Bio-Rad PowerPac HC High-Current Power Supply User Manual

Page 1
PowerPac™ HC
Power Supply
Instruction Manual
Catalog Number
164-5052
Page 2
Page 3
Table of Contents
Page
Safety ...........................................................................................i
1.1 Overview.................................................................................................1
1.2 Unpacking...............................................................................................2
Section 2 Control Features.........................................................................3
Section 3 Setup and Operation ..................................................................4
3.1 Setup and Programming a Run...............................................................4
3.2 Editing Run Parameters During a Run ....................................................5
3.3 Ending a Run..........................................................................................6
3.4 Changing System Defaults using the Setup Menu ..................................6
Section 4 Maintenance and Troubleshooting............................................8
4.1 Maintenance ...........................................................................................8
4.2 Troubleshooting ......................................................................................8
4.2.1 Basic Troubleshooting................................................................8
4.2.2 Power Failure Detection .............................................................8
4.2.3 User Serviceable Error Messages ..............................................9
4.2.4 Bio-Rad Serviceable Error Messages.......................................11
4.3 Replacing a Fuse..................................................................................11
4.4 Expediting Technical Support................................................................11
Appendix A Specifications ...........................................................................12
Appendix B Warranty and Ordering Information ........................................13
List of Figures
1. Front View.
2. Rear View.
3. Front View with Legs in Lowered Position.
4. Front Panel.
5. Power Leads Connected Correctly.
6. Power Leads Connected Incorrectly.
7. Rear View Showing Fuse Drawer with Notches.
© 2013 Bio-Rad Laboratories, Inc. All rights reserved.
Page 4
Page 5
Safety
!
Caution/Warning
PowerPac power supplies use high output voltages that are electrically isolated from earth ground through a protective impedance to minimize the risk of electrical shock to the user. The following guidelines should be observed and followed when using a PowerPac power supply.
PowerPac power supplies have passed test for operation at temperatures between 0° and 40°C, with relative humidity between 0 and 95% non-condensing. Operating the power supply outside these conditions is not recommended by Bio-Rad and will void the warranty.
1. To ensure adequate cooling of the power supply, be sure that there is at least 6 cm
2. Always connect the power supply to a 3-prong, grounded AC outlet, using the
3. Bio-Rad electrophoresis cells have molded two-prong plugs that are inserted
4. Do not operate the power supply in extreme humidity (>95%) or where
5. When taking the power supply into a cold room, the unit can be operated
6. Never connect a high voltage output lead to earth ground. This defeats the floating
clearance around the power supply. Do not block the fan vents at the rear of the unit.
3-prong AC power cord provided with the power supply.
into the power supply's high voltage output jacks. These plugs have been EN 61010* certified for safety compliance for use with PowerPac power supplies. Use of other plugs or banana jacks is done at the user's own risk and is not recommended by Bio-Rad. When inserting and removing the molded two-prong plug, always grasp the plug by the molded support at the rear of the plug. Do not grasp the individual prong ends.
condensation can short the internal electrical circuits of the power supply.
immediately. However, when removing the power supply from the cold room, let the unit equilibrate to room temperature for a minimum of 2 hours before using it.
electrical isolation of the power supply and exposes the user to potentially lethal high voltages.
Important
This instrument is intended for laboratory use only. This product conforms to the class A standards for Electromagnetic Emissions,
intended for laboratory equipment applications. It is possible that emissions from this product may interfere with some sensitive appliances when placed nearby or on the same circuit as those appliances. The user should be aware of this potential and take appropriate measures to avoid interference.
Bio-Rad's PowerPac power supplies are designed and certified to meet EN 61010* safety standards. Certified products are safe to use when operated in accordance with the instruction manual. This safety certification does not extend to electrophoresis cells or accessories that are not EN 61010 certified, even when connected to this power supply.
This instrument should not be modified or altered in any way. Alteration of this instrument will void the manufacturer's warranty, void the EN 61010 certification, and create a potential safety hazard for the user.
Bio-Rad is not responsible for any injury or damage caused by the use of this instrument for purposes other than those for which it is intended, or by modifications of the instrument not performed by Bio-Rad or an authorized agent.
*EN 61010 is an internationally accepted electrical safety standard for laboratory instruments.
