SafetyBefore you use the SiCalPro sof tware progr am, make sure th at
you read and understand all the rela ted data. This inc ludes: the
applicable local safety procedur es, the instructions for the
equipment you are using with the software, and this
publication.
When you use equipment with this software, make sure that it
is serviceable and that it is in its normal condition for safe
operation.
Before you start an operation or procedure in this publication,
make sure that you have the necessary ski lls (if necessary, with
qualifications from an approved training establishment). Follow
good engineering practice at all times.
TrademarksAll product names are trademarks of their respective
companies.
Microsoft, Excel, and Windows are either registered
trademarks or trademarks of Microsoft Corporation in the
United States and/or other countries.
InstallShield is a registered trademark and service mark of
InstallShield Software Corporation in the United States and/or
other countries.
ii
Issue 1
Software purposeThe SiCalPro software is a virtual i nstrument panel that lets you
control and/or monitor a range of GE instruments from your
computer. Modem and internet facilities are avail able. The
software works with these instruments:
change the configuration, monitor a set of units (Safe area
only: up to 99 in a “daisy chain”)
• IDOS pressure module (IDOS module): monitor output, set
tare, set the zero/span
• Digital pressure module (DPM): monitor outp ut; set t are, set
the zero/span
You can also use the software for these tasks:
• Save records of the instrument operation in log files. You
can then see and/or print the result s (as text or as a graph).
• Design, save and print a calibration ce rtif icate. This can use
live data or data from saved log file s.
• Convert data files to *.xls and *.doc format.
• PC6-IDOS/PC6-IS only: For applicable units under test
(UUT), set up or change a calibration procedure. The UUT
can then import the new procedure file.
RegistrationThe SiCalPro software is automatically registered when you
purchase it. The copy of the software that you receive will wor k
with all the instruments specified at the time of purchase.
Note: To use the software, it must have the serial number for
each instrument you want to use it with. To add a new
instrument, please contact us at www.gesensing.com.
To use the SiCal Pro sof t ware, this is t he mini mum speci ficati on
for your computer:
• Operating system: Windows 95, Windows 98,
Windows 2000, Windows NT 4.0 or Windows XP
• 66 MHz processor (Intel® Pentium® processors
recommended)
• 128 MB of RAM (256 MB recommended)
• CD-ROM drive or DVD-ROM drive
Before you install the software, close all other Windows
applications.
1. Insert the CD into the CD-ROM drive
Note: If your computer is not set up to open a CD
automatically, you must open the file fr om the CD yourself.
Double-click the file setup.exe
2. When the InstallShield® Wizar d op ens on yo ur computer,
follow the on-screen instructions.
To prevent error mes sages, we recommend that you do not
change the default directories:
• C:\sicalpro
• C:\WINNT\sytem32\CVIRTE
• C:\Program Files\National Instr uments
TM
3. To install the National Instruments
Run-Time Engine, the installation can include these steps:
LabWindowsTM/CVITM
• Insert Source: Click on Cancel. To use the SiCalPr o
software, the Device Drivers CD is not necessary.
• Installation Complete: Click on Finish. To use the SiCalPro
software, there is no additional activation.
Installation1-1
Issue 1
1.3 After installationWhen the installation is complete, you can do the following:
• Chapter 2: Start the SiCalPro software. You can run more
than one copy of the software at the same time; one for
each instrument connected to the computer.
• Chapter 3: Connect the applicable instru ment or instrument s
to your computer.
Caution! To prevent elect ri cal damage when you
connect or disconnect the RS232 connection on your
computer , make sure the instrument power is set to off.
• Chapter 4: Set up and use the software with the applicable
instruments.
1.4 Remove the
SiCalPro software
If it is necessary to remove the software from a computer , ther e
are two programs to remove: SiCalPro and the related Real Time Engine. Use these step s to remove each program from
your computer:
1. On the Windows taskbar , click on the Start button and select
Settings > Control Panel > Add/Remove Programs.
2. Select the software program (SiCalPro or National
Instruments Soft ware) and select the “Remove” option.
3. Follow the on-screen instructions.
1-2Installation
Chapter 2: Getting started
2.1 IntroductionThis chapter gives a description of these items:
• the procedure to start the SiCalPro software
• the start up displays
• the procedure to set a different language
• the menu structure and options
2.2 Start the
software
2.2.1 General start up
procedure
You can run more than one copy of the software at the same
time; one for each instrument connected to the computer.
1. On the Windows task bar, click on the Start button and
select Programs > SiCalPro > SiCalPro.
