The following safety precautions apply to both operating and maintenance personnel and must
be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in this manual.
Apply all the listed safety precautions.
Verify that the voltage selector at the line power cord input is set to the correct line voltage.
Operating the instrument at an incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power.
Do not operate the instrument in ways not specified by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the
safety standards of design, manufacture, and intended use of the instrument. B&K Precision
assumes no liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard defines safety category ratings that specify the amount of electrical
energy available and the voltage impulses that may occur on electrical conductors associated
with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating
is also accompanied by a maximum voltage of the circuit to be tested, which defines the voltage
impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to
be connected to the mains supply. The voltages in the environment are typically derived from a
limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be
connected to the mains supply at a standard wall outlet or similar sources. Example
measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be
connected to the mains installation of a building. Examples are measurements inside a
building's circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be
connected to the primary power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what
is specified in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal
to or higher than the instrument's category rating to maintain the instrument's category rating.
Failure to do so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The
mains power should be 120 V RMS or 240 V RMS. Use only the power cord supplied with the
instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an
electrical safety ground. This instrument is grounded through the ground conductor of the
supplied, three-conductor AC line power cable. The power cable must be plugged into an
approved three-conductor electrical outlet. The power jack and mating plug of the power cable
meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all
accessible conductive parts (including control knobs) may provide an electric shock. Failure to
use a properly-grounded approved outlet and the recommended three-conductor AC line
power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument's front or rear panel is for a
reference of potential only and is not to be used as a safety ground.
Do not operate in an explosive or flammable atmosphere
Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finelydivided particulates.
The instrument is designed to be used in office-type indoor environments. Do not operate the
instrument
In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finelydivided particulates.
In relative humidity conditions outside the instrument's specifications.
In environments where there is a danger of any liquid being spilled on the instrument or where
any liquid can condense on the instrument.
In air temperatures exceeding the specified operating temperatures.
In atmospheric pressures outside the specified altitude limits or where the surrounding gas is
not air.
In environments with restricted cooling air flow, even if the air temperatures are within
specifications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The
operating temperature range is 0 °C to 40 °C and the operating humidity range is up to 80%
relative humidity with no condensation allowed.
Measurements made by this instrument may be outside specifications if the instrument is used
in non-office-type environments. Such environments may include rapid temperature or
humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise,
strong electric fields, or strong magnetic fields.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other
material gets on or inside the instrument, remove the instrument's power cord, remove the
instrument from service, label it as not to be operated, and return the instrument to B&K
Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives,
lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a
clean dry lint-free cloth or as instructed in this manual.
Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a
component in a life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and
internal adjustments must be made by qualified service-trained maintenance personnel who
are aware of the hazards involved when the instrument's covers and shields are removed.
Under certain conditions, even with the power cord removed, dangerous voltages may exist
when the covers are removed. To avoid injuries, always disconnect the power cord from the
instrument, disconnect all other connections (for example, test leads, computer interface
cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any
conductors by measurements with a properly-operating voltage-sensing device before touching
any internal parts. Verify the voltage-sensing device is working properly before and after
making the measurements by testing with known-operating voltage sources and test for both
DC and AC voltages. Do not attempt any service or adjustment unless another person capable
of rendering first aid and resuscitation is present.
Do not insert any object into an instrument's ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a
fault condition in the circuit exists.
Fuse replacement
Fuse replacement must be done by qualified service-trained maintenance personnel who are
aware of the instrument's fuse requirements and safe replacement procedures. Disconnect the
instrument from the power line before replacing fuses. Replace fuses only with new fuses of
the fuse types, voltage ratings, and current ratings specified in this manual or on the back of the
instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a
fire. Failure to use the specified fuses will void the warranty.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument.
Return the instrument to B&K Precision for service and repair to ensure that safety and
performance features are maintained.