Page 6
Page 7
Section 1 Introduction
1.1 Overview
The PowerPac HC power supply is designed to provide constant voltage, current or power for electrophoresis applications. The power supply operates at the value specified for the constant parameter, with limits for the other parameters. To prevent damage to the electrophoresis cell, the PowerPac HC provides automatic crossover to constant current, constant voltage, or constant power, depending on which set limit is first reached. If the set limit of the non-constant parameter is reached, the power supply will switch, making the limiting non-constant parameter the new constant.
Output specifications:
Voltage: Adjustable from 5 to 250 volts (V), in 1 V increments
Current : Adjustable from 0.01 to 3.00 amperes (A), in 0.01 A increments
Power: Adjustable from 0.05 to 300 watts (W) in 1 W increments
Output jacks: Four sets of output jacks are provided to facilitate connection of
up to 4 identical electrophoresis cells simultaneously
Fig. 1. Front View.
Fig. 2. Rear View.
stop/home
setup
FRONT PANEL
OUTPUT JACKS
abc
12
jklghi
45
tuvpqrs
7
8
EDIT
0CE
def
3
mno
6
wxyz
9
1
Page 8
The PowerPac HC power supply has the following features:
Programmable constant voltage, constant current or constant power operation with automatic crossover.
LCD screen displays all running parameters at once.
Timer control to 99 hr, 59 min.
Programmable methods with up to 3 steps.
Pause mode for editing run parameters.
Safety Features include automatic detection of no-load, short circuit, rapid resistance change, ground leak, fan failure and system overheating.
Automatic completion of a run interrupted by power failure.
Adjustable LCD display contrast.
Input power 100–120/220–240 VAC, auto-switching
Four output terminals.
Stackable case with adjustable viewing angle via flip down legs (See Figure 3)
Fig. 3. Front View with Legs in Lowered Position.
1.2 Unpacking
When you receive your power supply, carefully inspect the container for any damage which may have occurred in shipping. Severe damage to the container may indicate damage to the power supply itself. If you suspect damage to the unit, immediately file a claim with the carrier in accordance with their instructions before contacting Bio-Rad Laboratories.
After unpacking the PowerPac HC, remove the plastic film from the translucent green top case. The plastic film may leave a residue. If so, clean with a soft, damp cloth. Also remove the die-cut plastic film covering the display window. Contents include:
PowerPac HC power supply
Power cord
Instruction manual
Warranty card
Declaration of conformity If any part is missing or damaged, contact Bio-Rad Laboratories immediately.
2
Page 9
Section 2 Control Features
RUN/PAUSE KEY
STOP/HOME KEY
SETUP KEY
SOFT KEYS
EDIT KEY
Fig. 4. Front Panel.
Key Description
Run/Pause Key
Starts or pauses a run. Pausing allows editing of the run parameter values.
The corresponding LED indicates the status of the power supply.
Stop Key
Stops the run. The display will show final run values alternating with the "Run Stopped" screen.
A second stroke to the stop key clears the final run values of the previous run and prepares the screen to repeat the run.
Reset Key
The reset key can be used at any time during programming to clear previous run parameters and reset to the default startup screen.
ARROW KEYS
ALPHANUMERIC KEYPAD
CLEAR ENTRY KEY
IR PORT
In the setup menu, the reset key returns settings to default values.
Setup Key
Setup key accesses the set up menu. The setup menu allows changes to defaults such as power failure detection, rapid resistance change detection, alarm, and display contrast. See pages 6–7 for details.
Constant Key
Selects constant voltage, constant current or constant power.
The lighted LED below the parameters on the screen indicates the constant parameter selected.
Cursor Keys
Navigates among the different entry fields on the screen.
Values Keys
Changes the value of the selected parameter (indicated by the cursor). If the values key is pressed continuously for more than 10 counts in either direction, +/-, the value will increase/decrease in increments of 10 to reach the desired value faster.
3
Page 10
Section 3 Setup and Operation
3.1 Setup and Programming a Run
This section guides the user through setting up the power supply for a run, connecting the electrophoresis cell(s), programming, and running.
