Covidien Nellcor Portable SpO2 Service manual

4.5 (2)
Covidien Nellcor Portable SpO2 Service manual

Service Manual

NellcorTM

Portable SpO2 Patient Monitoring System

COVIDIEN, COVIDIEN with logo, and Covidien logo and Positive Results for Life are U.S. and internationally registered trademarks of Covidien AG. ™* brands are trademarks of their respective owner. Other brands are trademarks of a Covidien company.

Table of Contents

1Introduction

1.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-1

1.2

Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-2

1.2.1

Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-2

1.2.2

Explosion, Shock, and Toxicity Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-2

1.2.3

Service Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-3

1.2.4

Monitoring System Operation and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-4

1.2.5

Patient Monitoring and Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-5

1.2.6

Monitoring System Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-5

1.2.7

Sensors, Cables, and Other Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-6

1.2.8

Electromagnetic Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-6

1.2.9

Connections with Other Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-7

1.2.10

Monitoring System Storage, Transport, and Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-7

1.3

Obtaining Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-8

1.3.1

Technical Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-8

1.3.2

Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-8

1.4

Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-9

2Service Menu Options

2.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-1

2.2

Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-1

2.3

Service Menu Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-1

2.3.1

Power On Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-3

2.3.2

Permission to Mute Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-3

2.3.3

Alarm Silence Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-4

2.3.4

Power Saving Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-4

2.3.5

Battery Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-4

2.3.6

Date/Time Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-4

2.3.7

Communication Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-4

2.3.8

Alarm Priority Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-5

2.3.9

Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-5

2.3.10

Dynamic Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-5

2.3.11

Enable PCSync Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-5

2.3.12

About Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-5

2.3.13

Covidien Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-6

3Data Management

3.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-1

3.2

External Data Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-1

3.2.1

Monitoring History Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-2

Service Manual

i

3.3

Firmware Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-11

4Modification and Testing

4.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-1

4.2

Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-2

4.3

System Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-2

4.3.1

Power-On Self-Test (POST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-2

4.3.2

Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-3

4.3.3

Patient Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-4

4.3.4

Homecare Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-5

4.3.5

Sleep Study Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-6

4.3.6

Dynamic Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-7

4.3.7

Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-9

4.4

Operational and Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-10

4.4.1

General Operation Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-10

4.4.2

Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-26

4.5

Verification Check Sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-45

5Troubleshooting

5.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-1

5.2

Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-1

5.2.1

Error Conditions by Category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-2

5.2.2

System Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-6

5.3

Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-8

6Repair

6.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-1

6.2

Spare Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-2

6.3

Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-4

6.4

Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-4

6.4.1

Remove the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-5

6.4.2

Replace the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-5

6.5

Disassembly and Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-6

6.5.1

Front and Rear Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-6

6.5.2

NELL1SR Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-8

6.5.3

Main Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-9

6.5.4

Coin Cell Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-11

6.5.5

LCD Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-12

6.5.6

PI Cable and Cable Housing Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-13

ii

Service Manual

List of Tables

Table 1-1. Safety Symbol Definitions...............................................................................

1-2

Table 2-1. Service Menu Settings......................................................................................

2-2

Table 3-1. Monitoring Status Codes.................................................................................

3-6

Table 4-1. Required Test Equipment ...............................................................................

4-2

Table 4-2. Patient Modes .....................................................................................................

4-5

Table 4-3. Functional Tests with SRC-MAX.................................................................

4-27

Table 5-1.

Error Conditions and Resolutions ................................................................

5-2

Table 5-2.

System Error Codes ...........................................................................................

5-6

Table 6-1.

Spare Parts List by Callout Number.............................................................

6-2

Table 6-2.

Monitoring System Accessories....................................................................

6-4

Service Manual

iii

Page Left Intentionally Blank

iv

Service Manual

List of Figures

Figure 2-1.

Service Menu ....................................................................................................

............. 2-3

Figure 3-1.

Mini-USB Port ................................................................................................................

3-3

Figure 3-2. Transfer Data Type ......................................................................................................

3-4

Figure 3-3. Transfer Data by USB ..................................................................................................

3-4

Figure 3-4. Sample Monitoring History Printout ....................................................................

3-5

Figure 3-5. Sample Bridge Driver Installer Window ..............................................................

3-7

Figure 3-6. Sample New Hardware Wizard Screen ................................................................

3-7

Figure 3-7. Device Manager Button, Hardware Tab ..............................................................

