Bio-Rad 2110 Fraction Collector User Manual

Page 1
Model 2110
Fraction
Collector
Instruction
Manual
Catalog Numbers
731-8120
731-8122
Page 2
Model _________________________________________
Serial No. ______________________________________
Date of Delivery_________________________________
Warranty Period _________________________________
Unless otherwise specified, instruments sold by Bio-Rad Laboratories are under warranty for 1 year against defects in materials and workman­ship.
If any defects should occur during this warranty period, Bio-Rad will replace the defective parts without charge. However, the following defects are specifically excluded:
1. Defects caused by improper operation
2. Repair or modification done by anyone other than Bio-Rad
Laboratories or their authorized agent.
3. Use of fittings or other spare parts supplied by anyone other than Bio-
Rad Laboratories.
4. Damage caused by deliberate or accidental misuse.
5. Damage caused by disaster.
6. Damage due to use of improper solvent or sample.
This warranty does not apply to tubing and fuses.
For inquiry or request for repair service, contact Bio-Rad Laboratories after confirming the model and serial number of your instru­ment.
For Technical Service Call Your Local Bio-Rad Office or in the U.S. call 1-800-4BIORAD(1-800-424-6723)
Page 3
Table of Contents
Section 1 Safety .............................................................................. 1
Section 2 Introduction ................................................................... 2
Section 3 Unpacking and Voltage Conversion ............................ 3
3.1 Unpacking Instructions ............................................................ 3
3.2 Voltage Conversion ..................................................................
3
Section 4 Physical Description and Control Features................ 5
4.1 Physical Description.................................................................
5
4.2 Front Panel Functions ..............................................................
7
4.3 Rear Panel Connections ...........................................................
8
Section 5 Instrument Set-Up......................................................... 9
5.1 Tubing Installation ...................................................................
9
5.2 Carousel Installation.................................................................
11
5.3 Collection Tube Installation.....................................................
11
5.4 Waste Collection Port...............................................................
12
Section 6 Stand Alone Operation ................................................. 12
6.1 Stop (Standby) and Run ...........................................................
12
6.2 Collection Modes .....................................................................
12
6.3 Operation without Direct Connection to a System..................
14
6.4 Manual Advance.......................................................................
15
6.5 Reset/End of Run ..................................................................... 15
Section 7 Remote Operation of the Model 2110
Fraction Collector ......................................................... 15
7.1 Remote Operation with Model EP-1 Econo Pump.................. 15
7.2 Remote Operation with BioLogic LP ...................................... 16
7.3 Remote Operation with BioLogic system................................ 16
7.4 Remote Operation with other Instruments............................... 16
Section 8 Accessories ..................................................................... 18
8.1 Instrument Dust Cover ............................................................. 18
8.2 Micro Tube Adaptor ................................................................. 18
Section 9 Cleaning and Maintenance........................................... 19
9.1 Drop Sensor.............................................................................. 19
9.2 Instrument Case, Carousel, and Drainage Tube....................... 19
Section 10 Troubleshooting ............................................................. 19
Appendix A Technical Specifications ................................................ 21
Page 4
Appendix B Model 2110 Fraction Collector Procedure
for Adjusting Drop Detector Signal Level .................. 22
Appendix C Ordering Information................................................... 24
Page 5
Section 1 Safety
Disconnect supply before servicing. No user serviceable parts inside, refer servicing to Bio-Rad service personnel.
High leakage current, ensure proper grounding.
This instrument is intended for laboratory use only.
This product conforms to the “Class A” standards for electro­magnetic emmissions intended for laboratory equipment applications. It is possible that emissions from this product may interfere with some sensitive appliances when placed nearby or in the same circuit as those appliances. The user should be aware of this potential and take appropriate measures to avoid interference.
1
!
!
Page 6
Section 2 Introduction
Fig. 2.1. Model 2110 Fraction Collector.
The Model 2110 Fraction Collector (Figure 2.1), shown with the optional instrument dust cover, is designed to be used with gravity feed columns, preparative electrophoresis and liquid chromatography systems.