STEP PROCEDURE DESCRIPTION
1. Connect cell(s) The power leads are color coordinated to
Fig. 5. Power Leads Connected Correctly. Fig. 6. Power Leads Connected Incorrectly.
the output terminals.
indicates high voltages.
Note: Power leads must be inserted perpendicular to the curve of the case.
2. Turn power on. Press the power switch located on the
right side of the unit to the ON ( I ) position.
The default settings are indicated in the home screen below:
0
0
0
0
Note: The firmware version number appears on the screen momentarily when the unit is turned ON.
3. Select constant parameter Press the constant key which toggles
between volts (V), amperes (A), or watts (W).
An LED indicates the constant parameter: V, A, and W. The cursor also indicates the selected constant parameter, activating the field to enter a value. The non-constant parameters display maximum values as a default and will act as output limits (see step 6 to change output limits).
00
00
Note: Constant voltage is the default setting in the startup screen.
4
Page 11
4. Enter constant value Press the Values keys to increase or
decrease the selected value. The keys change the value in increments of 1, however after a few seconds of holding the key in either direction, +/-, the value changes in increments of 10. If the run is going to be untimed or continuous, and the limits are the maximum output, proceed to step # 9.
5. Program time (optional) Move the cursor to the time field using
the left/right arrow keys and enter the values as in step 4.
0:00 Note: Time is comprised of two fields for hr:min hours and minutes. If values increase
past 59 min in the minutes field, an hour will be added in the hours field and the minutes field will restart at 00.
If no time is entered the run will be considered untimed or continuous.
6. Change limits (optional) Move the cursor to the fields of the
non-constant parameters and enter the values as in step 4.
7. Add a step or steps (optional) After time has been entered in the first
step, up to 2 additional steps may be added. Move the cursor to the step field in the upper left of the screen and press the + Value key to add a step, then proceed to enter run parameters as in steps 3–6 above. Note: To review a multistep program, use the cursor keys to move the cursor to the step field and use +/- Values keys to scroll through the steps.
8. Reset The Reset key may be pressed at any
time during the programming process to clear all entries and return to the startup screen (single step with default values).
9. Start the run Press the start/pause key.
The run may be started at any time after step 4.
3.2 Editing Run Parameters During a Run
STEP PROCEDURE DESCRIPTION
1. Pause Press the start/pause key while a run is
in progress.
Note: The LED next to the standing man lights up and the timer on the screen stops.
5
Page 12
STEP PROCEDURE DESCRIPTION
2. Editing Use the cursor keys and values keys to
change parameters. For multiple-step runs, future steps may be accessed and changed, yet the original number of steps cannot be changed.
3. Resume the run Resume the run by pressing the
start/pause key again.
3.3 Ending a Run
A run may be terminated in several ways: by user intervention, automatically if time was programmed, or due to a power failure (if power failure detection (PFd) is not enabled). The display will indicate how the run was ended as well as its final parameters.
Run MESSAGE EXPLANATION
Timed or Run Stopped Run manually stopped by pressing the continuous stop key
Timed Run Completed Timed run ends automatically
Timed Run Completed Run ends automatically, a power failure PFd ON after power fail occurred during the run which was
resumed automatically.
Timed or Stopped after Run manually stopped by pressing the continuous power fail stop key, a power failure occurred during PFd ON the run which was resumed automatically.
PFd OFF Aborted after A power failure occurred and the run was
power fail terminated.
When the stop key (or any key) is pressed again, the original settings will appear and the PowerPac is ready to start the run again (except after a power failure with PFd off. See Troubleshooting for details). Changes made during Pause mode will be lost.
3.4 Changing System Defaults using the Setup Menu
The setup menu allows the user to activate the automatic power recovery function, turn off the rapid resistance change safety function, turn off the alarm, and adjust the display contrast.
Power Failure Detection (PFd)
The default setting for Power Failure Detection (PFd) is OFF. If a power failure (or multiple power failures) occurs during a run when PFd is ON, the PowerPac HC will automatically resume the run when power is restored. This occurs whether the run is timed or continuous. After the run has been completed or stopped, PFd is reset to its default setting, OFF.
Note: The PowerPac HC cannot distinguish between a power failure and being turned off at the power switch. If PFd is ON, and the PowerPac HC is turned off at the power switch during a run, the power supply will resume the run when turned back on, possibly creating an unsafe condition. Therefore, it is important always to stop a run before turning off the power supply.
