A Beckman Coulter Company
Miami, Florida 33196-2500 USA
Page 2
LEGAL NOTICES
Beckman Coulter, Inc. makes no representation that, upon furnishing this service manual, the holder of the manual will have
the necessary technical capabilities and know-how to properly troubleshoot and repair any of the equipment specified in the
manual. Beckman Coulter, Inc. assumes no liability whatsoever, including consequential and incidental damages, resulting
from improper operation of Beckman Coulter instruments after maintenance of Beckman Coulter instruments has been
performed by persons not employed by Beckman Coulter, Inc. Furthermore, Beckman Coulter, Inc. assumes no liability
whatsoever for any personal injury or property damage resulting from maintenance and/or repair of Beckman Coulter
instruments performed by persons not employed by Beckman Coulter, Inc.
READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED PERSONNEL
BEFORE ATTEMPTING TO OPERATE INSTRUMENT.
HAZARDS AND OPERATIONAL PRECAUTIONS AND LIMITATIONS
WARNINGS, CAUTIONS, and IMPORTANTS alert you as follows:
WARNING-Might cause injury.
CAUTION-Might cause damage to the instrument.
IMPORTANT -Might cause misleading results.
Beckman Coulter, Inc. urges its customers to comply with all national health and safety standards such as the use of barrier
protection. This may include, but it is not limited to, protective eyewear, gloves, and suitable laboratory attire when
operating or maintaining this or any other automated laboratory analyzer.
"This Service Manual contains confidential information of Beckman Coulter, Inc. and its receipt or possession does not
convey any rights to reproduce, disclose its contents, or to manufacture, use, or sell anything it may describe.
Reproduction, disclosure, or use without specific written authorization of Beckman Coulter, Inc. is strictly forbidden."
The material in the revision B change pages was updated for software version 1.03 and for any hardware
changes since revision A. The changes include updating the adjustment procedures for the bath assembly,
HGB blank, RBC/PLT gain, WBC/BASO, motor current, thresholds, and the optical bench; updating the
replacement procedures for the heater assembly, power supply, start switch, reagent syringes, count
syringe, sample prove, waste syringe, 5diff syringe, flow-cell coax, optical bench lamp, diluent reservoir,
and sample syringe; updating the procedure for testing and configuring the bar-code reader; updating the
parts lists; updating the tubing lists and associated circuit connections; adding procedures for balancing
+
the WBC count, setting the diff
information on the LX300 + printer.
The change page packet also includes the latest revision of the Pneumatic/Hyraulic Schematic, 7616069B.
Changes that are part of the most recent revision are indicated in the printed copy by a bar in the margin of
the amended page.
This document applies to the latest software listed and higher versions. When a subsequent software version affects the
information in this document, the changes will be included on minor revision change pages or summarized on a Notice of
Information Update form and will be released by service memo.
PN 4237616B
iii
Page 4
REVISION STATUS
iv
PN 4237616B
Page 5
LEGAL NOTICES, 2
REVISION STATUS, iii
CONTENTS, v
1INTRODUCTION, 1.1-1
1.1MANUAL DESCRIPTION, 1.1-1
Scope, 1.1-1
Notification of Updates, 1.1-1
Intended Audience, 1.1-1
Organization, 1.1-1
Numbering Format, 1.1-2
Special Headings, 1.1-3
Identifying a Normal Distribution, 2.6-3
Interference on the Lower End of the Platelet Distribution Curve, 2.6-3
Microcytic Interferences on the Upper End of the Platelet Distribution
Curve, 2.6-3
Microcytic Interference with a Distinct Valley between 18 fL and 25 fL, 2.6-4
Microcytic Interference with a Valley below 18 fL, 2.6-4
Interference with No Distinct Valley, 2.6-5
Inspection, 3.2-1
Unpack the Analyzer, 3.2-1
Unpack the Installation Kit, PN XEA484A, 3.2-1
Unpack the Waste Alarm Kit, PN 6912680, 3.2-1
Verify All Caution and Compliance Labels are in Place, 3.2-1
Connect the Waste System, 3.2-3
Connect the Waste Tubing, 3.2-3
Install the Waste Alarm, 3.2-3
Connect the Reagents, 3.2-5
Connect the Diluent Tubing, 3.2-5
Install the Reagent Bottles, 3.2-6
Install the Printer, 3.2-7
Power On the Instrument, 3.2-7
Enter Reagent Lot Numbers, 3.2-8
Prime the Instrument, 3.2-8
Configure the Instrument Printer Settings, 3.2-9
Set the User Mode, 3.2-9
Verification, 3.2-10
viii
3.3PRINTER INSTALLATION, 3.3-1
®
EPSON
LX™- 300 and LX™- 300+ Printer Connection, 3.3-1
Unpack the Printer, 3.3-1
Install the Knob, 3.3-1
Install the Ribbon Cartridge, 3.3-2
Connect the Printer, 3.3-3
Paper Feed Options, 3.3-4
Single Sheet Paper Feed Setup, 3.3-4
Loading Continuous Feed Paper Feed, 3.3-5
Configure the Printer, 3.3-7
LX300 Printer, 3.3-7
LX300+ Printer, 3.3-7
Complete the Instrument Installation, 3.3-9
PN 4237616B
Page 9
4SERVICE AND REPAIR PROCEDURES, 4.1-1
4.1GUIDELINES FOR SERVICING THE A
ANALYZER, 4.1-1
General Guidelines, 4.1-1
4.15 REAGENT TEMPERATURE CHECK AND ADJUSTMENT, 4.15-1
Purpose, 4.15-1
Tools/Supplies Needed, 4.15-1
Reagent Temperature Check, 4.15-1
Reagent Temperature Adjustment, 4.15-3
Preparation, 4.15-3
Adjustment, 4.15-4
PN 4237616B
4.16 BATH ENCLOSURE TEMPERATURE CHECK AND ADJUSTMENT, 4.16-1
Purpose, 4.16-1
Tools/Supplies Needed, 4.16-1
xi
Page 12
CONTENTS
Bath Enclosure Temperature Check, 4.16-1
Bath Enclosure Temperature Adjustment, 4.16-2
This manual provides the reference information and procedures needed for servicing and
maintaining the BECKMAN COULTER™ A
C
referred to as the A
online and in hard copy. The online manual is released on the Service Resource Kit CD-ROM,
PN 6417471.
This manual is to be used in conjunction with the following customer documents and does
not contain information and procedures already covered in these documents:
DocumentLanguagePart Number
•T 5diff hematology analyzer or the instrument). It is available both
INTRODUCTION
1
C
•T™ 5diff hematology analyzer (hereafter
1
Operator’s GuideEnglish
French
Italian
German
Spanish
Chinese
Host Transmission SpecificationEnglish4277065
Notification of Updates
Any service memo that affects the information in this manual will include either minor
revision change pages or a Notice of Information Update form for this manual. A Notice of
Information Update form will summarize the changes and will list the specific headings,
figures, and tables affected.
4237615
4237630
4237631
4237632
4237633
4237634
Intended Audience
To use this manual effectively, you need the following:
rAn operator’s knowledge of the A
rA thorough understanding of -
tBasic electronic and pneumatic principles and devices
tReagent systems
C
•T 5diff hematology analyzer
PN 4237616B
tQuality control
tTroubleshooting concepts
rThe ability to -
tUse basic mechanical tools and understand related terminology
tUse a digital voltmeter (DVM)
tRead pneumatic/hydraulic schematics and understand related terminology
tRead electronic schematics and understand related terminology
Organization
The material in this manual is organized into eight chapters and two appendices. To make it
easier to access the information:
1.1-1
Page 34
INTRODUCTION
MANUAL DESCRIPTION
rIn the online manual, each page has a Contents button linked to a master table of
contents and an Index button linked to an alphabetic index.
rIn the printed manual, there is a master table of contents at the beginning of the manual,
a chapter-specific table of contents at the beginning of each chapter, and an alphabetic
index at the end of the manual.
The chapters / appendices contain:
Chapter 1, INTRODUCTION - A brief description of this manual and essential safety
information.
Chapter 2, INSTRUMENT DESCRIPTION - An introduction to the A
analyzer and a description of how it functions.
Chapter 3, INSTALLATION PROCEDURES - Installation and verification procedures.
Chapter 4, SERVICE AND REPAIR PROCEDURES - The procedures for servicing/repairing
C
the A
•T 5diff hematology analyzer.
C
•T5 diff hematology
Chapter 5, MAINTENANCE PROCEDURES - The procedures for maintaining the
C
•T 5diff hematology analyzer.
A
Chapter 6, SCHEMATICS - The schematic diagrams and tubing lists.
Chapter 7, TROUBLESHOOTING - An error message table.
Chapter 8, PARTS LISTS - The master parts list followed by the illustrated parts list.
Appendix A, QUICK REFERENCE INFORMATION - Quick reference information:
tolerances and limits; connectors, test points and jumpers for the circuit cards; the software
menu trees; location diagrams and summarized functions for main analyzer components.
Appendix B, SOFTWARE INTERFACE - Tables of fatal and non-fatal error messages.
Appendix C, FLAG SENSITIVITY AND THRESHOLDS - An overview of the theory
including default values; also includes the setup procedures.
ABBREVIATIONS - A list of abbreviations, acronyms, and reference designators used in this
manual.
Numbering Format
Each chapter of this manual is further divided into topics that are numbered sequentially,
beginning at one. The numbering format for the topic heading, which is called the primary
heading, is chapter number, decimal point, topic number. For example, the primary heading
number for the third topic covered in Chapter 2 is 2.3.
1.1-2
The page, figure, and table numbers are tied directly to the primary heading number. For
example, Heading 2.3 begins on page 2.3-1, the first figure under Heading 2.3 is Figure 2.3-1
and the first table under Heading 2.3 is Table 2.3-1.
PN 4237616B
Page 35
INTRODUCTION
MANUAL DESCRIPTION
Primary headings always begin at the top of a right-hand page.
Note:
Special Headings
Throughout this manual, WARNING, CAUTION, IMPORTANT, ATTENTION, and Note
headings are used to indicate potentially hazardous situations and important or helpful
information.
WARNING
A WARNING indicates a situation or procedure that, if ignored, can cause serious personal
injury. The word WARNING is in bold-faced text in the printed manual and is red in the
online manual.
CAUTION
A CAUTION indicates a situation or procedure that, if ignored, can cause damage to the
instrument.The word CAUTION is in bold-faced text in the printed manual and is red in the
online manual.
IMPORTANT
An IMPORTANT indicates a situation or procedure that, if ignored, can result in erroneous
test results.The word IMPORTANT is in bold-faced text in the printed manual and is red in
the online manual.
1
ATTENTION
An ATTENTION contains information that is critical for the successful completion of a
procedure and/or operation of the instrument.The word ATTENTION is in bold-faced text in
the printed manual and is red in the online manual.
Note
A Note contains information that is important to remember or helpful in performing a
procedure.
Conventions
This manual uses the following conventions.
C
rInstrument or analyzer refers to the A
rMain card refers to the motherboard in the instrument.
rMain Menu refers to the initial menu displayed on the instrument after Startup.
rEach menu option consists of an item number followed by bold, uppercase text. For
example,
Note:
number next to the menu item indicates the numeric pushbutton on the front of the
analyzer that can be pressed to select the menu option.
rKeys on the analyzer keypad are in bold, uppercase letters. For example, press
indicates the operator should press the
3. REAGENTS
Both the menu item number and text are displayed on the LCD screen. The item
is the third option on the Main Menu.
•T 5diff hematology analyzer.
pushbutton on the instrument keypad.
