Brooks Instrument Mfi User Manual

Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Brooks® Mf Series
NEMA 4X and IP66 Mass Flow Controllers and Meters
Mfx Series Controllers: Mf50x, Mf51x and Mf53x Meters: Mf60x, Mf61x and Mf63x
Mf61
Mass Flowmeter
Mf50
Mass Flow Controller
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
®
Mf Series
Part Number: 541B074AAG
August, 2009Brooks
Essential Instructions
Read this page before proceeding!
Brooks Instrument designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using and maintaining Brooks Products.
Read all instructions prior to installing, operating and servicing the product. If this instruction manual is not the correct manual, please see back cover for local sales office contact information. Save this instruction manual for future reference.
If you do not understand any of the instructions, contact your Brooks Instrument representative for clarification.
Follow all warnings, cautions and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation and maintenance of the product.
Install your equipment as specified in the installation instructions of the appropriate instruction manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
T o ensure proper performance, use qualified personnel to inst all, operate, update, program and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Brooks Instrument. Unauthorized parts and procedures can affect the product's performance and place the safe operation of your process at risk. Look-alike substitutions may result in fire, electrical hazards or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
Pressure Equipment Directive (PED)
All pressure equipment with an internal pressure greater than 0.5 bar (g) and a size larger than 25mm or 1" (inch) falls under the Pressure Equipment Directive (PED). The Directive is applicable within the European Economic Area (EU plus Norway, Iceland and Liechtenstein). Pressure equipment can be traded freely within this area once the PED has been complied with.
Section 1 of this manual contains important safety and operating instructions related to the PED directive.
Meters described in this manual are in compliance with EN directive 97/23/EC module H Conformity Assessment.
All Brooks Instrument Flowmeters fall under fluid group 1.
Meters larger than 25mm or 1" (inch) are in compliance with category I, II, III of PED.
Meters of 25mm or 1" (inch) or smaller are Sound Engineering Practice (SEP).
ESD (Electrostatic Discharge)
Handling Procedure:
1. Power to unit must be removed.
2. Personnel must be grounded, via a wrist strap or other safe, suitable means before any printed circuit card or other internal device is installed, removed or adjusted.
3. Printed circuit cards must be transported in a conductive container . Boards must not be removed from protective enclosure until immediately before installation. Removed boards must immediately be placed in protective container for transport, storage or return to factory .
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic designs contain components that utilize metal oxide technology (NMOS, SMOS, etc.). Experience has proven that even small amounts of static electricity can damage or destroy these devices. Damaged components, even though they appear to function properly, exhibit early failure.
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Dear Customer , We appreciate this opportunity to service your flow measurement and control requirements with a Brooks
Instrument device. Every day, flow customers all over the world turn to Brooks Instrument for solutions to their gas and liquid low-flow applications. Brooks provides an array of flow measurement and control products for various industries from biopharmaceuticals, oil and gas, fuel cell research and chemicals, to medical devices, analytical instrumentation, semiconductor manufacturing, and more.
The Brooks product you have just received is of the highest quality available, offering superior performance, reliability and value to the user. It is designed with the ever changing process conditions, accuracy requirements and hostile process environments in mind to provide you with a lifetime of dependable service.