3-8

Figure 3-8. Sample Hardware List in Device Manager Window ........................................

3-9

Figure 3-9. Sample Initial USB to UART Bridge Properties Window ..............................

3-10

Figure 3-10. Baud Rate List, Port Settings Tab .......................................................................

3-11

Figure 3-11. Firmware Upgrade Mode .....................................................................................

3-12

Figure 3-12. Firmware Upgrade, PCSync™* Utility ...............................................................

3-12

Figure 3-13. Firmware Upgrade Process .................................................................................

3-13

Figure 3-14. POST Screen and Firmware Version .................................................................

3-14

Figure 4-1. Power-On Self-Test Sequence ................................................................................

4-3

Figure 4-2. Change Patient Mode Menu ...................................................................................

4-4

Figure 4-3. Homecare Mode Menu Item ...................................................................................

4-5

Figure 4-4. Homecare Mode Monitoring Screen ....................................................................

4-6

Figure 4-5. Sleep Study Mode Menu Item ................................................................................

4-7

Figure 4-6.

Service Menu .................................................................................................................

4-8

Figure 4-7. Dynamic Password Menu (Homecare to Standard Example) ......................

4-8

Figure 4-8. Password Set or Reset (Homecare to Standard Example) ............................

4-9

Figure 4-9.

Date/Time Settings .....................................................................................................

4-9

Figure 4-10.

Sensor Port ................................................................................................................

4-10

Figure 4-11.

Sensor Port ................................................................................................................

4-11

Figure 4-12. Low SpO2 Alarm Limit of 99% ............................................................................

4-12

Figure 4-13. Low Pulse Alarm Limit of 160BPM ....................................................................

4-13

Figure 4-14. Alarm Silence Duration Setting of 30 Seconds ............................................

4-14

Figure 4-15. Permission to Mute Alarm ...................................................................................

4-15

Figure 4-16. Confirmation for Muted Alarm ...........................................................................

4-15

Figure 4-17. Alarm Volume Default Setting of 2 ...................................................................

4-17

Figure 4-18. Key Beep Volume Default Setting of 0 ............................................................

4-18

Figure 4-19. Pulse Volume Default Setting of 0 ....................................................................

4-19

Figure 4-20.

Brightness Setting ..................................................................................................

4-20

Figure 4-21. “Spot Reading Saved” Message .........................................................................

4-21

Figure 4-22. Monitoring History Spot Data ............................................................................

4-21

Figure 4-23. Sensor Disconnect Alarm Priority Setting ......................................................

4-22

Figure 4-24. Sensor Off Alarm .....................................................................................................

4-23

Figure 4-25. Screen Saver Time Setting ...................................................................................

4-24

Figure 4-26. Auto Power Off Time Setting ..............................................................................

4-26

Figure 4-27. SRC-MAX Oximetry Tester ...................................................................................

4-28

Figure 4-28.

SRC-MAX Tester-Generated Waveform ..........................................................

4-29

Service Manual

v

Figure 4-29. SRC-MAX Increase to 200 BPM ..........................................................................

4-30

Figure 4-30. SRC-MAX Decrease to 60 BPM ...........................................................................

4-30

Figure 4-31. SRC-MAX %SpO2 Increase to 90 .......................................................................

4-31

Figure 4-32. SRC-MAX %SpO2 Decrease to 75 .....................................................................

4-32

Figure 4-33. SRC-MAX High Modulation .................................................................................

4-33

Figure 4-34. BPM of 200 with High Modulation ...................................................................

4-34

Figure 4-35. BPM of 60 with High Modulation ......................................................................

4-35

Figure 4-36. %SpO2 of 90 with High Modulation ................................................................

4-36

Figure 4-37. %SpO2 of 75 with High Modulation ................................................................

4-37

Figure 4-38. %SpO2 of 75 with Low Modulation .................................................................

4-38

Figure 4-39. High Light Condition .............................................................................................

4-39

Figure 4-40. BPM of 200 with High Light Condition ...........................................................

4-40

Figure 4-41. BPM of 60 with High Light Condition ..............................................................

4-41

Figure 4-42. %SpO2 of 90 with High Light Condition ........................................................

4-42

Figure 4-43. %SpO2 of 75 with High Light Condition ........................................................

4-43

Figure 4-44. High Modulation and High Light Condition .................................................

4-44

Figure 5-1.

Return Packaging .........................................................................................................

