Drager Konica Minolta Hill-Rom Air-Shields JM-103 User manual

Page 1
USER MANUAL
Jaundice Meter (JM-103)
From Konica Minolta and Hill-Rom Air-Shields
Page 2
Jaundi ce Meter (Mode l JM - 1 03) User Manual (usr 07 0r b )
Page 3

Table of Contents

Symbol Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Section 2: Introduction, Features, and Specifications
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Measuring Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Explanation of the Test . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 3
Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 6
Controls, Indica tors, and C on n e ctions . . . . . . . . . . . . . . . .2 - 6
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 8
Standard Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 9
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -10
Standard Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 10
Regulations, Standards, and Codes . . . . . . . . . . . . . . . . . .2 - 11
Section 3: Precautions and Safety Tips
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Safety Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 3
Warning and Caution Labels . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 4
Section 4: Installation and Assembly
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1
Charging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1
Selecting the Unit of Measurement . . . . . . . . . . . . . . . . . . .4 - 3
Operational Chec kout of the Jaundice Meter . . . . . . . . . . .4 - 4
Section 5: Instructions f or Use
Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
Taking Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
Setting the Number of Average Measurements . . . . . . . . .5 - 4
Taking Average Measurements. . . . . . . . . . . . . . . . . . . . . .5 - 5
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page i
Page 4
Section 6: Cleaning, Maintenance, and Re placement Parts
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Steam Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Cleaning Hard to Clean Spots . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Section 7: Troubleshooting
Service Calls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
Section 8: Storage and Handling
Storage and Handling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 - 1
Section 9: Warranty
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1
Appendix A: Clinical Performance Summary
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 1
Selection Criteria . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 1
Performance Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 2
Error Plots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 12
Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 16
Appendix B: Medical and Scientific References on Transcutaneous Bilirubinom etr y (1979 - 1997)
Page ii Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 5
Section 1

Symbol Definition

and Intended Use
Symbol Definition
This manual contains different typefaces and icons des igned to improve readabili ty and i ncrease unde rsta nding of it s cont ent. No te the foll owing examples:
• Standard text—used for regular information.
Boldface text—emphasize s a word or phrase.
NOTE:—sets apart s p ec ial info r mation o r import an t instruc tion
clarification.
• The symbol below highlights a WARNING or CAUTION:
Warning and Caution
A WARNIN G identif ies situations o r actions th at may affect
patient or user safety . Disregarding a warning could result in patient or user injury.
A CAUTION points out s pecial procedures or precautions that
personnel must follow to avoid equipment damage.
• The symbol below highlights a type BF applied part:
T ype BF Appli ed Part
The instrument provi des a specified degree of protec tion
against electric shock, particularly the lea kage current and reliability of the protective ground connection with an F-type applied part . An F-type applied part indic ates an applied part isolated fr om all other parts of the instrument to such a degree that the p at ient leak ag e cur r e n t al lo w able in a sin g l e- fa u l t condition is no t exce eded whe n a volta ge equa l t o 1.1 ti mes the highest-rate d main s volta ge is applie d be tween t he appli ed part and ground.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 1 - 1
Page 6
• The symbol below highlights an ELECTRICAL SHOCK HAZARD WARNING:
Electrical Shock Hazard Warning
• The symbol below indicates INPUT RATING:
Input Rating Symbol
• The symbol below ind icates that the product uses a RECHARGEABLE BATTERY:
Rechargeable Battery Symbol
• The symbol below ind icates RESET:
RESET Button Symbol
• The symbol below, when applied to the device, indicates:
ATTENTION: Consult Accompanying Documents
Page 1 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 7

Intended Use

Indications for the Use of the Jaundice Meter (JM-103)
The Jaundice Meter (JM-103) is a non-invasive transcutaneous bilirubi nom eter. It measures yellowness of subcutaneous tissue in newborn infants and displays a measured value which has been shown to correlate with serum bilirubin. The device is for use in the hospital to assist clinicians in monitoring the status of newborn infants for the development of hyperbilirubinemia. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical s igns and laboratory measurements. Newborn infants whose Jaundice Meter (JM-103) test results are indicative of hyperbilirubinemia are evaluated by their doctor(s) for appropriate patient management. Bilirubin levels should be confi r me d by other methods, such as se rum bilirubin, prior to treatment determinations.

Intended Use of the User Manual

This manual provides instructions for installation, use, operator maintenance , and troubleshooting of the Jau ndice Meter (JM-103) from Konica Minolta and Hill-Rom Air-S hields. Konica Minolta/Hill-Rom Air-Shields cannot be responsi ble for the performance of the Jaun dice Meter if the user does not oper ate the unit in accordance with the instr uc tions, f ails to fol l ow m ainten ance rec ommen dation s , or makes any repairs with unauthorized compon ents. Only qualified se rvice personnel shoul d perform repair . Service information is available through your local distributor or Konica Minolta/Hill-Rom Air-Shields.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 1 - 3
Page 8
NOTES:
Page 1 - 4 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 9
Section 2
Introduction, Features,
and Specifications

Introduction

To prevent kernict erus in newborn infants, it is very im portant to detect jaundice in i ts earl y st ages. The Jaundice Meter (JM-103) is a non­invasive tra nscut aneous bi li rubi nometer. Th is hand-he ld d evice allows a quick, non-invasive estimate of bilirubin concent r ation, to be used as an aid for the managemen t of jaundice in newborn infant s. The measur ements are taken au to matically when you place the in strument ’s measuring probe against the forehea d or sternum of the inf ant and press it gently; the measured value is then displayed.
The Jaundice Meter (JM-103) measures the yellowness of the subcutaneous tissue in infants through the detection of the optical density di fference at two wavelength s. It displays a measured value that has been sho wn to correla te wit h ser um bili rubi n. The devic e is inten ded for use in the h ospital to assist clinicians in monit or ing the sta tus of infants for the deve lopment of hyperbil irubinemia. The device is not intended a s a standalone for diagnosis of hyperbilirubinemia. It is to be used in c onjunction with other clinical signs and laboratory measurements. Infants whose Jaundice Meter (JM-103) test results are indicative of hyperbiliru binemia are evalu ate d by their doctor(s) for appropriate patient manag em ent. Bilirubin levels should be confirmed by other methods (e.g. serum bilirubin) prior to treatment determinations.
SPECIFICATIONS

Measuring Point

Measurements must be taken only on the infant’s forehead or sternum where a suf fici ent am ount of b lood is ci rcula ted. A po ssibil it y exis ts th at the bilirubin in the subcutaneous tissue may measure low for are as with minimal blood flow or areas in which the subcutaneous tissue is subject to keratiniz ation.
The measuring points for this device include the forehead and the sternum of the infant. Although c orrelation with serum bilirubin was observed for both forehe ad and ster num measurements, the clinical studies performed with the Jaundice Meter (JM-103) show consis tently better results with measurement s ta ken at the sternum versus the forehead.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 2 - 1
Page 10

Precocious Jaundice

Do not use this device on infants with precocious jaundice. If the re is a possibility that the infant is suffering from precocious jaundice, as a result of an incompatible blood type or hemolytic jaundice, it is recommended that the total serum bilirubin be measured.
Phototherapy
WARNING:
Do not use the Jaundice Meter after initiation of phototherapy or after an exchange transfusion. Patient injury could occur.
Page 2 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 11

Explanation of the Test

Measurin g Principle
The Jaundice Meter determines the yellowness of an infant’s subcutaneous tissue by measuring the difference in the optical densities for light in the blue (450 nm) and green (550 nm) wavele ngth regions. The measuring probe has two optical paths. This method allows for a more precise measurement of yellowness in an infant’s subcutaneous tissue by minimiz ing the infl uence s of the melan in pigment and th e skin maturity.
SPECIFICATIONS
When the measuring probe is pressed against the forehead or sternum of the infant, the built-in xenon lamp flashes. The light fr om th e xenon lamp passes thr ough the glass fiber and illuminates the skin. The light scatters an d is ab so r b ed in th e sk in an d su b cu taneous tissue rep e at ed l y, and then finall y returns to the sensor side of the glass fiber. Of the light that returns, the part scattered from the shallow areas of the subcutaneous tissue passes through the inner core, or short-optical path, of the fiber. The part scattered from the deep areas of the subcutaneous tissue passes through the outer core, or long-optical pa th, and then reaches it s corresponding photodiode.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 2 - 3
Page 12
By calculating the difference in the optical densities, the pa rts that are common to the epidermis and dermis are d educted, and as a re sult, the differe nce in the optical densities between the two wavelength re gions can be obtained for the subcutaneous tissue only. Since the optical density difference shows a linear correlation with the total serum bilirubin concentration, it is converted to the estimated bilirubin concentration and is indicated digitally.
The Jaundice Meter (JM-103) de vice software uses a correlation coeff icient to convert the measurement difference from the dual optical path to an estimated bilirubin concentration. The calculation formula used includes the correlation coefficients α and γ. These coefficients were determined in pre-clinical testing. The equation used is as follows:
J
= α(L-S) + γ
sample
Where L and S are the long and short optical path measurements.

