All rights reserved. No part of this document may be
reproduced or transmitted in any form or by any means,
electronic, mechanical, photocopying, recording, or
otherwise, without prior written permission from
Beckman Coulter, Inc.
Beckman Coulter, the stylized logo, and Allegra are
trademarks of Beckman Coulter, Inc. and are registered in
the USPTO.
All other trademarks, service marks, products, or services
are trademarks or registered trademarks of their
respective holders.
Find us on the World Wide Web at:
www.beckmancoulter.com
Made in U.S.A.
Printed in U.S.A.
Page 3
Read all product manuals and consult with Beckman Coulter-trained personnel before attempting
to operate instrument. Do not attempt to perform any procedure before carefully reading all
instructions. Always follow product labeling and manufacturer’s recommendations. If in doubt as
to how to proceed in any situation, contact your Beckman Coulter Representative.
Alerts for Warning, Caution, and Note
WARNING
WARNING indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
CAUTION indicates a potentially hazardous situation, which, if not avoided,
may result in minor or moderate injury and/or mechanical damage.
Safety Notice
NOTE
NOTE is used to call attention to notable information that should be followed during installation, use,
or servicing of this equipment.
Safety During Installation and/or Maintenance
This centrifuge weighs 128 kg (283 lb). Do not attempt to lift or move it without assistance.
Any servicing of this equipment that requires removal of any covers can expose parts that involve
the risk of electric shock or personal injury. Make sure that the power switch is off and the
centrifuge is disconnected from the main power source by removing the Mains (power) plug from
the outlet receptacle, and refer such servicing to qualified personnel.
Do not replace any centrifuge components with parts not specified for use on this instrument.
Electrical Safety
To reduce the risk of electrical shock, this equipment uses a three-wire electrical cord and plug to
connect the centrifuge to earth-ground. To preserve this safety feature:
•Make sure that the matching wall outlet receptacle is properly wired and earth-grounded.
Check that the line voltage agrees with the voltage listed on the name-rating plate affixed to
the centrifuge.
•Never use a three-to-two wire plug adapter.
PN GXR-IM-4AC
iii
Page 4
Safety Notice
Safety Against Risk of Fire
•Never use a two-wire extension cord or a two-wire non-grounding type of multiple-outlet
receptacle strip.
Do not place containers holding liquid on or near the chamber door. If they spill, liquid may get into
the centrifuge and damage electrical or mechanical components.
Safety Against Risk of Fire
This centrifuge is not designed for use with materials capable of developing flammable or explosive
vapors or hazardous chemical reactions. Do not centrifuge such materials (such as chloroform or
ethyl alcohol) in this centrifuge nor handle or store them within the 30-cm (1-ft) area surrounding
the centrifuge.
Mechanical Safety
For safe operation of the equipment, observe the following:
•Use only the rotors and accessories designed for use in this centrifuge.
•Do not exceed the maximum rated speed of the rotor in use.
•NEVER attempt to slow or stop the rotor by hand.
•Do not lift or move the centrifuge while the rotor is spinning.
•NEVER attempt to override the door interlock system while the rotor is spinning.
•Maintain a 7.6-cm (3-in.) clearance envelope around the centrifuge while it is running. During
operation you should come within the envelope only to adjust instrument controls, if
necessary. Never bring any flammable substances within the 30-cm (1-ft) area surrounding the
centrifuge. Never lean on the centrifuge or place items on the centrifuge while it is operating.
Chemical and Biological Safety
Normal operation may involve the use of solutions and test samples that are pathogenic, toxic, or
radioactive. Such materials should not be used in this centrifuge, however, unless all necessary
safety precautions are taken.
•Observe all cautionary information printed on the original solution containers prior to
their use.
•Handle body fluids with care because they can transmit disease. No known test offers complete
assurance that they are free of micro-organisms. Some of the most virulent—Hepatitis (B and
C) and HIV (I–V) viruses, atypical mycobacteria, and certain systemic fungi—further emphasize
the need for aerosol protection. Handle other infectious samples according to good laboratory
procedures and methods to prevent spread of disease. Because spills may generate aerosols,
observe proper safety precautions for aerosol containment.
Do not run toxic, pathogenic, or radioactive materials in this centrifuge without taking
iv
PN GXR-IM-4AC
Page 5
appropriate safety precautions. Biosafe containment should be used when Risk Group II
materials (as identified in the World Health Organization Laboratory Biosafety Manual) are
handled; materials of a higher group require more than one level of protection.
•Dispose of all waste solutions according to appropriate environmental health and safety
guidelines.
It is your responsibility to decontaminate the centrifuge and accessories before requesting service by
Beckman Coulter.
RoHS Notice
These labels and materials declaration table (the Table of Hazardous Substance's Name and
Concentration) are to meet People’s Republic of China Electronic Industry Standard SJ/T11364-2006
“Marking for Control of Pollution Caused by Electronic Information Products” requirements.
China RoHS Caution Label
Safety Notice
RoHS Notice
This label indicates that the electronic information product contains certain toxic or hazardous
substances. The center number is the Environmentally Friendly Useful Period (EFUP) date, and
indicates the number of calendar years the product can be in operation. Upon the expiration of the
EFUP, the product must be immediately recycled. The circling arrows indicate the product is
recyclable. The date code on the label or product indicates the date of manufacture.
China RoHS Environmental Label
This label indicates that the electronic information product does not contain any toxic or hazardous
substances. The center ‘e’ indicates the product is environmentally safe and does not have an
Environmentally Friendly Useful Period (EFUP) date. Therefore, it can safely be used indefinitely.
The circling arrows indicate the product is recyclable. The date code on the label or product
indicates the date of manufacture.
PN GXR-IM-4AC
v
Page 6
Safety Notice
RoHS Notice
vi
PN GXR-IM-4AC
Page 7
Safety Notice,iii
Introduction,xiii
CHAPTER 1:Description,1-1
Introduction,1-1
Centrifuge Function and Safety Features,1-1
Centrifuge Function,1-1
Safety Features,1-2
Contents
Centrifuge Chassis,1-2
Housing,1-2
Door,1-3
Rotor Chamber,1-3
Temperature Sensing and Control,1-3
Drive,1-3
Name Rating Plate,1-3
Controls and Indicators,1-3
Power Switch,1-3
Control Panel,1-4
Digital Displays,1-4
Run-Parameter Keys,1-5
Specifications,1-7
Available Rotors,1-9
CHAPTER 2:Operation,2-1
Introduction,2-1
Summary of Run Procedures,2-2
Manual Run,2-2
Programmed Run,2-3
Preparation,2-3
Installing the Rotor,2-3
Manual Operation,2-4
Selecting a Rotor,2-5
vii
Page 8
Contents
Entering Run Speed,2-5
Entering Run Time,2-5
Entering Run Temperature,2-6
Entering Acceleration and Deceleration Rates,2-6
Starting a Run,2-7
Pulse Function,2-8
Programmed Operation,2-9
Creating a New Program,2-9
Recalling a Program,2-9
Disabling/Enabling the Audible Tones,2-10
CHAPTER 3:Troubleshooting,3-1
Introduction,3-1
User Messages,3-1
Other Possible Problems,3-4
Retrieving Your Sample in Case of Power Failure,3-6
CHAPTER 4:Care and Maintenance,4-1
Introduction,4-1
Instrument Care,4-1
Maintenance,4-1
Cleaning,4-2
Tube Breakage,4-2
Decontamination,4-3
Sterilization and Disinfection,4-3
2.1Acceleration and Deceleration Rates (in Minutes:Seconds) to
and from Maximum Speed,2-7
3.1Diagnostic Message Chart,3-2
3.2Troubleshooting Chart,3-5
Tables
xi
Page 12
Tables
xii
Page 13
Certification
Beckman Coulter Allegra X-15R centrifuges are manufactured in a facility that maintains
certifications to both ISO 9001:2008 and ISO 13485:2003. They have been designed and tested to be
compliant (when used with Beckman Coulter rotors) with the laboratory equipment requirements
of applicable regulatory agencies. Declarations of conformity and certificates of compliance are
available at www.beckmancoulter.com.