ENTER
ENTER
PN 4237616B
1.1-3
Page 36
INTRODUCTION
MANUAL DESCRIPTION
rTo select a menu item,
tUse the arrow keys to highlight the desired menu item then press the
tSimply press the numeric pushbutton (on the front of the analyzer) that correlates
ENTER
pushbutton on the front of the analyzer to select the highlighted option.
or
with the desired option. This is the faster way to select a menu item.
For example, to select the
3. REAGENTS
tUse the down arrow to highlight the 3. REAGENTS
menu item from the Main Menu, you may:
option and then press the
ENTER
pushbutton on the front of the analyzer.
or
tPress the pushbutton labeled
on the front of the analyzer.
3
rSelect menu item tt sub-menu item indicates the software options you have to select, as
well as the order in which you should select them. For example, to prime the diluent
reagent:
From the Main Menu, select
3. REAGENTS
3. PRIME
tt
1. DILUENT
tt
.
rItalics us used to indicate screen messages. For example:
The message CYCLE IN PROGRESS. PLEASE WAIT . . . appears on the screen.
C
rA
rA
rA
rA
rA
•T 5diff Rinse reagent is sometimes referred to as Rinse.
C
•T 5diff Fix reagent is sometimes referred to as Fix.
C
•T 5diff Hgb Lyse reagent is sometimes referred to as Hgb Lyse.
C
•T 5diff WBC Lyse reagent is sometimes referred to as WBC Lyse.
C
•T 5diff Diluent reagent is sometimes referred to as Diluent or diluent.
rIn the electronic version of the manual:
tLinks to additional information are in blue and are underlined. To access the linked
information, select the blue underlined text.
1.1-4
tThe material is divided into many small sections (electronic files) to enhance the
loading and accessibility features.
tEvery primary heading is a separate file and whenever possible the amount of
material contained within one primary heading is limited to four to ten pages.
tIf a primary heading must be large, such as an illustrated parts list (IPL), invisible
breaks are added to the electronic file to further divide it.
Note: Unless you are scrolling, these divisions are invisible. If you choose to scroll
through the IPL, you will encounter stop points. When you scroll to the end of a
section and encounter a stop point, use the navigation bar to access the next
section.
tTo move from one section (electronic file) to the next in the HTML version of the
manual, use the right and left arrows on the navigation bars displayed at the top and
bottom of each section.
Graphics
All graphics, including screens and printouts, are for illustration purposes only and must not
be used for any other purpose.
PN 4237616B
Page 37
1.2SAFETY PRECAUTIONS
Electronic
INTRODUCTION
SAFETY PRECAUTIONS
1
WARNING
attached to power can cause personal injury from electric shock. Power down completely before removing
covers to access electronic components.
WARNING
service on the instrument, rings and other metal jewelry can become caught in the instrument. To avoid
personal injury or damage to the instrument, remove rings and other metal jewelry before performing
maintenance or service on the electronic components of the instrument.
CAUTION
circuit cards and components, the instrument could be damaged. To prevent damage to electronic
components, always be sure power is OFF before removing or replacing printed circuit cards and
components.
CAUTION
circuit cards and other electronic components. If there is a possibility of ESD damage with a procedure,
then perform that procedure at an ESD workstation, or wear an antistatic wrist strap attached to a metal part
of the chassis connected to an earth ground.
Risk of personal injury. Contacting exposed electronic components while the instrument is
Risk of personal injury or damage to electronic components. While performing maintenance or
Risk of damage to electronic components. If the power is ON while removing or replacing printed
Risk of damage to electronic components. Electrostatic discharge (ESD) can damage add-in
Biological
WARNING
the instrument with the doors open, you may become injured or contaminated. To prevent possible injury or
biological contamination, you must wear appropriate safety glasses, a lab coat, and gloves when servicing
the instrument with the doors open.
Risk of personal injury or contamination. If you do not properly shield yourself while servicing
PN 4237616B
Use care when working with pathogenic materials. Means must be available to decontaminate
the instrument, provide ventilation, and to dispose of waste liquid. Refer to the following
publications for further guidance on decontamination:
rBiohazards Safety Guide, 1974, National Institute of Health.
rClassifications of Etiological Agents on the Basis of Hazards, 3d ed., June 1974, Center
for Disease Control, U.S. Public Health Service.
Troubleshooting
Bring the following Warning to the customer’s attention before advising that customer to
perform any service, maintenance or troubleshooting procedures on the A
hematology analyzer. Also, make sure customers are aware of the Warning and information
labels shown in Figure 1.2-1.
C
•T 5diff
1.2-1
Page 38
INTRODUCTION
SAFETY PRECAUTIONS
WARNING
Risk of personal injury or contamination. If you do not properly shield yourself while performing
service, maintenance, and troubleshooting procedures, residual fluids in the instrument could injure or
contaminate you. Beckman Coulter recommends that you wear barrier protection, such as appropriate
safety glasses, a lab coat, and gloves throughout the performance of service, maintenance, and
troubleshooting procedures to avoid contact with cleaners and residual fluids in the instrument.
Figure 1.2-1 Warning and Information Label
MOD
c
A
T 5diff
MOD
NO.
ASSY
xxxxxx
NO.
50/60100-240
VOLTS
HZ AMPS
MANUFACTURED BY COULTER CORPORATION
A BECKMAN COULTER COMPANY
TO REDUCE THE RISK OF ELECTRICAL SHOCK DO NOT REMOVE THE COVER
OR BACK.
REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.
ELECTRIC SHOCK HAZARD. DISCONNECT UNIT FROM POWER SOURCE
PRIOR T O S E R VICING .
FOR CONTINUED PROTECTION AGAINSTR FIRE HAZARD, REPLACE ONLY
WITH SAME TYPE AND RATING OF FUSE.
FOR SAFETY REASONS, EQUIPMENT REQUI4RES CONNECTION TO
PROTECTIVE EARTH GROUND.
S/N
xxxxxx
WATTS
•
NO.
ASSY
NO.
VO LTS
xxxxxx
B
ECKMAN
C
OULTER
50-60100-240
HZAMPS
MANUFACTURED FOR BECKMAN COULTER INC.
11800 SW 147 AVENUE, MIAMI, FLORIDA 33196-2500 U.S.A.
PATENTS ISSUED AND/OR PENDING
MADE IN FRANCE
S/N
xxxxxx
0.9-2.0200
WATTS
AUTOMATED DIFFERENTIAL CELL COUNTER
F O R IN V IT R O D IA G N O S T IC U S E
CAUTION:
TO REDUCE THE RISK OF ELECTRICAL SHOCK DO NOT REMO VE THE CO VER
OR BACK.
REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.
ELECTRIC SHOCK HAZARD. DISCONNECT UNIT FROM PO W ER SOURCE
PRIOR TO SERVIC IN G.
F O R C O N T IN U E D P R O T E C T IO N A G A IN S T R F IR E H A Z A R D , R E P L A C E O N L Y
W ITH SAM E TYPE AN D RATIN G OF FUSE.
F O R S A F E T Y R E A S O N S , E Q U IP M E N T R E Q U IR E S C O N N E C T IO N T O
PROTECTIVE EARTH GROUND.
T H IS A R E A M A Y C O N T A IN
B I O H A Z A R D O U S M A T E R IA L
REFER TO PRODUICT REFERENCE
MANUAL FOR PROPER HANDLING
ALL CO VER S/PAN ELS M UST BE
SECUR ED IN PLACE PRIOR TO
IN STRUM ENT OPER ATIO N.
REFER TO PRO DUCT REFERENCE
M ANUAL FOR PROPER INSTALLATION.
Identifying a Normal Distribution, 2.6-3
Interference on the Lower End of the Platelet Distribution Curve, 2.6-3
Microcytic Interferences on the Upper End of the Platelet Distribution
Curve, 2.6-3
Microcytic Interference with a Distinct Valley between 18 fL and 25 fL, 2.6-4
Microcytic Interference with a Valley below 18 fL, 2.6-4
Interference with No Distinct Valley, 2.6-5
Diluent Input (Figure 2.8-10), 2.8-11
5diff Syringe and Flow Cell, 2.8-12
Probe and Probe Rinse, 2.8-13
Diluent to Baths, 2.8-13
Waste System, 2.8-15
2.9ELECTRONIC SYSTEM, 2.9-1
Plug/Jack Labels, 2.9-1
Optical Preamplifier Card, 2.9-1
LCD and Keypad Card, 2.9-1
LED Card, 2.9-1
Motor Interconnect Card, 2.9-1
Traverse Interconnect Card, 2.9-1
CONTENTS
2.10 SOFTWARE STRUCTURE, 2.10-1
Overview, 2.10-1
Menu Trees, 2.10-1
How to Select a Menu Item, 2.10-1
ILLUSTRATIONS
2.1-1User Interfaces on the AC•T 5diff Hematology Analyzer, 2.1-2
2.2-1Coulter Principle, 2.2-2
2.2-2Dual Focused Flow Process, 2.2-3
2.2-3Signal Processing, 2.2-4
2.2-4Basophil Thresholds, 2.2-5
2.2-5Bath Assembly, 2.2-6
2.2-6Sample Delivery Using Tangential Flow, 2.2-6
2.2-7CBC/DIFF Mode -
Sample Partitions inside the Probe, 2.2-7
2.2-8CBC Mode -
Sample Partitions inside the Probe, 2.2-7
2.3-1Sample Probe and LED at Start of a Cycle, 2.3-1
2.3-2Baths Assembly at Start of a Cycle, 2.3-1
2.3-3Rinsing Probe Exterior After Aspiration, 2.3-2
2.3-4Making the RBC/PLT First Dilution, 2.3-2
2.3-5Making the WBC/BASO Dilution, 2.3-3
2.3-6Making the DIFF Bath Dilution, 2.3-3
2.3-7Double Rinse of the Sample Probe, 2.3-4
2.3-8Aspirating from the First Dilution, 2.3-4
2.3-9Rinsing the Outside of the Probe, 2.3-5
PN 4237616B
2-iii
Page 42
CONTENTS
2.3-10 Making the RBC/Plt Dilution, 2.3-5
2.4-1Bath Assembly, 2.4-1
2.4-2Bath Assembly, 2.4-4
2.4-3Flow Cell Operation, 2.4-5
2.4-4DiffPlot Regions, 2.4-6
2.5-1Typical RBC Histogram, 2.5-1
2.5-2RBC1 and RBC2 Positions - RBC Histogram, 2.5-2
2.6-1Typical Plt Histogram, 2.6-1
2.6-2Area of the Plt Histogram Used to Determine the PDW Parameter Result, 2.6-2
2.6-3Typical Platelet Distribution Curve, 2.6-3
2.6-4Microcytic Interference with a Valley between 18 fL and 25 fL, 2.6-4
2.6-5Microcytic Interference with a Valley below 18 fL, 2.6-4
2.6-6Interference with no Distinct Valley, 2.6-5
2.7-1Areas Used to Determine WBC and BASO Parameter Results, 2.7-1
2.7-2DiffPlot Regions, 2.7-2
2.7-3Volume Thresholds, 2.7-5
2.7-4Absorbance Thresholds / NL, NE and MN Alarms, 2.7-6
2.8-1Valve 1 through Valve 16 Locations, 2.8-1
2.8-2Valve 17 and 18 Location, 2.8-1
2.8-3Valve 20 to Valve 31 Locations, 2.8-2
2.8-4Hgb Lyse Reagent Circuit, 2.8-5
2.8-5Fix Reagent Circuit, 2.8-6
2.8-6WBC Lyse Reagent Circuit, 2.8-7
2.8-7Rinse Reagent Supply Circuit, 2.8-8
2.8-8Probe Rinse Reagent Circuit, 2.8-9
2.8-9WBC/BASO Rinse Reagent Circuit, 2.8-10
2.8-10 Diluent Reagent Circuit, 2.8-11
2.8-11 Probe Diluent Reagent Circuit, 2.8-13
2.8-12 Bath Diluent Reagent Circuit, 2.8-14
2.8-13 Waste Circuit, 2.8-15
2.10-1 User Menu Tree, 2.10-2
2.10-2 Service Menu Tree, 2.10-3
2-iv
TABLES
2.1-1AC•T 5diff Hematology Analyzer Reagent Consumption, Software Version
1.03, 2.1-4
C
2.2-1A
2.4-1Technical Characteristics for Obtaining RBC and Platelet Counts, 2.4-2
2.4-2Technical Characteristics for the Measurement of the Hemoglobin, 2.4-3
2.4-3Characteristics Required to Obtain WBC and BASO Results, 2.4-4
2.4-4Technical Characteristics for Acquisition of the DiffPlot, 2.4-6
2.4-5Summary of Dilutions, 2.4-7
2.8-1Valves and their Functions, 2.8-3
•T 5diff Analyzer Measurement Technologies, 2.2-1
PN 4237616B
Page 43
INSTRUMENT DESCRIPTION
2
2.1INTRODUCTION TO THE AC•T 5diff HEMATOLOGY ANALYZER
Purpose
The purpose of the AC•T 5diff hematology analyzer is to identify normal patient results with
all normal system-generated parameters and to flag or identify patient results that require
additional studies.
