WARNING!
Each servicing and/or testing of the device
requires full understanding of this Technical Documentation. Carefully read the Instructions for Use prior to any use of the
device.
A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death
or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially
hazardous situation which, if not avoided, may result in minor or moderate
injury to the user or patient or in damage to the equipment or other property.
NOTE
A NOTE provides additional information intended to avoid inconvenience
during operation or servicing of the equipment.
Definitions according to German standard DIN 31051:
Inspection= examination of actual condition
Maintenance= measures to maintain specified condition
Repair= measures to restore specified condition
Servicing= inspection, maintenance, and repair
2Notes
This Technical Documentation conforms to the IEC 60601-1 standard.
Read each step in every procedure thoroughly before beginning any test.
Always use the proper tools and specified test equipment. If you deviate from
the instructions and/or recommendations in this Technical Documentation,
the equipment may operate improperly or unsafely, or the equipment could
be damaged.
Dräger recommends that only Dräger provided parts be used for maintenance. Otherwise the correct functioning of the device may be compromised.
The maintenance procedures described in this Technical Documentation may
be performed by trained service personnel only. These maintenance procedures do not replace inspections and servicing by the manufacturer.
This Technical Documentation is for the purpose of information only. Product
descriptions found in this Technical Documentation are in no way a substitute
for reading and studying the Instructions for Use.
NOTE
Unless otherwise stated, reference is made to laws, regulations or standards (as amended) applicable in the Federal Republic of Germany for
equipment used or serviced in Germany. Users or technicians in all other
countries must verify compliance with local laws or applicable international
standards.
•Weight information, collected from external sensors and cables
9
Theory of OperationIsolette® C2000/C2000e
This information is periodically updated and transmitted to the main controller
upon request.
The sensor module P.C. board does not require manual calibration. To provide safe monitoring and control, both the temperature information and oxygen information have redundant circuitry to prevent single-fault failures.
All signals are transmitted through serial data communication.
The sensor module connects to a sensor P.C. board that has the following
parameters required for the system:
•Air temperature
•Oxygen
•Humidity
•Fan operation
Power Supply
The power to the sensor module P.C. board is provided through connector
J4, providing ±12 V AC for the system. U21, U19, and U15 regulate the volt-
ages by providing +5 V, digital +5 V and analog -5 V, respectively. U20 provides a precision +5 V source for analog signal conversions.
Sensor Position Detection
Hall effect sensors sense the magnets in the slide mechanism. The sensors,
U9 and U2, determine the calibration position. The sensors, U16 and U22,
determine the hood’s position during normal operation. The output is normally
high. These devices provide a low output if a significant south pole magnetic
field is applied to their surface.
Remote Light Alarm Indicator
DS1 provides an alarm indication with a high field of view. A positive signal,
RMLITE, at the gate of transistor Q3 illuminates DS1.
Scale Interface
The connector J3 provides the interface and power for the scale module. The
scale module supports serial data communication. The signal SCCLK is used
for scale communication only and is normally high when the scale is disconnected. The signal CDATA is bi-directional and is normally in the low state
when the scale is not connected.
Fan Control/Feedback Circuit
To drive the DC fan on the sensor board, the signal FANON pulses Q1 on its
gate at a 50% duty cycle at approximately 48 Hz to maintain proper speed
and to increase fan life. Every 4 seconds, the microcontroller asserts FANON
for 42 milliseconds, and the fan pulse detection begins.
10
U1A handles the pulse detection and, through the resistors R3 and R7, samples the current spikes from the fan produced across R1. The amplifier operates as a differentiator, providing high gain for the current spikes. D1 and C16
then rectify and filter this signal and then feed it to the Analog/Digital (A/D)
converter as signal FANPUL. This provides a semi-DC level as a function of
the fan rotation.
Temperature Measurement
The temperature data acquisition circuit starts with the analog multiplexers,
U6 and U5, each allowing an 8-to-1 signal switching. The microcontroller
selects the multiplexer channel by the signals MSEL0, MSEL1, and MSEL2.
