Alaris MedSystem III 2860, MedSystem III 2863 User manual

..
MedSystem
III™
For
most
Troubleshooting
errors.
Tip
Sheet
Error
Codes
(6,3):
(16,X):
fere
SO
(31,3):
1.
Check
disconnected.
2.
Use a known
to
Air-in-line
Air-in-line
1.
Clean
2.
Replace
1.
Reseat
2.
Check
3.
Check
4.
Replace
5.
Replace
Opto
Top Fuse. Battery.
Troubleshooting
see
if
any
Air-in-Line
good
Air-in-Line
sensors
sensors*.
Module
Opto
are
defective.
and
Module*.
Board.
Power
MEA*.
Supply*.
do
CID/LIP
Steps
Sensors
Replace
have
become
to
find
out
which
defective
calibration.
Xz
(33,3)
(40,X):
(95,X):
Chana
O-A 3
(259):
(263):
A
1.
Reseat
2.
Replace
1.
2.
Replace
1.
Check
Top
Top
to
disconnected.
2,
Replace causes
1.
Clean
2.
Switch
with
the
channel
drive
Drive
channel
4.
Do
Pressure
Giso
Check
Board
MEA*.
Power
see
the
Supply",
if
any
Air-in-Line
pump
and
module
Module
replace
MEA*.
calibration.
Lihumim
and
EPOM.
Air-in-Line
Sensor*
to
go
into
do
FSOD
with
another
replace
Drive
bet
Sensors
on
whichever
alarm
when
calibration.
channel.
Module”.
have
become
channel
started.
if
error
If
error
moves
stays
with
in
same
(272):
1.
Do
CID/LIP
calibration.
(278):
(276);
(289)/(290)
(292)
uN
1.
2.
1.
1.
2.
1.
(293)
ON
Clean
with channel
Do
FSOD
Clean
Do
FSOD
with
channel
Reset
1.
Reset
and
drive
Drive
replace
Encoder
drive
Drive
replace
faults.
faults.
.
Lubricate
Module
channel,
with
replace
MEA*.
calibration.
Disk.
Make
calibration.
with
replace
MEA*.
do
another
Drive
sure
another
Drive
FSOD
channel.
calibration.
Module.
dust
cover
channel.
Module”.
0000
If
error
If
error
is
properly
If
error
If
error
moves
stays
in
installed.
moves
stays
with
in
same
with
same
(301)/(24)
(302)/(47)
1.
.
>
.
.
.
BOND
(305)/(88)
(303)/(67)
1.
2.
(311)
Replace
Check
Replace
Replace
Replace
.
Reseat
.
Check
.
Check
.
Replace
AWN
.
Reseat
.
Check
.
Check
.
Replace
BONS
MEA*.
Top
Fuse.
Battery.
Power
Power
MEA*.
Top
Board. Fuse. Battery.
MEA*.
Top
Board.
Fuse.
Battery.
Power
Supply
Board”,
Board".
(313)
1.
Replace
Air-in-Line
Sensor
*.
(314)
(315)
(326)
sofico
5,0
*
MK
1.
1.
2.
level
Replace
Replace
1.
Do
FSOD.
Replace
of
testing
Air-in-Line
Air-in-Line
MEA*.
Guide
Sensor*.
Sensor”.
Lines
in
Service
Bulletin
404A.
il
¡IVAC
#TO*CAL
\
SYSTEMS
Inquiries
fluid
the
1.
.
.
have
been
made
side
occlusion
subject,
but
SOLUTION
Viscous
algorithm
but
instead
solutions
written
certain
factors
of a 20
Suggestions:
alarms.
intended
VISCOSITY
obviously
into
can
and
knowledge
If
this
MOTOR
It is
possible
would
A
occupied.
this
observe
and B are
channel
Suggestions: of
999
for a period
that
stiff
as
selected.
The
that
This
over
in
any
that
is
associated
issue
of 4 to 5 hours - this
DISPOSABLE
The
pumping
inserted. will
have
lubricant will
also
Suggestions:
for
whatever
manufacturing
CALIBRATION
Annual
systems. A variety
failing
to
Suggestions:
chamber
The
amount
an
impact
was
applied
degrade
the
Once
reason, I recommend
problem
calibration
perform
the
Calibrate
Medsystem
as
to
the
The
quick
a
as
requires
the
FSOD
complicate
regarding
appears
to
be
extended
mechanical
typically,
is
for a single
On
the
that
ch.
average - ch. C is
with
is
easily
is
sealed
of
energy
on
FSOD.
in
the
piston
you
have
take
is
recommended
of
symptoms
An
manufacturing mobility.
determined
action
calibration.
the
instrument.