Function
The AC•T 5diff analyzer is a quantitative, fully automated (microprocessor controlled)
hematology analyzer and leukocyte differential counter
C
clinical laboratories. The A
(CBC) and white blood cell differential (DIFF) on open-vial, whole-blood specimens.
The CBC consists of white blood cell count (WBC), red blood cell count (RBC), hemoglobin
(Hgb), hematocrit (Hct), mean cell volume (MCV), mean cell hemoglobin (MCH), mean cell
hemoglobin concentration (MCHC), red cell distribution width (RDW), platelet count (Plt),
and mean platelet volume (MPV).
The DIFF (a 5-part leukocyte differential) consists of the percentage (%) and absolute
number (#) of the following WBC populations: neutrophils (NE% and NE#), lymphocytes
(LY% and LY#), monocytes (MO% and MO#), eosinophils (EO% and EO#), and basophils
(BA% and BA#).
•T 5diff hematology analyzer reports a complete blood count
For In Vitro Diagnostic Use
2
in
Six parameters are qualitative and are
Procedures
(PDW), percentage and absolute number of immature cells (IMM% and IMM#), and
percentage and absolute number of atypical lymphocytes (ATL% and ATL#).
. These parameters include the plateletcrit (Pct), platelet distribution width
For Research Use Only. Not For In Vitro Diagnostic
Description
Components
C
The A
only for the external printer, the diluent reagent container, and a waste container (if used).
Interaction with the A
The A
(Figure 2.1-1) initiates a cycle. When the sample probe is submerged in a whole-blood
specimen and the aspirate switch is pressed, sample is pulled from the specimen tube into the
sample probe. As the cycle continues, the instrument then dilutes and analyzes this sample.
When the analysis is complete, results appear on a LCD (Figure 2.1-1) and are available to
the printer.
•T 5diff hematology analyzer is contained in one unit, with additional space needed
C
•T 5diff Hematology Analyzer
C
•T 5diff analyzer uses an Open-Vial mode of operation. Pressing the aspirate switch
PN 4237616B
2.1-1
Page 44
INSTRUMENT DESCRIPTION
INTRODUCTION TO THE A
C
Figure 2.1-1 User Interfaces on the AC•T 5diff Hematology Analyzer
LC D
Keypad
Aspirate
sw itch
•T 5diff HEMATOLOGY ANALYZER
7616094A
You also interact with the instrument through the use of a menu system displayed on a 128 by
240 pixels LCD and a control panel keypad with buttons that are used to setup and operate
the instrument (Figure 2.1-1).
Since most input/output functions of the operating system software are controlled by the user,
the pushbutton keypad and LCD screen are particularly important because they provide the
physical user interface with the software.
See Heading 2.10, SOFTWARE STRUCTURE for more specific information as well as a
graphic representation of the available menus and menu items (or options).
2.1-2
PN 4237616B
Page 45
INSTRUMENT DESCRIPTION
C
INTRODUCTION TO THE A
•T 5diff HEMATOLOGY ANALYZER
Modes of Operation
The AC•T 5diff hematology analyzer has two operating modes: CBC and CBC/DIFF.
CBC Mode
Twelve parameters are generated in the CBC mode of operation - 10 parameters
Diagnostic Use
Vitro Diagnostic Procedures
and two qualitative parameters that are
:
For Research Use Only. Not For In
2
For In Vitro
Parameters
(For In Vitro Diagnostic Use)
WBCPct
RBCPDW
Hgb
Hct
MCV
MCH
MCHC
RDW
Plt
MPV
Parameters
(For Research Use Only)
CBC/DIFF Mode
26 parameters are generated in the CBC/DIFF mode of operation - 20 parameters
Diagnostic Use
Vitro Diagnostic Procedures
(For In Vitro Diagnostic Use)
and six qualitative parameters that are
:
Parameters
Parameters
(For Research Use Only)
For Research Use Only. Not For In
For In Vitro
PN 4237616B
WBCPct
RBCPDW
HgbIMM% and IMM#
HctATL% and ATL#
MCV
MCH
MCHC
RDW
Plt
MPV
NE% and NE#
LY % and LY #
MO% and MO#
EO% and EO#
BA% and BA#
2.1-3
Page 46
INSTRUMENT DESCRIPTION
INTRODUCTION TO THE A
C
Reagent Consumption
Table 2.1-1 shows the instrument reagent consumption by cycle.
For one background count only. The maximum is three.
2.1-4
PN 4237616B
Page 47
2.2OPERATION PRINCIPLES
Overview
The AC•T 5diff analyzer is a fully automated hematology analyzer providing a complete WBC
five-part differential, which is determined simultaneously by the A
(Absorbance Cytochemistry and Volume Technology) and the white blood cell / basophil
(WBC/BASO) methodologies.
C
The A
WBC/BASO methodology uses differential lysis, impedance technology, and differential
thresholds. See Table 2.2-1.
V Technology uses absorbance, cytochemistry, and focused flow impedance. The
Volume apertureCoulter PrincipleVolume and countRBC count, platelet count,
C
V Technology
A
Coulter Principle
me
Light absorbance of
cytochemically-stained
cells
Volume and countWBC count, basophil
Lymphocytes, monocytes,
neutrophils, eosinophils,
immature cells, and
atypical lymphocytes
percentage, and basophil
count
and hematocrit
Measurement Principles
Coulter Principle
In the A
platelet (RBC/Plt) dilution and the WBC/BASO dilution. This electronic method of counting
and sizing particles is based on the fact that cells, which are poor conductors of electricity,
will interrupt a current flow. The impedance variation generated by the passage of
nonconductive cells through a small, calibrated aperture is used to determine the count
(number of particles) and size (volume) of the particles passing through the aperture within a
given time period.
C
•T 5diff analyzer, the Coulter Principle is used to analyze the final red blood cell and
PN 4237616B
Aperture Sensor System
The RBC/Plt aperture sensor system determines the cell count and size of red blood cells and
platelets. The WBC/BASO aperture sensor system determines the cell count and size of white
blood cells. Additionally, the differentiation between basophils and other white blood cells is
C
related to the A
•T 5diff WBC Lyse-specific lytic action on the white blood cells in the
WBC/BASO bath.
2.2-1
Page 48
INSTRUMENT DESCRIPTION
V
OPERATION PRINCIPLES
To sense particles using the Coulter Principle (Figure 2.2-1), a current flow is established so
changes in that flow can be monitored. In this sensing system, an electrode is placed on each
side of the aperture (Figure 2.2-1). The most visible electrode is referred to as the counting
head. These electrodes are the conductive metallic housings attached to the front of the RBC
and WBC/BASO baths. The second electrode, referred to as the bath electrode, is not as
conspicuous. This electrode is located inside the bath. The aperture is located between the
counting head and the bath electrode.
Figure 2.2-1 Coulter Principle
Solution to be analyzed
Vacuum
constant
Constant
current
olts
Pulse
Electrodes
Time
Analyzing
electronic
circuit
7616035A
When the count circuit is activated and an electronically conductive reagent is in the RBC or
WBC/BASO bath, an electric current continuously passes through the aperture. Current
moving between the two electrodes establishes the electronic flow through the aperture.
Once a sample is aspirated, an aliquot of that aspirated sample is diluted with reagent (an
electrolyte) and is delivered to the RBC or WBC/BASO bath using tangential flow, which
ensures proper mixing of the dilution. When the cells suspended in the conductive reagent
are pulled through a calibrated aperture, the electrical resistance between the two electrodes
increases proportionately with the cell volume (Figure 2.2-1).
The resistance creates a pulse that is sensed and counted as a particle by the instrument. The
amount of resistance (amplitude of each pulse) is directly related to the size of the particle
that produced it.
The generated pulses have a very low voltage, which the amplification circuit increases so
that the electronic system can better analyze the pulses and eliminate the background noise.
Applying the Coulter Principle
C
The A
•T 5diff analyzer makes several dilutions of an aspirated whole-blood sample. The
RBC/Plt dilution begins in the DIL1/HGB (first dilution/hemoglobin) bath but is actually
analyzed in the RBC bath. The final dilution in the RBC bath is used to determine the cell
count and size of red blood cells and platelets.
The WBC/BASO aperture sensor system is directly responsible for determining the cell count
and size of white blood cells. The differentiation between basophils and other white blood
C
cells is also related to the A
cells.
2.2-2
•T 5diff WBC Lyse-specific lytic action on these white blood
PN 4237616B
Page 49
INSTRUMENT DESCRIPTION
OPERATION PRINCIPLES
Thresholds, which are electronically set size limits, exclude unwanted particles, such as
debris, from the analysis. Particles above the threshold are analyzed, and particles below the
threshold are excluded.
ACV Technology
In the DIFF (differential) bath, 25 µL of whole blood is mixed with 1,000 µL of AC•T5diff
C
Fix reagent for 12 seconds, then stabilized with 1,000 µL of A
additional three seconds. This reaction lyses the red blood cells, preserves the leukocytes at
their original size, and differentially stains the lymphocytes, monocytes, neutrophils, and
eosinophils, with eosinophils staining most intensely. The instrument maintains the reagents
and reaction at a regulated temperature of 35°C (95°F).
Lymphocytes, monocytes, neutrophils, and eosinophils each have a unique nuclear and
morphology structure and staining intensity; therefore, each cell type absorbs light differently.
Each stained cell is individually focused by the Dual Focused Flow (DFF) system and
transported through the flow cell using sample pressure and diluent sheath flow.
Dual Focused Flow (DFF)
DFF fluid dynamics uses a hydrodynamic focusing process to focus individual cells or
particles in a stream of diluent (Figure 2.2-2). The focused sample stream of the A
analyzer is about 40 µm in diameter.
•T 5diff Diluent for an
C
•T5diff
2
Figure 2.2-2 Dual Focused Flow Process
DFF uses sheath fluid to surround and force cells suspended in diluent to pass one at a time
through the center of the flow cell. The first sheath flow focuses the sample through the
impedance aperture. The second sheath flow maintains the focused flow of cells as they exit
the aperture into the optical flow cell. Hydrodynamic focusing in the flow cell enables
accurate and rapid cell-by-cell measurements on a large number of individual cells.