The Model 2110 Fraction Collector will accurately collect up to 80 fractions, at a flow rate of up to 40 ml/min. The Model 2110 Fraction Collector accepts standard 13 x 100 mm test tubes or, with an optional adaptor, 1.5 ml micro test tubes. Tubes are advanced by a motor-driven
22
Page 7
carousel where tube changes are initiated upon reaching a predetermined drop count (1-999 drops) or time interval (0.05-99.9 minutes), or upon command from an external source.
The silicone drop former insures uniform drop size for accurate drop
count measurement. The drop former will accommodate either 1¼
1
¼16-inch inlet tubing, without the need for additional fittings.
The Model 2110 Fraction Collector is spill-proof and solvent resistant. A drain tube at the rear of the case channels all spills to waste. The mem­brane front panel is sealed to the case and the carousel drive is positioned in such a way that it will not be affected by spills.
Section 3 Unpacking and Voltage Conversion
3.1 Unpacking Instructions
Lift the instrument out of its packing by grasping the instrument case. Do not lift by the drop arm. Remove the power cord and bag of acces­sories. Remove the fraction collector from its protective bag. Check off all parts against the supplied packing list. You may want to retain the origi­nal packaging material in the event the instrument needs to be returned for service. If repackaging is necessary, be careful to position the carousel so that it is not resting against the motor drive gear.
8-inch or
Warning: The Model 2110 Fraction Collector is shipped in its 120 V
version. For 220 V operation, refer to the procedure below.
3.2 Voltage Conversion
Prior to connecting the power cord to the power entry module and power mains, make sure the voltage indicated on the fraction collector power entry module matches your line voltage. If it does not, use the following procedure to make the conversion. (See Figure 3.1.)
1. Insure that the unit is disconnected from the wall socket.
2. Remove the fuse drawer with a small-blade screwdriver or similar tool.
3. Pull the fuse holder out of the fuse drawer and ensure that the voltage
shown in the window corresponds to the voltage you are using (100, 120,
220, 240 V ). To change the voltage, grasp the fuse drawer and withdraw
it (see Figure 3.1). Rotate the fuse drawer so that the voltage shown is cor-
rect, and reinsert it. The fuse holder is keyed and cannot be installed
improperly. The Model 2110 Fraction Collector uses 0.20 A fuses for
100/120 V operation, and 0.10 A fuses for 220/240 V operation.
3
Page 8
4. Reinsert the fuse drawer in the power entry module, with the locking tab to the left. The voltage indicator will read right-side-up if the drawer is oriented correctly. Push the fuse drawer in gently until it snaps into place.
5. Connect the power cord to the power entry module and power mains.
6. There is no power switch on the fraction collector. When connected to an AC power source, the Model 2110 Fraction Collector will come on in the Stop or Set mode.
Fig. 3.1. Voltage conversion.
Fuse drawer (110 V configuration)
Fuse drawer (220 V configuration)
4
Page 9
Section 4 Physical Description and Control Features
4.1 Physical Description
Fig 4.1. Model 2110 Fraction Collector features.
Feature Function Drop arm Directs flow to the collection tubes.
• Collection position, rotated forward, the drop arm is directly above each collection tube.
• Standby position, rotated rearward, the drop arm allows installation of the carousel or allows flow to drip onto the case and to waste. To change positions, gently lift and rotate the drop arm.
5
Inlet tubing
Arm cap
Drop former
Drive seal
Drop arm
Drive gear
Pivot shaft
Slider
Carousel positioning guides
Page 10
Feature Function Carousel Holds the collection tubes. The tube positions are
numbered 1-80 on the flange of the carousel, and line up with the center of each tube. The carousel is designed so that any liquid spilled in it will flow to the center and out the rear of the base. The carousel may also be used for storage of test tubes.
Drop former Seals against the inlet tubing (16
" or
1
¼8
" OD) to
form uniform size drops.
Arm cap Holds the drop former in place.
Drive gear Engages the carousel and advances it.