UM
possible
following
reference
system
the
flow.
the
cause,
periods
assembly.
C has
nuisance
overcome
by a silicone
required
example
FSOD
information
guide.
more
is
designed
These
the
use
suggest a 20dpm
of
infusion,
FSOD
by
is
normally
piston
by
the
of
of
the
that
you
install a new
to
return
to
including
the set
account
nuisances
and
suggested
is
in
to
factors
of
the
confirm
non-use,
If
we
ch. A is
chosen
periods
simply
motor
this
the
of
alarms.
running
done
sleeve
to
would
cassette.
piston
cassette,
for
for
any
FSOD
remedies
not
an
all
inclusive
through
compensate
include
set
the
motor
analyze
selected,
only
non-use
by
biomed.
in
which
drive
be
in
for
using a 60
key.
be
used.
and
channel
for
after
chs. A and B are
and
the
instrument
the
the
piston
cases
where
Contamination,
portion
of
//
the it
analysis.
wear
or
drift
could
May
8,
for
nuisance
study
the
set.
fluid
viscosity
drop
drive
belt
utilization two
lines
quite
often
at a rate
plastic
piston
up
and
insufficient
spills,
cassette
appears
in
to
the
1996
on
The
set
may
we
chs.
it
is
is
down
etc...,
is
stiff
be
a
sensor
from
NOTE - In
Box
Gites
Office
85335
hit
Corporate
Р.О.
Web.
CONTAMINATION Spills,
the
moving Suggestions:
the
important
danny
leaking
FSOD
parts.
recommended
any
Wateridge
10221
a
Diego.
San
ae
com
system.
Encourage
case
that
California
nef
sets
and
The
increased
cleaning
where
the
set
San
a
Circle
92186-5335
other
As
the
the
procedure
the
be
returned
Diego.
(619)
m
forms solution
load
institution
of
contamination
on
the
motor
to
for
MSIL,
administration
to
IVAC-Creedmoor
California
92124-1579
458-7000
will
have a negative
to
dry,
friction
triggers
FSOD
maintain a cleaning
hospitals
set
is
found
for
develops
alarms.
regimen
that
do,
to
be
analysis.
between
as
benefit
leaking,
Lee
Sr.
greatly.
it
Wengronowitz
Support
Tech.
outlined
is
on
the
in
vitally
Spec.
om
ire
©
ALARIS
MEDICAL
MEMO
Date
11-19-98
To:
Technical
Cc:
Rob
SYSTEMS
Support
Pecsar,
Tam
Tran
From:
Subject:
1.
At
password.
can
The file.
installed
2.
Hiep
Background
the
first
not
be
only
The dle.pwi
(default
Resetting
Exit
DLE
Launch
Open
Click
Click Click Record Click on
.
Uncheck
OBNOASWH>
410.
Click
11.
Click
12.
On
the
13.
Hit
the
14.
Click
15.
Verify
16.
Click
Dinh
Resetting
time
of
running
This
password
changed.
way
to
change
file
at
cAmsiiidle)
program.
Windows
‘Msiiidle’
or
‘View’. ‘Options’. ‘View’
‘Apply’ ‘OK’
‘Yes’
tab.
the
settings
‘Show
‘Hide
button.
button.
right
hand
‘Delete’
button
that
there
ail
file
key
‘View’.
the
DLE
(Version
DLE,
will
be
the
DLE
is
located
Password
NT
Explorer
‘msiiidie’
on
files’
extensions
side
on
on
is
no
folder,
this
button.
window,
the
keyboard.
the
‘Confirm
dle.pw
the
program
saved
password
in
the
directory
(Start,
or
window.
for
known
click
File
file
in
2.00)
Password
asks
in
the
is
for
Program,
the
file
the
dle.pwl
Delete’
the
right
the
users
to
where
reset
the
file,
it
Windows
Windows
where
types’
hand
box.
icon
popup
side
Instruction
to
set
named
by
deleting
DLE
DLE
to
highlight
die.pwi,
application
NT
Explorer).
was
window.
window.
the
DLE
and
the
dle.pwl
was
NT
installed.
it.
