Druck RPS8100, DPS 8100 Operating Manual

Page 1
Pressure measurement
for research & industry
Druck Limited
Leicester LE6 0FH
England
Tel: 0116 231 7100
Trench Etched
Resonant Pressure Sensor
TERPS 8000 Series
User Manual
Groby
K0473
Druck Limited 2013 This document is the property of Druck Limited and may not, either in part or whole, be copied or otherwise reproduced, communicated in any way to third parties, nor stored in any data processing system, without the express written authority of Druck Limited.
Page 1 of 42 K0473 Issue 5
DO NOT PRINT THIS PAGE
Page 2
mendment Record
A
Iss No Date C/N No Originator Typed
1 20/12/10 N/A Robert Lee Robert Lee 156219 New document
2 24/06/11 26758 Robert Lee Robert Lee 165022
3 26/10/11
4 30/01/12 27572 Robert Lee Robert Lee 176971
5 08/11/12
Approvals Engineering
Technical and operational accuracy
M COLE
Marketing Customer/market suitability
I ABBOTT
27245 27300
28898 29134
Glenn Roles Glenn Roles 171244
Robert Lee Robert Lee 192333
Technical Communications Compliance with style guidance and presentation
G ROLES
Workflow No.
Amendments
Delete two 80xx pressure ranges, change power supply voltage from 5 to 6 Vdc. Change Pascal to hector-Pascal, -
0.2mV/°C to -2mV/°C. Equations changed on page 5. Pressure range changed on page 4 25 to 40khz. Page 11, 7.9b, add 5% of span after `An applied pressure’. Add Hastelloy versions and up-date RS232/RS485. Add section 6.5 Stored Coefficients and appendix A
Page 2 of 42 K0473 Issue 5
DO NOT PRINT THIS PAGE
Page 3
Print Instructions
Finished Size: A5
• Print in colour on white throughout (covers + text), saddle stitched.
• Cover to 285 gsm, content to 100 gsm.
Translate into French, German, Italian, Spanish, Portuguese Brazilian, Simplified Chinese and Russian.
THIS HARDCOPY IS NOT TO BE USED AS CAMERA COPY.
Page 3 of 42 K0473 Issue 5
DO NOT PRINT THIS PAGE
Page 4
Page 4 of 42 K0473 Issue 5
DO NOT PRINT THIS PAGE
Page 5
GE
Measurement & Control
Trench Etched Resonant Pressure Sensor
8000 Series
User Manual - K0473
Page 6
© 2013 General Electric Company. All rights reserved.
Page 7
Safety The manufacturer has designed this sensor to be safe when
operated using the procedures detailed in this manual. Do not use this sensor for any other purpose than that stated.
This publication contains operating and safety instructions that must be followed for safe operation and to maintain the sensor in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage.
Use qualified* personnel and good engineering practice for all procedures in this publication.
Pressure WARNING:
Do not apply pressure greater the maximum safe working pressure to the sensor.
Toxic Materials There are no known toxic materials used in this sensor.
Maintenance The sensor must be maintained using the manufacturer’s
procedures and these should be carried out by authorised service agents or the manufacturer’s service departments.
Technical Advice For technical advice contact the manufacturer.
* A qualified technician must have the necessary technical
knowledge, documentation, special test equipment and tools to carry out the required work on this equipment.
EC Directives This equipment complies with:
BS EN 61000-6-1:2007 BS EN 61000-6-2:2005 BS EN 61000-6-3:2007 BS EN 61000-6-4:2007
BS EN 61326-1:2006 This equipment complies with the requirement of the Pressure Equipment Directive 97/23/EEC.
For further details see the Sales Data Sheet. A full conformity certificate is available from the manufacturer.
Contact GE Measurement & Control:
www.ge-mcs.com
i K0473 Issue 5 - [EN] English
Page 8
Abbreviations
The following abbreviations are used in this manual.
Note: Abbreviations are the same in the singular and plural.
aAbsolute ASCII American Standard Code for Information Interchange atm Atmosphere bps Bits per second cmHg Centimetre of mercury CR Carriage return DC Direct current EEPROM Electrically Erasable Programmable Read Only Memory FS Full-scale ftH
OFeet of water
2
hPa Hecto Pascal inHg Inch of mercury inH2OInches of water
kg Kilogram
2
kg/cm
kg/m
Kilogram per centimetre squared
2
Kilogram per metre squared
kPa Kilo Pascal
lb/ft2 Pounds per foot squared LF Line feed mA Milli Ampere mbar Millibar mbar a Millibar absolute mHg Metre of mercury mH
OMetres of water
2
mm Millimetre mmHg Millimetre of mercury mmH2O Millimetres of water
MW Mega ohm MPa Megapascal n/a Not applicable
[EN] English - K0473 Issue 5 ii
Page 9
Abbreviations (continued)
Pa Pascal pc Personal computer PIN Personal identification number ppm Parts per million psi Pound per square inch RS232 Serial data communication standard RS485 Serial data communication standard TERPS Trench etched resonant pressure sensor VVolt VA Volt amp °C Degrees centigrade
iii K0473 Issue 5 - [EN] English
Page 10
[EN] English - K0473 Issue 5 iv
Page 11
1Introduction
1.1 General The 8000 Series uses TERPS (trench etched resonant sensor)
technology. The RPS 8000 produces a frequency and diode voltage output. The DPS 8000 includes a microprocessor to produce a serial digital output.
Figure 1 TERPS General view
1.2 Configuration The following variables are specified at the time of ordering.
a. Pressure range b. Pressure connector c. Cable length d. Option A - Increased accuracy e. Accessories
i. User Manual (this publication) ii. Calibration certificate.
1 K0473 Issue 5 - [EN] English
Page 12
2 Installation TERPS 8XXX is a harsh media isolated product. Isolation is achieved
by hermetically sealing the sensor chip in an oil filled chamber. The weight of this oil gives a g-sensitivity as a pressure offset error.
