ALTANA Z7, Z10 User Manual

Measure what you see.
A5-A1 / A-T / U-Z6 / Z7-Z10
Bubble Viscometers
A member of
Manual
Additives & Instruments
Bubble Viscometers A5-A1 / A-T / U-Z6 / Z7-Z10
Manual
English
PV-0500 Series PV-0600 Series
BYK - Gardner USA
9104 Guilford Road Columbia, MD 21046 USA Phone 800-343-7721 301-483-6500 Fax 800-394-8215 301-483-6555
www.byk.com/instruments
BYK-Gardner GmbH
Lausitzer Str. 8 D-82538 Geretsried Germany Tel. 0-800-gardner (0-800-4273637) +49-8171-3493-0 Fax +49-8171-3493-140
Directions for Using the Gardner Bubble Viscometer
DIRECTIONS FOR USING THE GARDNER BUBBLE VISCOMETER A-5 to A-1 CAT. NO. PV-0500
This instrument is based on the principle that the viscosity is directly proportional to the bubble speed. The viscosity of a sample of varnish or other transparent liquid is therefore determined by reference to the standard tubes in which an air bubble rises with the same speed as it does in a tube of the sample being tested. For this purpose two carefully calibrated empty tubes are furnished with each set.
The bubble size of the sample is rst adjusted as
nearly as possible to that in the standard tubes. The tubes are then brought to a temperature of 25ºC (77ºF). This is essential due to the difference
in the temperature coefcient of viscosity of
mineral oils and varnishes. It is suggested that a water bath be used, which quickly brings the tubes to the proper temperature and insures greater accuracy.
The bubble should be started from the at end
of the tube held in a true vertical position. The bottom meniscus of the bubble should be used for comparing the samples. It is only necessary that they start together. Owing to unavoidable differences in bubble sizes and in the shapes of the ends of the tubes, it may be necessary to raise one of the tubes so that the points of comparison will be on the same level. The observation is made after the bubbles are completely formes and moving upwards along a true vertical line. This takes place about 1½ to 2 cm from the bottom of the tube.
The standard empty grade “A” tubes have been carefully selected to and I.D. of 10.65 ± 0.0025 mm. and should be used for exact determinations. However, this close tolerance is relatively expensive, and for the sake of economy, tubes
2
Directions for Using the Gardner Bubble Viscometer
having the same approximate diameter with a tolerance of ± 0.150 mm. Grade “B” are used by many manufacturer for general factory control tests.
This viscometer meets the requirements of ASTM D 154, D 555, D 1131, D 1545, D 1725, FTMS 14la-Method 4271, and AOCS Method Ka6.
Approximate Viscosity in Stokes at 25º C (77º F)
A5 0.051
A4 0.071
A3 0.140
A2 0.213
A1 0.310
3
Loading...
+ 8 hidden pages