Use of the Device

Patient Population
The Jaundice Meter (JM-103) is indicated for use in neonatal patients born >35 weeks gestati on who have not undergone transfusion or photother apy treatment.
Page 2 - 4 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 13
Averaging of Measurem ents
There was no significant difference between the averaged measurements and the single measurement approaches in the largest study for sternum measurements. The mean of three measure ments showed the highes t degree of correlation (r=0.965), however, the difference was minimal with a single measurement (r=0.963).
The advantages of using average measurements should be assessed at each facility.
Action Levels
Each facility should deter mi ne their own action levels based on studies of performance of the device on their population. Appropriate a ction levels may vary depending on performance of the device, such as precision or correlation with serum bilirubin, in the hands of the user. Some factors that could affect performance of the device or appropriate action levels include skin color, age, or measurement site. Careful selection of ac tion levels should be made so that false negat ives do not prevent appropriate follow up measures.
Calibration
There is no user calibration of this device. The system doe s include a checker that me asure s the int ensi ty of l ight fro m the de vice to ensure t he light output is within ran ge.
Processing of Measured Values
The Jaundice Meter (JM-103) determines the yellowne ss of the subcutaneous tissue by measuring the difference in the optical densities for light in the blue and green wavelengt h regions. The optical density diffe rence has been shown to have a linear correlation with serum bilirubi n concentration. T he device computes an estimated bilirubin concentra tion based on this linear correlation and provides the val ue on the display. The correla tion coefficient used in this computation was derived from clinical data.
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 2 - 5
Page 14

Features

Controls, Indicators, and Connections

Controls, Indicators, and Connect ions
Name Function
A Power switch Turns the Jaundice Meter on and off.
When used with the Reset button, the device switches to Check Mode and changes the unit of measurement.
B Ready lamp Illuminates to indicate that the J aundice Meter
is ready for the next measurement. C Display Displays the measured value. D Measuring prob e Takes the measurement when pres s ed against
the measuring point.
Page 2 - 6 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 15
Name Function
E Charger section Connect the cha rger unit to the charger sec-
tion.
F Reset button Deletes the currently displayed measured
value and prepares for the next measurement. When used with t he Power switch, the device switches to Chec k Mo de and chang es the un it
of measurement. G Str ap attachment area Is where the s trap attaches. H Checker cover Open this c hecke r cove r to c heck the Ja undice
Meter. I Charger lamp Illuminates t o indicate that the Jaundice Met er
is chargi n g. J DC jack Connect the AC adapter to this jack. K Charger jack Connect the main body to this jack. L Checker Checks for the intensity of light output by tak-
ing measurements in Check Mode. M Standard checker
For reference.
values N DC plug Connect the charger’s DC jack to this. O DC Plug (interna-
Connect the charger’s DC jack to this.
tional)
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 2 - 7
Page 16

Display

Name Function
A AVG Illuminates duri ng averaging measurement . B Optical path indicator
(•)
C Value Displays the measured value.
D Unit of measur em ent Displays the unit of measurement in either
E Battery indicator When the battery power is low, the battery
When performing inspections with the checker, • illuminates when the L-value appears and extingui shes when the S-value appears.
NOTE: When the measured value is greater than 20 mg/dl, the displ ay shows “---” an d the physician should be contacted.
milligrams per deciliter (mg/dL) or mi cro­moles of solute per liter (µmol/L)
indicator blin ks. Ch ar ge the bat tery a s soo n as possible (see “Charging the Battery” on page 4-1). If only the battery indicato r lights, the battery has run out. Go to “Charging the Battery” on page 4-1. If the power is on and the displ ay is blank, the battery is completely exhausted. Go to “Charging the Battery” on page 4-1.
Page 2 - 8 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 17

Standard Features

• Jaundice Meter (JM-103)
• Charger unit (Model JM-A30) with a checker
• AC adapter (Model JM-A31)
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 2 - 9
Page 18

Specificat ions

Standard Features
Feature Dimension
Model name JM-103 Measuring method Determines the yellowness of the
subcutaneous tissue by using two optical paths to measure the optical density difference at two wave­lengths
Measurement range 0.0 mg/dL to 20 mg/dL or 0 µmol/L
to 340 µmol/L
RMSE
Light source Pulse xen on arc lamp Light source life 150000 measurements Sensors Silicon phot odiodes Power source Special Ni-MH ba ttery Protectio n type and level Internally-powered instrument, BF-
Minimum number of measur ements when fully charged
Operating temperature range 10°C (50°F) to 40°C (104°F) Operating rela tive humidity range 30% to 95% Storage temperature range -10°C (14°F) to 50°C (122°F) Storage rela tive humidity range 30% to 95% Dimensions 48 mm (1.9") wide x 15.5 cm (6.1")
Weight, including battery 150 g (5.3 oz) AC adapter input for North America 120V, 50/60 Hz, 10 W AC adapter input for
International use AC adapter output 9V, 500 mA
± 1.5 mg/dL or ± 25.5 µmol/L
type 400 single measurements
high x 33 mm (1.3") deep
200V-240V, 50/60 Hz, 12.5VA
* This accuracy is based on the aggregate results from both clinical study groups (n=613). Refer to Appendix A for a detai led descri ption of results by clinical site, measurement loca tion, and patient demographics.
Page 2 - 10 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 19

Regulations, St a ndar ds, and Codes

In No rth Am e r ica With respect to electrical shock, fire, and mechanical hazards only, this instrument complies with UL 2601-1 and CAN/CSA C22.2 No. 601.1.
In Europe, this instr ument complies with EN60 601-1, EN60601-1-2, and EN ISO13485, and EN ISO14971.
SPECIFICATIONS
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 2 - 11
Page 20
NOTES:
Page 2 - 12 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 21
Section 3
Precautions and Safety Tips
Precautions
WARNING:
Do not use the instrument in areas where flammable or combustible gases, such as anesthetic gases, are present. Doing so could result in a fire. Personal injury or equipment damage could occur.
WARNING:
If the instrument, the charger unit, or the AC adapter are damaged, or if smoke or an odd smell occurs, do not use the instrument, the charger unit, or the AC adapter. In such situations, immediately turn off the instrument, unplug the AC adapter from its power source, and contact the nearest authorized service facility. Failure to do so could resul t in fire, personal injury, or equipment damage.
SHOCK HAZARD:
Always plug the instrument into an AC outlet of the correctly rated voltage and frequency. Fai lure to do so could result in fire, personal injury, or equipment damage.
SHOCK HAZARD:
Do not disassemble or modify the instrument, the charger unit, or the AC adapter. Fire, personal injury, or equipment damage could occur.
CAUTION:
Do not use a mobile telephone when using the Jaundice Meter. A measurement error could occur.
CAUTION:
The Jaundice Meter interferes with Magnetic Resonance Imaging (MRI) procedures. Inaccurate readings could occur.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 3 - 1
Page 22
CAUTION:
Do not place the instrument on an unstable or sloping surface. The instrument or charger unit could drop or overturn. Equipment damage could occur.
CAUTION:
The Jaundice Meter is a precision instrument. When using it, do not drop it, expose it to shocks or strong vibrations, or place heavy objects on it. Equipment damage could occur.
CAUTION:
Do not allow blood or other liquids to come in contact with the instrument. Should blood or other liquids come in contact with the instrument, immediately clean the instrument (see “Cleaning” on page 6-1). Failure to do so could result in equipment damage.
CAUTION:
Do not use the instrument in direct sunlight. Equipment damage could occur.
CAUTION:
The instrument has a built-in, non-replaceable battery. Do not disassemble the instrument to replace the battery. To replace the battery , contact your dealer or authorized Hill-Rom service center. Failure to do so could result in equipment damage.
Page 3 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 23