To calibrate the TERPS 8XXX, the unit is mounted vertically with the pressure port at the lowest point. Orientation other than this produces a pressure offset error as specified in the datasheet. The error is most noticeable at lower pressure ranges.
: The g-sensitivity will also create an error in a high vibration
Note
environment and the unit should be mounted accordingly.
TERPS 81XX is not a harsh media isolated product. There is negligible change in offset due to mounting position and vibration. Because the pressure media comes directly into contact with the sensor chip, care must be taken to ensure the pressure media does not damage the sensing chip.
2.1 Connecting TERPS to pressure source
When mounting the sensor, seal the mating surfaces. Failure to properly seal may affect performance or calibration accuracy.
Male threaded pressure connectors must not be sealed or constrained against the face at the base of the thread. The forward cone or flat face should always be used as indicated below.
Depth versions should not be used at hyperbaric pressures above 70 bar (approximately 700 m of water).
[EN] English - K0473 Issue 5 2
Page 13
3 TERPS media
compatibility
3.1 TERPS 81XX 0 to 3.5 bar: non-condensing dry gases compatible with silicon,
silicon dioxide, RTV adhesive, stainless steel 316L and glass.
3.2 TERPS 80XX- Ranges 0 to 200 bar: Fluids compatible with Stainless Steel 316L
and Hastelloy C276. Ranges 201 to 350 bar: Liquids and group II gases compatible with
Stainless Steel 316L and Hastelloy C276. Fluid classification complies with EC directive 67/548/EEC.
Statements comply with Pressure Equipment Directive 97/23/EC.
3.3 TERPS 82XX- Ranges 0 to 200 bar: Fluids compatible with Hastelloy C276.
Ranges 201 to 350 bar: Liquids and group II gases compatible with Hastelloy C276.
Fluid classification complies with EC directive 67/548/EEC. Statements comply with Pressure Equipment Directive 97/23/EC.
3.4 TERPS 83XX- Ranges 0 to 200 bar: Fluids compatible with Hastelloy C276.
Ranges 201 to 350 bar: Liquids and group II gases compatible with Hastelloy C276.
Fluid classification complies with EC directive 67/548/EEC. Statements comply with Pressure Equipment Directive 97/23/EC.
4 Pressure
containment
4.1 TERPS 81XX Ranges 0 to 3.5bar 7 bar maximum.
4.2 TERPS 8XXX- Ranges up to 7 bar: 70 bar maximum.
Ranges >7 to 100 bar: 200 bar maximum.
3 K0473 Issue 5 - [EN] English
Page 14
5 Applying power to
TERPS
5.1 Frequency and diode versions:
5.2 RS485 and RS232 Versions
6 Measuring TERPS
6.1 Frequency and diode version
6.2 Measuring the Frequency
The TERPS sensor should be connected to stable power supply between 6 and 28 V
Low jitter version require a less than 10 mA during normal operation.
Low power versions require less than 3.5 mA during normal operation.
During power-up the sensor draws more than current than above, it is recommended that the power supply can supply a short term peak of at least 20 mA.
The TERPS sensor should be connected to a stable power supply between 11 and 28 V
at 32 mA.
The frequency and diode version of the TERPS sensor requires the user to measure a frequency and a voltage to calculate pressure. The TERPS sensor has a very high level of repeatability that needs to be matched by the measurement system.
The frequency of the TERPS die is output as a TTL square wave referenced to ground in the range of 25 to 40 kHz. The frequency of the square wave needs to be measured to a better than 6.5 digits (i.e. 30 KHz to better than 0.05 Hz) to allow the sensor to meet quoted specification.
To make sure the calculated output is correct, the measurement device should be regularly calibrated refer to the manufacturer’s instructions against a traceable standard.
.
DC
. Current drawn is a nominal 16 mA peaking
DC
6.3 Measuring the Diode Voltage
[EN] English - K0473 Issue 5 4
The diode signal is referenced to ground. It is a nominal 0.5 V at room temperature, and changes with a nominal -2mV/°C. To achieve the quoted specification of the sensor this signal must be measured to better than 0.01mV. To make sure the calculated output is correct, the measurement device should be regularly calibrated refer to the manufacturer’s instructions against a traceable standard.
: Best practice, when measuring a TERPS device, take both
Note measurements together. Where this is not possible the measurements should be taken as closely together as possible. Control the environment in which the TERPS is situated so it is not subjected to sudden changes in temperature and pressure.
Page 15
6.4 Calculating
Where:
P = Applied pressure (psi) K
y
= Calibration coefficient
x = Pressure signal in Hz minus normalising factor X y = Temperature signal in mV minus normalising factor Y
[] [] []
)()()()(
)()()()(
)()()()(
32
23
22
22
12
21
02
20
31
13
21
12
11
11
01
10
30
03
20
02
10
01
00
00
yxKyxKyxKyxK
yxKyxKyxKyxK
yxKyxKyxKyxKP
+++
++++
++++=
Pressure
During manufacture, a large number of pressures and temperatures are applied to the sensor to build-up a mathematical description of its behavior. This data is then used to generate a polynomial equation that relates the measured outputs from the sensor (bridge output and temperature output) to the applied pressure.
nd
The following example assumes a 2 order temperature signal (T
3
). If other orders of fit are used, these
order pressure (P2) and a 3rd
equations will need to be expanded to suit. This equation is in the form:
Therefore:-
The x variable is the pressure sensitive, measured output from the sensor in Hz minus the normalising factor X.
The y variable is the temperature sensitive, measured output from the sensor in mV minus the normalising factor Y.
The coefficients Kij, X and Y are printed on the top right-hand corner of the calibration certificate supplied with the pressure sensor. A sample set is shown below:
SN refers to the serial number of the unit, CS is a check sum of the
5 K0473 Issue 5 - [EN] English
data and can be ignored.
