The following safety precautions apply to both
operating and maintenance personnel and
must be observed during all phases of
operation, service, and repair of this
instrument.
DO NOT OPERATE IN AN EXPLOSIVE
ATMOSPHERE
Do not operate the instrument in the presence
of flammable gases or fumes. Operation of
any electrical instrument in such an
environment constitutes a definite safety
hazard.
KEEP AWAY FROM LIVE CIRCUITS
Instrument covers must not be removed by
operating personnel. Component
replacement and internal adjustments must be
made by qualified maintenance personnel.
DO NOT SUBSTITUTE PARTS OR MODIFY
THE INSTRUMENT
Do not install substitute parts or perform any
unauthorized modifications to this instrument.
Return the instrument to B&K Precision for
1
service and repair to ensure that safety
features are maintained.
WARNINGS AND CAUTIONS
WARNING and CAUTION statements, such
as the following examples, denote a hazard
and appear throughout this manual. Follow all
instructions contained in these statements.
A WARNING statement calls attention to an
operating procedure, practice, or condition,
which, if not followed correctly, could result in
injury or death to personnel.
A CAUTION statement calls attention to an
operating procedure, practice, or condition,
which, if not followed correctly, could result in
damage to or destruction of part or all of the
product.
Safety Guidelines
To ensure that you use this device safely,
follow the safety guidelines listed below:
2
This meter is for indoor use, altitude up to
2,000 m.
The warnings and precautions should be
read and well understood before the
instrument is used.
When measuring in-circuit components, first
de-energize the circuits before connecting to
the test leads.
Discharge capacitor before testing.
The meter is safety-certified in compliance
with EN61010 (IEC 1010-1) Installation
Category II (CAT. II) 50 V, Pollution Degree
2 environment.
Use the meter only as specified in this
manual. Otherwise, the protection provided
by the meter may be impaired.
The power for the meter is supplied with a
single standard 9V battery. But also a line
operation is possible using a 12V AC to DC
adaptor. If a power adaptor is selected,
please be sure it fulfills the safety
requirements of a relevant IEC standard.
3
outside is negative (-)
Safety Symbols
This symbol is a warning and
indicates that the user should refer to
the operating instructions located in
the manual.
DC Current
Indicates inside pin is positive (+),
Compliance Statements
Disposal of Old Electrical & Electronic Equipment
(Applicable in the European
Union and other European countries with separate
collection systems)
4
This product is subject
Council of the European
electrical and electronic
equipment (WEEE) , and
ipal waste. Please
to Directive 2002/96/EC
of the European
Parliament and the
Union on waste
in jurisdictions
adopting that Directive,
is marked as being put
on the market after
August 13, 2005, and
should not be disposed
of as unsorted
munic
utilize your local WEEE
collection
facilities in the
disposition of this
product and otherwise
observe all applicable
requirements.
B&K Precision’s 878B and 879B 40,000-count L/C/R
hand-held meter is designed ideally for measuring
inductance, capacitance and resistance components.
Simple to operate, the instrument not only takes
absolute parallel mode measurements, but also series
mode measurement. The meter provides direct and
accurate measurements with selectable testing
frequencies.
Front panel push buttons maximize the convenience
of function and feature selection such as data hold,
maximum, minimum and average record mode,
relative mode, tolerance sorting mode, frequency and
L/C/R selection.
The test data can be transferred to PC through a Mini
USB connection, great for applications that require
data logging
A tilt stand provides position flexibility for viewing and
operating the meter. The over-molding rubber case
protects the meter for better durability. Additionally,
top rubber visor molding above the screen is designed
to prevent scratches on the display when meter is
positioned upside down.
10
A single 9V battery or the included DC 12V power
adaptor (model 879B only) can be used to power the
meter. This gives user flexibility for portable or benchtop use.
PACKAGE CONTENTS
Each 878B and 879B LCR meters are shipped with
the following contents.
• 878B or 879B LCR meter
• Instruction Manual
• Mini USB Interface Cable
• Red & Black Banana to Alligator Test Leads
• 9V Battery
• *AC Adapter (model 879B only)
*This can be purchased as an optional accessory for
model 878B.
Please locate them from the original packaging to
ensure nothing is missing. If in the case that an item
is missing, please contact B&K Precision immediately.
11
+-
R
!