Safety Tips

WARNING:
This instrument emits intense light to take its measurements. Take measurements only at the forehead or sternum, and do not press the measuring probe when it is directed toward the eyes. Damage to the eyes could occur.
WARNING:
Before use, clean the measuring probe by wiping it with medicinal alcohol. Failure to do so could result in the spread of infection or infant inju r y.
WARNING:
The charger unit (JM-A30) and the AC adapter (JM-A31) are solely designed for use with the Jaundice Meter (JM-103). Use them only when chargin g the instrumen t. Using them to charge other equipment could result in personal injury or equipment damage.
WARNING:
Only facility-authorized personnel should troubleshoot the Jaundice Meter. Troubl es hooting by unaut horized personnel could result in personal injury or equipment damage.
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could result in personal injury or equipment damage.
WARNING:
Do not use the Jaundice Meter after initiation of phototherapy or after an exchange transfusion. Patient injury could occur.
SHOCK HAZARD:
Do not plug or unplug the AC power cord’s plug with wet hands. Personal injury or equipment damage could occur.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 3 - 3
Page 24
SHOCK HAZARD:
Before cleaning, maintenance, or parts replacement, unplug the charger unit from its power source. Failure to do so could result in personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for liquid pooling. Personal injury or equipment damage could occur.
CAUTION:
Do not use harsh cleansers/detergents, such as scouring pads and heavy duty grease removers, or solvents, such as toluene, xylene, and acetone. Equipment damage could occur.

Warning and Caution Labels

Page 3 - 4 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 25
Section 4
Installation and Assembly

Installation

Before using th e ins trument, charge and inspect the instrument.

Charging the Battery

When using the instrument for the first time, ensure that it is fully charged. To maintain a full charge at all times, place the instrument on the charger unit when it is not being used for measurements. When the battery power is low, the Battery display blinks.
If the Jaundice Met er is left uncharged for a long period of time, the power of the battery dimini shes; e nsure t hat it i s charge d pri or to use . To charge the Jaundice Meter, perform the following:
WARNING:
The charger unit (Model JM-A30) and the AC adapter (Model JM­A31) are solely designed for use with the Jaundice Meter (JM-
103). Use them only when charging the instrument. Using them to charge other equipment could result in personal injury or equipment damage.
1. Plug the AC adapter into the DC jack
of the c harger u n it. Use only the charger unit and AC adapter supplied with the Jaundi ce Meter.
NOTE:
The shape of the AC adapter varies according to region.
NOTE:
120V model shown.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 4 - 1
Page 26
SHOCK HAZARD:
Do not plug or unplug the AC power cord’s plug with wet hands. Personal injury or equipment damage could occur.
2. Plug the AC ad apter’s plug into an AC out let. Never do so with wet hands.
3. Place the Jaundic e Meter on the charger unit so that its displa y faces you. When the Jaundice Meter is set on the charger unit properly, the
Charger lamp lights up.
NOTE:
With a fu lly char ge d batte ry, approxi mately 400 measurements can be taken. W ith six hours worth of char ging, approximately 100 measurements can be taken.
4. Allow approximately 32 hours for charging to complete.
CAUTION:
The instrument has a built-in, non-replaceable battery. Do not disassemble the instrument to replace the battery. To replace the battery , contact your dealer or authorized Hill-Rom service center. Failure to do so could result in equipment damage.
5. To replac e the battery, contact your dealer or an authoriz ed service facility.
Page 4 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 27

Selecting the Unit of Measurement

1. Hold down the Reset button, and turn on the Power switch.Do not release the Reset button.
2. While continuing to press the Reset button, allow approximately 15 seconds for the unit of m easurement to switc h f ro m mg/dL to µmol/L, or vice versa.
3. Visually ensure that the unit’s display has changed.
4. Releas e th e Reset button. The Ready lamp lights up, indicating that the instrument is rea dy to take a measurement.
5. To change the unit of measurement once more, turn off the Power switch, and repeat step 1.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 4 - 3
Page 28

Operational Checkout of the Jaundice Meter

Do not touch the checker’s surface with your fing ers . If the checker gets dirty, wipe it with a soft cloth dampened with water, and then wipe it with a dry cloth.
Using the checker supplied with the charger unit, check the instrument to ascertain that the meter lig ht output is within range. Check the instrument once or more each da y it is us ed . To inspect the Jau n di c e Meter, perform the following:
1. Hold the Reset button down, and set the Power switch to the On position.
NOTE:
If the Reset button is held down for longer than 15 seconds, the unit of measurement switches.
2. After CHE appears on the display window, immediately release the Reset button. If the Reset button is held down for longer than 15 seconds, switch the un it of measure ment back to its previ ous setting (see “Selecting the Unit of Measurement” on page 4-3).
3. Visually confirm that CHE appea rs in the display and that the Ready la mp illuminates.
4. Open the co v er of th e che ck er. Use only the checker su pplied with the Jaundice Meter (JM-103).
Page 4 - 4 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 29
5. Place the measuring probe perpendicul ar to the checker, and push it gently until a click soun ds .
6. If the measuring probe contacts the checker at an angle, place it perpendicular, and take the measur ement again.
NOTE:
The display interchanges between the L­value, the measured value of the long­optical path, and the S-value, the measured value of the short-optical path. When the L­value is displayed, “•” ap pears in the upper left-hand corner of the displa y.
7. Confirm the measured value. If both the L-va lu e an d th e S- v al ue are with in ± 1.0 of the reference values indicated on the che ck er, the values ar e acceptable.
8. If the measured value exceeds ± 1.0 of the reference value , per f orm the following:
a. Clean both the chec ker and the
measuring probe.
b. Place the measuring probe perpendicular to the checker, and
push it gently until a c lick sounds.
c. If the measured value still exceeds ± 1.0 of the reference value,
contact the nearest authorized service facility.
9. Close the cover of the checker.
10. Set the Power switch to the Off position.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 4 - 5
Page 30
NOTES:
Page 4 - 6 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 31

Instructio ns for Use

Taking Measurements

1. Remove the Jaundice Meter from the charger unit.
WARNING:
Before use, clean the measuring probe by wiping it with medicinal alcohol. Failure to do so could result in the spread of infection or infant injury.
2. Using medicinal alcohol, clean the measuring probe.
3. Set the Power switch to the On position. The measured value for a sin g le measurement, n-1, appears on the display.
Section 5
Instructions for Use
4. Visually check that the Ready lamp illuminates.
5. If the batt ery in dicat or bl ink s, c har ge t he battery (see “Charging the Battery” on page 4- 1).
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 5 - 1
Page 32
WARNING:
This instrument emits intense light to take its measurements. Take measurements only at the forehead or sternum, and do not press the measuring probe when it is directed to the eyes. Damage to the eyes could occur.
6. Perform the following: a. Place the measuring probe
vertically against the infant’s forehead or sternum. Avoid any bruises or disc olored areas of the skin.
b. Push the measuring probe gently
until a click sounds. The instrument’s xenon lamp flashes momentarily, and the mea sured value appear s on the display.
If the measured value is outside the measurement range of 20 mg/dL or 340 µmol/L, the display shows “--­” and user should cont ac t the physician.
NOTE:
If the instrument is not operate d for mor e than 60 seconds, the backlight on the display goes out.
7. To take another measurement, press
the Reset button, and continue from step 4.
Page 5 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 33
8. To stop measuring, perform the following:
a. Set the Power switch to the Off position. b. Using medicinal alcohol, clean the measuring probe . c. Place the Jaundice Meter on the charger unit. When the
Jaundice Meter is not in use, keep it on the charger unit.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 5 - 3
Page 34

Setting th e N u mber of Average Measurements

1. Set the Power switch to On or press the Reset button to prepare the instrument for measurement.
n-1, n-2, and so on (up to n-5)
will appear.
2. Press the Reset button for 5 s. The number of average measur e m en t s will switc h as f o ll o w s:
3. Release the Reset button when the required number of avera ge measurements is displayed.
If n-2 through n-5 is selected, AVG appears in the upper left corner of th e display.
The selected number of average measurements will be recorded.
Page 5 - 4 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 35