Each multiplexer output can be inhibited by either signal TM1SEL or
TM2SEL, depending on the multiplexer; only one multiplexer is active at a
time. With each multiplexer output into the amplification under control, this
data acquisition is viewed as a 16-to-1 analog temperature selector. The
microcontroller selects a new temperature channel every 21 milliseconds.
The resistor R14 provides the constant voltage drive required for each thermistor as it is selected from the appropriate multiplexer. This voltage is amplified by a factor of 2.1083, and is sent to the A/D converter as signal TCOM.
To provide automatic calibration of the circuit and eliminate temperature and
aging drifts, R44 and R45 introduce precise calibration values into each multiplexer. These values are read through the multiplexers and are used in software processing to eliminate the gain and offset errors of each
multiplexer/amplifier combination. These values equate to 120.87 °F
(49.37 °C) and 72.72 °F (22.62 °C), which allow for precise circuit calibration.
Resistor R43 is an additional check to the circuit, which provides a resistance
simulating 98.57 °F (36.98 °C).
The sensor module supports the following three air temperature sensors on
the sensor board:
•AIRTE
•AIRTC
•AIRTM
These signals interface through J2-6 to J2-8. The thermistors then route to
temperature multiplexers U5 and U6, which provide analog signal processing
into the A/D converter.
The skin temperature probes contain dual thermistors. The sensor module
supports two probes that plug into connectors J6 and J7. The two thermistors
connect to SKNT1M and SKNT1C or SKNT2M and SKNT2C, with a common
connector at AGND. Both probes have high frequency filtering by inductor
networks LN1 and LN2. In addition, each skin probe has a resistor that is
input to the multiplexers. The microcontroller uses these signals, SKNT1D
and SKNT2D, to determine if the probes are installed.
Humidity measurement originates with the humidity sensor on the sensor
P.C. board, connected to J2-5 and J2-14 (AGND and HS2, respectively). The
sensor is a capacitive-type that changes capacitance as a function of humidity; the net range of capacitances is from approximately 160 pF to 200 pF.
The sensor connects to the amplifier U7A/U8A, which is set up as a multivibrator. The sensor capacitance charges up through R20 and R21 to a
threshold voltage established by R30 and R26.
11
Theory of OperationIsolette® C2000/C2000e
When the capacitor voltage reaches the threshold, the capacitor U7A/U8A
goes high to 2.5 V as controlled by R19 and R28, and turns on Q2. This discharges the humidity sensor through R20 until it reaches the lower threshold
established by R24, R26, and R30. The capacitor voltage goes from approximately 0.2 V to 0.7 V. At this point, the comparators output goes low, releasing the drive to Q2 and allowing the humidity to start charging again. This
produces a frequency output as a function of capacitance, such as humidity.
The output signal, which is only 2.5 V peak, is then input into U7B/U8B to
condition the 5 V signal. Hysteresis is provided through the use of R23, R29,
and R32 to ensure stable frequency switching. The output signal, JUMPUL, is
then sent to the microcontroller for processing. A typical frequency would be
around 37 KHz.
Microcontroller
The microcontroller is a Priority-Interrupt Controller (PIC) 16C73, used for
signal-processing and for control of all signals on the sensor module. The
device has three external ports, configurable as inputs and outputs. The
microcontroller operates from a precise time-base of crystal Y1, operating at
4 MHz. The instruction cycle time of the PIC is ¼ of that, namely at 1 MHz or
1 microsecond.
To ensure a clean power-up, U10 provides a fixed power-up reset to the
microcontroller. This integrated circuit also generates a rest in the event of a
brownout condition when the D+5 falls below a predetermined threshold.
If the main controller determines that the sensor module requires reset intervention, the reset line of the microcontroller, SMRES, is available to the main
controller.
The PIC device operates with an internal watchdog timer device that asserts
SMRES if the program execution operates outside normal conditions.
