MMD 540H Hybrid Multimeter
Instruction Manual
540H Hybrid Multimeter | Table of Contents | 1
Contents
2 Introduction
3 Safety
7 Getting Started
8 Digital LCD Display 8 Front Panel Description 9 LCD Symbols
10 Push-button Functions
10 Mode Button 10 Manual range & Stroke 4/2,Hz,%,ms,CYL Button 10 Data Hold 11 Peak Hold 12 MAX/MIN 12 Backlight Button
13 Meter Functions
13 Voltage (V) 14 Resistance (Ω) 15 Diode Check( 16 Capacitance(CAP) 17 Audible Continuity ( 18 AC or DC Current (A) 19 Temperature (°C/°F) 20 Frequency (Hz) 21 Dwell ( 22 Duty Cycle (%) 23 Ms – PULSE. 25 RPM ( )
)
)
)
26 General Specifi cations
27 Electrical Specifi cations
30 Replacing the Battery 30 Battery Installation 31 Replacing the Fuses 32 Trouble Shooting
33 Warranty Information
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
2 | MMD 540H Hybrid Multimeter | Introduction
This chapter covers brief, introductory information, You will fi nd:
Safety precautionsÑ
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Safety | 3
Safety
DANGER
Engines produce carbon monoxide which
Ñ
is odorless, causes slower reaction time, and can lead to serious injury. When the engine is operating, keep service areas WELL VENTILATED or attach the vehicle exhaust system to the shop exhaust removal system.
Set the parking brake and block the
Ñ
wheels before testing or repairing the vehicle. It is especially important to block the wheels on front-wheel drive vehicles; the parking brake does not hold the drive wheels.
Wear an eye shield when testing or
Ñ
repairing vehicles.
Exceeding the limits of this meter is dangerous. This will expose you to serious or possibly fatal injury. Carefully read and understand the cautions and the specifi cation limits of this meter.
Voltage between any terminal and ground must not exceed 600V
Ñ
DC or 600V AC.
Use caution when measuring voltage above 25V DC or 25V AC.
Ñ
Circuit tested must be protected by a 20A fuse or circuit breaker.
Ñ
Do not use the meter if it has been damaged.
Ñ
Do not use the test leads if the insulation is damaged or if metal
Ñ
is exposed.
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
4 | MMD 540H Hybrid Multimeter | Safety
Safety Cont’d…
DANGER
Avoid electrical shock: do not touch the
Ñ
test leads, tips or the circuit being tested.
Do not try a voltage measurement with the
Ñ
test leads in the 20A or the mA terminal.
When testing for the presence of voltage or current, make sure
Ñ
the meter is functioning correctly. Take a reading of a known voltage or current before accepting a zero reading.
Ñ
Choose the proper range and function for the measurement. Do not try voltage or current measurements that may exceed the ratings marked on the Function/Range switch or terminal.
Ñ
When measuring current, connect the meter in series with the load.
Ñ
Never connect more than one set of test leads to the meter.
Ñ
Disconnect the live test lead before disconnecting the common test lead.
Ñ
The mA and the 20A terminals are protected by fuses. To avoid possible injury or damage, use only in circuits limited to 400mA or 20A for 30 seconds.
See also…
Fuse Replacement
Ñ
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Safety | 5
Safety Cont’d…
IMPORTANT
To maintain accuracy of the meter, replace the discharged
Ñ
battery immediately when the symbol “BAT” appears on the meter display.
Ñ
Avoid measuring error from outside interference. Keep the meter away from spark plug or coil wires.
Ñ
Avoid damaging the meter when testing voltage. Disconnect the test leads from the test points before changing functions.
Ñ
Do not exceed the limits shown in the table below:
Function Terminal Input limit
DC/AC Volts, Ohm†/Continuity /Diode,CAP.,IR- TEMP. Adaptor, Type-K TEMP.,Hz,%Duty, Ms Dwell, RPM
AC/DC μA mA μA / mA 400mA DC/AC AC/DC20A 20A *20A DC/AC
V-Ω-RPM 600VoltsAC DC
*20 Amp measurement for 30 seconds maximum.
