"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
How to use these Instructions for Use
In the header is the subject…of the main chapter.
Underneath is the title of the subchapter, to help you find your way around quickly.
On each page…the instructions for use
combining text with illustrations. The information istranslated directly into actions to enable the user to learn „hands-on“ how to use the workstation.
Left-hand column: the text…
provides explanations and guides the user clearly andergonomically with brief directions on the use of theproduct. Bullet points indicate the steps to be followed, and where there are several steps, numbers refer to theillustration and indicate the sequence.
Right-hand column: the illustration…
provides the link with the text and serves as a guide to the workstation. Points mentioned in the text areemphasised and non-essential information is omitted.
Screen displays guide the user and confirm the steps to be followed.
How to use these Instructions for Use
2
1Press the »MAN/SPONT« key,2and confirm with the rotary knob.
Display (example):
O2 flush
–For flushing and rapidly filling the breathing
system and breathing bag with O2 whilebypassing the Vapor unit.
3Press the »O2 +« button.
O2 flows into the breathing system withoutanaesthetic gas as long as the button is helddown.
Julian
1
2
OperationSelecting the ventilation modeMAN/SPONT
Julian
D
3
MAN/SPONT
Pleth
40
0
CO
2
4036
0.80.65856
38
9867
5.28
4.0030
Volumeter starten:Bestätigen!
O
2
SpO
2
N2O
Hal.
Fi
Fet
etCO
2
AMV
Freq
GrenzenCO
2
-Al.
aus/einAlarm
InfoListe
Kurven
Konfig.
Frischgas O
2
+ N2O
O
2
%
L/min
VT
0.35
0
Volumeter
--s
--
05
10
0.5
1
Frischgas
internextern
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
3
Contents
Index142
For Your Safety and that of Your Patients5
Intended Use7
Operating Concept11
Before Using for the First Time19
Preparation21
Starting Up25
Operation35
Monitoring55
Configuring in Standby Mode81
Care95
Julian as Wall-mounted Unit113
Maintenance Intervals115
What's what125
Technical Data131
Fault – Cause – Remedy119
Abbreviations / Symbols139
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
4
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
For Your Safety and that of YourPatients
Strictly follow the Instructions for Use
Any use of the apparatus requires full understanding andstrict observation of these instructions. The apparatus is only to be used for purposes specifiedhere.
Maintenance
The apparatus must be inspected and serviced regularlyby trained service personnel at six monthly intervals (and a record kept). Repair and general overhaul of the apparatus may only be carried out by trained service personnel.We recommend that a service contract be obtained withDrägerService and that all repairs also be carried out bythem. Only authentic Dräger spare parts may be used formaintenance. Observe chapter "Maintenance Intervals".
Accessories
Do not use accessory parts other than those in the order list.
Not for use in areas of explosion hazard
This apparatus is neither approved nor certified for use inareas where combustible or explosive gas mixtures arelikely to occur.
Safe connection with other electrical equipment
Electrical connections to equipment which is not listed inthese Instructions for Use should only be made followingconsultations with the respective manufacturers or anexpert.
Liability for proper function or damage
The liability for the proper function of the apparatus isirrevocably transferred to the owner or operator to theextent that the apparatus is serviced or repaired bypersonnel not employed or authorized by DrägerServiceor if the apparatus is used in a manner not conforming toits intended use.
Dräger cannot be held responsible for damage causedby non-compliance with the recommendations givenabove. The warranty and liability provisions of the termsof sale and delivery of Dräger are likewise not modifiedby the recommendations given above.
Dräger Medizintechnik GmbH
For Your Safety and that of Your Patients
5
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
–Peak pressure, mean pressure, plateau pressure and
PEEP–Expiratory minute ventilation
Tidal volume VT
Breathing rate
Patient compliance–Inspiratory and expiratory concentration of O2, N2O,
anaesthetic gas and CO2optional:–Functional oxygen saturation (SpO2) and pulse rate.
The following parameters are displayed as curves:
–Airway pressure–Expiratory flow–Inspiratory and expiratory concentration of O2, CO2
and anaesthetic gasoptional:–PlethysmogramTrend curves and measured value lists are also available.
Monitoring
by means of adjustable alarm limits that are automaticallyadapted to the ventilation mode.
The workstation may only be used under thesupervision of qualified medical personnel, so thatassistance can be provided immediately in the eventof any malfunctions.
Explosive anaesthetics, such as ether or cyclo-propane, must not be used due to the risk of fire.
Mobile radio telephones must not be used within 10 metres of the workstation!
Mobile telephones may interfere with the operation ofelectrical and electronic medical equipment.
Julian must not be used with nuclear spintomography (MRT, NMR, NMI)!
Operation of the apparatus may be impaired.
Intended Use
9
Intended Use
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
10
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Three basic screens for monitoring...........................................................17
The standard screen.................................................................................17
The data screen.......................................................................................17
The trend screen......................................................................................18
Operating Concept
Contents
11
Operating Concept
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Operating Concept
Screen ergonomics
All the settings required for–Fresh gas delivery–Ventilation–Monitoringare entered on the system screen using the appropriate
keys and the rotary knob.The keys are grouped in function fields:
Left-hand fieldFresh gas delivery
Right-hand fieldVentilation
Middle fieldMonitoring
The main functions for anaesthesia, e.g. selection of N2O or AIR, or selection of ventilation modes, can beselected directly by keys with permanently definedfunctions ("hardkeys"):
1Left-hand block
The »N2O« or »AIR« keys are used to select the gas to be mixed with O2 for the fresh gas mix.
2Right-hand block
The »MAN SPONT«, »IPPV«oder »PCV«keys select the ventilation mode.
These function keys are located in the bottom row on thecontrol panel.Left-hand block for fresh gas settingRight-hand block for ventilation
Operating ConceptScreen ergonomics
12
.
12...
.
NO
2
AIR
MAN
SPONT
IPPVPCV
Monitoring
Ventilation
Fresh gas delivery
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
.
12...
.
NO
2
AIR
MAN
SPONT
IPPVPCV
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
1
2
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Complementary "softkeys" with variable functions areprovided at the bottom edge of the screen, above eachgroup of hardkeys. These softkeys are used to set thefresh gas delivery parameters and ventilation parameters.
1Left-hand block:
The keys for setting the O2 concentration and freshgas flow.
2Right-hand block:
The keys for setting the parameters for the relevantventilation mode. The example shows the parametersfor IPPV controlled ventilation.
These softkeys have different functions, depending onthe operating status or ventilation mode.
The current parameter values are displayed in the softkeyfield.
In a prominent position at the bottom right-hand side:The "turn-and-push" rotary knob
is the main operating control of the apparatus and has thefollowing functions in all setting operations:
3select = turn4confirm = press
–to confirm the selected carrier gas or a ventilation
mode,
–to set and confirm the parameters for fresh gas and
ventilation modes,
–to set and confirm the monitoring functions.
Beside the rotary knob:The standby key E for switching over to standby mode.5Press standby key E and confirm by pressing the
rotary knob.
13
Operating Concept Screen ergonomics
.
12...
.
NO
2
AIR
MAN
SPONT
IPPVPCV
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
12
_
CO
2
Ve
.
12...
.
NO
2
AIR
MAN
SPONT
IPPVPCV
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
3
5
4
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operating ConceptScreen ergonomics
14
Selecting / setting ventilation parameters
Example: PEEP ventilation parameter
1Press the softkey »PEEP«.2Select the PEEP value = turn the rotary knob.3Confirm the PEEP value = press the rotary knob.
The keys for the various monitoring functions are locatedon the right-hand side of the screen.
These keys also have different functions = softkeys,depending on the monitoring screen required.
Julian
1
2
3
.
12...
.
NO
2
AIR
MAN
SPONT
IPPVPCV
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
15
Operating Concept Screen ergonomics
Selecting / setting monitoring functions
For example, to change the lower alarm limit of the end-tidal CO2 concentration.
1Press the »alarm limits« softkey. The alarm limits
menu is displayed on the screen.
●Select the alarm limit value = turn the rotary knob.
Confirm the selection = press the rotary knob.The limit value is highlighted.
Example: lower alarm limit etCO2: <30
●Set the alarm limit value = turn the rotary knob.
●Confirm the new alarm limit = press the rotary knob.
The cursor returns to the zsymbol.
Julian
1
Pleth
IPPV
CO2
40
0
PAW
60010
1:2
100251.0050
986738
6.029
0.8
--
95
3.0
--
20
1.0
0.040
8
120
50
--
Select limit: Turn and confirm ! !
4230
0
20
FiO
2
FiHal.
PAW
SpO
2
etCO
2
MV
alarmlimits
auto set limits
alarminfo
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
Pleth
IPPV
CO
2
40
0
60010
1:2
100251.0050
986738
6.029
0.8
--
95
3.0
--
20
1.0
0.040
8
120
50
--
Adjust limit value: Turn and confirm !
42
30
PAW
0
20
FiO
2
FiHal.
PAW
SpO
2
etCO
2
MV
alarmlimits
auto set limits
alarminfo
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O2 + N2O
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operating ConceptScreen ergonomics Screen layout
16
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
➀ ➁ ➂ ➃
➄ ➅ ➆ ➇
Exit the alarm limits menu:1Press the rotary knob
or
2Press the Q key.
The function keys for standard functions are located onthe right-hand side of the control panel.
GSuppress the acoustic alarm for 2 minutes.SSelect the screen page.QBack to standard page.
Screen layout
➀Status field:
Displays information on the current operating mode
➁Graphics field:
For curves and bargraphs
➂Alarm field:
Displays information on the alarms, warnings, etc. and their priority
➃Numerical value field:
For numerical values
➄Lower softkeys – for fresh gas parameters➅Prompt field:
For user guidance
➆Lower softkeys – for ventilation parameters➇Right-hand softkeys – for monitoring:
For rapid selection of the desired monitoring function
.
12...
.
NO
2
AIR
MAN
SPONT
IPPVPCV
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
Julian
2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
17
Operating Concept
Screen layout
Three basic screens for monitoringTo call up the standard screen, data screen and trend
screen in succession:1Briefly press the S key until the desired screen is
displayed.
Back to the standard screen:2Press the Qkey.
The fresh gas and ventilation parameters can be set bymeans of the softkeys in each of the three basic screens.
The standard screen
with three selectable curves.The most important numerical values are displayed ingroups on the right-hand side of the screen.
The data screen
contains all the numerical values with their units ofmeasure.
Julian
21
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
IPPV
PAW
40
20
0
60010
1:2
100251.0050
O
2
%
N
2
O
%
Hal.%
036
4033
0.80.65856
Fi
Fet
CO
2
mmHg
9867
6.0
0.6010
2720
51565
MVV
T
freq
peakplat.PEEPmeancompliance
mbar
mL/mbar
L/min
L
34
AW-temp.
°C
1/min
01. Nov. 98
sys. comp. 6.40 mL/mbarleckage 10 mL/min
01-10
8 : 00
12 : 00
SpO
2
%
1/min
alarmlimits
auto se
t
limits
alarminfo
list
config.
Freshgas O
2
+ N2O
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operating ConceptScreen layout
18
The trend screen
displays the recorded progress of the numerical valuesover time since measurement started. The currentmeasured values are displayed on the right-hand side.
Display (example):Trends for CO2 and minute ventilation MV
IPPV
4036
0.80.65856
38
9867
6.010
60010
1:2
100251.0050
PAW
40
20
0
11:0012:0013:00
60
30
0
15
10
5
0
Select zoom area:
Turn !
Zoom in: Press!
CO
2
MV
O
2
SpO
2
N2O
Iso.
Fi
Fet
etCO
2
MV
freq
alarmlimits
AGasN
2
O
CO
2
MV
O
2
compl.SpO
2
pulsefull
trend
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Charging the battery for emergency operation......................................20
When Julian is not in use........................................................................20
19
Before Using for the First Time
Contents
Before Using for the First Time
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Before Using for the First Time
Fit the enclosed O2 sensor, page 117 "Replacing O2 sensor"Fit the flow sensor, page 106
Charging the battery for emergencyoperation
Julian has a built-in uninterruptible power supply UPSwhich maintains the power supply for 30 minutes in theevent of a mains failure, provided that the battery ischarged. The switch-over to the battery-powered UPS is automaticand is indicated on the screen by the message:POWER FAIL!