Expansion Devices
The digital multiplexer, U3, allows additional digital signals for processor control. It is a dual, 4-to-1 multiplexer that allows the microcontroller to use two
ports for 8-bits of information. The signals, DVSEL0 and DVSEL1, control
U18.
The buffered line-drivers, U13A/B and U14A, are used for signals that are
going off-board, namely SMDATA, SCDATA, and SCCLK. The SMDATA line
is used as a bi-directional line that can change from input or output “on-thefly” for data communication to the main controller. The SCDATA is similar,
with connection to the scale at connector J3. SCCLK is the buffered clock line
used for scale communications.
12
Buffer U14B provides an inversion for TM1SEL, producing TM2SEL to alternately enable and disable the temperature multiplexers.
Analog/Digital (A/D) Conversions
The A/D converter, U11, is an eight-channel, 12-bit, serial, interface device.
Control for the channels is software-configurable by the serial communication
line SSPCLK, ADCDIN, and COMOUT. The signal, ACENI, enables the A/D
converter for signal processing and is asserted twice every 21 milliseconds;
the A/D converter is read twice. In addition, the ADCIN and COMOUT are
driven at 21 millisecond intervals. The SSPCLK is shared with the EEPROM
on the sensor board and scale clock; therefore, the timing is not periodic.
All temperature information appears as a multiplexed signal on Channel 0,
and oxygen information appears as a multiplexed signal on Channel 2. Channel 4 enables the A/D converter to read its maximum input, and Channel 5
enables the A/D converter to read its minimum input to determine proper A/D
functioning.
The analog representation of fan pulses apply to Channel 6.
4.1.2ControllerThe controller accepts input voltages between the range of 90 V AC and
264 V AC through a universal input switching power supply. Voltages above
the safe operating range are clamped using a transorb diode.
The controller accepts input frequencies between the range of 48 Hz to 62 Hz
through a universal input switching power supply.
The stand supplies the input power and protective ground to the controller
and incorporates a 15 A circuit breaker and electromagnetic interference
(EMI) filtering components.
The controller provides AC power to the heater and the humidifier. These outputs are fused in the controller to protect the controller in the event of a short
circuit or electrical overload.
•Maximum heater voltage—264 V AC
•Maximum heater current—4.8 A
•Maximum humidifier voltage—264 V AC
•Maximum heater current—1.2 A
•Heater/humidifier fuse rating—6.3 A
The controller provides DC power to the following:
•The fan
•The sensor module
•The scale
•The SPO2 module, if available
•The airflow sensors
•The door switches
These outputs are current-limited in the controller to protect the controller and
the powered device if a circuit shorts or electrically overloads. These outputs
are regulated to ensure the output voltage is within the voltage specification
for the powered device. The microprocessor feeds and monitors the outputs
1 and 2 into the A/D converter.
The Impeller Movement Detector (IMD) P.C. board is positioned so that magnets pressed into the bottom of the impeller pass directly over a Hall effect
sensor mounted to the IMD P.C. board. The IMD circuit monitors the Hall
effect sensor’s pulse train, produced by the magnets when the impeller
rotates. The speed of the impeller is measured and compared with the predetermined maximum and minimum acceptance limits. If the impeller’s rotational speed is too fast or too slow, an impeller error signal generates. The
IMD circuit also detects if one, two, or all three magnets are missing or if an
old impeller without magnets is used. These errors produce the same error
signal to the controller as for low or high impeller speeds.
The controller’s +12 V supplies the power for the IMD circuit through a 301 Ω
resistor. This resistor and the load of the IMD circuit form a voltage divider
that sets the LONG signal voltage that remains constant. U1, a voltage regulator, supplies +5 V DC to the Hall effect sensor (U3) and the microcontroller
(U2). As the magnets pass above the Hall effect sensor, its open-drain output, U3-2, goes low, detecting the magnets’ field. When the magnets’ field is
removed, the oscillator is used as a reference. When an error condition is
detected, U2-2, the microcontroller’s output, drives low, which accomplishes
the following:
•Turns off Q1.