Taking Average Measurements

1. Set the numb er o f average measurements ne eded.
2. Ensur e th at the Ready lamp is illum i n at ed once n-5 appears.
NOTE: n-5 is use d in the steps below as an
example. The number of measurements you require and set is the number that should appear in the display.
NOTE:
Each measurement must be take n individually by the user. The first measurement is complete when the measuring probe is pressed agains t the patient and the unit clicks. The probe must then be li fted from the patient and reapplied for the total number of measurements selected (in this exampl e 5 measurements).
WARNING:
This instrument emits intense light to take its measurements. Take measurements only at the forehead or sternum, and do not press the measuring probe when it is directed toward the eyes. Damage to the eyes could occur.
NOTE:
It is recommended that all of the measurem ents taken for averagi ng be taken from the same measuring point, forehead or s ternum.
3. Place the measuring probe vertic ally on the measuring p oint, and then apply gentle pressure until the probe clicks.
The measurement will be taken, and the number of remaini ng measurements will be displayed.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 5 - 5
Page 36
4. While ensu r ing tha t th e Ready lamp is illuminated, repeat the
measuring until the number of remaining measurements is 0.
When the remaining number of measurements is completed, the average of the measured values appears in the display.
If the instrument is left without the preset number of measurements taken, the setting will be canceled without the measurement value being displayed. To set the average measurements again, press the Reset button and reset the number of average measurements needed.
To change the number of average measurements, refer to “Setting the Number of Average Measurements” on page 5-4.
Page 5 - 6 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 37
Section 6
Cleaning, Maintenance,
and Replacement Parts

Cleaning

WARNING:
Follow the product manufacturer’s instructions. Failure to do so could result in personal injury or equipment damage.
SHOCK HAZARD:
Before cleaning, maintenance, or parts replacement, unplug the charger unit from its power source. Failure to do so could result in personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for liquid pooling. Personal injury or equipment damage could occur.
CAUTION:
Do not use harsh cleansers/detergents, such as scouring pads and heavy duty grease removers, or solvents, such as toluene, xylene, and acetone. Equipment damage could occur.
If there is no visible soila ge with possible body fluids, we recommend that you clean the uni t with a mild detergent and warm water. If disinfec tion is desired, you may use a combination cleanser/disinfectant as explained in “Disinfecting” on page 6-2.

Steam Clea n ing

Do not use any steam cleaning device on the unit . Do not autoclave the unit. Excessive mo isture can damage mechanism s in this unit.

Cleaning Hard to Clean Spots

T o re move dif fic ult spots or s tai ns, we rec ommend th at you us e stan dard household clea nsers and a soft bristle brush. To loosen heavy, dried-on soil, you may first need to saturate the spot.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 6 - 1
Page 38

Disinfecting

When there is visible soilage and bet ween patients, we rec ommend that you disinfect the unit with a tuberc ulocidal disinfectant, such as Kleenaseptic B®. For customers in the US, the d is infectant should be register ed with the Environmental Protection Agency.
Dilute the d isinfectant according to the manufacturer’s instructions, if necessary.

Maintenance

There are no user serviceable parts and no maintenance or calibration required. Serv ic e is o nly re qu i r ed if the un i t ce ases to fun ct i on as intended or fail s the checker reading (see “Operational Checkout of the Jaundice Mete r” on page 4-4).

Replacement Parts

CAUTION:
The instrument has a built-in, non-replaceable battery. Do not disassemble the instrument to replace the battery. To replace the battery , contact your dealer or authorized Hill-Rom service center. Failure to do so could result in equipment damage.
There are no user-replaceable parts. To replace the battery, contact your dealer or an authorized service facility.
Page 6 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 39
Section 7
Troubleshooting

Service Calls

When you call Hill-Rom about your unit, be prepared to give the serial number from the product identification label. When you give the seria l number , the Hill -Rom re presenta ti ve can identi fy your unit and give you the information you need more quickly.

Error Messages

For warnings that may appear on the display window, refer to the table below.
Error Messages
Warning Cause Solution
Er1 The measured value is
abnormal. In the case of an averaging measurement, the measurement fluctuation is excessively large.
Er2
through
Er6
A measurement error may have occurred during an aver­aging me as u r em en t , or th e hardware is not functioning properly.
Place the measuring probe perpendicular to the infant’s forehead or sternum, a nd take the measurement again. If Er1 still appears, contact the nearest author i zed s er v i ce facility.
Set the Power switch to the Off position, and then return it to the On position. If the warning stil l appears, contact the near est a uth oriz ed se rvic e facility.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 7 - 1
Page 40

Troubleshooting

WARNING:
Only facility-authorized personnel should troublesh oot the Jaundice Meter. Troubleshooting by unauthorized personnel could result in personal injury or equipment damage.
If an abnormality occurs with the Jaundic e Met er, perform the following:
1. Refer to the table below, and take the necessary action giv en.
2. If the ab no r m a lity st il l ap p ears, se t the Power sw it ch to th e Off position, and then return it to the On position.
3. If the abnormality still continues, contact th e nea rest authorized service facility.

Troubleshooting

Symptom Possible Cause Action
The display is bla nk when the Power switch is in the On position.
The display suddenly goes blank during a measurement.
The Charger lamp does not illuminate when the Jaundice Meter is placed on the charger unit.
It is impossible to take measurements.
The batteries are exhausted.
The batteries are exhausted.
The charger unit and the AC ada p ter are not plugged into an AC outlet correctly.
The Jaundice Meter is not plac ed in th e charger unit corre ctly.
The Ready lamp is not illuminated.
The batteries are exhausted.
Charge the battery (see “Char ging the Battery” on page 4-1).
Charge the battery (see “Char ging the Battery” on page 4-1).
Plug the cha rger unit and the AC adapter into an appropriate power source correctly.
Place the Jaundice Meter perpendic ular to the measuring point in the charger unit.
Before taking a measur em en t , en s ur e that th e Ready lam p is illuminated.
Charge the battery (see “Char ging the Battery” on page 4-1).
Page 7 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 41
Section 8

Storage and Handling

Storage and Handling
When storing the instrument, pay attention to the following conditions:
• Store the instrument at a temperature range of -10°C (14°F) to 50°C
(122°F), and at a non-condensing relative humidity range of 30% to 95%.
• Keep the instrument dry.
• Do not store the inst rument in locations tha t may have an adverse
effect on its perform ance, such as locatio ns subject to the f o llowing:
High atmospheric pressure – Direct sunlight—do not store near windows. – Extreme dust—do n ot stor e in close ts or bi ns where dust or l int
can gath er. – Air having sali nity or sulphur content – Strong magnetism—do not store near MRI or other imaging
equipment, a nd do not store near operating rooms.
• Do not subject the instrument to vibration or impact.
• Do not s to re th e instrum e n t in lo cations where chem ic als are stor e d or where g as may be emit ted.
• To ensure no problems will exist the next tim e the main body and charger are used, thoroughly clean the main body and char ger, and then store them together.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 8 - 1
Page 42
NOTES:
Page 8 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 43
Section 9

