Braun ThermoScan IRT 6030, ThermoScan IRT 6520, IRT 6030, IRT 6520 User Manual

ThermoScan
®
Infrared ear thermometer
IRT 6030 IRT 6520
百靈耳溫槍
This product is manufactured by Kaz Europe Sàrl under license to the ‘Braun’ trademark. ‘Braun’ is a registered trademark of Braun GmbH, Kronberg, Germany.
此產品由歐洲卡舒公司(Kaz Europe Sàrl)以百靈牌商標認可生產,   德國百靈公司(Braun GmbH, Kronberg, Germany)為百靈牌商標持有人。
ThermoScan is a registered trademark of Kaz Europe Sàrl ThermoScan是Kaz Europe Sàrl的註冊商標
Manufactured by Kaz Europe Sàrl Place Chauderon 18 CH-1003 Lausanne Switzerland © 2014, Kaz Europe Sàrl. All rights reserved.
Patents /專利:
www.kaz.com/patents/braun Designed in Germany, Made in Mexico.
0297
IRT 6030/6520AP
P/N: 31IM65AP190
13JUN14
English .................................1
繁體中文..............................14
2 3
English
The Braun ThermoScan® thermometer has been carefully developed for accurate, safe and fast temperature measurements in the ear.
The shape of the thermometer probe prevents it from being inserted too far into the ear canal which can hurt the eardrum.
However, as with any thermometer, proper technique is critical to obtaining accurate temperatures. Therefore, read the instructions carefully and thoroughly.
The Braun ThermoScan
®
thermometer is indicated for intermittent measurement and monitoring
of human body temperature for people of all ages. It is intended for household use only.
Use of IRT 6520 along with AgeSmart
feature is not intended as a substitute for consultation with
your physician.
Product description
IRT 6030
Lens filter (Box of 20)
Probe tip
Probe
ExacTemp light
Lens filter ejector
Display
Power button
Start button
Battery door
Protective case
1 2
3
4
5
6
7
8
9
10
1 2
3
4
5
6
7
8
9
11
10
IRT 6030
3
2
5
9
4
8
6
7
10
1
IRT 6520
3
2
5
10
4
9
6
7
8
11
1
WARNINGS AND PRECAUTIONS
Please consult your doctor if you see symptoms such as unexplained irritability, vomiting, diarrhea, dehydration, changes in appetite or activity, seizure, muscle pain, shivering, stiff neck, pain when urinating, etc. in spite of any color of the background light or absence of fever.
Please consult your doctor if the thermometer shows elevated temperature (yellow or red background lights for IRT 6520).
The operating ambient temperature range for this thermometer is 10 – 40 °C (50 – 104 °F). Do not expose the thermometer to temperature extremes (below –25 °C / –13 °F or over 55 °C / 131 °F) or excessive humidity (> 95 % RH). This thermometer must only be used with genuine Braun ThermoScan
®
lens filters (LF 40).
To avoid inaccurate measurements always use this thermometer with a new, clean lens filter attached.
If the thermometer is accidentally used without a lens filter attached, clean the lens (see «Care and cleaning» section). Keep lens filters out of reach of children.
IRT 6520
Lens filter (Box of 20)
Probe tip
Probe
ExacTemp light
Lens filter ejector
Display
AgeSmart and Memory buttons
Power button
Start button
Battery door
Protective case
4 5
English
Body temperature
Normal body temperature is a range. It varies by site of measurement, and it tends to decrease with age. It also varies from person to person and fluctuates throughout the day. Therefore, it is important to determine normal temperature ranges. This is easily done using Braun ThermoScan
®
. Practice taking temperatures on yourself and healthy
family members to determine the normal temperature range.
Note: When consulting your physician, communicate that the ThermoScan
®
temperature is a temperature measured in the ear and if possible, note the individual’s normal ThermoScan
®
temperature range as additional reference.
This thermometer is intended for household use only. This product is not intended to diagnose any disease, but is a useful screening tool for temperature. Use of this thermometer is not intended as a substitute for consultation with your physician.
The AgeSmart
TM
feature is not intended for pre-term babies or small-for-gestational age
babies. The AgeSmart™ feature is not intended to interpret hypothermic temperatures. Do not allow
children to take their temperatures unattended.
Do not modify this equipment without authorization of the manufacturer.
Parents/guardians should call the pediatrician upon noticing any unusual sign(s) or symptom(s). For example, a child who exhibits irritability, vomiting, diarrhea, dehydration, seizure, changes in appetite or activity, even in the absence of fever, or who exhibits a low temperature, may still need to receive medical attention.
Children who are on antibiotics, analgesics, or antipyretics should not be assessed solely on temperature readings to determine the severity of their illness.
Temperature elevation as indicated by AgeSmart™ may signal a serious illness, especially in adults who are old, frail, have a weakened immune system, or neonates and infants. Please seek professional advice immediately when there is a temperature elevation and if you are taking temperature on:
• Neonatesandinfantsunder3months(consultyourphysicianimmediatelyifthetemperature
exceeds 37.4 °C or 99.4 °F)
• Patientsover60yearsofage
• Patientshavingdiabetesmellitusoraweakenedimmunesystem(e.g.HIVpositive,cancer
chemotherapy, chronic steroid treatment, splenectomy)
• Patientswhoarebedridden(e.g.nursinghomepatient,stroke,chronicillness,recoveringfrom
surgery)
• Atransplantpatient(e.g.liver,heart,lung,kidney)
Fever may be blunted or even absent in elderly patients.