5-9

Figure 6-1.

Exploded View ...............................................................................................................

6-2

Figure 6-2. Standard Cover (3 Shown) and Ambulatory Cover .........................................

6-4

Figure 6-3.

Battery Replacement ..................................................................................................

6-5

Figure 6-4. Front and Rear Assembly Replacement ...............................................................

6-7

Figure 6-5. NELL1SR Board Replacement ..................................................................................

6-9

Figure 6-6. Main Board Replacement .......................................................................................

6-10

Figure 6-7. Coin Cell Battery Replacement .............................................................................

6-11

Figure 6-8.

LCD Replacement .....................................................................................................

6-12

Figure 6-9.

PI Cable and Cable Housing Replacement ......................................................

6-13

vi

Service Manual

1Introduction

1.1Overview

This manual contains information for servicing the Nellcor™ portable SpO2 patient monitoring system.

This manual applies to the following product:

PM10N

Note:

Before use, carefully read this manual, the Operator’s Manual, accessory Instructions for Use, and all precautionary information and specifications.

Reference the Operator’s Manual for the following information:

Intended Use statement

Operations-related warnings and cautions

Overviews of the display and operating buttons

Descriptions of product and packaging symbols

Installation and operation instructions

Alarms management

Preventive maintenance

Performance considerations

Accessories

Theory of operations

Product specifications

Clinical studies

1-1

Introduction

1.2 Safety Information

This section contains important safety information related to general use of the Nellcor™ portable SpO2 patient monitoring system. Other important safety information appears throughout the manual. The Nellcor™ portable SpO2 patient monitoring system is referred to as the “monitoring system” throughout this manual.

1.2.1 Safety Symbols

Table 1-1. Safety Symbol Definitions

Symbol

Definition

WARNING

Alerts users to potential serious outcomes (death, injury, or adverse events) to the patient, user, or environment.

Caution

Identifies conditions or practices that could result in damage to the equipment or other property.

Note

Provides additional guidelines or information.

1.2.2Explosion, Shock, and Toxicity Hazards

WARNING:

Explosion hazard — Do not use the monitoring system in the presence of flammable anesthetics.

WARNING:

Shock hazard—Do not pour or spill liquids onto the monitoring system.

WARNING:

Shock hazard—Firmly close the battery cover to prevent moisture from entering the monitoring system.

WARNING:

The LCD panel (display) contains toxic chemicals. Do not touch broken LCD panels. Physical contact with a broken LCD panel can result in transmission or ingestion of toxic substances.

1-2

Service Manual

Safety Information

1.2.3Service Procedures

WARNING:

To avoid possible injury, do not attempt to service the monitoring system if there are any signs of burning or smoking coming from the monitoring system.

WARNING:

Before attempting to service the monitoring system, disconnect it from the patient to avoid possible injury to the patient.

WARNING:

Before attempting to disassemble the monitoring system, remove the batteries to prevent possible injury.

WARNING:

Ensure that conductive portions of the electrodes, leads, and cable do not come into contact with any other conductive parts.

WARNING:

High voltage is generated by the LCD backlight driver. Exercise caution when operating the monitoring system with covers open.

WARNING:

Extreme care must be taken in modifying default or other settings to ensure they are appropriate to the intended use.

WARNING:

Make sure to complete all performance tests in Chapter 4, Modification and Testing before placing the monitoring system into operation after repair or maintenance. Failure to perform all tests could result in erroneous monitoring system readings.

WARNING:

Any connections between this monitoring system and other devices must comply with applicable medical systems safety standards such as IEC 60601-1. Failure to do so could result in unsafe leakage current and grounding conditions.

WARNING:

To ensure accurate performance and prevent device failure, do not expose the monitoring system to extreme moisture, such as direct exposure to rain. Such exposure may cause inaccurate performance or device failure. Reference the Operator’s Manual for fluid ingress specifications.

Service Manual

1-3

Introduction

1.2.4Monitoring System Operation and Service

WARNING:

Inspect the monitoring system and all accessories before use to ensure there are no signs of physical damage or improper function. Do not use if damaged.

WARNING:

To ensure accurate performance and prevent device failure, do not expose the monitoring system to extreme moisture, such as direct exposure to rain. Such exposure may cause inaccurate performance or device failure. Do not immerse in water, solvents, or cleaning solutions, since the monitoring system and pulse oximetry sensors and connectors are not waterproof.

