4. Centrifuge the spin column at ~80 x g (1,000 rpm)*
Life Science
Group
06-0735 1206 Sig 1106
10008254 US/EG Rev D
Bio-Rad
Laboratories, Inc.
Web site www.bio-rad.com USA 800 4BIORAD
Australia 61 02 9914 2800 Austria 01 877 89 01
Belgium 09 385 55 11 Brazil 55 21 3237 9400
Canada 905 712 2771 China 86 21 6426 0808
Czech Republic 420 241 430 532
Denmark 44 52 10 00 Finland 09 804 22 00
France 01 47 95 69 65 Germany 089 318 84 0
Greece 30 210 777 4396
Hong Kong 852 2789 3300
Hungary 36 1 455 8800 India 91 124 4029300
Israel 03 963 6050 Italy 39 02 216091
Japan 03 5811 6270 Korea 82 2 3473 4460
Mexico 52 555 488 7670
The Netherlands 0318 540666
New Zealand 0508 805 500 Norway 23 38 41 30
Poland 48 22 331 99 99
Portugal 351 21 472 7700
Russia 7 495 721 14 04 Singapore 65 6415 3188
South Africa 27 861 246 723
Spain 34 91590 5200
Sweden 08 555 12700
Switzerland 061 717 95 55
Taiwan 886 2 2578 7189
United Kingdom 020 8328 2000
for 30 seconds to remove the buffer.
5. Replace the bottom cap of the spin column.
6. Add 200 µl 50 mM Tris buffer, pH 9 to the column,
then replace the top cap on the column.
7. Vortex the column to mix the sorbent with
the buffer.
8. Remove the top and bottom caps of the spin
column; set the column upright in a new 1.5 ml
microcentrifuge tube.
9. Centrifuge the spin column for 30 seconds to
remove the buffer.
10. Repeat steps 5–9 twice, for a total of three
buffer washes.
11. Replace the bottom cap on the spin column.
The sample can now be added to the spin column.
12. If the column is not used right away, add 0.5 ml
buffer, pH 9 to prevent dehydration.
* Recommended speeds for an Eppendorf 5417R centrifuge.
Ordering Information
Catalog # Description
C54-00017
C57-30080
© 2006 Bio-Rad Laboratories, Inc.
ProteinChip Q Spin Columns, 20
ProteinChip Q10 Arrays, A-H format, 12
Eppendorf is a trademark of Eppendorf-Netheler-Hinz GmbH. HyperD is
a trademark of Pall Corporation.
The SELDI process is covered by US patents 5,719,060, 5,894,063,
6,020,208, 6,027,942, 6,124,137, 6,225,047, 6,528,320, 6,579,719,
and 6,734,022. Additional US and foreign patents are pending.
ProteinChip®Q
Spin Columns
Instruction Manual
Catalog #C54-00017
For technical support,
call your local Bio-Rad office, or
in the US, call 1-800-4BIORAD
(1-800-424-6723).
Introduction
Ion exchange chromatography (IEC) is the most
frequently used chromatography technique for the
separation of proteins, peptides, and other charged
biomolecules. IEC achieves separation by taking
advantage of the difference in net surface charge of these
elements in a complex biological mixture.
Surface charge is determined by the presence of weak
acidic and basic groups on a protein and as such is
highly pH dependent. Typically, at pH values below a
protein’s isoelectric point (pI), the protein acquires a net
positive surface charge and will adsorb to cation
exchange media. At pH values above the protein’s pI, the
protein acquires a net negative surface charge and will
adsorb to anion exchange media. Variations in mobile
phase pH are most often used to selectively bind a
molecule of interest to an IEC support. The ideal pH value
is one that creates a large net charge difference among
the different sample components.
Desorption of bound proteins from IEC media is achieved
by either altering the pH or increasing the salt
concentration (introducing competing ions) of the mobile
phase. The higher the surface charge of the protein, the
higher the ionic strength that is needed to bring about
desorption. Variations in the mobile phase ionic strength
are used that take advantage of the charge differences in
species bound to the IEC support.
ProteinChip Q (quaternary amine) spin columns contain
Q ceramic HyperD F sorbent, an anion exchange
chromatography support. This material has been
specifically designed to rapidly capture and separate
proteins and peptides from crude biological mixtures
under physiological conditions of ionic strength.
Applications
Crude biological samples contain thousands of proteins
and peptides. Their complexity presents a significant
analytical challenge for researchers evaluating in vitro
and in vivo experiments from cell lysates and plasma
samples. Strategies to reduce the complexity of these
samples are essential to gaining a better understanding
of cellular mechanisms and disease states.
IEC is a powerful approach to prefractionation of
biological samples. This technology can reduce the
presence of abundant species, greatly increase the
number of less abundant species, and enrich samples
in elements of interest. ProteinChip Q spin columns are
designed to facilitate the analysis of crude biological
samples by fractionating proteins on the basis of net
surface charge. Biological samples can be further
simplified by fractionation with ProteinChip arrays.
Q ceramic HyperD F sorbent mimics the binding and
elution characteristics of ProteinChip Q10 arrays.
ProteinChip Q spin columns can be used for small-scale
purifications of proteins discovered on array surfaces.
Storage
Store columns at 4°C.
Technical Considerations
n
Recommended sample binding time is 20–40 min
n
Each spin column containing Q ceramic HyperD F
sorbent has a minimal protein binding capacity of
7.7 mg
Column Equilibration
ProteinChip Q spin columns contain a sorbent that
requires buffer equilibration prior to use. This equilibration
buffer should be the same as the intended sample
binding buffer. To ensure that most proteins will bind to
the column, the salt concentration should be <100 mM,
pH ~9.
1. Tap the spin column lightly to settle the sorbent to
the bottom (near the tapered end) of the column.
2. Remove the caps on the top and bottom of the
spin column.
3. Set the column upright in a 1.5 ml microcentrifuge tube.
© 2006 Bio-Rad Laboratories, Inc. © 2006 Bio-Rad Laboratories, Inc. © 2006 Bio-Rad Laboratories, Inc.