Drive seal Protects the electric drive from direct spills.
Waste line Collects drainage or spills and directs it to waste.
See Section 5.5 for connection.
Inlet tubing Inlet tubing (not included) can be any rigid or
flexible tubing with
1
¼16" or 8" OD, usually
coming directly from a detector.
6
Page 11
4.2 Front Panel Functions
MODEL 2110 FRACTION COLLECTOR
Fig. 4.2. Front panel controls.
Feature Function Run/Stop key The Run/Stop key is used to select
Time/Drop key The Time/Drop key is used to select
standby or active mode. In the active mode, the collector will collect by time or drops, or accept an external signal. When the collector is con­trolled by a system, both the standby and active mode lights will be lit, and the LED display will read “---”.
collection by time or by drops.
Arrow keys The Arrow keys are used for
increasing or decreasing the length of time for collection in each tube, or changing the number of drops to be collected in each tube.
ml
LED multi This three-digit display will show
min
ml/min
%
display decimal minutes or drop count that
has progressed for the tube currently collecting. When the collector is con­trolled by a system, both the standby and active mode lights will be lit, and the LED display will read “---”.
Advance key Pressing the advance key will
advance the carousel one position.
7
Page 12
Fig. 4.3 Rear Panel connections.
4.3 Rear Panel Functions
Connection Function Power entry Grounded 3-pin receptacle for the
module power cord. Also contained within is
9-pin “D” For remote operation of the collector. connector The Model 2110 can be controlled by
a four-position 100 V/120 V/ 220 V/ 240 V line voltage selector. The collector is shipped in its 120 V or 220 V version. For operation at other voltages, refer to Section 3.2.
Bio-Rad’s Model EP-1 Econo Pump, the BioLogic LP low pressure chromatography system, or the BioLogic System, a high resolution liquid chromatography system. The Model 2110 can also be controlled by other non-Bio-Rad equipment. See Section 7 for additional information.
8
Page 13
Section 5 Instrument Set-Up
5.1 Tubing Installation
1. Insert the drop former, narrow end downward, into the drop arm, pressing it firmly in the hole (see Figure 4.1).
2. Using a twisting motion, press the arm cap onto the drop arm to hold the drop former in place.
3. Insure that the end of the inlet tubing is cleanly cut and not flattened or deformed.
Insert tubing into the drop former, pushing down on the tubing until the drop former starts to stretch. Do not use excessive force. Then, gently pull out on the tubing until the drop former returns to its original shape.
4. The inlet tubing may be routed through the tubing holder fitting at the rear of the movable portion of the drop arm.
9
Page 14
ab
21
c
Fig. 5.1. Carousel installation.
10
Page 15
5.2 Carousel Installation
The Model 2110 Fraction Collector fills tubes from the outside of the
carousel inward. Load as many tubes as desired into the carousel, starting at the outside end of the spiral and moving inward. The carousel need not be fully loaded, since any eluant flowing after the last tube will drain through the carousel into the well in the case and out the central drain tube located at the rear of the instrument.
1. Swing the drop arm to the waste (rear) position by gently lifting and rotating (Figure 5.1a).
2. Hold the carousel so that the notch in its outline is toward the rear of the instrument.
3. Place the carousel on the pivot shaft, holding it about 1¼2" above its seated position.
4. Push the carousel and the pivot shaft toward the rear of the instrument, settling it down over the drive gear so that the gear is trapped near the start of the spiral.
5. Holding the gear teeth free from engagement, turn the carousel coun­terclockwise until it stops against the gear (Figure 5.1b). Release the carousel and it will spring into engagement on the drive gear. Engagement may be confirmed by trying to move the carousel gently (Figure 5.1c).
6. Check the alignment with the two positioning guides located on the top front of the instrument case. The number corresponds to the first tube that will be filled.
Swing the drop arm into the collect (front) position. Make sure the
drive gear is fully seated into the grooved underside of the carousel. This automatically aligns the drop former over the first tube.