LCR
REL
ESR
MAX
AVG
MIN
¦ÈQ
AUTO
D
1%5%10%20%
deg
%
kHz
PAL
SER
¦ÌH
p F
¦Ì
RMT
Mk
TOL
DH
n
n
@
OFF
Z
1
2
4
5
3
6
7
8
9
10
13
12
11
USB
D/Q/
/ESR
L/C/R/Z
POWER
REC
REL
HOLD
CAL
P
S
FREQ
TOL
UTIL
LCR Meter
879B
FRONT PANEL OVERVIEW
Figure 1 - Front Panel Display (model 879B shown)
12
Front Panel Display Descriptions
1. LCD Display
2. Power ON/OFF Button
3. USB Communication / *Back light button
4. Secondary Display mode (for dissipation
factor(D), quality factor (Q), *phase angle (θ),
*equivalent series resistance (ESR)
measurement) selection button
5. Primary Display mode (for inductance,
capacitance, resistance, and *impedance
measurements) / Parallel or Series
measurement method selection button
6. Tolerance mode / Utility down arrow selection
button
7. Hold Display mode / Record mode selection
button
8. Utility menu button
9. Test Frequency / Utility up arrow selection
button
11. 12V DC adapter input (use with an external
power adapter (rated 12VDC, 150mA, 4mm
power plug))
Note: Use with included power adapter only.
Use with improper power adapters may
damage instrument.
13
WARNING: Before connecting an external
power adapter, please check the battery
compartment in the rear side of the unit. If
a battery is installed, be sure that the
polarity matches the (+) and (-) labels as
indicated inside the battery compartment.
If it is not installed correctly, please
remove the battery and install it with
correct matching polarity as indicated in
the compartment. See “Installing Battery
section for details. DO NOT, at any time,
connect an external power adapter when a
battery is installed incorrectly. Doing so
will damage the instrument and void its
warranty.
12. Input sockets (banana jack inputs) and
terminals for positive, negative, and guard
(see “Guard Terminal” in “SUPPLEMENTAL
INFORMATION” section for details)
13. Standard mini USB port (for remote
controllability)
For model 879B only. Not included on model 878B.
*
”
Front Panel Buttons
All front panel buttons have specific colored labels on
them. They are all marked in white, blue, or yellow
14
color. Each color has a specific representation, as
described below:
White – With the exception of the
button, all white colored labels
represent the primary function of that
button; meaning that function will be
set or configured upon pressing it.
Blue – Some of the buttons have a blue label
underneath a white label. This means
the function indicated by the blue
label will be set or configured if that
button is pressed and hold down for 2
seconds.
Yellow – There are total of 3 buttons with
yellow labels. They are
. These functions are
exclusively for use when entering
UTIL menu only. See “Utility Menu”
section for details.
15
L
C
R
REL
ESR
MAX
AVG
MIN
¦ÈQ
AUTO
D
1%5%10%20%
deg
kHz
PAL
SER
¦Ì
H
p F
¦Ì
RMT
Mk
TOL
DH
n
n
@
OFF
Z
1
25
910
12
14
16
2021
26
27
28
8
3
2324
25
4
15
11
13
6 7
22
17
18
19
%
29
LCD DISPLAY OVERVIEW
Figure 2 - LCD Indicator Display
LCD Display Descriptions
1. LCRZ – Primary display function indicator (*Z
display)
2. MAX – Maximum reading indicator
3. AVG – Average reading indicator
4. MIN – Minimum reading indicator
5. REL – Relative mode indicator
16
6. Θ – *Phase angle indicator for secondary
display
Before beginning to operate the instrument, a power
source is necessary for it to turn on. There are two
methods to power the instrument: Battery and
external source.
Installing Battery
The 878B and 879B LCR meters can use a battery to
provide power to the instrument so that it can be
portable.
The meters use a standard 9V size battery (or
NEDA 1604, JIS006P, IEC6F22 carbon-zinc or
alkaline battery).
To install the battery:
1. Place the meter upside down. Open up the
back-flip stand, and locate the screw that
tightens the battery compartment cover as
indicated in Figure 3. Use a screwdriver to
unscrew and remove the cover.
19
Figure 3 - Back Cover
2. Insert 9V battery into compartment. Note the
positive (+) and negative (-) terminals as
indicated inside the battery compartment (See
Figure 4). Be sure to insert the battery with
matching polarity.
20
Figure 4 - Battery Compartment
3. Place the battery compartment cover piece by
sliding it into the top slid first. Place screw at
the bottom of the cover piece and tighten
down with a screw driver.
4. Push and hold down the button for 2
seconds to turn on the instrument.
Connecting External Power Source
The 878B and 879B can also be powered using an
external AC adapter. The model 879B comes with
this adapter included in the package, while it is
optional for model 878B.
21
For external power, use AC adapter rated for
output 12VDC, 150mA, with a standard 4mm
connector only.
WARNING: Use of incorrect adapters may damage
the instrument. Please use B&K Precision’s adapter
only.