This instrument contains one or more cooling fans. For continued safe operation of the
instrument, the air inlet and exhaust openings for these fans must not be blocked nor must
accumulated dust or other debris be allowed to reduce air flow. Maintain at least 25 mm
clearance around the sides of the instrument that contain air inlet and exhaust ports. If
mounted in a rack, position power devices in the rack above the instrument to minimize
instrument heating while rack mounted. Do not continue to operate the instrument if you
cannot verify the fan is operating (note some fans may have intermittent duty cycles). Do not
insert any object into the fan's inlet or outlet.
Use correctly sized wires
To connect a load to the power supply, use a wire diameter large enough to handle the
maximum continuous output short-circuit current of the power supply without the wire
overheating.
For continued safe use of the instrument
Do not place heavy objects on the instrument.
Do not obstruct cooling air flow to the instrument.
Do not place a hot soldering iron on the instrument.
Do not pull the instrument with the power cord, connected probe, or connected test lead.
Do not move the instrument when a probe is connected to a circuit being tested.
Certification
We certify that this product met its published specifications at time of shipment from the
factory.
Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other
European countries with separate collection systems).
This product is subject to Directive 2002/96/EC of the European
Parliament and the Council of the European Union on waste electrical and
electronic equipment (WEEE), and in jurisdictions adopting that Directive,
is marked as being put on the market after August 13, 2005, and should
not be disposed of as unsorted municipal waste. Please utilize your local
WEEE collection facilities in the disposition of this product and otherwise
observe all applicable requirements.
B&K Precision models 9103 and 9104 are multi-range power supplies differing from
conventional power supplies because voltage and current limits are varied to allow maximum
rated power to be delivered at any operating point. Contrast this to a conventional power
supply which has a maximum voltage and maximum current rating -- all operating points must
fall inside these limits.
This family of switching mode power supplies comes with a small form factor, auto cross-over
for constant voltage (CV) and constant current (CC), 3 voltage/current presets for frequentlyused settings, and remote control offering versatility for various applications.
The dual action (coarse/fine) tuning makes setting the voltage and current levels smooth,
precise, and fast. Setting, changing, and checking the current limit level can be done without
shorting the output.
Operating software through the USB interface allows turning the output power on or off and
the voltage and current can be adjusted without using the front panel. Up to 20 timed steps of
voltage and current, each with its own time duration from 1 s to 600 s, can be programmed for
up to 999 cycles.
A DC ramp or pulsed waveforms can be generated by the front panel or by the supplied
software with a preview of the final waveform. The 3 user defined presets facilitate quick
access to frequently used VI settings.
1.2 Package Contents
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items
from the shipping carton, and check for any obvious signs of physical damage that may have
occurred during transportation. Report any damage to the shipping agent immediately. Save
the original packing carton for possible future reshipment. Every instrument is shipped with the
following contents:
Disconnect all cables including the power cord from the instrument when changing the
instrument's line voltage. After changing the line voltage setting, ensure the
instrument has fuses of the proper ratings and types for the selected line voltage
before applying line power.
Figure1-Replacing a Fuse
Verify that all items above are included in the shipping container. If anything is missing, please
contact B&K Precision.
1.3 Input Power
The power supply has a universal AC input that accepts a line voltage with a nominal input of
110 V or 220 V AC. Use the line voltage selector switch in the back to switch between 110 V and
220 V operation.
1.4 Fuse Requirements
An AC input fuse is necessary when powering the instrument. Below is a table of the fuse
required for all models operating with either 110 VAC or 220 VAC input.
Table 1 - Fuse Specifications
1.5 Fuse Replacement
Follow the steps below to replace or check the fuse.
1. Locate the fuse box next to the AC input connector in the rear panel.
2. With a small flat blade screwdriver, insert into the fuse box slit to pull and slide out the
• Do not operate or touch this power supply with wet hands.
• Do not open the casing of the power supply when it is connected to AC mains.
• All servicing by qualified service personnel.
• Before replacing the AC fuse, determine and fix the cause of the over-current condition.
• Replace the AC fuse with the same type and rating as the original fuse.