Warranty

Warranty
HILL-ROM COMPANY, INC.
LIMITED WARRANTY
Hill-Rom Company, Inc. (Hill-Rom) has a long tradition of providing superior products and service to our customers. Our goal is “Total Customer Satisfaction”. In that spirit, Hill-Rom is proud to offer the following warranty.
GENERAL WARRANTY (APPLICABLE UNLESS A SPECIFIC WARRANTY IS LISTED)
Hill-Rom warrants to the original purchaser that its products and replacement parts shall be free from defects in material and workmanship for a period of one (1) year from date of delivery. Hill-Rom’s obligation under this warranty is expressly limited to supplying replacement parts and/or service for, or replacing, at its option, any product which is, in the sole discretion of Hill-Rom, found to be defective. In addition to the foregoing one year warranty, Hill-Rom warrants to the original purchaser that the frame and welds on its products will be free from structural defects for the life of the product. Any product upgrade or modification initiated by Hill-Rom does not affect the original product warranty.
SPECIFIC WARRANTIES
MAT TRESS WARRANTIES
Hill-Rom warrants to the original purchaser that its mattress product shall be free from defects in material and workmanship for a period of two (2) years from date of delivery. However, electro mechanical mattress components (compress ors, valves, pri nted circ uit boards, hoses , and couplers) are covered by the general one (1) year warranty.
EXPENDABLES WARRANTIES
A sixty (60) day limited warranty from date of delivery applies to expendable parts such as cushions, coverlets, software diskettes, locator badge batteries, dome light incandes cent bulbs, overhead fluorescent tubes, heating elements, temperature probes, filter sheets, and microspheres. This warranty is limited to replacement of the parts covered.
TO OBTAIN PARTS AND SERVICE
In the United States and Canada, call Hill-Rom Technical Support Department at (800) 523-5756, Monday through Friday. Outside the United States and Canada, call your authorized Hill-Rom Distributor. In order to expedite service, we re quest you furnish the following information: customer identificat ion number, product model number, serial number, and description of problem. A qualified specialist will provide, via telephone (United States and Canada), or F AX (Outside the United States and Canada), troubleshooting assistance for facility personnel and provide necessary parts to make repairs. If troubleshooting determines the need for on­site technical service, a qualified service representative will be dispatched. Replacement of non-technical items will be the responsibility of the customer. If requested by Hill-Rom, products or parts for which a warranty claim is made shall be returned prepaid to Hill-Rom’s factory.
OUT OF WARRANT Y EXCHANGE POLICY
After the expiration of the original warranty, upon request, Hill-Rom will ship as a replacement, components such as selected: motors and printed circuit boards, for like units returned to Hill-Rom by the original purchaser at a substantial savings. Please call Hill-Rom Technical Support Department for current pricing.
PARTS AV AILABILITY POLICY
Hill-Rom will offer parts for new and remanufactured products for ten (10) years from date of sale; for communi cations products for five (5) yea r s from date of sale. Note: Some original component parts and assemblies may not be available; functional equivalents may be substituted.
Jaundice Meter (Model JM-103) User Manual (usr070rb) Page 9 - 1
Page 44
THE FOREGOING WARRANT IES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES AND IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS OF PURPOSE. HILL-ROM’S OBLIGATION UNDER THESE WARRANTIES SHALL NOT INCLUDE ANY LIABILITY FOR LOSS OF PROFITS, DIRECT, INDIRECT OR CONSEQUENTIAL DAMAGES OR DELAYS. Some
states, provinces, or countries do not allow the exclusion or limitation of incidental or consequential damages, so the above exclusion or limitation may not apply. Any improper or negligent use, any alterations or repairs not in accordance with Hill-Rom’s manuals or performed by others in such manner as in Hill-Rom’s sole judgment affects the product materially and adversely, shall void these warranties. These warranties do not cover failures due to misuse, abuse, neglect, or lack of routine maintenance. No employee or representative of Hill-Rom is authorized to change these warranties in any way or grant any other warranty unless in writing and signed by a Hill-Rom officer. These warranties provide specific legal rights; but, there may be other available rights, which vary from state to state, province to province, or country to country. Revise d October 1, 2003
Hill-Rom Company, Inc., 1069 State Route 46 E, Batesville, IN 47006-9167
Page 9 - 2 Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 45
Appendix A
Clinical Performance Summary

Introduction

The Jaundice Meter (JM-103) has been the subject of three clinical studies. One study in Japan, the other two studies in the United States. The following information is a summary of results from the two US studies.
The purpose of the JM 103 is to measure the yellownes s of subcutaneous tissue in the neonate. This measurement can be correlated to serum bilirubin concentration and is therefore an aid in the management of neona tal jaundice. Since this device performs measurement through the use of light it is non-invasive and painless for the infa nt. The ob je ctive of th e clinica l stu dies was to confir m th a t th e device measurement, di splayed in units of estimated bilirubin concentration, does correlate with the serum bilirubin concentration sufficiently well to warrant its use as a screening tool.
The following is a su mmary of the study protoc ols:

Selectio n Criteri a

The patient selection crite ria used for the study included infants less than 30 days old and weighing grea ter than 1000 grams. The test was performed on i nfants who were determine d by their physician to requir e a serum bilirubin test. A tabulation of weight distribution is provided in grap h 2 2 Infant Weight Distribution. Error plot s by w e i ght are pr o v ided in graphs 23 Clini cal S tu dy Site A Bir thwei ght Error Plot, Forehead and 24 Clinical Study Site A Birthweight Error Plot, Sternum.

Demographics of Patient Population

All patie nts meet ing th e ab ov e cri te ria were inc lu ded in th e study. There was significant eff o rt to ensure that there was sufficient representation of all skin pigmentation to verify that the JM 103 could be used across all populations with consistent results. The demographics of the patient population included Caucasian, African-American, and other.
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 1
Page 46

Sample Size

The total number of infants in the sample population was 613.

Measurement Selection

One study followed a protocol where three measurements were take n, each measurement was recorded and then the three measurements we re averaged. The sec ond study took one measurement only. Estimated bilirubin meas urement s take n during th e stud ies range d from 1.1 t o 20.8 mg/dL.

Body Sites Tested

In both studies the meas urements were taken on the forehe ad and sternum each time the measurements were take n for a give n patient.

Number of Clinical Sites

There were two clinical sites that participated in the study. Clinical study site A with 513 patients studied. Clinical study site B with 100 patients studied.

Performance Data

The following data is provided to demonstrate the performance of the JM 103 in the clinica l use environment. This series of graphs show the correlatio n of the estima ted bilirub in concentr ation take n non-invasivel y with the JM 103 to the actual serum bilirubin concentration measured from a blood sample taken from the pa tient (TSB). The device operates over a range of 0.0 – 20.0 mg/dL (0 – 340 µmol/L). The data in clude graphs in the form of x- y plots where x is the total serum bilirubin concentration measured and y is the JM 103 estimated bilirubin measurement. Forehea d and s ternum were not combined because the data would be too difficult to distinguish. The serum bilirubin measurements were take n using direct spectrophotometry in Clini ca l Study A a nd specifically with the Bea kman-Colter Synchron LX-20by in Clinical Study B.
Page A - 2 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 47

Graph 1 – Clinical Study Site A All Patients, Forehead

Graph 2 – Clinical Study Site A All Patients, Sternum

Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 3
Page 48

Graph 3 – Clinical Study Site A African-American Patients, Forehead

Graph 4 – Clinical Study Site A African-American Patients, Sternum

Page A - 4 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 49

Graph 5 – Clinical Study Site A Caucasian Patient s, Forehead

Graph 6 – Clinical Study Site A Causasian Patients, Sternum

Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 5
Page 50

Graph 7 – Clinical Study Site A Other Patients, Forehead

Graph 8 – Clinical Study Site A Other Patients, Sternum

Page A - 6 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 51

Graph 9 – Clinical Study Site B All Patients, Forehead

Graph 10 – Clinical Study Si te B All Pati ents, Sternum

Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 7
Page 52

Graph 1 1– Clinical Study Site B Afri can-American Patients, Forehead

Graph 12 – Clinical Study Site B African-American Patients, Sternum

Page A - 8 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 53

Graph 13 – Clinical Study Site B Caucasian Patients, Forehead

Graph 14 – Cli nical Study Site B Causasian Patients, Sternum

Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 9
Page 54

Graph 15 – Clinical Study Site B Other Patients, Forehead

Graph 16 – Clinical Study Site B Other Pati ents, Sternum

Page A - 10 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 55
Graph 17 – Summary Table of Slope, Intercept, Standard Deviation, and
Correlation Coeffi cients for Each Graph
Study site / patient population
SITE A Forehead All (n=513) SITE A Sternum All (n=513) SITE A Forehead African American (n=65 ) SITE A Sternum African American (n=6 5 ) SITE A Forehead Caucasian (n=399) SITE A Sternum Caucasian (n=399) SITE A Forehead Other (n=49 ) SITE A Sternum Other (n=49 ) SITE B Forehead All (n=100) SITE B Sternum All (n=100) SITE B Forehead African American (n=48 ) SITE B Sternum African American (n=4 8 ) SITE B Forehead Caucasian (n=35) SITE B Sternum Caucasian (n=35) SITE B Forehead Other (17) SITE B Sternum Other (17)
Slope Intercept Correlation
Coefficient (r)
Standard Deviation (RM SE)
1.05 -0.35 0.914 1.29
1.07 -0.74 0.946 1.02
1.15 -0.5 0.908 1.59
1.11 -0.4 0.908 1.55
1.01 -0.1 0.916 1.2
1.04 -0.6 0.956 0.88
1.06 -0.5 0.941 1.04
1.10 -1.0 0.977 0.65
1.07 -0 .00 0.84 2.14
1.16 -0 .43 0.89 1.85
1.40 +0.46 0.84 2.27
1.21 -0 .17 0.89 1.9
1.10 -1 .04 0.87 1.72
1.22 -1 .69 0.88 1.81
1.03 -0 .56 0.94 1.49
1.03 0.65 0.97 0.94
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 11
Page 56
Graph 18 – Clinical Study Sit e A Error Plot, Forehead
Graph 19 – Clinical Study Site A Error Plot, Sternum
Page A - 12 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 57