WARNING:

Do not sterilize the monitoring system by irradiation, steam, or ethylene oxide.

WARNING:

The monitoring system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, observe the monitoring system to verify normal operation in the desired configuration.

WARNING:

The only user-serviceable parts inside the monitoring system are the four AA batteries. While users can open the battery cover to change the batteries, only qualified service personnel should remove the cover or access internal components for any other reason. Users should not modify any components of the monitoring system.

WARNING:

Do not spray, pour, or spill any liquid on the monitoring system, its accessories, connectors, switches, or openings in the casing, since this may cause damage to the monitoring system. Never place fluids on the monitoring system. If fluid spills on the monitoring system, remove batteries, wipe all components dry immediately, and have the monitoring system serviced to ensure no hazard exists.

WARNING:

Do not damage the batteries by applying pressure. Do not throw, hit, drop, or impact the batteries.

WARNING:

Keep the monitoring system and batteries out of the reach of children.

Caution:

The monitoring system may not operate properly if it is operated or stored at conditions outside the ranges stated in this manual, or if it is subjected to excessive shock or dropping.

1-4

Service Manual

Safety Information

1.2.5Patient Monitoring and Safety

WARNING:

Always disconnect and remove the monitoring system and sensors during magnetic resonance imaging (MRI) scanning. Attempting to use the monitoring system during an MRI procedure could cause burns or adversely affect the MRI image or the monitoring system's accuracy.

WARNING:

Keep patients under close surveillance when monitoring. It is possible, although unlikely, that radiated electromagnetic signals from sources external to the patient and the monitoring system can cause inaccurate measurement readings.

WARNING:

As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.

WARNING:

Do not lift or carry the monitoring system by the pulse oximetry sensor or pulse oximetry interface cable. The cable may disconnect and cause the monitoring system to drop on a patient or cause damage to monitoring system surfaces.

1.2.6Monitoring System Readings

WARNING:

The monitoring system may remain attached to the patient during defibrillation or during use of an electrosurgical unit; however, the monitoring system is not defibrillator-proof, and readings may be inaccurate during defibrillation and shortly thereafter.

WARNING:

Check the patient's vital signs by alternate means should there be any doubt about the accuracy of any measurement. Request a qualified service technician confirm the monitoring system is functioning correctly.

WARNING:

For best product performance and measurement accuracy, use only accessories supplied or recommended by Covidien. Use accessories according to their respective Instructions for Use.

Service Manual

1-5

Introduction

1.2.7Sensors, Cables, and Other Accessories

WARNING:

Before use, carefully read the pulse oximetry sensor Instructions for Use, including all warnings, cautions, and instructions.

WARNING:

Use only the Covidien-approved pulse oximetry sensors, interface cables, and accessories. Use of other sensors, cables, and accessories can result in inaccurate readings and increased monitoring system emissions.

WARNING:

Do not use any other cables to extend the length of the Covidien-approved interface cable. Increasing the length will degrade signal quality and may lead to inaccurate measurements.

WARNING:

To prevent damage, avoid undue bending of the sensor cable.

WARNING:

The sensor disconnect error message and associated alarm indicate the pulse oximetry sensor is either disconnected or has faulty wiring. Check the connection and, if necessary, replace the sensor, the pulse oximetry cable, or both.

1.2.8Electromagnetic Interference

WARNING:

Any radio frequency transmitting equipment or other nearby sources of electrical noise may result in disruption of the monitoring system.

WARNING:

The monitoring system is designed for use in environments in which the signal can be obscured by electromagnetic interference. During such interference, measurements may seem inappropriate or the monitoring system may not seem to operate correctly.

WARNING:

Large equipment using a switching relay for its power on/off may affect monitoring system operation. Do not operate the monitoring system in such environments.

1-6

Service Manual

Safety Information

Caution:

This device has been tested and found to comply with the limits for medical devices related to IEC 60601-1-2: 2007 and IEC 60601-1-2:2014. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.

Caution:

This monitoring system generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. If interference is suspected, move pulse oximetry cables away from the susceptible device.

Caution:

Be aware of possible interference from sources of electromagnetic interference, such as cellular phones, radio transmitters, motors, telephones, lamps, electrosurgical units, defibrillators, and other medical devices. If pulse oximetry readings are not as expected for the patient’s condition, remove the sources of possible interference.