Page 16
6.5 Stored Coefficients
The coefficients are also stored internally on a serial EEPROM. See Appendix A for details on data format and communication information.
The internal EEPROM is only available on some electrical connector variants of the RPS 8000 series, see datasheet or calibration certificate for electrical connection details.
7 Using RS485 and
Output is compensated pressure in user specified units.
RS232 versions
7.1 Serial Data Communications
7.2 Applications The TERPS may be used as a single transducer directly connected
7.3 RS485 Connections
RS485 bi-directional digital communication, no handshaking, user programmable baud rate between 300 and 9600 baud, 8 bit data, 1 stop bit. Format: ASCII text, pressure reading including pressure units. User programmable reading rate from one reading per second to one reading per 999999 seconds.
to a serial interface. It can also be part of a network of devices using the programmed addressing facility. Each device on the network is a slave with a master device controlling the communications.
The installed transducer is completely controlled by the serial interface. The TERPS transducer contains its characteristic data in
an integral E
Stand Alone RS485 Electrical connections RS485-B (or RS485 -) +ve supply 0V supply RS485-A (or RS485 +) Transducer body (screen)
Notes
2
PROM.
:
Other connectors available refer to manufacturer.
For electrical connections refer to the sales data sheet.
[EN] English - K0473 Issue 5 6
Page 17
Figure 2 RS485 Direct connections
Figure 3 RS485 Network connections
7.4 RS232 Connections
7 K0473 Issue 5 - [EN] English
Stand Alone RS232 Electrical connections
Tx +ve supply 0V supply Rx Data ground (GRD) connects to 0V
Page 18
Figure 4 RS232 Direct connections
[EN] English - K0473 Issue 5 8
Page 19
7.5 Command Summary
The following functions can be performed using the serial data link:
FUNCTION COMMAND
Measurement
Transmit pressure reading........................................ ...............R
Trigger new reading and transmit........................ ...............G
Read raw data................................................................ ...............Z
Information
Identity and set-up information............................. ...............I
General set-up
Set automatic transmission interval.................... ...............A
Set filter............................................................................. ...............F
Set device address........................................................ ...............N
Pressure measurement speed................................ ...............Q
Set output pressure units.......................................... ...............U
PIN Set-up
Digital output calibration........................................... ...............C
Set full-scale ................................................................... ...............H
User message................................................................ ...............M
Communication settings............................................ ...............O
Change PIN...................................................................... ...............P
Set offset........................................................................... ...............S
Command Terminators.............................................. ............... <CR>
.............................................. ............... <CRLF>
:
Notes
Waiting 20 seconds after entering a valid command has the same effect as <CR>.
The backspace character deletes typing errors.
Commands are processed on receipt of a <CR> or a 20 second time out.
Adding a * before a command character causes a text reply.
All <LF> characters are removed from the incoming string before processing, so <CR> = <CRLF>, number of terminating characters set in ‘O’ command effects the reply.
9 K0473 Issue 5 - [EN] English
Page 20
FUNCTION QUERY
General set-up
Automatic transmission interval?.......................... ...............A,?
Filter setting?.................................................. ............... F,?
Device address?............................................. ............... N,?
Pressure measurement speed?.............. ............... Q,?
Output pressure units?............................... ...............U,?
PIN Set-up
Digital output calibration........................... ............... C,?
Full-scale value?............................................ ...............H,?
User message?............................................... ...............M,?
Communication settings?......................... ............... O,?
PIN set? ............................................................. ...............P,?
Offset value?................................................... ...............S,?
Factory Set-up (Query only)
Crystal Reference Frequency................... ...............E,?
Load Calibration Coefficients.................. ...............L,?
Analogue/Digital Convertor Set-up...... ...............T,?
Transducer Type and Settings................. ...............V,?
Write to EEPROM........................................... ...............W,?
7.6Measuring Pressure
7.7 Communication Modes
The frequency output from the sensor is counted for a fixed nu mber of cycles by the microprocessor. At atmospheric pressure this takes a nominal 400 milliseconds. The output from the integral temperature sensing diode takes 400 milliseconds for a single conversion through the discrete converter. Measurement cycles are continuous, each cycle delivers new values of frequency and temperature.
The TERPS smart transducer operates in one of two operating modes: direct and addressed. These modes are controlled by sending the transducer an address number.
Direct mode
The valid address in direct mode is zero. In this mode, the transducer continuously transmits data at an interval set by the user. Before any command can be sent to the transducer, the data stream must be stopped by sending a single byte character or back space. The stop character is lost and not parsed with the instruction.
Note: On shipment, sensors are set to direct mode as default.
Addressed mode
The valid address in addressed mode is a number between 1 and
32. In this mode, data has to be polled from the transducer using the valid address. All devices on the network respond to the address 0 for the G, R, I and Z commands. The reply timing depends on the address of the sensor, each sensor delays response by a
[EN] English - K0473 Issue 5 10
Page 21
time equal to the message length times (it's address - 1) times a character transmit time. For example sensor address 20 will delay by a time to allow 19 sensors to reply with the same message length. The command 0:I returns the sensor serial number only.
Special Note
Transducers are shipped with the following initial settings (unless alternatives have been requested):
Units PIN Address Long error messages Communications set Update rate Filter User message
-
mbar
-
000 (not set)
-
Direct mode
-
Present
-
9600, 8, N, 1
-
1 reading/second
-
disabled
-
none
7.8General Command Format
A sequence of single letter commands control the transducer. Some commands require one or more numeric values to follow the command letter. The correct number of parameters must be sent for this particular command. A comma separates all command fields. The transducer accepts upper-case or lower-case letters. Single commands should be in the form:
Command, P1, P2......................................................... ...............Pn
Where, Command is a single letter command
P1 to Pn are the required numeric values in the form:
(-) mmm.ddd
or
(-)m.ddddE(-)xx
e.g. 123.456 or 1.23456E02
Example 1 - Single Command
A,3<CR> selects the output update of one reading every 3 seconds (alternative command - a,3<CRLF>)
Command Strings
The transducer also recognises a string of commands each separated by a semicolon. After receiving a valid terminator, the transducer accepts the command string. Example 2 - Multiple Commands (Command String)
P, PIN, new PIN; U, units;A,5 <CRLF> This command string sets the new PIN, the units in which the pressure reading is transmitted and the interval between transmissions.
