Braun IRT 4520 User Manual [ko]

IRT 4520 IRT 4020
I / O
Type 6022 Type 6023
Type: 6022
1-800-327-7 226
ThermoScan
Braun Infolines
English 4
13
Internet:
www.braun.com
GB
IRL
0800 783 70 10
1 800 509 448
Manufactured by: Braun GmbH Frankfurter Str. 145 61476 Kronberg / Germany
6-022-435/00/VIII-06 GB/Korea Printed in Germany
IRT 4520
1
2
c
T
a
e
x
m
7
8
p
start
3 4
ThermoScan
5
I / O
6
mem
ThermoScan
b
a
p
m
E
e
T
x
c
a
Type: 6022
9
1-800-327
-
7
226
IRT 4020
1
2
3 4
5
c
T
a
e
x
m
7
8
p
start
ThermoScan
10
11
1-800-327
Type: 6022
-
9
7
226
English
The Braun ThermoScan thermometer has been carefully developed for accurate, safe and fast temperature measurements in the ear. The shape of the thermometer prevents it from being inserted too far into the ear canal to hurt the eardrum.
However, as with any thermometer, proper technique is critical to obtaining accurate temperatures. Therefore, read the use instructions carefully and thoroughly.
Important
• The operating ambient temperature range for this thermometer is 10–40 °C (50–104 °F).
• Do not expose the thermometer to temperature extremes (below –20 °C / –4 °F or over 50 °C / 122 °F) nor excessive humidity (> 95 % RH).
• This thermometer must only be used with genuine Braun ThermoScan Lens Filters (LF 40). Never use this thermometer without a new, clean lens filter attached.
• Keep lens filters out of reach of children.
• This thermometer is intended for household use only.
• Use of this thermometer is not intended as a substitute for con­sultation with your physician.
How does Braun ThermoScan work?
Braun ThermoScan measures the infrared heat generated by the eardrum and surrounding tissues. To avoid gross temperature dif­ferences, the sensor itself is brought to a temperature close to that of the human body. When the Braun ThermoScan is placed in the ear, it continuously monitors the infrared radiation. The measurement will be finished and the result is displayed, when an accurate reading can be assured.
Why measure in the ear?
The goal of thermometry is to measure core body temperature which is the temperature of the vital organs. Ear temperatures accurately reflect core body temperature, since the eardrum shares blood supply with the temperature control center in the brain, the hypothalamus. Therefore, changes in body temperature are reflected sooner and more accurately in the ear than at other sites.
• Axillary temperatures only measure skin temperature and therefore, are not a reliable indicator of core body temperature.
• Oral temperatures are influenced by drinking, eating and breathing.
• Rectal temperatures lag behind changes in core body temperature and there is a risk of cross-contamination.
4
Body temperature
Normal body temperature is a range. The following table shows that this normal range also varies by site. Therefore, readings from different sites should not be directly compared.
Normal ranges by site:
Axillary: 34.7 – 37.3 °C 94.5 – 99.1 °F
Oral: 35.5 – 37.5 °C 95.9 – 99.5 °F
Rectal: 36.6 – 38.0 °C 97.9 – 100.4 °F
ThermoScan: 35.8 – 38.0 °C 96.4 – 100.4 °F
• As well, a person’s normal temperature range tends to decrease with age. The following table shows normal ThermoScan ranges by age.
Normal ThermoScan ranges by age:
0 – 2 years 36.4 – 38.0 °C 97.5 – 100.4 °F
3 – 10 years 36.1 – 37.8 °C 97.0 – 100.0 °F
11 – 65 years 35.9 – 37.6 °C 96.6 – 99.7 °F
> 65 years 35.8 – 37.5 °C 96.4 – 99.5 °F
However, the range of normal also varies from person to person and fluctuates throughout the day. It is therefore important to deter­mine your normal temperature range. This is easily done using Braun ThermoScan. Practice taking temperatures on yourself and healthy family members to determine their normal temperature ranges.
Note: When consulting your physician, communicate that the ThermoScan temperature is a temperature measured in the ear and if possible, note the individual’s normal ThermoScan temperature range as additional reference.
5
Product description
1. Lens filter
2. Probe
3. Lens filter detector
4. Lens filter ejector
5. Display
6. «I/O» button (On/memory function – IRT 4520 only)
7. «ExacTemp» light
8. «start» button
9. Battery door
10. Protective cover (IRT 4520)
11. Cap (IRT 4020)
How to use your Braun ThermoScan?
1. To achieve accurate readings, make sure a new,
clean lens filter (1) is in place before each
start
measurement.
IRT 4020
a
c
T
e
m
p
IRT 4520
2. IRT 4020:
IRT 4520:
Push the «start» button (8). Push the «I/O» button (6).
During an internal self-check, the display shows
all segments. Then the last temperature taken will be displayed together with «MEM».
Wait for the ready signal beep and the ready symbol in the display.
3. Fit the probe snuggly into the ear canal, then push
the «start» button (8).
ThermoScan
n
e
T
c
a
x
E
6
If the probe has been fitted securely into the ear canal during the complete measuring process, a long beep will signal the end of the measuring process. You can be assured to have taken an accurate temperature reading. The result is shown on the display (5).
If you take the temperature of another person, the «ExacTemp» light (7) will be of help. It flashes during the measuring process as long as the probe is securely positioned, and lights up con­tinuously when an accurate reading has been taken.
4. If the probe has not been constantly placed in a stable position in the ear canal, a sequence of short beeps will sound, the «ExacTemp» light will go out and the display will show an error message («POS» = position error).
5. For the next measurement, eject the used lens filter (push ejector (4)) and put on a new, clean lens filter. IRT 4020:
Clear the display by pushing the
«start» button once.
IRT 4520: Clear the display by pushing the
«I/O» button once. Wait for the ready signal. Fit the probe snuggly into the ear canal, then push the «start» button.
The Braun ThermoScan ear thermometer turns off automatically after 60 seconds of inactivity. The IRT 4520 can also be turned off by pressing the «I/O» button for at least three seconds. The display will shortly flash «OFF» and after releasing the button it will go blank.
Temperature taking hints
• The right ear reading may differ from the reading taken at the left ear. Therefore, always take the temperature in the same ear.
• The ear must be free from obstructions or excess earwax build-up to take an accurate reading.
• External factors may influence ear temperatures, including when an individual has: – been lying on one ear or the other – had their ears covered – been exposed to very hot or very cold tempera-
tures, or
– been recently swimming or bathing.
n
e
T
c
a
x
E
7
In these cases, remove the individual from the situation and wait 20 minutes prior to taking a temperature.
• Use the untreated ear if prescription ear drops or other ear medications have been placed in the ear canal.
Memory mode
The last temperature taken is stored in its memory and will be automatically displayed when it is turned on again. The display will show «MEM».
IRT 4520: This model stores the last 8 temperature readings. To display the stored readings, the thermometer must be turned on. Then press the «I/O» button for at least 1 second. The display shows the memory number (e. g. MEM 1), and when releasing the «I/O» button, the stored temperature for that memory number is displayed, together with «MEM». If «I/O» button is pressed too long, the thermometer will be switched off. Each further pressing of the «I/O» button displays the remaining memory numbers (up to MEM 8). MEM 1 is the latest reading, MEM 8 would be the oldest. The memory mode is left automatically after displaying the oldest reading, or after pressing the «I/O» button for at least 1 second.
8
Loading...
+ 17 hidden pages