Introduction
Scope of Manual
This manual is designed to familiarize you with the Beckman Coulter Allegra X-15R centrifuge, its
functions, specifications, operation, and routine operator care and maintenance. Beckman Coulter
recommends that you read this entire manual, especially the Safety Notice and all safety-related
information, before operating the centrifuge or performing instrument maintenance.
•CHAPTER 1, Description contains system specifications and a brief physical and functional
description of the centrifuge, including the operating controls and indicators.
•CHAPTER 3, Troubleshooting lists diagnostic messages and other possible malfunctions,
together with probable causes and suggested corrective actions.
•CHAPTER 4, Care and Maintenance contains procedures for routine operator care and
maintenance, as well as a brief list of supplies and replacement parts
•APPENDIX A, Installation provides requirements for preparing laboratory facilities and
installing the centrifuge.
NOTE
If the centrifuge is used in a manner other than specified in this manual, the safety and performance
of this equipment could be impaired. Further, the use of any equipment other than that recommended
by Beckman Coulter has not been evaluated for safety. Use of any equipment not specifically
recommended in this manual and/or the applicable rotor manual is the sole responsibility of the user.
PN GXR-IM-4AC
xiii
Page 14
Introduction
Conventions
Conventions
Certain symbols are used in this manual to call out safety-related and other important information.
These international symbols may also be displayed on the centrifuge and are reproduced on the
inside of the back cover.
Typographic Conventions
Certain typographic conventions are used throughout this manual to distinguish names of user
interface components, such as keys and displays.
•Key names (for example,
CFC-Free Centrifugation
To ensure minimal environmental impact, no CFCs are used in the manufacture or operation of
Allegra X-15R centrifuges.
Radio Interference
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with this instruction manual, may cause interference to radio communications.
Operation of this equipment in a residential area is likely to cause interference, in which case the
user will be required to correct the interference at his own expense.
Canadian Regulations
START or ROTOR) appear in capital letters in boldface type.
xiv
This equipment does not exceed the Class A limits for radio noise emissions from digital apparatus
as set out in the radio interference regulations of the Canadian Department of Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites
applicables aux appareils numériques de Classe A prescrites dans le reglement sur le brouillage
radioelectrique édicté par le Ministère des Communications du Canada.
PN GXR-IM-4AC
Page 15
Recycling Label
A28219-AA
This symbol is required in accordance with the Waste Electrical and Electronic Equipment (WEEE)
Directive of the European Union. The presence of this marking on the product indicates:
1. the device was put on the European market after August 13, 2005 and
2. the device is not to be disposed via the municipal waste collection system of any member state
of the European Union.
It is very important that customers understand and follow all laws regarding the proper
decontamination and safe disposal of electrical equipment. For Beckman Coulter products bearing
this label please contact your dealer or local Beckman Coulter office for details on the take back
program that will facilitate the proper collection, treatment, recovery, recycling and safe disposal
of the device.
Introduction
Recycling Label
PN GXR-IM-4AC
xv
Page 16
Introduction
Recycling Label
xvi
PN GXR-IM-4AC
Page 17
Introduction
This chapter provides a brief physical and functional description of the Beckman Coulter Allegra X-15R
centrifuge. The operating controls and indicators are also described; instructions for their use are in
CHAPTER 2, Operation. Chemical compatibilities of materials listed in this manual can be found in
Chemical Resistances (publication IN-175). Refer to the applicable rotor manuals for rotor descriptions.
Centrifuge Function and Safety Features
CHAPTER 1
Description
Centrifuge Function
The Beckman Coulter Allegra X-15R centrifuge (see Figure 1.1) is a refrigerated benchtop centrifuge
that generates centrifugal forces required for a wide variety of applications. Together with the
Beckman Coulter rotors designed for use in this centrifuge, the centrifuge applications include:
•Routine processing such as sample preparations, pelleting, extractions, purifications,
concentrations, phase separations, receptor binding, and column centrifugations.
•Cell isolation.
•Binding studies and separation of whole blood.
•Processing large numbers of small-volume samples in multiwell plates for concentrating tissue-
culture cells, cloning and replicate studies, in-vitro cytotoxicity studies, receptor binding, and
genetic engineering experimentation.
•Rapid sedimentation of protein precipitates, large particles, and cell debris.
The Allegra X-15R is microprocessor-controlled, providing interactive operation. The instrument
design features a brushless asynchronous motor, automatic rotor identification system, program
memory that enables repeated run conditions, temperature control system, and a choice of
acceleration and deceleration rates. User messages alert the operator to conditions that may need
attention.
PN GXR-IM-4AC
1-1
Page 18
Description
Centrifuge Chassis
Figure 1.1 The Centrifuge
Safety Features
The Allegra X-15R centrifuge has been designed and tested to operate safely indoors at altitudes up
to 2000 m (6562 ft). Safety features include the following.
•An electromechanical door lock system prevents operator contact with spinning rotors and
prevents run initiation unless the door is shut and locked. The door is locked when a run is in
progress and can be opened only when the rotor is stopped and the LED (light-emitting diode)
next to the
sample recovery.
•A steel barrier surrounds the rotor chamber to provide full operator protection.
•A rotor identification system prevents the installed rotor from running above its maximum
rated speed. During acceleration the microprocessor checks the magnetic rotor identification.
Speed is limited to the maximum safe speed of the identified rotor. If the system identifies a
rotor different than the one entered by the user, the system displays an error message and
reduces the speed to the maximum allowable speed of the rotor.
•An imbalance detector monitors the rotor during the run, causing automatic shutdown if rotor
loads are severely out of balance. At low speeds, an incorrectly loaded rotor can cause
imbalance. Rotor instability can also occur if the centrifuge is moved while running, or if it is
not resting on a level and secure surface (see CHAPTER 3, Troubleshooting).
Centrifuge Chassis
Housing
DOOR key is lit. If there is a power failure, the door lock can be manually tripped for
1-2
The centrifuge housing is made of sheet steel and structural foam and is finished with acrylic baking
enamel. The control panel overlay is made of coated polystyrene copolymer.
PN GXR-IM-4AC
Page 19
Door
The aluminum and structural foam door is secured to the housing by solid hinges. A window in the
center allows strobe viewing. When the door is closed, the locking system engages. The door is
locked when a run is in progress and can be opened only when the rotor is stopped and the LED next
to the
DOOR key is lit. If there is a power failure, the door lock can be manually tripped for sample
recovery (see CHAPTER 3, Troubleshooting).
Rotor Chamber
The aluminum rotor chamber is coated with epoxy enamel for corrosion resistance. The chamber
is sealed by a foam gasket.
Temperature Sensing and Control
With the power on, the temperature control system is activated when the door is closed and locked.
A sensor in the rotor chamber continuously monitors chamber temperature. The microprocessor
calculates the required chamber temperature to maintain the selected rotor temperature. Run
temperature can be set between –10 and +40°C.
Description
Name Rating Plate
1
NOTE
To avoid chamber icing, refrigeration is off when the door is open. The centrifuge door must be closed
and locked for the refrigeration system to begin operating.
Drive
The asynchronous direct-drive motor is brushless for clean, quiet operation. The resilient
suspension ensures that loads will not be disturbed by vibration, and prevents damage to the drive
shaft if an imbalance occurs during centrifugation. Maximum acceleration and deceleration may be
selected to allow fast processing of samples; alternately, delicate gradients may be preserved using
slower acceleration and deceleration.
Name Rating Plate
A name rating plate is affixed to the rear of the instrument. Always mention the serial number and
model number when contacting Beckman Coulter regarding your centrifuge.
Controls and Indicators
Power Switch
The power switch, located on the left side of the centrifuge, controls electrical power to the
centrifuge. It is also a circuit breaker that will trip to cut off power in the event of a power overload.
The power switch must be turned on before the chamber door can be opened.
PN GXR-IM-4AC
1-3
Page 20
Description
PULSE
ENTER/
SAVE
Controls and Indicators
Control Panel
The control panel (see Figure 1.2) is mounted at an angle on the centrifuge front for easy visibility
and access. It is used to enter run parameters via function keys and a keypad, and to display run
parameters, program information, and user messages.