•Releases a portion of the load on the short signal, allowing it to rise above
the LONG signal voltage.
R2, the switch portion of the SHORT signal load, and R4, the unswitch position of the SHORT signal load, adjust to provide an approximate 1V swing
between the error and non-error outputs. The capacitors, C1 and C2, filter the
system’s supply.
The IMD P.C. board supports in-circuit programming (ICP) of the microcontroller. Programming is done after the unprogrammed microcontroller is populated with all the other components by connecting a programmer to the pads
labeled VPP, +5 V, CLK, DTA, and GND.
4.1.4Fan MotorThe controller sets the fan motor speed if the watchdog is not tripped. The
microprocessor supplies a pulse-width modulation (PWM) signal to an optocoupler for isolation. The output connects to an integrator circuit that converts
the PWM signal to an analog signal for the motor controller. The motor incorporates the Hall effect sensors for monitoring and control. One of the Hall
effect sensor outputs is fed to the microprocessor for measuring the motor
speed. If the watchdog timer trips, the fan motor speed is maintained at
1500 rpm ± 450 rpm.
The controller provides an alarm to indicate a failure of the fan to rotate.
When this occurs, the heater and humidifier disable, and an audible alarm
with a visual indication activates.
4.1.5Heater PowerThe controller monitors the heater power. A current transformer is in series
with the power to the heater and the humidifier. The output of the current
transformer connects to the A/D converter.
14
The system enables control of the incubator’s heater. The microprocessor
controls a solid state relay that controls the power to the heater. The microprocessor and the watchdog circuit control the safety relay, K3. The release
of the safety relay removes power from the heater regardless of the functionality of the heater triac.
4.1.6Humidity Heater PowerThe controller monitors the humidity heater power. A current transformer is in
series with the power to the heater and humidifier. The output of the current
transformer connects to the A/D converter.
The system enables control of the humidity heater. The microprocessor controls a solid state relay that controls the power to the humidity heater. The
microprocessor and the watchdog circuit control the safety relay, K3. The
release of the safety relay removes power from the humidifier heater regardless of the functionality of the humidity heater triac.
Oxygen Control
The system enables control of the oxygen pneumatics. The microprocessor
provides a PWM signal to the solenoid’s metal oxide semiconductor fieldeffect transistor (MOSFET).
The voltage to the oxygen solenoid is monitored and fed into the A/D converter. This circuit monitors the 12 V power supply and thermal fuse.
Light-Emitting Diodes (LEDs)
The microprocessor drives each light-emitting diode (LED). The hardware
watchdog timer circuit drives the alarm/system fail indicator. The power fail
detection circuitry drives the Power Fail indicator.
Audio Alarms
The audible alarm circuit incorporates an oscillator circuit to generate the
three alarm frequencies used:
•600 Hz
•1500 Hz
•2500 Hz
The microprocessor, the watchdog circuit, and the power failure detection circuitry drive the audible alarm circuit.
The audio volume is capable of three discrete sound levels. An analog
switch, incorporated in the audible alarm amplifier circuit, selects a 57 dB,
62 dB, or 65 dB output, as measured by International Electrotechnical Commission (IEC) 601-19-2:102.3. The microprocessor, the watchdog circuit, and
the power failure detection circuit control the analog switch.
The controller provides an audio output for power fail conditions. The alarm
oscillator is set for 600 Hz at 65 dB output, as measured by IEC 601-192:102.3. A timer circuit generates the cadence tone during power failures.
When a Power Failure alarm is activated, the following occurs:
•The Power Fail indicator on the front panel illuminates.
•An alarm sounds.
A high energy storage capacitor powers the power failure detection circuitry
and supplies power to the audible alarm and indicator for a minimum of 10
minutes. This capacitor charges while the unit is operating. When power is
15
Theory of OperationIsolette® C2000/C2000e
lost to the controller and the Power switch remains in the On position, the
storage capacitor supplies power to the power failure circuitry. The power failure circuitry incorporates a timer circuit that periodically enables the audible
alarm and Power Fail indicator at a cadence of 520 milliseconds off and 98
milliseconds on until one of the following occurs:
•The Power switch is turned off.