The battery recharges automatically when the workstationis plugged into the mains.
The battery must be charged for 10 hours before usingthe Workstation for the first time.
●Plug the mains plug of the Julian workstation into the
mains socket. The mains voltage must correspond tothe voltage specified on the nameplate.
1The green LED
NN
NN
lights up.
●Leave Julian connected to the mains for 10 hours.
The workstation does not have to be switched on.
The devices connected to auxiliary power sockets willnot be powered by the UPS in the event of a powerfailure.
When Julian is not in use
●Charge battery at least every 4 weeks.
Allowing it to run down can lead to damage.
If Julian is out of use for an extended period:
●Leave the workstation connected to the mains at all
times.
1The green LED
NN
NN
is lit.
20
Before Using for the First TimeCharging the battery for emergency operationWhen Julian is not in use
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Connecting the gas supply.....................................................................22
Connecting the backup gas cylinders for O2 and N2O .............................23
Caution when handling O2 cylinders.........................................................23
Connecting the anaesthetic gas scavenging system.............................23
Connecting the power supply.................................................................24
Connecting the power supply...................................................................24
21
Preparation
Contents
Preparation
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Preparation
Use only cleaned and disinfected parts!
Connecting the gas supply
1Screw the central gas supply pressure hoses for O2,
AIR and N2O into the ports on the back of the gassupply block.The two ports at the front are reserved for the backupgas cylinders.
2 The vacuum supply block (VAC) is available as option
for secretion aspiration.Plug the other end of the pressure hoses into the wall
supply points.
●Make sure the gas pressures of the central gas
supply are between 2.7 and 5.5 bar:
3All three pressure gauges in the green zone.
22
PreparationConnecting the gas supply
N2O AIR O2 VAC
12
Julian
D
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Connecting the backup gas cylinders for O2 and N2O
Even if the workstation is connected to a central gassupply, the cylinders must remain on the apparatus asbackup supply.
●On the back of the workstation:
place full cylinders in the cylinder holders and securein position.
1Screw the pressure reducing adapters onto the
cylinder valves.
2Screw the gas hoses into the ports on the front of the
gas supply block.
Switch over immediately to cylinder supply if the centralgas supply fails:
●Open the cylinder valves.
Caution when handling O2 cylinders
●Do not oil or grease the O2 cylinder valves or O2
pressure reducing adapters, and do not handle withgreasy fingers.
Danger of explosion!
●The cylinder valves must be opened / closed by hand.
Do not use tools.
●If a cylinder valve is leaky or difficult to operate, it
must be repaired by an expert.
Connecting the anaesthetic gas scavengingsystem
3Connect the transfer hose to the waste gas port and
to the port of the scavenging system.
4Connect the scavenging hose to the port of the
scavenging system.
5Connect the anaesthetic waste gas probe to the
aspirating hose.
6Ensure that the second connection to the scavenging
system is sealed by a screw plug.
●Follow the Instructions for Use of the anaesthetic gas
scavenging system.
23
Preparation
Connecting the gas supply
Connecting the anaesthetic gas scavenging system
N2O
1
1
O2
22
6
3
3
5
4
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Connecting the power supply
Connecting auxiliary systems1connect to auxiliary sockets on the back of the
workstation.
The auxiliary sockets are not powered by theuninterruptible power supply UPS in the event of apower failure!
Do not connect HF surgical devices to the auxiliarysockets!
Connecting equipment to the auxiliary sockets may causethe patient leakage current to rise above the permittedvalues in the event of failure of a protective earthconductor.The risk of electric shock cannot be excluded in suchcases.Additional power adapter sockets must not be connectedto the auxiliary sockets.
Equipotential bonding
e.g. for intracardial or intracranial operations.2Connect one end of the earth cable to one of the
connecting pins on the back of the workstation.
●Connect the other end of the earth cable to the
specified equipotential bonding point, e.g. on theoperating table or ceiling lamp.
Connecting the power supply
The mains voltage must be the same as specified on thenameplate on the back of the workstation.Voltage range: 90 to 265 V
●Plug the mains plug into the wall socket.
24
PreparationConnecting the power supply
1
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Checking the workstation against the checklist....................................26
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Starting Up
Checking the workstation against the checklist
Preconditions:The workstation must be prepared as described onpages 95 to 105 and assembled ready for operation asdescribed on pages 106 to 111.The gas supply and power supply must be connected.
Switching on1Switch on Julian: press the power switch » «.
Version without rotary knob for O2 emergencymetering:
After switching on, Julian delivers an O2 flow of 4 L/minfor manual ventilation.This O2 flow is switched off automatically by Julian assoon as the »cancel test« key appears afteracknowledging the checklist.
The opening screen appears with the current softwareversion after about 10 seconds. Julian now loads itssoftware and tests its internal memory.
The checklist is displayed after about 40 seconds.
●Check the components as instructed in the checklist
on the screen.
If the self-test has to be interrupted, e.g. for a quick startin an emergency:
● Acknowledge the checklist:
Press the rotary knob.After a few seconds, the »cancel test« softkey appears on the right-hand edge of the screen.
●Press the »cancel test« key. See "Emergency start"
on page 33 to continue.
Starting UpChecking the workstation against the checklist
26
Julian
D
1
Technology for life
SW-version 2.0
01. Nov. 98
Please wait !
Julian
Check-list Julian
VaporSafety fillVapor MontInterlock system
Handwheel set to zero,fill level o.k.Stopper inserted and locked into positionVaporiser present and locked into positionnon used vapor blocked
WatertrapGas scavenging systemBreathing systemSoda limeEmergency breathing bag
level ok ?connected and ready fir usecomplete and fixed in placeno colour changepresent and functional
Central gas supplyBack up cylindersSuction unitO
2
-Flush
O
2
pressure, AIR, N2O pressure, 2,7 bar
O
2
pressure > 50 bar, N2O pressure > 30 barfunctionalaudible flow, breathing bag fills
Julian o.k. according to check-list? Please confirm !
Examine machine according to check-list prior to operation
Before commencing test, close y-piece, connect gas sample-line
and set APL valve to 30 mbar and "Man"
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
"Vapor" anaesthetic vaporizer (example: Vapor 19.3)Only use Vapor 19.3, Vapor 2000 or Devapor.
Follow the Instructions for Use for the particularvaporizer used.
1Handwheel to "0" and engaged.2Filled to sufficient level.3Sealing slide valve inserted and screwed tight.
●The plug-in adapter must fit evenly on the connection
block.
4The locking lever must point to the left = locked.5The unused Vapor unit must be locked by the
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
AGS anaesthetic gas scavenging system1The transfer hose from the waste gas port must be
connected.
2The scavenging hose must be connected;
the anaesthetic waste gas probe must be pluggedinto the Dräger wall socket and the indicator must begreen.
3The float must be between the two marks.
Breathing system
●Complete and engaged, with both valve discs
inserted, bellows fitted and hoses securelyconnected.
●Soda lime renewed, no violet discoloration.
28
Starting UpChecking the workstation against the checklist
3
21
2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Emergency ventilating bag
Example: Dräger Resutator 2000
●Ready for operation.
Central gas supply Gas pressures:1All pressure gauges in the green zone.
Backup gas cylinders 2Slowly open the cylinder valves.
Check the cylinder pressure gauges:
3O2 pressure greater than 50 bar,4N2O pressure greater than 30 bar.
2Reclose the cylinder valves.
29
Starting Up
Checking the workstation against the checklist
2
2
4
3
N2OO2
Julian
D
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Suction system1Switch to I.2Set suction pressure with rotary knob »Vac.«.3Block the secretion viewing window or kink the
suction hose.
4Measure the suction pressure on the pressure gauge.1Switch to 0.
O2 flush5Press »O2 +« button.6The breathing bag inflates.7O2 flows from the Y-piece.
Version with rotary knob for O2 emergency metering8Turn on rotary knob »Safety O2« for O2 emergency
metering.
6The breathing bag inflates.7O2 flows from the Y-piece.8Turn off rotary knob »Safety O2« for O2 emergency
metering again.
30
Starting UpChecking the workstation against the checklist
7
Julian
D
6
5
8
Julian
D
4
1
3
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Preparing Julian for the self-test
●Make sure that the breathing system is securely
connected.
1Close the Y-piece = plug firmly on to the cone.2Ensure that the sample line is connected to the
Y-piece and to the water trap on the back of theworkstation.
Self-test
If all points on the checklist are OK:
●Confirm = press the rotary knob.
The self-test is started. The self-test is run automatically and takes 3 to 4 minutes.
Julian carries out the following automatic tests andactions:
Electronics
–Testing and calibration of gas metering bench–Testing and calibration of O2 sensor–Testing of pressure sensor for airway pressure
measurement–Testing the flow sensor–Testing the battery–Activating the default settings for alarm limits,
monitoring parameters and the default settings of
the ventilator and fresh gas settings
Fresh gas mixer
–Testing of gas supply O2, N2O, AIR –Testing of gas inlet valves–Calibration of pressures sensors in the compressed
air supply–Testing of O2 emergency metering –Testing of fresh gas failure monitoring (optional)
Ventilator and breathing system
–Testing and calibration of PEEP/Pmax valves–Testing of safety valve–Testing of MAN/AUTO switch-over valve –Calibration of flow monitoring valve–Testing of time monitoring valve–Testing of bellows
31
Starting Up
Checking the workstation against the checklist
2
2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
System compliance
Julian determines the current system compliance.Depending on the breathing hoses used, the systemcompliance is 5 to 6 mL/bar.
Leakage
Julian determines the current leakage of the breathingsystem and breathing hoses.The system tolerates leaks of up to 150 mL/min.
For leaks of more than 150 mL/min:
●Check the breathing system and repeat the leak test.
Display (example):A tick (✓) indicates that the relevant test point has been
completed successfully.The clock symbol uuuu indicates the test stage is currentlyin progress.
Malfunctions detected in the self-test and any gas failuresare displayed on the screen. Some malfunctions can beaccepted by confirming = pressing the rotary knob, forexample no AIR supply (AIR FAILURE!!!).Other malfunctions need to be corrected before startingup the workstation, for example no O2 supply (O2 FAILURE!!!).
Julian switches to standby after the self-test.Display:
Starting UpChecking the workstation against the checklist
32
Self-test
electronics
freshgasmixer
ventilatorandbreathing system
systemcompliance
leakage
✓
✓✓
3.40 mL/mbar
20 mL/mbar
canceltest
deletetrend
leakagetest
config.
Julian
SW-version 2.0
01. Nov. 98
Standby
4.00
L/min
30
Operation with,or
MAN
SPONT
IPPV
PCV
O
2
%
Freshgas O2+N2O
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Emergency start
This procedure should only be used when there is notime for the self-test!
1Switch on the workstation.
Version without rotary knob for O2 emergency
metering:
After switching on, Julian delivers an O2 flow of
4 L/min for manual ventilation.
Version with rotary knob for O2 emergency
metering:2Set rotary knob »Safety-O2« for O2 emergency
metering to the desired O2 flow.
Range 0 to 12 L/min.
●Wait for the software to be internally loaded and for
the electronics to be self-tested.
The checklist appears after about 40 seconds.
●Confirm the checklist: press the rotary knob.
After a few seconds, the »Cancel test« softkey is
displayed on the right-hand edge of the screen.
Display:
●Press the softkey »Cancel test«.
The apparatus only runs through a minimal test.
Julian is ready for operation about 1 minute afterswitching on. Calibration of the O2 sensor is completeafter about 2 minutes.
The leakage and compliance test is not performed.The accuracy levels specified in the "Technical Data"cannot be guaranteed.
To prevent abuse of this facility, the self-test can onlybe cancelled 10 times in succession.
The self-test cannot be cancelled the next time that Julianis started and a complete self-test must be run through.
33
Starting Up
Emergency start
Julian
D
1
2
Self-test
electronics
freshgasmixer
ventilatorandbreathing system
✓
canceltest
39
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
The workstation switches to standby mode aftercompleting the minimal test.