Figure 1.2 Control Panel
Digital Displays
The displays provide run information, user messages, and diagnostics.
•During a run, they show the actual (real-time) operating conditions.
•When the run parameters are being entered, they show the set values selected. Set values can
be recalled during operation by pressing
SHOW SET. After 5 seconds, the displays return to
actual conditions.
•Error messages (see CHAPTER 3, Troubleshooting) also appear on the displays, when applicable.
The centrifuge emits a series of audible tones to alert the user to an error condition.
(Instructions for disabling the tones are in CHAPTER 2, Operation.)
Indicates rotor speed in revolutions per minute (RPM) or in relative centrifugal field, a
term describing the ratio of the centrifugal acceleration at a specified radius and
SPEED
TIME
speed to the standard acceleration of gravity (RCF ´g).
• In a timed run, indicates the remaining run time in hours and minutes. The time
display begins counting down when ENTER/SAVE then START is pressed and
continues counting down to 0. The run automatically ends when the set time
reaches 0 and deceleration begins (a tone sounds when the rotor stops spinning).
If a malfunction shuts down the centrifuge during a run, when the instrument
is turned back on the time will not remain and 00:00 will be displayed in the time
window. By comparing this time with the original set time, you can determine
when the run ended.
• In a hold run, indicates the time elapsed since START was pressed. After 99 hours
and 59 minutes the timer resets to 0 and continues counting elapsed time.
• In a pulse (momentary) run, indicates the time (in minutes and seconds) elapsed
since PULSE was pressed. Time continues to count up until the PULSE key is
released. The total elapsed time is then displayed until another key is pressed.
1-4
TEMPERATURE
Indicates rotor temperature in degrees C at thermoequilibrium.
PN GXR-IM-4AC
Page 21
ROTOR
ACCEL
DECEL
PROGRAM
Description
Controls and Indicators
Displays the name of the selected rotor.
When the ACCEL key is pressed, one of ten available acceleration rates can be
entered. Displays the selected acceleration rate throughout the run.
When the DECEL key is pressed, one of eleven available deceleration rates can be
entered. Displays the selected deceleration rate throughout the run.
• During setup— when the PROGRAM key is pressed, a program number (one of ten
possible) can be selected and program parameters entered.
• During centrifugation— the number of the program being run is displayed.
1
Run-Parameter Keys
Run-parameter keys are used to enter specific run information as follows. The centrifuge emits an
audible tone when each key is pressed. (Instructions for disabling the tone are in CHAPTER 2,
Operation).
Pressed to display scrollable list of compatible rotors. Can be pressed repeatedly to
scroll through the rotor list (or the arrow keys can be used for scrolling).
If the display shows speed in RPM, the key is pressed once to enter run speed (using
the keypad) in RPM (revolutions per minute); pressed twice to enter run speed in RCF
(relative centrifugal field). If the display shows speed in RCF, the key is pressed once
to enter speed in RCF or twice to enter speed in RPM.
Pressed to enter run time using the keypad. When TIME is pressed twice, HOLD mode
is selected. (Pressing it again reverts to time-entry mode.)
• Timed run—Run time up to 99 hours and 59 minutes can be set. If more than
59 minutes are entered in the minutes field, the system automatically converts
the entry to hours and minutes. Deceleration begins when the set time counts
down to 0.
• Continuous run (hold)— For runs of unspecified lengths, hold mode is used. When
99 hours, 59 minutes is reached, the system resets to 0 and continues counting.
A hold run continues until the STOP key is pressed.
PN GXR-IM-4AC
Pressed to enter temperature (using the keypad), from –10 to +40°C. If a new
temperature is not entered, the centrifuge uses the temperature set for the
previous run.
Enables selection of one of ten preset acceleration rates (see Table 2.1 in CHAPTER 2,
Operation) available to protect specific gradient and sample-to-gradient interfaces. If
no acceleration rate is selected, the centrifuge defaults to maximum acceleration.
1-5
Page 22
Description
Controls and Indicators
Enables selection of one of eleven preset deceleration rates (see Table 2.1 in
CHAPTER 2, Operation) available to maintain optimum separation. If no deceleration
rate is selected, the centrifuge defaults to maximum deceleration. (Deceleration rate
0 is always coast to complete stop.)
A toggle switch used to enter into and exit out of the program mode to create or recall
programs. When the key is pressed, the program number can be selected using the
keypad or arrow keys.
Up and down arrow keys, which can be pressed to increment parameter values up
or down.
The ENTER/SAVE key can be pressed to save parameter inputs to system memory
during manual run setup or to a program during a program set up, and to start the
centrifuge (ENTER/SAVE, then START must be pressed to start the centrifuge).
The keypad is used to enter numerical run parameters, to enter or recall a program
number, and to select acceleration and/or deceleration rates. In addition to keys 0
through 9, the keypad includes a plus/minus key and a CE (clear entry) key.
The ± key is used to enter a temperature setting below 0°C (for example, –2°C).
The CE (clear entry) key clears any parameter on the display. Pressing CE will also
clear some diagnostic messages. If a diagnostic message does not clear when CE is
pressed, see CHAPTER 3, Troubleshooting.
Pressed during a run to display set values for 5 seconds.
Causes the rotor to accelerate, at maximum rate, to set speed for short-duration runs
(as long as the key is pressed). Deceleration, at maximum rate, begins when the key
is released.
Pressing ENTER/SAVE, then START begins the run. (START must be pressed within
5 seconds of pressing ENTER/SAVE.) This key combination can also be used to abort
a deceleration process and restart the centrifuge.
Can be pressed at any time while the rotor is spinning to terminate a run. The
centrifuge decelerates to a complete stop according to the deceleration setting. The
green light next to the key flashes while the rotor decelerates. A tone sounds when
the rotor comes to a complete stop. (Instructions for disabling the tone are in
CHAPTER 2, Operation.)
Unlocks and opens the door. (The instrument will not accept the unlock command if
the rotor is spinning.)
1-6
PN GXR-IM-4AC
Page 23
Specifications
Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data,
without guarantee.
SpecificationsDescription
Speed • Set Speed
Time • Set time
Temperature • Set temperature
Description
Specifications
— 200 to 10,200 RPM in 10-RPM increments
• Speed display
— actual rotor speed in 10-RPM increments or in RCF (when
selected)
— to 99 hours 59 minutes or continuous (hold)
• Time display
— Timed run: indicates run time remaining
— Continuous (hold) or pulse run: indicates elapsed time
— –10 to +40°C in 1° increments
• Temperature control (after equilibration)
— ±2°C of set temperature
• Temperature display (after equilibration
— chamber temperature in 1° increments
• Ambient temperature range
— 10 to 35°
• Ambient temperature range for optimum operation
— 10 to 25°C
6800 Btu/hr (2 kW)
room under steady-state
conditions
Noise level 0.91 m (3 ft) in front of
£ 68 dBa
centrifuge
Installation (overvoltage) category II
Pollution degree
a. During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. To reach
temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber during
operation. At low run speeds or low ambient temperatures, the centrifuge may not be able to achieve some higher
temperatures.
b. Normally only nonconductive pollution occurs; occasionally, however, a temporary conductivity caused by condensation
must be expected.
b
2
1-8
PN GXR-IM-4AC
Page 25
Available Rotors
The following Beckman Coulter rotors can be used in the Allegra X-15R centrifuge. The rotors are
described in individual manuals that accompany each rotor.