•The power is restored.
•The storage capacitor is depleted.
The Power Failure alarm silence is hardware-controlled. Pressing the AlarmSilence key during power failure silences the alarm for the duration of the
power failure. The Power Fail indicator flashes until one of the following
occurs:
•The storage capacitor is depleted.
•The Power switch is turned off.
•The power is restored.
The System Failure alarm is unaffected by the Alarm Silence key.
Interfacing
An interface port enables an RS-232 serial communication link. The serial
port is fully isolated from the remaining controller circuitry. The power to the
serial port interface circuitry derives from an isolated winding on the power
supply transformer. The RS-232 interface connector is a female DB-9,
mounted on the rear of the controller. An RS-232 transceiver converts the
RS-232 to logic voltage levels and vice versa. Optocouplers provide the isolation barrier and interface the RS-232 transceiver to the PC16550 UART. The
UART interfaces the serial port to the microprocessor bus. All lines connected to the RS-232 connector are filtered to block EMI. The RS-232 transceiver incorporates electrostatic discharge (ESD) protection.
An interface enables communication between the controller module and the
sensor module. The sensor module interface connector is a female DB-9,
mounted on the rear of the controller and comprised of a bi-directional data
line, a clock output line, and a reset output line. The data lines are fully isolated and optocoupled to the microprocessor. The controller provides isolated
power to the sensor module.
Door Switches
The controller connects to the two door switches that are wired in parallel.
The controller performs the following:
16
•Provides no more than 5 milliamperes (mA) of current to the switches
•Provides less than 6 V of power to the switches
•Monitors the return current to determine if either door is open
The switches are open when the door is closed. The input is protected with
transorb diodes and is filtered to block EMI and prevent ESD damage to the
controller.
The cooling fan provides a continuous flow of air through the controller to
remove heat generated by the various components inside the controller
enclosure. The cooling fan operates whenever power is applied to the controller. The cooling fan is equipped with a tachometer output signal that is
supplied to the microprocessor.
Ambient Temperature Sensors
The temperature sensors, located in the airflow of the cooling fan, are NTC
thermistors. The output signals of the redundant sensors feed into the A/D
converter.
Watchdogs
The first watchdog timer is internal to the microprocessor. If the software
does not update the watchdog timer within the required time frame, the internal watchdog resets the microprocessor and all peripherals connected to the
external reset line.
The second watchdog timer circuit attaches to the microprocessor bus. The
microprocessor continuously writes the following data to the watchdog timer:
•Data 55 hex (01010101 binary) to watchdog register #1.
•Data AA hex (10101010 binary) to watchdog register #2.
The watchdog timer trips in 1 second ±0.4 second unless the above
sequence is completed. Once the watchdog timer trips, the following occurs:
•The safety relay turns off, removing power from the heater and the
humidifier.
•The fan control reverts to closed loop control, maintaining a constant fan
speed regardless of the door’s position.
•The oxygen solenoid control from the microprocessor is overridden, and
the oxygen solenoid turns off so that no oxygen enters the hood.
•A constant alarm sounds for a minimum of 500 milliseconds.
•The system failure indicator illuminates.
The microprocessor resets the watchdog timer after a watchdog trip by sending the above data sequence.
Factory defaults are stored in program memory, flash EEPROM. System
parameters are configured and stored in the real time clock (RTC) module or
serial EEPROM. The RTC memory and random access memory (RAM) are
protected against corruption during power failures and are battery-backed for
a period of time.
The program is stored in reprogrammable memory and may be reprogrammed through a cable connected to the serial port of a computer. The
program memory is stored in a flash EEPROM. The RS-232 serial port operates at speeds of 115,200 baud to expedite the speed of the program download.
17
Theory of OperationIsolette® C2000/C2000e
Power Supply
The power supply is so designed that 1 second after disconnection of the
plug, the voltage between the supply pins of the plug and between either supply pin and the enclosure does not exceed 60 V by using a bleeder resistor
across the mains filter capacitor, if necessary.