Figure 2-1: Example display - OFFLINE
2. Connect the instrument to a RS232 communic ations po rt on
your computer. Refer to Chapter 3.
Caution! To prevent elect ri cal damage when you
connect or disconnect the RS232 connection on your
computer , make sure the instrument power is set to off.
3. To st art communication between the SiCalPro software and
the instrument, click on OFFLINE.
Getting started2-1
Issue 1
2.2.2 Start up - Program
group
Figure 2-2 shows the program group (Start > Programs >
SiCalPro):
NILWCVIRTE***: This re-installs the Run Time Engine.
SiCalPro: This starts the SiCalPro software. If you right-click
on this option, you can make a shortcut to put on your desktop.
vbdde: This starts a server connection t o use the softwar e over
a computer network.
Figure 2-2: SiCalPro program group
2.2.3 Start up - DisplaysWhen you start the software, Figure 2- 3 gives an example of
the OFFLINE window that opens for each type of instrument.
Table 2-1: Key to Figure 2-3
ItemDescription
(1)
(2)
(3)
(4)
(5)
(6)
Menu bar (All instruments): To select a task from a menu list. There is a
standard set of menu options plus special options for the different instruments.
Refer to Section 2.3.
Analogue pressure gauge (PC6, IDOS modul e/DPM inst rument s onl y): To show
the pressure values measured by the instrument.
Simulation display and keypad (PC6 instruments only). To simulate the display
and control functions on the instrument . The keypad is not available (gr eyed out)
in MONITOR mode.
Maximum and Minimum (All): To show the maximum and minimum pressures
measured by the instrument.
Unit mode (PC6, DPI 104 instruments only): The mode gives an indication of the
operations available.
• PC6 instruments: CONTROL (the PC6 simulation keypad is available) or
MONITOR ( the monitor and log operations are av ailable). This is se t by menu
Mode.
• DPI 104 instruments: NORMAL (the monitor and log operations are
available) or Daisy Chain (the daisy chain functions are avail able).
Communications status (All): The ONLINE/OFFLINE button starts or stop s the
SiCalPro communications with the inst rument. I f the i nstrument po wer i s of f and
it has an internal power on function, ONLINE sets the power on.
(7)
2-2Getting started
QUIT button (All): To close the SiCalPro software.
Table 2-1: Key to Figure 2-3 (Continued)
11
9
10
8
2
1
34
56
7
(a) PC6-IDOS/PC6-IS:
Special and common SiCalPro functions
(b) DPI 104/DPI 104-IS (Part only):
Special SiCalPro functions
(c) IDOS module (Part only)*:
Special SiCalPro functions
Note: The DPM
has almos t th e
same display as
the IDOS module
*
ItemDescription
Issue 1
(8)
(9)
(10)
(11)
Simulation display (DPI 104 instrument s only): To simulate the display functions
on the instrument.
Output panel (IDOS module/DPM instruments only): To show the output from
the instrument.
Tare button (IDOS module/DPM instruments only): To make an adjustment for
atmospheric pressure and s et the pressure value t o zero. The t are valu e is reset
each time the instrument goes ONLINE.
Reset Max/Min button (IDOS module/DPM instruments only): To reset the
maximum and minimum pressure values to zero.
Figure 2-3: Example st art up displays - Main functions
Getting started2-3
Issue 1
2.2.4 Start up - LanguageThe software can operate in these languages: Engli sh,
German, French, Spanish, Ital ian.
To change the language, use the menu bar:
Select *** Setup > Language and set the applicable option.
(Where *** = PC6 Pro; DPI 104, IDOS, DPM)
Note: Fo r correct operation, make sure you are usin g the same
language on the instrument.
2.3 The menu options Table 2-2 gives a description of the menu options available with
all the instrument types.
Note: Some of the fields that you are permitted to change
contain hidden characters. If you want t o change these fields
(text or numeric values), highlight all the applicable characters
and overwrite them.
Table 2-2: Menu Bar Functions
Menu Heading/
Options
Description
Mode: Sets the operations available with the instrument. The default mode of operation is
Monitor (Text) mode. To use this menu, the instrument must be ONLINE.
Control(PC6 instruments only)
In this mode, you can use the MENU, UP,
DOWN, ENTER, PRINT/LOG, and ON/OFF buttons to simula te
operations on the instrument keypad. Refer to the instrument user
manual.
Note: The Log! menu is not available in Control mode. If necessary,
store the data on the PC6 instrument and export it to the SiCalPro
software later.
2-4Getting started
Table 2-2: Menu Bar Functions (Continued)
Issue 1
Menu Heading/
Options
MonitorTwo options: Graphical or Text.