• The maximum output voltage of the Model 9104 can be over 60 volts. Avoid touching
the output terminals.
• Series or parallel connection of 2 power supplies is not allowed.
• Use a grounded 3 prong AC receptacle.
• This unit is for indoor use only.
• Do not operate or place this unit in a humid or dusty environment, in direct sunlight, or
near any heat source.
•Before plugging into local AC power, check that the rating label on the back of the unit
matches the power outlet's characteristics.
• Do not block any ventilation openings of the unit.
• This unit must be used within its specified ratings. Regular excessive loading may
damage the power supply.
• When using the Remote Sense function, never short the Remote Sensing Terminal or
use the output on-off switch on the front panel or via Analog Remote Control or Remote
Programming.
•The input power cable conductors must be at least 0.75 mm2 and the total length of the
power cable must not exceed 3 m.
An activated button will be illuminated. The three buttons in the lower row have dual functions
which work in conjunction with other lower buttons or with the voltage, current control knobs
as defined by the green labels.
Throughout this manual the Menu button also acts as the Shift button. When used to enter the
menu mode, the button is pressed and held, when used for a Shift function, it is pressed
momentarily.
Control Knobs
The rotary encoder control knobs provide fine and coarse tuning with detents.
Push the knobs to toggle through the adjustable digit values (note the brightness change in the
affected LED to signify the selected digit).
Adjust the knobs as needed. The display will resume its normal brightness after few seconds to
confirm your adjustment.
3.3 Menu/Shift button and Key Lock button functions
Upper Voltage Limit (UVL)
a) To set the UVL, press the “Shift” button and then press the Voltage control knob.
b) The voltmeter will show the present limit and the Ammeter will show SUuL.
c) Tune the Voltage control knob to your desired limit value.
d) Press the “Shift” key to confirm and return to normal operation.
Upper Current Limit
a) To set the UCL, press the “Shift” Key and then press the Current control knob.
b) The Ammeter will show the present limit and the Voltmeter will show SUCL.
c) Tune the Current control knob to your desired limit value.
d) Press the “Shift” key to confirm and return to normal operation.
b) Turn the Voltage control knob until the panel meter displays wording rC SEt as below.
Note: After activating the Remote Control Mode, all function keys on the front panel will be Locked and the
Lock/Unlock key and Remote LED light up.
Figure 5 - Factory Reset Display
Figure 4 - Remote Control Display
c) Press the Voltage control knob to confirm the entry into remote control mode.
d) Turn the Current control knob to select remote ON or OFF.
e) Press the Voltage control knob to confirm and return to Set menu.
f) Press the Menu key to leave the Set menu.
Deactivating Remote Control Mode
a) Press the Lock/Unlock key to unlock all the function keys.
b) Repeat step above, but set to OFF at steps d, e.
Factory Reset
a) Press and hold the “Menu” Key for 3 seconds.
b) Turn the Voltage control knob until the panel meter displays “FACS SEt” as below.
c) Press the Voltage control knob to confirm to entry the Factory Reset mode.
d) Turn the Current control knob to select Yes or No.
e) Press the Voltage control knob to confirm.
f) Press the Menu key to leave the Set menu.
The power supply has returned to the factory settings of the 3 Presets and Upper Voltage and
Current limits.
This function is only valid when the output on/off is set to on.
a) Press Output ON/OFF to ON.
b) Press the “Shift”/ (Manual) key and then press the “Lock/Unlock” key.
c) Note the orange “W” next to the green “CV” to indicate the power in watts is being
displayed.
d) To change from Wattmeter to Ammeter, press the “Shift” key and then press
“Lock/Unlock” key.
The Key Lock button
a) Press the “Lock/Unlock” key to lock or unlock all function keys and the control knob on
front panel.
b) The “Lock/Unlock” key lights up when the keys are locked.
These power supplies have the following modes of operation:
Normal
Preset
Transient Mode
Remote Control
Remote Programming
The power supply is factory preset to Normal Mode with the maximum specified current level
and the supply is in constant current mode.