We recommend that you read this manual in its entirety. Should you require any additional information concerning Brooks products and services, please contact your local Brooks Sales and Service Office listed on the back cover of this manual or visit www.BrooksInstrument.com
Y ours sincerely , Brooks Instrument
Brooks® Mf Series
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
®
Mf Series
Part Number: 541B074AAG
August, 2009Brooks
Mf Series Summary Installation Overview (shows cable gland option)
Terminal Strip Label:
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Paragraph Page Number Number
Section 1 Introduction
1-1 Description .................................................................................................................................1-1
1-2 Design Features.........................................................................................................................1-1
1-3 Principle of Operation.................................................................................................................1-1
1-4 Specifications .............................................................................................................................1-2
1-5 Electrical Specifications..............................................................................................................1-4
Section 2 Installation
2-1 General ......................................................................................................................................2-1
2-2 Receipt of Equipment.................................................................................................................2-1
2-3 Recommended Storage Practice ...............................................................................................2-1
2-4 Return Shipment ........................................................................................................................2-2
2-5 Transit Precaution ......................................................................................................................2-2
2-6 Removal from Storage ...............................................................................................................2-2
2-7 Gas Connections........................................................................................................................2-3
2-8 In-Line Filter ...............................................................................................................................2-3
2-9 Installation (Refer to Table 4-1, Figure 1-4 and Quick Start Instructions)...................................2-3
2-10 Electrical Interfacing (Electrical Specifications Refer to Section 1-5).........................................2-8
2-1 1 Configuring the PC Board ..........................................................................................................2-9
Brooks® Mf Series
Contents
Section 3 Operation
3-1 Theory of Operation ...................................................................................................................3-1
3-2 Operating Procedure..................................................................................................................3-2
3-3 Zero Adjustment.........................................................................................................................3-3
3-4 Calibration Procedure ................................................................................................................3-4
3-5 Response ...................................................................................................................................3-8
Section 4 Maintenance & Troubleshooting
4-1 General ......................................................................................................................................4-1
4-2 Troubleshooting..........................................................................................................................4-2
4-3 Sensor Tube...............................................................................................................................4-7
4-4 Disassembly and Assembly .......................................................................................................4-7
4-5 Gas Conversion Factors ...........................................................................................................4-18
4-6 Restrictor Sizing........................................................................................................................4-22
4-7 Orifice Sizing.............................................................................................................................4-25
Section 5 Parts List
5-1 General ......................................................................................................................................5-1
Section A CE Certificate
CE Certificate of Mass Flow Equipment ................................................................................................A-1
Warranty, Local Sales/Service Contact Information....................................................................... Back Cover
Figures
i
Contents Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
Brooks® Mf Series
Figure Page Number Number
1-1 Principle of Operation.................................................................................................................1-2
1-2 Maximum Allowable Output Loop Resistance............................................................................1-5
1-3 T erminal Strip Labeling...............................................................................................................1-6
1-4 Mf Series Typical Hookup to Brooks 0151i Power Supply..........................................................1-6
1-5 Model Mf50 Controller Dimensions ............................................................................................1-7
1-6 Model Mf51 Controller Dimensions ............................................................................................1-7
1-7 Model Mf53 Controller Dimensions ............................................................................................1-8
1-8 Model Mf53 Controller with Flange Dimensions.........................................................................1-8
1-9 Model Mf60 Meter Dimensions ..................................................................................................1-9
1-10 Model Mf61 Meter Dimensions ..................................................................................................1-9
1-11 Model Mf63 Meter Dimensions .................................................................................................1-10
1-12 Model Mf63 Meter with Flange Dimensions ..............................................................................1-10
2-1 Common Electrical Hookups......................................................................................................2-6
2-2 Recommended I/O Wiring Configuration for Current Signals (Non-Isolated Power Supply)......2-7
2-3 Recommended I/O Wiring Configuration for Current Signals (Isolated Power Supply) .............2-7
August, 2009
3-1 Flow Sensor Operational Diagram .............................................................................................3-2