Graph 20 – Clinical Study Sit e B Error Plot, Forehead

Graph 21 – Clinical Study Site B Error Plot, Sternum

Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 13
Page 58

Graph 22 – Infant Weight Dist ri bution

Infant Weight
(grams)
Clinic a l Study A Clinic a l Study B
Up to 999 6 2 1000 - 1499 15 0 1500 - 1999 23 13 2000 - 2499 37 23 2500 - 2999 68 19 3000 - and up 337 43 Unknown 27 0 Tota l 513 100

Graph 23 – Clinical Study Site A Birthwei ght Error Plot, Forehead

Page A - 14 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 59

Graph 24 – Clinical Study Site A Birthweight Error Plot, Sternum

Reproducibility

Reproducibility of the li ght output of the devi ce was tested daily using the checker. The checker determines the intensity of the light outpu t of the device. The accuracy of th e device is determ ine d by how well the detectors in the unit measure the returning light. The results of the device testing using the checker show that the device produced output within the required range over the course of both clinical studies.
Reproducibi lity testin g in the pati ent popul ation can be derived from the data taken in Clinical s tudy site A. The reproducibili ty da ta is based on 467 babies. Three indepe ndent measurements were taken at each sit e, forehead and sternum. Each measurement was recorded, the mean and standard deviation co mputed and recorded. The average standard deviation for forehe ad an d sternum measurements was 0.3 for both.
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page A - 15
Page 60
Table 1: Reproducibility Data
Mean
Estimated
Bilirubin
Measure
ment
Body Site
Deviation
Range
Deviation
Mean
Deviation
Median
Range
(mg/dL)
Sternum 0.0 - 2.2 0.3 0.3 0.8 - 18.5 Forehead 0.0 - 2.6 0.3 0.2 0.1 - 19.5

