Amsco Reliance 444 HTM User manual

0 (0)
UNCRATING/INSTALLATION INSTRUCTIONS
AMSCO
®
RELIANCE
®
444 HTM
SINGLE-CHAMBER WASHER/DISINFECTOR
(02/28/02) P-920012-041
Rev. 0
IN NO CASE, WHETHER AS A RESULT OF A BREACH OF CONTRACT, BREACH OF WARRANTY,
OR TORT (INCLUDING STERIS’S OR CUSTOMER’S WILLFUL ACTS, NEGLIGENCE, OR STRICT
LIABILITY) SHALL STERIS OR CUSTOMER BE LIABLE TO THE OTHER FOR ANY CONSEQUEN-
TIAL OR INCIDENTAL DAMAGES INCURRED BY THE OTHER, INCLUDING, BUT NOT LIMITED
TO, LOSS OF REVENUE, PROFITS, OR GOODWILL. HOWEVER, NOTHING CONTAINED IN THIS
MANUAL IS INTENDED TO RELIEVE EITHER STERIS OR CUSTOMER FROM CLAIMS, LIABILITY,
DAMAGES, OR EXPENSES RESULTING FROM BODILY INJURY, INCLUDING DEATH, OR
FROM PROPERTY DAMAGE INCURRED DUE TO THE WILLFUL ACTS, NEGLIGENCE, OR
STRICT LIABILITY OF THAT PARTY.
STERIS AGREES TO DEFEND, INDEMNIFY, AND HOLD CUSTOMER HARMLESS FROM ANY
AND ALL CLAIMS, LIABILITY, DAMAGES, OR EXPENSES DUE TO PERSONAL INJURIES,
INCLUDING DEATH, TO EMPLOYEES OF STERIS AND CUSTOMER AND TO THIRD PARTIES
AND FOR PROPERTY DAMAGE TO THE EXTENT CAUSED BY THE WILLFUL ACTS, NEGLI-
GENCE, OR STRICT LIABILITY OF STERIS . CUSTOMER AGREES TO DEFEND, INDEMNIFY,
AND HOLD STERIS HARMLESS FROM ANY AND ALL CLAIMS, LIABILITY, DAMAGES, OR
EXPENSES DUE TO PERSONAL INJURIES, INCLUDING DEATH, TO EMPLOYEES OF CUS-
TOMER AND STERIS AND TO THIRD PARTIES AND FOR PROPERTY DAMAGE TO THE EXTENT
CAUSED BY THE WILLFUL ACTS, NEGLIGENCE, OR STRICT LIABILITY OF CUSTOMER.
WARNING - COPYING PROHIBITED
This manual is protected by Federal Copyright Law, which
provides for damages of up to USD $20000, as well as
criminal fines and imprisonment, for unauthorized copying.
i
Uncrating/Installation Instructions P-920012-041
© 2002, STERIS Corporation. All rights reserved. Printed in Canada
A WORD FROM STERIS CORPORATION
Follow each step of the uncrating/installation instructions in the
order presented. Open the carton carefully to avoid damage to the
equipment inside. If you find any indication of damage to the
equipment (no matter how slight), show it to your supervisor.
To properly install this unit, you will need the Equipment Drawings
(previously furnished), showing all utility service and space require-
ments. If drawings cannot be located, replacement copies may be
obtained by writing, faxing, or telephoning STERIS, giving the serial
and model numbers of your equipment.
Once installed, equipment operation should be tested by a qualified
service technician prior to your usage of the equipment.
If STERIS supervision is desired, for installing and starting up this
equipment, contact your local STERIS representative.
Advisory
Introduction
IMPORTANT: A listing of Safety Warnings and Cautions to be
observed when uncrating, installing, and operating this washer can
be found on Page 1-1 of these instructions. Do not begin uncrating/
installing this equipment until you have become familiar with this
information.
IMPORTANT: Be sure to check local occupational health and
safety regulations, as well as electric and plumbing codes, for any
special requirements that may pertain to the installation of this unit.
Any alteration of the equipment not authorized by STERIS Engineer-
ing Service which could affect its operation will void the warranty,
could adversely affect washing efficacy, could violate federal,
state, and local regulations, and could jeopardize your insurance
coverage.
STERIS supervision is recommended for installing and starting up
this equipment. Contact the STERIS Regional Office nearest you for
availability information on this service.
Indications for Use
The Amsco
®
Reliance
®
444 HTM Single-Chamber Washer/
Disinfector is indicated for use in the cleaning and low-level
disinfection of reusable utensils, trays, glassware, bedpans and
urinals, rubber and plastic goods, simple hard-surfaced rigid
surgical instruments, such as forceps and clamps, and other similar