: A command with the *prefix causes a reply to change to text.
Note
i.e., Command: *R Reply: <Pressure value><pressure units>
11 K0473 Issue 5 - [EN] English
Page 22
Command String Limitations
A command string must not be more than thirty characters. The transducer ignores all the commands in a string containing more than thirty characters.
Network Mode Syntax
In this mode an address field must be added to the start of the general format as follows:
Address: command,P1,P2.......................................... ...............Ppn
where: ‘Address’ is the address of the transducer to send the command
‘:’ is the end of the address field ‘command’ is the single letter instruction ‘P1’ to ‘Pn’ are the required numeric values
Example 3 - 1:R<CR>
Get device @ address 1 to transmit pressure
7.9 Error Message Error messages are generated from two sources:
a. User commands
In a command string, the transducer accepts correct commands and rejects incorrect commands. An incorrect command causes the transducer to send an error message related to the command.
Error Message Format
The form of the error message:
ERROR nn - where nn is a decimal number
Defined error numbers are detailed in Table 1, page 28.
b. Transducer fault
A transducer fault sends an error message blocking the transmission of pressure data and sending the fault message in place of data.
Error messages
An applied pressure, 5% of span out of the calibrated range, causes:
*Over Pressure*
*Under Pressure*
A damaged transducer or no frequency output from the transducer causes:
**** NO RPT ****
[EN] English - K0473 Issue 5 12
Page 23
8Command
Definitions
The commands have been defined for direct mode operation (address 0). In the addressed mode an address field will be added to the start of the command and reply. For example:
Get reading command
Command
direct : R<CR> addressed : Address:R<CR>
: Address:*R<CR>
Reply
direct : Pressure value pressure units<CR> addressed : Address:Pressure value pressure units<CR>
: Address*:<Pressure reading><units><CR>
8.1 Measurement Commands
R - Get Reading This command polls the devices in the network for a reading.
Although used mainly in the addressed mode it can also be used in direct mode.
Command: R<CR>
Reply: Pressure value pressure units<CR>
Result: A new pressure reading is transmitted.
Command: *R<CR>
Reply: <Pressure value> <pressure units><CR>
Result: A new pressure reading and the units of pressure measurement are
transmitted.
Query: n/a
G - Get New Reading and Transmit
This command starts a new "measurement cycle" and, when completed, transmits the reading.
Command: G<CR>
Reply: <Pressure reading>(CR)
Result: Transmission of the reading takes 1.5 times the current
measurement interval set by the Q command.
13 K0473 Issue 5 - [EN] English
Page 24
Command: *G<CR>
Reply: <Pressure reading>,<Units>(CR)
Result: Transmission of the reading takes 1.5 times the current
measurement interval set by the Q command.
Z - Read Raw Data This command sends the last complete "measurement cycle" of the
transducer as frequency and voltage.
Note: The frequency value is proportional to the applied pressure
and the voltage is proportional to the temperature sensed by the diode.
Command: Z<CR>
Reply: <Frequency><Diode voltage>(CR)
Command: *Z<CR>
Reply: <Frequency (Hz)><Diode voltage (mV)>(CR)
Response: In direct mode changes between automatic transmission mode
(see A command) and transmission of the frequency and voltage. In network mode sends the last complete "measurement cycle" of the transducer as frequency and voltage.
Query: n/a
8.2 Information Commands
I - Transducer Identity This command gets the identity and set-up data from the
transducer. Either formatting character can be used with this command.
Command: I<CR>
Reply: <Unit type>, <Serial Number>,<Style>,<Range Unit
Number>,<Minimum Pressure>,<Maximum Pressure>,<Calibration Date>,<Software Version>,<Transmission Interval>,<Units sent Y/N>,<Measurement Speed>,<Filter Factor>,<Filter Step>,<User Message>,<Units number>,<Pin set Y/N>,<User Zero Y/N>,<>(CR)
Command: *I<CR> or <I<CR>
Reply: "Device : Device type" (TERPS-1234)<CR>
"S/N : Transducer serial number"<CR> "Range : Pressure range"<CR> "Address : Transducer address"<CR> <CR> "Press any key to continue . . . "<CR>
Query: n/a
[EN] English - K0473 Issue 5 14
Page 25
8.3 General Set-up Commands
A - Auto-send Reading This command sets the interval between each transmission. The
interval can be to one decimal place, i.e., 1.5 seconds. Entering a negative number causes an error.
: Setting this interval to less than the measurement interval (Q
Note
Command), causes the output of the same reading twice or more.
Command: A,nnnnnn<CR>
Where: ‘nnnnnn’ is a number between 0 and 999999. This is the interval in
seconds between successive transmissions.
Reply: None in network mode.
In direct mode: <Pressure reading>(CR)
<Pressure reading>(CR)
Result: A reading is transmitted every nnnnnn seconds in the format set in
the set-up commands. The interval input is non-volatile.
Command: *A,nnnnnn<CR>
Where: ‘nnnnnn’ is a number between 0 and 999999. This is the interval in
seconds between successive transmissions.
Reply: None in network mode.
In direct mode: <Pressure reading><Units>(CR)
<Pressure reading><Units>(CR)
Result: A reading is transmitted every nnnnnn seconds. The interval input
is nonvolatile.
Query: A,?