3-2 Flow Control System Block Diagram..........................................................................................3-3
3-3 Mf Series Calibration Connections.............................................................................................3-5
3-4 Adjustment Potentiometer Location ...........................................................................................3-5
3-5 PC Board Jumper Location & Function ......................................................................................3-6
3-6 Fast Response Adjustment ........................................................................................................3-8
4-1 Torque Sequence for the Valve Ret ainer Plate (Controller)........................................................4-5
4-2 Voltmeter Connections for V alve Adjustment .............................................................................4-9
4-3a Valve Adjusting Sp acer Locations (Normally Closed Valve) .....................................................4-12
4-3b Valve Adjusting Sp acer Locations (Normally Open Valve)........................................................4-13
4-4 Model Mf51/61 Restrictor Element Assembly............................................................................4-24
4-5 Model Mf51/61 Restrictor Element Orientation in Controller/Meter Body..................................4-25
5-1 Model Mf50/60Parts Drawing ..................................................................................................... 5-2
5-2 Model Mf51/61 Parts Drawing ....................................................................................................5-3
5-3 Model Mf53/63 Parts Drawing ....................................................................................................5-4
ii
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Tables
Table Page Number Number
1-1 Flow Ranges ..............................................................................................................................1-2
1-2 Power Requirements..................................................................................................................1-4
1-3 Mf Series Controllers and Meters Terminal Strip Hookup ..........................................................1-5
2-1 Recommended Filter Size ..........................................................................................................2-3
4-1 Bench Troubleshooting ..............................................................................................................4-3
4-2 Sensor Troubleshooting .............................................................................................................4-4
4-3 Mf Series Torque Requirements................................................................................................4-10
4-4 Orifice Capacities ......................................................................................................................4-16
4-5 Conversion Factor( Nitrogen Base)...........................................................................................4-19
4-6 Model Mf50/60 Series Standard Restrictors..............................................................................4-22
4-7 Model Mf51/61 Restrictor Selection Guide................................................................................4-24
Brooks® Mf Series
Contents
5-1 Mf Series Common Replacement Parts.....................................................................................5-5
5-2 Model Mf50/60 Replacement Parts ............................................................................................5-6
5-3 Model Mf51/61 Replacement Parts ............................................................................................5-7
5-4 Model Mf53/63 Replacement Parts ............................................................................................5-8
5-5 Mf Series Tool and Spare Part Kits.............................................................................................5-9
iii
Contents Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
Brooks® Mf Series
August, 2009
THIS PAGE WAS INTENTIONALLY
LEFT BLANK
iv
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
1-1 Description
1-2 Design Features
Section 1 Introduction
Brooks® Mf Series
The Brooks® MfTM Series mass flow controllers and meters are specifically designed to be used in an industrial environment. These controllers and meters offer high accuracy, control and measurement of industrial gases with the added integrity of NEMA 4X, IP66 industrial p ackaging. The heart of the controller and meter is the flow sensor which produces an electrical output signal linear with mass flow rate. This output can be used for indicating, recording and/or control purposes. The Brooks Mf Series Mass Flow Controller has an integral valve and accepts a remote setpoint which makes it a simple and easy to install flow control system. Many options are offered to provide a versatile system of mass flow control and measurement.
• NEMA 4X, IP66 watertight construction
• UL and cUL listed and recognized for Class1 Division 2 hazardous locations (Mfx )
• Wide FS flow range for measurement and control for gases from 3 sccm to 1000 slpm N2 (control) and 1000 slpm N2 ( measurement)
Flow control to 1000 slpm N Flow metering to 1000 slpm N
• Sensor: Removable (Mfi) / Welded (Mfx)
• Insensitive to mounting attitude
• Selectable response time
• Electrically activated valve override
• Low command flow cutoff
• Corrosion resistant valve
2
2
1-3 Principle of Operation
The operating principle of the Brooks mass flow controller and meter is thermodynamic. A wire wound heating element directs heat to the midpoint of the bypass sensor tube. A predetermined portion of the total flow , flows through the bypass sensor tube. On the same tube, equidistant upstream and downstream of the heat input, are resistance temperature measuring elements.
With no flow, the heat reaching each temperature element is equal. With increasing flow the flow stream carries heat away from the upstream element, T1, and an increasing amount towards the downstream element, T2. An increasing temperature dif ference develop s between the two elements and this difference is proportional to the amount of gas flowing or the mass flow rate. A bridge circuit interprets the temperature difference
1-1
Section 1 Introduction
s s
Brooks® Mf Series
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
August, 2009
and positions the precision solenoid control valve. When the command signal is below 1% of full scale, the control valve is positioned to fully closed. The control valve can be latched fully open or closed by activating the valve override circuit.
To Power Supply
T1
Upstream
Temperature
Sensor
Heater
T2 Downstream Temperature
Sensor
Bypass
Sensor Tube
1-4 Specifications
Flow
+ Vdc
Bridge
for
T Detection
Δ
Amplifier
0-5 Vdc
and
4-20 mA
Figure 1-1 Principle of Operation.
WARNING
Do not operate this instrument in excess of the specification listed below. Failure to heed this warning can result in seriou personal injury and/or damage to the equipment.
PERFORMANCE CHARACTERISTICS:
Table 1-1 Flow Ranges
Mass Flow Mass Flow Flow Ranges PED
Controller Meter Nitrogen* Module H
Model Model Min. F. S. Max. F. S. Max. Press.