Display (example):
If no key is pressed in standby mode for two minutes, the standby screen is switched off and a screen saverwith the Dräger logo is displayed instead.
To switch the standby screen on again:
●Press the rotary knob or any other key.
To start Julian:
Version with rotary knob for O2 emergency metering:
●Turn rotary knob for O2 emergency metering to 0.
●Select fresh gas setting and ventilation mode.
See »Operation«, page 36 onwards.
34
Starting UpEmergency start
deletetrend
leakage test
config.
Julian
SW version 2.0
01. Nov. 98
Standby
2.00
L/min
100
Operation with, or
MAN
SPONT
IPPV
PCV
Freshgas O
2
O
2
%
First interruption of self-test
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Setting the fresh gas concentrations.....................................................36
Connecting the breathing hoses...............................................................49
Using non-rebreathing systems..............................................................50
In the event of a power failure................................................................52
In the event of a gas failure....................................................................53
In the event of a Major Hardware Fault..................................................53
End of Operation.....................................................................................54
35
Operation
Contents
Operation
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Operation
Setting the fresh gas concentrations
The fresh gas settings are displayed in the Standbyscreen:
–»Fresh gas O2 + N2O«
or
»Fresh gas O2 + AIR«–O2 concentration »O2 %«–Fresh gas flow »L/min«The settings correspond to the configurable default
values after switching on and after entering standbymode. The fresh gas settings can be modified beforeselecting the ventilation mode.Fresh gas does not flow in standby mode. The fresh gasflow is not enabled until a ventilation mode has beenstarted.
Adjustment ranges and default settings on delivery
Selecting the carrier gas1Press the »N2O« or »AIR« hardkey.
The yellow LED in the selected key flashes.
●Press the rotary knob to confirm. The yellow LED
lights up constantly.
The selected fresh gas components are displayed on thescreen.
Setting the O2 concentration2Press the softkey »O2%«.
The key field is highlighted against a whitebackground.
●Set and confirm the O2 concentration by means of the
rotary knob.
* Only in MAN/SPONT mode
OperationSetting the fresh gas concentrations
36
Julian
1
2
deletetrend
leakagetest
config.
Julian
SW-version 2.0
01. Nov. 98
Standby
4.00
L/min
30
Operation with,or
MAN
SPONT
IPPV
PCV
O
2
%
Freshgas O2+N2O
Fresh gas parameters
Adjustment range
Default setting
on deliveryCarrier gasAIR or N2ON2OO2 %25 to 100100Fresh gas flow
L/min
0* ; 0,5 to 122
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Setting the fresh gas flow1Press the softkey »L/min«.
The key field is highlighted against a light background.
●Set and confirm the fresh gas flow by means of the
rotary knob.
Julian is fitted with an electronic O2 minimum dosingsystem to avoid hypoxic gas mixtures. For fresh gas flowsettings below 1 L/min, the O2 concentration isautomatically increased to a value corresponding to anO2 flow of 250 mL/min. If this control system isactivated, the O2 value is highlighted in addition to theactive softkey »L/min«.
Version with fresh gas failure detection (optional):During operation, Julian checks that the bellows aresufficiently full.
If the message „Fresh gas ? !!“ appears:
●Increase the fresh gas flow
The default settings valid whenever Julian is switched oncan also be modified; see "Setting default values" onpage 82.
Selecting the ventilation mode
MAN/SPONT ventilation mode
Choose between manual ventilation MAN and sponta-neous breathing SPONT on the APL pressure-limitingvalve.
Manual ventilation2Set the lever of the APL pressure-limiting valve to
MAN.
3Set the pressure limit = rotate lever.
Spontaneous breathing4Set the lever of the APL pressure-limiting valve to
SPONT.
The valve is open for a free spontaneous breathingstroke regardless of the set pressure limit.
37
Operation
Setting the fresh gas flow
Selecting the ventilation mode
MAN/SPONT
Julian
1
MAN
MAN
3
2
SPONT
SPONT
4
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
1Press the »MAN/SPONT« key,2and confirm with the rotary knob.
Display (example):
O2 flush
–For flushing and rapidly filling the breathing system
and breathing bag with O2 while bypassing the Vapor unit.
3Press the »O2 +« button.
O2 flows into the breathing system withoutanaesthetic gas as long as the button is held down.
OperationSelecting the ventilation modeMAN/SPONT
38
Julian
1
2
Julian
D
3
MAN/SPONT
Pleth
40
0
CO
2
4036
0.80.65856
38
9867
5.28
Start volumeter:Confirm !
O
2
SpO
2
N2O
Hal.
Fi
Fet
etCO
2
MV
freq
alarmlimits
CO
2
-al.
on/offalarm
infolist
curves
config.
Freshgas O
2
+ N2O
fresh gas
O
2
%
L/min
VT
0.35
0
Volumeter
--s
--
0510
0.5
1
_
CO
2
Ve
internalexternal
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
39
Operation
Selecting the ventilation mode
MAN/SPONT
Certain alarms are automatically switched off inMAN/SPONT ventilation mode in order to avoid artefacts.The deactivated alarms are identified by the grey back-ground in the table.
– – :The default setting on delivery is outside the
monitoring range and the corresponding alarm limitis switched off.
To set the values of the alarm limits during operation seepage 60.
To set the default alarm limits see page 86.
Alarm limitsin MAN/SPONT
mode
Default setting on delivery
SpO2 ON
ON
– –
92
Pulse
H
ONON
120
50
etCO2ON
ON
50 (mmHg)
– –
FiCO2 OFF
MV OFF
OFF
– –– –
FiO2 OFF
ON
– –
20
Fi Hal. ON
OFF
1.5– –
Fi Iso. ON
OFF
2.3– –
Fi Enf. ON
OFF
3.4– –
Fi Des.ON
OFF
12.0– –
Fi Sev. ON
OFF
3.4– –
PAWON
OFF
40
– –Apnoea pressureOFFApnoea flowOFFApnoea CO2
ON
activated after
60 seconds
_
_
_
_
_
_
_
_
_
_
_
>
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
Volume-controlled ventilation with fixed mandatory minutevolume (MV), set with the tidal volume (VT), breathingrate (Freq.) and the ratio of inspiration time to expiration(TI:TE ).
Presetting the ventilation parameters for IPPV1Press the »IPPV« hardkey; its LED flashes.
●The six ventilation parameters for IPPV are displayed
on the screen with their default settings.
2Press the softkey for the desired ventilation
parameter.
3Set and confirm the ventilation parameter with the
rotary knob.
Display (example):
Adjustment ranges and default settings on delivery
For tidal volumes VT of less than 200 mL:
●Use the paediatric hose set,
see "Ventilating children", page 49.
In IPPV ventilation mode, automatic system compliancecompensation takes place, and the applied tidal volumetherefore corresponds to the set volume.
*The default values can be set specifically for each hospital,
see page 82.
40
OperationSelecting the ventilation modeIPPV
Julian
2
3
1
Ventilation parametersAdjustment
range
Default setting on delivery *
Pressure limit Pmax [mbar]
10 to 7025
Tidal volume VT[mL]
50 to 1400600
Frequency Freq.[1/min]
6 to 6012
Insp/Exp ratioTI : TE
2 : 1, 1 : 1, 1 : 2, 1 : 3, 1 : 4
1 : 2
Insp. pause time :Insp. time TIP : TI [%]
0 to 5010
PEEP [mbar]0 to 200
deletetrend
leakagetest
config.
Julian
SW- version 2.0
01. Nov. 98
Standby
To Start IPPV: Confirm !
4.00 30
60012
1:2
100
25
O2 %
L/min
V
T
mL
PEEP
mbar
Freq.
1/min
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O2+N2O
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Starting IPPV1Press the »IPPV« hardkey2and confirm with the rotary knob.
Display (example):
The preset ventilation parameters are displayed on thescreen.
The symbol rotates in the softkey »L/min« to indicatethat fresh gas is flowing.
If a change in ventilation parameters is required:
●Press the corresponding softkey. Set and confirm the
ventilation parameter with the rotary knob.
41
Operation
Selecting the ventilation mode
IPPV
Julian
1
2
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
OperationSelecting the ventilation modeIPPV
42
Alarms effective in IPPV mode
– – : The default setting on delivery is outside the
monitoring range and the corresponding alarm limitis switched off.
To set the values of the alarm limits during operation seepage 60.
To set the default alarm limits see page 86.
Alarm limitsIn IPPV modeDefault setting
on delivery
SpO2 ON
ON
– –
92
Pulse
H
ONON
120
50
etCO2ON
ON
50 (mmHg)
– –
FiCO2ON Fixed upper alarm
limit 5 mmHg
MV ON
ON
– –
3.0
FiO2 ON
ON
– –
20
Fi Hal. ON
ON
1.5– –
Fi Iso. ON
ON
2.3– –
Fi Enf. ON
ON
3.4– –
Fi Des.ON
ON
12.0– –
Fi Sev. ON
ON
3.4– –
PAWON
ON
40
8
Apnoea pressureONActive after
15 seconds
Apnoea flowONActive after
15 seconds
Apnoea CO2ONActive after
30 seconds
_
_
_
_
_
_
_
_
_
_
>
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
In pressure controlled ventilation mode, the applied tidalvolume depends on the ventilation parameters Pmax,insp. Flow, insp. Time, PEEP and the lung compliance.Changes in ventilation parameters or lung complianceinfluence the tidal volume and the MV minute volumemust therefore be monitored constantly.
Presetting ventilation parameters for PCV1Press »PCV« its LED flashes.
●The five ventilation parameters for PCV appear on the
screen with the default settings.
2Press the softkey of the relevant ventilation parameter.3Set and confirm the ventilation parameter with the
rotary knob.
Display (example):
Adjustment ranges and default settings on delivery
When ventilating infants weighing less than 20 kg:
●Use the paediatric hose set;
see page 49 "Ventilating children".
–––––––––––*The default settings can be set specifically for each hospital,
see page 82.
**Can be set by DrägerService.
43
Operation
Selecting the ventilation mode
PCV
1
Julian
2
1
3
Ventilation parameterAdjustment
range
Default settingon delivery*
Pressure limitPmax [mbar]
(PEEP+1)
to 70
15
Frequency Freq.[1/min]
6 to 6012
Insp./Exp. ratio [TI : TE]
2: 1, 1: 1, 1: 2,
1: 3, 1: 4
1 : 2
Insp. flow [L/min]5 to 50 (75**)30PEEP [mbar]0 to 200
Julian
SW version 2.0
01. Nov. 98
Standby
4.00
30
12
1:2
300
15
deletetrend
leakage test
config.
To Start PCV : Confirm !
O2 %L/min
PEEP
mbar
Freq.
1/min
TI:TEInsp.Flow
L/min
Pmax
mbar
Freshgas O2+N2O
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
44
OperationSelecting the ventilation modePCV
Starting PCV1Press »PCV«. 2Confirm by pressing the rotary knob.
Display (example):
The preset ventilation parameters appear on the screen.Fresh gas flows, as indicated by the rotating symbol
in the softkey »L/min«.If a change in ventilation parameters is required:
●Press the relevant softkey. Set and confirm the
ventilation parameter with the rotary knob.
Julian
1
2
Pleth
PCV
CO
2
40
0
PAW
0
20
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
4.310
MV
freq
10
1:2
300184.0030
alarmlimits
alarminfo
list
curves
config.
PEEP
mbar
Freq.
1/min
TI:TEInsp. flow
L/min
Pmax
mbar
Freshgas O
2
+ N2O
O
2
%
L/min
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
45
Operation
Selecting the ventilation mode
PCV
Alarms effective in PCV mode
– – :The default setting on delivery is outside the
monitoring range and the corresponding alarm limitis switched off.
To set the values of the alarm limits during operation seepage 60.
To set the default alarm limits see page 86.