Reply: <Transmission interval set>,<Y if Units displayed, N if units not
displayed>(CR)
Query: *A,?
Reply: Interval = <Transmission interval set>(CR)
Units = <Yes if Units displayed, No if units not displayed>(CR)
15 K0473 Issue 5 - [EN] English
Page 26
F - Change Filter Settings This command sets the filter applied to the pressure reading (R
Command, network mode) or direct mode pressure reading.
Command: F, (Filter factor),(Filter step)
Where: The last two pressure readings are filtered using a % factor of the
sum of the two values. The filter step takes a value expressed as an integer percentage of full-scale. If step = 0, then the filter is disabled. If the percentage change between last pressure reading and the current pressure reading exceeds the filter step value the current reading passes through unfiltered. The last two pressure readings are filtered using a % factor of the sum of the two values,
. (1-F/100) + PN . F/100
P = P
o
Where
P = New returned pressure Po = Last returned pressure
= New measured pressure
P
N
F = Filter factor as integer between 1 and 99
The filter step . . . . . .”
Reply: <None>
Result: A filtered value within the filter step band, this change is
non-volatile.
Query: F,?
Reply: <Filter Factor>,<Filter Step>
Query: *F,?
Reply: Filter Factor = <Filter Factor>
Filter Step = <Filter Step>
N - Set Device (transducer) Address
This command sets the address of the transducer. A valid address is a number between 0 and 32. Setting the address to zero causes the transducer to go into direct mode. All other address settings cause the transducer to go into network mode. This command has a secondary function and uses N,0 or *N,0 to change the length of error messages (where: 0 = direct mode).
Command: N, (new address)
Reply: (none)
Command: N,0
Reply: (none)
Result: Changes to short error messages.
where: 0 = direct mode
[EN] English - K0473 Issue 5 16
Page 27
Command: *N,0
Result: Changes to long error messages.
Example: Address 1 change to long error messages:1:*N,1
Query: N,?
Query: *N,?
Q - Set Measurement Speed
Command: Q (speed setting)
Query: Q,?
Query: *Q,?
Reply: (none)
where: 0 = direct mode
Reply: <Device Address> (CR)
Reply: Device Address =<Device Address> (CR)
This command changes the rate of measuring by changing the number of output cycles counted for each reading. The RPT produces an output at 16000 cycles per second and by selecting one of 6 settings (0 to 5), a sample rate can be made. The highest rate (5) causes the most electrical noise.
: Increasing the Q speed may, for the next two seconds, cause
Note
reading errors or !020 No Frequency error message to be returned.
Reply: (none)
Reply: <Measurement Speed>(CR)
Reply: Measurement Speed =<Measurement Speed>(CR)
Q Command Setting
2[1] 16000 360 530 1.0
[1] Default setting for the accuracy stated in the specification. [2] This interval varies between minimum and maximum.
U - Set Units This command sets the units of pressure measurement.
17 K0473 Issue 5 - [EN] English
Resonant Sensor Cycles Counted
0 64000 1400 2100 4.0 1 32000 710 1050 2.0
3 8000 180 260 0.50 4 4000 90 130 0.25 5 2000 50 75 0.25
Measurement Interval milliseconds [2]
minimum maximum
G Command fixed setting
Page 28
Either formatting character can be used with this command.
Command: U,nn<CR>
Where: ‘nn’ is a number between 0 and 24 used to select the required units.
Reply: <none>
Result: Changes the pressure units of the transmitted reading.
Query: U,?
Reply: <unit number> (CR)
Command: *U
Reply: (Interactive)
Current units are <Units> (<unit number>) (CR) (CR)
0 - mbar<CR> 13 - mH 1 - Pa<CR> 14 - torr <CR> 2- kPa<CR>15- atm <CR> 3- MPa<CR>16- psi <CR>
4- hPa<CR>17- lb/ft2<CR> 5- bar<CR>18- inHg <CR>
2
6- kg/cm
<CR> 19 - inH2O04<CR>
7- kg/m2<CR> 20 - ftH2O04 <CR> 8 - mmHg<CR> 21 - mbar <CR>
9- cmHg<CR>22- inH 10-mHg<CR>23-ftH 11 - mmH
O<CR> 24 - mbar <CR>
2
12 - cmH2O<CR>
2
O <CR>
O20<CR>
2
O20<CR>
2
: mbar can be selected using code 0, 21 or 24.
Note
Result: Each of these non-volatile commands changes the pressure units
of the transmitted reading.
Query: *U,?
Reply: Units = <Units> (<unit number>) (CR)
[EN] English - K0473 Issue 5 18
Page 29
8.4 PIN Protected Set-up Commands
C - User Calibration This command allows the user to perform a two point calibration on
the output of the device. The calibration routine requires two pressures to be applied within the operating extremes and in units of pressure measurement set in the transducer. The software calculates an offset and gain correction to all subsequent readings.
Command: C or *C
Reply: (Interactive)
Reply: Enter PIN’ => (enter PIN)
Reply: “Apply first pressure, then send <CR>” (when stable, send <CR>)
Reply: “Enter applied pressure (<units>)” =>(enter pressure value in
specified units) <CR>
Reply: “Apply second pressure, then send <CR>” (when stable, send <CR>)
Reply: “Enter applied pressure (<units>)” =>(enter pressure value in
specified units) <CR>
Reply: “Pressure gain = <new pressure gain>” (CR)
“Pressure offset = <new pressure offset>” (CR) “Accept these values? (Y/N) =>(Enter Y or N)
(Y answered) “EEPROM updated” <CR> (N answered) “Current settings left unchanged” <CR>
(If invalid data entered) “!023 Bad Cal Pres” <CR>
Command: *C,PIN,mode,?<CR>
Non-interactive calibration requires the C command to be used twice, i.e. two calibration pressures must be entered for a straight line fit.