6
Page 13
Rapid Resistance Change Detection (RRCd)
The Rapid Resistance Change Detection safety feature alerts the user when the resistance load detected by the PowerPac HC has changed by more than 25% over a brief time-period, possibly indicating an unsafe condition. Certain applications, such as use with the DCode or DGene systems, might exhibit resistance changes ³ 25% and trigger a shutdown erroneously before completion of the run if RRCd is not turned off. For these applications, RDCd should be turned off. After a run has been completed or stopped with RRCd turned OFF, RRCd is reset to its default setting, ON.
Alarm and Contrast Settings
The PowerPac HC has an alarm that signals when a timed run is started, completed, or when an error message occurrs. The alarm can be turned off in the setup menu, if desired. If the alarm is turned off, it will not signal error messages.The display contrast can also be adjusted through the PowerPac HC’s setup menu. Both of these user adjustments are preserved in memory, regardless of completion of the run or power shutdown. The Reset key may be used to return these settings to their defaults in the setup menu.
STEP PROCEDURE DESCRIPTION
1. Access the setup menu Press the Setup key. This key is not active during run and pause modes.
Default settings are as follows:
PFd: RRCd:OFF
ON
Note: The setup menu cannot be accessed during a run.
2. Select feature Use the cursor keys to select a feature to modify. Four can be adjusted.
1. PFd: Power failure detection.
2. RRCd: Rapid Resistance Change Detection
3. : Alarm
4. : Contrast
3. Change values The Values keys toggle between ON/OFF for Pfd, RRCd, and alarm setting, and adjust screen contrast from 1 through 9.
4. Reset, if needed. Press the reset button to return to default values:
PFd: RRCd:OFF
ON
YES 5
YES 5
5. Exit the Setup menu Press the Setup key again.
7
Page 14
Section 4 Maintenance and Troubleshooting
4.1 Maintenance
The PowerPac HC requires little maintenance to assure reliable operation. To clean the case, first unplug the power supply. Use a damp cloth to wipe down the outer case. Ethanol may be used to wipe the case, however, prolonged contact or use may cloud the green translucent case.
4.2 Troubleshooting
4.2.1 Basic Troubleshooting
Problem Cause Solution
No display/lights/fan 1. No AC power. 1. Check if PowerPac HC is unplugged,
2. Blown fuse. 2. Replace fuse. See Section 4.3 for
Repeated blown fuses Hardware failure Contact Bio-Rad Technical Resources. Leads from cell are Output terminals for the Use the PowerPac Adaptor,
not long enough to fit PowerPac HC are (Catalog number 165-5061) which output jacks recessed 16 mm to accommodates most standard 4 mm
meet safety regulations. banana plugs, to make a secure Some leads are not electrical connection. long enough to make Note: Use of the PowerPac Adaptor electrical connection. voids EN61010 safety provisions.
or problem with AC power source, or power switch is in off position "O".
details.
4.2.2 Power Failure Detection (PFd)
Power failure related messages provide information in the event of a power interruption during a run. The power supply and the electrophoresis cell are not at fault, and there is no actual corrective action.
The power failure detection default setting is OFF. That means that if a power outage occurs during a run, the run will be aborted and the run parameters will be lost. According to the application and user preferences, PFd may be turned ON, which will enable the PowerPac HC to resume the run automatically when power is restored.
8
Page 15
If a power failure occurs during a run, there are several possible scenarios which are described as follows:
Message Displayed Explanation
RUN ABORTED PFd was OFF. AFTER POWER FAIL The run, either timed or continuous, was aborted after a power
*PRESS STOP KEY After pressing the stop key to reset, the original run parameters TO RESET are no longer available.
RUN STOPPED PFd was ON when a power failure occurred and the run was AFTER POWER FAIL resumed automatically.
(Message alternates After pressing the stop key to reset, the original run parameters with final run parameters) are available for re-run.
RUN COMPLETED PFd was ON when a power failure occurred and the run was AFTER POWER FAIL resumed automatically.
(Message alternates run time was completed. final run parameters) After pressing the stop key to reset, the original run parameters
failure occurred.
The run, either timed or continuous, has been terminated by pressing the stop key.