To connect the adapter, do the following:
1. If a battery is installed, please check the
battery compartment again that the polarity of
the battery matches the polarity as indicated
by the labels inside the compartment. If it is
not, please remove and insert the battery with
matching polarity. If a battery is not installed,
continue to the next step.
WARNING: DO NOT, at any time, connect
an external power adapter when a battery
is installed incorrectly (reverse polarity or
non-matching polarity to indicator of
battery compartment). Doing so will
damage the instrument and void its
warranty.
2. Connect the AC adapter connector into the
right side panel of the instrument. See Figure
5 below.
22
12VDC Input
AC Adapter
Figure 5 - Connecting AC Adapter to Meter
3. Now, connect the AC Adapter socket into an
electrical outlet.
4. Push and hold down the button for 2
seconds to turn on the instrument.
Note: The meter can be operated with a battery
installed while an AC adapter is plugged in at the
same time (As long as the battery is inserted properly
with correct polarity). In this event, the meter will
automatically switch to consume power from the AC
adapter instead of the battery to preserve battery life.
Low Battery Indication
The LCR meter has a low battery indicator to notify
the user when to replace battery. When the display
23
starts flashing the indicator, the battery voltage
is below normal working voltage. In this case,
accuracy of the meter will also decrease. It is
recommended that the battery be replaced as soon as
possible before continuing operation. See “Installing
Battery” for instructions.
Backlit Display (model 879B only)
Model 879B LCR meter has a backlit display that
allows you to see the LCD display in dark conditions.
To turn on the back light, press and hold down
button for 2 seconds. Back light will turn on and
brighten the LCD display.
To turn off the back light at any time, press and hold
down button for 2 seconds again. Back light
will turn off and return to normal display.
When Using Battery Power
When the meter is powered using 9V battery, the back
light display will turn on upon holding down the
button for 2 seconds. It will stay at maximum
24
brightness for 15 seconds. After another 15 seconds
(30 seconds total from the time of turning on), the
back light will automatically turn off to conserve
battery power.
When Using External Power
When the meter is powered using an external AC
adapter, the back light display will turn on upon
holding down the button for 2 seconds. It will
stay at maximum brightness continuously until the
user presses and holds down the button for 2
seconds again.
Note: If a battery is installed while using an AC
adapter simultaneously, unplugging the AC adapter
will automatically turn off the back light after it has
been lit for 30 seconds.
25
OPERATION INSTRUCTIONS
Data Hold
The data hold function allows the user to freeze the display
when pressed, holding the measured value until data hold is
turned off.
Turn On Data Hold
To use data hold, press the button once. The “DH”
indicator will display on the screen when data hold is active.
Turn Off Data Hold
To disable the data hold, press again. The “DH”
indicator will disappear on the screen, and meter will remain
in normal operation mode.
Note: Changing the primary function, secondary function,
or test frequency will automatically turn off the data hold.
Static Recording
This mode is used for recording maximum, minimum, and
average values. It is often useful for testing the range of
values in which a component is expected to fall within upon
measurement.
26
Enable Static Recording
Press and hold down the button for two seconds to
enter the static recording mode. The display should indicate
“MAX AVG MIN” simultaneously. This indicates the meter is
in static recording mode and recording is performed
immediately.
Using Static Recording
There are four different modes that can be selected in
static recording. They are indicated by the
descriptions below. These modes can be changed
with each button press. Per each press of the
button, the modes will change and repeat in
the following order:
Recording Mode Maximum Mode Minimum
Mode Average Mode
Recording Mode
This is the default mode when first enabling static
recording. In this mode, the screen will display “MAX AVG MIN” indicator. At this point, the meter will start
making recordings based on measured values from
27
the input sockets or terminals. As recording is
performed, maximum, minimum, and average values
will be stored after a brief moment. A beep tone will
sound once a recording has been stored.
Note: Subsequent beep tones may occur in this mode if
there are new values that are recorded. For example, if a
new maximum is detected, it will beep once again to indicate
that the new value has been stored. Any previously stored
values will be overwritten with the new recorded values.
Maximum Mode
In this mode, the “MAX” indicator will be shown on display.
This indicates that the value in the primary display
represents the recorded maximum value.
Minimum Mode
In this mode, the “MIN” indicator will be shown on display.
This indicates that the value in the primary display
represents the recorded minimum value.
Average Mode
In this mode, the “AVG” indicator will be shown on display.
This indicates that the value in the primary display
represents the recorded average value. This average value
is obtained by taking the maximum and minimum recorded
values and taking the average of the two values.
28
Loading...
+ 74 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.