Memo:
Resetting
the
DLE
(Version
2.00)
Password
Instruction
Page 1 of
3
ALARIS
MEDICAL
SYSTEMS
17.Click
18.
Click
19.
Reset
20.
Click
21.Click
22.
Close
23.
Start
3.
Resetting
+
Exit
DLE
Launch
Open Click
AN
Click Glick
Record
Uncheck
OND
Folders.
9.
Click
10.
Click
11.
Click
12.
On
the
413.
Hit
the
14.
Click
15.
Verify
16.
Click
17.
Click
18.
Click
24.
Reset
19.
Click
20.
Click
21.Close
22.
Start
‘Options’.
‘View’
‘Apply’ ‘OK’
the
the
settings
button.
the
‘Windows
DLE
tab.
button.
program.
program.
'Windows
‘Msilidle’
‘View’.
‘Folder ‘View’
on
‘Apply’ ‘OK’
Options...’
tab.
the
settings
‘Hide
‘Show
button.
button.
right
hand
‘Delete’
‘Yes’
button
that
there ‘View’. ‘Folder ‘View’
the ‘Apply’ ‘OK’
the
the
Options...’.
tab.
settings
button.
button.
‘Windows
DLE
on
this
NT
Password
Explorer’
or
‘msilidie’
on
this
file
extensions
all
files’
side
window,
key
on
the
on
the
is
no
dle.pwi
on
this
Explorer’
program.
Explorer’
Enter
and
(Start,
folder,
window.
for
button
under
click
keyboard.
‘Confirm
file
window
window.
Enter and
as
recorded
window.
confirm a new
for
Program,
or
the
known
Hidden
the
File
Delete’
in
the
as
recorded
confirm a new
Windows
file
files.
dle.pwi
right
hand
earlier.
password.
Windows
Explorer).
where DLE
box
icon
popup
side
earlier.
password.
was
under
to
highlight
window.
window.
95
installed.
Files
and
it.
4.
Exit
Launch
Open
Click
DONT
Memo:
Resetting
DLE
program.
‘File
Manager’.
‘msilidle’
‘View’.
Resetting
the
folder,
DLE
(Version
Password
or
the
2.00)
where
Password
for
Instruction
Windows
DLE
was
installed.
3.1X
Page 2 of
3
ALARIS
Click Record
Check
Click
On
oo
10.
Hit
the
11.
Click
12.
Click
13.
14,
Click
15.
Click
16.
Reset
17.
Click
18.
Close
19.
Start
[end]
MEDICAL
‘By
File
the
‘Show
‘OK’
the
right
‘Delete’ ‘OK’ ‘Yes’
that
View.
'By
the
‘OK
the
the
SYSTEMS
Type...’.
settings
Hidden/System
button.
hand
side
key
button
button
there
File
settings
on
on
is
no
Type...”.
on
button.
‘File
Manager’
DLE
program.
on
this
window,
on
the
the
the
die.pwl
this
window.
Files’.
click
keyboard.
‘Delete’
popup
‘Confirm
file
window.
Enter
and
the
die.pwt
window.
File
Delete’
in
the
right
as
recorded
confirm a new
icon
papup
hand
earlier.
password.
ta
window.
side
window.
highlight
it.
Memo:
Resetting
the
DLE
(Version
2.00)
Password
Instruction
Page 3 of
3
Ни.
al
IVAC
MEDICAL
SYSTEMS
NOTIFICATION
OF
PRACTICES
TO
REDUCE
THE
OCCURRENCE
THE
Attention:
IVAC alarms
e
9
s
Medical
can
Fully
extending
as
illustrated
Turing
Stopping
Removing
IVAC®
Sales
be
reduced
the
the
the
OF
MEDSYSTEM
Force
Systems,
in
Figure 1 below.
pump
channel
cassette
with
the
cassette
on
Inc.
the
before
from
PUMP
has
determined
following
slide
clamp
inserting
removing the
pump
INCORRECT
LATCH
III™
that
practices:
when
the
cassette
the
cassette
CASSETTE
SLIDE
CLOSED
INFUSION
the
occurrence
removing
into
from
turning
the
<>
Em
the
the
pump
of
the
cassette
pump.
pump.
off.