Graphical: To monitor the instrument output in a two-graph window
format. Top graph: pressure values; bottom graph: an additional value
applicable to the instrument.
• PC6 instruments: Use Control mode to set the bottom graph value
(volts, mV, mA, °C, maximum, minimum, % and tare). Other PC6
values are not available.
• DPI 104 instruments: Use the Bottom Line menu to set the bott om
graph value (maximum, minimum, %, switch).
• IDOS module/DPM instruments: Use the Function menu to set the
bottom graph value (maximum, minimum, tare, %, °C).
Buttons: EXIT (To go back to the normal display), STOP (stop a graph
record), START (st art a new graph record), SAVE (save a graph record
to a *.mon file).
Text (Default): To monitor the instrument output on a normal SiCalPro
display (Figure 2-3). The PC6 keypad is not available in Monitor mode.
Refer to Section 5.2.
Description
File: To work with the data files saved for an instrument (*.imp, *.mon, *.log files only). The
*.imp files from a PC6 instrument contain data from a completed calibration procedure or
normal log data.
Two options: Graphical or Text.
View
Graphical: To show a graph and related data for all the records saved
in a file. You can go through the records (
Text: To show the data for each record saved in a file. You can go
through the records (Prev., Next) and Close.
DeleteTo delete a file (with a two-step approval)
Import(PC6 instruments only) To import a log file (*.imp) from a connected
PC6 instrument. To prevent errors, make sure you use the file export
procedure for the PC6 instrument. Refer to the instrument user manual.
PrintTo print a file to the default printer for your computer. The file prints as
unformatted text. If the output is incorrect, adjust the printer settings or
use the option File > View > Graphical > Print.
▲ ▼), Print and Cancel.
Getting started2-5
Issue 1
Table 2-2: Menu Bar Functions (Continued)
Menu Heading/
Options
Description
ConvertTo convert a file (*.imp, *.mon, *.log) to Microsoft® Excel® or Microsoft®
Word format (Default file location: c:\sicalpro\data). The file paths for
these programs are set in: *** Setup > File Convert Path.
Convert File Type (Excel): St arts the Microsoft Excel program and
puts the data into the default file: SICALPRO.XLS (new data replaces
old data). To keep the data, save the file with a differ ent filename and/or
save it to another directory.
Note: SICALPRO.XLS must stay in the defaul t file location.
Convert File Type (Word): Starts the Microsoft Word program and
saves the data in the f ile: *.doc (* is your or iginal filename) . If necessary,
you can change the filename and move it to another file location.
Log! To make a permanent record of the instrument operation ( normal Log or Leak Test). T o
use this menu, the instrument must be ONLINE. The data is colle cted in a Text or Graphic
window format. The Log Type (Timed or Keyed) does not affect Leak Test operations.
Log! opens the Logging Set up window. Refer to Section 5.4.
PC6 instruments: Timed or Keyed; Timed - Logging Interval and Leak
Duration are all set with the instrument Menu button. Refer to the
instrument user manual.
To do a Leak Test, set the instrument to show Leak: Start Set-Time Exit
DPI 104, IDOS module/DPM instruments: Timed or Keyed; Timed Logging Interval and Leak Duration are all set in menu:
Certificate: To set up and control the calibration cer tificates for an instrument. In menu
*** Setup, you can set up a password for this funct ion. You can also set standard
header/footer text; set a standard adjustment to the printed layout; set a standard user
name for Calibrated by: ... .
NewTwo certificate options: Normal or Leak
Type in the Test and Standard instrument data; add certificate
headers/footers and a user name; use Stored data (*.log or *.imp files)
or Live data from an ONLINE instrument.
You can View, Print
or Save the certificate. The Save function makes
two files: *.cer and *.txt. Refe r to Section 5.5.
EditTo change a certificate.
Note: It is not possible to change these items: Test Instrument Details
(Range); Standard Instrument Detail s (Serial No, Accuracy); Results
(Stored Data).
2-6Getting started
Table 2-2: Menu Bar Functions (Continued)
Issue 1
Menu Heading/
Options
Description
PrintTo print a certificate to the default printer for your computer. Refer to
Section 5.5
DeleteTo delete a certificate. This deletes the *.cer and the related *.txt file
(with a two-step approval).
RS232: To set up the RS232 communications and do tests for corr ect operation, or to set up
communications for a modem. The modem f acili ty l et s you communica te wit h an inst rument
through the telephone system.
ConfigTo set up the communications between the SiCalPro software and the
RS232 port used by the instrument (for example: COM1). The settings
must be correct for the instrument. Refer to Chapter 3.
