BENNING MA 2-16
CEE 5-Pin Measuring Adapter
Instruction Manual
Translation of German original version
5057 en
Version 1.00
10193764
03.07.2017

Measuring Adapter BENNING MA 2-16
1
1. Application
With the BENNING MA 2-16 measuring adapter you can test
devices equipped with a 5-pin 16 A/ 6h CEE plug quickly and
efficiently. To do this connect the adapter to a tester for
testing portable devices in compliance with
DIN VDE 0701-0702
or for testing in compliance with
DIN EN 62353 (VDE 0751-1)
or
DIN EN 60974-4 (VDE 0544-4)
,
that has only a shock-proof socket for connecting to the test
sample and that can measure protective conductor current
directly.
The adapter is suitable for connection to the appliance testers
BENNING ST 755 and BENNING ST 760.
The following tests can be carried out with the measuring
adapter on appliances with CEE plugs:
Test of the continuity of the protective conductor system
Test of insulation
Measurement of the protective conductor current with
the method:
direct measurement method/ differential and alternative
leakage current
Functional test
Attention!
Restriction of the application
The BENNING MA 2-16 measuring adapter may only
be used for testing purposes in combination with tester
types BENNING ST 755 and BENNING ST 760. Any
other use of the adapter, in particular for permanent
connection of three-phase appliances to a threephase mains, is not permissible under any
circumstances!
The maximum permissible thermal continuous
current in all cases is 16 A AC per phase.
2. Safety features and measures
The measuring adapter is constructed and tested in
compliance with the safety regulations
IEC 61010-1/ EN 61010-1/ VDE 0411-1. The safety of users
and the device is guaranteed if it is used for its intended
purpose.
The device may only be connected to a TN, TT or IT power
supply of max. 240 V/ 400 V, which complies with the
applicable safety regulations (
e. g. IEC 60346, VDE 0100
) and is
secured with a maximum nominal current of 16 A.
Before use please read this instruction manual carefully and
completely and in particular the instruction manual of the
tester with which you intend to use this measuring adapter.
Observe and follow all points included in the instructions.
The adapter may only be used for testing appliances with a
5-pin CEE plug and a maximum current consumption of
3 x 16 A (AC-1).
The measuring adapter must not be used:
with the housing open
if there are any signs of external damage
if the CEE socket or connection cable are damaged
after severe overuse, i. e. exceedance of the load limits
in the technical specifications
after storage over a longer period of time under
unfavourable conditions
(e. g. humidity, dust, temperature)
Meanings of the symbols on the device
Attention: Observe documentation!
This symbol means that the instructions in the
instruction manual must be observed in order to
prevent danger.
Device in the measuring category II
This symbol on the measuring adapter means
that the measuring adapter conforms to the EU
directives.
The device must not be disposed of with
household waste. You will find more information
about the WEEE symbol on the Internet under
the search term WEEE.
3. Connecting the BENNING MA 2-16
Before you can start testing, you have to connect the
BENNING MA 2-16 to your tester and the mains. To do this
proceed as follows:
Connect the mains cable of your tester to the shock-
proof socket on the BENNING MA 2-16 marked “mains
connection tester“.
Connect the protective contact line of the
BENNING MA 2-16 to the test/ mains socket of your
tester. The connection on the BENNING MA 2-16 is
marked “for test/ mains socket tester”.
Connect the CEE connection lead of the BENNING
MA 2-16 to a suitable CEE socket 16 A/ 6h in your
electrical installation. The connection on the
BENNING MA 2-16 is marked
“
3~/ N/ PE, 400 V, 50-60 Hz, 16 A
”. From this time onwards
the BENNING MA 2-16 and your tester will be supplied
with electrical power.
Finally connect your test sample to the CEE socket of
the BENNING MA 2-16. The socket is marked
“Test/ mains socket test sample 16 A”.
You can now start the tests. (s. Chapter.4)
4. Testing with the BENNING MA 2-16
All tests that you can carry out with your tester on devices
with a shock-proof plug and that are supported by the
BENNING MA 2-16 adapter can now be carried out in an
identical manner on devices with a CEE plug. Proceed here
as indicated in the operating manual of your tester.
The following special features of the
BENNING MA 2-16 adapter must be taken into
account during testing with the tester.
In the case of an insulation or alternative leakage
current test, the three phase connections L1, L2, L3 and
N of the test sample are short-circuited in the adapter
and then the tests carried out.
In the case of a protective conductor resistance test, the
value of the measured protective conductor resistance
is increased by the proportion that is caused by the
protective conductor system of the BENNING MA 2-16
measuring adapter itself.