Note: PFd defaults automatically to OFF after pressing the stop key.
The run was timed and automatically terminated when the set
are available for re-run. Note: PFd defaults automatically to OFF when the run is completed.
4.2.3 User Serviceable Error Messages
The following user serviceable messages may appear, interrupting a run. Refer to this section for their cause and solution. Follow instructions on the screen to clear the error message and resume operation.
Problem/error message Cause Solution
E01 NO LOAD No load detected. Verify all electrical connections. DETECTED • Electrophoresis cell Verify buffer levels are appropriate.
CELL & CABLES power supply. Verify the electrophoresis application's
* PRESS STOP KEY power requirements match PowerPac
E02 SHORT CIRCUIT Short circuit detected • Insure all electrical connections
CELL & CABLES current is greater cables and wire electrodes are in
* POWER OFF/ON than 3.0 A). good shape.
E04 REGULATING Regulation Error. Ensure the electrophoresis application BELOW 5V Current drawn by the settings and power requirements
METHOD VALUES exceeds the set current
* PRESS STOP KEY limit of the unit, at less
not connected to the
• The current load is HC output range. less than 0.01 A.
(over current, load are making good contact and the
• Accidental shorting of • Verify buffer levels are appropriate. output leads.
• Shorting due to wrong connections.
electrophoresis system match PowerPac HC output range
than 5 V.
9
Page 16
Problem Cause Solution
E09 RESISTANCE Change in load Check and correct any potential CHANGE DETECTED resistance detected. resistance problem.
CELL & CABLES detects drastic changes Verify buffer levels are appropriate.
* PRESS STOP KEY in resistance which
E11 GROUND LEAK Insulation failure in the Check electrical connections and check DETECTED electrical connections cell/chiller system for leaks. Ensure
CELL & CHILLER supply has caused a an insulated, dry surface.
* POWER OFF/ON current flow that may
E16 FAN FAILURE The fan is not Inspect the fan grid to check if the fan
FAN GRID working, either due to a supply is turned "ON". Make sure no
* POWER OFF/ON electrical failure or foreign material has penetrated the
E17 SYSTEM Fan vents blocked and or Clear a space of at least 6 cm around OVERHEATING ambient temperature the fan vents to improve airflow. Press
FAN & RM. TEMP restart. Operate the unit within the
* POWER OFF/ON recommended temperature range
The PowerPac HC Verify all electrical connections.
may indicate failure of the cell's power leads or a loose output connection.
• Loose ouput connections leading to intermittent connection to the loads.
• Cells added or removed during the run.
Note: Certain applications exhibit intrinsic fluctuations in resistance that can trigger the Resistance Change error message. If this is the case, the Resistance Change detection feature can be deactivated to allow an uninterrupted completion of the run. See Section 3.4 for details on turning off RRCd.
Caution: Deactivating this safety feature increases the possibility that a failure of the power leads or loose output connections will not be detected.
outside the power that the electrophoresis cell rests on
create an unsafe condition.
blade is rotating when the power
physical obstruction. fan vent. If problem persists, contact
Bio-Rad.
too high. any key to go to the home screen to
(0 to 40°C)
10
Page 17
4.2.4 Bio-Rad Serviceable Error Messages
Error code Cause Solution
ERROR 03 Output regulation Restart the unit by turning the power * POWER OFF/ON problem switch OFF and ON.
ERROR 08 Output regulation * POWER OFF/ON problem
ERROR 99 Hardware failure * POWER OFF/ON
ERROR 98 Hardware failure * POWER OFF/ON
Contact Bio-Rad Technical Resources if problem persists.
4.3 Replacing a fuse
If there is no display, lights, or fan, and the PowerPac HC is plugged into a working AC power outlet with the power switch in the "ON" position, the fuse may need to be replaced.
1. Disconnect the power cord from the electrical outlet.
2. Insert a flat blade screwdriver into the notches of the power entry module's fuse drawer to release it. See Figure 7.
3. Remove the fuse from the fuse drawer. Replace it with 8A, 250 V, 5x20 mm, type T fuse (Bio-Rad part number 900-8933).
4. Re-insert the fuse drawer into its position. Press the fuse drawer gently until it snaps into place.
The unit is now ready for use.