ALARMS
PUMP
pump
latch
from
the
FOR
pump
7
To
address a pump
+
Using as
illustrated
e
{f
the
pumping
qualified
If
you
require
Corporate
only
service
STAT
latch
your
finger,
in
Figure 2 below.
latch
jaw
personnel.
further
line
information
using
push
is
visibly
our
down
toll
free
the
procedure
the
broken,
PUMPING
press
LATCH
regarding
number
1
pumping
JAW
2
this
(800)
below
the
latch
“SERVICE”
notification, 547-7828.
be
followed:
jaw
until
softkey
please
it
snaps
and
call
open
contact
IVAC's
Corporate
PO. Box
35335
Web
Site:
Office @ 10221
dl
San
http:/
/www.ivac.com/
Wateridge
Di
Circle 8 San
California
92186-5335 @ 1819)
-wac/
Diego.
California
92121-1579
453-7000
MedSystem
III
P/N
2860505 2860521
2860539 2860554
2860604
2860620
144980
143013
136615
.
160653
2860703
2860530
2860729
2860745 2860752 2860760
145799
2860802 2860516
2860950
2860828
2860836
2860844 2860851
2860869
2860877
2860885
2860893 2860919
Description
Top Lower
Lower Main
Main
Graphic Pad
Dust
Display,
Power
Power
Main Main Main
MEA
Audio/Connector Audio/Connector
3A
Kit,Battery
NiCad
‘Assy,Memory,
- Drive
Kit,
SPARE PARTS
Label
Kit
FuseKit,(5pcs)
Drive
Motor
Valve/Actuator
Air-In-Line
Air-In-Line
Air-In-Line
Slide
Serv
Serv Serv
Chassis
Knob
Slide
Kit,Pressure
Kit(2ea)
Housing
Housing
Housing
Housing
Label
Cap
Kitf,(5ea=1Pump)
Module
PCBAKit
PCBAKit
Electronics Electronics
Electronics
kit
2863/5
Pack
Housing
Kit
Sensor
Links(15pes=5Pumps)
Pump
Assy,
Pump
Assy, Assy,Module,Opto
Mount
Assembly
Assembly
Knob
Connector,
Transducer
Kit
Clip(2es)
Partial
Partial Kit
Assembly
Assembly
(2860/3)
|
(Reconditioned)
(2863/5)
Assy
Assy Assy
PCB PCB
2860, 2860,
Kit Kit
2860,
5.0
With
3.5
Software Software
Assy
Assy
Kit
Battery,
Backup
Kit
Kit@sets=1Pump)
Kit
2860
Kit,2863/5
Screws(soscrews=30sets=10Pumps)
Shaft
Seal(15pes)
Shaft
(2sets-1Pump)
Service
Kit,
Kit
RepairGsets=1Pump)
LIST
Qty!
2 2
2 2
2 2 2
2
2
1
1
1 1
1
1
1 1
2
„5
1
2
2
1
2
2
2
1
1
2
2
2
5
2
2
2
2 2
Revised
Unit
Price
15.48
26.24
12.41
167.71
20.25
14.51
89.10
46125
250.00
486.48
507.04
36.54
90.00
32.46
60.14
34.88
143.01
23.97
115.00
80.46
141.21
100.81
80.26
105.84
16.29
55.67
121.00
11
May
00
2860794
"860968 2861065 2861089 2861123 2861131 2863079 2861461 2860525
2860526
2861107
2861115
2863087
2861462
2863012
2863277
15370
2860529
2863277
8631A
2863351
Tape,
Pressuer
Pump
Pole
Clip,
Locking
Universal
LCD Cap,
Calibration
FMS2.2c,
For
Latch
Clamp
Adapter,
AC
Pole
Backlight
Service
use
with
pump
Kit
Chass
AC
Adapter
Knob
Device
FMS2.2c,Service
For
use
with
pump
FMS, For
FMS,
For
Service
use
with
Service
use
with
Kit,
pump
Kit,
pump
FMS2.24Service
For
use
with
pump
Kit
Communication
Adaptor
Manual
Manual,
Kit,
Kit,
5.08-A
Communication
2.1
2.2 Syringe
Die
MSIII
Front
Front
Quick Carrying
9D
Service,
Maintenance,
DFU,
2863
DFU,
286X
Upgrade
FMS
FMS
cast
Chassis
Operating
Ref
Case
FEM
Key
Key
Guide
Trasducer
Mount
Gpes)
Clamp
Assembly
Assy(Not
Kit,
With
software
software
software
software
software
versions
Kit,
W/o
versions
With
Cablesq.1p(Not
versions
W/o
Cables.1p(Wot
versions
Kit
with
versions
Cable
to
25
Technical
Field
For
use
with
(For
use
with
Kit(ASIC
Protocol
Inst.