TestTo do a test for RS232 communications between the computer and the
instrument:
OK - Port settings correct; instrument ONLINE
No communication - Instrument OFFLINE
Failed - Port settings, power supply or connections are not correct
Modem SettingsTo use this menu, the communications must be OFFLINE. Set the
applicable values for COMS port, Telephone Number (with area codes
but no space characters), Modem Usage, Dialling Mode, Result Code
Format. Refer to Chapter 3.
*** Setup: There is a standard set of options plus special options for the different
instruments.
Change Password To set, change or remove the password for menu Certificate.
Passwords are case sensitive. Default: no password.
You must confirm a New Password. Use the tab key or click on an
available option (Enter New Password, Re-enter New Passwo rd;
RemovePassword ... No
▼).
If you lose your password, contact us at www.gesensing.com.
Getting started2-7
Issue 1
Table 2-2: Menu Bar Functions (Continued)
Menu Heading/
Options
Description
*** Setup To set and/or see configuration options for the connected instrument;
set an instrument certificate number (for example: the last calibration
certificate); set applicable special instrument options:
DPI 104: Daisy Chain. If you have more than one DPI 104 instrument in
a network, these options are avai lable: Abort Daisy Chain (t o stop daisy
chain communications); Set Auto Address (to set addresses 00 to 99);
Select DPI ... (to select the instrument you want to work wit h).
IDOS module/DPM instruments: Zero/Span (to do a zero and span
calibration); Abort Zero/Span (to stop the cali bration); History (to show
the last calibration date, span date and the overload history)
Refer to Chapter 4.
Software SetupThis gives data about the Si CalPro sof t ware v ersion an d the inst ru ment
it communicates with:
Software Serial No (the serial number of the instrument);
Issue date (the software version date);
Language (the software language specified in * ** Setup > Language);
Type (the instrument type).
Edition (Full or Demo). You can use a Demo version 30 times.
Internet Copy (Yes or No). An internet copy can send data to an internet
server. Refer to Chapter 3
Certificate Layout To set the Header offset (1 to 10) for your certificates. If you use
preprinted paper, set a value that stops the first line of the certificate
overwriting standard text and logos.
1 = Certificate starts on line 1.
Certificate
Header/Footer
To set up standard header and footer text for New certificates. Header
text goes above the Results section; footer text goes below the Results
section. Set the applicable text On or Off and click on Update.
User NamesTo set up user names (1 to 10) for the Certificate function.
To change a name, highlight the text and overwrite it. To set a default
name for New certificates, use the Active User buttons (
▲ ▼).
File Convert Path To set up the file paths for the Microsoft Excel and Microsoft Word
programs. If you use menu File > Convert, the SiCalPro software uses
the specified locations to find these programs.
Language
To change the default language for the software.
Note: For correct operation, make sure you are using the same
language on the instrument.
2-8Getting started
Table 2-2: Menu Bar Functions (Continued)
Issue 1
Menu Heading/
Options
Description
Units(IDOS module/DPM instruments only) To set up two more pressure
units:
User defined unit 1 (or unit 2) text: Up to 6 characters.
User defined unit 1 (or unit 2) conversion factor: An applicable
value that changes the pressure (in bar) to another pressure unit.
To update the standard Units list (on the menu b ar), Save the new unit s.
If a name is already used by a standard unit, the new unit is ignored.
Note: These units are only availab le in the SiCalPro sof tware. They are
not saved in the instrument. If you use the software on more than one
computer, set up the new units in each copy.
Internet(Optional software for a Windows NT network that supports Active
Server Pages - contact us at www.gesensing.co m) This option lets you
send instrument display data to an internet server. Another user can
then use the internet to log in and monitor the data.
You must set applicable val ues for Web address, Password, File
Update rate, Screen Refresh rat e , Maximum no. of records. Refer to
Chapter 3.
Units (DPI 104 and IDOS module/DPM instruments only): To change the instrument’s
pressure units to another unit on the list (mbar, psi, MPa ...). To use this menu, the
instrument must be ONLINE.
UNIT 1
UNIT 2
(IDOS module/DPM instruments only). Two pressure units specified in
menu *** Setup > Units.
Bottom Line (DPI 104 instruments) OR
Function (IDOS module/DPM instruments): To set the value for the bottom graph that
you see in menu Mode > Monitor > Graphical. The pressure units are the same as in menu
Units.
MaximumMaximum pressure
MinimumMinimum pressure
PercentagePressure as a percentage of the full scale output
Switch(DPI 104 instruments only) Pressure values to open/close a switch