1.2.9Connections with Other Equipment

Caution:

Accessory equipment connected to the monitoring system's data interface must be certified according to IEC 60950-1 for data-processing equipment. All combinations of equipment must be in compliance with IEC 60601-1 Requirements for Medical Electrical Systems. Anyone who connects additional equipment to the signal input or signal output port configures a medical system and is therefore responsible for ensuring the system complies with the requirements of IEC 60601-1, IEC 60601-1-2:2007, and IEC 60601-1-2:2014.

Caution:

When connecting the monitoring system to any instrument, verify proper operation before clinical use.

Caution:

Anyone who connects a PC to the data output port configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of IEC 60601-1-1 and the electromagnetic compatibility IEC 60601-1-2.

1.2.10Monitoring System Storage, Transport, and Disposal

Caution:

Remove the batteries from the monitoring system before placing it in storage or when not using it for a long period.

Service Manual

1-7

Introduction

Caution:

Do not short-circuit the batteries, as they may generate heat. To avoid short-circuiting, do not let the batteries come in contact with metal objects at any time, especially during transport.

Caution:

Follow local government ordinances and recycling instructions regarding disposal or recycling of the monitoring system and its components, including batteries and accessories.

1.3Obtaining Technical Assistance

1.3.1 Technical Services

For technical information and assistance, contact Covidien or a local Covidien representative.

Covidien Technical Services: Patient Monitoring

15 Hampshire Street

Mansfield, MA 02048 USA

1.800.635.5267, 1.925.463.4635,

or contact a local Covidien representative

www.covidien.com

When calling Covidien or a local Covidien representative, have the monitoring system serial number available. The serial number label is located on the bottom of the monitoring system. Provide the firmware version number displayed during the power-on self-test (POST).

1.3.2 Related Documents

Nellcor™ Portable SpO2 Patient Monitoring System Operator’s Manual — Provides basic information for operating the monitoring system and troubleshooting errors or malfunctions.

Nellcor™ Pulse Oximetry Sensor Instructions for Use — Guides sensor selection and usage. Before attaching any of the various Covidien-approved pulse oximetry sensors to the monitoring system, refer to the individual Instructions for Use.

Nellcor™ Oxygen Saturation Accuracy Specification Grid — Provides sensor-specific guidance related to desired SpO2 saturation accuracy measurements. Available online with the product manuals for the monitoring system at www.covidien.com.

1-8

Service Manual

Warranty Information

1.4 Warranty Information

The information contained in this document is subject to change without notice. Covidien makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Covidien shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Service Manual

1-9

Introduction

Page Left Intentionally Blank

1-10

Service Manual

2Service Menu Options

2.1Overview

This chapter describes Service Menu settings for the Nellcor™ portable SpO2 patient monitoring system. A password is required to access the Service Menu. Contact Technical Services (reference Technical Services, page 1-8) for the default password.

2.2Safety Reminders

WARNING:

For best product performance and measurement accuracy, use only accessories supplied or recommended by Covidien. Use accessories according to their respective Instructions for Use.

WARNING:

The sensor disconnect error message and associated alarm indicate the pulse oximetry sensor is either disconnected or has faulty wiring. Check the connection and, if necessary, replace the sensor, the pulse oximetry cable, or both.

WARNING:

Do not use damaged pulse oximetry sensors. Do not use with exposed optical components. Do not immerse completely in water, solvents, or cleaning solutions, since pulse oximetry sensors and connectors are not waterproof. Do not sterilize by irradiation, steam or ethylene oxide. Refer to the cleaning instructions in the Instructions for Use for reusable sensors.

Caution:

Do not attach any cable intended for computer use to the sensor port connector.

2.3Service Menu Overview

WARNING:

Only qualified service personnel should open the monitoring system housing, remove and replace parts, or make adjustments.

2-1

Service Menu Options

The monitoring system contains a Service Menu that is accessible only by password. Contact Technical Services (reference Technical Services, page 1-8) for the default passwords for all restricted menus and functions.

The monitoring system ships with the Service Menu factory default settings listed in Table 2-1. For a list of the factory default settings for other menus, reference the Operator’s Manual.