4.1 Normal Mode
Table 2 - Power Ranges
Maximum values of voltage and current are limited by the maximum output power of 320 W.
Decrease the voltage setting to get a higher current and vice versa for high voltage.
Powering on the power supply
After the unit is turned on, the CV, V and A LED indicators are lit up, displaying the set voltage.
The current is set to 00.00. The output on-off button is set to off by default. Push the
voltage/current knob or turn one click in either direction to see the present settings. Both
displays return to 00.00 after a few seconds.
Connect the device under test to the power supply
Red (+) is connected to the positive terminal and black (-) is connected to the negative terminal.
Connect equipment to be powered with the power supply output off. Connect a grounding wire
as needed (be cautious to avoid ground loops).
Due to the high current rating of the power supply, proper wire sizes are necessary for safe
connectivity and to prevent wires from overheating. Refer to the table below as a reference for
proper wire sizes according to the amount of current used for operation:
Before connecting wires to the output terminals, turn OFF the power supply to avoid
damage to the instrument and the device under test (DUT). For safety, load wires must
have a wire gauge size large enough to prevent overheating when the power supply
operates at maximum short circuit output current. It will also prevent large voltage drops
from resistances in the wires.
Hazardous voltages may exist at the outputs and the load connections when using a power
supply with a rated output greater than 40V. To protect personnel against accidental contact
with hazardous voltages, ensure that the load and its connections have no accessible live parts.
Ensure that the load wiring insulation rating is greater than to the maximum output voltage of
the power supply.
Turn on the Power Supply
Press the Output On/Off button.
Voltage and current will be displayed continuously and CV or CC will be lit according to the
settings and the value of the connected device under test.
When operation or testing is finished, turn off the device under test first and then switch off
the power supply.
Changing Ammeter to Wattmeter
This function is only valid when the output on/off is set to on.
Press the “Shift” (Menu) key and then press the “Lock/Unlock” key to change Ammeter
to Wattmeter.
The unit is in Wattmeter display mode.
To change back to Ammeter display mode, press the “Shift” key and then press the
“Lock/Unlock” key.
Key Lock Function
This function prevents any inadvertent changes on the voltage and current settings.
Press the “Lock/Unlock” key to lock or unlock all function keys and control knob on the
The “Lock/Unlock” key lights up when the keys are locked.
Note: The UVL and UCL settings are factory preset at maximum rated output values.
Changing the values of UVL and UCL
These functions are designed to give additional protection for voltage sensitive and/or currentsensitive loads. When the voltage or current at the output terminal exceed the UVL or UCL set
limits, the output will be turned off.
UVL setting: Press the SHIFT “Menu” Key to light up, then press the VOLTAGE control
knob.
The voltmeter will show the present voltage limit and the Ammeter will show SUuL.
Turn the Voltage control knob to your desired limit voltage. Press the SHIFT “Menu” key
to confirm with light out and return to normal operation.
UCL setting: Press the SHIFT “Menu” Key to light up, then press the CURRENT control
knob. The voltmeter will show the SUCL and the ammeter will show the present current
limit. Turn the CURRENT control knob to your desired limit current. Press the SHIFT
“Menu” key to confirm with light out and return to normal operation.
The UVL and UCL settings will affect the usable voltage range and current range and will trigger
the respective protection if the set limits are exceeded.
If the remote sense function is used, allowance must be included in the setting of UVL for the
gain in output voltage.
4.2 Preset Mode
There are 3 preset outputs (Preset A/ Preset B/ Preset C)
•Press the Preset key A or B or C to use the preset voltage and current settings; the
key will light up.
•To check the preset values without affecting the output, press the “Shift” Key and
then the selected Preset Key.
• The V and I settings of corresponding PA, PB, PC will be show on the panel meters.
• To change the values of the Presets, press the desired preset key (it will light up)
and adjust the voltage/current level, the set values will be brighter than the
measured values.
