For U.S.A.········································································································ (1)
For EU Countries····························································································· (1)
Labels and Markings on the Instrument ························································ (2)
Upright Stand Type Sensor Unit······································································ (2)
Universal Stand Type Sensor Unit··································································· (2)
Table Type Sensor Unit ··················································································· (3)
For U.S.A. and Canada ···················································································· (4)
For EU Countries····························································································· (4)
Labels and Markings on the Instrument ························································ (5)
Power Box ······································································································· (5)
Power Box/Auto-Collimation·········································································· (6)
Manual Control No. : BY8-2280-0E9
Name of Product :
Distribution Control No. :
Issued on :
CXDI-40EG/40EC
Service Manual Introduction
This service manual belongs to a series of after-sales guides Canon Inc. published as part of its
comprehensive product quality guarantee program, and will make a useful tool in promoting the sales
of the product, let alone repairing it.
This service manual consists of two chapters; “GENERAL”, “INSTALLATIONT”, “FUNCTION EXPLANETION”, “REPAIR GUIDE”, “TROUBLESHOOTING”, “PARTS CATALOG”,
“SERVICE MANUAL REPORT” and “TOOL”
If the product undergoes a large modification, a new service manual of revised edition will be sent to
you.
In other cases, service manual report will be sent to you to update the manual.
Note 1:
This service manual is published by Canon Inc. in accordance with Article 6 (Furnishing the
Referring Materials) of the Service Assignment Contract concluded with your company.
The contract prohibits the exposure of the contents of this service manual in any from to the
third party without a written consent of Canon Inc.
Note 2:
This service manual is property of Canon Inc. and the company may seek to have it returned,
depending on circumstances. You are expected to keep it until then.
Note 3:
Your inquiries, suggestions etc. about the contents of this service manual should be
addressed to:
Medical Products Technical Service Dept.
Canon Inc. Headquarters
30-2, Shimomaruko 3-chome, Ohta-ku, Tokyo 146-8501, Japan
CAUTION
Follow the safety instructions indicated below. Ignoring them may result in injury or accident.
1. Disassembly, Assembly, Adjustment and Maintenance
Disassembly, assembly, adjustment and maintenance must be done only by a service person who has
attended a service training designated by Canon.
2. Removal of Covers
Be sure to turn OFF the power of the instrument before removing the covers for maintenance and repair.
Also, do not touch the instrument with wet hands. Otherwise, you may get an electric shock that may
result in death or serious injury.
3. Fuse
When the fuse is going to be replaced, be sure to turn OFF the power of the instrument and solve the
problem which caused the fuse to blow. Be sure to replace the fuse with the specified type only.
Otherwise, fire or electric shock may result.
4. Ground Wire
Be sure to ground the instrument to an indoor grounded connector. Otherwise, fire or electric shock may
result due to leakage.
5. Modification
Never modify the product as it may result in fire or electric shock.
6. Moving Parts
This instrument incorporates moving parts. Do not perform any operation carelessly while the parts are
moving. Otherwise, you may get injured.
7. Interface Connector
When other equipment is going to be connected to the instrument using the connector for interface, be
sure to check after connection that leakage current is within the tolerable value.
i
VORSICHT
Befolgen Sie die unten angegebenen Sicherheitsanweisungen.
Mißachtung kann zu erletzungenoder Unfällen führen.
1.Zerlegung, Zusammenbau, Einstellung und Wartung
Zerlegung, Zusammenbau, Einstellung und Wartung dürfen nur von einem Wartungstechniker
durchgeführt werden, der an einem von Canon vorgeschriebenen Wartungslehrgang teilgenommen hat.
2.Entfernen von Abdeckungen
Schalten Sie unbedingt die Stromversorgung des Instruments aus, bevor Sie die Abdeckungen zwecks
Wartung und Reparatur entfernen.
Vermeiden Sie auch eine Berührung des Instruments mit nassen Händen.
Anderenfalls können Sie einen elektrischen Schlag erleiden, der zum Tod oder schwerer Verletzung
führen kann.
3.Sicherung
Wenn die Sicherung ausgewechselt werden muß, schalten Sie unbedingt die Stromversorgung des
Instruments aus, und beheben Sie die Ursache für das Durchbrennen der Sicherung.
Ersetzen Sie die Sicherung nur durch den vorgeschriebenen Typ.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag kommen.
4.Erdleiter
Erden Sie das Instrument unbedingt an einer Schukosteckdose.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag durch Leckstrom kommen.
5.Umbau
Jeder Umbau des Produktes ist strengstens untersagt, da dies zu einem Brand oder elektrischen Schlag
führen kann.
6.Bewegliche Teile
Dieses Instrument enthält bewegliche Teile.
Führen Sie während der Bewegung der Teile keine unachtsame Tätigkeit aus.
Anderenfalls können Sie verletzt werden.
7.Schnittstellenanschluß
Wenn andere Geräte über den Schnittstellenanschluß an das Instrument angeschlossen werden, prüfen Sie
nach dem Anschluß, daß der Leckstrom innerhalb des zulässigen
Bereichs liegt.
ii
Caution regarding the setup
According to the “IEC60601-1-1:2000”, devices installed in the patient environment are
restricted to “electric medical devices conforming to IEC60601-1”.
The Control PC, operation unit, and the magnetic card reader, etc. options that are parts of the
CXDI-C3S are classified under the data processing device standard (IEC60950), therefore
these items should not be installed in the patient environment.
Otherwise the Control PC is only classified in CXDI-C3S.