(slpm) psi (Bar)
Mf50 Mf60 0.003 30 1500 (100) SEP
N/A Mf61 0.003 30 4500 (300) Mf51 Mf63 10 100 1000 (6 8) Mf53 N/A 100 1000 1000 (68) 1 for all 150 lbs.Flanges
* Standard temperature and pressure equals 0°C and 101kPa (760 Torr).
These mass flow controllers and meters can be calibrated to other conditions. Specify at time of ordering.
1-2
Control/Usable Range
50 to 1
Accuracy
±1% full scale including linearity at calibrated conditions ±1.5% full scale including linearity for Mf50 and Mf60 for flow ranges greater than 20 slpm
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Section 1 Introduction
Brooks® Mf Series
Repeatability
0.25% of rate
Sensitivity to Mounting Attitude
±0.5% full scale maximum deviation from specified accuracy after re-zeroing under 200 psig. Specify mounting attitude at time of order to insure optimum performance.
Temperature Sensitivity
Zero: Less than ±0.075% F.S. per degree C Sp an: Less than ±1.0% F.S. shift from original calibration over 50°F to 122°F range (10°C to 50°C)
Pressure Sensitivity
±0.03% per psi up to 200 psig (N2)
Ratings: Operating Pressure
Refer to Table 1-1
Differential Pressure (Controllers)
Mf50: 5 to 50 psid pressure drop Mf51: 10 to 50 psid pressure drop Mf53: Standard high dif ferential valve > 30 to 290 p sid Optional: Low differential valve 7.5 to 30 psid (<500 slpm)
1 1.8 to 30 psid (>500 slpm)
Response Time
Less than 6 seconds to within 2% of full scale of final value for a 0 to 100% command change. Refer to Figure 1-2, typical performance curve.
Temperature Ambient/Gas
41°F to 149°F (5°C to 65°C)
Leak Integrity , Outboard
1 x 10-9 atm cc/sec. He (excluding permeation)
PHYSICAL:
Materials of Construction
St andard wetted parts: S tainless steel with V iton® fluoroelastomers or Buna-N Optional: Kalrez
®
Dimensions
Controllers: Refer to Figures 1-6 through 1-9 Meters: Refer to Figures 1-10 through 1-12
Process Connections
Refer to the table at the bottom of each dimensional drawing for process connection sizes and options specific to each size controller and meter .
1-3
Section 1 Introduction
V
V
*
Brooks® Mf Series
1-5 Electrical Specifications
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
August, 2009
Setpoint Command Requirements (Controllers)
4-20 mA (75 ohms input resist ance). The 4-20mA setpoint signal must be supplied from the customers side(sourcing type). The input load for this signal is 75 ohms. For 0 to 5 Vdc (220 K ohms input resistance)
Output Signals
0/4-20 mA, loop resistance is power supply dependent, refer to Figure 1-2, or 0 to 5 Vdc into 2000 ohms, or greater load. Maximum ripple 3 mV.
Power Requirements
Refer to Table 1-2.
Electrical Connections
Wire hookup is through a Pg1 1 water tight cable gland suitable for cable diameters of .20 to .39 inches or 1/2" FNPT conduit fitting. Wiring termination's are pluggable moving vise clamp with screw type terminations. Refer to Figure 1-3 for termination points and Table 1-3 for terminal identification and functions.
Typical Electrical Configuration
Figure 1-4 illustrates typical electrical hookup to Brooks 0151i Power Supply/Set Point Controller/Readout.