• The newly set values of the Preset will be kept even after the power supply has
Factory
Output Voltage
Output Current
Preset A
9103: 5V
9104: 5V
9103: 20A
9104: 10A
Preset B
9103: 13.8V
9104: 13.8V
9103: 20A
9104: 10A
Preset C
9103: 20V
9104: 20V
9103: 16A
9104: 8A
Always check the voltage & current level of the Preset before setting output to on when
connected to a load.
been turned off.
•The factory default settings of the 3 Presets are shown in the following table.
Table 4 - Preset Mode
4.3 Transient Mode
This feature uses the 3 Presets from the previous section to generate the Ramp Up/Down and
other waveforms such pulse, triangular, trapezoid, etc. by varying the voltage level, duration
time (Func) at each voltage level and time interval (Δt) to move from one voltage to another
voltage level. The number (Runt) of repetitive cycles of the waveform can be set as well. At the
end of the run cycle run, the output will be turned off.
It is recommended that the computer interface be used instead of panel programming to
operate this function as the simulated open circuit output wave forms can be previewed at the
PC and settings are made more conveniently and be saved. The output waveform data can be
logged for reference.
The Transient Mode is independent from the output state. It is recommended to set the voltage and
current to zero before enabling this mode.
Turn the output off by pressing the OUTPUT ON button (make sure the backlight is off).
Set the voltage and current to 0 V and 0 Amps.
Turn the output on by pressing the OUTPUT ON button (make sure the backlight is on).
Run the transient: Press the Menu button (Shift function)+ Toggle button.
Turn the output off by pressing the OUTPUT ON button (make sure the backlight is off).
Turn the transient off by pressing the Menu button (Shift function) +Toggle button.
Figure 7 - Step size (seconds)
Figure 6 - Sequence Menu: dvdt set
The voltage knob will allow the user to navigate through the options of the Transient Mode and
the Current knob will set the values of the options. Press the voltage knob to enter the value
and exit the current menu and to exit the main menu, press the Menu Button.
Step Size (seconds)
This menu allows the user to enter the step size (duration of a Preset value). Range: 0 to 600 s
Transition time:
This is the time the power supply will take from the first preset value to the second. Range: 0 to
20 s.
Turn the output off by pressing the OUTPUT ON button (make sure the backlight is off).
Set the voltage and current to 0 V and 0 Amps.
Turn the output on by pressing the OUTPUT ON button (make sure the backlight is on), now, current
and voltage are set to zero.
Run the sequence: Press the Menu button (Shift function) + Toggle button.
Turn the output off by pressing the OUTPUT ON button (make sure the backlight is off), sequence will
continue running, but there will be no output.
Turn the sequence off by pressing the Menu button (Shift function) +Toggle button.
Example 1: Pulse waveform
1. Press PRESET A button and set 10 V using the Voltage knob and using the current knob
enter 1 Amp. Following this procedure also configure PRESET B with values 5 V and 1 Amp.
2. Press and hold the Menu button (~4 seconds).
3. Use the Voltage knob to find the dudt Set then press the Voltage knob to enter the menu.
4. Use the Voltage knob to navigate to the FunA screen and enter a time of 3 seconds (003) by
rotating the current knob and press the Voltage knob to enter this value.
5. Rotate the Voltage knob to navigate to the FUNB screen and adjust the time following the
procedure in step #4, but this time set the FUNB value to 003.
6. Rotate the Voltage to navigate to the FUNC screen and adjust the time following the
procedure in step #4, but this time set the FUNC value to 000 since this preset will not be
used.
7. In the same menu, dudt Set, use the Voltage knob to set the transition time from preset A
to preset B y setting A_b_ to 0000 and b_A_ also 0000.
8. Also, set the run to A_b to indicate a transition from PRESET A to PRESET B.
9. Set rUNt to 0000 which means “cycle indefinitely “.
10. To generate the waveform, push the “Shift” key and then press the “Output on” button.
The waveform will repeat cyclically if rUNt 0000 is set to 0000 or according to number of
cycles entered. The output will show a pulse waveform with 3 seconds at 10 V and 1
second at 5 V.