Lbl.
Pad
(2860/3)
Pad
5.0
Cover
Screws(4osccows=20sets=10Pumps)
in
any Service
Cables@.1p(Wot
4.0
or
Cables¢.1j(Not
4.0
or
3.3 & 3.5.
3.3 & 3.5.
Cables
4.0
or
pump
software
pump
software
Pumps
Model
Agreement)
greater.
greater.
in
any Service
in
any
and
DLE
greater
version
version
2863)
in
any
Service
in
any
Service
Agreement)
Service
Agreement)
3.5.)
4.0 & greater.)
Agrmnt)
Agrmní)
gor
front
of
pump)
0.70
29.32
17.87
27.33
61.29
58.50
95.00
95.00
10.00
85.00
25.00
25.00
25.00
95.00
25.00
25.00
25.00
13.50
10.12
9.92
MEDICAL
SYSTEMS
Technical
IVAC®
MedSystem
Service
Manual
III“
Infusion
System
This
of
the
device
is
following
protected
U.S.
Patents:
under
one
or
more
4,814,685 4,818,190
4,832,299
4,850,817
_
4,856,340 4,863,425 4,872,813
D
309,188
D
309,662
Other
U.S.
and/or
Copyright O 1999
All
Rights
Reserved.
Reproduction English
or
in in prior
reviews
or
written
other scientific
permission
ALARIS
in
any
languages,
4,878,896 4,884,065 4,919,596 4,940,399 4,950,235 5,000,663 5,006,110 5,009,641 5,041,086
foreign
manner,
Medical
papers,
of
ALARIS
patents
in
except is
may
Systems,
whole
or
for
prohibited
Medical
be
Inc.
in
part,
brief
excerpts
without
Systems.
5,053,747 5,064,412 5,126,616 5,149,603 5,176,343 5,176,631
pending.
in
Certification
has
(DE
been
granted
by
the
following
p
©
Standard Risk
agencies:
C22.2
Class
2G
No.
125
ALARIS”
Medical
Systems,
Inc.
IVAC®
Models
MedSystem
2860
Multi-Channel
Technical
Service
III"
and
Infusion
2863
Manual
Pump
MedSystem
Ill?
is
the
trademark
of
ALARIS
Medical
Systems,
Inc.
formerly
known
as
IVAC
Corporation.
MedSystem
111%
Technical
Service
Manual
Copyright O 1999
This
document
possession
described. authorization
.
This
Product Diego,
Printed
Manual
does
Reproduction,
service
Support
CA
92126-5335,
in
the
part
from
manual
U.S.A.
number
ALARIS
contains
not
convey
ALARIS
is
subject
at
(800)
139981B
Medical
proprietary
any
rights
disclosure,
Medical
to
change
854-7128
attention
Product
Systems,
information
Systems,
ext.
to
or
use
without
6003,
Support.
Inc.
reproduce
other
Inc.
of
ALARIS
its
than
for
is
strictly
notification.
or
write
ALARIS
All
Medical
contents,
the
intended
forbidden.
For
Medical
Rights
Systems,
or
to
manufacture
purpose
current
Systems,
Reserved.
Inc.,
and
without
technical
Inc.,
its
receipt
or
sell
any
specific
information
P.O.
Box
or
product
written
please
85335,
call San
WARRANTY
“ALARIS
A.
Each
use
Medical
Medical
new
and
Medical
B.
Each
new
use
and
Medical
If
any
product
'
this
expense.
packaged
repair in
consequential Systems warranty
Furthermore, responsible any.