5.3 Collection Tube Installation
Load the carousel with up to eighty 13 x 100 mm test tubes. You may
find a tube which fits too tightly into the carousel. It is best to discard these rather than risk breaking the tube upon installation. Tubes 39 through 53 are raised above tubes 1-38 to make tube removal easier.
To locate an individual tube in a carousel, find its number on the
angled flange. The tube is in line with the number and the center of the carousel. The top row of numbers corresponds to the outermost row of tubes, the middle row corresponds to the next row of tubes, and the bot­tom row to the innermost row of tubes. To divert a large amount of eluant to waste at the beginning of a run, you can simply remove the first
11
Page 16
collection tube, and the eluant will flow through the carousel to waste out the rear of the collector.
5.4 Waste Collection Port
All spills on the top of the case and carousel, as well as waste eluant within the carousel, drain to a central waste-collection port at the rear of the instrument case. From this port, a drain tube passes out through the rear of the case. This tube may be directed into a waste-collection vessel, or an extension tube may be attached to the provided barbed union fitting.
Note: To provide proper drainage, the waste-collection vessel must be
positioned so that the exit of the drain tube is no higher than the bot-
tom of the fraction collector case.
Section 6 Stand Alone Operation
This section describes the stand-alone operation of the Model 2110 Fraction Collector. Refer to Section 7 for information about connecting the Model 2110 to the Model EP-1 Econo Pump, the BioLogic LP, the BioLogic System, or Bio-Rad’s preparative electrophoresis systems..
6.1 Stop (Standby) and Run
The Model 2110 Fraction Collector powers up automatically in the Stop or Standby mode, with the LED lit. When in the Stop mode, the collection mode, either Time or Drops can be selected, and the time inter­val or number of drops to be collected per tube can be set. When in the Run mode, the Arrow keys are inactivated.
6.2 Collection Modes
The Model 2110 Fraction Collector will collect fractions by counting the number of drops or by a specified time interval per tube. The collec­tion mode is selected using the key on the front panel. In either mode, time and drop count are continuously monitored, so that you can switch between the two collection modes, monitoring time and drop count since the last tube advance.
12
Page 17
Drop Collection Mode
The fraction collector counts drops at flow rates up to ~5 ml/min. Drop counting is performed by the sensor in the drop arm. If the flow rate exceeds 5 ml/ min., counting will not be accurate, and use of the Time mode is recommended.
1. Plug the Model 2110 Fraction Collector into a grounded wall outlet. The collector will power up in the Stop, or Standby mode, with the LED lit.
2. Press the Time/Drop key until the LED is lit.
3. Select the number of drops to be collected per collection tube by pressing the keys. The LED display will display the number of drops to be collected.
4. Insure that the carousel is in the desired posi­tion, and press the Run/Stop key to begin col­lection. The LED will be lit.
5. The fraction collector will reset the display to ‘0’ and begin counting drops up to the drop limit set, and then advance. The display reads ‘A’ during each carousel advance.
13
Page 18
Time Mode
The fraction collector collects fractions at intervals of .01 minutes, for times per tube in the range of .05 to 9.99 minutes, and at.10 minute inter­vals for times per tube in the range of 10.0 to 99.9 minutes.
1. Press the Run/Stop key until the LED is lit.
2. Press the Time/Drop key until the LED is lit.
3. Select the time interval per tube by pressing the keys. The LED multi-display will display
the time interval selected.
4. Insure that the carousel is in the desired position.
Press the key to begin the program. The LED will be lit.
5. The fraction collector will reset the display to ‘0’,
and begin counting time up to the set time inter­val, and then advance.
Note: When the Model 2110 Fraction Collector reaches the preset time interval in a time mode collection scheme, it will collect one extra drop prior to advancing. This minimizes the chance of a drop falling between tubes during carousel advance. If the delay between drops is greater than 5% of the preset time interval, the collector will advance to the next fraction instead of waiting for one extra drop.
6.3 Operation without Direct Connection to a System
The Model 2110 can be used in stand-alone mode with any system. Simply set up the collector in time mode or drop mode and press “start” at the initiation of the method.