11. Press the OUTPUT ON button to turn the output off (backlight should be off, too).
12. To stop the cycle running, push the “Shift” key and then press “Output on” button.
a) Make sure that the voltage and current of the Presets have been set first.
In this example, we use Preset A = 5V, Preset B = 10V
b) Press and hold the Menu button (~4 seconds).
c) Using the voltage knob, navigate to the “dudt SEt” option (transient mode).
d) Press the VOLTAGE KNOB to enter the transient mode.
e) Set these parameters with the following values:
Func A=0 sec. Displays “FUNCA 0000”. Step size for Preset A.
Func b=15 sec. Displays “FUNCN 0015”. Step size for Preset B.
ΔV a-b = Displays “rUN A _ b”. This step selects the first step.
Δt a-b = 5 sec: Displays “A_b_0005”. This step sets the transition time from A to B to 5
seconds.
Run t = 3 cycles: Displays “runt 0003”. Set the sequence to run 3 times.
f) In each option, make sure to press the Voltage Knob to enter the values. Press the Menu
Key to exit this Menu.
g) To generate the waveform, push the “Shift” key and then press the “Output on” button.
h) Press the OUTPUT ON button to turn the output off (backlight should be off, too).
i) To stop the cycle running, push the “Shift” key and then press “Output on” button.
After entering into Remote Control Mode, the following external set ups can be selected.
There are two methods for remote control of current and voltage adjustment.
Both methods require both the current remote control part and the voltage remote control
part to be set up and in use at the same time in order for the analog remote control mode to be
functional.
Otherwise the power supply will be in CC mode and the analog remote control will not be
functional.
Setting up the provided remote connector plug:
a) Remove the black portion of the remote control connector plug by removing the screw as
shown in Figure 17 - Remote Control Connector.
Figure 17 - Remote Control Connector
b) Solder 5 wires (22AWG) to pins 1, 2, 3, 4, and 5 of pin plug. Refer to Figure 18 for pin
numbers.
Figure 18 - Pin Numbers
c) Make sure the load is disconnected and the power supply is OFF.
d) Plug the remote connector plug into the analog remote control terminal of the power
PIN
FUNCTIONS
REMARKS
1
Internal DC +5 V
Less than 50 mA
2
Voltage Adjust
0 – 5 V
3
Current Adjust
0 – 5 V
4
Ground
5
Output OFF
Short to Ground
6
N/A
supply.
e) Secure the remote connector plug to the terminal socket by screwing in the connector ring
(Figure 19).
Figure 19 - Connector Ring
Then, you can choose one of the following two methods to use the analog remote control
feature:
Using two external variable DC voltage sources, or
Using two 5 kΩ variable resistors.
Analog Remote Control: Voltage Control
A variable external DC voltage source of 0 – 5 V is fed into the analog remote control terminal
to adjust the output voltage level of both Main and Auxiliary output.
Table 5 – Analog Remote Control Connector Pinout: Voltage Control
Do not input higher than 5 V, otherwise the overvoltage protection (OVP) will be
triggered.
Steps to configure Voltage Control:
1. Make sure the load is disconnected and the power supply is OFF.
2. Connect pin 2 to positive polarity of first external voltage source and pin 4 to negative
polarity of first external voltage source.
3. Connect pin 3 to positive polarity of second external voltage source and pin 4 to negative
polarity of second external voltage source.
4. Turn the remote control ON/OFF switch to ON position.
5. Switch on the power supply.
6. Check the output voltage range of the power supply by varying the external voltage source
for voltage adjustment from 0 to 5 V.
7. Short circuit the main output with a 12AWG wire and check the display for CC setting by
varying the external voltage source for current adjustment from 0 to 5 V.