(a)
communicate
CA)
warranty,
The
and
facility
no
event
product.
shall
repaired representative;
Systems™,
Systems”)
instrument
service
Systems.
for a period
to
accessory
service
Systems
to
shall
shall
this
for,
Medical
by
for a period
requires
determine
repair
product
postage
be
ALARIS
damages
This
not
warranty
any
Inc.
warranis
is
free
the
original
is
free
to
the
original
service
directly
the
appropriate
or
replacement
requiring
prepaid
at
purchaser's
Medical
in
connection
warranty
apply
loss
Systems
to
shall
or
other
formerly
that:
from
defects
of
one
purchaser.
from
defects
of
purchaser.
with
the
service
by
purchaser.
risk.
Systems
shall any not
apply
product
than
IVAC®
in
material
(1)
year
in
material
(90)
days
the
applicable
will
with the
apply
subsequent
arising
which
an
authorized
be
carried
to,
and
in
has
be
facility.
purchase
connection
Corporation
and
from
the
and
from
the
warranty
Medical
Loss
liable
owner
been;
Systems
out
at
be
returned
or
for
or
to
the
Medical
or
(hereinafter
workmanship
date
of
delivery
workmanship
date
of
delivery
as
ALARIS
any
use
original holder
with
the
Medical
period,
provided
in
incidental,
of
Systems
the
headquarters
Medical
promptly,
any
of
purchase
Systems
referred
under
under
normal
by
normal
by
purchaser
otherwise
Systems
properiy
shipment
indirect
Medical
the
product.
shall
not
or
use
service
to
as
(San
This
be
in
to
or
of
(b)
altered
product's
(c)
subjected serial
(d)
improperly
written
This
warranty
obligations not
give
person
with
the
ALARIS
in
stability
or
lot
instructions
or
or
to
assume
sale
or
MEDICAL OR IMPLIED, FITNESS
See
FOR A PARTICULAR
packing
inserts
anyway
to
maintained
is
liabilities
use
so
or
altered,
furnished
in
lieu
of
directly
on
behalf
of
reliability;
or
SYSTEMS
INCLUDING
for
international
as
to
affect,
negligence
effaced
or
used
by
of
all
other
Medical,
or
indirectly,
of
ALARIS
Medical
DISCLAIMS
ANY
PURPOSE
in
ALARIS
or
accident,
or
removed;
or
in
any
Medical
warranties,
Systems
the
Medical
Systems
WARRANTY
OR
warranty,
Medical
or
which
other
Systems.
express
and
authority
Systems
products.
ALL
to
any
OTHER
OF
MERCHANTABILITY
APPLICATION.
if
applicable.
Systems
- has
had
than
in
accordance
or
implied,
Medical
any
representative
other
liability
WARRANTIES,
judgement,
the
product's
with
and
all
other
Systems
in
connection
or
does
other
EXPRESS
OR
the
the
OF
iii
The
purpose trained troubleshoot,
All
KEY
instrument.
This
manual Summary, the
instrument.
in
the
NAMES
Screen
Appendix
of
analysis
contains
this
and
are
Technical
and
repair
the
written
messages
guidelines
H,
for
MedSystem
Servcie
servicing
MedSystem
in
regular type
are
written
for
general
precautions
III?
Technical
Service
Foreword
Manual
of
microprocessor-controlled
user
M1?
Infusion
as
they
in
italics.
and
patient
is
to
appear
before
enable
System.
safety.
beginning
personnel
on
Manual
who
instrumentation
the
computer
We
suggest
any
maintenance
are
experienced
to
keyboard
you consult
or
and
maintain,
or
the
Safety
repair
on
Specific applicable.
If local
If
If
you
you
you
advisory
need
ALARIS
need
need
and
cautionary
Contacting
assistance
Medical
technical
Customer
Systems
assistance,
with the
Service
information
Medical
installation
Account
call
or
Parts,
Manager.
toll
free
call
is
contained
or
operation
(800)
toll
free
Systems
of
854-7128
(800)
throughout
Technical
MedSystem
extension
482-4822.
the
manual
Support
Ill
equipment,
6003.
where
contact
your
:
Chapter
1:
INTRODUCTION
MAIN
TABLE
OF
CONTENTS
Chapter
contents.
