology for life
SW-version 2.0
01. Nov. 98
Please wait !
Julian
Check-list Julian
Vapor Safety fill Vapor Mont Interlock system
Handwheel set to zero, fill level o.k. Stopper inserted and locked into position Vaporiser present and locked into position non used vapor blocked
Watertrap Gas scavenging system Breathing system Soda lime Emergency breathing bag
level ok ? connected and ready fir use complete and fixed in place no colour change present and functional
Central gas supply Back up cylinders Suction unit O
2
-Flush
O
2
pressure, AIR, N2O pressure, 2,7 bar
O
2
pressure > 50 bar, N2O pressure > 30 bar functional audible flow, breathing bag fills
Julian o.k. according to check-list? Please confirm !
Examine machine according to check-list prior to operation
Before commencing test, close y-piece, connect gas sample-line
and set APL valve to 30 mbar and "Man"
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
"Vapor" anaesthetic vaporizer (example: Vapor 19.3) Only use Vapor 19.3, Vapor 2000 or Devapor.
Follow the Instructions for Use for the particular vaporizer used.
1 Handwheel to "0" and engaged. 2 Filled to sufficient level. 3 Sealing slide valve inserted and screwed tight.
The plug-in adapter must fit evenly on the connection
block.
4 The locking lever must point to the left = locked. 5 The unused Vapor unit must be locked by the
interlock slider (example: left-hand Vapor locked).
Water trap
Remove water trap container by pulling downwards
and empty if necessary.
Observe the hygiene regulations of the hospital.
Replace container from below.
When using the "Waterlock" water trap
See separate Instructions for Use
27
Starting Up
Checking the workstation against the checklist
33
4
5
4
1
2 2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
AGS anaesthetic gas scavenging system 1 The transfer hose from the waste gas port must be
connected.
2 The scavenging hose must be connected;
the anaesthetic waste gas probe must be plugged into the Dräger wall socket and the indicator must be green.
3 The float must be between the two marks.
Breathing system
Complete and engaged, with both valve discs
inserted, bellows fitted and hoses securely connected.
Soda lime renewed, no violet discoloration.
28
Starting Up Checking the workstation against the checklist
3
21
2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Emergency ventilating bag
Example: Dräger Resutator 2000
Ready for operation.
Central gas supply Gas pressures: 1 All pressure gauges in the green zone.
Backup gas cylinders 2 Slowly open the cylinder valves.
Check the cylinder pressure gauges:
3 O2 pressure greater than 50 bar, 4 N2O pressure greater than 30 bar.
2 Reclose the cylinder valves.
29
Starting Up
Checking the workstation against the checklist
2
2
4
3
N2O O2
Julian
D
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Suction system 1 Switch to I. 2 Set suction pressure with rotary knob »Vac.«. 3 Block the secretion viewing window or kink the
suction hose.
4 Measure the suction pressure on the pressure gauge. 1 Switch to 0.
O2 flush 5 Press »O2 +« button. 6 The breathing bag inflates. 7 O2 flows from the Y-piece.
Version with rotary knob for O2 emergency metering 8 Turn on rotary knob »Safety O2« for O2 emergency
metering.
6 The breathing bag inflates. 7 O2 flows from the Y-piece. 8 Turn off rotary knob »Safety O2« for O2 emergency
metering again.
30
Starting Up Checking the workstation against the checklist
7
Julian
D
6
5
8
Julian
D
4
1
3
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Loading...
+ 114 hidden pages