Table 2-1. Service Menu Settings

Parameter

Ranges/selection

 

Factory default

 

 

 

 

 

Adult

Pediatric

 

Neonatal

 

 

 

 

 

 

 

 

 

Power On Settings

Factory Defaults, Institutional Defaults, Last Setting

 

Factory Defaults

 

 

 

 

 

 

Permission to Mute Alarm

No, Yes

 

No

 

 

 

 

 

 

Alarm Silence Duration

30, 60, 90, 120 seconds

 

120 seconds

 

 

 

 

 

 

Power Saving Settings

Screen Saver Time:

 

3 minutes

 

 

Never, 1 - 10 minutes

 

 

 

 

 

 

 

 

 

 

 

Screen Saver Brightness:

 

20%

 

 

 

0% - 50% in 10% increments

 

 

 

 

 

 

 

 

 

 

Auto Power Off Time:

 

3 minutes

 

 

Never, 1 - 10 minutes

 

 

 

 

 

 

 

 

 

Battery Type

Lithium, Alkaline

 

Lithium

 

 

 

 

 

 

Date/Time Settings

Date Format:

 

YY/MM/DD

 

 

YY/MM/DD, DD/MM/YY, MM/DD/YY

 

 

 

 

 

 

 

 

 

Communication Settings

Serial Connectivity Settings:

 

ASCII, 19200

 

 

ASCII 19200, ASCII 115200,

 

 

 

 

 

SPDout 19200, SPDout 115200

 

 

 

 

 

 

 

 

 

Alarm Priority Settings

Low, Medium, High:

 

Low

 

 

- Sensor Disconnect

 

 

 

 

 

- Sensor Off

 

 

 

 

 

- Sensor Failure

 

 

 

 

 

 

 

 

 

 

Low, Medium, High:

 

Medium

 

 

- SpO2 High

 

 

 

 

 

- Pulse Rate High

 

 

 

 

 

 

 

 

 

 

Medium, High:

 

Medium

 

 

- SpO2 Low

 

 

 

 

 

- Pulse Rate Low

 

 

 

 

 

 

 

 

 

Languages

Chinese, Czech, Danish, Dutch, English, Finnish,

 

English

 

 

French, German, Greek, Hungarian, Italian,

 

 

 

 

 

Japanese, Korean, Norwegian, Polish, Portuguese,

 

 

 

 

 

Russian, Slovakian, Spanish, Swedish, Turkish

 

 

 

 

 

 

 

Dynamic Password

Passwords can be changed for entry into Homecare

(Contact Technical Services)

 

Mode and Sleep Study Mode and back to Standard

 

 

 

 

 

Mode

 

 

 

 

 

 

 

 

 

Enable PCSync Mode

On, Off

 

Off

 

 

 

 

 

 

 

2-2

Service Manual

Service Menu Overview

Figure 2-1. Service Menu

2.3.1Power On Settings

Caution:

Last Settings should only be used in a service situation, not during normal operation.

Note:

Once changes are made to any settings, you must choose Last Settings or Institutional Defaults to have those settings saved when the monitoring system is powered off.

Factory Defaults — The settings for the monitoring system when shipped from the manufacturer. Choose this option to reset all values to factory defaults. (Default)

Institutional Defaults — The settings of the monitoring system for an individual institution or patient situation. Set all desired parameters, then go to the Service Menu and select Institutional Defaults. The monitoring system powers off once you exit the Service Menu. When you power on the monitoring system again, the institutional defaults you set will be active. When you switch between operating modes (for example, between Sleep Study Mode and Standard Mode), the institutional defaults are retained.

Last Settings — The settings of the monitoring system are saved just before power off.

2.3.2Permission to Mute Alarm

No — User cannot disable the audible alarm for SpO2 or pulse rate alarms. (Default)

Yes — User can disable the audible alarm for SpO2 or pulse rate alarms.

Service Manual

2-3

Service Menu Options

Note:

The visual alarm (the yellow background behind the SpO2 or pulse rate reading) is always enabled, even when the audible alarm is disabled.

2.3.3Alarm Silence Duration

The amount of time the audible alarm remains silent after the Silence Alarm button is pressed. Options include 30, 60, 90, and 120 seconds (default).

2.3.4 Power Saving Settings

Provides options for prolonging battery life during operation:

Screen Saver Time — By default, the screen dims after 3 minutes if no button is pressed. This value can be changed to 1 - 10 minutes or Never. When the screen dims, pressing any button returns the screen to normal brightness.