†Ohms can not be measured if voltage is present, ohms can be measured only in a non-powered circuit. However, the meter is protected to 600 volts.
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
6 | MMD 540H Hybrid Multimeter | Notes
Notes:
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Getting S
tarted | 7
Getting Started
This chapter will help you get started. It describes the basic functions of the Meter.
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
8 | MMD 540H Hybrid Multimeter | Meter Basics
Meter Basics
1. Digital LCD display
Display features:
1a. Symbols to identify function 1b. Four character digital
display The digital display is best for stable inputs. The bar graph is best for rapidly changing inputs.
2. Function buttons
Press the button to select a function. A symbol will display to verify your choice.
3. Rotary Selector Switch
Turn this switch to select a function or turn the meter OFF.
4. Volts Terminal
The Red test lead is used to measure; Amps, Volts, Ohms, TEMP, Hz, RPM, Cap, mS and Dwell.
The Black test lead is used in the Common(COM) terminal for all tests.
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | LCD Display Symbols | 9
LCD Display Symbols and Annunciators
-8.8.8.8 LCD digits
Continuity
BAT Low Battery
Diode
DATA HOLD Data Hold
AUTO AutoRanging
AC Alternating Current or Voltage
DC Direct Current or Voltage
-9
n nano (10
μ micro (10
m milli (10
) (capacitance)
-6
) (amps, cap)
-3
) (volts, amps)
A Amps
3
k kilo (10
) (ohms)
F Farads (capacitance)
6
M mega (10
) (ohms)
Ω Ohms
V Volts
Hz Hertz (frequency)
% Percent (duty ratio)
ºF Degrees Fahrenheit
ºC Degrees Centigrade
IR TEMP Infrared Temperature
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
10 | MMD 540H Hybrid Multimeter | Push-button Functions
Push–button Functions
Mode Button
Press the mode pushbutton to select the following functions; DC/AC Voltage, DC/AC
Current Resistance, Diode, Continuity & Capacitance
Manual Range & Stroke 4/2(DIS), z, %, ms +, CYL Button
Press to this button to select: STROKE 4, 2DIS, Hz ,%, ms+ , CYL range & V/A/ Resistance manual Range
Manual Ranging
The meter turns on in the autoranging mode.
Press the Range button to go to manual ranging.
The display icon “ range button will step to the next range as indicated by the units and decimal point location. Press and hold the Range button for two seconds to return to autoranging.
” will appear. Each press of the
Data Hold
The Data Hold function allows the meter to “freeze” a measurement for later reference.
Press the DATA HOLD button to
1.
“freeze” the reading on the indicator. The indication of “HOLD” will appear in the display.
Press the DATA HOLD button again to return to
2.
normal operation.
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Push-button Functions | 11
Peak Hold
The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture negative or positive peaks as fast as 1 millisecond in duration.
Set the function switch to the Amps or
1. Volts.
Press and Hold the PEAK button until
2. “CAL” appears in the display. This procedure will zero the range selected & the meter will go into manual ranging mode.
Press the PEAK button, Pmax will display.
3.
The display will update each time a higher positive peak occurs.
4.
Press the PEAK button again, Pmin will display. The display will
5. now update and indicate the lowest negative peak.
To return to normal operation, press and hold the PEAK button
6. until the Pmin or Pmax indicator switches off.
NOTE: If the Function switch position is changed after a calibration the Peak Hold calibration must be repeated for the new function selected.
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
12 | MMD 540H Hybrid Multimeter | Push-button Functions
MAX/MIN
Press the MAX/MIN key to activate the
1. MAX/MIN recording mode. The display icon “MAX or MIN” will appear. The meter will go into manual ranging mode & display and hold the maximum or minimum reading and will update only when a new “max or min” occurs.
2.
Press the MAX/MIN key and a blinking “MAX MIN” will appear. The meter will display the present reading, but will continue to update and store the max and min readings.
To exit MAX/MIN mode press and hold the MAX/MIN key for 2 seconds.
Backlight Button
Press the BACKLIGHT button to turn
1. the backlighting ON.