Rotor ProfileDescription
Rotor
Selection
Code
Max
RPM
a
Max RCF
(´g)
at r
max
b
Number
of Tubes ´
Nominal
Capacity
Description
Available Rotors
Rotor
Manual
Number
1
FX6100 Fixed Angle
r
= 98.0 mm
max
SX4750 Swinging Bucket
Tube-and-bottle buckets,
r
= 207.8 mm
max
Multiwell-plate carriers,
= 183.2 mm
r
max
SX4750A Swinging
Bucket (ARIES)
Tube-and-bottle buckets,
r
= 207.8 mm
max
Multiwell-plate carriers,
= 183.2 mm
r
max
FX610010,20011,4006 ´ 100 mLGX-TB-005
GX-TB-003
SX4750
SX4750m
4750
4450
5250
4060
4 ´ 750 mL
4 ´ 96 mL
GX-TB-004
SX4750A
SX4750A
4750
4450
5250
4060
4 ´ 750 mL
4 ´ 96 mL
a. Maximum speeds are based on a solution density of 1.2 g/mL. At upper temperature and humidity ambient conditions, swinging bucket rotor
speed may require reduction.
b. Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius and speed (rw2) to the standard acceleration of
gravity (g) according to the following formula: RCF = rw2/g — where r is the radius in millimeters, w is the angular velocity in radians per second
(2 p RPM/60), and g is the standard acceleration of gravity (9807 mm/s2). After substitution: RCF = 1.12 r (RPM/1000)2.
PN GXR-IM-4AC
1-9
Page 26
Description
Available Rotors
1-10
PN GXR-IM-4AC
Page 27
Introduction
This section contains centrifuge operating procedures. A summary is provided at the start of this section.
If you are an experienced user of this centrifuge, you can turn to the summary for a quick review of
operating steps. Refer to the applicable rotor manual for instructions on preparing the rotor for
centrifugation.
CHAPTER 2
Operation
WARNING
Handle body fluids with care because they can transmit disease. No known test
offers complete assurance that they are free of micro-organisms. Some of the
most virulent—Hepatitis (B and C) and HIV (I–V) viruses, atypical mycobacteria,
and certain systemic fungi— further emphasize the need for aerosol protection.
Handle other infectious samples according to good laboratory procedures and
methods to prevent spread of disease. Because spills may generate aerosols,
observe proper safety precautions for aerosol containment.
Do not run toxic, pathogenic, or radioactive materials in this centrifuge without
taking appropriate safety precautions. Biosafe containment should be used when
Risk Group II materials (as identified in the World Health Organization Laboratory Biosafety Manual) are handled; materials of a higher group require more than one
level of protection.
Do not use the centrifuge in the vicinity of flammable liquids or vapors, and do not
run such materials in the instrument. Do not lean on the centrifuge or place items
on it while it is operating.
PN GXR-IM-4AC
2-1
Page 28
Operation
Summary of Run Procedures
Summary of Run Procedures
For runs at other than room temperature, refrigerate or warm the rotor beforehand for fast equilibration.
Manual Run
1
Turn the power switch on (I).
2
Press DOOR to open the chamber door; lift the door open.
3
Install the rotor according to the applicable rotor manual, then close the chamber door.
4
Press ROTOR, then press ROTOR again or use the arrow keys to select the rotor.
5
Press RPM/RCF, then use the keypad to enter the run speed. Press RPM/RCF a second time to
toggle between RPM and RCF modes.
6
Press TIME, then use the keypad to enter the run time (to 99 hrs, 59 minutes); or press TIME
twice to toggle between hold (continuous) run and set time mode.
7
Press TEMP °C, then use the keypad to enter the required run temperature –10 to +40°C.
8
Press ACCEL, then use the keypad or arrow keys to enter the selected acceleration rate number,
1 (SLOW) through 10 (MAX).
9
Press DECEL, then use the keypad or arrow keys to enter the selected deceleration rate number,
0 (OFF) through 10 (MAX).
10
Check that all parameters are correct and that the door is closed (door LED will be lit). Press
ENTER/SAVE, then press START (within 5 seconds).
2-2
11
Wait for the set time to count down to zero, or end the run by pressing STOP.
12
When the rotor stops (a tone sounds and the DOOR LED lights), press DOOR to open the chamber
door; lift the door.
PN GXR-IM-4AC
Page 29
Programmed Run
1
Turn the power switch on (I).
2
Press DOOR to open the chamber door.
3
Install the rotor according to the applicable rotor manual, then close the chamber door.
4
Press PROGRAM, use the keypad or arrow keys to enter the required program number, then
press
5
Check that all parameters are correct and that the door is closed (door LED will be lit). Press
ENTER/SAVE, then press START (within 5 seconds).
ENTER/SAVE.
Operation
Preparation
2
6
Wait for the run to end, or end the run by pressing STOP.
7
When the rotor stops (a tone sounds and the DOOR LED lights), press DOOR to open the chamber
door; lift the door.
Preparation
Prepare the rotor for centrifugation as described in the applicable rotor manual. For runs at other
than room temperature, refrigerate or warm the rotor beforehand for fast equilibration.
Installing the Rotor
The power must be turned on before you can unlock and open the chamber door. (To end a run for
any reason, do not turn the power switch off; press
1
Turn the power switch on.
STOP instead.)
PN GXR-IM-4AC
The control panel display shows “
2
Press DOOR.
The centrifuge will accept this command only when the rotor is at rest and the
The door opens and the LED goes out.
Allegra X-15R” and the software version.
DOOR LED is lit.
2-3
Page 30
Operation
Manual Operation
3
Install the rotor according to directions in the rotor manual.
Ensure that the rotor is seated on the drive hub.
Avoid bumping the control panel keys during rotor installation or removal.
4
Close the chamber door by pressing down firmly on both sides.
The door locks.
When the latches engage, the LED next to the DOOR key lights up.
Manual Operation
NOTE When using multiwell-plate carriers on the SX4750 rotor, enter SX4750m as the rotor, when using
multiwell-plate carriers on the SX4750A rotor, enter SX4750A as the rotor and limit max speed of the
instrument to 4450 RPM (when running up to three stacked Beckman Coulter plates separated by cap
strips; reduce speed to 2700 RPM if running four stacked plates).
Set rotor speed using RPM values only. To set speed to a specific RCF, use the RPM-to-RCF conversion
chart (GX-TB-015) to determine the applicable RPM to enter.
1
When a run-parameter key (ROTOR, RPM/RCF, TIME, TEMP, ACCEL, or DECEL) is pressed, a
parameter can be entered or changed.
Entry is completed when
2
To change an entry before you’ve pressed ENTER/SAVE or another parameter key, enter a
different value.
a. Press
b. To change an entry after you’ve pressed
3
If an unacceptable value is entered, an out of range message is displayed in the message line.
Enter the correct value.
CE to abort changes and start over.
ENTER/SAVE or another function key is pressed.
ENTER/SAVE, press the run-parameter key again.
2-4
PN GXR-IM-4AC
Page 31
Selecting a Rotor
1
Press ROTOR, then use the arrow keys or press ROTOR again to select a rotor.
2
Press the next run-parameter key or press ENTER/SAVE.
Entering Run Speed
Enter a run speed up to the maximum speed of the rotor in use. Or, enter a relative centrifugal field
(RCF) value up to the maximum achievable RCF of the rotor.
1
Press RPM/RCF once to enter speed in RPM. Press RPM/RCF again to toggle between RPM and
RCF modes.
Operation
Manual Operation
2
Allowed speed range for the selected rotor is displayed.
2
Use the keypad to enter required speed.
Entered speed appears on the speed display.
(If the entered speed is out of the speed range of the installed rotor, an error message will
appear and the input will be cleared.)
Enter an acceptable value from the range displayed.
3
Press the next run-parameter key or ENTER/SAVE.
The rotor will accelerate or decelerate to the new speed.
You can change the set speed at any time during a run by repeating steps 1 through 3.
Entering Run Time
Run time can be set for up to 99 hours and 59 minutes or for a hold (continuous) run. The time
display begins counting down when the rotor starts to spin. For a timed run, the run automatically
terminates when the set time reaches zero and deceleration begins at the selected rate; a tone
sounds when the rotor has stopped. For a hold run, the elapsed time is displayed and the run
continues until
STOP is pressed.
PN GXR-IM-4AC
1
Press TIME.
HOLD is displayed, press TIME again to enter time.
If
2-5
Page 32
Operation
Manual Operation
2
Use the keypad or arrow keys to enter required time.
Entered time appears on the display.
If you entered a number higher than 59 minutes, the centrifuge automatically recalculates the
time in hours and minutes.
or
Press
TIME again for a hold run.
HOLD appears on the display.