Conclusion

What the data presented sh ows is that the JM 103 estimated bilirubin concentration measurement correlates to the serum bilirubin measurements. This data supports the use of this non-invasive device along with other clinica l ind ic ato rs as a aid in the managem ent of jaundice in the neonat al pati ent popula tion.
Page A - 16 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 61
Appendix B
Medical and Scientific References on
Transcutaneous Bilirubinometry (1979 - 1997)
1. Amato M, et al;
Transc utaneo us bilirubin de termination : correlati on in white prema­ture infants weighing less than 1500 gm.
Schweiz Med Wochenschr, 1985 Jul 9.
2. Amato M, et al;
Assessment of neonatal jaundice in low birth weight infants com­paring transcutaneous, capillary and arterial bilirubin levels.
Eur J Pediatr, 1990 Nov.
3. Amato M
Transcutaneous, capil la ry, and arterial bilirubin levels.
J. Pediatr. 125(2), 332, 1994
4. Amit Y, et al;
Effect of skinf old th ic kness on transcutaneous bilirubin measure­ments.
Biol Neonate, 1993.
5. Arimichi J, et al;
Reliability and future application of a transcutaneous bilirubinome­ter.
Josanpu Zasshi, 1984 Apr.
6. Berget M, et al;
Transcut aneous bilirubinometry in newborn infants.
Tidsskr Nor Laegeforen, 1984 May 20
7. Bhat V, et al ;
Correlation of transcutaneous bilirubinome try with serum bilirubin in south Indian neonates.
Indian J Med Res, 1987 Jul.
8. Bhut ta ZA, et al;
Transcut aneous bilirubinometry in Pakistani newborns: a prelimi­nary report.
JPMA J Pak Med Assoc, 1991 Jul.
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page B - 1
Page 62
9. Boo Nem Yu n, et al;
Transcutaneous bilirubi nom et ry in Malay, Chinese and Indian term neonates.
Med J Malaysia, 1984 Mar.
10. Bourchier D, et al;
Transcut aneous bilirubinometry: 22 months experience at Waikato W omen's Hospital.
N Z Med J, 1987 Sep 23.
11 . B ro wn A; et al;
Transcutaneous bilirubinometry in infants: influence of race and phototherapy.
Pediatr Res, 1981 , 15:653
12. Brown LP, et al;
Tra n scutaneous bilirubinometer: intermeter reliab ility.
J Perinatol, 1990 Jun.
13. Brown LP, et al;
Transcutaneous bilirubinometer: an instrument for clinical research.
Nurs Res, 1990 Jul-A ug.
14. Brown LP, et al;
Incidence and patter n of jaun dic e in heal thy breast-fed infants dur­ing the first month of life.
Nurs Res, 1993 Mar - Apr .
15. Brucker MC, et al;
Neonatal jaundi ce in the home: assessment with a noninvasive device.
J Obstet Gynecol Neonatal Nurs, 1987 Sep-Oct.
16. Cassady G
Transcutaneous mo nitoring in the ne w born in fant .
J Pediatr, 1983 Dec.
17. Chen ZL, et al;
Clinical use of transcutaneous bilirubinometry.
Chung Hua I Hsueh Tsa Chih, 1986 Jan.
18. Christo GG, et al;
Transcutaneous bilirubinometry in newborns.
Indian Pediatr, 1988 Nov.
Page B - 2 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 63
19. Corch ia C, et al;
Comment on transcut aneous bilirubin device of Yamanouchi [letter]
Pediatrics, 1981 Mar.
20. Dai J, et al;
Clinical impact of transcut aneous bilirubinometry as an adjunctive screen for hyperbiliru bine mi a
Clin Biochem 29 (6): 581-586, Dec 1996
21. Dai J, et al;
Transcutaneous bilirubinometry: its role in the assess ment of neo­natal jaundice
Clin Biochem 30 (1): 1-9, Feb 1997
22. Derksen-Samsom JF, et al;
The reliability of trans cut aneous bilirubin measu rem ent : a clinical study with statisti ca l dat a and literature review.
Tijdschr Kindergeneeskd, 1984 Oct.
23. Dominguez Ortega F , et al;
Transcutaneous bilirubinometry: correlation of the reading site obtained with spectrophotometry and diaz oreaction technique.
An Esp Pediatr, 1993 Nov.
24. Douville P, et al;
Diagnostic value of sequential readings with the Minolta transcuta­neous bilirubi nometer in normal and low-birthweight infants [letter]
Clin Chem, 1983 Apr, 29:4, 740-1.
25. Fabr i s C , et al;
Evaluation of transcutaneous bilirubinometry in newborn infants.
Minerva Pediatr, 1984 Jun 15.
26. Fisher B, et al;
The cephalocaudal progressi on of jaundice in newborns in relation to the transfer of bilirub in from plasma to ski n.
Early Hum Dev, 1990 Apr, 22:1, 23-8
27. Foged N, et al;
Transcutaneous bil irubinometry. A non-invasive method of measur­ing physiologica l jaundice.
Ugeskr Laeger, 1983 Sep. 19.
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page B - 3
Page 64
28. Fok TF, et al;
Transcu taneou s bilirubi nometer: i ts use in Chines e term infants and the effect of haema toc rit and phototherapy on the TcB index.
Aust Paediatr J, 1986 May.
29. Galletto P, et al;
Supervision of neonat al jaun dice by use of the transcutaneous bilirubinometer.
Minerva Pediatr, 1983 Jan 31.
30. Goldman SL, et al;
Jaundice meter: evaluation of new guidelines.
J Pediatr, 1982 Aug.
31. Grande R, et al;
Physiological va ria ti ons in the pigm ent at ion of ne w born infants.
Hum Biol, 1994 Jun.
32. Hanneman R; et al;
Determination of serum bi lirubin by skin reflectance : ef fect of pig­mentation
Pediatr Res 13 (12): 1326-1329 (Dec 1979)
33. Hanneman R; et al;
Evaluation of the Minolta bil irubin meter as a screening device in white and black infants.
Pediatrics, 1982, 69, 107-109
34. Hegyi T, et al;
Tra n scutan eo us bilirubinometr y. I. Corre lations in term infants.
J Pediatr, 1981 Mar.
35. Hegyi T, et al;
Transcutaneous bilirubinometry. The cephalocaudal progression of derma l icterus .
Am J Dis Child, 1981 Jun.
36. Hegyi T
Transcutaneous bilirubinometry: a new light on an old subje ct .
Pediatrics, 1982 Jan.
37. Hegyi T, et al;
Transcu taneous bilirubin ometry II. Dermal bil irubin kinetics duri ng phototherapy.
Pediatr Res, 1983 Nov.
Page B - 4 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 65
38. Hegyi T
Transcutaneous bilirubinometry in the newborn infant : sta te of th e art.
J Clin Monit, 1986 Jan.
39. Hegyi T, et al;
Transcutaneous bilirubinometry. III. Dermal bilirubin kinetics under green and blue light phototherapy.
Am J Dis Child, 1986 Oct.
40. Heick C, et al;
Transcut aneous bilirubin determi nation in the newborn infant. Rec­ommendations of the Swiss Neonatology Group.
Helv Paediatr Acta, 1982.
41. Hodr R , et al;
Transcutaneous measuremen t of the severi ty of icte rus and bil iru­binemia in normal neon at es.
Cesk Pediatr, 1990 Nov.
42. Hung WT, et al;
Diagnosis of atreti c prol onged obstructive jaundice; technetium 99m hepatolite excr etion study.
J Pediatr Surg, 1990 Jul.
43. Karr ar Z, et al;
Transcutaneous bilirubin measurements i n Saudi infants: the use of the jaundice meter to identify significant jaun dic e.
Ann Trop Paediatr, 1989 Mar.
44. Kenny M, et al;
Transcutaneous bilirubin monitoring of newborns .
Ann N Y Acad Sci, 1984.
45. Keshishjan E.S., et al;
Diagnostics and control of newborn hype rbilirubinemia with use noninvasive trans cutaneous photometric analyzer "Bilitest" of a type AHP-02.
The methodical r e commendat ion s. Russia mini s try of public health. 1992 Feb.
46. Keshishjan E.S., et al;
Transcutaneous bilirubinometry method at newborn jaundice diag­nost ics and c o ntrol.
The Russian bulletin perinatology and pediatrics, V38, 5, 1993
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page B - 5
Page 66
47. Keshishjan E.S., et al;
Use of the method of transcutaneous bilirubinometry in hyperbi­lirubinemia in the newbo rn.
Clinical laboratory diagnostics. 5, 1994 Moscow, Medicine
48. Kivlahan C, et al;
The natural history of neonat al jaun dice.
Pediatrics, 1984 Sep.
49. Knudsen A.
Prediction of the deve lopm en t of neonatal jaundice by increa s ed umbilical cord blood bil irubin.
Acta Pediatr Scand 1989;78:217-21.
50. Knudsen A, et al;
Skin colour and bilirubin.
Arch Dis Child 1989;64:605-9.
51. Knudsen A, et al;
Maternal bili rubin , c ord bilirubin, and placenta function at delivery and the development of jaundice in mature newborns.
Acta Obstet Gynecol Scand 1989:68:719-24.
52. Knudsen A.
The cephalocaudal pr ogre s si on of ja undic e in newborns in relation to the transfer of bilirubin from plasma to skin.
Early Human Development 1990;22:23-28.
53. Knudsen A.
Measurement of the yellow colour of the skin as a test of hyperbi­lirubinemia in mature ne wborns .
Acta Pediatr Scand 1990;79:1175-81.
54. Knudsen A.
Maternal smokin g and the serum bilirubin concentrati on in the first three days of life.
Eur J Obstet Gynecol Reprod Bid 1991; 40:123-2 7.
55. Knudsen A.
The influence of the reserv e album in con ce ntra tion and pH on the cephalocaudal progression of jaundice in newborns.
Early Human Development 1991;25:37-41.
56. Knudsen A.
The influence of the basal yellow colour of the skin at birth on l at er jaundice meter readings in mature newborn infants.
Acta Pediatr Scand 1992;81:494-97.
Page B - 6 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 67
57. Knudsen A.
Prediction of later hyperbilirubinemi a by measure ment of s kin colour at first postnata l day and cor d bili rubin.
Dan Med Bull 1992;39:1 93-6.
58. Knudsen A
Predicting the need for phototherapy in healthy mature neonates using transcutan eous bilirubinometry on th e first postnatal day.
Biol Neonate, 1995.
59. Knudsen A, et al;
Transcutaneous evaluation of icterus in healthy matu re newb orn infants. A substitution for blood tests.
Ugeskr Laeger, 1995 Mar 13.
60. Knudsen A, et al;
Transcutaneous bilirubinometry in neonatal intensive care units.