Flow Cell
Sequential analyses for cell volume (impedance) and light absorbance are performed in the
flow cell. A total of 72 µL of sample is injected through the flow cell for 15 seconds. The flow
cell incorporates a 60 µm aperture for cellular volume analysis and about a 40 µm
measurement area for light absorbance.
PN 4237616B
2.2-3
Page 50
INSTRUMENT DESCRIPTION
OPERATION PRINCIPLES
Focused Flow Impedance
Focused flow impedance technology measures the electrical resistance of a cell as it passes
through the aperture in the flow cell. The change in resistance is directly proportional to the
volume of the cell.
Absorbance Cytochemistry
As a cell passes through the optical portion of the flow cell, light is scattered in all directions.
A sensor detects only forward scattered light. The optical measurement is derived as a
function of the amount of light lost due to diffraction and absorbance, as compared to full
transmission when no cell is present.
The collected signals are converted into voltage pulses and are processed. The magnitude of
the voltage pulses are proportional to the physical and chemical characteristics of the cells
being analyzed. Light absorbance is related to cellular contents (granularity, nuclear content,
and so forth) after cytochemical staining. These measurements provide the information for
lymphocytes, monocytes, neutrophils, and eosinophils, and their precursors.
Signal Processing
The signals from the flow cell aperture and from the optical measurement are correlated by a
window of time. The optical pulse must be detected within 100 to 300 microseconds of the
impedance pulse; otherwise, the signal is rejected.
The output signals from the focused flow impedance and the light absorbance measurements
are combined to define the WBC differential population clusters. See Figure 2.2-3.
Figure 2.2-3 Signal Processing
Thresholds
Most of the population partition thresholds are fixed and give the limits of the morphological
normality of leukocytes. Changes in the morphology of a population are expressed on the
DiffPlot by a shifting of the corresponding population. Volume and absorbance thresholds are
used to detect shifting populations.
2.2-4
PN 4237616B
Page 51
INSTRUMENT DESCRIPTION
OPERATION PRINCIPLES
WBC/BASO Methodology
In the WBC/BASO bath, 10 µL of whole blood is mixed with 2,000 µL of AC•T5diff WBC
Lyse reagent. This reaction lyses the red blood cells and specifically differentiates between
basophils and other leukocytes by volume. The instrument maintains the reagents and
reaction at a regulated temperature of 35°C (95°F).
Using a constant vacuum, the instrument then pulls the sample through an 80 µm aperture.
As each cell passes through the aperture, a pulse is generated proportional to the cellular
volume. The total leukocyte count and basophil percentage are determined by specific
thresholds on the WBC/BASO histogram (Figure 2.2-4).
Figure 2.2-4 Basophil Thresholds
2
Sample Analysis Overview
Aspiration
When the sample probe is immersed in a whole-blood specimen and the aspirate switch is
pressed, sample is pulled from the tube into the sample probe. Depending on the selected
C
mode of operation, the A
(CBC/DIFF mode) of sample.
The volume of sample aspirated into the sample probe is sufficient to make all the dilutions
needed to develop parameter results in the selected mode of operation.
•T 5diff analyzer aspirates either 30 µL (CBC mode) or 53 µL
PN 4237616B
2.2-5
Page 52
INSTRUMENT DESCRIPTION
A
OPERATION PRINCIPLES
Dilution
Using the Sequential Dilution System (SDS) technique, the aspirated sample is partitioned as
it is distributed to make a series of dilutions in a series of baths (Figure 2.2-5).
Figure 2.2-5 Bath Assembly
Delivery
In the CBC and the CBC/DIFF modes, each aliquotted sample is delivered to its appropriate
bath using a tangential flow (Figure 2.2-6) of reagent, which mixes the diluted sample and
minimizes viscosity problems.
Figure 2.2-6 Sample Delivery Using Tangential Flow
Probe
Reagent
input
Bath
Tangential flow
7616002
2.2-6
PN 4237616B
Page 53
INSTRUMENT DESCRIPTION
OPERATION PRINCIPLES
Sample Partitioning
Figure 2.2-7 shows the sample partitioning that occurs in the CBC/DIFF mode. Notice there
are three aliquots of the aspirated whole-blood sample that will be used to make dilutions.
2
Figure 2.2-7 CBC/DIFF Mode Sample Partitions inside the Probe
CBC/DIFF Mode
After aspiration in the CBC/DIFF mode, aliquots of
the whole-blood sample are distributed to the
Diluent
various baths as follows (Figure 2.2-5):
rThe 3 µL sample aliquot at the tip of the probe
Air bubble
is discarded into the rinse chamber as the
exterior of the sample probe is rinsed,
ensuring sample integrity.
Not used
DIFF dilution
r10 µL of sample is delivered to the DIL1/HGB
bath for use in preparing the primary RBC/Plt
dilution and for measuring the Hgb value.
r10 µL of sample is delivered to the WBC/BASO
WBC/BASO dilution
RBC/PLT/HGB first dilution
Not used
7616001A
bath for the WBC/BASO count.
r25 µL of sample is delivered to the DIFF bath
for development of the DiffPlot.
r5 µL of remaining sample is discarded into the
rinse chamber.
Figure 2.2-8 shows the sample partitioning that occurs in the CBC mode. Notice there are
only two aliquots of the aspirated whole-blood sample that will be used to make dilutions in
this mode of operation. (The DIFF aliquot is not needed in the CBC mode.)
PN 4237616B
Figure 2.2-8 CBC Mode Sample Partitions inside the Probe
Diluent
Air bubble
Not used
WBC/BASO dilution
RBC/PLT/HGB first dilution
Not used
CBC Mode
After aspiration in the CBC mode, aliquots of the
whole-blood sample are distributed to the various
baths as follows (Figure 2.2-5):
rThe 3 µL sample aliquot at the tip of the probe
is discarded into the rinse chamber as the
exterior of the sample probe is rinsed,
ensuring sample integrity.
r10 µL of sample is delivered to the DIL1/HGB
bath for use in preparing the primary RBC/Plt
dilution and for measuring the Hgb value.
r10 µL of sample is delivered to the WBC/BASO
bath for the WBC/BASO count.
r7 µL of remaining sample is discarded into the
rinse chamber.
7616056A
2.2-7
Page 54
INSTRUMENT DESCRIPTION
OPERATION PRINCIPLES
2.2-8
PN 4237616B
Page 55
2.3CYCLE DESCRIPTION
This cycle description focuses on the sequence of the sample probe movement among the
baths. It also focuses on the volume of sample and reagents being delivered to make the
dilutions needed for sample analysis.
Cycle Start Conditions
INSTRUMENT DESCRIPTION
CYCLE DESCRIPTION
2
Figure 2.3-1 Sample Probe and LED at Start of a Cycle
Figure 2.3-2 Baths Assembly at Start of a Cycle
rThe sample probe is in its home position.
rThe green LED is glowing indicating the
instrument is ready.
rAll the baths (except the rinse chamber) are
filled with clean diluent.
PN 4237616B
2.3-1
Page 56
INSTRUMENT DESCRIPTION
CYCLE DESCRIPTION
Sample Flow
rTo initiate a cycle, submerge the sample probe in
a well-mixed whole-blood specimen and press
the aspirate switch to start the cycle.
rAll the baths drain.
rA sample of the whole-blood specimen is
aspirated.
t53 µL in the CBC/DIFF mode.
t30 µL in the CBC mode.
Figure 2.3-3 Rinsing Probe Exterior After Aspiration
Figure 2.3-4 Making the RBC/PLT First Dilution
rThe horizontal traverse assembly positions the
sample probe over the rinse chamber.
r3 µL sample aliquot at the tip of the sample
probe is discarded into the rinse chamber as the
exterior of the sample probe is rinsed.
Discarding this aliquot helps ensure sample
integrity.
rThe horizontal traverse assembly positions the
sample probe over the DIL1/HGB (first
dilution/Hgb) bath.
rThe vertical traverse assembly moves the probe
downward into the bath. The probe tip is
positioned to produce a tangential flow when the
sample and diluent are simultaneously dispensed
into the bath. For a more detailed description of
tangential flow, see Delivery under Heading 2.2,
OPERATION PRINCIPLES.
2.3-2
r10 µL of the whole-blood partitioned for making
the first dilution is delivered to the DIL1/HGB
bath using a tangential flow of 1.7 mL of diluent.
rThe tangential flow of reagent mixes the sample
and the diluent. Mixing bubbles enter the bath to
make a uniform suspension of cells. This 1:170
dilution is commonly referred to as the first
dilution.
PN 4237616B
Page 57
INSTRUMENT DESCRIPTION
CYCLE DESCRIPTION
2
Figure 2.3-5 Making the WBC/BASO Dilution
Figure 2.3-6 Making the DIFF Bath Dilution
rThe horizontal traverse assembly positions the
sample probe over the WBC/BASO bath
rThe vertical traverse moves the probe downward
into the bath. The tip of the probe is positioned
so that a tangential flow occurs as the 10 µL of
the whole-blood sample and 2.0 mL of WBC
Lyse are simultaneously dispensed into the bath.
rThe tangential flow of reagent mixes the sample
and reagent. Mixing bubbles enter the bath to
make a uniform suspension of cells. The WBC
Lyse destroys the red blood cells and the specific
lytic action on the white blood cells
differentiates the basophils from other WBCs.
WBC/BASO Bath Dilution
Whole-blood volume10 µL
C
Volume of A
Dilution ratio1:200
•T 5diff WBC Lyse reagent
2000 µL
rThe horizontal traverse assembly moves the
sample probe over the DIFF bath.
rThe vertical traverse assembly moves the probe
downward into the bath.
rThe tip of the probe is positioned so that a
tangential flow occurs as 25 µL of the
whole-blood sample and 1.0 mL of Fix reagent
are simultaneously dispensed into the bath.
rThe tangential flow of reagent mixes the sample
and the Fix reagent. Mixing bubbles enter the
bath to make a uniform suspension of cells. The
Fix reagent lyses the red blood cells, stabilizes
the WBCs in their native form, and differentially
stains the lymphocytes, monocytes, neutrophils,
and eosinophils, with the eosinophils staining
most intensely.
rAfter 12 seconds of incubation, the staining
process inside the DIFF bath is completed by
adding another 1.0 mL of diluent which stops
the cytochemical reaction.
DIFF Bath Dilution
Whole-blood volume25 µL
C
Volume of A
Volume of A
Final dilution ratio1:80
•T 5diff Fix reagent
C
•T 5diff Diluent
1000 µL
1000 µL
PN 4237616B
2.3-3
Page 58
INSTRUMENT DESCRIPTION
CYCLE DESCRIPTION
Figure 2.3-7 Double Rinse of the Sample Probe
Figure 2.3-8 Aspirating from the First Dilution
rThe horizontal traverse assembly moves the
sample probe over the rinse chamber.
rA double rinsing (interior and exterior) of the
probe removes residual whole-blood sample
from inside the probe.
tIn the CBC/DIFF mode, 5 µL is discarded in
the rinse chamber.
tIn the CBC mode, 7 µL is discarded in the
rinse chamber.
rThe horizontal traverse assembly moves the
sample probe over the DIL1/HGB bath.
rThe vertical traverse assembly moves the probe
downward into the bath.
r42.5 µL of the 1:170 first dilution is aspirated
into the sample probe.