4.2Air System
4.2.1Overall Functional
Description
The controller displays the air temperature and the skin temperature on an
electroluminescent display. Optional displays of the humidity and oxygen
concentration levels within the hood environment and the infant’s weight are
available. In addition, Trend displays of 2, 4, 8, 12, and 24 hours of all parameters (except weight, which is presented in days) are user-selectable.
To indicate which mode of operation, Air Mode or Skin Mode, is in control, the
set temperature of the controlling parameter remains on adjacent to the
actual displayed temperature. In addition, the rotating wheel in the Air or
Skin softkey designator rotates.
The forced air circulation system controls the temperature, humidity, and oxygen concentration (see Figure 4). The motor-driven impeller in the shell
draws a controlled amount of approximately 7 liters per minute (lpm) of room
air through the air intake filter.
The impeller also provides the internal circulation at a much greater flow than
that of the fresh gas inflow. The total flow of fresh and circulated air is
directed past the airflow sensor and around the heater. The air enters the
infant compartment up through the slots at the front and rear of the main deck
and then passes between the front and rear inner walls. The air circulates
past the sensor module, which contains the temperature sensing probe that
encapsulates the air temperature control thermistor and a high air temperature alarm thermistor. After circulating within the infant compartment, the air
then re-circulates down through a slot in the right end of the main deck, and
back to the impeller. When the access panel of the hood is open, the air continues to flow upward past the opening, Impeller ramps up to 2200 RPM to
creating a warm air curtain to minimize the drop in air temperature in the incubator. The temperature is regulated using either the incubator’s air or the
infant’s skin temperature as the controlling parameter; the desired mode is
selected by the front panel keys.
In either mode of operation, the heater output is proportional to the amount of
heat required to maintain the desired temperature.
4.2.2Air ModeIn Air Mode, the air temperature is maintained from 68 °F to 99 °F (20 °C to
37 °C) (99 °F to 102 °F (37 °C to 39 °C) in Temperature Override Mode), as
selected by the Air Set Temperature Up and Down arrow keys on the front
panel. A probe located in the sensor module monitors the incubator’s air temperature and compares it with the air’s set temperature setting. The probe
supplies this information to the heater control circuitry, which regulates the
heater output to maintain the air temperature setting. The actual air temperature appears on the Air Temperature display. A second sensor within the air
temperature probe serves as a backup to limit the maximum incubator temperature. If the high temperature limit activates, the heater shuts off.
In Air Mode, the infant’s temperature is a function of the air temperature and
the infant’s ability to establish and maintain its own temperature. A small
infant, or one with underdeveloped homeostatic control, may not be able to
maintain a stable temperature at the desired level.
In Air Mode, there is a 15-minute setpoint retention. When you first power the
unit on, the air setpoint temperature is 95 °F (35 °C); for example, if you
change the air setpoint temperature to 95.9 °F (35.5 °C), and a power failure
occurs, the air set temperature comes up to 95.9 °F (35.5 °C) if the unit turns
on again before 15 minutes elapse.
4.2.3Skin ModeIn Skin Mode, the infant’s temperature is selected from 93 °F to 99 °F (34°C
to 37 °C) (99 °F to 100 °F (37 °C to 38 °C) in Temperature Override Mode) by
the Skin Set Temperature Up and Down arrow keys on the front panel. A
temperature sensing probe attaches directly to the infant’s skin. The probe
supplies information to the heater control circuitry, which proportions the heater output to maintain the skin set temperature.
The air temperature still appears in Skin Mode, but as information only. If Air
Mode is selected while the skin probe remains connected, the Skin Tempe-rature display continues to display actual skin temperature, but it does not
control.
The sensor module accepts two skin probes. However, when the second skin
probe connects to the sensor module in Skin Mode, an alarm sounds, and the
message Remove Skin 2 Probe appears. To connect the second skin probe,
first select Air Mode. The controller then displays the two temperatures.