Table 1-2 Power Requirements
Mass Flow Controllers Mass Flow Meters
Model
Mf50 15 to 28 Vdc 240 mA @ 15 Vdc Mf60 15 to 28 Vdc 90 mA
Mf51*
Mf53
** Note minimum voltage for Mf51 is 22 Vdc
oltage Current Model
370 mA @ 28 Vdc
15 to 28 Vdc 309 mA @ 22 Vdc Mf61 15 to 28 Vdc 90 mA
370 mA @ 28 Vdc
15 to 28 Vdc 240 mA @ 15 Vdc
370 mA @ 28 Vdc
Mf63
oltage Current
15 to 28 Vdc 90 mA
1-4
Figure 1-2 Maximum Allowable Output Loop Resistance
Installation and Operation Manual
r
r
r
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Section 1 Introduction
Brooks® Mf Series
Table 1-3 Mf Series Controllers and Meters Terminal Strip Hookup
TB-1 Label Function Colo
Terminal 1-6 Identification Code**
1 VSUP Supply Voltage P l us (+) See Table 3 Orange 2 SUPCOM Supply Voltage Common Grn/Blk 3 NC Not Used Blue 4 SUPCOM Signal Common Org/Blk 5 VSIG Voltage Signal O utput White 6 ISIG Current Si gnal Output G reen
TB-2 Label Function Colo
Terminal 1-9 Identification Code**
1 NC Not Used Blu/Wht 2 GND Chasis Grou nd Grn/Wht 3 NC Not Used Red 4 VREF Reference Output +5 Vdc Blu/Blk 5 VOR Valve Override Input Blk/Wht* 6 NC Not Used Red/Wht 7 CMDCOM Command Common Black* 8 VCMD Voltage Command Input (Setpoint) Red/Blk* 9 ICMD Current Comman d Input (Setpoin t) Wht/Blk*
TB-3 Label Function Colo
Termin al 1 & 2 Identification Code**
1 N ONE Valve Hookup Orange* 2 N ONE Valve Hookup Orange*
*These connections used only for controllers (Models Mf50, Mf51 and Mf53) **Brooks reference
Certifications: EMC Directive 89/336/EEC:
Per EN 61326
Hazardous Location Classification
Non-Incendive Enclosure T ype 4X 1/IP66 Ambient T emperature: 50C > Tamb < 650C (T3C) or
50C > Tamb < 550C (T4)
United States and Canada
UL Listed: E73889 Volume 1, Section 17 UL Recognized: E73889 Volume 3, Section 1
Class I, Division 2, Groups A, B, C and D; Class II, Division 2, Groups F and G; Suitable for Class III, Division 2
Per UL 1604 and CSA-213
Pressure Equipment Directive (97/23/EC)
See Table 1-1 for further information
1-5
Section 1 Introduction
Brooks® Mf Series
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
August, 2009
TB-1
TB-2
Figure 1-3 Terminal Strip Labeling
VSUP+
SUPCOMNCSIGCOM
ICMD
VCMD
CMDCOMNCVOR
VSIG
ISIG
VREF
NC
GND
VALVE
NC
TB-3
Figure 1-4 Mf Series Typical Hookup to Brooks 0151i Power Supply
1-6
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Figure 1-5 Model Mf50 Controller Dimensions
Section 1 Introduction
Brooks® Mf Series
Connections "A" Dim.
9/16 -18 UNF 5.28/134 1/4" Tube Compr. 7.30/185 1/8" Tube Compr. 7.13/181
1/4" VCR 1/4" VCO
6mm Tube Compr. 7.32/186
TM TM
1/4" NPT 6.89/175
Dimensions in inches/mm
7.17/182
6.85/174
Figure 1-6 Model Mf51 Controller Dimensions
Connections "A" Dim.
9/16 -18 UNF 5.85/148.5 1/4" Tube Compr. 7.87/199.9 3/8" Tube Compr. 7.99/203.0
1/4" VCR 7.74/196.5 1/4" VCO 7.42/188.5
Dimensions in inches/mm
1-7
Section 1 Introduction
Brooks® Mf Series
Figure 1-7 Model Mf53 Controller Dimensions
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
August, 2009
Connections "A" Dim.
1/2" Tube Compr. 11.96/303.8 3/4" Tube Compr. 11.96/303.8
1" Tube Compr. 12.31/312.8
1/2" VCO 10.86/275.8 3/4" VCO 11.57/293.8
1/2" VCR 11.17/283.8
0.5", 0.1", 1.5" NPT or 1-1/16"-12 9.28/235.8
Dimensions in inches/mm
Figure 1-8 Model Mf53 Controller with Flange Dimensions
1-8
Flange Class
DIN DN14 PN 40 DIN DN25 PN 40 DIN DN40 PN 40 DIN DN50 PN 40
ANSI 0.5" 150 LBS ANSI 0.5" 300 LBS
ANSI 1" 150 LBS
ANSI 1" 300 LBS ANSI 1.5" 150 LBS ANSI 1.5" 300 LBS
ANSI 2" 150 LBS
ANSI 2" 300 LBS
Dimensions in inches/mm
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng Part Number: 541B074AAG August, 2009
Figure 1-9 Model Mf60 Meter Dimensions
Section 1 Introduction
Brooks® Mf Series
Connections "A" Dim.