Alarm limitsIn PCV modeDefault setting on
delivery
SpO
2 ON
ON
– –
92
Pulse
H
ONON
120
50
etCO2ON
ON
50 (mmHg)
– –
FiCO2ON Fixed upper alarm
limit 5 mmHg
MV ON
ON
– –
3.0
FiO2 ON
ON
– –
20
Fi Hal. ON
ON
1.5– –
Fi Iso. ON
ON
2.3– –
Fi Enf. ON
ON
3.4– –
Fi Des.ON
ON
12.0– –
Fi Sev. ON
ON
3.4– –
PAWON
ON
40
8
Apnoea pressureONActive after
15 seconds
Apnoea flowONActive after
15 seconds
Apnoea CO2ONActive after
30 seconds
_
_
_
_
_
_
_
_
_
_
>
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Setting the Vapor unit
1Lock the unused Vapor unit
= push lever fully in the direction of the unused Vapor(example: left-hand Vapor locked).
2Press the O button and3set the handwheel to the required anaesthetic agent
concentration.
Aspirating secretion
●Swing the bottles forwards.
1Set the switch to I.2Seal the "fingertip" or kink the suction hose and3Set the appropriate suction pressure for the patient
with the rotary knob »Vac.« and
4check on the pressure gauge.
Observe the hygiene regulations of the hospital!
After aspirating
●Rinse the suction hose with the water purifier.
1Set the switch to 0.
Julian
D
2
4
1
3
46
OperationSetting the Vapor unitAspirating secretion
2
1
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Changing patients
To switch Julian to standby:
●Press the Standby key E and confirm with the
rotary knob.
The functions of the workstation are switched off.The set alarm limits are cancelled and the default alarmsettings are valid again.
The default settings are loaded for dosing anaestheticgas and for the ventilation parameters.
If the ventilation hoses have been changed:
●Perform a leakage test, page 48.
Changing soda lime
–If 2/3 of the absorber charge has turned violet.–If the inspiratory CO2 concentration FiCO2 equals
5 mmHg or more.
●Press the Standby key E and confirm by pressing
the rotary knob.
1Raise the writing top.2Pull the catch and draw out the breathing system at
the same time.
3Turn the absorber counterclockwise and pull it out
downwards.
●Drain the used soda lime and dispose of with
domestic waste.
●Fill the absorber with fresh soda lime up to the MAX
mark.
●Fit the absorber to the breathing system from below
and turn it clockwise as far as it will go.
●Push the breathing system inwards until it clicks into
place.
●Fold down the writing top.
47
Operation
Changing patients
Changing soda lime
1
2
MAX
3
deletetrend
leakagetest
config.
Julian
SW-version 2.0
01. Nov. 98
Standby
4.00
L/min
30
Operation with,or
MAN
SPONT
IPPV
PCV
O
2
%
Freshgas O2+N2O
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Leakage test
Must not be performed when a patient is connected tothe workstation!
–When the soda lime has been changed
or
–when the breathing hoses have been changed.
1Close the Y-piece = place firmly on the cone.2Ensure that the sample line is connected to the
Y-piece and to the water trap at the back of theworkstation.
●Press the softkey »leakage«.
Julian performs the leakage test for IPPV/PCV, durationapprox. 30 seconds.The breathing bag and its hose are not included in thetest.
Display:After the test has been successfully completed, the clock
symbol disappears and Julian displays the values forleakage and system compliance for a few seconds. The results of the leakage test are continually displayedon the data screen.
To return to the standby screen:
●Press the rotary knob.
48
OperationLeakage test
2
2
1
Leakage test
System-compliance
Leakage
Please wait.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Ventilating children
For tidal volumes VT of less than 200 mL:
●Use paediatric hoses.
Connecting the breathing hoses1Use Y-piece with connection for sample line.
●The inspiratory and expiratory
microbial filter 654 St should not be used – thisreduces the system compliance.
2Connect 0.5 L breathing bag with socket to the
breathing hose with the large connection sleeves.Slip the breathing hose over the angled socket. Hang the 0.5 L breathing bag onto the hook.
3Slip the breathing hoses with the large sleeves onto
the inspiratory and expiratory sockets and connect thesmall sleeves to the Y-piece.
4Connect the sample line to the Y-piece and water
trap.
●Set the sample rate to 200 mL/min, page 84.
If the measured gas is not recirculated:
● Increase the tidal volume VT in accordance with the
sample rate.
Volumeter function:
●Select »Child« scale, page 67.
To determine system compliance and leakage:5Firmly connect the Y-piece to the cone.
●Determine the system compliance and leakage,
see „Leakage test“ on page 48.Set a correspondingly higher tidal volume VT inaccordance with the airway pressure PAW.
●Use PCV or IPPV ventilation mode.
49
Operation
Ventilating children
3
2
1
4
5
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Using non-rebreathing systems
Example: Bain system
●Prepare the Bain system in accordance with the
separate Instructions for Use.
For the specified monitoring of O2, CO2 and anaestheticagents:
1Screw the sample line to the Luer lock connection of
the mask manifold pipe and to the water trap on theback of the workstation.
For mask pipes without sample line connector:
●Place a T-piece with filter between the mask pipe and
fresh gas connection port.
or:
●where applicable, use the Luer lock connection of a
filter.
2Connect the fresh gas hose of the Bain system to the
fresh gas outlet.
3Connect the anaesthetic gas scavenging system AGS
to the Bain system.
●Note the Instructions for Use of the Bain system.
Display (example):
In MAN/SPONT ventilation mode:
●Press the softkey »Fresh gas internal/external«,
select »Fresh gas external« and confirm by pressingthe rotary knob.
50
OperationUsing non-rebreathing systems
3
2
1
MAN/SPONT
Pleth
40
0
CO
2
4036
0.80.65856
38
9867
5.28
4.0030
Start volumeter:Confirm !
O
2
SpO
2
N2O
Hal.
Fi
Fet
etCO
2
MV
freq
alarmlimits
CO
2
-al.
on/offalarm
infolist
curves
config.
Freshgas O
2
+ N2O
fresh gas
O
2
%
L/min
VT
0.35
0
Volumeter
--s
--
0510
0.5
1
_
CO
2
Ve
internalexternal
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
51
Operation
Using non-rebreathing systems
Display (example):The airway pressure (PAW), minute volume (MV) and
frequency are not measured.
●Set the fresh gas flow. The fresh gas supply must be
at least twice the minute volume in order to preventrebreathing.
Certain alarms are automatically switched off in order toavoid artefacts.The inactive alarm limits are highlighted by the greybackground in the following table.
FRESHGAS EXTERNAL
CO
2
40
0
30 28
0.80.8 58 57
38
9867
--
--
6.0030
Pleth
O
2
SpO
2
N2O
Hal.
Fi
Fet
etCO
2
MV
freq
CO
2 al.
on/off
alarmlimit
alarminfo
list
curves
config.
Freshgas O
2
+ N
2
O
O
2
%
L/min
fresh gas
_
CO
2
Ve
internalexternal
Alarm limitsIn external fresh
gas mode
Default setting on delivery
SpO2 ON
ON
– –
92
Pulse
H
ONON
120
50
etCO2ON
ON
50 (mmHg)
– –
FiCO2 ON Fixed upper alarm
limit 5 mmHg
MV OFF
OFF
– –– –
FiO2 OFF
ON
– –20
Fi Hal. ON
OFF
1.5– –
Fi Iso. ON
OFF
2.3– –
Fi Enf. ON
OFF
3.4– –
Fi Des.ON
OFF
12.0– –
Fi Sev. ON
AUS
3.4– –
PAWON
OFF
40
– –Apnoea pressureOFFApnoea flowOFFApnoea CO2ON
_
_
_
>
_
_
_
_
_
_
_
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Switch off the external fresh gas output:
●Press the softkey »Fresh gas internal/external« and
confirm by pressing the rotary knob. Julian switches over to MAN/SPONT ventilation modeor
●Press the »IPPV« or »PCV« key and confirm with
the rotary knob.Julian then switches over directly to controlled
ventilation via the original rebreathing system.
When changing from the external non-rebreathing systemto the rebreathing system of the Julian:
●Reconnect the sample line to the Y-piece.
In the event of a power failure
Julian automatically switches over to the integrateduninterruptible power supply (UPS).If the battery is charged, operation will be maintained foran average of 30 minutes.
The message
POWER FAIL!
appears on the screentogether with the remaining available battery capacity in
percent.Example:
JJ
JJ
95 %
If the battery is almost completely discharged, themessage BATTERY LOW !!! appears on the screen.
In the event of a power failure if the battery is discharged,Julian switches over automatically to MAN/SPONTventilation mode. The fresh gas mixer is deactivated.
Version without rotary knob for O2 emergencymetering:
Julian automatically delivers an O2 flow of 4 L/min. This O2 flow bypasses the Vapor unit, and so the patientno longer receives any anaesthetic agent.
Version with rotary knob for O2 emergency metering:
●Set rotary knob »Safety-O2« for O2 emergency
metering to the desired O2 flow. Range 0 to 12 L/min. This O2 flow flows through the Vapor unit.
●Check Vapor unit setting.
●Ventilate patient manually!
OperationUsing non-rebreathing systemsIn the event of a power failure
52
Julian
D
Pleth
IPPV
POWER FAIL !
alarmlimits
95 %
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
J
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:T
I
%
25
Pmax
mbar
Freshgas O2+N2O
1.00
O
2
%
L/min
50
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
In the event of a gas failure
Julian activates an alarm if the O2 or AIR or N2O supplyfails.
●Open the valve on the corresponding backup gas
cylinder on the rear of the workstation.
●Reconnect the central gas supply.
If there is no backup supply of gas available, Julianreplaces the set gas mixture with AIR or O2:
Julian delivers 100 % AIR if the O2 supply fails.Julian delivers 100 % O2 if the N2O supply fails.Julian delivers 100 % O2 if the AIR supply fails.
The amount of fresh gas flow (L/min) remains constant.Fresh gas metering is still possible even in the event of
failure of one of the gases. For example, if the N2Osupply fails, AIR can be selected as the carrier gas.
In the event of a Major Hardware Fault
such as the failure of the fresh gas mixer, the message:
SYSTEM FAULT!!!
will be displayed. In the event of total system failure including the screen,
an alarm will sound for about 30 seconds. In such cases Julian switches to MAN/SPONT ventilationmode.
Version without rotary knob for O2 emergencymetering:
Julian automatically delivers an O2 flow of 4 L/min*. This O2 flow bypasses the Vapor unit, and so the patientno longer receives any anaesthetic agent.
Version with rotary knob for O2 emergency metering:
●Set rotary knob for O2 emergency metering to the
desired O2 flow. Range 0 to 12 L/min. This O2 flow flows through the Vapor unit.
●Check the Vapor unit setting.
●Ventilate the patient manually!
* Can be adjusted to up to 8 L/min O2 by DrägerService.
N2OO2
53
Operation
In the event of a gas failure
In the Event of a Major Hardware Fault
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
End of Operation
To switch Julian to Standby:
●Press the standby key E and confirm by pressing
the rotary knob.The workstation is now on standby. The fresh gas flow is switched off.
Manual ventilation is not possible!
The following default values are activated in standbymode:Fresh gas default valuesVentilation parametersAlarm limits
To switch Julian off:
●Press the power switch » « down completely.
The apparatus is switched off after 10 seconds.
●Remove the gas supply probes from the wall sockets.
●Leave Julian plugged into the power supply in order to
charge the uninterruptible power supply (UPS).
To switch off the anaesthetic gas scavenging systemAGS:
●Pull back the black sleeve on the connector of the
suction hose to release the connector.
Julian is equipped with a switch-off time-lag. Themessage "System will be closed down and switched off"appears for 10 seconds after pressing the main powerswitch. Julian can immediately be restarted by pressingthe main power switch again during this time.The switch-off time-lag is not active during the self-test.
54
End of Operation
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Contents
Selecting the standard screen................................................................56
Selecting the data screen.......................................................................57
Selecting the trend screen.....................................................................58
Tips to prevent artefacts...........................................................................78
Applying the Durasensor DS-100 A..........................................................79
Airway temperature measurement (optional)..........................................80
55
Monitoring
Contents
Monitoring
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Monitoring
Selecting the standard screen
The standard screen is automatically displayed on thescreen whenever a ventilation mode is selected(MAN/SPONT or IPPV or PCV). This page can alwaysbe selected during operation:
1Press the Q key
or
2Press the S key several times.