Proceed as follows:
Command: *C,PIN,1,?<CR>
Where: ‘PIN‘ is the PIN of the transducer and ‘1‘ is the first pressure value.
Apply pressure 1, when stable send:
C,(PIN),1,<CR>
Note: Sending *C,PIN,1,?<CR> causes the current stored data for
pressure 1 to be sent.
Apply pressure 2, when stable send:
C,(PIN),2,<CR>
19 K0473 Issue 5 - [EN] English
Page 30
Query: C,PIN,mode?
original slope
Output reading
Applied pressure
0
12
3
4
0
1
2
3
4
new set-point (span) using the H command
new off-set (zero) using the S command
Reply: <Measured>,<Applied><Got first point (Yes/No)> (CR)
Query: *C,?
Reply: “Measured Pressure = <Measured>” (CR)
“Applied Pressure = <Applied>” (CR) “First Point Done=<Yes/No>(CR)
8.5 H - Change Slope This command changes the slope of the output function of the
transducer. The relationship between the applied pressure and the output
signal can be shown in graphic form:
[EN] English - K0473 Issue 5 20
Figure 4-1 Output to Applied Pressure Graph
:
Notes
If S remains at 0 (the before value) and the H value is changed, the slope "pivots" around zero.
If H remains the same (slope) and the S value is changed the whole slope goes higher or lower by the S value.
Page 31
H and S Commands Proceed with caution when using these commands.
Do not use both commands together. Use one command and then assess the result
before continuing.
Command: H,(PIN),(pressure)
Reply: <none>
Command: H,?
Reply: <slope>,<setpoint>(CR)
Command: *H,?
Reply: Slope =<slope>(CR)
Set At =<setpoint><Units>(CR)
Where: The slope is a gradient and set-point is the reading value entered for
an applied pressure.
M - User Message Th is comman d s et s a us er de fi ned message of 16 characters stored
in non-volatile memory. Only ASCII characters 32 to 128 are stored, all others will be removed from the final string. ASCII 0 indicates the end of message and stores this in memory but does not transmit this character. The (CR) character is transmitted with the message but not stored in memory. Attempting to store the colon character causes an error '!022 Bad Message'.
Command: M,(PIN),(message)
*M,(PIN),(message)
Reply: <none>
Command: M,?
Reply: <message>(CR)
Command: *M,?
Reply: Message =<message>(CR)
21 K0473 Issue 5 - [EN] English
Page 32
O - Communication Settings
This command sets the communications for the transducer. The settings can be either interactive or non-interactive.
Where:
Baud rate: number between 0 and 6 or an actual rate.
Parity: set with a single letter:
I = ignore, N = None, O = Odd, E = Even Character length: 7 or 8 Stop bits: 1 or 2 Software handshaking:Y or N (Yes or No) Termination characters:1 or 2
(1 = transmitted lines must end in CR, received lines end in either CR or LF). (2 = transmitted lines must end in CRLF, received lines end in CRCR or CRLF or LFLF or LFCR).
Command: O, PIN,n <CR>
where: O, (PIN), (Baud Rate), (Parity), (Character Length), (Stop Bits),
(Software Handshaking), (Number of Termination Characters)
Reply: <none>
Command: *O, PIN,n <CR>
where: O, (PIN), (Baud Rate), (Parity), (Character Length), (Stop Bits),
(Software Handshaking), (Number of Termination Characters)
Reply: <none>
Command: O,?
Reply: <Baud Rate>, <Parity -NIEO>, <Character Length>, <Stop Bits>,
<Software Handshaking - Y/N>, <Number of Termination Characters> (CR)
Command: *O,?
Reply: Baud Rate = <Baud Rate>
Parity = <Parity -Ignore/None/Odd/Even (CR) Data Bits = <Char Length> (CR) Stop Bits = <Stop Bits> (CR) Handshake = <Software Handshaking - Yes/No> (CR) Term Chars = <Number of Termination Characters> (CR)
[EN] English - K0473 Issue 5 22
Page 33
Command: *O
Reply: (Interactive)
Enter PIN - >(Enter PIN) 0 - 19200 bps. (CR) 1 - 9600 bps. (CR)2 - 4800 bps. (CR) 3 - 2400 bps. (CR) 4 - 1200 bps. (CR)5 - 600 bps. (CR) 6 - 300 bps. (CR) Select Baud rate (0 - 6) = >(enter selection) Enter parity ([N]one/[I]gnore/[E]ven/[O]dd) = >(Enter parity) Enter number of data bits (7/8) = >(Enter number of data bits) Enter number of stop bits (1/2) = >(Enter number of stop bits) Software Handshaking? (Y/N) = >(Enter Y or N) Number of termination characters (1 = CR, 2 = CRLF) = >(Enter
number of termination characters) Accept these values? (Y/N) = >(Enter Y or N)
If Y entered - EEPROM updated. (CR)
If N entered - Current settings unchanged. (CR)
Result: These settings take effect after switching off then on the power
supply for the transducer.
P - Change PIN This command sets the PIN of the transducer that must be used to
access the PIN protected settings. To change this setting the current PIN must be used and can be integers between 0 and 999.
Command: P, (current PIN), (new PIN)
Where: PIN’ is the PIN of the transducer.
‘NewPIN’ is the replacement PIN.
Reply:
Result:
<none>
The PIN of the transducer changes to new PIN.
Command: *P, (current PIN), (new PIN)
Where:
Reply: <none>
Result: The PIN of the transducer changes to new PIN.
‘current PIN’ is the PIN of the transducer. ‘new PIN’ is the replacement PIN.
Command: P,?
Reply: (Y if PIN set, N if PIN not set) (CR)
23 K0473 Issue 5 - [EN] English
Page 34
S - Set Offset
Command: *P,?
Reply: Pin Set = (Yes if PIN set, No if PIN not set) (CR)
This command adds an offset value to the output of the transducer.