Alternatively, a light-weight gravity feed column (such as 2 ml Poly-Prep
®
columns, Econo-Pac®columns, and small Econo-Column
®
chromatography columns) can be connected directly to the Model 2110 Fraction Collector, avoiding the need for interconnecting tubing. A two-way stopcock is used to make the connection and control eluant flow.
14
ml
min
ml/min
%
Page 19
1. Attach the stopcock into the hole in the top of the drop arm. Position the stopcock so that its handle overhangs the end of the arm and is free to turn.
2. Attach the column to the stopcock, making sure both stopcock and column are firmly connected.
6.4 Manual Advance
The Model 2110 Fraction Collector may be manually advanced to the
next tube by pressing the key. This key is functional in both Stop/Standby and Run modes. When pressed during a run, the drop and time counts will automatically reset to 0, and restart the time or drop count.
6.5 Reset/End of Run
If the Model 2110 Fraction Collector is allowed to complete 80
advances, the display will read END. To reset, push twice. (Note: Manual advances are not counted.) Reposition the carousel to begin collection again.
Section 7 Remote Operation of Model 2110
The Model 2110 Fraction Collector may be controlled remotely.
7.1 Remote Operation with Model EP-1 Econo
Pump
Use System Cable 1 (catalog number 731-8261) to connect the Econo
Pump to the fraction collector. Connect the 9-pin “D” connector to rear of the Model 2110, and the mini-DIN plug to the rear port of the Econo Pump with the collector icon. (See Econo Pump manual if further explanation is required.) In order for the Econo Pump to remotely operate the Model 2110, a method must be programmed into the Econo Pump that calls for fraction collection. You will be prompted to program the fraction size (volume), the void volume, and the total volume collected. Please note that to correctly use the void volume function, a Model SV-3 Diverter/Bypass valve (catalog number 731-8322) must be connected. When programming is complete, the LED display underneath the fraction collector icon on the front panel of the Econo Pump will light and the Model 2110 display will read “---” indicating that the collector is controlled by a system.
15
Page 20
7.2 Remote Operation with BioLogic LP
Use System Cable 1 (catalog number 731-8261) to connect the BioLogic LP to the fraction collector. Connect the 9-pin “D” connector to rear of the Model 2110, and the mini-DIN plug to the rear port of the BioLogic LP controller in the port marked FRACTION COLLECTOR. (See BioLogic LP manual if further explanation is required.) The BioLogic LP enables sophisticated collection, including programmed collection windows, peak detection, and a combination of peak detection and collection windows. The BioLogic LP sends “advance” commands to the fraction collector.
7.3 Remote Operation with BioLogic System
Use System Cable 5 (catalog number 731-8265) to connect the BioLogic System to the fraction collector. Connect the 9-pin “D” connec­tor to rear of the Model 2110. Connect the black wire to the FC ADV port of the AUX connector, which is port #5. Connect the white wire to the GND port of the AUX connector, which is port #9. Refer to the separate instruction sheet for the wiring diagram and color code for System Cable
5.
To connect the wires to the AUX port of the BioLogic System, turn the small flathead screwdriver counterclockwise to back out the screw. Insert the bare wire into the port and turn the screwdriver clockwise to secure the connection. The BioLogic System enables sophisticated collection, including programmed collection windows, peak detection, and a combination of peak detection and collection windows. The BioLogic System sends “advance” commands to the fraction collector.
7.4 Remote Operation with other Instruments
Use System Cable 5 (catalog number 731-8265) to connect the Model 2110 Fraction Collector to other instruments. Signals are accessible via the 9-pin “D” connector on the back of the instrument. All of the input and output signals, with the exception of the Advanced Contact pins 7 and 8, are TTL active low.
Drop counting is performed by the sensor in the drop arm.