3
Press the next run-parameter key or ENTER/SAVE.
Entering Run Temperature
Run temperature can be set from –10 to +40°C. If no value is entered, the centrifuge selects the last
entered temperature.
1
Press TEMP.
The cursor and the temperature range (–10 to 40°C) appear.
2
Use the keypad or arrow keys to enter required temperature.
Entered temperature appears on the display.
(If the entered temperature is outside the valid temperature range, an error message
will appear and the input will be cleared.)
Enter a valid temperature.
3
Press the next run-parameter key or ENTER/SAVE.
Entering Acceleration and Deceleration Rates
You can select one of ten acceleration rates and eleven deceleration rates to protect the gradient
and sample-to-gradient interface. Select rates depending on the type of run you are performing—
for pelleting runs, where sample mixing is not a concern, maximum acceleration and deceleration
can be used. If running delicate gradients, a lower setting may be needed. Typical acceleration and
deceleration rates are listed in Table 2.1. If no rate is selected, the centrifuge automatically
accelerates and decelerates at maximum.
2-6
PN GXR-IM-4AC
Page 33
Operation
Manual Operation
2
Table 2.1 Acceleration and Deceleration Rates (in Minutes:Seconds) to and from Maximum Speed
SX4750SX4750mSX4750AFX6100
(4750 RPM)(4450 RPM)(4750 RPM)(10,200 RPM)
Rate
10 (MAX)
9
8
7
6
5
4
3
2
1 (SLOW)
0 (OFF)
a. Times shown are with the rotor fully loaded in the Allegra X-15R. (Times are approximate; actual times will vary depending
on the rotor in use, the rotor load, run speed, and voltage fluctuations.)
AccelDecelAccelDecelAccelDecelAccelDecel
1:27
2:17
2:31
2:46
3:04
3:30
4:02
4:49
5:58
7:55
—
1:29
2:19
2:34
2:48
3:11
3:34
4:08
4:57
6:05
8:04
15:34
1:21
2:08
2:19
2:34
2:52
3:15
3:46
4:30
5:35
7:23
—
1:25
2:12
2:24
2:41
2:57
3:21
3:55
4:35
5:44
7:37
14:22
1:27
2:17
2:30
2:45
3:04
3:29
4:02
4:49
5:59
7:55
—
1:29
2:21
2:36
2:51
3:08
3:34
4:08
4:54
6:04
8:03
16:20
1:25
2:24
3:27
4:31
5:32
6:33
7:23
8:35
9:38
10:08
—
a
1:26
2:27
3:28
4:25
5:27
6:30
7:29
8:32
9:31
10:04
17:18
1
Press ACCEL, then use the keypad or arrow keys to select the required acceleration rate.
Entered acceleration rate appears on the display.
2
Press the next run-parameter key or ENTER/SAVE.
3
Press DECEL, then use the keypad or arrow keys to select the required deceleration rate.
Entered deceleration rate appears on the display.
4
Press the next run-parameter key or ENTER/SAVE.
Starting a Run
NOTE Wipe out moisture from the chamber and the chamber gasket before each run.
PN GXR-IM-4AC
2-7
Page 34
Operation
Manual Operation
1
2
Press ENTER/SAVE and START.
The light next to the START key flashes and the rotor begins to spin.
NOTE To begin a run, ENTER/SAVE must always be the last key pressed before pressing START. If you
wait more than 5 seconds, the START key will not activate. If this happens, press ENTER/SAVE and
START again to begin the run.
The run will end when the time display counts down to zero.
Press
STOP to terminate a run for any reason.
•The green light next to the
STOP key will flash while the rotor decelerates.
•A tone sounds when the rotor comes to a complete stop.
(Instructions for disabling the tone are at the end of this section.)
After the rotor has stopped and the LED lights, press DOOR.
The door opens.
3
Remove the rotor (or containers).
NOTE
equipment to decelerate or reset with the possible loss of in-process operation. If you suspect this has
occurred, simply repeat the runs.
Pulse Function
The pulse function, accessed by pressing the PULSE key, is used for short-duration runs as follows.
•When the
continues to spin as long as the
deceleration settings are overridden by the pulse function.) When the
the rotor begins decelerating to 0 RPM using maximum deceleration. Using the pulse function
eliminates the need to press the
CAUTION
If disassembly reveals evidence of leakage, you should assume that some fluid
escaped the rotor. Apply appropriate decontamination procedures to the
centrifuge and accessories.
Depending upon the duration of the event, transient power-line interruptions may cause this
PULSE key is pressed, the rotor accelerates at maximum rate to the set speed and
PULSE key is pressed. (The current run time, acceleration, and
PULSE key is released,
START and STOP keys.
2-8
•When the
the
PULSE key is pressed, the TIME display begins displaying the elapsed seconds. When
PULSE key is released, the seconds stop accumulating.
•The centrifuge memory retains the parameters of the last run performed before the
was pressed. At the end of a pulse run, after the centrifuge door is opened and closed the
previous run parameters are displayed.
PN GXR-IM-4AC
PULSE key
Page 35
Programmed Operation
The instrument internal memory can store up to 10 programs which can be recalled by keypad
numbers 0 through 9. Saved programs are retained in memory even if the power is turned off.
Procedures for entering run parameters are the same for programmed operation as those described
above for manual operation.
Creating a New Program
1
Press PROGRAM, then use the keypad or arrow keys to select a program number that is not in
use, or one that you want to overwrite, and press
The program number flashes.
2
Enter run parameters (rotor, speed, time, temperature, acceleration/deceleration rate
settings). Press
If you entered an unacceptable parameter, an error message will be displayed.
ENTER/SAVE.
ENTER/SAVE.
Operation
Programmed Operation
2
Press
CE and enter an acceptable value, then press the next run-parameter key.
3
When all parameters are entered, press ENTER/SAVE.
•The program is saved into memory and “
•The program number displayed will stop flashing.
If any parameters are then modified, the program number will show a dash
4
Press ENTER/SAVE and then START to run the programmed run.
NOTE
If the PROGRAM key is pressed while the program number is flashing, program mode is exited
without saving entered parameters into program memory.
Recalling a Program
1
Press PROGRAM, then use the keypad or arrow keys to enter the number of the program.
ENTER/SAVE to recall a program or press PROGRAM to exit program mode.
Press
Program saved” appears on the display.
PN GXR-IM-4AC
The program parameters are displayed if
ENTER/SAVE was pressed.
2-9
Page 36
Operation
Disabling/Enabling the Audible Tones
2
Press ENTER/SAVE and START to begin the programmed run.
The run starts, the green light next to the
NOTE To begin a run, ENTER/SAVE must always be the last key pressed before pressing START. If you
wait more than 5 seconds, the STARTSTART again to begin the run.
The run will end automatically when the time display counts down to zero.
Press
STOP to terminate a run at any time.
•The green light next to the
•A tone will sound when the rotor has come to a complete stop.
3
After the rotor has stopped and the LED is lit, press DOOR.
The door opens.
4
Remove the rotor (or containers).
key will not activate. If this happens, press ENTER/SAVEand
STOP key will flash while the rotor decelerates.
Disabling/Enabling the Audible Tones
START key flashes, and the rotor begins to spin.
The audible tones (beeps) that sound when a run-parameter key is pressed, or at the end of each
run, or if an error occurs may be turned on and off by following the steps below (rotor must be
at 0 RPM).
1
To disable the key or error tone press ENTER, then press 2 3 6.
The words “
2
To disable the end-of-run tone, press ENTER, then press 2 2 3.
The words “Disable RUN beep” appear on the display.
Tones that have been disabled can be enabled as follows.
1
To enable a key or error tone press ENTER, then press 2 3 5.
The words “
2
To enable the end-of-run tone press ENTER, then press 2 2 2.
The words “Enable RUN beep” appear on the display.
Disable KEY beep” appear on the display.
Enable KEY beep” appear on the display.
2-10
PN GXR-IM-4AC
Page 37
Introduction
This section lists possible malfunctions, together with probable causes and corrective actions.