2.3-4
PN 4237616B
Page 59
INSTRUMENT DESCRIPTION
CYCLE DESCRIPTION
2
Figure 2.3-9 Rinsing the Outside of the Probe
rWhile still inside the DIL1/HGB bath, the
exterior of the sample probe is rinsed with
0.4 mL of diluent.
rThe vertical traverse assembly moves the probe
up out of the bath.
r0.4 mL of Hgb Lyse is added to the bath. The
Hgb Lyse reagent rapidly destroys the red blood
cells and converts a substantial proportion of the
hemoglobin to a stable pigment so a hemoglobin
value can be determined.
rMixing bubbles enter the bath to ensure a
uniform dilution.
DIL1/HGB Bath Dilution
First dilution1:170
Volume of first dilution removed 42.5 µL
C
Volume of A
Volume of A
Final dilution ratio1:250
•T 5diff Hgb Lyse reagent
C
•T 5diff Diluent reagent
400 µL
400 µL
Figure 2.3-10 Making the RBC/Plt Dilution
rThe horizontal traverse assembly moves the
sample probe over the RBC bath.
rThe vertical traverse assembly moves the probe
downward into the bath.
rThe tip of the probe is positioned so that a
tangential flow occurs as the 42.5 µL of 1:170
dilution obtained from the first dilution in the
DIL1/HGB bath and 2.0 mL of diluent are
simultaneously dispensed into the bath
rAn additional 0.5 mL of diluent is dispensed
through the probe at the end of the second
dilution.
RBC Bath Dilution
Volume 1:170 dilution from
C
Volume of A
Final dilution ratio1:10,000
•T 5diff Diluent reagent
DIL1/HGB
bath 42.5 µL
2500 µL
PN 4237616B
2.3-5
Page 60
INSTRUMENT DESCRIPTION
CYCLE DESCRIPTION
2.3-6
PN 4237616B
Page 61
2.4SAMPLE ANALYSIS
RBC and Platelet Analysis
The RBC/Plt dilution analyzes red blood cells and platelets. This dilution is prepared in two
stages – the primary (first) dilution and the secondary (last) dilution.
The primary dilution is made in the DIL1/HGB bath, and the secondary dilution is made in
the RBC bath (Figure 2.4-1).
Figure 2.4-1 Bath Assembly
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
2
PN 4237616B
Table 2.4-1 summarizes the technical characteristics required to obtain RBC and Platelet
results.
2.4-1
Page 62
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
Table 2.4-1 Technical Characteristics for Obtaining RBC and Platelet Counts
Dilution Characteristics
Primary Dilution for RBC and Plt:
Initial volume of whole-blood10 µL
Volume A
Primary dilution ratio1:170
Secondary Dilution for RBC and Plt
Volume of primary dilution42.5 µL
Volume A
Secondary dilution ratio1:58.8
Final dilution for RBC and Plt results1:170 x 1:58.8 = 1:10,000
Reaction temperature35°C (95°F)
Measurement Characteristics
C
•T 5diff diluent
C
•T 5diff diluent
1700 µL
:
2500 µL
Method of analysisCoulter Principle
Aperture diameter50 µm
Count vacuum200 mb (5.9 in. Hg)
Count period2 x 5 seconds
Parameter Results Obtained from the RBC/Plt Dilution
This final 1:10,000 RBC/Plt dilution is used to:
rDetermine the RBC count.
rDevelop the RBC histogram, which is needed to obtain the Hct, MCV, and RDW results.
rDetermine the Plt count.
rDevelop the Plt histogram, which is needed to obtain the MPV, Pct, and PDW results.
2.4-2
PN 4237616B
Page 63
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
Hgb Measurement
Hemoglobin is determined from the dilution in the DIL1/HGB bath (Figure 2.4-1). This
dilution is prepared in two stages – the primary (first) dilution and the secondary (last)
dilution.
The primary dilution is made and 42.5 µL of that dilution is removed for making the RBC/Plt
C
dilution. A
dilution.
The Hgb concentration is based on the transmittance of light through the optical part of the
DIL1/HGB bath using a spectrophotometric technique at a wavelength of 550 nm. The
transmittance of the sample dilution is compared to the transmittance of a reagent blank. The
system calculates the Hgb using the blank and sample readings.
Table 2.4-2 summarizes the technical characteristics required for measuring hemoglobin.
•T 5diff Hgb Lyse and additional diluent are added to make the final 1:250
2
Table 2.4-2 Technical Characteristics for the Measurement of the Hemoglobin
Dilution Characteristics
Volume of whole-blood10 µL
C
Volume A
Preliminary dilution ratio1:170
Volume of the 1:170 dilution removed
(for making the RBC/Plt dilution)
Additional volume of A
Volume of A
Final dilution for Hgb determination1:250
Reaction temperature35°C (95°F)
Measurement Characteristics
Method of analysisSpectrophotometry
Wavelength550 nm
•T 5diff diluent
C
C
•T 5diff Hgb Lyse
•T 5diff diluent
.
1700 µL
42.5 µL
400 µL
400 µL
PN 4237616B
2.4-3
Page 64
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
WBC Count and Differential
The WBC count is determined twice using two different methodologies:
rThe reference WBC count is the count obtained in the WBC/BASO bath (Figure 2.4-2).
The WBC count and the BASO count are determined simultaneously.
rA second WBC count is determined in the flow cell during acquisition of the DiffPlot.
The dilution analyzed in the flow cell is prepared in the DIFF bath (Figure 2.4-2).
The WBC counts from the two methodologies are compared and if the results exceed the
predefined limits, they will be flagged.
Figure 2.4-2 Bath Assembly
2.4-4
Table 2.4-3 summarizes the technical characteristics required to obtain WBC and BASO
results.
Table 2.4-3 Characteristics Required to Obtain WBC and BASO Results
Method of analysisCoulter Principle
Aperture diameter80 µm
Count vacuum200 mb (5.9 in. Hg)
Count period2 x 6 seconds
C
•T 5diff WBC Lyse
2,000 µL
PN 4237616B
Page 65
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
Parameter Results Obtained from the WBC/BASO Dilution
The final 1:200 dilution is used to:
rDetermine the WBC count, and
rDevelop the WBC/BASO histogram, which is needed to obtain the BASO count.
Differential
Twenty-five microliters (25 µL) of whole blood is delivered to the DIFF bath in a flow of
C
A
•T 5diff Fix reagent, which lyses the red blood cells, stabilizes the WBC in their native
forms, and differentially stains the lymphocytes, monocytes, neutrophils, and eosinophils,
with the eosinophils staining most intensely.
The solution is then stabilized with diluent for three seconds and transferred to the
measuring bath. See Figure 2.4-3. Each cell is measured in absorbance (cytochemistry) and
resistivity (volume).
Figure 2.4-3 Flow Cell Operation
2) Second focused flow for optical detection
2
PN 4237616B
1) Primary focused flow for impedance
Table 2.4-4 summarizes the technical characteristics required for acquisition of the DiffPlot.
2.4-5
Page 66
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
Table 2.4-4 Technical Characteristics for Acquisition of the DiffPlot
Dilution Characteristics
Volume of whole-blood25 µL
Volume A
Volume A
Final dilution ratio1:80
Reaction temperature35°C (95°F)
Incubation duration12 seconds
Measurement Characteristics
Method of analysisImpedance with hydrofocus
Aperture diameter60 µm
Diameter of the flow 42 µm
Volume injected72 µL
C
•T 5diff Fix
C
•T 5diff Diluent
1000 µL
1000 µL
Injection duration15 seconds
Data accumulation12 seconds
Parameter Results Obtained from the DIFF Dilution
From these measurements, a DiffPlot is developed with optical transmission (absorbance) on
the X-axis and volume on the Y-axis. Figure 2.4-4 shows the DiffPlot regions.
From the DiffPlot, four out of five leukocyte (white blood cell) populations are determined:
lymphocytes, monocytes, neutrophils, and eosinophils.
In a typical whole-blood sample, the basophil population (determined in the WBC/BASO
bath) is very small compared to the other four white blood cell populations.
Figure 2.4-4 DiffPlot Regions
2.4-6
PN 4237616B
Page 67
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
Dilution Summary
Table 2.4-5 summarizes the dilution characteristics required to obtain CBC and CBC/DIFF
parameter results.
Table 2.4-5 Summary of Dilutions
2
Technical Characteristics
for
WBC Count and BASO count
(in the WBC/BASO bath)
Differential Acquisition with
Differential WBC Count
(in the DIFF bath)
Hemoglobin Measurement
(in the DIL1/HGB bath)
RBC and PLT Count
(in the RBC bath)
Note:
The primary dilution
(1:170) is made in the
DIL1/HGB bath.
Whole-Blood
VolumeReagent(s)
10 µL
25 µL
10 µL
42.5 µL of the
1:170 dilution
(from the
DIL1/HGB bath)
C
A
•T 5diff WBC Lyse
C
•T 5diff Fix
A
C
•T 5diff Diluent
A
C
A
•T 5diff Diluent
After removing 42.5 µL
of the 1:170 dilution:
C
A
•T 5diff Diluent
C
•T 5diff Hgb Lyse
A
C
•T 5diff Diluent
A
Reagent
VolumeDilution Ratio
2000 µLFinal
1:200
1000 µL
1000 µL
1700 µL
400 µL
400 µL
2500 µLSecondary
Final
1:80
Preliminary
1:170
Final
1:250
1:58.8
1:170 x 1:58.8 =
Final
1:10,000
Reaction
Temperature
35°C (95°F)
35°C (95°F)
35°C (95°F)
35°C (95°F)
PN 4237616B
2.4-7
Page 68
INSTRUMENT DESCRIPTION
SAMPLE ANALYSIS
2.4-8
PN 4237616B
Page 69
2.5RBC PARAMETER DEVELOPMENT
A
RBC/Plt Dilution
The final 1:10,000 dilution in the RBC bath contains red blood cells, white blood cells, and
platelets. Thresholds are used to separate the platelet pulses, which are much smaller, from
the red and white blood cell pulses. Since white blood cells fall in the red blood cell size
range, they are counted and sized as RBCs. The WBCs are not sorted out because any
interference is usually insignificant; there are normally very few WBCs (thousands) in
comparison to the number of RBCs (millions). Only when the white count is markedly
elevated is the red cell count or histogram influenced.
RBC Count
The AC•T 5diff hematology analyzer uses duplicate counting criteria, voting criteria, and
proprietary flagging information to confirm the parameter result prior to reporting it. To
obtain an RBC count result, the instrument compares the data from the two 5-second count
periods then votes and rejects any questionable data.
RBC count = Number of cells counted per volume unit x Calibration factor
INSTRUMENT DESCRIPTION
RBC PARAMETER DEVELOPMENT
2
6
The RBC count is displayed and printed as: RBC = N x 10
Cells per microliter (cells/µL) is the US reporting unit format. See Heading A.7 in the
Note:
Operator’s Guide for the other available formats.
cells /µL.
RBC Histogram
In addition to being counted, red blood cells are categorized according to size (from 30 fL to
300 fL) by a 256-channel pulse-height analyzer. The pulse-height analyzer uses a number of
thresholds to sort the particles into several size (volume) categories and to develop a size
distribution curve of the particles. The RBC distribution curve shows cells in their native size.
Figure 2.5-1 is an example of an RBC histogram with a normal RBC size distribution.
Figure 2.5-1 Typical RBC Histogram
30300
7616036
PN 4237616B
2.5-1
Page 70
INSTRUMENT DESCRIPTION
A
RBC PARAMETER DEVELOPMENT
Parameter Results Obtained Using the RBC Histogram
Hct Measurement
The height of the pulse generated by the passage of a cell through the aperture is directly
proportional to the volume of the analyzed red blood cell. The hematocrit (Hct) is the sum of
all the digitized pulses. The Hct is displayed and printed as a percentage (%).
