Sample Container Sizes
Primary tubes:
13 x 75 mm (5 mL)
13 x 100 mm (7 mL)
16 x 75 mm (7 mL)
16 x 100 mm (10 mL)
Sample cups:
0.5 mL, 2.0 mL
SYNCHRON Microtubes
™
Reporting Units Available
ng/mL, µg/mL,
ng/dL, µg/dL, mg/dL, g/dL,
mg/L, µg/L, g/L, nmol/L, µmol/L, mmol/L
mEq/L, KU/L
nKat/L, µKat/L
IU/mL, U/mL, µIU/mL, mIU/L, mIU/mL
U/L, IU/L, mA, mA/min
%, Ku/u
Communication Modes
Unidirectional, bidirectional, bidirectional
with true Host Query
RS-232C Serial
Power and Environmental Requirements
Analyzer:
200 to 240 VAC
50/60 Hz
20A Nema L6-20R twistlock connector
Computer console and monitor:
100 to 240 VAC
50/60 Hzl
Printer:
120 or 240 VAC
50/60 Hz
Heat output:
4,972 BTU/hour
Ambient operating environment:
Ambient temperature: +18°C to +30°C
Relative humidity: 30 to 85%, non-condensing
Dimensions and Weight
Analyzer only:
Height: 69 inches (176 cm)
Length: 47 inches (120 cm)
Depth: 30 inches (77 cm)
Weight: 785 pounds (356 kg)
Detection Systems
Multi-wavelength, diffraction grating
spectrophotometer
Light source: Pulsed xenon lamp
Available wavelengths: 340, 380, 410, 470, 520,
560, 600, 650, 670, 700 nm
Measurement Principles
Immuno-turbidimetric
CEDIA*
Bidentate
Enzyme Immunoassay
Rate
Endpoint
Throughput
225 tests/hour maximum
Reaction Carousel Temperature
37°C (± 0.1°C)
Reagent Cartridge Compartment
Temperature
+2°C to +8°C
Reagent Capacity
24 methods simultaneously on board
Bar Code Reagent Features
Automatic tracking of:
Available tests
Expiration date
Lot number
Calibration expiration
Sample Management/Capacity
63 samples in STANDBY mode, continuous loading
(56 in STAT mode)
Bar-code symbologies:
Code 39 (Code 3 of 9, or SD-3)
Code 128 (USD-6)
Interleaved 2 of 5 (USD-1)
Codabar (USD-4)
Sample volume as low as 3 µL
Sample Types:
Serum
Plasma
Urine
CSF
Whole Blood Hemolysate
Chemistry Menu Capacity
> 90 preprogrammed, bar-coded methods
currently available
On-board data storage for 100
user-defined methods
Calibration Types
Single-point, 2-point, multi-point
Linear and non-linear models (chemistry method
dependent)
System Console Computer
3.5 inch, 1.44 MB floppy diskette
≥ 2.0 GB Hard disk drive
128 MB RAM
CD-ROM
Water Requirements
Deionized water, NCCLS Type II
(Total Bacteria Count <10 cfu)
Drain Requirements
6.9 L/hour minimum continuous flow rate
< 39 inches (97.5 cm) above the floor
1.5 inches (3.81 cm) internal diameter
* CEDIA is a registered trademark of Roche Diagnostics.
SYNCHRON CX®4 PRO Specifications
S IMPLIFY • A UTOMATE • I NNOVATE
Eastern Europe, Middle East, North Africa: Switzerland, Nyon (41) 22 994 0707. Australia, Gladesville (61) 2 9844 6000. Canada, Mississauga (1) 905 819 1234.
China, Beijing (86) 10 6515 6028. Hong Kong (852) 2814 7431, 2814 0481. France, Villepinte (33) 1 49 90 90 00. Germany, Krefeld (49) 2151 33 35. Italy, Milan (39) 02 953921.
Japan, Tokyo (81) 3 5404 8424. Latin America (1) (305) 380 4709. Mexico, Mexico City (525) 575 6805. Netherlands, Mijdrecht(31) 297 230630. Singapore (65) 6339 3633.
South Africa/Sub-Saharan Africa, Johannesburg(27) 11 805 2014. Sweden, Bromma (46) 8 564 85 900. Switzerland, Nyon 0800 850 810. Taiwan, Taipei (886) 2 2378 3456.
Turkey, Istanbul (90) 216 309 1900. UK, High Wycombe (44) 01494 441181. USA, Brea, CA (1) 800 352 3433, (1) 714 993 5321.
www.beckmancoulter.com © Copyright 2003 Beckman Coulter, Inc. PRINTED IN U.S.A.