Pin Function Signal Type
1 External Advance Input 2 Drops Output 3 Advanced Output 4 Ready Output 5 Alarm Output
16
Page 21
Pin Function Signal Type
6 External Start Input 7 Advanced Contact 1 Output 8 Advanced Contact 2 Output 9 Logic Ground
External Advance (Pin 1)
The External Advance line performs two functions. It causes the instrument to enter the Remote mode when held low for longer than two seconds, and it causes the carousel to advance one position each time a high-going pulse is applied. To use the control line to advance the carousel, connect pin 1 of the 9-pin “D” connector to pin 9 and hold for at least 100 msec, then release. Alternatively, the carousel will advance by applying an external signal (TTL low logic level ) to pin 1 for at least 100 msec (with a return to TTL high logic level state). Be sure to connect the controlling instrument’s ground to the fraction collector’s logic ground, pin 9. When remote mode is entered (by holding pin 1 low for greater than 2 seconds) the time and drop mode features of the collector are disabled and the front panel control function is lost, with the exception of the man­ual tubing advance key, which will function normally. The remote mode is indicated by three horizontal bars on the fraction collector display.
Drops
The fraction collector will generate a 50 msec TTL pulse on pin 2 every time a drop is detected.
Advanced
Each time the carousel is advanced, a low-going TTL pulse is gener­ated on pin 3. This signal is useful for triggering the event mark input of a chart recorder to indicate fraction intervals. If the recorder being used does not support TTL signals, then see below under “Advanced Relay Contacts to trigger the event input.
Ready
A low level signal on pin 4 indicates that the fraction collector is func­tioning properly and that it is ready to accept input from a controlling device.
Alarm
If the fraction collector is unable to carry out an external advance command due to a malfunction (such as a jammed carousel), a low level will be held on pin 5 until the problem is cleared.
17
Page 22
External Start
The external start line is used to allow another instrument to remote­ly start and stop the fraction collector’s programmed functions. Applying a low pulse of at least 100 msec to pin 6 has the same effect as pressing the Run/Stop key on the front panel. That is, the first pulse will start the previously selected mode (drops or time) and the second pulse will stop the program.
Advanced Relay Contacts
Pins 7 and 8 provide a normally open set of contacts that are momen­tarily closed each time the carousel is advanced. This signal is a conve­nient way to interface with equipment that does not support TTL level sig­nals. The maximum current and voltage which can be put across these pins is 24 V DC and .25 A.
SSeeccttiioonn 88 AAcccceessssoorriieess
8.1 Instrument Dust Cover
The dust cover (catalog number 731-8136) is intended for use only when the drop arm is in the collect (forward) position. The inlet tube may be directed out the top opening or through the small opening at the bot­tom/rear of the instrument.
8.2 Micro Tube Adaptor
The micro tube adaptor is intended to convert a carousel for use with
1.5 ml micro test tubes (catalog number 223-9500).
Align the step in the outline of the adaptor with the step of the outline of the carousel. Position the adaptor so that its four legs fit into the four openings in the inner wall of the carousel. Push the adaptor downward until it is fully seated on the carousel.
Operation with micro tubes is identical to operation with standard test tubes. To prevent splashing, collect no more than 1 ml of eluant in each micro tube.
18
Page 23
Section 9 Cleaning and Maintenance
The Model 2110 Fraction Collector requires very little maintenance to
insure reliable operation. Two procedures are outlined below.
9.1 Drop Sensor
The Model 2110 Fraction Collector counts drops through an optical sensor located in the drop area inside the drop former arm. Precipitates or other foreign debris may obscure the optical path and cause erroneous drop counts. To prevent this problem, the drop sensor should be cleaned periodically. An ordinary cotton swab moistened with water can be used to clean the sensor.
9.2 Instrument Case, Carousel, and Drainage Tube
During normal operation, spills and splashes may result in precipi­tates forming on the case and carousel. Use a squirt bottle and soapy water to wash down carousel and case. The water will drain out the back of the unit through the drainage tube. Prior to squirting the instrument case, make sure the rubber seal around the gear device is in place. Insure that the waste tube is draining properly to avoid instrument damage.