The patient environment described below is an example cited from the “IEC60601-1-1:2000”
– the measurements are only guidelines. However, the “IEC60601-1-1:2000” example must
be treated as the standar
Therefore, the CXDI-C3S must be installed in a location further than the measurements
below (outside of the patient environm
*Areas where the patient moves (not only during imaging but when entering and leaving
the room, etc.) are also considered as part of the patient environment, therefore the
installation location should be determined upon consultation with the user regarding
areas outside of the patient environment.
Example of patient environment
d.
ent).
Note: The measurements are only guidelines.
iii
Safety Information
Regulations
This instrument conforms to IEC 60601-1-2:2001.
For U.S.A. and Canada
This instrument is a CLASS I EQUIPMENT according to UL60601-1.
WITH RESPECT TO ELETRIC SHOCK, FIRE MECHANICAL
AND OTHER SPECIFIED HAZARDS ONLY IN
ACCORDANCE WITH CAN/CSA C22.2 NO. 601.1,
MEDICAL EQUIPMENT CERTIFIED FOR CANADA
MEDICAL EQUIPMENT WITH RESPECT
TO ELETRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH UL60601-1
<CONTROL NUMBER 41C4>
Grounding reliability can only be achieved when the equipment is connected to an equivalent
receptacle marked “Hospital Only” or “Hospital Grade”.
For EU Countries
The following mark shows compliance of the instrument with Directive 93/42/EEC.
This instrument has been classified into EN55011 Group 1/Class A.
This instrument is a CLASS I EQUIPMENT according to EN 60601-1.
(1)
Safety Information
Labels and Markings on the Instrument
The CXDI-40EG has a few labels and markings on it.
Contents of those and positions where they are attached are indicated below.
The illustrations below are examples of sensor units with the grid unit slot on the left hand side.
Upright Stand Type Sensor Unit
This mark indicates that this is a Type B
Applied Part according to UL60601-1
and EN60601-1.
This unit can be installed in the patient
environment.
Universal Stand Type Sensor Unit
This mark indicates that this is a Type B
Applied Part according to UL60601-1
and EN60601-1.
This unit can be installed in the patient
environment.
(2)
Table Type Sensor Unit
Safety Information
This mark indicates that this is a Type B
Applied Part according to UL60601-1
and EN60601-1.
This unit can be installed in the patient
environment.
(3)
Safety Information
Regulations
This instrument conforms to IEC 60601-1-2:2001.
For U.S.A. and Canada
This instrument is a CLASS I EQUIPMENT according to UL60601-1.
WITH RESPECT TO ELETRIC SHOCK, FIRE MECHANICAL
AND OTHER SPECIFIED HAZARDS ONLY IN
ACCORDANCE WITH CAN/CSA C22.2 NO. 601.1,
MEDICAL EQUIPMENT CERTIFIED FOR CANADA
MEDICAL EQUIPMENT WITH RESPECT
TO ELETRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH UL60601-1
<CONTROL NUMBER 41C4>
Grounding reliability can only be achieved when the equipment is connected to an equivalent
receptacle marked “Hospital Only” or “Hospital Grade”.
For EU Countries
The following mark shows compliance of the instrument with Directive 93/42/EEC.
This instrument has been classified into EN 55011 Group 1/Class A.
This instrument is a CLASS I EUIPMENT according to EN 60601-1.
(4)
Safety Information
Labels and Markings on the Instrument
The Power Box has a few labels and markings on it.
Contents of those and positions where they are attached are indicated below.
Name label
Rating label
(for CXDI-40EG)
Rating label
(for CXDI-40EC)
(5)
Labels and Markings on the Instrument
The Power Box / Auto-Collimation has a few labels and markings on it.
Contents of those and positions where they are attached are indicated below.
Name label
Rating label
(for CXDI-40EG)
Rating label
(for CXDI-40EC)
(6)
C
C
h
h
a
a
G
G
ptt
p
E
E
err
e
N
N
1
1
E
E
R
R
A
A
Printed by Canon
December 2007 Rev. 06
L
L
Canon Inc. Japan
Copyright(C) Canon Inc. Medical Technical Service Dept. All rights reserved.
In the former CXDI-40EG, Aux. Shift Register Driver (PCB-D-EP), Signal Read-out Circuit (PWB40EA) and A/D Conversion (PWB-40EAD) are located in both sides of LANMIT. Image data was
read out from two directions. However, in new CXDI-40EG, one set of Aux. Shift Register
Driver(PCB-D-EP), Signal Read-out Circuit (PWB-40EA) and A/D Conversion (PWB-40EAD) are
located in LANMIT. Image data was read out from one direction.
Since the appearance of new CXDI-40EG is the same as that of the former CXDI-40EG, they are
identified by the different serial number of the main unit.
CXDI System Software Ver.6.4 and later
Body No.
CXDI-40EC (revised read-out circuit)
In the former CXDI-40EC, Aux. Shift Register Driver (PCB-D-EP), Signal Read-out Circuit (PWB40EA) and A/D Conversion (PWB-40EAD) are located in both sides of LANMIT. Image data was
read out from two directions. However, in new CXDI-40EC, one set of Aux. Shift Register
Driver(PCB-D-EP), Signal Read-out Circuit (PWB-40EA) and A/D Conversion (PWB-40EAD) are
located in LANMIT. Image data was read out from one direction.
Since the appearance of new CXDI-40EC is the same as that of the former CXDI-40EC, they are
identified by the different serial number of the main unit.