Note: Percentage (%) is the US reporting unit format. See Heading A.7 in the Operator’s
Guide for the other available formats.
MCV Calculation
The MCV (Mean Cell Volume) is calculated using the Hct and the RBC count. The MCV is
displayed and printed in femtoliters (fL).
Note: Femtoliters (fL) is the US reporting unit format. See Heading A.7 in the Operator’s
Guide for the other available formats.
RDW Calculation
The RDW (Red cell Distribution Width) is an index of the variation or spread in the size of
the red blood cells. The study of the RBC distribution detects erythrocyte anomalies linked to
anisocytosis and enables the clinician to follow the evolution of the width of the curve
relative to the cell number and average volume. Displayed and printed as a percentage, RDW
is calculated using the standard deviation (SD) of the RBC population and the MCV.
K SD
---------------
RDW (%)=
MCV
where:
K = System constant
SD = Calculated standard deviation based on the red cell distribution
MCV = Mean Cell Volume of the red cells
RBC Distribution Flags
Once the RBC distribution curve is developed, two positions on the distribution curve are
located (Figure 2.5-2):
Figure 2.5-2 RBC1 and RBC2 Positions - RBC Histogram
RBC1RBC2
30300
7616057
2.5-2
PN 4237616B
Page 71
INSTRUMENT DESCRIPTION
RBC PARAMETER DEVELOPMENT
RBC1 and RBC2 Thresholds
Thresholds RBC1 and RBC2 define the
MICRO and MACRO regions and are calculated based
on standard deviation (SD) of the RBC population.
The RBC1 threshold (monitoring area for microcytes) and the RBC2 threshold (monitoring
area for macrocytes) identify the points on the curve that are ±2 SD from the mean
(Figure 2.5-2).
Flags
Note:
MICRO
and
MACRO
flags will be activated in software version 1.0 and higher.
2
MICRO
The
to the total number of RBCs exceeds the preset default limit of 5%. The
flag is generated when the percentage of cells in the microcytic region compared
MACRO
flag is
generated when the percentage of cells in the macrocytic region compared to the total number
of RBCs exceeds the preset default limit of 7.5%. A laboratory may establish its own limits to
replace the preset default values.
Note:
The
MICRO
and
MACRO
flags are independent of the
Microcytosis
Macrocytosis
and
flags that are generated from the Low and High patient limits.
Hgb Determination
The hemoglobin (Hgb) released by the lysis of the red blood cells combines with the
potassium cyanide to form a stable cyanmethemoglobin compound.
This compound is measured through the optical part of the DIL1/HGB bath using a
spectrophotometric technique at a wavelength of 550 nm. Transmittance of the sample
dilution is compared with the transmittance of a reagent blank. The system calculates the Hgb
using both the blank and sample readings.
The final Hgb result in g/dL represents: absorbance value obtained x Calibration factor.
Hgb is displayed and printed as: Hgb = N g/dL.
Grams per deciliter (g/dL) is the US reporting unit format. See Heading A.7 in the
Note:
Operator’s Guide for the other available formats.
PN 4237616B
Hgb Blank Reading
The Hgb blank value measured during the first patient cycle after a Startup cycle is stored as a
reference blank. This blank must be greater than 2.5 Vdc. During each analysis cycle, the
instrument checks the measured Hgb blank against the stored Hgb blank reference value
using the following formula:
(Blank
Ref
x 1/3) + (BlankS x 2/3) = Blank
NR
where:
Ref
Blank
Blank
Blank
Note:
= Hgb blank reference value
S
= Hgb blank value from the current cycle
NR
= New Hgb blank reference value for comparison
If the new Hgb blank reference value is within 3% of the old reference value,
the Hgb blank reference value is changed to this new value.
2.5-3
Page 72
INSTRUMENT DESCRIPTION
RBC PARAMETER DEVELOPMENT
Sample Reading
This value is based on the sample, diluent, and Hgb Lyse reagent mixture in the DIL1/HGB
bath during sample measurement.
Hgb Specific Flags
If the Hgb blank value is less than 2.5 Vdc, a reject (
If the difference between the new Hgb blank reference value and the original Hgb blank
reference value is greater than 3%, a review (
R
(
) flags occur on the Hgb blank reference value, the (
For each Hgb sample read value, the instrument takes three readings. If the difference
between these readings exceeds the predefined limits (default setting is 60 A to D units), a
V
voteout (
) flag is generated.
MCH and MCHC Calculations
MCH Calculation
The MCH (Mean Cell Hemoglobin) is calculated from the Hgb value and the RBC count and
describes the average weight of hemoglobin in a red cell. The calculation for MCH is:
R
) flag occurs on the Hgb value.
R
) flag is generated. If three consecutive review
. . . . .
) code replaces the Hgb result.
Hgb
-----------RBC
Picograms (pg) is the US reporting unit format. See Heading A.7 in the Operator’s
Note:
10 ×MCH (pg)=
Guide for the other available formats.
MCHC Calculation
The MCHC (Mean Cell Hemoglobin Concentration) is calculated using the Hgb and Hct
values and describes the average concentration of hemoglobin in the red blood cells. The
calculation for MCHC is:
Hgb
----------
Note:
100 ×MCHC (g/dL)=
Hct
Grams per deciliter (g/dL) is the US reporting unit format. See Heading A.7 in the
Operator’s Guide for the other available formats.
2.5-4
PN 4237616B
Page 73
2.6PLATELET PARAMETER DEVELOPMENT
RBC/Plt Dilution
Platelet counting and sizing is also done in the RBC bath. Thresholds separate the platelet
pulses, which are much smaller, from the red and white blood cell pulses.
Plt Count
The AC•T 5diff hematology analyzer uses duplicate counting criteria, voting criteria, and
proprietary flagging information to confirm the parameter result prior to reporting it. To
obtain a Plt count result, the instrument compares the data from the two 5-second count
periods then votes and rejects any questionable data.
Plt count = Number of cells counted per volume unit x Calibration factor.
INSTRUMENT DESCRIPTION
PLATELET PARAMETER DEVELOPMENT
2
The Plt count is displayed and printed as: Plt = N x 10
Cells per microliter (cells/µL) is the US reporting unit format. See Heading A.7 in the
Note:
Operator’s Guide for the other available formats.
3
cells /µL.
Platelet Distribution Curve
Platelets are categorized according to size by a 256-channel pulse-height analyzer. A
pulse-height analyzer uses a number of thresholds to sort the particles into several size
(volume) categories and to develop a size distribution curve of the particles between 2 fL and
30 fL. The Plt distribution curve shows cells in their native size. Figure 2.6-1 is an example of
a Plt histogram with a normal Plt size distribution.
Figure 2.6-1 Typical Plt Histogram
PN 4237616B
3
25fL
30
2.6-1
Page 74
INSTRUMENT DESCRIPTION
PLATELET PARAMETER DEVELOPMENT
Parameter Results Obtained Using the Plt Histogram
MPV Measurement
The MPV (Mean Platelet Volume) is measured directly from analysis of the platelet
distribution curve. The MPV is displayed and printed in femtoliters (fL).
Note: Femtoliters (fL) is the US reporting unit format. See Heading A.7 in the Operator’s
Guide for the other available formats
Pct Calculation
The Pct (plateletcrit or thrombocrit) is calculated according to the formula:
The Pct parameter result is displayed and printed as a percentage (%).
PDW Calculation
PDW (Platelet Distribution Width) is calculated from the Plt histogram as the width of the
curve between S1 and S2.
/µL()MPV (fL)×
10 000,
Pct%=
As shown in Figure 2.6-2, S1 and S2 are placed so that:
r15% of the platelets occur between 2 fL and S1.
r15% of the platelets occur between S2 and the variable upper threshold.
Note: This threshold is explained under the Detecting Abnormal Platelet Distributions
heading that follows.
rThe PDW result is determined on the platelets between S1 and S2.
Figure 2.6-2 Area of the Plt Histogram Used to Determine the PDW Parameter Result
15%
PDW
S1S2
15%
7615002A
2.6-2
The PDW parameter result is displayed and printed as a percentage (%).
PN 4237616B
Page 75
INSTRUMENT DESCRIPTION
PLATELET PARAMETER DEVELOPMENT
Detecting Abnormal Platelet Distributions
Particles of approximately platelet size can interfere with the platelet histogram and count.
Small particles, such as microbubbles or dust, can overlap the low end. Microcytic red cells
can intrude at the upper end.
Identifying a Normal Distribution
When a platelet histogram is being evaluated, a mobile threshold can move from its starting
position at 25 fL to 18 fL (Figure 2.6-3). The computer searches for a valley between the
platelet and red cell populations. If no valley is detected between 18 fL and 25 fL, the
threshold remains at 25 fL and no flag is generated.
Figure 2.6-3 Typical Platelet Distribution Curve
2
3
Interference on the Lower End of the Platelet Distribution Curve
Particles that are approximately platelet size can interfere with the platelet histogram and
count. Small particles, such as microbubbles or dust, can interfere at the low end. If the
number of pulses in the 2 to 3 fL region is higher than the predefined limits, an
appears to alert the operator that a significant number of small cells or interference, such as
microbubbles, are present.
Microcytic Interferences on the Upper End of the Platelet Distribution Curve
Microcytic red cells can intrude at the upper end of the platelet distribution curve. If the
specimen contains microcytes, the A
the influence of this interference by repositioning the variable threshold (25 fL threshold)
and excluding the microcytes.
C
25fL
•T 5diff analyzer may be able to successfully eliminate
30
SCL
flag
PN 4237616B
2.6-3
Page 76
INSTRUMENT DESCRIPTION
PLATELET PARAMETER DEVELOPMENT
Microcytic Interference with a Distinct Valley between 18 fL and 25 fL
If the intrusion of microcytes creates a valley between the 25 fL and the 18 fL thresholds
(Figure 2.6-4). The 25 fL threshold is repositioned at the valley to minimize interference to
the platelet parameter results. Therefore, the reported platelet results are acceptable. The
(microcytes) flag appears to alert the operator that microcytes are present.
Figure 2.6-4 Microcytic Interference with a Valley between 18 fL and 25 fL
MIC
2
18
25fL
30
Microcytic Interference with a Valley below 18 fL
If the microcytes are extremely small so that the valley between the platelet population and
the microcyte population falls below the 18 fL limit, the threshold is placed at the 18 fL limit
MIC
(Figure 2.6-5). The
flag appears and the platelet count is flagged to alert the operator that
the extremely small microcytes present in this sample could not be eliminated. The platelet
count and associated parameters are not reliable and should be verified by an alternative
method. To effectively eliminate the microcytes, the Operator’s Guide suggests the customer
use platelet rich plasma (PRP) or a manual count to verify the results.
Figure 2.6-5 Microcytic Interference with a Valley below 18 fL
2.6-4
3
18
25fL
30
PN 4237616B
Page 77
INSTRUMENT DESCRIPTION
PLATELET PARAMETER DEVELOPMENT
Interference with No Distinct Valley
Interference present in the upper area of the platelet distribution curve that blends with the
platelet population so that there is no clear distinction between the platelets and the
interference suggest the presence of schistocytes (fragmented red cells) or platelet aggregates
(platelet clumps).
If the threshold cannot be positioned in the 25 fL to 18 fL region, the threshold defaults to the
SCH
18 fL position (Figure 2.6-6). The
flagged to alert the operator that the interference (which is most likely either schistocytes or
platelet clumps) could not be eliminated. The platelet count and associated parameters are
not reliable and must be verified using an alternative method.