Section 10 Troubleshooting
Problem Cause Remedy
Drops miss tubes Carousel misaligned Reposition carousel
Drop former missing Replace drop former Drop head misaligned Align Unit not level Place unit on level surface
Drops miss tube Flow rate greater Reduce flow rate during advance than 5 ml/min
Drops not counted Drop sensor is dirty Clean drop sensor or E08 error code
Flow rate greater than Use Time collection mode 5 ml/min
Operating in a windy Try relocating unit, or put environment a piece of tape over slot in
No power Fuses Check fuses, replace if
drop arm
necessary
19
Page 24
Fraction Collector Error Codes
There are several self-diagnostics built into the software of the Model 2110 Fraction Collector. Below are the principal error numbers and codes which occur. If other error codes appear, try unplugging and replugging the instrument, or call Bio-Rad Technical Service.
20
Page 25
Error Number Cause Remedy
E00 Hardware failure Unplug and replug to E01 mains power. If error per­E04 sists, contact your local
Bio-Rad representative.
E02 Motor jam Reposition carousel;
unplug and replug to mains power. If problem persists, apply WD-40 to the motor shaft. If motor must be replaced, order part number 8006484.
E08 Drop sensor If problem persists after E10 cleaning the drop sensor
with a wetted cotton swab, follow the Procedure for Adjusting Drop Detector Signal Level. See Appendix B. If drop sen­sor must be replaced, order part number
8006481.
Appendix A Technical Specifications
Capacity
Collection basis 1 to 999 drops in 1 drop increments
Materials of construction Case and carousel: polypropylene
Eighty 13 x 100 mm glass or plastic test tubes, or 1.5 ml micro test tubes, when using adaptor (catalog number 731-8135)
0.05 to 9.99 minutes in 0.01 minute increments
10.0 to 99.9 minutes in 0.1 minute increments
See Econo Pump manual for other collec­tion options when connected to the Econo System
Front panel: hard-coated polycarbonate Drop former: silicone rubber Waste outlet tube: polyurethane
21
Page 26
Operating temperature + 4 to + 40 °C
Storage temperature - 5 to + 71 °C
Line voltage 90-132 VAC or 180-265 VAC, 47 Hz
to 63 Hz
Power consumption 25 VA max, 15 W max, fused Fuse rating
100,120 V Operation 200 mA, 250 V, 5 x 20 mm, Fast Acting,
UL/CSA approved Ref: Schuter 034.3923 or Bussmann GMA200mA
220,240 V Operation 100 mA,250 V, 5 x 20 mm, Time Lag
(FST), VDE approved Ref: Schuter 034.3107
Size 24.0 x 33.0 x 25.0 cm (W x D x H)
Carousel does not overhang base at any point in operation
Weight 2.3 kg
Output signals See Section 6.5 of the manual
Regulatory CSA: C22.2 No. 151-M1986
Compliance TÜV:EN61010-1: 1993
CE: EN55011: 1991 CLASS A,
EN50082-2: 1995
Appendix B Model 2110 Fraction Collector Procedure for Adjusting Drop Detector Signal Level
This procedure should be performed:
1. After replacing the drop arm on the Model 2110 Fraction Collector.
2. Whenever "EO8" is displayed. In this situation, clean the optical window and light source prior to performing this procedure.
Safety: The Model 2110 Fraction Collector should be purged of all hazardous chemicals and be disconnected from the main power supply prior to conduct­ing this procedure. Ensure the case is dry. Perform this procedure on a dry sur­face. During the actual adjustment of the signal level (step 10), potentially
dangerous voltages are present inside the case - do not touch exposed wires. Use a plastic-handled screwdriver to minimize the risk of electric shock.
22
Page 27
Tools required: Small phillips-head screwdriver, small slot-head screw
driver.
Procedure:
1. Disconnect the Model 2110 collector from the power source.
2. Disconnect any tubing leading to the fraction collector, remove the yellow plastic arm cap, white rubber drop former, and carousel.