9/16 -18 UNF 3.89/98.7 1/4" Tube Compr. 5.93/150.7 1/8" Tube Compr. 5.74/145.7
1/4" VCR 5.78/146.7
1/4" VCO 5.46/138.7
1/4" NPT 5.50/139.7
6 mm Tube Compr. 5.93/150.7
Dimensions in inches/mm
Figure 1-10 Model Mf61 Meter Dimensions
Connections "A" Dim.
9/16 -18 UNF 4.45/113.1 1/4" Tube Compr. 6.48/164.5 3/8" Tube Compr. 6.60/167.6
1/4" VCR 6.34/161 1/4" VCO 6.02/152.8
Dimensions in inches/mm
1-9
Section 1 Introduction
Brooks® Mf Series
Figure 1-11 Model Mf63 Meter Dimensions
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
August, 2009
Connections "A" Dim.
1/2" Tube Compr. 10.23/260 3/4" Tube Compr. 10.23/260
1" Tube Compr. 10.59/269
1/2" VCO 9.13/232 3/4" VCO 9.84/250 1/2" VCR 9.44/240
0.5", 1", 1.5" NPT or 1-1/16" - 12 7.55/192
Dimensions in inches/mm
Figure 1-12 Model Mf63 Meter with Flange Dimensions
1-10
Flange Class
DIN DN15 PN 40 DIN DN25 PN 40 DIN DN40 PN 40 DIN DN50 PN 40 ANSI 0.5" 150 LBS ANSI 0.5" 300 LBS
ANSI 1" 150 LBS
ANSI 1" 300 LBS ANSI 1.5" 150 LBS ANSI 1.5" 300 LBS
ANSI 2" 150 LBS
ANSI 2" 300 LBS
Dimensions in inches/mm
Installation and Operation Manual
X-TMF-Mfi-Mfx-eng Part Number: 541B074AAG August, 2009
2-1 General
2-2 Receipt of Equipment
Section 2 Installation
Brooks® Mf Series
This section contains the procedures for the receipt and installation of the instrument. See Section 1 for dimensional and connection requirements. Do not attempt to start the system until the instrument has been permanently installed. It is important that the start-up procedures be followed in the exact sequence presented.
When the instrument is received, the outside packing case should be checked for damage incurred during shipment. If the packing case is damaged, the local carrier should be notified at once regarding his liability . A report should be submitted to your nearest Product Service Department.
Brooks Instrument
407 W. V ine Street P.O. Box 903 Hatfield, PA 19440 USA Toll Free (888) 554 FLOW (3569) Tel (215) 362 3700 Fax (215) 362 3745 E-m ail: BrooksAm @BrooksInstrument.com www.BrooksInstrument.com
2-3 Recommended Storage Practice
Brooks Instrument Brooks Instrument
Neonstraat 3 1-4-4 Kitasuna Koto-Ku 6718 WX Ede, Netherlands Tokyo, 136-0073 Japan P.O. Box 428 Tel +81 (0) 3 5633 7100 6710 BK Ede, Netherlands Fax +81 (0) 3 5633 7101 Tel +31 (0) 318 549 300 Email: BrooksAs@BrooksInstrument.com Fax +31 (0) 318 549 309 E-mail: BrooksEu@BrooksInstrument.com
Remove the envelope containing the packing list. Carefully remove the instrument from the packing case. Make sure spare part s are not discarded with the packing materials. Inspect for damaged or missing parts.
If intermediate or long-term storage of equipment is required, it is recommended that the equipment be stored in accordance with the following conditions:
a. Within the original shipping container. b. Stored in a sheltered area, preferably a warm, dry, heated warehouse. c. Ambient temperature 21°C (70°F) nominal, 32°C (90°F) maximum,
7°C (45°F) minimum.
d. Relative humidity 45% nominal, 60% maximum, 25% minimum.