Command: S, (PIN), (pressure)
Reply:
Result: The transducer stores the value set with the S command in
<none>
non-volatile memory.
Command: S,?
Reply:
<Offset>,<Setpoint>(CR)
Command: *S,?
Reply:
Offset = <Offset><Units>(CR) Set at = <Setpoint><Units>(CR)
The transducer stores the value entered as <Setpoint>. The H command also uses this set-point value. To remove the set-point send:
S,(PIN),(X)
This sets the stored set-point value to zero.
[EN] English - K0473 Issue 5 24
Page 35
8.6 Factory Set Commands ­Query Only
The following command codes can only be accessed using a factory PIN. These command codes can be queried by the user. The transducer sends the data/settings stored in non-volatile memory for the queried commands code.
E - Crystal Reference Frequency
Command: n/a
Query: E,?
Query: *E,?
L - Load Calibration Coefficients
Command: n/a
Query: L,?
This direct mode only command sets the measured frequency of the crystal oscillator in non-volatile memory.
Reply: <Reference Frequency>(CR)
Reply: Reference Frequency = <Reference Frequency>kHz(CR)
This direct mode only command sets the calibration coefficients in non-volatile memory.
Reply: <Coefficient0,0>,<Coefficient 0,1>,<Coefficient0,2>,<Coefficient
0,3>,<Coefficient 0,4>,<Coefficient 1,0>, ... ,<Coefficient 5,3>,<Coefficient 5,4>,<Frequency Offset>,<Diode Offset>,<Calibration Date>(CR)
Reply: K[0][0] = <Coefficient 0,0> K[0][1] = <Coefficient 0,1>
K[0][2] = <Coefficient 0,2>(CR) K[0][3] = <Coefficient 0,3>
K[0][4] = <Coefficient 0,4> (CR) K[1][0] = <Coefficient 1,0> ...... ...
K[5][4] = <Coefficient 5,4>(CR) (CR)
Frequency Offset (X) = <Frequency Offset> Diode Offset (Y)= <Diode Offset>(CR)
Serial Number = <Serial Number> Calibration Date= <Calibration Date>(CR)
Curve Fit = <X order> x <Y order>(CR) (CR) Type <CR> to continue(CR) (Waits 20 seconds or for CR)
25 K0473 Issue 5 - [EN] English
Page 36
T - Analogue/Digital Converter Set-up
Command : n/a
This direct mode only command sets the internal analogue to digital converter used to measure the ambient temperature value in non-volatile memory.
Query: T,?
Reply: <Diode Calibration>(CR)
Query: *T,?
Reply: Diode Cal = <Diode Calibration>(CR)
V - Transducer Type
and Settings
Command: n/a
Query: V,?
Reply: <Type>,<Serial Number>,<Sensor Serial Number>,
Query: *V,?
Reply:
This command sets the transducer type, pressure range and serial number in non-volatile memory.
Transducer data
Type: 5 character text string Serial Number: ‘AA/1/2’ format Sensor Serial Number: serial number is a 7 digit integer Style: G for Gauge, A for Absolute Pressure Unit Number: 0 to 24, see the U command Minimum Pressure: real numbers in pressure units Maximum Pressure: real numbers in pressure units
Direct mode only
<Style>,<Pressure Unit Number>,<Minimum Pressure>, <Maximum Pressure>(CR)
Type = <Type>(CR) Serial Number = <Serial Number>(CR) Sensor SN = <Sensor Serial Number>(CR) Style = <Style>(CR) Minimum Pressure = <Minimum Pressure> <Pressure Units>(CR) Maximum Pressure = <Maximum Pressure> <Pressure Units>(CR)
[EN] English - K0473 Issue 5 26
Page 37
W - Write to EEPROM This command sets the non-volatile memory, as listed below:
Address Meaning
-1 Wipe the first bank of the EEPROM.
-2 Wipe the second bank of the EEPROM.
-3 Wipe the both banks of the EEPROM.
-4 Wipe the calibration coefficients in both banks.
-5 Copy from bank 2 to bank 1.
-6 Copy from bank 1 to bank 2.
Command: n/a
Query: W,?
Reply:
<Byte 0>,<Byte 1>,<Byte 2>,<Byte 3>,<Byte 4>, <Byte5>,<Byte 6>, ... ,<Byte 510>,<Byte 511>(CR)
Query: *W,?
Reply:
000=> <Byte 0>,<Byte 1>, ... ,<Byte 15>(CR) 010=> <Byte 16>,<Byte 17>, ... ,<Byte 31>(CR) ... 0F0=> <Byte 240>,<Byte 241>, ... ,<Byte 255>(CR) (CR) Send <CR> to continue(CR) (Waits 20 seconds or for CR) 100=> <Byte 256>,<Byte 257>, ... ,<Byte 271>(CR) ... 1F0=> <Byte 496>,<Byte 497>, ... ,<Byte 511>(CR) Send <CR> to continue(CR) (Waits 20 seconds or for CR)
27 K0473 Issue 5 - [EN] English
Page 38
Table 1, Fault Finding
Error Message Probable cause Action
!001 Buf Overflow
!002 EEPROM Error Failure of non-volatile memory. Return to GE for repair or replacement. !004 Bad
Command
!005 Bad Char Illegal character sent in a command string.
!006 Bad Param(s)
!008 Bad Format
!009 Miss’g Param
!010 Invalid PIN Incorrect PIN for the command sent. Check command string sent. !011 Bad Value Out of range value sent to the transducer. Send in range value.
!012 Bad BUS Cmd
!013 Cal Error
!014 Press Range
!015 Under Press Pressure applied below calibrated limit.
!016 Over Press Pressure applied above calibrated limit.
!017 Bad Global
!018 Bad Response
!019 Timed Out
!021 Bad Checksum
!022 Bad Message
!023 Bad Cal Pres
More than 30 characters sent without an end of the line character.