3. Clean the optical window and light source (located in drop arm), using the a cotton swab moistened with water.
4. Set fraction collector on its side with underside of the instrument facing toward you.
5. Remove the four phillips-head screws from the underside of the instru­ment. Pull gently on the rubber feet to remove the underside panel.
6. Locate the large circuit board with the power transformer. On this board, locate a small rectangular potentiometer, identified as "R3" on the circuit board. R3 is located near where the ribbon cable connects to the circuit board.
7. Read safety warning above. Use a plastic-handled screwdriver to min-
imize the risk of electric shock.
8. Press and hold down both the "up" and "down" arrow keys on the front panel.
9. While continuing to press down arrow keys, connect the fraction collec­tor to the power source. Continue to press down both keys for five sec­onds, then release.. The "time" LED will glow steadily, and the display will show a decimal point, followed by three characters (._ _ _). The frac­tion collector is now in a "diagnostic" mode.
10. Using a small flat-bladed screwdriver, adjust R3 by turning the small screw until the display reads between "0A0" and "0A5".
If a reading between "0A0" and "0A5" cannot be obtained, contact your local Bio-Rad representative.
11. Disconnect unit from the power source.
12. Replace underside panel.
13. Connect fraction collector to a power source. The Model 2110 Fraction Collector is now ready for operation.
23
Page 28
Appendix C Ordering Information
Model 2110 Fraction Collector
Catalog Number Product Description
731-8122 Model 2110 Fraction Collector, 100/120 VAC (USA
power cord), includes 80-tube carousel and tubing connectors
731-8120 Model 2110 Fraction Collector, 220/240 VAC (no
power cord), includes 80 tube carousel and tubing connectors
731-8130 Carousel, 80 tube capacity
731-8131 Replacement Drop Former Kit, includes 2 drop
and 2 drop caps
731-8135 Micro Tube Adaptor
731-8136 Instrument Dust Cover
731-8261 Econo System Cable 1, 8-pin Mini-DIN to DB-9 con-
nector, to connect the Model 2110 Fraction Collector to the Econo Pump or BioLogic LP
731-8265 Econo System Cable 5, DB-9 connector to bare
wires, to connect the Model 2110 Fraction Collector to the BioLogic System or other external instruments
731-8268 Econo System Cable 8, 8-pin standard DIN to DB-9
connector, for connecting the Model 2110 Fraction Collector to Model 1327 Recorder
731-8322 Model SV-3 Diverter/Bypass Valve
Model EP-1 Econo Pump
731-8140 Model EP-1 Econo Pump, 100/120 VAC (USA power
cord), includes tubing set and starter fittings kit
731-8142 Model EP-1 Econo Pump, 220/240 VAC (no power
cord), includes tubing set and starter fittings kit
Bio-Rad sells a complete line of integrated chromatography systems for protein purification. For information on the BioLogic System family of products, contact your local Bio-Rad representative.
24
Page 29
Life Science Group
Web site www.bio-rad.com USA (800) 4BIORAD Australia 02 9914 2800 Austria (01) 877 89 01 Belgium 09-385 55 11 Brazil 55 21 507 6191 Canada (905) 712-2771 China 86-10-8201-1366/68 Denmark 45 44 52-1000 Finland 358 (0)9 804 2200 France 01 47 95 69 65 Germany 089 318 84-177 Hong Kong 852-2789-3300 India (91-124) 6398112/113/114 Israel 03 951 4124 Italy 34 91 590 5200 Japan 03-5811-6270 Korea 82-2-3473-4460 Latin America 305-894-5950 Mexico 52 5 534 2552 to 54 The Netherlands 0318-540666 New Zealand 64-9-4152280 Norway 47-23-38-41-30 Russia 7 095 979 98 00 Singapore 65-2729877 Spain 34-91-590-5200 Sweden 46 (0)8-55 51 27 00 Switzerland 061-717-9555 United Kingdom 0800-181134
00-000 0000 Sig 1200Bulletin 0000 US/EG Rev A
Bio-Rad Laboratories
M7318120 Rev G
Loading...