Display (example):The most important parameters are grouped together on
the right-hand side of the screen.Three curves are displayed on the left-hand side
(to select other curves, see page 66).
Measured values shown in grey
Calibration is performed automatically by Julian. Themeasured values are shown in grey until sufficient dataare available and until the sensors have reached theiroperating temperature.In this way, Julian indicates that the accuracy of thesedata is still reduced in comparison with the "Technicaldata". For maximum accuracy the sensors can be calibratedmanually – see page 75.
When measuring CO2, anaesthetic agent and N2O, theirnumerical values are shown in grey during the warm-upphase whenever the workstation is switched on.
The display CAL appears instead of the numerical valueduring automatic calibration of the O2 sensor (every 24 hours).
The flow sensor is automatically calibrated as soon as aCO2 signal is detected by the gas measuring system.Julian takes account of the relevant gas mixture whencalculating the volumetric concentrations.
MonitoringSelecting the standard screen
56
Julian
21
Pleth
IPPV
alarmlimits
auto setlimits
alarminfo
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
N2O
Hal.
4036
0.80.65856
Fi
Fet
etCO
2
38
9867
6.010
MV
freq
600
V
T
mL
10
1:2
100
PEEP
mbar
Freq.
1/min
TI:TETIP:TI
%
25
Pmax
mbar
Freshgas O2+N2O
4.00
O
2
%
L/min
30
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Selecting the data screen
●Press the S key several times until the data page
appears.Display (example):All numerical values are displayed on the data screen
with their units of measure. The bar display on the left ofthe screen shows the change in airway pressure PAWover time. The lines cutting across the bar graph markthe upper and lower alarm limits. To set the alarm limitsfor PAW, see page 60.
The system compliance (Sys.-Compl.) and leakage aredisplayed in the right-hand field, together with the time ofthe last test.
57
Monitoring
Selecting the data screen
IPPV
PAW
40
20
0
60010
1:2
100251.0050
O
2
%
N
2
O
%
Hal.%
036
4033
0.80.65856
Fi
Fet
CO
2
mmHg
9867
6.0
0.6010
2720
51565
MVV
T
freq
peakplat.PEEPmeancompliance
mbar
mL/mbar
L/minL
34
AW-temp.
°C
1/min
01. Nov. 98
sys. comp. 6.40 mL/mbarleckage 10 mL/min
01-10
8 : 00
12 : 00
SpO
2
%
1/min
alarmlimits
auto set limits
alarminfo
list
config.
Freshgas O
2
+ N2O
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Selecting the trend screen
Displays the measured values over time since the start ofmeasurement.
Maximum storable time: 8 hours.The following display combinations can be selected:
– CO2/MV–AGas/N2O–O2/compliance*–SpO2/pulse rate
●Press the S key several times until the trend screen
appears.
AMV and compliance trends are scaled in accordancewith the setting in the configuration menu for adults orchildren.
AMV trend scale:0 to 3 L/min for children
0 to 15 L/min for adultsCompliance trend scale:0 to 20 mL/mbar for children
0 to 100 mL/mbar for adults
Display (example):Trend CO2 and MV
Selecting other display combinations
●Press the appropriate softkey:
»AGas N2O«or»O2 compl.«or»SpO2 pulse«
The softkey disappears if a measuring function is notavailable.
____________*Lung properties can be monitored continuously with the compliance
trend display so that ventilation can be adjusted accordingly.Patient compliance is not measured in MAN/SPONT ventilationmode.
58
MonitoringSelecting the trend screen
IPPV
4036
0.80.65856
38
9867
6.010
60010
1:2
100251.0050
PAW
40
20
0
11:0012:0013:00
60
30
0
15
10
5
0
Select zoom area:
Turn !
Zoom in: Press!
CO
2
MV
O
2
SpO
2
N2O
Iso.
Fi
Fet
etCO
2
MV
freq
alarmlimits
AGasN
2
O
CO
2
MV
O
2
compl.SpO
2
pulsefull
trend
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Zoom function
The trend display can be magnified with the zoomfunction after half-an-hour of operation.
To select the range:
●Turn the rotary knob = the dashed frames move.
To enlarge the selected area to the full width of thedisplay:
●Press the rotary knob.
A new dashed frame appears after a correspondingperiod of operation and can also be enlarged.
To return to the overall trend:
●Press the softkey »full trend« and the complete trend
is displayed again.
To delete the trend memoryOnly possible in standby mode.
The trend memory and list are deleted together!
In standby mode:
●Press the softkey »delete trend«.
The system asks you to confirm that the trend is really tobe deleted.
To delete:
●Press the softkey »delete«.
The stored trend for the anaesthetic gas concentration isdeleted automatically when a new anaesthetic gas isselected.
59
Monitoring
Selecting the trend screen
IPPV
4036
0.80.65856
38
9867
6.010
60010
1:2
100251.0050
PAW
40
20
0
11:0012:0013:00
60
30
0
15
10
5
0
Select zoom area:
Turn !
Zoom in: Press!
CO
2
MV
O
2
SpO
2
N2O
Iso.
Fi
Fet
etCO
2
MV
freq
alarmlimits
AGasN
2
O
CO
2
MV
O
2
compl.SpO
2
pulsefull
trend
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
_
CO
2
Ve
deletetrend
leakagetest
config.
Julian
SW-version 2.0
01. Nov. 98
Standby
4.00
L/min
30
Operation with,or
MAN
SPONT
IPPV
PCV
O
2
%
Freshgas O2+N2O
do notdelete
config.
Julian
SW-version 2.0
01. Nov. 98
Standby
4.00
L/min
30
Operation with,or
MAN
SPONT
IPPV
PCV
O
2
%
Freshgas O2+N2O
delete
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Alarms
Displaying and setting alarm limits
Alarms can be displayed and set from all three basicscreens (standard, data and trend screen) duringoperation.
The default alarm limits for the ventilation mode areactivated automatically when changing from standby to aventilation mode.
These default alarm limits can be used unchanged or theycan be set individually for the patient.
The individual settings are deleted on changing fromventilation mode to standby and the default alarm limitsare reactivated.
To call up alarm limits:
●Press the softkey »alarm limits«.
Display (example):The measured values and their alarm limits are shown on
the right-hand side.The curves with the alarm limits represented by dashedlines are shown on the left.
Example etCO2:The upper alarm limit (42 mmHg) and the lower alarmlimit (30 mmHg) are shown alongside the measured value(38 mmHg).
A deactivated alarm limit is indicated by two dashes (– –).
To set an alarm limit:
●Place the cursor bar on the alarm limit by turning the
rotary knob and press to confirm.The alarm limit is highlighted.
●Set the new value by turning the rotary knob and
press to confirm.The new alarm limit is now effective. The cursor barreturns to the zzzzsymbol.
MonitoringAlarms
60
Pleth
IPPV
CO
2
40
0
60010
1:2
100251.0050
986738
6.029
0.8
--
95
3.0
--
20
1.0
0.040
8
120
50
--
Adjust limit value: Turn and confirm !
42
30
PAW
0
20
FiO
2
FiHal.
PAW
SpO
2
etCO
2
MV
alarmlimits
auto set limits
alarminfo
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O2 + N2O
38
42
30
etCO
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Adjustment range of the alarm limits
To quit the alarm limits menu:
●Press the rotary knob
or
●Press the Q key.
61
Monitoring
Alarms
Alarm limitAdjustment rangeSpO2
[%]
51 to 100
50 to 99
Pulse
H
[1/min]
21 to 25020 to 249
etCO2[mmHg]
1 to 750 to 74
MV[L/min]
0.1 to 390 to 38.9
FiO2[Vol.%]
19 to 100
18 to 99
Fi Hal. [Vol.%]
0.1 to 70 to 6.9
Fi Iso.[Vol.%]
0.1 to 70 to 6.9
Fi Enf. [Vol.%]
0.1 to 70 to 6.9
Fi Des.[Vol.%]
0.1 to 21.90 to 21.8
Fi Sev. [Vol.%]
0.1 to 9.90 to 9.8
PAW[mbar]
5 to 980 to 35
_
_
_
_
_
_
_
_
_
_
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Adapting limits – AutoSet
When the ventilation settings have been made, Julian canautomatically adapt the alarm limits for the minute volume(MV) and airway pressure (PAW) to the currentventilation settings in IPPV mode.
●Press the softkey »auto set limits«.
The new alarm limits for the set MV and PAW arecalculated by Julian as follows:
To reset individual alarm limits for MV and PAW:
●See "Displaying and setting alarm limits"
on page 60.
To reuse all default alarm limits:
●See "Configuring in standby mode,
setting default alarm limits" on page 72.
62
MonitoringAlarms
Upper alarm limitLower alarm limit
MV[L/min]
VT x f x 1.4At least 2.0
VT x f x 0.6At least 0.5
PAW[mbar]
(Pmax + 10)PEEP +
Pmax – PEEP
4
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
If an alarm occurs
Alarm messages are assigned to three priority classes by Julian, depending on their urgency, and identified byexclamation marks:
Warning!!!= message with top priorityA warning message requires immediate action
Caution!!= message with medium priorityA caution message requires prompt action
Advisory!= message with lowest priorityAn advisory message requires to be noted and actiontaken if necessary
If several alarms occur simultaneously, Julian displays thealarm with the highest priority. The others are notdisplayed.
All warning, caution and advisory messages that havebeen emitted can be displayed by pressing the »alarm Info« softkey.
Each time an alarm message is displayed, the alarm lampeither flashes or lights up continuously and an audibletone sequence is emitted to indicate the alarm priorityclass:
Warning:1Red lamp flashes, accompanied by a continuous tone.Caution:1Yellow lamp flashes, accompanied by an intermittent
tone every 30 seconds.
Advisory:1Yellow lamp lights up continuously, accompanied by a
single tone.
The alarm tone can be suppressed for two minutes:2Press the G key and its yellow LED lights up.To reactivate the alarm tone:2Press the G key and its yellow LED goes out.A list of alarm messages can be found on page 120,
"Message – Cause – Remedy".
63
Monitoring
If an alarm occurs
Julian
2
1
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Alarm info
This function is used to list all active alarms / warnings inorder of priority.
●Press and hold the softkey »alarm info«.
Display (example):Active alarms:
FIO2<<
<<
!!!
Active warning:ETCO2<<<<!!
The display only remains visible as long as the softkey ispressed.
CO2 alarms on /off
The alarm limits for etCO2, insp. CO2 and CO2 apnoeamonitoring can be switched off with the softkey »CO2-al. on/off« in MAN/SPONT mode.
In this case, the symbol CO2gappears alongside themeasured end-expiratory CO2 concentration.
The CO2 alarm limits are reactivated by pressing thesoftkey again. The deactivated CO2 alarm limits arereactivated automatically when changing to a differentventilation mode.
The CO2 alarm limits can also be activated anddeactivated under "Alarm limits" in the configurationscreen in all ventilation modes, see page 72.
National and European standards require a minimummonitoring with alarm function. These standards may notbe met if the alarm function of the etCO2 monitoringparameter is deactivated.
MonitoringCO2 alarms on/offAlarm info
64
Pleth
IPPV
CO
2
40
0
PAW
0
20
60010
1:2
100251.0050
ETCO2!!
FIO2!!!
FIO2!!!
alarmlimits
auto setlimits
alarminfo
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
MAN/SPONT
CO
2
40
0
4036
0.80.65856
38
9867
5.28
4.0030
VT
0.35
0
Volumeter
--s
Start volumeter:
Confirm !
--
05
10
0.5
1
Pleth
SpO
2
etCO
2
freq
MV
O
2
Hal.
N
2
O
FiFet
alarmlimits
CO
2-al.
on /off
alarmInfo
list
curves
config.
O
2
%
L/min
Freshgas O
2
+ N
2
O
fresh gas
_
CO
2
Ve
internalexternal
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
List display
Record of measured values, alarms and ventilationmodes to facilitate compilation of the anaesthetic record.The control criteria for entries can be configured, see"Record" on page 71.
