Features
DMS-20PC Series, 3½ Digit, LED Display, Digital Panel Voltmeters combine a
precision A/D converter; a factory-trimmed, highly stable, voltage reference; and a
large (0.37"/9.4mm), easy-to-read LED display in a single package that is only
slightly larger than the display itself. Displays are offered in either red, orange,
amber, yellow or green colors. High-intensity and low-power (35mW total) red LED's
are also optional.
subminiature (1.38" x 0.88" x 0.48"), epoxy-encapsulated cases incorporate built-in
color filters and bezels; are moisture and vibration proof; and function well in the
harshest environments. Their 12-pin, dual-in-line configuration offers component-like,
plug-in convenience and maximum versatility. Operating temperature range is 0 to
+60°C.
The meters come with one of four, differential, input voltage ranges: ±200mV, ±2V
±20V or ±200V. Input impedance is a minimum 800k. CMRR is typically 86dB (dc
to 60Hz), and CMV is ±2V. Input overvoltage protection (on the non-inverting input)
is ±250V. Devices are fully calibrated at the factory to an accuracy of ±1 count
(±0.05% of full scale range) and never require calibration or adjustment.
A DISPLAY ENABLE function permits the display to be disabled for "power-
down" operation. All models have a DISPLAY TEST function. Standard red LED
models offer an optional DISPLAY HOLD function.
Small size, low cost and adjustment-free reliability make the DMS-20PC Series
the best choice for all your 3½ digit, LED, DPM applications.
Figure 1. DMS-20PC Series Simplified Schematic
Lowest-cost LED meters
Subminiature size:
1.38" x 0.88" x 0.48"
35mm x 22mm x 12mm
Large (0.37"/9.4mm) LED display
Choice of 5 LED colors
High-intensity or low-power (7mA) red
LED's optional
Epoxy-encapsulated, 12-pin DIP package
with built-in color filter and bezel
4 differential input voltage ranges
Factory calibrated, ±1 count accuracy
Single +5V power supply
User-selectable decimal point placement
DISPLAY ENABLE function for
"power-down" mode
DISPLAY TEST and HOLD (optional)
functions
0 to +60°C temperature range
3½ Digit, LED Display
Low-Cost, Subminiature
Digital Panel Voltmeters
DMS-20PC Series
Actual Size
(+) INPUT HI
() INPUT LO
ANALOG
COMMON
REFERENCE OUT
REFERENCE IN
+5V SUPPLY
R1
DP 1
DP 2
DP 3
6
5
4
11
12
10
8
7
1
3
DC/DC
CONVERTER
2
5V RETURN
+5V
5V
R2
»
+2.0V
DATA
909k
V+
V
A/D
CONVERTER
0 Vdc
0.01
µF
9
DISPLAY
ENABLE
+5V
R2 is not used on ± 2 0 0 mV (-0) models or ±2V (-1) mode ls.
R2 = 100k on ±20V (-2) models and 9.1k on ±200V (-3) models.
Only used on ±200mV (-0) and ±2V (-1) models.
DISPLAY
TEST/HOLD
11
Pin 9 is DISPLAY TEST on all but eight models. On those models
(
-H option
)
, it is DISPLAY HO L D.
1
2
3
2
1
3
®®
www.datel.com
2
DMS-20PC
3½ DIGIT, LED DISPLAY DIGITAL PANEL VOLTMETERS
Performance/Functional Specifications
Typical at TA = +25°C and supply voltage = +5V using the single-ended input circuit,
unless otherwise noted.
Applies for transient or continuous overvoltages applied to (+) INPUT HI (pin 11)
with () INPUT LO (pin 12) properly connected. Pin 12 is not overvoltage protected
(see Figure 1). Voltages applied to pin 12 should not exceed the supply voltage.
See Technical Notes.
The DISPLAY HOLD function is optional on standard red and green LED models only.
See Ordering Information.
Leave blank for standard
models.
Add -H for DISPLAY
HOLD option (available on
standard red and green
LED models only).
DMS-20PC-1- RS
LED Color:
AS = Standard Amber
GS = Standard Green
OS = Standard Orange
RS = Standard Red
YS = Standard Yellow
RH = High-Intensity Red
RL = Low-Power Red
Accessories:
DMS-20-CP Panel cutout punch
DMS-BZL3 DMS-20 bezel assembly
DMS-BZL4 DMS-20 bezel assembly with sealing gasket
DMS-EB2 Application/evaluation board with standard
MOLEX connector, decimal point solder pads
and attenuation resistor pads.