8. Switch off the power supply.
Analog Remote Control: Resistance Control
Using two variable resistor of 5 kΩ can be used to adjust the output voltage level of both Main
and Auxiliary output.
Steps to configure Resistance Control:
1. Make sure the load is disconnected and the power supply is OFF.
2. Prepare two 5 kΩ variable resistors and connect wires from pins 1, 2, 3, and 4 as shown in
6. Short circuit the main output with 12AWG wire and check the display for CC setting by
Note: using the 8 pin remote plug provided and connect with 22 AWG wires.
Do not use remote output on/off control when power supply is in Remote Sense
connection.
The power supply can be damaged due to the high current going through the thin remote
sense wires.
Note: The power supply does not need to have Remote control enabled to be controlled via the PC interface.
Pin numbers are marked
varying the 5 kΩ variable resistor for current adjustment.
7. Switch off the power supply.
Remote Enable and Disable the Output
This remote output on/off control can be activated in any of the following modes: Normal,
Preset, Remote, and Set mode.
a. By default, Pin 5 is open and output is on.
b. Shorting Pin 5 to Pin 4 (ground) and output is off.
c. When the output is off, the CV and CC LEDs will flash. The current output voltage and
current setting will show on the panel meter.
d. You also can adjust the output by the voltage and current control knobs to the desired
value when the output is off.
Figure 21 - Analog Remote Control Connector
4.5 PC Interface Control
This PC Interface Mode connects the power supply to a PC via USB PORT. Application software
for wave form generator, external timed programs, data logging of output with I, V, W with
time span and the setting of miscellaneous parameters limits.
An incorrect disconnection sequence will damage the Power Supply.
Always disconnect the Remote Sensing Terminal first.
DO NOT USE Output On Off on front panel or by remote operation.
It is recommended you use the key lock function to prevent accidental switching off
the output by the output on/off switch.
Never connect the Remote Sensing Terminals in reverse polarity.
Figure 22 - Remote Sensing
Connections:
1. Make the power connections between the power supply and the equipment.
2. Check and make sure the power connections are secure.
3. Make the connections between the Remote Sensing terminals and the equipment.
Figure 22 shows the connections between the Remote Sensing terminals, Power output
terminals, and the equipment.
Disconnecting Remote Sensing
1. First disconnect the remote sense connections.
2. Then disconnect the power connections between the power supply and equipment.
Remote Sense commonly compensates for output line voltage drops of about 0.25 to 0.75V.
Use Twisted Remote Sense wires and shielding to minimize noise.
This unit has a built-in over voltage protection feature. In the event the output voltage exceeds
the set OUP voltage value (see the specified range from specifications table), the protection will
be triggered and the output power will be turned off and the OUP warning appears as below.
To reset the warning:
1. Remove the connection to the load and press the Output ON/OFF key.
2. Reconnect the load and normal operation should be resumed. If this problem persists,
please consult with BK Precision.
6.2 OTP: Over Temperature Protection
There is a temperature sensor inside the unit to prevent the power supply from getting too hot.
At OTP, there is no output and the following warning will appear on the LED display.
When you get this warning, switch off the unit and disconnect the load.
Check the load and output setting. Allow the unit to cool down for at least 30 minutes.
Ensure no vents are blocked and check that there is enough clearance around the power
supply.
Listen for the self-test fan noise from the cooling fan when you turn on the unit again.
If you cannot hear this self-test fan noise on power-up, then the fan is at fault; do not use the
power supply and contact BK Precision.
Normally the over-current protection is provided by the constant current mode.
If constant current mode fails and is not detected, it may cause serious damage to your load.
The OCP is to minimize the damage to your load if the power supply's constant current control
fails.
Figure 25 - OCP Warning
To reset the warning:
1. Switch off your power supply as soon as you see the OCP warning as shown below.
2. To reset this warning, disconnect the load and press the Output ON/OFF key.
3. Reconnect the load and the power supply should resume normal operation once power
is turned on again.
4. If this problem cannot be fixed, please contact and consult BK Precision.