Arc h Dis Child Fetal Neonat al Ed, 1996 Jul.
61. Kuma r A
Micro-invasive management of neonatal bilirubinemia.
Indian Pediatr, 1992 Sep.
62. Kuma r A , et al;
Transcut aneous bilirubinometry in the management of bilirubine­mia in te rm neonates.
Indian J Med Res, 1994 May.
63. Laeeq A, et al;
Transcutaneous bilirubinometry: clinical applic at ion.
JPMA J Pak Med Assoc, 1993 Feb.
64. Leb ru n F, et al;
Clinical evaluation of transcutaneous bilirubinometry.
Ann Pediatr (Pari s), 1982 Sep.
65. Lin YJ, et al;
The clinical applica ti on of tra ns cutaneous bilirubinom etry in full­term Chinese infants.
Acta Paediatr Sin, 1993 Mar-Apr.
66. Linder N, et al;
Noninvasive det ermination of neonatal hyperbilirubinemia: s ta n­dardization for variation in skin color.
Am J Perinatol, 1994 May.
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page B - 7
Page 68
67. Loong EP, et al;
Changes in neonatal t ranscutaneous bilirubinometer index follow ­ing intravenous flu id and oxytoc in in fus ion during labour.
Asia Oceania J Obstet Gynaecol, 1988 Dec.
68. Maisels MJ, et al;
Transcutaneous bilirubin measurements in full-term infants.
Pediatrics, 1982 Sep.
69. Maisels MJ, et al;
Tra n scutaneous bilirubin measurements: variation in me ter response.
Pediatrics, 1983 Mar.
70. Maisels MJ, et al;
Transcutaneous bilirubinometry decreases the need for serum bilirubin measurements and saves money
Pediatrics 99 (4): 599-601, Apr 1997
71. Moscicka A, et al;
Usefulness of transc uta neous measurements of bili rubin in infants with jaundice.
Ginekol Pol, 1994 Jun.
72. Myara A, et al;
Early changes in cutane ous bili rubin and serum bilirubin isom ers during intensive phototherapy of jaundiced neonates with blue and green light
Biol Neonate 71 (2): 75-82, 1997
73. Onks D, et al;
Effect of melanin, oxyhemoglobin and bili rubin on transcutaneous bilirubinometry.
Acta Paediatr, 1993 Ja n.
74. Pallas Alonso C, et al;
Transcutaneous bilirubin measurement in neonates.
An Esp Pediatr, 1993 Jan.
75. Palmer DC, et al;
Transcutaneous bilirubinometry: use in Australia.
Aust Paediatr J, 1982 Dec.
76. Palmer DC; et al;
Jaundice: A 10 year review of 41,000 live born infa nts. Australian Pediatrri c Journal, 1983, 19(2),86-8 9
Page B - 8 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 69
77. Peng DJ, et al;
A study of the effect of pathologic factors on the TcBM readings.
Hua Hsi I Ko Ta Hsueh Hsueh Pao, 1989 Mar.
78. Romo - P inales J, et al;
Preliminary s tudy in Mexico on the use of the transcutaneous biliru­binometer.
Bol Med Hosp Infant Mex, 1984 Oct.
79. Ruchala PL, et al;
Validating assessme nt of neonatal jaundic e with transcutaneous bilirubin measure ment.
Neonatal Ne tw, 1996 Jun.
80. Salmeron Escobar FJ
Differential diagnosis of jaundi ce .
Rev Esp Enferm Dig, 1991 Jan.
81. Schubiger G, et al;
Transcut aneous determination of bilirubin in the neonatal depart­ment: an analysis of its uses.
Helv Paediatr Acta, 1986 Aug.
82. Schumacher RE, et al;
Transcutaneous bilirubinometry: a comparison of old and new methods.
Pediatrics, 1985 Jul.
83. Schumacher RE
Noninvasive measurements of bilirubin in the newborn.
Clin Perinatol, 1990 Jun.
84. Serrao PA, et al;
Significance of ant i-A an d anti -B isohemagglutinins in cord blood of ABO incompatible newborn in fant s : correlation with hyperbi­lirubinemia
J Perinatol, 1989 Jun.
85. Sharma JN, et al;
Transcutaneous bilirubinometry in newborns.
Indian Pediatr, 1988 Aug.
86. S heridan-Perei r a M, et al;
Transcutaneous bilirubinometry: an evaluation.
Arch Dis Child, 1982 Sep.
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page B - 9
Page 70
87. Smith DW, et al;
Use of noninvasive tests to pre dic t significant jaundice in full-term infants: preliminary studies.
Pediatrics, 1985 Feb.
88. Str an ge M, et al ;
Neonatal transcutaneous bilirubinometry.
Clin erinat ol, 1985 Feb.
89. Suckling RJ, et al;
Transcutaneous bilirubinometry as a screening tool for neonat al jaundice.
Scott Med J, 1995 Feb.
90. Taha SA, et al;
Transcutaneous biliru bin mea su r ement in evaluating neonat al jaun ­dice among Saudi newborns.
Ann Trop Paediatr, 1984 Dec .
91. Tan KL
Tra n scutan eo us bilirubinome tr y in Chinese and Malay neonate s.
Ann Acad Med Singapore, 1985 Oct.
92. Tan KL, et al;
Transcutaneous bilirubinometry in preterm very low birthwe ight infants.
Acta Paediatr Scand, 1988 Nov.
93. Tan KL, et al;
Efficacy of phototherapy for hyperbilirubinaemia in infants with the respiratory distre ss syndrome
J Paediatr Child Health 31 (2): 127-129, Apr 1995
94. Tan KL, et al;
Transcutaneous bilirubinometry in Chinese, Malay and Indian infants
Acta Paediatr 85 (8): 986-990, Aug 1996
95. Tsai LT, et al;
Clinical evaluat ion of tra ns cu ta neous jaundice meter in full-te r m newborns.
Acta Paediatr Sin, 1988 Nov-Dec.
96. Tudehope DI, et al;
Non-invasive method of measuring bilirubin leve ls in ne wb orn infants.
Med J Aust, 1982 Feb 20.
Page B - 10 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 71
97. Tude hope DI, et al;
Multiple site readings from a transcutaneous bilirubinometer. Aust Paediatr J, 1982 Jun.
98. Uchida K, et al;
Fundamental studi es on trans cutan eous bilir ubinometry in newborn infants using an organ scanning spectrophotometer.
Nippon Sanka Fujinka Gakkai Zasshi, 1988 Feb.
99. Utchajkin V.F., et al;
Clinical importance of noninvasive bilirubinemia monitoring by children viral hepat iti s .
2-nd Symposi ums "No ninvas iveMet hods of Dia gnosti cs" . The t hesis of the reports. Moscow. 1995 Oct. 30 - Nov. 2.
100. Vaisman S, et al;
Clinical trial of a trans cut ane ous bilirubinometer.
Rev Chil Pediatr, 1983 Mar-Apr.
101. Veleminsky M
The significance of transcutaneous measurement of "bilirubi n" in neonates.
Cesk Pediatr, 1992 Nov.
102. Vocel J, et al;
Transcut aneous bilirubinometry using the Minolta/Air-Shields in neonates.
Cesk Pediatr, 1985 Nov.
103. Vocel J, et al;
Transcutaneous bilirubinometry using the Minolta/ Air-Shields apparatus in neonates with a low birthweight.
Cesk Pediatr, 1987 Jan.
104. Vocel J, et al;
Transcutaneous bilirubinometry in neonates with hyperbilirubine­mia treated with photot he rapy.
Cesk Pediatr, 1988 Dec.
105. Weiss WJ
A completely implante d left ventricular assist de vic e. Chronic in vivo testing.
ASAIO J 39(3), M427-M432, 1993
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page B - 11
Page 72
106. Yamanouchi I, et al;
Transcutaneous bilirubinometry: prelimina ry s tudi es of noninva­sive transcutaneous bilirubin meter in the Okayama N ational Hos­pital.
Pediatrics, 1980 Feb.
107. Yamauchi Y, et al;
Transcutaneous bilirubino me try. Effect of irradiation on the sk in bilirubin index. Biol Neonate, 1988.
108. Yamauchi Y, et al;
Transcuta ne o us bilirubi n ometry. Evaluatio n of accuracy and re li ­ability in a large popul at ion.
Acta Paediatr Scand, 1988 Nov.
109. Yamauchi Y, et al;
Transcutaneous bilirubinometry: bilirubin kine tics of the skin and serum during and after phot othe rapy.
Biol Neonate, 1989.
110. Yamauchi Y, et al;
Transcut aneous bilirubinometry: serum bilirubin measurement using transcutane ous bil irubinometer (TcB). A prelim ina ry study.
Biol Neonate, 1989.
111. Yamauchi Y, et al;
Transcutaneous bilirubino me try in norma l Japa nese infants.
Acta Paediatr Jpn, 1989 Feb.
112. Yamauchi Y, et al;
Tra n scutan eo us bilirubinometr y. Interin strumental varia b ility of TcB instruments.
Acta Paediatr Scand, 1989 Nov.
113. Yamauchi Y, et al;
Clinical ap plicatio n o f transc utane ous bili rubin m easure ment. E arly prediction of hyperbilirubinemia.
Acta Paediatr Scand, 1990 Apr.
114. Yamauchi Y, et al;
Transcutaneous bilirubinometry: eff ect of postnatal ag e.
Acta Paediatr Jpn, 1991 Oct.
115. Yamauchi Y, et al;
Transcutaneous bilirubinometry: variability of TcB measurements on the forehead with crying.
Acta Paediatr Jpn, 1991 Oct.
Page B - 12 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 73
1 16. Yamauchi Y, et al;
Factors affecting transcutaneous bilirubin measurement : ef fect of daylight.
Acta Paediatr Jpn, 1991 Oct.
1 17. Yip WC, et al;
Transcutaneous bilirubin ome try [letter]
Acta Paediatr Scand, 1983 Mar.
Jaund ice Meter (JM-103) User Manual (usr0 70rb) Page B - 13
Page 74
NOTES:
Page B - 14 Jaundice Meter (JM-103) User Manual (usr070rb)
Page 75
Jaundi ce Me t er ( Mo d el JM - 1 03 ) User M an ual (usr07 0)
Page 76
Manufactured by: Konica Minolta Sensing, Inc. 3-91, Daisennishimachi Sakai, Osaka 590-8551 Japan
Authorized Representative: Hill-Rom Europe Director Quality and Regulatory Affairs Les Rabelais - Paris Nord II 22 Avenue des Nations BP 50436 -VILLE PINTE 95 944 - Rois sy CDG CEDEX France tel. +01 49 89 40 20 fax +01 49 89 40 40
Distributed by:
330 Jacksonvill e Road, Hatboro, PA 19040 USA USA: (800) 523-5756 International: Contact your distributor. Website: www.hill-rom.com
Part number Revisions Date
usr070 rb October, 2003
Hill-Rom reserves the right to make changes wit hout notice in design, specifications, and models. The only warranty Hill-Rom makes is the express written warrant y extended on the sale or r ental of its products. The informati on in th is manual is confidential and may not be disclosed to third partie s without the prior writt en consent of Hill-Rom. © 2003 by Hill-Rom Services, Inc. ALL RIGHTS RESERVED.
Loading...