Maintenance procedures are given in CHAPTER 4, Care and Maintenance. For any problems not covered
here, call Beckman Coulter Customer Service at 1-800-742-2345 (U.S.A. or Canada) or contact your local
Beckman Coulter Representative for assistance.
CHAPTER 3
Troubleshooting
NOTE
It is your responsibility to decontaminate the instrument, as well as any rotors and/or accessories,
before requesting service by Beckman Coulter Field Service.
User Messages
User messages appear on the display to communicate information about the instrument or to alert
you to abnormal conditions that need attention.
•Help and informational messages caused by incorrect input or certain operating conditions can
be cleared by pressing
•Diagnostic messages that result from abnormal operating conditions or equipment
malfunction require troubleshooting. Refer to Table 3.1 to determine the nature of the
condition and any recommended actions. If a problem persists after you have performed the
recommended action, call Beckman Coulter Field Service at 1-800-742-2345 (U.S.A. or Canada)
or contact your local Beckman Coulter Representative. To help the field service representative
diagnose and correct the problem, gather as much information about the situation as you can,
including:
— the diagnostic number and message,
— the operating situation when the diagnostic condition occurred (such as rotor in use,
— any unusual environmental and/or operating conditions (such as ambient temperature or
CE and following the instructions in the message.
speed, or load type), and
voltage fluctuations).
PN GXR-IM-4AC
3-1
Page 38
Troubleshooting
User Messages
Table 3.1 Diagnostic Message Chart
Diagnostic Number/
Message
ProblemResultRecommended Action
P1–
Power failure
occurred, see manual
L1 through L6–
Door latch, call
Service
L7–
Can’t open door, see
manual
C1–
Rotor temp exceeds
4C above set
C2–
Rotor temp exceeds
8C above set, call
Service
System or facility
power failure
Latches are not
operating properly
Door latches do not
release
Rotor temperature
exceeds temperature
setting by more than
4°C but less than 8°C
after equilibration
Rotor temperature
exceeds temperature
setting by more than
8°C after equilibration
Run shuts downPress CE to clear message. Restart run.
Remaining time shows on display.
Error message
appears; run shuts
down with no brake
Press down on both sides of the door until the
DOOR LED comes on. If you close the door
repeatedly and the problem continues, call
Beckman Coulter Field Service.
Door doesn’t open
when DOOR key is
pressed
Press down on the door, then press DOOR
again. If door still doesn’t open call
Beckman Coulter Field Service. (Procedures for
retrieving your sample are at the end of this
section.) If the door opens but the message
stays on the display, press CE to clear.
Run continues; the
Press CE to clear message.
message is
informational
Run shuts down
Call Beckman Coulter Field Service.
a
C3 and C5–
Temp error, call
Service
T1 and T2–
Temp, door locks for
45 minutes, call
Service
T3–
Temp, door locks for
45 minutes, see
manual
D1 through D9–
Drive error, call
Service
Refrigeration system
error
Run shuts down
with maximum
brake
System temperature
problem
Run shuts down
with no brake; door
locks for 45 minutes
Motor overheated Run shuts down
with no brake; door
locks for 45 minutes
Drive system problem Run stops, usually
with no brake
Call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
Wait 45 minutes for motor to cool down, then
restart run. If problem persists, call
Beckman Coulter Field Service.
Call Beckman Coulter Field Service. Before
trying to open the door, listen carefully and
make sure that no sound is coming from the
chamber (indicating a spinning rotor). Follow
the directions under Retrieving Your Sample in
Case of Power Failure, below.
3-2
PN GXR-IM-4AC
Page 39
Table 3.1 Diagnostic Message Chart (Continued)
Diagnostic Number/
Message
ProblemResultRecommended Action
Troubleshooting
User Messages
3
R1 and R2–
Rotor ID problem, call
Service
R3–
Rotor ID mismatch,
check rotor
R4 and R5–
No rotor match, see
manual
S1–
System error, call
Service
S8–
System error, door
locks for 45 minutes,
see manual
No magnets
identified, or magnets
incorrectly identified
The entered rotor
number is not the
same as the rotor
identified
The system cannot
identify the rotor
There is a problem
with the system
control software,
EPROM, or RAM
There is a problem
with the system
control software
Run continues,
speed may be
derated
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
System shuts down,
no brake
Run shuts down
with no brake, door
locks for 45 minutes
Press CE to clear message. If problem repeats,
call Beckman Coulter Field Service.
Press CE to clear message. Enter the correct
rotor entry identification.
• Make sure the rotor in use is a compatible
Beckman Coulter rotor (see Available
Rotors in CHAPTER 1). Check that magnets
are in place on rotor bottom. Press CE to
clear the message.
• Call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
Wait 45 minutes then press CE to clear
message, restart run. If problem persists, call
Beckman Coulter Field Service.
H1, H2, and H4–
Speed error, call
Service
H3, H5, and H6–
Speed error, door
locks for 45 minutes,
call Service
Speed control
problem
Speed control
problem
Run shuts down
with maximum
brake.
Run shuts down
with no brake; door
locks for 45 minutes
Call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
PN GXR-IM-4AC
3-3
Page 40
Troubleshooting
Other Possible Problems
Table 3.1 Diagnostic Message Chart (Continued)
Diagnostic Number/
Message
ProblemResultRecommended Action
I1–
Imbalance, see
manual
Rotor load is severely
out of balance
Rotor tie-down is not
properly fastened
Centrifuge is
misaligned (tilted)
Centrifuge was
moved during
operation
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
• Make sure that tubes or bottles are
balanced and loaded symmetrically in the
rotor as described in the applicable rotor
manual.
• With swinging bucket rotors, remove the
buckets and clean and lubricate the pivot
pin sockets where the buckets contact as
described in the applicable rotor manual.
Unlubricated pivot pin sockets can prevent
the buckets from reaching horizontal
position, which can cause imbalance.
• Press CE to clear message.
Make sure rotor lid is properly aligned and tiedown T-bar is securely fastened. Press CE to
clear message.
Level the centrifuge on the bench or table.
Press CE to clear message.
After the rotor comes to a complete stop, open
and close the centrifuge door, then restart.
Press CE to clear message.
Drive error
(mechanical damage)
I4, 15, and l6–
Imbalance error,
call Service
a. Call 1-800-742-2345 (U.S.A. or Canada) or contact your local Beckman Coulter Representative.
Imbalance sensor
failed or improperly
calibrated
Run shuts down
with maximum
brake
Cannot start runCall Beckman Coulter Field Service.
Other Possible Problems
Operating problems that may not be indicated by diagnostic messages are described in Table 3.2,
along with probable causes, listed in the probable order of occurrence, and corrective actions.
Perform the recommended corrective action in sequence, as listed. If you are unable to correct the
problem, call Beckman Coulter Field Service at 1-800-742-2345 (U.S.A. or Canada) or contact your
local Beckman Coulter Representative.
Call Beckman Coulter Field Service.
3-4
PN GXR-IM-4AC
Page 41
Table 3.2 Troubleshooting Chart
ProblemProblem/ResultRecommended Action
Troubleshooting
Other Possible Problems
3
Rotor cannot achieve
set speed
Door will not openRotor spinningWait until the rotor stops.
Displays are blankPower not onPlug in the power cord; turn power on.
Chamber does not
reach selected
temperature
Line voltage below ratingHave a qualified service person measure line
voltage while the instrument is operating.
Electrical failureMake sure the power cord is securely connected;
call Beckman Coulter Field Service.
Motor failureCall Beckman Coulter Field Service.
Power not onPlug in the power cord; turn power on.
Source power failureSee Retrieving Your Sample in Case of Power
Failure, below.
Electrical failureMake sure the power cord is securely connected;
call Beckman Coulter Field Service.
Centrifuge cannot
maintain selected
temperature for rotor in
use at speed selected
• Precool or prewarm rotors before running at low
or high temperatures.
• Precool rotor chamber by running a 30-minute
cycle at the required temperature with the
speed set at 2000 RPM.
• Make sure the air intake vent is clear.
• If using multiwell-plates for runs exceeding
2 hours in a warm and/or humid environment,
speed reduction may be required to maintain
low sample temperature.