Press the BACKLIGHT button again to
2. turn the backlighting OFF.
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Meter Functions | 13
Meter Functions – Voltage (V)
Select the Volts V” range with
Ñ
the rotary switch.
The meter will automatically
Ñ
select the best voltage (V) range.
Select DCV or ACV with the
Ñ
MODE button.
Insert:
Black lead in COM terminal.
Ñ
Ñ
Red lead in V-Ω-RPM terminal
Touch the Black probe to ground or to the negative (-) circuit
Touch the Red probe to the circuit coming from the power source
IMPORTANT: Voltage must be measured in parallel (Red probe measuring circuit from power source).
Black– Red+
WARNING
When measuring voltage, be sure the Red test lead is in the terminal marked “V”. If the test lead is in an Amp (A) or Milliamp (mA) terminal, the meter can be damaged and risk of personal injury.
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
14 | MMD 540H Hybrid Multime
ter | Meter Functions
Meter Functions – Resistance (Ω)
IMPORTANT: If you are testing an application that has capacitors in the circuit, be sure to turn the power OFF on the test circuit and discharge all capacitors. Accurate measurement is not possible if external or residual voltage is present.
Select the Resistance
Ñ
” range with the rotary switch.
Select the Resistance
Ñ
function with the Mode button. Please note, the unit defaults to resistance.
Insert:
Black lead in COM
Ñ
terminal.
Red lead in V-Ω-RPM
Ñ
terminal.
Touch the test lead probes across the resistor to be tested.
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
Black– Red+
MMD 540H Hybrid Multimeter | Meter Functions | 15
Meter Functions – Diode Check ( )
IMPORTANT: Turn the power OFF to the test circuit
Select the Diode Check
Ñ
” function with the rotary switch and mode button.
Insert:
Black lead in COM terminal.
Ñ
Red lead in V-Ω-RPM terminal.
Ñ
Touch the Black test probe to the negative (-) side of the diode.
Touch the Red test probe to the positive (+) side of the diode. Reverse the probes: Black to the positive (+) side and Red to the negative (-) side.
NOTE: A “good” diode will read low in one `direction and high in the other direction when the probes are reversed (or vice versa).
A defective diode will have the same reading in both directions or read between 1.0 to 3.0 V. in both directions.
Black– Red+
Reverse Probes
Diode - to +
Good
Bad
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
.4 to .9V OL OL .4 to .9V OL 1.0 to 3.0V
1.0 to 3.0V OL .4 to .9V .4 to .9V OL OL .000V .000V
+ to -
16 | MMD 540H Hybrid Multime
ter | Meter Functions
Meter Functions – Capacitance (CAP)
IMPORTANT: Turn the power OFF to the test circuit
Select the Capacitance “CAP
Ñ
function with the rotary switch and mode button
Insert:
Black lead in COM terminal.
Ñ
Red lead in V-Ω-RPM terminal
Ñ
CAUTION: When checking in­circuit capacitance, be sure that the circuit has all power removed and all capacitors are fully discharged.
Touch the test lead probes across the capacitance circuit to be tested. Read the measured value from the LCD display.
NOTE:
The bar graph is disabled in
1.
capacitance measurement mode. However, since the measurement time of 4mF and 40mF modes is quite long (3.75s and 7.5s respectively, to be precise), the bar graph is instead used to display the time required to accomplish the measurement.
In order to obtain an accurate reading, a capacitor must be
2.
discharged before measurement begins. The meter has a built­in discharge mode to automatically discharges the capacitor. In discharge mode, the LCD displays “DIS.C”
Discharging through the chip is quite slow. We recommend the
3.
user to discharge the capacitor with some other apparatus.
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
Black– Red+
MMD 540H Hybrid Multimeter | Meter Functions | 17
Meter Functions – Audible Continuity ( )
IMPORTANT: Turn the power OFF on the test circuit
Select the Audible Continuity
Ñ
” function with the rotary
“ switch and mode button.
Insert:
Black lead in COM terminal.
Ñ
Red lead in V-Ω-RPM terminal.