FUSE
DRAWER
NOTCHES
Fig. 7. Rear View Showing Fuse Drawer with Notches.
Note: Repeated blowing of the fuse indicates a hardware failure. Contact Bio-Rad
Technical Support.
4.4 Expediting Technical Support
Make sure the following information is readily available before contacting Bio-Rad. Product model no: Located in the sticker on the bottom of the unit
(example: PowerPac HC Power Supply). Serial number: Located in the sticker on the bottom of the unit
(example: xxxBRxxxxx). Software version: The PowerPac HC displays the software version momentarily
after switching the power ON (example: Firmware 1.00). State clearly the error code, error message or anomaly, and the conditions that
originated the problem, including run parameters (V, A, and W) as well as the electrophoresis cell and buffer system.
11
Page 18
Appendix A Specifications
Output specifications 250 V, 3.0 A, 300 W Output range (programmable) 5–250 V, fully adjustable in 1 V increments
increments
Type of output Constant voltage, current, or power with
Output terminals 4 pair recessed banana jacks in parallel Timer up to 99 hr, 59 min Pause/resume function yes Display 2 line, 16 character backlit LCD Operating conditions 0–40°C; 0–95% humidity in the absence of
Safety compliance EN61010 EMI Conforms to CE standards for emissions and
Safety features No load detection; sudden load change
Input protection Fuse on hot and neutral Input power (nominal) 100–120/220–240 VAC, autoswitching Dimensions (D x W x H) 28.5 X 25 X 8 cm Weight 2.0 kg
0.01–3.0 A, fully adjustable in 0.01 A
1–300 W, fully adjustable in 1 W increments
automatic crossover
condensation
immunity class A, tested only at 230 V. TÛV EMC certification
detection; ground leak detection; overload/short circuit detection; overvoltage protection; overheating protection
12
Page 19
Appendix B Warranty and Ordering Information
Warranty
The PowerPac HC power supply is covered by a standard Bio-Rad Laboratories warranty. Contact your local Bio-Rad representative for details of the warranty. If any defects should occur during this warranty period, Bio-Rad Laboratories will replace the defective parts without charge. However, the following defects are specifically excluded:
1. Defects caused by improper operation.
2. Repair or modification done by anyone other than Bio-Rad Laboratories or their authorized agent.
3. Use with cables or connectors not specified by Bio-Rad Laboratories for this power supply.
4. Deliberate or accidental misuse.
5. Damage caused by disaster.
For inquiry or request for repair service, contact your local Bio-Rad office.
Warranty Information
Model:
Serial Number:
Date of Delivery:
Warranty Period:
Ordering Information
Catalog Number Description
164-5052 PowerPac HC power supply, 100–120/220–240V 165-5061 PowerPac Adaptor, qty 1 165-5066 PowerPac Adaptor, qty 2 900-8933 Replacement Fuse, 8 A, 250 V, 5 x 20 mm, Type T, 1
13
Page 20
Page 21
Page 22
Page 23
Page 24
Life Science Group
Sig 1212
4006222 Rev D US/EG
Bio-Rad Laboratories, Inc.
Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austria 01 877 89 01 Belgium 09 385 55 11 Brazil 55 11 5044 569 9 Canada 905 364 343 5 China 86 21 6169 8500 Czech Republic 420 241 430 532 Denmark 44 52 10 00 Finland 09 804 22 00
France 01 47 95 69 65 Germany 089 31 884 0 Greece 30 210 9532 220 Hong Kong 852 2789 3300 Hungary 36 1 459 6100 India 91 124 4029300
Israel 03 963 6050 Italy 39 02 216091 Japan 03 6361 7000 Kore a 82 2 3473 4460 Mexico 52 555 488 7670 Th e Netherlands 0318 540666 New Zealand 64 9 415 2280 Norway 23 38 41 30 Poland 48 22 331 99 99 Portugal 351 21 472 7700 Russia 7 495 721 14 04 Singapore 65 6415 3188 South Africa 27 861 246 723 Spain 34 91 590 5200 Sweden 08 555 12700 Switzerland 026 674 55 05 Tai wa n 8 86 2 2578 7189 Thailand 800 88 22 88 United Kingdom 020 8328 2000
Loading...