PN GXR-IM-4AC
Chamber icing Excess humidity in
chamber
Condensation build-up
between runs
Wipe out moisture from the chamber and the
chamber gasket before each run.
• Leave door open between runs.
• Set temperature to a setting higher than
ambient temperature.
• Turn centrifuge power off.
3-5
Page 42
Troubleshooting
1
Retrieving Your Sample in Case of Power Failure
Retrieving Your Sample in Case of Power Failure
If facility power fails, you will have to restart the run when the power is restored. In the event of an
extended power failure, it may be necessary to trip the door-locking mechanism manually to
remove the rotor and retrieve your sample.
1
Turn the power off and disconnect the power cord from the main power source.
No displays or LEDs are lit.
2
Make sure that the rotor is not spinning.
No sound or vibration comes from the centrifuge.
WARNING
Never attempt to override the door interlock system while the rotor is spinning.
3
Locate the two joined release cords underneath the bottom center of the centrifuge (not visible
from the centrifuge front). See Figure 3.1.
Figure 3.1 Door Latch Override
1. Release cord
4
Pull on the cords or insert a screwdriver through the two cords and pull to release.
The latch releases and the door opens.
If the rotor is still spinning, close the door and wait until it stops before attempting to remove it.
3-6
WARNING
Never attempt to slow or stop the rotor by hand.
PN GXR-IM-4AC
Page 43
Introduction
This section contains care and maintenance procedures that should be performed regularly. For
maintenance not covered in this manual, call Beckman Coulter Customer Service at 1-800-742-2345 (U.S.A.
or Canada) or contact your local Beckman Coulter Representative for assistance. User messages and
recommended actions are discussed in CHAPTER 3, Troubleshooting.
CHAPTER 4
Care and Maintenance
NOTE
It is your responsibility to decontaminate the centrifuge, as well as any rotors and/or accessories,
before requesting service by Beckman Coulter Field Service.
Instrument Care
WARNING
Any maintenance procedure requiring removal of a panel exposes the operator to
the possibility of electrical shock and/or mechanical injury. Therefore, turn the
power off and disconnect the instrument from the main power source, and refer
such maintenance to service personnel.
Maintenance
Perform the following procedures regularly to ensure continued performance and long service life
of the centrifuge.
1
Inspect the centrifuge chamber for accumulations of sample, dust, or glass particles from
broken sample tubes.
Clean as required (see Cleaning, below).
PN GXR-IM-4AC
2
Check the air intake and exhaust for obstructions. Keep vents clear and clean.
4-1
Page 44
Care and Maintenance
Instrument Care
3
4
NOTE
Cleaning
Wipe condensation out of the rotor chamber between runs with a sponge or clean cloth to
prevent chamber icing.
If chamber icing occurs, defrost the system and wipe moisture out of the chamber before use.
To defrost the system, set the temperature to 30°C for 20 minutes.
(These are suggested settings which may be adjusted as appropriate for your laboratory
conditions.)
Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed method will not damage the
equipment.
Clean the centrifuge frequently. Always clean up spills when they occur to prevent corrosives or
contaminants from drying on component surfaces.
1
To prevent accumulations of sample, dust, and/or glass particles from broken sample tubes,
keep the chamber clean and dry by frequent wiping with a cloth or paper towel.
a. For thorough cleaning, wash the chamber using a mild detergent such as
Solution 555 (339555)
b. Dilute the detergent with water (10 parts water to 1 part detergent).
c.Rinse thoroughly and dry completely.
2
Clean the centrifuge exterior surfaces by wiping with a cloth dampened with Solution 555.
Dilute the detergent with water (10 parts water to 1 part detergent).
Do not use acetone.
3
Remove the rotor from the centrifuge and clean the drive hub regularly using Solution 555 and
a soft brush.
Dilute the detergent with water (10 parts water to 1 part detergent).
Rinse thoroughly and dry completely.
4-2
Tube Breakage
1
If a glass tube breaks, and all the glass is not contained in the bucket or rotor, be sure to
thoroughly clean the chamber.
PN GXR-IM-4AC
Page 45
WARNING
Be careful when examining or cleaning the chamber and chamber gasket, as sharp
glass fragments may be embedded in their surfaces.
2
Examine the chamber gasket to make sure that no glass particles are retained in it.
Carefully remove any glass particles that may remain.
3
Carefully wipe away any glass particles that remain in the chamber.
Decontamination
Care and Maintenance
Circuit Breaker and Fuses
4
If the centrifuge and/or accessories are contaminated with radioactive or pathogenic solutions,
perform all appropriate safety and decontamination procedures. Refer to Chemical Resistances
(publication IN-175) to be sure the decontamination method will not damage any part of the
instrument.
Sterilization and Disinfection
The centrifuge is finished with acrylic baking enamel. Ethanol (70%) may be used on this surface.
See Chemical Resistances for chemical compatibilities of centrifuge and accessory materials.
CAUTION
Ethanol is a flammability hazard. Do not use it in or near operating centrifuges.
While Beckman Coulter has tested ethanol (70%) and found that it does not damage the centrifuge,
no guarantee of sterility or disinfection is expressed or implied. When sterilization or disinfection
is a concern, consult your laboratory safety officer regarding proper methods to use.
Circuit Breaker and Fuses
PN GXR-IM-4AC
There are no user-replaceable fuses in the centrifuge.
If the centrifuge circuit breaker trips for any reason, the power switch will move to the off (O)
position. Reset the circuit breaker by turning the power switch back to the on (
again immediately, do not reset it. Call Beckman Coulter Field Service at 1-800-742-2345 (U.S.A. or
Canada) or contact your local Beckman Coulter Representative.
I) position. If it trips
4-3
Page 46
Care and Maintenance
Storage and Transport
CAUTION
Repeated attempts to reset the centrifuge circuit breaker can cause substantial
damage to electrical and electronic components.
Storage and Transport
Storage
To ensure that the centrifuge does not get damaged, contact Beckman Coulter Field Service for
specific instructions and/or assistance in preparing the equipment for transport or long-term
storage. Temperature and humidity conditions for storage should meet the environmental
requirements described under Specifications
in CHAPTER 1.
Supply List
Call Beckman Coulter Customer Service at 1-800-742-2345 (U.S.A. or Canada) or contact your local
Beckman Coulter office for detailed information on ordering parts and supplies. You may also
obtain information at www.beckmancoulter.com. For your convenience, a partial list is given
below.
Replacement Rotor Parts
DescriptionPart Number
Anti-rotation kit392988
Other
NOTE For MSDS information, go to the Beckman Coulter website at www.beckmancoulter.com.
DescriptionPart Number
Silicone vacuum grease (1 oz)335148
4-4
Solution 555 (1 qt)339555
Program library record233679
PN GXR-IM-4AC
Page 47
Introduction
Preinstallation requirements have been sent prior to shipment of your centrifuge.
APPENDIX A
Installation
WARNING
This centrifuge weighs 128 kg (283 lb). Do not attempt to lift or move it without
assistance.
Space and Location Requirements
WARNING
Do not place the centrifuge near areas containing flammable reagents or
combustible fluids. Vapors from these materials could enter the centrifuge air
system and be ignited by the motor.
The centrifuge ships in a cardboard box on a wooden pallet. For easy access, remove the top of the
box, the foam insert on top of the centrifuge, and then the upper part (sides) of the box and set them
aside. Then, with the help of another person, move the centrifuge from the pallet to its final position.
1
Position the centrifuge on a level surface, such as a sturdy table or laboratory bench that is able
to support the weight of the centrifuge (128 kg/283 lb) and resist vibration.
Make sure that the centrifuge front feet are fully supported on the table.
2
Locate the centrifuge in an area with sufficient ventilation to allow for heat dissipation.
3
Check that there are 7.6-cm (3-in.) clearances at the sides and back of the centrifuge to ensure
sufficient air circulation.
Dimensions are shown in Figure A.1.
PN GXR-IM-4AC
Additional clearance is required on the left side to allow access to the power switch.
A-1
Page 48
Installation
cm
in
96.5
38.0
76.2
30.0
7.6
3.0
7.6
3.0
34.3
13.5
62.2
24.5
Electrical Requirements
4
Ambient temperatures during operation should not be lower than 10°C (50°F) or higher than
35°C (95°C).