Measuring Adapter BENNING MA 2-16
2
If in doubt, i. e. in cases where the results of a measurement
are close to the permissible limit value, measure the
protective conductor resistance of the adapter on the PE
connection of its CEE socket and subtract this value from
the measured value of the system.
To measure the protective conductor current with the
differential current method set the measurement type
switch to ΔI (Differential Current). In the case of a 3phase differential current measurement, the resulting
differential current is transmitted to the tester and
recorded as protective conductor current on the
protective conductor.
To measure the protective conductor current directly set
the measurement type switch to I
direkt
Before carrying out the functional test and all tests in
which the test sample has to be supplied with supply
voltage and put into operation, it is imperative that there
are no short circuits in the test sample within the phases
L1, L2, L3 and the neutral conductor N! Nonobservance could result in the BENNING MA 2-16
measuring adapter, or even your test sample, being
damaged or destroyed!
The values for the power and current consumption of
the test sample cannot be measured with the
BENNING MA 2-16 adapter within the scope of the
functional test. The values displayed relate to the power
consumption of the BENNING MA 2-16 adapter.
Attention!
Sequence of the test –
Problem starting currents
Starting the test
To put your test sample into operation it is imperative
that you first start the test of your tester (orange lamp
on the BENNING MA 2-16 illuminated) and then
switch on your test sample!
Ending the test
It is imperative that you first switch off the test
sample and then end the test on your tester.
In the case of extreme starting currents of your test
sample or in the case of inductances in the circuit,
non-observance of this sequence could result in the
BENNING MA 2-16 adapter being damaged or
destroyed!
Attention!
Short circuit test of the test sample
A short circuit of two or all three phases in the test
sample cannot be determined by the tester before
the differential current test! In such a case the
BENNING MA 2-16 adapter can be severely
damaged by the arsing current surge when carrying
out this test!
For this reason test that there is no short circuit
between the phases of the test sample, e. g. with
a continuity tester, before connecting a device
to the BENNING MA 2-16!
Attention!
Correct phase connection of the test sample
Take care, especially after a repair of three-phase
devices or reconnection of a CEE plug, that the
phases are connected in the correct sequence.
Attention!
Interruption of the protective devices – max.
current consumption
As soon as the orange signal lamp on the
BENNING MA 2-16 lights up, the connections of the
supplying mains socket are unprotected or a
different type of protective element is connected to
the test/mains socket of the measuring adapter.
Take care that only devices designed for a threephase current with a max. current consumption of
3 x 16 A (AC-1) are connected to this socket.
5. Technical Data
Nominal voltage: 3 x 400 V ± 10 %
Current carrying capacity 16 A three-phase current
Intrinsic connected load
“Mains active” 7 VA, cos φ ~ 0.4
Protection class: I in compliance with: IEC 61010-1/
EN 61010
-1 / VDE 0411 -1
Measuring category II
Contamination level 2
EMC EN 61326-1
Differential current
Measuring range 0.08 mA ... 10.0 mA AC
Intrinsic deviation 4 % of measured value ± 40 µA
Operating measurement
deviation
6 % of measured value± 60 µA
Direct measurement method
See tester specifications
Mechanical design
Protection type: Housing IP40, connections IP20
Dimensions L x W x H: 290 mm x 120 mm x 105 mm
(without cables and sleeves)
Weight 2.0 kg
Part number 044160

Text
und
Abbildungen
entsprechen
dem technischen
Stand bei
Drucklegung.
Technische
Änderungen
vorbehalten.
Keine
Haftung
für Druckfehler
3-447-001-01
1/7.17
Support / Helpdesk
BENNING Helpdesk-Team
Telefon: +49 2871 93-555
Telefax: +49 2871 93-417
E-Mail: info@benning.de
Internet: www.benning.de