Ñ
Connect one test probe to each end of the circuit to be tested.
Circuit complete, the meter
Ñ
will beep continuously.
Circuit open, there is no beep
Ñ
and the display shows to OL (over limit).
Black– Red+
Circuit open
- no beep
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
18 | MMD 540H Hybrid Multimeter | Meter Functions
Meter Functions – AC or DC Current (A)
IMPORTANT: All current measured fl ows through the meter.
It is important that you do not:
Measure current greater than 600
1. Volts AC or DC, with respect to ground.
Do Not Exceed 30 seconds when
2. measuring continuous current between 1A-20A. Allow fi ve minutes for cool down before continuing.
Select the “20A” or “mA” range with
Ñ
the rotary switch.
Press the Mode button to select AC or
Ñ
DC.
Insert:
Black lead in COM terminal.
Ñ
Round ide
+
Black–Red
Red lead in the 20A or mA terminal
Ñ
(select 20A if you are unsure of the current draw).
IMPORTANT: Turn OFF all power to the circuit or disconnect the circuit from the power source.
Connect:
The Red probe to the side of the circuit closest to the power
Ñ
source.
The Black probe to the side of the circuit to ground.
Ñ
Turn the power ON and test.
Ñ
NOTE: Current must always be measured with the meter test probes connected in series, as described
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Meter Functions | 19
Meter Functions – Temperature (°C/°F )
IMPORTANT: To avoid heat damage to the meter, keep it away from sources of very high temperature. The life of the Temperature Probe is also reduced when subjected to very high temperatures. Probe operating range is –58° to 482°F.
Select the Temper ature
Ñ
°C or °F” function with the rotary switch.
Insert the temperature
Ñ
probe connector into the K-type thermocouple adapter. Insert the adapter into the front of the meter as shown.
Temperature Sensor
emperature
T Probe
Touch the end of the temperature sensor to the area or surface
Ñ
of the object to be measured.
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
20 | MMD 540H Hybrid Multime
ter | Meter Functions
Meter Functions – Frequency(Hz)
Select the “HzFrequency
Ñ
function with the rotary switch.
Insert:
Black lead in COM terminal.
Ñ
Red lead in V-Ω-RPM
Ñ
terminal.
Connect the Black test probe to ground.
Connect the Red test probe to the “signal out” wire of the sensor to be tested.
Black– Red+
“Signal O ut” Side
“Volta ge In” Side
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
Ground Side
MMD 540H Hybrid Multimeter | Meter Functions | 21
Meter Functions – Dwell ( )
Select the “DWELL” function
Ñ
with the rotary switch.
Insert:
Black lead in COM terminal.
Ñ
Red lead in V-Ω-RPM terminal.
Ñ
Connect the Black test probe to ground.
Connect the Red test probe to the wire that connects to the breaker points (see illustration).
Black– Red+
To br eake r Points
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
22 | MMD 540H Hybrid Multime
ter | Meter Functions
Meter Functions – Duty Cycle (%)
Select the “%DUTY”
Ñ
Cycle function with the rotary switch.
Insert:
Black lead in COM
Ñ
terminal.
Red lead in V-Ω-RPM
Ñ
terminal.
Connect the Black test probe to ground.
Connect the Red test probe to the signal wire circuit.
The illustration for a mixture control solenoid is shown with the metering rod in the closed position. The meter will display the percentage of time the plunger is in the closed position. (low duty cycle) during one duty cycle.
Black– Red+
Mixture Control Shown
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Met
er Functions | 23
Meter Functions – ms-PULSE (Pulse Width) & ms-PERIOD (Period)
Pulse Width is the length of time an actuator is energized. For example, fuel injectors are activated by an electronic pulse from the Engine Control Module (ECM).
This pulse generates a magnetic fi eld that pulls the injectors nozzle valve open.
The pulse ends and the injector nozzle is closed.
This open to close time is the Pulse Width and is measured in milliseconds( ms).
The most common automotive application for measuring pulse width is on fuel injectors.
You can also measure the pulse width of the fuel mixture control solenoid and the idle air control motor.