Figure 2.6-6 Interference with no Distinct Valley
(schistocytes) flag appears and the platelet count is
2
25fL18
2
30
PN 4237616B
2.6-5
Page 78
INSTRUMENT DESCRIPTION
PLATELET PARAMETER DEVELOPMENT
2.6-6
PN 4237616B
Page 79
2.7WBC PARAMETER DEVELOPMENT
Overview
WBC parameter results are generated from two different dilutions: the 1:200 WBC/BASO
dilution which is made and analyzed in the WBC/BASO bath and the 1:80 DIFF dilution
which is made in the DIFF bath but analyzed in the flow cell.
WBC/BASO Dilution
The WBC and basophil counts are determined from the 1:200 dilution made in the
WBC/BASO bath. To make this dilution, 10 µL of whole blood is mixed with 2,000 µL of
C
•T 5diff WBC Lyse reagent. The reaction that occurs lyses the red blood cells and
A
specifically differentiates between basophils and other leukocytes by volume.
WBC Count
The AC•T 5diff hematology analyzer uses duplicate counting criteria, voting criteria, and
proprietary flagging information to confirm the parameter result prior to reporting it. To
obtain an WBC count result, the instrument compares the data from the two 5-second count
periods then votes and rejects any questionable data. This is the reference WBC count, which
is also the count reported.
INSTRUMENT DESCRIPTION
WBC PARAMETER DEVELOPMENT
2
A second WBC count is determined in the flow cell during acquisition of the DiffPlot. The
two counts are compared and if they differ more than the predefined limit, a flag occurs.
WBC count: Number of cells per volume x calibration factor.
The WBC count is displayed and printed as: WBC = N x 10
Cells per microliter (cells/µL) is the US reporting unit format. See Heading A.7 in the
Note:
Operator’s Guide for the other available formats.
3
cells /µL.
BASO Count
Differentiation between basophils and other leukocytes is obtained by means of the
C
•T 5diff WBC Lyse-specific lytic action. See Figure 2.7-1.
A
Figure 2.7-1 Areas Used to Determine WBC and BASO Parameter Results
bcd
basophilsWBC
PN 4237616B
7615003A
In Figure 2.7-1, basophils are located in the area between the thresholds labeled candd.
One hundred percent (100%) of the leukocytes is represented by the total number of
nucleated particles plus the basophils within the area between the thresholds labeled
andd.
b
2.7-1
Page 80
INSTRUMENT DESCRIPTION
WBC PARAMETER DEVELOPMENT
The basophil percentage is calculated from the number of particles existing in the area
between the thresholds labeled
BASO count = Number of cells per volume x calibration factor in a percentage relative to the
number of counted cells (basophils plus other WBC nuclei).
BASO count
DIFF Dilution
The data for the DiffPlot is accumulated as the dilution made in the DIFF bath is injected into
the flow cell. To make the 1:80 DIFF dilution, 25 µL of the whole-blood sample is mixed with
C
1,000 µL of A
the white blood cells, and differentially stains the lymphocytes, monocytes, neutrophils, and
eosinophils, with the eosinophils staining most intensely. After 12 seconds of incubation,
1,000 µL of A
dilution is injected through the flow cell 15 seconds. For 12 of these 15 seconds, data for
developing the DiffPlot is accumulated.
•T 5diff Fix reagent. The Fix reagent lyses the red blood cells, stabilizes the
C
•T 5diff Diluent reagent is added to stop the cytochemical reaction. This
BASO%
---------------------WBC%
andd (Figure 2.7-1)
c
WBC count×=
DiffPlot Development
The DiffPlot analysis on the AC•T 5diff hematology analyzer is based on three essential
principles:
rDual Focused Flow (DFF) fluid dynamics, which is a process by which individual cells
or particles are focused in a stream of diluent (hydrodynamic focusing).
rThe volume measurement (Coulter Principle).
rThe measurement of transmitted light with zero degree (0°) angle, which permits a
response proportional to the internal structure of each cell and its absorbance.
From these measurements, a DiffPlot is developed with optical transmission (absorbance) on
the X-axis and volume on the Y-axis. Figure 2.7-2 shows the DiffPlot regions.
Figure 2.7-2 DiffPlot Regions
2.7-2
PN 4237616B
Page 81
INSTRUMENT DESCRIPTION
WBC PARAMETER DEVELOPMENT
DiffPlot Regions Defined
The study of the DiffPlot permits the clear differentiation of four out of five leukocyte
populations. In a typical whole-blood sample, the basophil population is very small when
compared with the other four white cell populations.
Neutrophil (Neut)
Neutrophils, with their cytoplasmic granules and segmented nuclei, scatter light according to
their morphological complexity. A hypersegmented neutrophil gives an increased optical
response when compared to a young neutrophil population. The higher the complexity of the
cell, the further to the right they appear in the DiffPlot (Figure 2.7-2).
Lymphocyte (Lymph)
Lymphocytes, typically being small with regular shape are smaller in volume and lower in
absorbance than the other cells, and are positioned in the lower region of the DiffPlot
(Figure 2.7-2). Normal lymphocyte populations typically have a homogeneous volume with a
Gaussian (bell-shaped) distribution.
Large lymphocytes, reactive lymphoid forms, stimulated lymphocytes and plasma cells are
found in the upper portion of the lymphocyte region (Figure 2.7-2).
2
The lower area of the lymphocyte zone is normally empty; however, when small lymphocytes
are present, a population may exist in this area (Figure 2.7-2).
The presence of platelet aggregates is indicated by a distribution pattern that moves from the
DiffPlot origin into the lymphocyte region (Figure 2.7-2).
NRBC cytoplasmic membranes lyse like those of mature erythrocytes. The small nuclei that
remain appear in the debris and small lymphocyte regions (Figure 2.7-2).
Monocyte (Mono)
Monocytes are typically large cells with a kidney-shaped nucleus and agranular (granule-free)
cytoplasm. These cells neither scatter nor absorb large amounts of light and, therefore, are
positioned in the lower end of the absorbance axis. Due to their size, the monocytes are
clearly positioned high on the volume axis (Figure 2.7-2).
Very large monocytes may be found in the IMM (immature cell) region.
Eosinophil (Eos)
With the reagent action, eosinophils are the most intensely stained cells for optical
separation. Due to the staining and their size, the eosinophils will show higher absorbance
than the neutrophils, but will be of similar volume (Figure 2.7-2).
Debris
Platelets and debris from erythrocyte lysis represent the background debris population
located in the lower region of the DiffPlot.
PN 4237616B
2.7-3
Page 82
INSTRUMENT DESCRIPTION
WBC PARAMETER DEVELOPMENT
Immature White Blood Cells
Immature Granulocytes
Immature granulocytes are detected by their larger volume and by the presence of granules
that increase the intensity of the scattered light.
Due to their increased volume and similar absorbance, promyelocytes, myelocytes, and
metamyelocytes are located above the neutrophil population and are typically counted as
IMM cells. IMM cells are included in the reported neutrophil value.
Band Cells
Band cells are typically larger or of similar size to the neutrophils; however, due to their low
level of cellular complexity, they absorb less light. As a result, band cells tend to appear in the
region between the neutrophils and the monocytes.
Blast Cells
Blast cells are generally larger than monocytes and have similar absorbance. When blast cells
are present, they are generally located above the monocytes, which means they will be
included in the IMM cell count.
Small blasts will be located between the normal lymphocyte and monocyte populations.
DiffPlot Thresholds
Most of the population partition thresholds are fixed and give the limits of the morphological
normality of leukocytes. Changes in the morphology of a population are expressed on the
DiffPlot by a shifting of the corresponding population. Volume and absorbance thresholds are
used to detect shifting populations. Volume thresholds and definitions are shown in
Figure 2.7-3. Absorbance thresholds and definitions are shown in Figure 2.7-4. The NL, NE
and MN alarms are also included in Figure 2.7-4.
2.7-4
PN 4237616B
Page 83
Figure 2.7-3
Volume Thresholds
INSTRUMENT DESCRIPTION
WBC PARAMETER DEVELOPMENT
2
DL Threshold
lymphocytes.
DN Threshold
neutrophils.
SL Threshold
lymphocytes.
LN Threshold
neutrophils.
DE Threshold
LMN Threshold
lymphocyte, monocyte, and neutrophil
thresholds.
AL Threshold
atypical lymphocytes.
LMU Threshold
slope between atypical lymphocytes and
monocytes.
LMD Threshold
slope between atypical lymphocytes and
monocytes.
- Separates debris and small
- Separates debris and lower
- Separates small lymphocytes and
- Separates neutrophils and lower
- Separates debris and eosinophils.
- Intersection point between the
- Separates lymphocytes and
- Lower point on the separation
- Upper point on the separation
MN Threshold
slope between monocytes and neutrophils.
UM Threshold
monocytes.
UN Threshold
neutrophils.
- Upper point on the separation
- Separates monocytes and upper
- Separates neutrophils and upper
PN 4237616B
2.7-5
Page 84
INSTRUMENT DESCRIPTION
WBC PARAMETER DEVELOPMENT
Figure 2.7-4
Absorbance Thresholds / NL, NE and MN Alarms
NL Threshold
- Separates lymphocytes and
neutrophils.
RMN Threshold
- Separates upper monocytes and
upper neutrophils.
NE Threshold
- Separates neutrophils and
eosinophils.
NL, NE and MN Alarms
FNL
- # of channels for NL alarm area.
FNE
- # of channels for NE alarm area.
FMN
- # of channels for MN alarm area.
2.7-6
PN 4237616B
Page 85
2.8PNEUMATIC/HYDRAULIC SYSTEM
Functions of Valves
Valve blocks are located close to the elements concerned. Five different blocks:
rValves 1 to 11 (Figure 2.8-1):
tIn the left side compartment behind the Main card,
tHorizontal block,
tAbove the 5diff syringe and reagent syringes assembly.
rValves 12 to 16 (Figure 2.8-1):
tIn the left side compartment behind the Main card,
tVertical block,
tBeside the count syringe.
Figure 2.8-1 Valve 1 through Valve 16 Locations
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
2
PN 4237616B
rValves 17 to 18 (Figure 2.8-2):
tIn the right side compartment,
tHorizontal block,
tAt the top of the vertical traverse (above the sample syringe assembly).
Figure 2.8-2 Valve 17 and 18 Location
2.8-1
Page 86
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
rValve 19 is not used on the A
rValves 20 to 26 (Figure 2.8-3):
tIn the right side compartment,
tVertical block,
tBeside the waste syringe.
rValves 27 to 31 (Figure 2.8-3):
tIn the right side compartment (bath enclosure area),
tHorizontal block,
tBelow the baths assembly.
Figure 2.8-3 Valve 20 to Valve 31 Locations
C
•T 5diff hematology analyzer.
See Table 2.8-1 for a description of the functions for each valve in the AC•T 5diff hematology
analyzer.