Command character not recognised.
Invalid parameters sent in command string ­out of range, too long, wrong type etc.
Invalid formatting character sent in command string or command is interactive.
At least one of the required parameters missing from command string.
Command sent to transducer in Network mode.
Corrupt calibration data detected, transducer pressure measurement unreliable.
Corrupt calibration data detected, pressure range not known, transducer pressure measurement unreliable.
Command sent to the transducer as a global 0: command not valid as a global command in network mode.
Invalid value entered in response to a prompt normally happens with inactive commands.
Key not pressed in the specified time period (usually 20 seconds) normally happens with inactive commands.
Error occurred in memory check sum.
Erroneous character detected in the middle of a message.
Divide by zero error occurred in calibration calculation.
Send 30 delete characters followed by an end of line to clear the buffer.
Check command string sent. Check the serial interface connection.
Check command string sent is valid. Check the serial interface connection.
Check command string sent.
Check command string sent requires a valid formatting character.
Check command string sent.
Send this command with transducer in Direct mode.
Return to GE for recalibration.
Return to GE for recalibration.
Check for system fault where transducer is installed.
Check for system fault where transducer is installed.
Send the command to each transducer in turn, using correct address in place of global 0.
Repeat command with valid values in response to each prompt.
Repeat command, entering response in less than 20 seconds between each character and send carriage return within the specified time.
Power cycle sensor, if fault remains return to GE for repair or replacement.
Check and re-send message.
Check data and repeat calibration.
[EN] English - K0473 Issue 5 28
Page 39
9 Appendix A:
RPS 8000 Series EEPROM format
The EEPROM device in the RPS 8000 Series is a Microchip 11LC040. The datasheet can be found at:
http://www.microchip.com/wwwproducts/Devices.aspx?dDocName=en535102
Only the first 511 bytes of memory are defined as follows:
Table 2: EEPROM Data format
Address
(Dec) (Hex)
0 0000 1
2 0002 4
8 0008 16
40 0028 2
44 002C 1
45 002D 1
46 002E 1
52 0034 4
56 0038 4
64 0040 4
68 0044 4
72 0048 1
73 0049 1
80 0050 1
81 0051 1
128 0080 4
132 0084 4
136 0088 4
140 008C 4
144 0090 4
148 0094 4
152 0098 4
156 009C 4
160 00A0 4
164 00A4 4
168 00A8 4
172 00AC 4
Size
(Bytes)
Field Name
Data field format code 1 8 bit Integer
Serial Number 32-bit Integer
Product ID Code RPS 8000 ASCII Text
Transducer Type Identifier &h 1F40 16 bit Integer
Day of calibration - dd 8 bit Integer
Month of calibration - mm 8 bit Integer
Year of calibration - yy 8 bit Integer
Customer Offset term 0 32 bit IEEE
Customer Gain term 1 32 bit IEEE
Upper pressure range 32 bit IEEE
Lower pressure range 32 bit IEEE
Pressure units code 8 bit Integer
Sensor type Abs/Gauge 0 / 1 8 bit Integer
Number of pressure coefficients 8 bit Integer
Number of temperature coefficients 8 bit Integer
X - Frequency datum 32 bit IEEE
Y - Temperature datum 32 bit IEEE
K00 32 bit IEEE
K01 32 bit IEEE
K02 32 bit IEEE
K03 32 bit IEEE
K04 32 bit IEEE
K10 32 bit IEEE
K11 32 bit IEEE
K12 32 bit IEEE
K13 32 bit IEEE
K14 32 bit IEEE
Contents
(Default)
Format
29 K0473 Issue 5 - [EN] English
Page 40
Address
(Dec) (Hex)
176 00B0 4
180 00B4 4
184 00B8 4
188 00BC 4
192 00C0 4
196 00C4 4
200 00C8 4
204 00CC 4
208 00D0 4
212 00D4 4
216 00D8 4
220 00DC 4
224 00E0 4
228 00E4 4
232 00E8 4
236 00EC 4
240 00F0 4
244 00F4 4
248 00F8 4
252 00FC 4
510 01FE 2
Size
(Bytes)
Field Name
K20 32 bit IEEE
K21 32 bit IEEE
K22 32 bit IEEE
K23 32 bit IEEE
K24 32 bit IEEE
K30 32 bit IEEE
K31 32 bit IEEE
K32 32 bit IEEE
K33 32 bit IEEE
K34 32 bit IEEE
K40 32 bit IEEE
K41 32 bit IEEE
K42 32 bit IEEE
K43 32 bit IEEE
K44 32 bit IEEE
K50 32 bit IEEE
K51 32 bit IEEE
K52 32 bit IEEE
K53 32 bit IEEE
K54 32 bit IEEE
Check sum 16 bit Integer
Contents
(Default)
Format
Where the entries are in the following number formats: 8 bit Integer signed integer 16 bit Integer signed integer, MSB in lowest address 32 bit Integer signed integer, MSB in lowest address ASCII Text series of ASCII characters 32 bit IEEE 4 byte IEEE-754 floating point number, lowest address holds sign and exponent,
highest address holds LSB of mantissa. All un-used locations below 01FE Hex are filled with zero. All un-used coefficients are populated with the value zero. The check sum is calculated so that addition of all locations between 0 and 01FF Hex will sum to
1234 Hex. The list of available values for 0048 Hex, pressure unit code are shown in Table 3.
[EN] English - K0473 Issue 5 30
Page 41
Table 3 RPS Pressure unit codes
Pressure
Code
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Units
Not defined
mBar
Bar
hPa
KPa
MPa
psi
mm H2O
in H2O
ft H2O
m H2O
mm Hg
in Hg
2
Kgf / cm
Atm
31 K0473 Issue 5 - [EN] English
Page 42
[EN] English - K0473 Issue 5 32
Loading...