Measuring Pulse Width on Port Fuel injectors is described on the following page.
(continued on next page)
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
24 | MMD 540H Hybrid Multimeter | Meter Functions
(continued from previous page)
To measure pulse width (mS):
Select the “mS-Pulse
Ñ
function with the rotary switch .
Press the ±TRIG button for
Ñ
2 seconds until the negative (-) trigger slope is displayed on the upper left side of the display.
NOTE: The applied time for most fuel injectors is displayed on the negative (-) slope.
Insert:
Black lead in COM jack.
Ñ
Red lead in V-Ω-RPM jack.
Ñ
Connect:
Jumper wires between the fuel injector and the harness connector.
Black test probe to a good
Black– Red+
Mixture Control Shown
Jumper Leads From Port Fuel Injector
Good Ground
ground at the fuel injector or the negative (-) vehicle battery post.
Red test probe to the fuel injector solenoid driver input on the jumper cable.
Start the engine. A pulse width in milliseconds should be read.
NOTE: Initially, the unit will read “OL”, then readings will descend and stabilize to the actual pulse width. If “OL” remains, re-check your connections.
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Meter Functions | 25
Meter Functions – RPM/×10RPM
Select the RPM range with the rotary
Ñ
switch.
OR
Select the×10RPM range with the rotary
Ñ
switch (1,000 to 12,000 RPM). Multiply the displayed reading by ten to get actual RPM.
Press STROKE through RPM 2-stroke and DIS ignitions.
Insert the inductive pickup connecting terminals into the meter.
Black lead in COM terminal.
Ñ
Red lead in V-Ω-RPM terminal.
Ñ
Connect the inductive pickup to a spark plug wire. If no reading is received, unhook the clamp, turn it over and connect again.
NOTE:
Position the inductive pick-up as far away from the distributor
Ñ
and the exhaust manifold as possible.
Position the inductive pick-up to within six inches of the spark
Ñ
plug or move it to another plug wire if no reading or an erratic reading is received.
RPM 4: For RPM of 4-stroke engines which have 1 ignition on every
4 engine strokes
/ DIS button to select
for 4-stroke, RPM for
RPM 2: For RPM of DIS ( Distributorless lgnition System) & 2-stroke
engines which Have 1 ignition on every 2 engine strokes
NOTE: The rpm pick-up has an adjustable sensitivity switch that can also be used to correct an unstable reading.
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
26 | MMD 540H Hybrid Multime
ter | General Specifi cations
General Specifi cations
The instrument complies with: IEC 1010-1 EN61010-1.
Insulation: Class2, Double insulation.
Overvoltage category: CATIII1000V/CATIV600V.
Display: 4000 counts LCD display with function indication.
Polarity: Automatic, (-) negative polarity indication.
Overrange: “OL” mark indication.
Low battery indication: The “BAT” is displayed when the battery
voltage drops below the operating level.
Measurement rate: 2 times per second, nominal.
Auto power off: Meter automatically shuts down after approx. 30
minutes of inactivity.
Operating environment: 0° C to 50° C (32° F to 122° F) at < 70 % relative humidity.
Storage temperature: -20° C to 60° C (-4° F to 140° F) at < 80 % relative humidity.
For inside use, max height: 2000m
Pollution degree: 2
Power: One 9V battery , NEDA 1604, IEC 6F22.
Dimensions: 182 (H) x 82 (W) x55 (D) mm
Weight Approx.: 375g.
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Electrical Specifi cations | 27
Electrical Specifi cations
*Accuracy is given as±([% of reading]+[number of least signifi cant digits]) at 18° to 28°(65°F to 83°F), with relative humidity up to 70%.