Screen Saver Brightness — By default, when the screen saver option is in effect, the screen dims to 20% of full brightness after the specified time without a button press. This value can be changed to 0 - 50% in 10% increments. When the screen dims, pressing any button returns the screen to normal brightness.

Auto Power Off Time — By default, the monitoring system powers off after 3 minutes if no button is pressed. This value can be changed from 1 - 10 minutes or Never.

2.3.5 Battery Type

Provides power optimization when using different battery types:

Lithium (default)

Alkaline

2.3.6Date/Time Settings

Provides options for the date format: YY/MM/DD (default), MM/DD/YY, or DD/MM/YY. Allows manual setting of the year, month, and day, and the hour, minute, and second.

2.3.7 Communication Settings

The serial connectivity settings to be used when trend data is transferred from the monitoring system to a PC using a USB cable. Options include:

ASCII, 19200 baud (default)

ASCII, 115200 baud

2-4

Service Manual

Service Menu Overview

SPDout, 19200 baud

SPDout, 115200 baud

2.3.8Alarm Priority Settings

Allows the setting of alarm priorities for Sensor Disconnect, Sensor Off, Sensor Failure, SpO2, and Pulse Rate alarms. Reference Table 2-1 on page 2-2.

2.3.9 Languages

Displays most menu items in the chosen language (reference Table 2-1 on page 2-2 for the complete list of languages). For some languages, certain menu items are not translated and always display in English.

2.3.10 Dynamic Password

Allows changes to the passwords for switching between the following operating modes:

Homecare to Standard Mode

Standard to Homecare Mode

Sleep Study to Standard Mode

Standard to Sleep Study Mode

Contact Technical Services (reference Technical Services, page 1-8) for the default passwords. Reference Dynamic Passwords, page 4-7 for instructions for changing these passwords.

2.3.11 Enable PCSync Mode

Enables use of the PCSync™* utility for testing the monitoring system and performing firmware upgrades. Choices are On or Off (default).

2.3.12 About Monitor

Model Name — PM10N

System Software Version — The version of the software (firmware) loaded on the monitoring system

SpO2 Module Version — The software version of the SpO2 circuit board installed in the monitoring system

Service Manual

2-5

Service Menu Options

2.3.13 Covidien Service

For Covidien personnel only.

2-6

Service Manual

3Data Management

3.1Overview

This chapter contains information for accessing, transmitting, and downloading patient monitoring data and history. This chapter also contains instructions for upgrading firmware for the Nellcor™ portable SpO2 patient monitoring system. The monitoring system supports the following types of data viewing and transmission:

Access stored monitoring history — Monitoring history (trend data) can be viewed anytime it is stored in the monitoring system. Reference the Operator’s Manual.

Download stored monitoring history — Monitoring history can be downloaded to a PC using the HyperTerminal™* program or other data transmission and analysis tools.

Upgrade the monitoring system’s firmware — Occasionally, Covidien will provide upgrades to the firmware for the monitoring system, which must be loaded via the mini-USB port.

3.2External Data Communication

WARNING:

Any connections between this monitoring system and other devices must comply with applicable medical systems safety standards such as IEC 60601-1 and applicable collaterals. Failure to do so may result in unsafe leakage current and grounding conditions.

Caution:

Do not attach any cable intended for computer use to the sensor port connector.

Caution:

Connect the monitoring system to a medical grade PC that is on an isolated AC circuit.

3-1

Data Management

3.2.1Monitoring History Download

WARNING:

Replacing the coin cell battery for the main board resets the monitoring system’s date and time settings. Integrity of existing patient data will be questionable. Reset the date and time after replacing this battery with a known good battery.

Caution:

Anyone who connects a PC to the data output port configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of IEC 60601-1-1 and the electromagnetic compatibility IEC 60601-1-2.

Caution:

Signal artifacts, secondary to a variety of external factors, may compromise the presence or accuracy of the displayed values.

Caution:

If the monitoring system does not contain its own isolation barrier, connect it to a medical grade PC that is on an isolated AC circuit.

The monitoring system presents monitoring history (trend data) in tabular format. The newest data values appear at the top.

To download monitoring history (trend data), connect by mini-USB port to a PC using the HyperTerminal™* program or other data transmission and analysis tools. Any PC connected to the data port must be certified according to IEC 60950. All combinations of equipment must be in compliance with IEC 60601-1-1 system requirements.