If Skin Probe 1 disconnects from its receptacle during Skin Mode, the Skin
Temperature display goes blank, an alarm sounds, and the heater turns off.
4.2.4Oxygen ControlAn oxygen sensor assembly mounted inside the sensor module adjusts the
flow of oxygen into the hood and controls the oxygen concentration level
within the incubator’s hood environment.
A valve regulates the flow into the incubator and periodically interrupts the
flow of oxygen into the incubator.
The sensor module houses two independent oxygen fuel cells that monitor
and control the oxygen concentration levels inside the incubator.
If the sensor module is outside of the hood environment during Oxygen
Mode, audible and visual alarms are enabled, and the flow of oxygen is interrupted.
In Oxygen Mode, the user sets the oxygen level control point from 21% to
65%. The high and low alarm limits automatically set to ±3% from the control
point. If the oxygen concentration level rises above or falls below the selected
setpoint limits, an audible and visual alarm occurs.
4.2.5Humidity Control ValveThe built-in humidifier provides humidification of the incubator from 30% to
95% RH in 1% increments. The humidifier reservoir permits visual inspection
of the water level.
If the water level in the chamber is depleted, an audible and visual Low
Humidity alarm occurs, indicating a need to replenish the water supply.
4.3Hardware
4.3.1Weighing ModeTwo load cells in a platform under the mattress perform the actual weighing
function. These cells provide a voltage that is proportional to the load on it.
The controller processes the voltage and displays it in either kilograms or
pounds/ounces on the Weight display.
The weighing routine is initiated by placing the infant on the mattress. If the
infant is already on the mattress, lift the infant off the mattress; when the system zeros, return the infant to the mattress to obtain the weight.
The Weigh key enables repeated weighing of the infant after the weighing
routine is initiated as described above.
4.3.2Trend DisplaysFour standard parameters are presented on Trend displays:
•Air temperature
20
•Skin temperature #1
•Skin temperature #2
•Heater power
Additional Trend displays are available when the unit is equipped with any of
the following options:
The Trend time is user-selectable in intervals of 2, 4, 8, 12, and 24 hours for
all parameters, except for weight, which provides a trend of 7 days.
4.3.3Interface ConnectionsA serial interface port is provided as a data terminal device and an RS-232
output.
The following parameters are available:
•Air and skin setpoint temperatures
•Current air and skin temperatures
•Oxygen setpoint
•Oxygen level
•Humidity setpoint
•Humidity level
•Infant weight
4.3.4RS-232 Serial Port Protocol
The RS-232 serial port connector is next to the AC power connector on the
front of the incubator. The serial port is configured for 2400 baud, 8 data, 1
stop, no parity, and is output only (see Figure 5).
During normal operation, a data packet transmits every 5 seconds. Each data
packet is entirely in American Standard Code Information Interchange
(ASCII) and is readable when displayed on any standard RS-232 terminal
device. A data packet consists of one 82-character line of text that is composed of a prefix, a data portion, a suffix, a checksum, and a carriage
return/line feed (CR/LF) pair.
The prefix identifies the data line. It consists of an opening bracket and an ID
character that are unique to the data line. The format of the data portion
depends on the specific data line. Any character positions within the data por-
21
Theory of OperationIsolette® C2000/C2000e
tion that are undefined transmit as spaces to enhance the readability of the
output. The suffix property limits the data portion and consists of a closing
bracket.
The checksum is two ASCII hexadecimal digits and represents an 8-bit acclamation of the ASCII characters from the prefix to the suffix, inclusive.
All monitored parameters, including temperature, oxygen, humidity, and
weight, transmit at the 5-second interval. Asynchronous events, such as
alarms or mode changes, transmit as they occur.
For example: If the air temperature and oxygen are low and Procedural
Silence is initiated, such as when an access door is open, the 10 character
value equals 000000405.
Certain fields, such as air temperature, have an implied decimal point. The
decimal point does not physically appear in the data stream.