RPM (Tach)
Range Resolution Accuracy
RPM 4 600~4000 RPM 1 RPM
1000~12000 RPM (X10 RPM)
RPM 2/DIS 300~4000 RPM 1 RPM
1000~6000 RPM (X10 RPM)
Effect Reading: >600 RPM
Dwell Angle
Cylinder Range Resolution Accuracy
4CYL 0~90.0° 0.1°
5CYL 0~72.0°
6CYL 0~60.0°
8CYL 0~45.0°
10 RPM
±2 % of rdg ±4 dgts
10 RPM
±2.0% of rdg ±4 dgts
DC Voltage
Range Resolution Accuracy
400.0mV 0.1mV
4.000V 1mV
40.00V 10mV
400.0V 100mV
600V 1V
Input Impedance: 10MΩ
±0.5% of rdg ±2 dgts ±1.5% of rdg ±2 dgts
±1.8% of rdg ±2 dgts
AC Voltage
Range Resolution Accuracy
4.000V 1mV
40.00V 10mV
400.0V 100mV
600V 1V
Input Impedance: 10MΩ.
Frequency Range: 50 to 60Hz
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC
±1.0% of rdg ±3 dgts ±1.5% of rdg ±3 dgts
±2.0% of rdg ±4 dgts
28 | MMD 540H Hybrid Multimeter | Electrical Specifi cations
DC Current
Range Resolution Accuracy
40.00mA 10uA ±1.5% of rdg ±3 dgts
400.0mA 100uA
20A 10mA ±2.5% of rdg ±5 dgts
Overload Protection: 0.5A / 250V and 20A / 250V Fuse.
Maximum Input: 400mA dc or 400mA ac rms on uA / mA ranges,20A dc or ac rms on 20A range.
AC Current
Range Resolution Accuracy
40.00mA 10uA ±1.8% of rdg ±5 dgts
400.0mA 100uA
20A 10mA ±3.0% of rdg ±7 dgts
Overload Protection: 0.5A / 250V and 20A / 250V Fuse.
Frequency Range: 50 to 60 Hz
Maximum Input: 400mA dc or 400mA ac rms on uA / mA ranges, 20A dc or ac rms on 20A range.
Resistance
Range Resolution Accuracy
400.0Ω 0.1Ω ±1.2% of rdg ±4 dgts
4.000kΩ 1Ω ±1.0% of rdg ±2 dgts
40.00kΩ 10Ω ±1.2% of rdg ±2 dgts
400.0kΩ 100Ω
4.000MΩ 1kΩ
40.00MΩ 10kΩ ±2.0% of rdg ±3 dgts
Capacitance
Range Resolution Accuracy
4.000nF 1pF ±5.0% of rdg±50 dgts
40.00nF 10pF ±5.0% of rdg ±7 dgts
400.0nF 0.1nF ±3.0% of rdg ±5 dgts
4.000uF 1nF
40.00uF 10nF
400.0uF 0.1uF
4.000mF 0.001mF ±10% of rdg ±10 dgts
40.00mF 10.00mF
Robert Bosch LLC MMD 540H Instruction Manual | Version 1 | February 2008
MMD 540H Hybrid Multimeter | Electrical Specifi cations | 29
Frequency
Range Resolution Sensitivity Accuracy
4.000kHz 1Hz >5V RMS ±1.5% of rdg ±3 dgts
40.00kHz 10Hz
400.0kHz 100Hz
40.00MHz 1kHz >10V RMS ±2.0% of rdg ±4 dgts
Duty Cycle
Range Resolution Accuracy
0.1%~99.9% 0.1% ±2% of rdg ±2 dgts
Pulse width: >100us, <100ms;
Frequency width: 5Hz – 100kHz
Sensitivity: >5V RMS
Pulse Width
Range Resolution Accuracy
1.0~ 20.0ms 0.1ms ±2% of rdg ±20 dgts
Type-K Temperature
Range Resolution Accuracy
-30° C~ 1000° C 1° C ±3% of rdg ±5° C/8° F (Meter
-22° F~1832° F 1° F
Sensor: Type K Thermocouple
only, probe accuracy not included)
Diode Test
Test current Resolution Accuracy
1.0mA typical 1 mV ±5% of rdg ±15 dgts
Open circuit voltage: 3.0V dc typical
Audible continuity
Audible threshold: Less than35Ω Test current: <1mA dc typical
MMD 540H Instruction Manual | Version 1 | February 2008 Robert Bosch LLC