Note:

Users may choose to import patient monitoring history to a spreadsheet program. To do so, export monitoring history using the ASCII format option. Have a qualified service technician set this option prior to attempting a data download.

System Compatibility Prerequisites

Windows™* operating system

HyperTerminal™* program or equivalent data transmission and analysis software installed on the PC

Hardware

Mini-USB data download cable

CD or thumb drive, if USB driver required

3-2

Service Manual

External Data Communication

Data transfer by USB port relies on existing communication software drivers for USB-based devices already on the computer, so should not require any modification of the drivers used by the USB interface. If, for some reason, the computer does not have the correct USB driver, use the device driver provided on the product CD or from Technical Services. Reference COM Port USB Driver Alternatives, page 3-6.

Note:

Any monitoring history download relies on either factory default settings or institutional default settings established by a qualified service technician prior to usage. This includes baud rate and communication protocol selection.

To download monitoring history using the HyperTerminal™* program

1.Configure the monitoring system’s serial connectivity settings appropriately.

2.Connect the monitoring system’s mini-USB port to the computer.

Figure 3-1. Mini-USB Port

3.Execute the HyperTerminal™* program.

Note:

If this is the first time the HyperTerminal™* program launches, it will prompt the user to set it as the default Telnet program. Depending on institutional requirements, choose Yes or No.

4.Set appropriate port setting values for the HyperTerminal™* program:

a.Set the baud rate (bits per second) to match the monitoring system’s baud rate.

b.Ensure the data bit is set to 8.

c.Ensure the parity bit is set to none.

d.Ensure the stop bit is set to 1.

e.Ensure the flow control is set to off.

5.From the monitoring system’s Transfer Data menu, select Spot Data or Continuous Data.

Service Manual

3-3

Data Management

Figure 3-2. Transfer Data Type

6.Select By USB.

Figure 3-3. Transfer Data by USB

The data is transferred, and a progress bar is displayed. If desired, select Cancel to abort the transmission.

The Output Complete message is displayed when the transmission is complete.

3-4

Service Manual

External Data Communication

To interpret downloaded monitoring history

1.Examine monitoring history on the HyperTerminal™* program screen, in a spreadsheet, or on a printout from the personal computer.

Figure 3-4. Sample Monitoring History Printout

1

Product column headings

Data source, firmware version, and system settings

2

Patient data column headings

Lists appropriate time and data headings

3

Time column

Real-time clock date and time stamp

4

Output Complete

Message indicating completion of monitoring history

 

 

download

5

%SpO2

Current saturation value

6

PR

Current pulse rate

7

PA

Current pulse ampltitude

8

Status

Operating status of the monitoring system

2.Ensure patient data settings coincide with expected settings. This would include the version of firmware and its CRC code, which should be all zeros; alarm limit settings; patient mode; and SatSeconds™ setting.

3.Scan the time, SpO2, or PR column until reaching the events of interest.

4.Reference Table 3-1 on page 3-6 for descriptions of the operating status codes.

Service Manual

3-5

Data Management

 

Table 3-1. Monitoring Status Codes

 

 

 

Status

 

Description

code

 

 

 

 

 

LM

 

Loss of pulse, patient motion

 

 

 

LP

 

Loss of pulse

 

 

 

CB

 

Critically low battery

 

 

 

LB

 

Low battery

 

 

 

SO

 

Sensor off

 

 

 

SD

 

Sensor disconnect

 

 

 

AO

 

Alarm off

 

 

 

AS

 

Alarm silenced

 

 

 

MO

 

Signal interference, patient motion

 

 

 

PS

 

Pulse search

 

 

 

COM Port USB Driver Alternatives

Load the appropriate driver from the product CD into the connected computer.

Contact Technical Services or a local Covidien representative.

To install a USB driver from the compact disc

1.Insert the Nellcor™ portable SpO2 patient monitoring system compact disc (CD) into the designated personal computer (PC).

2.Copy the COVIDIEN USB to UART Bridge Driver zip file to the PC, installing it in the desired program folder.

3.Right-click on the zipped folder.

4.Select Extract All.

5.Open the extracted folder.

6.Launch the Driver Installer executable file.

Note:

To change the location of the driver, select the desired mapping by clicking Change Install Location.

7.Click Install.

3-6

Service Manual

Loading...
+ 82 hidden pages