Benning MA 2-16 Instruction Manual

BENNING MA 2-16
CEE 5-Pin Measuring Adapter
Instruction Manual
5057 en Version 1.00 10193764
03.07.2017
Measuring Adapter BENNING MA 2-16
1

1. Application

With the BENNING MA 2-16 measuring adapter you can test devices equipped with a 5-pin 16 A/ 6h CEE plug quickly and efficiently. To do this connect the adapter to a tester for testing portable devices in compliance with DIN VDE 0701-0702
or for testing in compliance with
DIN EN 62353 (VDE 0751-1)
or
DIN EN 60974-4 (VDE 0544-4)
, that has only a shock-proof socket for connecting to the test sample and that can measure protective conductor current directly.
The adapter is suitable for connection to the appliance testers BENNING ST 755 and BENNING ST 760. The following tests can be carried out with the measuring
adapter on appliances with CEE plugs:
Test of the continuity of the protective conductor system Test of insulation Measurement of the protective conductor current with
the method: direct measurement method/ differential and alternative leakage current
Functional test
Attention! Restriction of the application
The BENNING MA 2-16 measuring adapter may only be used for testing purposes in combination with tester types BENNING ST 755 and BENNING ST 760. Any other use of the adapter, in particular for permanent connection of three-phase appliances to a three­phase mains, is not permissible under any circumstances!
The maximum permissible thermal continuous current in all cases is 16 A AC per phase.

2. Safety features and measures

The measuring adapter is constructed and tested in compliance with the safety regulations IEC 61010-1/ EN 61010-1/ VDE 0411-1. The safety of users and the device is guaranteed if it is used for its intended purpose.
The device may only be connected to a TN, TT or IT power supply of max. 240 V/ 400 V, which complies with the applicable safety regulations (
e. g. IEC 60346, VDE 0100
) and is
secured with a maximum nominal current of 16 A. Before use please read this instruction manual carefully and
completely and in particular the instruction manual of the tester with which you intend to use this measuring adapter. Observe and follow all points included in the instructions.
The adapter may only be used for testing appliances with a 5-pin CEE plug and a maximum current consumption of 3 x 16 A (AC-1).

The measuring adapter must not be used:

with the housing open if there are any signs of external damage if the CEE socket or connection cable are damaged after severe overuse, i. e. exceedance of the load limits
in the technical specifications
after storage over a longer period of time under
unfavourable conditions (e. g. humidity, dust, temperature)

Meanings of the symbols on the device

Attention: Observe documentation! This symbol means that the instructions in the
instruction manual must be observed in order to prevent danger.
CAT II
Device in the measuring category II
Earth (voltage to earth)
This symbol on the measuring adapter means that the measuring adapter conforms to the EU directives.
The device must not be disposed of with household waste. You will find more information about the WEEE symbol on the Internet under the search term WEEE.

3. Connecting the BENNING MA 2-16

Before you can start testing, you have to connect the BENNING MA 2-16 to your tester and the mains. To do this proceed as follows:
Connect the mains cable of your tester to the shock-
proof socket on the BENNING MA 2-16 marked “mains connection tester“.
Connect the protective contact line of the
BENNING MA 2-16 to the test/ mains socket of your tester. The connection on the BENNING MA 2-16 is marked for test/ mains socket tester”.
Connect the CEE connection lead of the BENNING
MA 2-16 to a suitable CEE socket 16 A/ 6h in your
electrical installation. The connection on the BENNING MA 2-16 is marked
3~/ N/ PE, 400 V, 50-60 Hz, 16 A
”. From this time onwards the BENNING MA 2-16 and your tester will be supplied with electrical power.
Finally connect your test sample to the CEE socket of
the BENNING MA 2-16. The socket is marked “Test/ mains socket test sample 16 A”.
You can now start the tests. (s. Chapter.4)

4. Testing with the BENNING MA 2-16

All tests that you can carry out with your tester on devices with a shock-proof plug and that are supported by the BENNING MA 2-16 adapter can now be carried out in an identical manner on devices with a CEE plug. Proceed here as indicated in the operating manual of your tester.

The following special features of the BENNING MA 2-16 adapter must be taken into account during testing with the tester.

In the case of an insulation or alternative leakage
current test, the three phase connections L1, L2, L3 and N of the test sample are short-circuited in the adapter and then the tests carried out.
In the case of a protective conductor resistance test, the value of the measured protective conductor resistance is increased by the proportion that is caused by the protective conductor system of the BENNING MA 2-16 measuring adapter itself.