and related articles found in healthcare facilities.
ii
P-920012-041 Uncrating/Installation Instructions
Introduction
A thorough preventive maintenance program is essential to safe
and proper operation of this equipment. You are encouraged to
contact your STERIS representative concerning extended service
maintenance agreements to give your washer planned mainte-
nance, assuring equipment performance according to factory
specifications.
A global network of skilled service specialists can provide periodic
inspections and adjustments to assure low-cost peak performance.
STERIS representatives can provide information regarding Annual
Maintenance Agreements.
STERIS carries a complete line of accessories for use in this
equipment. Your STERIS representative will gladly review these
with you.
Sales and Service Information: Technical Assistance:
STERIS Corporation STERIS Engineering Services
5960 Heisley Road 2424 West 23rd Street
Mentor, Ohio 44060 Erie, PA 16506
Tel.: 440 354 2600 Tel.: 814 452 3100
Fax: 440 639 8199 Fax: 814 870 8400
Manufacturing Plant: Authorized EU Representative
STERIS Canada Corporation STERIS Limited
415, rue des Alleghanys STERIS House
Beauport (Québec) Jays Close
Canada G1C 4N4 Viables
Tel.: 418 664 1549 Basingstoke
Fax: 418 664 0188 Hampshire RG22 4AX
United Kingdom
Tel.: 44 1256 840400
Fax: 44 1256 866502
STERIS Offices Worldwide:
Canada 800 661 3937
Germany 49 22 33 69990
Hong Kong 852 2 563 3623
Italy 39 01 41 590429
Japan 81 78 321 2271
Korea 82 2 517 1517
Latin America 305 442 8202
Singapore 65 841 7677
Spain 34 91 658 5920
United Kingdom 44 1256 840400
Web Site: www.steris.com
Service Information
iii
Uncrating/Installation Instructions P-920012-041
This Amsco
®
Reliance
®
444 HTM Single-Chamber Washer/Dis-
infector complies with the following regulations:
Governing Directive for the affixing of the CE mark:
Medical Device Directive (93/42/EEC), Annex II.
Standards applied to demonstrate conformity to the directives:
IEC-61010-1, IEC-61010-2-045, IEC-61326-1.
Compliant with the requirements mentioned in the guidance
document HTM 2030.
Introduction
Certification
iv
P-920012-041 Uncrating/Installation Instructions
This Page Intentionally Left Blank
v
Uncrating/Installation Instructions P-920012-041
1 LISTING OF WARNINGS, CAUTIONS,
AND SYMBOLS............................................................. 1-1
Symbols ..........................................................................1-3
2 UNCRATING INSTRUCTIONS...................................... 2-1
Open Crate .....................................................................2-1
Move Equipment and Remove Skid ................................2-5
Prepare Unit for Installation .............................................2-6
Manual Opening of Door (without Electric
and Air Supplies) .........................................................2-7
Manual Opening of Door (with Air Supply) ...................2-7
3 INSTALLATION INSTRUCTIONS ................................. 3-1
Technical Specifications .................................................3-1
Voltage, Amperage, and Power Consumption .............3-1
Permissible Environmental Conditions .........................3-1
Noise Level ..................................................................3-1
Installation Requirements ................................................3-2
Connect Utilities .............................................................. 3-5
Cleanup...........................................................................3-6
4 INSTALLATION CHECKLIST ....................................... 4-1
5 OPERATIONAL TEST ................................................... 5-1
Test Procedure................................................................5-1
Dip Switches Settings ..................................................5-5
Start-up Procedure for Independent Monitor...................5-7
SECTION TITLE PAGE
TABLE OF CONTENTS

