Dell PS-402-HG-OEM User Manual

Dell PS-402-HG-OEM User Manual

UL1363a, Paragraph 14.1(d) States "The SPRPT shall be permanently attached to the medical equipment by a means such that it is only removable with the use of a tool." To achieve compliance, one must use

the PSCLAMP.

Safe-IT UL 1363A Medical-Grade Power Strip for Patient-Care Vicinity, 4x 15A Hospital-Grade Outlets, 2 ft. Cord

MODEL NUMBER: PS-402-HG-OEM

Supplies power to movable plug-connected components mounted to IV poles, mobile carts, crash carts

and mobile medical workstations in operating rooms and patient-care vicinities.

Description

The PS-402-HG-OEM Medical-Grade Power Strip supplies power to movable plug-connected components mounted to a rack, table or pedestal in operating rooms, patient-care vicinities, therapy rooms, imaging environments and administrative areas. Compliant with UL 1363A specifications, this 120V, 50/60 Hz power strip features a NEMA 5-15P-HG hospital-grade plug and four NEMA 5-15R-HG hospital-grade outlets.

The power strip is protected by a patented antimicrobial powder coating with EPA-registered and FDAcompliant silver ionic technology. The antimicrobial surface protection is 99.9 percent effective in inhibiting the growth of bacteria and reducing the risk of HAIs in hospitals and other medical environments. The coating meets the requirements of JIS Z 2801:2000, an international standard for evaluating efficacy in antimicrobial products.

Two built-in resettable 15A circuit breakers control power to all outlets and protect the power strip against overloads. Switchless design prevents the unit from being accidentally shut off. The two-foot (0.6-meter) power cord eliminates excess cordage that needs to be wrapped.

Mounting tabs and flanges let you install the PS-402-HG-OEM on a wide variety of surfaces, such as IV poles, mobile carts, crash carts and mobile medical workstations. The all-metal housing ensures safety and is guaranteed to last a long time in demanding environments.

Features

Power Distribution Anywhere in Any Medical FacilityIdeal for use in operating rooms, patient-care vicinities, therapy rooms, imaging environments and administrative areasFor use on IV poles, mobile carts, crash carts and mobile workstations, when properly mounted120V, 50/60 Hz power strip complies with UL 1363A specifications

Exclusive, Patented Antimicrobial ProtectionSilver ionic powder coating is 99.9% effective in inhibiting the growth of bacteriaPatented coating helps to reduce the risk of healthcare-acquired infections (HAIs) in

Highlights

Patented antimicrobial coating is 99.9% effective in inhibiting the growth of bacteria

Approved for use in patient-care vicinities when properly mounted

Complies with UL 1363A specifications

Features 4 NEMA 5-15R-HG hospital-grade outlets

Includes 2 ft. (0.6 m) cord with NEMA 5-15P-HG hospital-grade plug

Package Includes

PS-402-HG-OEM UL 1363A Medical-Grade Power Strip with 4 15A Hospital-Grade Outlets and 2-ft. (0.61 m) Cord

Owner's Manual with maintenance requirements

1 / 3

hospitals and other medical environmentsCoating meets the requirements of JIS Z 2801:2000, an international standard for evaluating efficacy in antimicrobial products.

Versatile Medical-Grade Solution4 NEMA 5-15R-HG hospital-grade outlets provide consistent power2ft. (0.61 m) power cord eliminates excess cordage that needs to be wrapped upRugged, all-metal housing for long life in demanding environments

Premium Safety FeaturesDual, resettable built-in 15A circuit breakers protect against dangerous overloadsSwitchless design prevents accidental shut-off

Ready to Use Right Out of the BoxBuilt-in mounting tabs and flanges make the PS-402-HG-OEM easy to install on a variety of surfaces

Specifications

OVERVIEW

UPC Code

037332147998

 

 

INPUT

 

Nominal Input Voltage(s) Supported

120V AC

 

 

Recommended Electrical Service

15A, 120V

 

 

Input Plug Features

Hospital Grade Plug

 

 

Input Cord Gauge Type

14 gauge; SJT

 

 

Voltage Compatibility (VAC)

120

 

 

Input Plug Type

NEMA 5-15P-HG

 

 

Input Cord Length (ft.)

2

 

 

Input Cord Length (m)

0.61

 

 

Integrated Cord Clip

No

 

 

OUTPUT

 

Frequency Compatibility

50 / 60 Hz

 

 

Output (Watts)

1800

 

 

Output Receptacles

(4) 5-15R-HG

 

 

Circuit Breaker (amps)

15

 

 

Overload Protection

2 x 15amp breakers (1 line, 1 neutral)

 

 

Transformer Accommodation

(2) Yes

 

 

Outlet Orientation

Single row of 4

 

 

Right-Angle Outlets

No

 

 

USER INTERFACE, ALERTS & CONTROLS

Switches

No power switch

 

 

PHYSICAL

2 / 3

Loading...