This thermometer contains small parts that can be swallowed or produce a choking hazard to children. Always keep the thermometer out of children’s reach.
AgeSmart™ color-coded display
Clinical research shows the definition of fever changes as newborns grow into little kids and little kids grow into adults
1
.
Braun’s AgeSmart
color-coded display takes the guess work out of interpreting
temperature for the entire family.
Simply use the AgeSmart™ button to select the appropriate age setting, take a temperature and the color-coded display will glow green, yellow, or red to help you understand the temperature reading. A green screen suggests no fever, a yellow screen suggests moderate fever, and red signals high fever.
6520
How does Braun ThermoScan® work?
Braun ThermoScan® measures the infrared heat generated by the eardrum and surrounding tissues. To help avoid inaccurate temperature measurements, the probe tip is warmed to a temperature close to that of the human body. When the Braun ThermoScan
®
is placed in the ear, it continuously monitors the infrared energy. The measurement is finished and the result displayed when the thermometer detects that an accurate temperature measurement has been taken.
Why measure in the ear?
The goal of thermometry is to measure core body temperature1 which is the temperature of the vital organs. Ear temperatures accurately reflect core body temperature
2
, since the eardrum shares blood supply with the temperature
control center in the brain
3
, the hypothalamus. Therefore, changes in body temperature are reflected sooner in the ear than at other sites. Axillary temperatures measure skin temperature and may not be a reliable indicator of core body temperature. Oral temperatures are influenced by drinking, eating and mouth breathing. Rectal temperatures often lag behind changes in core body temperature and there is a risk of cross-contamination.
1
How to use your Braun ThermoScan
®
Remove the thermometer from the protective case.
1. Guyton A C, Textbook of medical physiology, W.B. Saunders, Philadelphia, 1996, p 919
2. Guyton A C, Textbook of medical physiology, W.B. Saunders, Philadelphia, 1996, p 754-5
3. Netter H F, Atlas of Human Anatomy, Novartis Medical Education, East Hanover, NJ, 1997, pp 63, 95.
2
Push the Power button .
During an internal self-check, the display shows all segments. Then the last temperature taken will be displayed for 5 seconds.
1.HerzogL,PhillipsS.AddressingConcernsAboutFever.ClinicalPediatrics.2011;50(#5):383-390.
6 7
English
0-3 Mos >3-36 Mos
>36 Mos –
adult
3
4
5
6520
Fit the probe snuggly into the ear canal, then push and release the Start button
.
Select age with AgeSmart™ button as shown below. Press button to toggle through each age group.
NOTE: You must select an age in order to take a measurement.
The thermometer is ready to take temperature when the display looks like the images below.
The lens filter indicator will blink to signal one is needed.
To achieve accurate measurements, make sure a new, clean lens filter is in place before each measurement.
Attach a new lens filter by pushing the thermometer probe straight into the lens filter inside the box and then pulling out.
Note: The Braun ThermoScan
®
will not work
unless a lens filter is attached.
ExacTemp light will pulse while temperature is in progress. The light will remain solid for 3 seconds to indicate that a successful temperature reading has been achieved.
NOTE: If the probe has been properly inserted into the ear canal during the measurement, a long beep will sound to signal the completed measurement.
If the probe has NOT been constantly placed in a stable position in the ear canal, a sequence of short beeps will sound, the ExacTemp light will go out and the display will show an error message (POS = position error).
See «Errors and troubleshooting» section for more information.
6
7
9
65206030
The confirmation beep indicates that an accurate temperature measurement has been taken. The result is shown on the display.
Temperature displays.
0-3 3-36 36+
Months
0-3 3-36 36+
Months
0-3 3-36 36+
Months
Age range
0-3 Months
>3-36 Months
>36 Months ­adult
Green
Normal temperature
35.8 – 37.4 °C (96.4 - 99.4 °F)
35.4 – 37.6 °C (95.7 - 99.6 °F)
35.4 – 37.7 °C (95.7 – 99.9 °F)
Yellow
Moderate fever
>37.6 – 38.5 °C (>99.6 – 101.3 °F)
>37.7 – 39.4 °C (>99.9 – 103.0 °F)
Red
High fever
>37.4 °C (>99.4 °F)
>38.5 °C (>101.3 °F)
>39.4 °C (>103.3 °F)
For the next measurement, press Eject button to remove and discard used lens filter, and put on a new, clean lens filter.
NOTE: Defaults to last age setting used if you do not change.
The Braun ThermoScan
®
ear thermometer turns off automatically after 60 seconds of inactivity. The thermometer can also be turned off by pressing the Power button .
The display will briefly flash OFF and it will go blank.
8
Temperature taking hints
Always replace disposable lens filters to maintain accuracy and hygiene. The right ear measurement may differ from the measurement taken in the left ear. Therefore, always take the temperature in the same ear. The ear must be free from obstructions or excess earwax build-up to take an accurate reading.
External factors may influence ear temperatures, including when an individual has:
Factor Yes affects
Poor probe placement
Used lens filter
Dirty lens
60306520
Loading...
+ 10 hidden pages