Table of Contents

vi
P-920012-041 Uncrating/Installation Instructions
This Page Intentionally Left Blank
1-1
Listing of Warnings, Cautions and Symbols Uncrating/Installation Instructions P-920012-041
The following is a listing of the safety precautions which must be observed when uncrating and installing equipment.
WARNING indicates the potential for danger to personnel and CAUTION indicates the potential for damage to
equipment. These precautions are repeated, where applicable, throughout the instructions.
WARNING - PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
When moving the unit, use a forklift. Do not use door(s) or door handle(s) for lifting.
WARNING - LACERATION AND/OR EYE INJURY HAZARD:
When removing bands, wear gloves and eye protection, and always use a tool specifically designed
to cut the bands. The bands used to secure this carton can cause personal injury when cut and
tension is released.
WARNING - LACERATION HAZARD:
When removing bolts, wear gloves to protect your hands.
WARNING - PERSONAL INJURY HAZARD:
The chamber door is heavy. Lifting it manually may require two people.
Door lifts at high speed; keep your head, hands, or fingers away from door opening.
WARNING - BURN HAZARD:
When checking the automatic stop feature, wear gloves and face protection and open chamber door
slowly. Hot water/steam may be sprayed through door opening if door is opened during a cycle.
Except for emergency, do not open door when cycle is in progress. In an emergency, first stop cycle
by pressing the STOP/RESET touch pad and wait for water flow to stop. Wear protective gloves and
face shield whenever reaching into chamber.
WARNING - FALL HAZARD:
To prevent falls, keep floors dry. Promptly clean up any spills or drippage.
WARNING - ELECTRICAL SHOCK HAZARD:
Fasteners and star washers are used to ensure protective bonding continuity. Always reinstall any
star washer which may have been removed during installation or servicing.
(See next page for additional Warnings and Cautions)
LISTING OF WARNINGS, CAUTIONS, AND
SYMBOLS
1
1-2
P-920012-041 Uncrating/Installation Instructions Listing of Warnings, Cautions and Symbols
WARNING - CHEMICAL BURN AND/OR EYE INJURY HAZARD:
Risk of explosion: Do not remove from pressurized lines. Do not exceed maximum temperature/
pressure specifications. Do not install/service without following installation instructions (see
sensor manual). Wear safety goggles and face shield during installation/service. Do not alter
product construction. Failure to follow safety instructions could result in severe personal injury.
WARNING - HEALTH HAZARD:
Vapors from solvents can be harmful. Use with adequate ventilation. Follow directions on the
container.
CAUTION - POSSIBLE EQUIPMENT DAMAGE:
After utilities are connected to equipment, slowly remove the protective adhesive paper from the
exterior cabinet panels to reduce the level of static discharge.
When removing adhesives from stainless steel, use a small amount of non-flammable cleaning
solvent. Rub in a back-and-forth motion (in same direction as surface grain). Solvent rubbed in a
circular motion or applied with a wire brush or steel wool on door and chamber assemblies can be
harmful to stainless steel. Do not use solvents on painted surfaces.
Once three-phase power is connected, check pump for correct rotation (indicated by arrow on pump
motor). Incorrect pump rotation may result in pump damage and improper cleaning action.
Before operating unit, always position each manifold or accessory header over a manifold
connector. If manifolds or accessory headers are not positioned correctly, damage may result and
unit will be unable to effectively wash loads.
1-3
Listing of Warnings, Cautions and Symbols Uncrating/Installation Instructions P-920012-041

Symbols

Symbol Definition
Symbols on the carte
Tip'N Tell Indicator
"If TIP'N TELL arrow point is blue, this package
has been on its side or tipped over in transit.
Make note on the bill of lading and check for
damage. Any claims for dipping on this notation."
Maximum Shipping and Storing
Temperature
Maximum Humidity Level
This Side Up
Do Not Stack
Fragile
Keep Dry
Information on the nameplate
MODEL Model number of the unit
S/N Serial number of the unit
kVA Kilovolt-Ampere
V_~ Volt, Number of phase (3 or 1 [blank]),
Alternate Current
A Ampere
YEAR Year of manufacture of the unit
Hz Hertz — Frequency of the unit
Loading...