2.8-2
PN 4237616B
Page 87
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Table 2.8-1 Valves and their Functions
ValveFunctionAction
1Differential diluentSelect flow cell sheath 2 / DIFF bath
2Differential diluentSelect flow cell sheath 1 / sheath 2
3Differential diluentSelect input/output for flow cell Diluent syringe
4Flow cell sample supplyOpens pathway from the DIFF bath to the flow cell
5Flow cell sample injectorOpens waste path for sample injector syringe
6Hgb Lyse syringe flowSelects input/output of Hgb Lyse syringe
7Rinse syringe flowSelects input/output of Rinse syringe
8Fix syringe flowSelects input/output of Fix syringe
9Diluent syringe flowSelects input/output of Diluent syringe
10Diluent output controlRoutes diluent to probe rinse block or heating coil
11WBC Lyse syringe flowSelects input/output of WBC Lyse syringe
12Rinse output controlSelects rinse to probe rinse block or WBC/BASO bath
2
13Count syringe ventOpens vent line of count syringe
14RBC/PLT count valveOpens vacuum count line for RBC bath
15Diluent reservoir ventSelects between vacuum and vent for diluent reservoir
16Count syringe drainOpens count syringe drain path
17Probe rinse drainOpens drain line for probe rinse block
18Probe diluentRoutes diluent/rinse to sample syringe or rinse block
19SpareNot used
20Waste syringe ventOpens waste vent (through the rinse chamber)
21Sweep flow diluentRoutes diluent to heating coil or sweep flow
22Diluent bath selectRoute diluent (via heating coil) to Hgb or RBC bath
23WBC/BASO count vacuumRoutes vacuum direct or through RBC/PLT count head
24Flow cell drainOpens path from flow cell output to DIFF bath for drain
25Diluent reservoir inputOpens diluent source to diluent reservoir
26Waste syringe controlSelects waste to syringe / syringe waste out
27Rinse chamber drainOpens drain path from rinse chamber
28HGB bath drainOpens drain path from Hgb bath
29DIFF bath drainOpens drain path from the DIFF bath
30RBC bath drainOpens drain path from the RBC bath
PN 4237616B
31WBC/BASO bath drainOpens drain path from WBC/BASO bath drain
2.8-3
Page 88
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Pneumatic Diagrams
To locate the pneumatic diagram for a desired reagent or the waste circuit, see the designated
figure:
rFor the Hgb Lyse reagent circuit, see Figure 2.8-4.
rFor the Fix reagent circuit, see Figure 2.8-5.
rFor the WBC Lyse reagent circuit, see Figure 2.8-6.
rFor the Rinse reagent supply circuit, see Figure 2.8-7.
rFor the Probe Rinse reagent circuit, see Figure 2.8-8.
rFor the WBC/BASO Rinse reagent circuit, see Figure 2.8-9.
rFor the Diluent reagent circuit, see Figure 2.8-10.
rFor the Probe Diluent reagent circuit, see Figure 2.8-11.
rFor the Bath Diluent reagent circuit, see Figure 2.8-12.
rFor the waste circuit, see Figure 2.8-13.
2.8-4
PN 4237616B
Page 89
Figure 2.8-4 Hgb Lyse Reagent Circuit
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
2
PN 4237616B
2.8-5
Page 90
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Figure 2.8-5 Fix Reagent Circuit
2.8-6
PN 4237616B
Page 91
Figure 2.8-6 WBC Lyse Reagent Circuit
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
2
PN 4237616B
2.8-7
Page 92
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Figure 2.8-7 Rinse Reagent Supply Circuit
2.8-8
PN 4237616B
Page 93
Figure 2.8-8 Probe Rinse Reagent Circuit
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
2
PN 4237616B
2.8-9
Page 94
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Figure 2.8-9 WBC/BASO Rinse Reagent Circuit
2.8-10
PN 4237616B
Page 95
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Diluter System
Figure 2.8-10)
Diluent Input
Diluent for the A
from a diluent container and is stored in the diluent reservoir. The input tubing from the
diluent container should be no longer than 2 meters (78.7 in.) and the top of the container
cannot be lower than 80 cm (31.5 in.) from the input fitting on the instrument. Vacuum,
produced by the count syringe, is used to draw diluent into the diluent reservoir from the
diluent container. This vacuum is applied to port 2 of the diluent reservoir. Solenoid valve 15,
Diluent Reservoir Vent, is activated to connect the diluent reservoir to the count syringe.
When in the normal inactive state, this valve vents the diluent reservoir to atmosphere. The
vent tubing from port 1 of valve 15 is notched to ensure it does not seal against any surface
and is routed to the left side drip tray. A float sensor located in the reservoir is used to
determine when the reservoir is full, or needs more diluent.
Figure 2.8-10 Diluent Reagent Circuit
(
C
•T 5diff hemtology analyzer enters a fitting at the rear of the instrument
2
PN 4237616B
2.8-11
Page 96
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Two ports distribute diluent from the diluent reservoir. Port 1 supplies diluent to be used by
the flow cell while port 3 supplies diluent for the baths and aspirate probe. Solenoid valve 3
normally connects port 1 of the diluent reservoir to the center syringe piston in the 5diff
syringe assembly, allowing the syringe to fill with diluent. For diluent output from the 5diff
syringe see 5diff Syringe and Flow Cell.
Solenoid valve 9 connects the diluent reagent syringe to port 3 of the diluent reservoir, and
when it is in a normal inactive state, this path is open, allowing the syringe to be filled. For
output from the main diluent syringe, see Diluent to Baths and Probe and Probe Rinse.
5diff Syringe and Flow Cell
Diluent for several flow cell requirements originates at the center syringe piston of the 5diff
syringe assembly. When solenoid valve 3 is active, valves 1 and 2 have control of the diluent
path. If valve 2 is energized, diluent is sent from the large center syringe piston to (and
through) the left injector piston. A short upward stroke is used to fill the small left injector
piston with clean diluent. Since the center syringe piston is much larger, even a short stroke
will produce excess diluent, which exits out the top of the left injector piston and through the
flow cell.
When valve 2 is in the normal state, valve 1 determines where diluent is routed. An energized
valve 1 allows diluent to be sent through the heater assembly to port 2 of the DIFF bath. This
is used for the second part of the DIFF dilution when 1 mL of diluent is added to the sample
12 seconds after the original dilution with Fix. This stops the staining action of the Fix.
Solenoid valve 1 in its normal inactive state creates a diluent path for the flow cell outer
sheath. During normal flow cell operation, the center syringe piston (solenoid valve 3
energized and solenoid valves 1 and 2 in a normal inactive state) sends diluent through an
electrical isolator and T-fitting to ports 2 and 4 of the flow cell. This creates an outer sheath or
fluid pipe around the stream of fluid exiting the flow cell aperture. This sheath is
approximately 140 µm in diameter.
During the upward stroke of normal flow cell operation, the 5diff syringe assembly creates the
inner sheath flow. Diluent exits from the top of the left injector piston and enters the flow cell
at port 5. It then forms the inner sheath, creating a fluid pipe around the injected sample. The
sample injector creates sample flow of 40 µm in diameter, while the sample and inner sheath
is forced through the aperture, which is 60 µm in diameter. On exiting the aperture, an outer
sheath is created and this double sheath around the sample is called Dual Focused Flow, or
DFF.
Sample is sent through the flow cell with the right injector piston (see Figure 2.8-13), but it
first must be positioned. The sample dilution, created in the DIFF bath, is drained with the
waste syringe. The vacuum path to the waste syringe, starting at the bath, is through fluid
sensor M1, an energized solenoid valve 4, through the right injector piston (note that the
piston is not being used at this time), through an energized solenoid valve 5, T-fitting T6,
waste fluid sensor M2, a fluidic/electrical isolator, and solenoid valve 26 in its normal inactive
state. Fluid is drained from the bath until sensor M1 detects air. The length and size of the
tubing ensures that the sample does not actually reach or enter the right injector piston when
sensor M1 detects air and stops flow. This is critical. When the injector piston pushes up,
sample in the tubing between T-fitting T2 and port 5 of the right injector piston is sent
through the flow cell.
2.8-12
PN 4237616B
Page 97
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Probe and Probe Rinse
When valves 9 and 10 are both energized (Figure 2.8-11), diluent output from the diluent
syringe is sent to the probe and probe rinse block by way of solenoid valve 18. An energized
solenoid 18 routes diluent through the sample syringe and out the probe for backwash and
sometimes dilution. Note that the sample syringe is not used at this time. Since the piston seal
is an O-ring at the base of the piston, rather than a seal at the top of the piston, the bottom
fitting and top fitting on the syringe assembly have an open fluid path. Diluent is sent
through an inactive sample syringe by the diluent syringe. The sample syringe is only used
when aspirating or dispensing sample.
Figure 2.8-11 Probe Diluent Reagent Circuit
2
PN 4237616B
With solenoid valve 18 in its normal inactive state, the diluent is sent to fitting 1 on the probe
rinse block. This fluid exits the rinse block from fitting 2 and passes through solenoid valve
17 to the waste system. Usually, this occurs while the probe is moving up through the rinse
block to clean blood off the exterior of the probe. However, this is not always the case. On
occasion, solenoid 17 is not opened, and the probe does not move. Diluent is forced to flow
down the outside of the probe. This is done once to clean the exterior of the probe without
moving it and another time to add a small amount of diluent to a dilution.
Diluent to Baths
Energizing solenoid valve 9 routes the diluent syringe output through solenoid valve 10
(Figure 2.8-12). In the normal state of valve 10, diluent is routed for use at the sample baths.
There are 3 uses of diluent at the baths, diluent for dilution in the Hgb bath, diluent for
dilution in the RBC bath, and rinse for the WBC/BASO counting head.
2.8-13
Page 98
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Solenoid valve 21 routes diluent for dilutions through valve 22 when in a normal inactive
state. Valve 22 selects the RBC bath (normal, inactive state) or the Hgb bath (energized state).
The fluid paths to the baths both go through a heater block. Each path has a 1 mL coil of fluid
in the heater block which allows 1 mL of reagent to be heated to 35°C prior to being delivered
to the bath.
Figure 2.8-12 Bath Diluent Reagent Circuit
2.8-14
When solenoid valve 21 is energized, fluid is sent to the WBC/BASO counting head to rinse
the pathway and counting head. This is necessary since fluid in this area is used to flush cells
from the rear of the RBC/Plt aperture, a technique called the Rinse Flow System (RFS).
During the RBC/Plt count, the WBC/BASO count head is connected to the RBC/Plt count
head through an inactive (normal state) valve 23. The vacuum applied to the RBC/Plt
aperture actually draws fluid (rinse solution, not diluent) from the WBC/BASO bath, through
the WBC/BASO aperture, and past the rear of the RBC/Plt aperture, sweeping away any RBC
cells from the rear of the aperture. When solenoid valve 23 is energized, the WBC/BASO
counting head is connected directly to the counting syringe, providing vacuum for the
WBC/BASO count.
PN 4237616B
Page 99
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
Waste System
The waste and drain system (Figure 2.8-13) comprises many components. The waste syringe
itself is used to drain the baths, and to expel waste from the baths. The count syringe expels
any waste that it accumulates during count directly into the waste system through normally
closed solenoid valve 16. Waste from the probe rinse block is pushed out by the diluent
syringe as well as being evacuated with the waste syringe.
The bath drain system connects each bath, including the rinse chamber, through a normally
closed solenoid valve to the waste system with a series of T-fittings. There is a fluid isolator
chamber between the baths and the waste syringe and container. This electrically isolates the
baths (and flow cell and aspirate probe) from any interference that can be picked up by the
external waste system. Draining waste from any bath involves opening the associated
solenoid valve while the drain syringe is filling. Solenoid valve 26 is then energized, which
connects the waste syringe to the external waste system, and waste is expelled.
There is also a fluid sensor, M2, in the waste system, just before the isolator. It is used to
detect whether the waste system has fluid or air at the appropriate times. It is not used to stop
a drain action, like M1 does when detecting that sample has been drained from the DIFF
bath.
2
Figure 2.8-13 Waste Circuit
PN 4237616B
2.8-15
Page 100
INSTRUMENT DESCRIPTION
PNEUMATIC/HYDRAULIC SYSTEM
2.8-16
PN 4237616B
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.