T echnical Notes
1. REFERENCE OUTPUT (Pin 8) and INPUT (Pin 7): Pin 8 is a
precision reference actively trimmed at the factory. In normal
operation, pin 8 must be tied to pin 7 to achieve all listed accuracy
and drift specifications.
2. ANALOG COMMON (Pin 10): This pin is connected to an internal,
low-noise, "relative" ground. It is used in certain differential and
"floating" measurements as described in the Applications section
of this data sheet and Ap Note 3 of the DATEL Panel Meter Catalog.
Pin 10 should not be connected to pin 3 (5V RETURN) or to
your system’s analog ground.
3. Decimal Point Placement: The location of the decimal point is
user-selectable, and the decimal point control pins (DP1-DP3) are
active low functions. Select the appropriate decimal point by tying
the appropriate pin (pin 4, 5 or 6) to pin 3 (5V RETURN). Unused
decimal point location pins should be left open.
Hard wiring is preferable, however, you can use logic gates to
exercise dynamic control over the location of the decimal point if the
following drive conditions are met:
Model Applied "0" Voltage Load Current*
DMS-20PC-X-RL +0.05V max. 0.7mA max.
All Others +0.4V max. 6mA max.
* The driving gates must be able to sink this much current
(I
OL) with a logic "0" output.
Ordering Information
Input Range:
0 = ±200mV
1 = ±2V
2 = ±20V
3 = ±200V
A panel-mount retaining clip is supplied with each model.
Analog Inputs Min. Typ. Max. Units
Full Scale Input Range:
DMS-20PC-0 -- ±200 -- m V
DMS-20PC-1 -- ±2 -- Volts
DMS-20PC-2 -- ±20 -- Volts
DMS-20PC-3 -- ±200 -- Volts
Input Impedance:
DMS-20PC-0, -1 100 1000 -- M
DMS-20PC-2, -3 0.8 1 -- M
Overvoltage Protection
-- -- ±250 Volts
Common Mode Voltage Range -- -- ±2 Volts
CMRR (dc to 60Hz) -- 86 -- dB
Control Inputs ➁ ➂
Decimal Point Placement (Pins 4-6) Tie to pin 3 to activate
Display Test (Pin 9) Tie to pin 1 to activate all segments
Display Hold (Pin 9)
Tie to pin 1 to hold last reading
Performance
Sampling Rate 2.5 samples per second
Accuracy (3 minute warm-up):
DMS-20PC-0 (VIN = +0.19V) -- ±1 ±3 Counts
DMS-20PC-1 (VIN = +1.9V) -- ±1 ±3 Counts
DMS-20PC-2 (VIN = +19V) -- ±2 ±3 Counts
DMS-20PC-3 (VIN = +190V) -- ±2 ±3 Counts
Zero Reading (VIN = 0 Volts) "001" "000" "001"
Temperature Drift (0 to +60°C) -- ±0.2 ±0.4 Cnts/°C
Power Supply Requirements
Supply Voltage + 4.75 +5.00 +5.25 Volts
Supply Current:
DMS-20PC-X-RS -- +60 +90 mA
DMS-20PC-X-RL -- +7 +12 mA
DMS-20PC-X-AS -- +90 +120 mA
DMS-20PC-X-GS -- +90 +120 mA
DMS-20PC-X-OS -- +90 +120 mA
DMS-20PC-X-YS -- +90 +120 mA
DMS-20PC-X-RS-H -- +60 +90 mA
DMS-20PC-X-GS-H -- +90 +120 mA
DMS-20PC-X-RH -- +6 0 +90 mA
Display
Display Type and Size 3½ Digit LED, 0.37"/9.4mm high
Polarity Indication Autopolarity ("" for negative VIN)
Overrange Indication "1___" for negative VIN
"1___" for positive VIN
Physical/Environmental
Operating Temperature 0--+60°C
Storage Temperature 40 -- +75 °C
Humidity (Non-condensing) 0 -- 95 %
Case Material Polycarbonate
Weight 0.4 ounces (11 grams)