●Press the softkey »list«.
Display (example):
"Open" the previous page:
●Select »previous page«by turning the rotary knob
and press to confirm.
"Open" the next page:
●Select »next page«by turning the rotary knob and
press to confirm.
Return to the standard page:
●Place cursor on the zz
zz
symbol by turning the rotary
knob and press to confirm.
or
●Press the Q key.
Delete listThe list and trend memory are deleted together!
Only possible in standby!
In standby:
●Press the softkey »delete trend«.
Delete trends and list:
●Press the softkey »delete«.
65
MonitoringList display
IPPV
next pageprevious page
alarmtimeHR/PulsNiBP/mmHg
Sys/M/Dia
SpO2etCO2MVO2
Fi/Fet
AGASFi/Fet
PAWPeak/Peep
999899
403840
6.0
5.9
6.1
30/2625/2030/26
1.6/0.6
0.8/0.7
0.8/0.7
33/1030/1030/10
FI HALFI O2
12:1312:1512:20
656566
!!!!!!
60010
1:2
100251.0050
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
PAW
40
20
0
deletetrend
leakagetest
config.
Julian
SW-version 2.0
01. Nov. 98
Standby
4.00
L/min
30
Operation with,or
MAN
SPONT
IPPV
PCV
O
2
%
Freshgas O2+N2O
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Selecting curves
Only possible from the standard screen!Three curves can be selected and positioned for the
momentary ventilation mode in this menu.The following curves can be selected:AGasCurve of anaesthetic agent concentration at
the Y-piece
CO2Curve of CO2 concentration at the Y-pieceFlowExpiration flowO2Curve of O2 concentration at the Y-piecePAWAirway pressurePlethThe plethysmogram derived from SpO2
measurement (optional)
VolumeterA display mode showing the minute volume
and tidal volume VT as bar graphs
PressuresDigital indication of the measured value for
PPEAK, PPLAT, PMEAN and PEEP
blankBlank field
●Press the softkey »curves«.
Display (example):The names of the curves in each curve field are displayed
together with the choice of possible parameters. Thecurrent curves are highlighted by a grey background.
In the example, the PAW curve is to be replaced by thevolumeter display.
●Turn the rotary knob to select the required
parameter (»volumeter« ) and press to confirm.
The cursor jumps to the zzzzsymbol of the positioningmenu.
●Select the desired position with the rotary knob and
press to confirm.Display (example):The curve is replaced and the cursor returns to the zz
zz
symbol in the selection menu.To return to the standard page:
●Press the rotary knob.
If a curve field is to remain blank:
●Select »blank« with the rotary knob and
press to confirm.
66
MonitoringSelecting curves
Pleth
IPPV
4036
0.80.65856
38
9867
6.010
60010
1:2
100251.0050
Select curve:Turn and confirm !
SpO
2
etCO
2
freq
MV
O
2
Hal.
N
2
O
FiFet
alarmlimits
auto set limits
alarminfo
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
CO
2
PAW
Flow
Volumeter
Pleth
O
2
AGas
blank
CO
2
PAW
Pleth
Pressures
_
CO
2
Ve
Pleth
IPPV
4036
0.80.65856
38
9867
6.010
60010
1:2
100251.0050
Select curve position:Turn and confirm !
SpO
2
etCO
2
freq
MV
O
2
Hal.
N
2
O
FiFet
alarmlimits
auto set limits
alarminfo
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N2O
CO
2
Volumeter
Pleth
CO
2
PAW
Flow
Volumeter
Pleth
O
2
AGas
blank
Pressures
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Using the volumeter function
To observe and assess ventilation in manual andpressure controlled ventilation modes and duringspontaneous breathing.
➀Upper bar:
Current tidal volume VT, preceded by the numerical value.
➁Lower bar:
Volumeter (minute volume measurement), preceded by the numerical value.
The scale of the bar display depends on the defaultsetting, see "Configuring in operation, selecting thescale" on page 69 and "Configuring in standby mode, selecting the scale"on page 83.
Starting the volumeter
●Press the rotary knob.
The values are deleted and volumeter restarted if therotary knob is pressed again before 60 seconds have elapsed.
The elapsed time in seconds is displayed above the barwith the summed volume on the left. The individualbreaths are separated by segments in the bar. Thevolumeter stops automatically with an acoustic signalafter 60 seconds.The measured values are displayed for 4 minutes andthen erased.
67
Monitoring
Using the volumeter function
Setting Tidal volume VT
in L
Minute
volume MV
in L/min
Scale for children
0.22
Scale for adults
1.010
MAN/SPONT
Pleth
40
0
CO
2
4036
0.80.65856
38
9867
5.28
4.0030
VT
0.35
0
Volumeter started. To re-start:
Confirm !
0.5
1
O
2
SpO
2
N2O
Hal.
Fi
Fet
etCO
2
MV
freq
alarmlimits
CO
2
-al.
on / offalarm
infolist
curves
config.
Freshgas O
2
+ N2O
O
2
%
L/min
Volumeter46 s
4.1
05
10
fresh gas
_
CO
2
Ve
internalexternal
➀
➁
VT
0.35
0
0.5
1
Volumeter46 s
4.1
05
10
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Configuring in operation
The configuration menus are used to select or modifyindividual monitoring functions for the measurements inprogress.The settings entered here are only valid while measu-rement is in progress and are deleted on switching tostandby mode.
In the standard screen (example) or data screen:
●Press the softkey »config.«.
Display (example):The softkeys for fresh gas and ventilation parameters
remain active.The values highlighted in grey represent the current
The anesthetic agent is automati-cally selected by gas identification
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Setting the monitoring functions
●Select the »settings« column by turning the rotary
knob and press to confirm.
Display (example):
Setting the pulse tone0 = OFF
9 = Maximum volume
●Select the line »pulse tone« by turning the rotary knob
and press to confirm. The cursor automatically jumpsto the last value selected (in the example: setting 5).
●Set the value with the rotary knob and press to
confirm.
Setting the alarm sound1 = Minimum volume
9 = Maximum volume
●Select the line »alarm sound« by turning the rotary
knob and press to confirm. The cursor automaticallyjumps to the last value selected (in the example:setting 4).
●Set the value with the rotary knob and press to
confirm.
Set a sufficiently loud alarm sound!
The audible »NO O2« and »NO FRESH GAS« alarms arealways sounded at maximum volume.
Julian complies with national regulations in certaincountries that stipulate a minimum volume of 45 dB (A).Settings 1 to 4 are programmed at 45 dB (A) for thesecountries.
Selecting the scale
To adapt the scale for the volumeter function and trendcurve.
●Select the line »scale« with the rotary knob and press
to confirm.
●Select »adult« or »child« with the rotary knob and
The anesthetic agent is automati-cally selected by gas identification
settings
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Setting the measuring parameters
●Select the line »parameters« with the rotary knob and
press to confirm.The »parameters« menu appears in the left-hand field.
Display (example):
Switch SpO2 measurement (optional) on/off:
●Select the line »SpO2 measurement« with the rotary
knob and press to confirm.
●Select »on«or»off« with the rotary knob and press to
confirm.
Select the sample rate for gas measurement:
●Select the line »sample rate« with the rotary knob and
press to confirm.
●Select »60«or»200 mL/min« with the rotary knob
and press to confirm.
For paediatric use:
●Select a sample rate of 200 mL/min in order to obtain
more accurate values on account of the high frequen-cies and time constant of the measuring system.
●Recirculate the measured gas, see page 109.
If the measured gas is not recirculated:
●Increase the minute volume accordingly in order to
maintain the volume balance.
For adult use:
●Select a sample rate of 60 or 200 mL/min.
Recommendation: select 200 mL/min and recirculatethe measured gas, see page 109.
Set CO2 unit:
●Select the line »CO2 unit«with the rotary knob and
press to confirm. The last value selected is highlightedby a grey background.
●Select the required unit of measure by turning the
rotary knob and press to confirm.
IPPV
0 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
adult child
60010
1:2
100251.0050
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
PAW
40
20
0
settings
puls tonealarm soundScale
parameter
recordactivate monitor defaults
SpO2-measure.sample rateCO
2
unit
on off
60 200 mL/min
mmHg kPa vol%
Menu for configuration of measurement parameters.
anesth. agent
calibration
alarms
70
MonitoringConfiguring in operationSetting the monitoring functions
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Record
This function is used to determine the type of event thattriggers an entry in the record list or a printout on theconnected logging printer.
●Select the line »record« with the rotary knob and
press to confirm. The »record« menu appears on the left-hand side.
Display (example):
Timeinterval-started (min)
A list entry is activated after a fixed time interval inminutes.
NiBP-startedAn entry is made after every NiBP measurement with newmeasured values. This entry can be read by an externalDräger monitor via the MEDIBUS interface.
warning-startedAn entry is made whenever an alarm is triggered.
caution-startedAn entry is made whenever a caution message istriggered.
●Turn the rotary knob to select the event triggering the
entry and press to confirm.
Activating monitor defaults
This function is used to activate the default monitoringsettings configured in standby mode and active wheneverthe apparatus is switched on.
●Select the line »activate monitor defaults« with the
rotary knob and press to confirm.
IPPV
0 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
adult child
60010
1:2
100251.0050
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
TI:TETIP:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
PAW
40
20
0
settings
puls tonealarm soundScaleparameter
record
activate monitor defaults
Menu for configuration of measurement parameters.
list entry: time intervall (min) NiBP started warning started caution started
printer entry: time intervall (min) NiBP started warning started caution started
12510off
onoffonoff
on
off
12510off
on
off
onoff
on
off
anesth. agent
calibration
alarms
71
Monitoring
Configuring in operation
Setting the monitoring functions
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Selecting alarms
●Turn the rotary knob to select the column »alarms«
and press to confirm.
Default
This function is used to activate the default alarm limitsset in the standby configuration which are effectivewhenever the apparatus is switched to standby.
Setting default alarm limits:
●Select the »default settings« menu by turning the
rotary knob and press to confirm.See page 86 for the further procedure.
CO2 alarms on/off
To activate / deactivate all CO2 alarms, including the
apnoea alarm.
●Turn the rotary knob to select »CO2 on« or »CO2 off«
and press to confirm.
SpO2 alarms on/off
To activate/deactivate all SpO2 alarms.
●Turn the rotary knob to select »SpO2 on« or »SpO2
off« and press to confirm.
Deactivated alarms are indicated by the symbol _ in themeasurement field.
National and European standards require minimummonitoring with alarm function. These standards may notbe met when the alarm function of the monitoringparameter etCO2 or SpO2 is deactivated.
72
MonitoringConfiguring in operationSelecting alarms
The anesthetic agent is automati-cally selected by gas identification
settings
alarms
CO
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Alarm mode for the heart-lung machine HLM
HLM alarm mode can be used independently of theactive ventilation mode for patient monitoring when usingthe heart-lung machine.
In HLM alarm mode,–all apnoea alarms are deactivated.– the gas concentrations are measured continuously
regardless of the respiratory phase.
–the SpO2 alarms are deactivated.
They are reactivated automatically when pulsationsare detected by the apparatus again after switchingoff the HLM alarm mode.
Selecting HLM alarm mode
●Press the softkey »config«.
●Turn the rotary knob to select the column »alarms«
and press to confirm.
●Turn the rotary knob to select the line »HLM on« and
press to confirm.
●Display (example):
The lung is kept open with constant airway pressure.
HLM alarm mode can be switched off at any time:
●Press the softkey »HLM off«.
In standby mode, the HLM alarm mode is automaticallydeactivated.
73
Monitoring
Configuring in operation
Selecting alarms
MAN/SPONT HLM
CO
2
40
0
3030
0.80.8
0.00.0
5
--
--
-.- --
6.0030
SpO
2
etCO
2
freq
MV
O
2
Hal.
N
2
O
FiFet
alarmlimit
HLMoff
alarminfo
list
curves
config.
O
2
%
L/min
Freshgas O
2
+ AIR
PAW
0
20
fresh gas
_
internalexternal
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Selecting the anaesthetic agent
Julian automatically recognizes the anaesthetic gas usedand switches measurement and monitoring of theanaesthetic gas concentration to the identified gas.