Relative humidity should not exceed 75% (noncondensing).
The centrifuge must have adequate air ventilation to ensure compliance to local requirements
for vapors produced during operation.
Figure A.1 Centrifuge Dimensions
Electrical Requirements
208-V, 60-Hz centrifuge187–229 VAC, 12 A, 60 Hz
220-V, 60-Hz centrifuge198–242 VAC, 12 A, 60 Hz
200-V, 50/60-Hz centrifuge180–220 VAC, 12 A, 50/60 Hz
230-V, 50-Hz centrifuge207–253 VAC, 12 A, 50 Hz
To reduce the risk of electrical shock, this centrifuge comes with a 2.5-m (8-ft) three-wire electrical
cord (attached to the power connector at the rear of the instrument) and plug to connect the
centrifuge to earth-ground. The plug for use with 208-VAC, 60-Hz centrifuges is the NEMA 6-15P;
a NEMA 6-15R socket should be available. In regions where the instrument is supplied with an
A-2
PN GXR-IM-4AC
Page 49
Installation
Securing the Centrifuge to the Bench
unterminated cord, a plug that meets local electrical and safety requirements must be supplied.
(Contact your local Beckman Coulter Service Office for specific information regarding these
requirements.)
1
Make sure that the matching wall outlet receptacle is properly wired and earth-grounded.
Check that the line voltage agrees with the voltage listed on the name rating plate affixed to the
centrifuge.
2
Never use a three-to-two wire plug adapter.
3
Never use a two-wire extension cord or a two-wire non-grounding type of multiple-outlet
receptacle strip.
4
If there is any question about voltage, have a qualified service person measure it under load
while the drive is operating.
A
To ensure safety, the centrifuge should be wired to a remote emergency switch (preferably outside
the room where the centrifuge is housed, or adjacent to the exit from that room) in order to
disconnect the centrifuge from the main power source in case of a malfunction.
Securing the Centrifuge to the Bench
Allegra X-15R centrifuges are certified to meet the requirements of the European CE mark. To meet
these requirements, the centrifuge must be secured to the bench using the anti-rotation kit shipped
with the instrument. This will prevent the centrifuge from moving in the unlikely event of a rotor
mishap.
Complete instructions for installing the anti-rotation kit are packaged with the hardware, which is
shipped with the centrifuge. The instructions (publication GXR-TB-002) include a full-size template
to be used as a guide for drilling holes in the bench. Refer to this document for installation
instructions.
The Allegra X-15R may also be secured to the benchtop surface by means of a non-invasive
centrifuge restraint system (NICRS), which uses an adhesive to secure the centrifuge to the
benchtop. The NICRS installation kit (A79118) contains two adhesive restraining plates, mounting
hardware, and easy to follow instructions to secure the Allegra X-15R. Contact your
Beckman Coulter representative for purchasing details.
PN GXR-IM-4AC
A-3
Page 50
Installation
Test Run
Test R u n
NOTE The centrifuge must be plugged in and the power switch turned to on position (I) before the door
can be opened.
We recommend that you make a test run to ensure that the centrifuge is in proper operating
condition following shipment. See CHAPTER 2, Operation for instructions on operating the
centrifuge.
A-4
PN GXR-IM-4AC
Page 51
Beckman Coulter, Inc.
Allegra X-15R Centrifuge Warranty
Subject to the exceptions and upon the conditions specified below and the warranty clause of the
Beckman Coulter, Inc. terms and conditions in effect at the time of sale, Beckman Coulter agrees to correct
either by repair or, at its election, by replacement, any defects of material or workmanship which develop
within one (1) year after delivery of an Allegra X-15R centrifuge (the product), to the original buyer by
Beckman Coulter or by an authorized representative, provided that investigation and factory inspection by
Beckman Coulter discloses that such defect developed under normal and proper use.
Some components and accessories by their nature are not intended to and will not function for as long as
one (1) year. A complete list of such components or accessories is maintained at the factory and at each
Beckman Coulter District Sales Office. The lists applicable to the products sold hereunder shall be deemed to
be part of this warranty. If any such component or accessory fails to give reasonable service for a reasonable
period of time, Beckman Coulter will repair or, at its election, replace such component or accessory. What
constitutes either reasonable service and a reasonable period of time shall be determined solely by
Beckman Coulter.
Replacement
Any product claimed to be defective must, if requested by Beckman Coulter, be returned to the factory,
transportation charges prepaid, and will be returned to Buyer with the transportation charges collect unless
the product is found to be defective, in which case Beckman Coulter will pay all transportation charges.
Conditions
Beckman Coulter makes no warranty concerning products or accessories not manufactured by it. In the event
of failure of any such product or accessory, Beckman Coulter will give reasonable assistance to the Buyer in
obtaining from the respective manufacturer whatever adjustment is reasonable in light of the manufacturer’s
own warranty.
Beckman Coulter shall be released from all obligations under all warranties, either expressed or implied, if
the product(s) covered hereby are repaired or modified by persons other than its own authorized service
personnel, unless such repair in the sole opinion of Beckman Coulter is minor, or unless such modification is
merely the installation of a new Beckman Coulter plug-in component for such product(s).
Disclaimer
IT IS EXPRESSLY AGREED THAT THE ABOVE WARRANTY SHALL BE IN LIEU OF ALL WARRANTIES OF FITNESS
AND OF THE WARRANTY OF MERCHANTABILITY AND THAT BECKMAN COULTER, INC., SHALL HAVE NO
LIABILITY FOR SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY KIND WHATSOEVER ARISING OUT OF THE
MANUFACTURE, USE, SALE, HANDLING, REPAIR, MAINTENANCE, OR REPLACEMENT OF THE PRODUCT.
PN GXR-IM-4AC
Warranty-1
Page 52
Beckman Coulter, Inc. Allegra X-15R Centrifuge Warranty
Warranty-2
PN GXR-IM-4AC
Page 53
English / Deutsch / Español / Français / Italiano / Portugués / Русский / 中文 / 日本語 / 한국어
Symbol
Symbol
Simbolo
Symbole
Simbolo
Simbole
символ
符号
記号
상징
Title / Titel / Titulo / Titre / Titolo / Titulo / Название / 标题 / タイトル / 제목
Dangerous voltage
Gefährliche electrische Spannung
Voltaje peligroso
Courant haute tension
Pericolo: alta tensione
Consult Instructions for Use
Konsultieren Sie Anwendungsvorschriften
Consulte las instrucciones para el uso
Consultez les instructions pour l'usage
Consulti le istruzioni per uso
Tensão perigosa
Опасное напряжение тока
危险电压
危険な電圧
위험한 전압
Cuidado, ulta documentos adjuntos
Внимание, советует с сопроводительными
документами
注意, 咨询附属单证
注意, 伴う文書に相談しなさい
주의, 동반 문서를 상담하십시오
Material infeccioso potencial
биологической опасности
可能的传染性物
潜在的な感染性物質
전염하는 물자
Fora (o poder)
На (мощности)
开 (电源)
ン (電源)
에 (힘)
Fora de (poder)
С (сила)
(电源)
ン (電源)
떨어져 (힘)
Terra de proteção (terra)
Защитное заземление (земля)
保护接地
保護アース (接地)
방어적인 지구 (지상)
Te r r a
Земли
接地
アース (接地)
지구 (지상)
Fabricante
производитель
制造商
メーカー
제조자
Consulte instruções para o uso
Советуйте с инструкциями для пользы
咨询使用说明书
使用説明に相談しなさい
사용 설명을 상담하십시오
Page 54
www.beckmancoulter.com
Related Documents
Chemical Resistances
for Beckman Coulter
Centrifugation Products
(IN-175)
RPM-to-RCF Conversion Chart
for Use when Running
Multiwell-Plate Carriers in the
SX4750A Rotor (GX-TB-015
Available in hard copy or
electronic pdf by request.
Available at
www.beckmancoulter.com
Anti-Rotation Kit for the
Allegra X-15R Centrifuge
(GXR-TB-002)