2-1
Section 2 Installation
Brooks® Mf Series
2-4 Return Shipment
Installation and Operation Manual
X-TMF-Mfi-Mfx-eng
Part Number: 541B074AAG
August, 2009
Prior to returning any instrument to the factory , contact your nearest Brooks location for a Return Materials Authorization Number (RMA#). This can be obtained from one of the following locations:
Brooks Instrument
407 W. V ine Street P.O. Box 903 Hatfield, PA 19440 USA Toll Free (888) 554 FLOW (3569) Tel (215) 362 3700 Fax (215) 362 3745 E-m ail: BrooksAm @BrooksInstrument.com www.BrooksInstrument.com
Brooks Instrument Brooks Instrument
Neonstraat 3 1-4-4 Kitasuna Koto-Ku 6718 WX Ede, Netherlands Tokyo, 136-0073 Japan P.O. Box 428 Tel +81 (0) 3 5633 7100 6710 BK Ede, Netherlands Fax +81 (0) 3 5633 7101 Tel +31 (0) 318 549 300 Email: BrooksAs@BrooksInstrument.com Fax +31 (0) 318 549 309 E-mail: BrooksEu@BrooksInstrument.com
2-5 Transit Precautions
2-6 Removal from Storage
Instrument must have been purged in accordance with the following:
All flow instruments returned to Brooks requires completion of Form RPR003-1, Brooks Instrument Decontamination S tatement, along with a Material Safety Data Sheet (MSDS) for the fluid(s) used in the instrument. Failure to provide this information will delay processing by Brooks personnel. Copies of these forms can be downloaded from the Brooks website www .BrooksInstrument.com or are available from any Brooks Instrument location listed above.
To safeguard the instrument against transportation damage, it is recommended to keep the instrument in its factory container until ready for installation.
2-2
Upon removal of the instrument from storage, a visual inspection should be conducted to verify its "as-received" condition. If the instrument has been subject to storage conditions in excess of those recommended (See Section 2-3), it should be subjected to a pneumatic pressure test in accordance with applicable vessel codes.
Installation and Operation Manual
X-TMF-Mfi-Mfx-eng Part Number: 541B074AAG August, 2009
2-7 Gas Connections
2-8 In-Line Filter
Section 2 Installation
Brooks® Mf Series
Refer to Tables 5-1 through 5-4 for the available process connection types and sizes. It is recommended that good tubing and piping practice be followed. It is also recommended that for very low flows in meter size Mf50's that tubing 1/4 or less be used and for flows over 10 slpm 3/8 tubing be applied. Prior to installation, make certain all piping is clean and free of obstructions. Install the piping in such a manner that permits easy removal if the instrument is to be removed for cleaning or test bench troubleshooting.
It is recommended that an Brooks in-line filter (refer to DS-5848 for appropriate models/flows) be installed upstream from the instrument to prevent the possibility of any foreign material entering the flow sensor or control valve. The filtering element should be periodically replaced or ultrasonically cleaned.
For Models Mf53/63, an in-line filter of 100 microns standard has been installed upstream from the meter or controller to prevent the possibility of any foreign material entering the flow sensor or control valve. The filtering element should be periodically replaced or ultrasonically cleaned.
Table 2-1 Recommended Filter Size
Maximum Flow Rate Recommended Filter Size
100 sccm 1 micron 500 sccm 2 micron
1 to 5 slpm 7 micron
10 to 30 slpm 15 micron
30 to 100 slpm 30 micron
Note: The above lists the max. recommended porosity for each flow range. It is recommended that the minimum micron porosity that does not limit the full scale flow rate be used.
2-9 Installation (Refer to Table 1-4, Figure 1-4, 2-1 and Quick Start Instructions)
CAUTION
When installing the controller, care should be taken that no foreign materials enter the inlet or outlet of the instrument. Do not remove the protective end caps until time of installation.
2-3
Section 2 Installation
Brooks® Mf Series
Installation and Operation Manual
X-TMF-Mfi-Mfx-eng
Part Number: 541B074AAG
August, 2009
Recommended installation procedures:
a. The Mf Series instrument should be located in an environment relatively
free from shock and vibration. b. Leave sufficient room for access to the electrical components. c. Install in such a manner that permits easy removal if the instrument
requires cleaning.