If a mixture of two anaesthetic agents is detected,
the system generates a caution message:
AGAS MIXTURE !!
The anaesthetic agent with the higher concentration ismonitored and displayed. As long as a mixture of twoanaesthetic agents is present (e.g. after changing theanaesthetic agent), the values of the anaesthetic agentconcentration are displayed in grey = reduced accuracy.
Displays Fi and Fet give information about theanaesthetic agent mixture situation:
Fi displayFet displaySituationgreylightAGAS mixture
inspiratory only
greygreyAGAS mixture
inspiratory and expiratory
lightgreyAGAS mixture
expiratory only
lightlightNo more AGAS mixture
The anaesthetic agent can also be selected manually:
●Turn the rotary knob to select the column »anasth.
agent« and press to confirm.
●Turn the rotary knob to select the line »man« and
press to confirm.
●Turn the rotary knob to select the column »gas« and
press to confirm.
●Turn the rotary knob until the required gas appears in
the display window.Press the rotary knob to confirm the selection.
The user is responsible for manual selection of theanaesthetic agent and for correctly filling of theanaesthetic vaporizer!
74
MonitoringConfiguring in operation Selecting the anaesthetic agent
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Manual calibration
Julian calibrates its sensors automatically duringoperation. The message CAL appears instead of thecurrent measured value during a calibration process.Calibration of the O2 sensor and zero calibration of theother gas sensors can be repeated manually, for instanceif automatic calibration was unsuccessful.
●Turn the rotary knob to select the »calibration«
column and press to confirm.
●Select and confirm the appropriate line with the rotary
knob, for example:»O2-Sensor21 vol.%«
The calibration process is started when the selection isconfirmed. The clock symbol UUUUappears after therelevant line, i.e. calibration is in progress.
When calibration is complete, a tick (✓) appears insteadof the clock symbol. A question mark (?) indicates thatcalibration must be repeated because the previouscalibration was defective.
Other settings can already be made during the calibrationprocess.
Patient1 to 20 kg10 to 50 kg>40 kg>30 kg>50 kgweight
DurationShort and long-termShort and long-termShort-termShort and long-termShort and long-termof usemonitoringmonitoringmonitoringzeitüberwachungmonitoring
PatientLimitedLimitedImmobileLimitedImmobilemobilitymobilitymobilitypatients onlymobilitypatients only
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
●Select the appropriate sensor.
On the back of the Julian:1Swing the flap over the socket »SpO2« to the left and
insert the sensor plug.
C-lock-ECG synchronization (optional)If the patient is highly agitated or the arterial flow is very
low, the measurement signals of the SpO2 measurementcan be improved with the C-Lock-ECG synchronization.In this case, Julian receives two separate signals thattransmit the heart activity:
–An optical signal from the SpO2 sensorand–An electrical signal from the separate ECG monitor.
Julian uses the R-wave of the ECG signal to identify thepulse and to synchronize with the SpO2 measurement.
2Channel the ECG signal of the ECG monitor with the
cable andyesck-plug to the back of the Julian (»Sync.« socket). See the "Technical data" on page 131 for therequirements for the electrical signal and plug layout.
In the event of a delayed ECG signal
Synchronization may be impaired if the ECG signal isdelayed by more than 40 milliseconds to the QRScomplex.
If you suspect a fault of this kind:
●Use the Julian without C-Lock-ECG synchronization.
1
77
Monitoring
SpO2 Measurement
2
EKGMonitor
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Monitoring SpO2 Measurement
78
Tips to prevent artefacts
Only use Nellcor sensors in the recommended positions,otherwise incorrect measurements and tissue damagemay result.
Damaged sensors with exposed electrical contacts mustnot be used – danger of electric shock.
Used adhesive strips of the Oxiband-OXI-A/N and OXI-P/I sensor must not be reused, as they may notadhere properly.The strips must not be stretched unduly.Never use two strips together, as this may lead to venouspulsation and failure of the pulse signal.
High intrathoracic pressure, pressure on the thorax andother consecutive impairments of the venous flow canlead to venous pulsation with failure of the pulse signal.
The pulse signal may fail in the presence of shock, lowblood pressure, severe vasoconstriction, major anaemia,hypothermia, arterial occlusion proximal to the sensor andasystolia.
The sensor must be protected from exposure to brightlight (e.g. surgical lamps and direct sunlight), otherwisethe pulse signal may fail or inaccurate results may beobtained.
The sensor should not be positioned on limbs togetherwith an arterial catheter, sphygmomanometer cuff orintravascular venous infusion, otherwise the pulse signalmay fail and measurement become inaccurate.
Measurement accuracy may be reduced in the presenceof significant concentrations of dyshaemoglobins, suchas carboxyhaemoglobin or methaemoglobin.
Intravascular dyes, such as methylene blue, may alsoimpair measurement accuracy.
Electrocautery can impair the measuring accuracy; theleads and sensor should therefore be positioned as faraway as possible from the electrocautery and its neutralelectrode.
Sensor performance may be impaired and lead toinaccurate results if the patient moves violently. Thesensor should be applied to a different site in such casesin order to reduce the risk of artefacts due to movement.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Applying the Durasensor DS-100 A
Reusable sensor for short-term monitoring of relativelyquiet patients weighing over 40 kg.
The sensor is preferably positioned on the index finger,although other fingers can also be used. The little fingershould be used if the patient is particularly large orobese.
●Open the clip slightly and slide the sensor onto the
finger. The tip of the finger must touch the end andthe soft padding should rest on the nail and tip of thefinger. The lead should be on top of the finger.
●Ensure that the finger is not compressed or hurt by
the clip.
●Change the application site after not more than
4 hours in order to avoid a build-up of blood pressure(blocked circulation).
Follow the specific Instructions for Use when using otherNellcor sensors!
79
Monitoring
SpO2 Measurement
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Airway temperature measurement (optional)
Parts required:–Temperature sensor84 05 371
–Y-piece with connection
for temperature sensorM 30 543–T-piece86 00 224–Filter 86 00 225
1Connect Y-piece to respiratory hoses.2Insert T-piece into Y-piece.3Insert filter into T-piece.4Insert temperature sensor into Y-piece.
5Connect temperature sensor plug to »Temp.« socket
at the back of the workstation.
●The airway temperature is displayed in the data
section of the screen:
AW temp. XX °C
The upper alarm limit is fixed. If the airway temperatureexceeds 40°C, the »AW Temp > !!!« alarm will beactivated.
80
Monitoring Airway temperature measurement
4
3
2
1
5
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
82
Configuring in Standby ModeSetting default values
Configuring in Standby Mode
The default settings for the monitoring functions and thefresh gas and ventilation parameters can be configured instandby mode. The default settings for the monitoringfunctions and the choice of anaesthetic gas are validwhenever the apparatus is switched on. The defaultvalues for the alarm limits, fresh gas parameters andventilation parameters are valid after each standby.
●Switch Julian to standby mode.
●Press the softkey »config.«.
Display (example):
The following settings can be made in the menu:–Default settings –Calibration–Anaesth. agentFields highlighted by a grey background contain the
currently valid settings.The symbol zzzzindicates a return to the previous
menu level.
●Select, set and confirm the new settings with the
rotary knob.
Setting default values
Default values are valid whenever the apparatus isswitched on. The default monitoring settings can beactivated during operation with activate monitordefaults, see "Activating monitor defaults" on page 71.
The anestheticagent is automati-cally selected bygas identification
Isoflurane
These default settings are passwordprotected.The default settings are activeon start-up.
To change the default settingsenter the password:
0123456789
Turn and confirm .
Standby / Configuration
default settings
* * *
anesth. agent
calibration
O2-sensor21 vol.%
gas sensormore
✓
✓
selectionautoman
The anestheticagent is automati-cally selected bygas identification
Isoflurane
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
The menu requests the user to enter a 4-digit passwordin order to prevent unauthorized tampering with the basicsettings. This password is notified during system training.The function can be disabled by DrägerService ifrequired.
●Turn the rotary knob to select the digits con-secutively
from the line provided and press to confirm. The pass-word is indicated underneath with asterisks (
* * * *
).The menu for selecting the default settings appearswhen the password has been entered correctly.
Display (example):
Setting the pulse tone (optional)0 = OFF
9 = Maximum volume
●Select the line »pulse tone« by turning the rotary knob
and press to confirm. The cursor automatically jumpsto the last value selected (example 5).
●Set the value with the rotary knob and press to
confirm.
Setting the alarm sound1 = Minimum volume
9 = Maximum volume
●Select the line »alarm sound« by turning the rotary
knob and press to confirm. The cursor automaticallyjumps to the last value selected (example 4).
●Set the value with the rotary knob and press to
confirm.
Set a sufficiently loud alarm sound!
The audible »NO O2« and »NO FRESH GAS« alarms arealways sounded at maximum volume.
Julian complies with national regulations in certaincountries that stipulate a minimum volume of 45 dB (A).Settings 1 to 4 are programmed at 45 dB (A) for thesecountries.
Selecting the scale
To adapt the scale for the volumeter function and trendcurve.
●Select the line »scale« with the rotary knob and press
to confirm.
●Select »adult« or »child« with the rotary knob and
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
84
Configuring in Standby ModeSetting default values
Setting the measuring parameters
●Select the line »parameters« with the rotary knob and
press to confirm.The »parameters« menu appears in the left-hand field.
Display (example):
Switch SpO2 measurement (optional) on/off:
●Select the line »SpO2 measurement« with the rotary
knob and press to confirm.
●Select »on« or »off« with the rotary knob and press
to confirm.
Select the sample rate for gas measurement:
●Select the line »sample rate« with the rotary knob
and press to confirm.
●Select »60« or »200 mL/min« with the rotary knob
and press to confirm.
For paediatric use:
●Select a sample rate of 200 mL/min in order to obtain
more accurate values on account of the high frequen-cies and time constant of the measuring system.
●Recirculate the measured gas, see page 109.
If the measured gas is not recirculated:
●Increase the minute volume accordingly in order to
maintain the volume balance.
For adult use:
●Select a sample rate of 60 or 200 mL/min.
Recommendation: select 200 mL/min and recirculatethe measured gas, see page 109.
Set CO2 unit:
●Select the line »CO2 unit«with the rotary knob and
press to confirm. The last value selected is highlightedby a grey background.
●Select the required unit of measure by turning the
rotary knob and press to confirm.
Standby / Configurationanesth. agent
calibration
0 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
adult child
SpO2-measure.sample rateCO
2
unit
on off
60 200 mL/min
mmHg kPa vol%
default settings
Menu for configuration of measurement parameters.
pulse tonealarm soundscale
parameters
recordinterfacesalarm limitscurvesbasic configurationsIPPV default settingsPCV default settingsfresh gas delivery
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
85
Configuring in Standby Mode
Setting default values
Record
This function is used to determine the type of event thattriggers an entry in the record list or a printout on theconnected record printer.
●Select the line »record« with the rotary knob and
press to confirm. The »record« menu appears on the left-hand side.
Display (example):
Timeinterval-started (min)
A list entry is activated after a fixed time interval inminutes.
NiBP-startedAn entry is made after every NiBP measurement with newmeasured values. This entry can be read by an externalDräger monitor via the MEDIBUS interface.
Warning-startedAn entry is made whenever an alarm is triggered.
Caution-startedAn entry is made whenever a caution message istriggered.
●Turn the rotary knob to select the event triggering the
entry and press to confirm.
Configuring the interfaces
Julian has three serial interfaces:COM1, COM2, COM3.COM2 and COM3 are configured as MEDIBUS*interfaces and COM1 as printer interface. The interfacescan be adapted to the connected equipment.
●Turn the rotary knob to select the line »interfaces«
and press to confirm.
The »interfaces« menu appears in the left-hand field.Display (example):COM2 is configured in the upper half of the menu and
COM1 and COM3 in the bottom half.