CAUTION
When used with a reactive (sometimes toxic) gas, contamination or corrosion may occur as a result of plumbing leaks or improper purging. Plumbing should be checked carefully for leaks and the controller purged with dry Nitrogen before use.
d. The Mf Series controllers and meters can be installed in any position.
However, mounting orientations other than the original factory calibra-
tion will result in a ±0.5% maximum full scale shift after re-zeroing. e. When installing controllers with full scale flow rates of 10 slpm or
greater, be aware that sharp abrupt angles in the system piping directly
upstream of the controller may cause a small shift in accuracy. If
possible, have at least ten pipe diameters of straight tubing upstream of
the Mf devices. Note: The control valve in the Mf controller provides precision control and
is not designed for positive shut-off. If positive shut-off is required, it is
recommended that a separate shut-off valve be installed in-line.
2-4
Special considerations to be taken when installing the Mf53 MFC:
The Model Mf53 has a valve design that is different from standard low flow Brooks TMFC’s. The Mf53 consists of a dual stage, pilot operated valve. The pilot valve (located on top of the MFC) controls a differential pressure across the main valve which, in turn controls the flow through the device. The main valve is a pressure operated valve that utilizes a bellows spring and diaphragm to control flow . This bellows and diaphragm assembly can be susceptible to damage by pressure spikes or surges. For this reason, it is recommended that process line startups are handled with care.
The bellows spring is offered in two levels. A low force for low differential pressures (delta P < 30psig), and a high force (delta P >30 and <300 psig).
The selection of the bellows spring is mainly determined by the differential pressure as specified on the customer order . This should reflect your actual process conditions. The low force bellows consists of a softer bellows spring which is required to allow flow control at lower differential pressures.
During startup conditions, when a process line is being pressurized, the
Installation and Operation Manual
X-TMF-Mfi-Mfx-eng Part Number: 541B074AAG August, 2009
Section 2 Installation
Brooks® Mf Series
pressures and/or pressure differentials that the Mf53 is exposed to may be different from the final process conditions. For higher pressure applications, and especially those with the low force bellows, it is important to bring the pressure up in a controlled manner in order to prevent a possible pressure spike to the bellows spring and main valve diaphragm. A pressure spike could deform the bellows, damage the diaphragm or blow out the bellows O-ring seal. This typically results in a failure to shutoff (leakby at zero setpoint).
One method to assure successful startups is to set a 100% setpoint command or valve override open command and then gently ramp the pressure up to operating conditions. This will allow you to bring your process pressures up to normal operating conditions and the Mf53 will then function as specified.
Another method is to utilize a bypass valve to allow pressure around the device while ramping up pressure to proper operating conditions.
The main point is to not instantly open a ball valve and allow a high upstream pressure or high back pressure surge into the Mf53 main valve. Proper process line venting is also important. If operating at pressures greater than 50 psig, be sure to perform a controlled pressure release from inlet and back pressure simultaneously in order to prevent bellows damage from excessive back pressure.
Following careful startup and venting procedures will contribute to a long problem free life of your Mf53 controller .
Stable Operating Conditions:
As stated above, the Mf53 model utilizes a pressure operated main valve. V alve performance is dependant on st able system pressures. Oscillating or unstable upstream or downstream pressures are likely to cause the device flow control to become unstable. For the best performance, it is important to create a stable pressure environment by utilizing quality inlet and back pressure regulators. In many cases, the addition of a back pressure regulator will isolate the Mf53 from the unstable downstream pressures inherent in many process designs.
For more information, please contact the Brooks Technical Service group.
2-5
Section 2 Installation
Brooks® Mf Series
Installation and Operation Manual
X-TMF-Mfi-Mfx-eng
Part Number: 541B074AAG
August, 2009
Figure 2-1 Common Electrical Hookups
2-6
Installation and Operation Manual
X-TMF-Mfi-Mfx-eng Part Number: 541B074AAG August, 2009
Section 2 Installation
Brooks® Mf Series
Figure 2-2 Recommended I/O Wiring Configuration for Current Signals (Non-Isolated Power Supply)
Figure 2-3 Recommended I/O Wiring Configuration for Current Signals (Isolated Power Supply)
2-7
Loading...
+ 57 hidden pages