_____________* MEDIBUS: Dräger communications protocol for medical equipment.
Standby / Configurationanesth. agent
calibration
0 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
adult child
default settings
list entry: time intervall (min) NiBP started warning started caution started
printer entry: time intervall (min) NiBP started warning started caution started
12510off
onoffonoff
on
off
12510off
on
off
onoff
on
off
Menu for configuration of list and printer entries.
pulse tonealarm soundscaleparameters
record
interfacesalarm limitscurvesbasic configurationsIPPV default settingsPCV default settingsfresh gas delivery
Standby / Configurationanesth. agent
calibration
0 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
adult child
default settings
Menu for configuration of Medibus and printer ports.
pulse tonealarm soundscaleparametersrecord
interfaces
alarm limitscurvesbasic configurationsIPPV default settingsPCV default settingsfresh gas delivery
COM 2 (Medibus):baud rateparitydata bitsstop bits
COM 1 (printer), COM 3 (Medibus):config. of: baude rateparitydata bitsstop bits
1.2
9.6
even81
COM 1COM 3
1.22.44.89.619.2
even none
odd
78
12
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Configuring in Standby ModeSetting default values
86
Interface
Choice between configuration of COM2 or COM3.
Baud rate
Transmission speed (variable, see Instructions for Use ofthe equipment to be connected).
Parity
This display is invariable for MEDIBUS and is shown onlyfor information.
Data bits
This display is invariable for MEDIBUS and is shown onlyfor information.
Stop bits
This display is invariable for MEDIBUS and is shown onlyfor information.
Setting default alarm limits
●Turn the rotary knob to select the line »alarm limits«
and press to confirm.The »alarm limits«menu appears in the left-hand field.
Display (example):
●Select and confirm the required alarm limit with the
rotary knob.The default alarm limit is highlighted against a whitebackground.
●Set and confirm the value with the rotary knob. The
value is then displayed in normal mode. The cursor bar jumps to the next default alarm limit.
●Set the next default alarm limit as described above.
Setting default alarm limits for anaesthetic agents
●Turn the rotary knob to select the line
»anaesth. agent« and press to confirm.The corresponding menu is displayed.
– – :The default value set on delivery is outside the
monitoring range and the corresponding alarm limitis therefore switched off.
The newly set default alarm limits are valid wheneverthe apparatus is switched on or started up fromstandby mode.
Certain alarm limits are switched off automatically inMAN/SPONT ventilation mode and when using theexternal fresh gas outlet: see tables, alarm limits for MAN/SPONT, page 39 and alarm limits for non-rebreathing systems, page 51.
Alarm limitAdjustment rangeDefault value
set on delivery
SpO
2
[%]
51 to 100
50 to 99
– –
92
Pulse
H
[1/min]
31 to 30030 to 299
120
50
etCO2[mmHg]
1 to 750 to 74
50 (mmHg)
– –
MV [L/min]
0.1 to 390 to 38.9
12
3.0
FiO2[Vol.%]
19 to 100
18 to 99
– –
20
Fi Hal. [Vol.%]
0.1 to 70 to 6.9
1.5– –
Fi Iso. [Vol.%]
0.1 to 70 to 6.9
2.3– –
Fi Enf. [Vol.%]
0.1 to 70 to 6.9
3.4– –
Fi Des. [Vol.%]
0.1 to 21.90 to 21.8
12.0– –
Fi Sev. [Vol.%]
0.1 to 9.90 to 9.8
3.4– –
PAW[mbar]
5 to 980 to 35
40
8
_
_
_
_
_
_
_
_
_
_
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Configuring in Standby ModeSetting default values
88
Configuring curves
Three standard curves can be configured for each of theventilation modes: MAN/SPONT, IPPV, PCV and Freshgas external.
●Turn the rotary knob to select the line »curves« and
press to confirm.The selection menu appears in the left-hand field.
Display (example):
●Select the required ventilation mode with the rotary
knob:
The »Standard IPPV screen« has been selected in theexample:
The currently selected parameters appear in the left-handfield, followed by the selection menu with the curveswhich can be selected.
●Select the required parameter in the selection menu
with the aid of the rotary knob and press to confirm.The parameter appears against a white background.
●Select and confirm the position of the new parameter
with the rotary knob.The corresponding position remains unused if »blank«is selected.
Setting basic configurations
for time, date, language and tone sequence.
●Turn the rotary knob to select the line
»basic configurations« and press to confirm.The selection menu appears in the left-hand field.
IPPV default settingsPCV default settingsfresh gas delivery
time
datelanguagetone sequence
08 : 15day month year27 - 10 - 98
English (GB)
Dräger Euro-stand.
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Setting the date/time:
●Select and confirm the »time« line with the rotary
knob.The cursor bar marks the hours.
●Press the rotary knob to confirm and the value is
highlighted against a white background. It can now beadjusted and confirmed with the rotary knob.The minutes are adjusted and confirmed in the sameway.
●Set and confirm the date as described above.
Setting the language of the display texts:The last language selected is highlighted by a grey
background.The following alternatives can be selected: English (GB)Swedish (SW)
Modify default settings or confirm ! Carrier gas selection via hardkey.
2.00100
O
2
%
L/min
Freshgas O
2
anesth. agent
calibration
O2-sensor21 vol.%
gas sensormore
✓
✓
selectionauto man
The anestheticagent is automati-cally selected bygas identification
Isoflurane
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
Configuring in Standby ModeManual calibration
92
Manual calibration
Julian calibrates its sensors automatically during opera-tion. The message CAL appears instead of the currentmeasured value during a calibration process. Calibration of the O2 sensor and zero calibration of theother gas sensors can be repeated manually, for instanceif automatic calibration was unsuccessful.
In standby / configuration mode:
●Turn the rotary knob to select the »calibration«
column and press to confirm.
●Select and confirm the appropriate line with the rotary
knob, for example:»O2-sensor21 vol.%«
The calibration process is started when the selection isconfirmed. The clock symbol Uappears after the relevantline, i.e. calibration is in progress.
When calibration is complete, a tick (✓) appears insteadof the clock symbol. A question mark (?) indicates thatcalibration must be repeated because the previouscalibration was defective.
To improve the measuring accuracy of high O2 concen-trations, an additional 100 vol% O2 calibration and alinearity test of the O2 sensor can be performed underthe menu option »more«.
Other settings can already be made during the calibrationprocess.
100 Vol.% O2 calibration
●Use a separate O2 source, e.g. O2 flowmeter.
O2 from the Julian breathing system is not suitable, asit still contains traces of anaesthetic gas.
●Unscrew the sample line from the Y-piece and place it
in the continuous flow of the O2 source.
●Turn the rotary knob to select the line
»O2-sensor100 vol.%«
and press to confirm.
●Allow the O2 flow to continue flowing into the sample
line.
A tick (✓) appears after the field when calibration iscomplete.
The anestheticagent is automati-cally selected bygas identification
Isoflurane
✓
✓
Standby / Configuration
O2-sensor21 Vol.%
gas sensor
more
O2-sensor100 Vol.%
linearitytest O
2
calibration
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
O2 linearity test
●First perform the "100 vol.% O2 calibration".
Then:
●Turn the rotary knob to select the line
»linearity test O2«
and press to confirm.
After approx. 20 seconds:
●Display (example):
O221 vol.%
A display in the range between 18 and 24 vol.% iscorrect.If the display is outside this range, the O2 sensor isdefective or spent (replace O2 sensor, see page 117).
Selecting anaesthetic agent
Julian automatically recognizes the anaesthetic agentused and switches measurement and monitoring of theanaesthetic agent concentration to the identified gas.
If »auto« is selected in the menu »anesth. agent«, theanaesthetic agent is selected automatically.
The anaesthetic agent can also be selected manually:
●Turn the rotary knob to select the column
»anaesth. agent« and press to confirm.
Under the »selection« line:
●Turn the rotary knob to select »man« and press to
confirm.
●Turn the rotary knob to select the line »gas« and press
to confirm.
●Turn the rotary knob until the required gas appears in
the display window.Press the rotary knob to confirm the selection.
●Check that the manually selected anaesthetic agent is
compatible with the Vapor unit.
The user is responsible for manual selection of theanaesthetic agent and for correctly filling theanaesthetic vaporizer!
Connecting the anaesthetic gas scavenging system AGS.......................111
95
Care
Contents
Care
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Care
Dismantling components
Dismantling sample line
●Unscrew the sample line from the Y-piece and water
trap on the back of the unit.The sample line is not reusable and can be disposedof with ordinary domestic waste.
Removing water trap container
●Pull the container of the water trap down and off, and
empty it.
Note the hygiene regulations of the hospital!
●Prepare the container of the water trap for disinfection
and cleaning in a washing machine.
When using the "Waterlock" water trap
●See separate Instructions for Use.
Removing breathing hoses
●Remove the breathing hoses and breathing bag.
●Dismantle Y-piece and connector.
●Prepare the breathing hoses, breathing bag and
Y-piece for disinfection and cleaning in a washingmachine.
96
CareDismantling components
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
all-guides.com
●Do not damage the breathing hoses!
Always hold the hoses by the connection sleeve andnot by the spiral ribbing when removing andconnecting the breathing hoses, otherwise the spiralribbing may be torn off the sleeve.Breathing hoses with damaged spiral ribbing caneasily be kinked and interrupt ventilation!Always check the breathing hoses for damage prior touse.Damaged breathing hoses must not be used.
Removing microbial filters
On the sleeve of the microbial filter:
●Press together the surfaces marked »PRESS« and
pull the filter off the connection at the same time.Do not use force, otherwise the sleeve may bedamaged.
●Prepare the microbial filter for conditioning in
accordance with the separate Instructions for Use.
97
Care
Dismantling components
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Dismantling the suction system1Remove the suction hose and vacuum hose.2Grip the silicone sleeve of the bottle cap and pull it
off.
3Remove the secretion collecting bottle and flushing
bottle from the bracket and empty them.
Note the hygiene regulations of the hospital.
4Remove the silicone sleeve from the lid of the
secretion collecting bottle.
5Remove the overflow protection float from the rising
pipe.Store the parts together for disinfection and cleaningin a washing machine.
6Replace the bacterial filter in the back panel of the
Julian – if soaked through or if contamination issuspected.The filter must be replaced at the latest every twoweeks.
●The bacterial filter must be disposed of as infectious
special waste. It can be incinerated with little pollutionat temperatures of more than 800 °C.
2
33
1
1
98
CareDismantling components
4
5
6
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Dismantling the breathing system7Fold up the writing top.8Pull the catch and at the same time pull out the slot-in
module.
9Hold breathing system by the handle and pull
upwards to remove.
Removing the flow sensor1Unscrew the expiration port.2Remove the flow sensor.The flow sensor cannot be disinfected / cleaned in a
washing machine, nor can it be sterilized in high-temperature steam.
●The flow sensor must be disinfected in 70 % ethanol
solution for approx. 1 hour.Leave the sensor to dry in air for at least 30 minutes,otherwise the remaining alcohol can damage thesensor when calibrated.
●The flow sensor can be reused as long as it can be
calibrated successfully.
Removing the absorber and bellows3Turn the absorber counterclockwise and pull it
downwards.
●Empty the soda lime; it can be disposed of with
ordinary domestic waste.
4Remove the insert from the absorber.5Turn the container of the bellows counterclockwise
and pull it downwards.
6Remove the bellows.
●Prepare the absorber, container of the bellows and
the bellows for disinfection and cleaning in a washingmachine.
99
Care
Dismantling components
4
MAX
3
6
5
1
2
7
8
9
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Opening the breathing system
–Weekly1Loosen the five sealing screws a quarter-turn with the
key supplied.
●Remove the cover.
2Remove the two valve discs.
●Lift off the metal valve plate.
●Keep the valve discs in a cassette to protect them
against damage and prepare them for disinfection and cleaning in a washing machine.
●Prepare the housing parts for disinfection and
cleaning in a washing machine.
Dismantle the anaesthetic gas scavenging systemAGS
3Remove the scavenging hose.4Remove the transfer hose.5Remove the anaesthetic gas scavenging system.
●Prepare the individual parts for disinfection and
cleaning in a washing machine.
11
11
1
100
CareDismantling components
2
2
5
4
4
3
"Only for internal use by Draeger. Not applicable as Instructions for Use!"
All manuals and user guides at all-guides.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.