Chapter
Chapter
for
the
Chapter
Chapter Three
after
Chapter
Chapter operating
Chapter
Chapter optimum
Chapter
One
contains
2:
PRINCIPLES
Two
MedSystem
3:
repair
or
4:
MAINTENANCE
Four
efficiently
5:
CALIBRATION
Five
performance.
6:
TROUBLESHOOTING
general
provides
III.
an
overview
FUNCTIONAL
describes
calibration,
provides
contains
the
and
procedures
and
to
references
information
OF
OPERATION
functional
annually
detect
the
for
about
of
the
system
TESTS
tests
recommended
as
part
of
for
properly
need
for
calibrating
the
MedSystem
and
details
preventive
maintaining
repair
or
adjustment.
and
adjusting
Ill
the
principles
`
prior
to
maintenance.
the
the
and about
of
using
the
instrument
instrument
the
manual
operation
instrument,
to
keep
it
to
maintain
Chapter possible procedures
Chapter
Chapter
reassembling
and
Six
contains
malfunctions.
in
the
event
7:
Seven
contains
the
instrument,
subassemblies.
troubleshooting
References
of a malfunction.
detailed
and
are
for
step-by-step
procedures
included
removing
to
to
help
guide
procedures
and
installing
isolate
service
replaceable
and
identify
personnel
for
disassembling
causes
to
the
proper
and
components
of
MedSystem
ili
Technical
Service
Manual
Appendix
This
appendix
instrument.
Appendix
This
appendix
depicting
Appendix
This
appendix
Appendix
This
appendix
useful
troubleshooting.
Appendix
in
MAIN
A:
TECHNICAL
contains
B:
REQUIRED
is a general
special
C:
RECOMMENDED
lists
D:
BATTERY
provides
performing
E:
ILLUSTRATED
TABLE
the
EQUIPMENT
listing
tooling
and
recommended
HISTORY
details
battery
OF
CONTENTS
SPECIFICATIONS
physical,
electrical,
AND SUPPLIES
of
required
equipment.
SPARE PARTS
spare
parts,
LOG
about
the
Battery
operating
PARTS
time
functional
LIST
and
equipment
as
well
as
History
checks,
(Continued)
mechanical
and
the
order
Log,
specifications
supplies,
numbers
which
contains
annual
of
as
well
as
figures
for
each.
information
maintenance,
the
and
This
appendix
each.
Appendix
This
appendix
Appendix
This
appendix
Appendix
This
appendix
the
manual.
Appendix
This
appendix
assembly
is
an
illustrated
F:
LIST
OF
provides a list
G:
GLOSSARY
defines
H:
SAFETY
provides
I:
SCHEMATICS
provides
and
the
power
listing
ABBREVIATIONS
of
the
abbreviations
OF.
TERMS
some
words
SUMMARY
general
supply,
board
safety
assemblies
along
of
instrument
and
terms
precautions
with
parts,
AND
an
ACRONYMS
and
which
may
that
and
schematics
interconnect
as
well
acronyms
not
be
are
not
diagram.
as
the
order
used
in
the
well-known
included
for
the
within
main
numbers
manual.
to
the
user.
the
electronics
for
text
of
TABLE
INTRODUCTION
1.0
Systeminiroducton........................
1.1
Objeciives................
1.2
UsinghisManual..........................
1.3
Newinstrument
1.3.1
1.3.2
1.4
Customizing
15
PointsofContaci.........................
oPhysicalinspection
Functionallesis.........................
Checkout.
.................
............................
Parameters...................
OF
CONTENTS
cee
00mm
444
00...
eserse
eee
0.0.
eee
ων
een
κών
eee
enes
ence
eens нина
κενο ν 1-4
ώ
ον
Page
1-3 1-3 14
1-4 14 1-5 1-5
MedSystem
III
Technical
Service
Manual
1-1
Introduction
1-2
MedSystem
III
Technical
Service
Manual
Chapter
1
Introduction
1.0
System
The
MedSystem
*
based
venous, the bottles operates and
The
fluid
equivalent
or
programmed
air
in a line,
tored
using
following
Upgradeable
»
Low
e
Calibration
*
CMOS
*
Modular Easy
Maintenance,
of
MedSystem
user
to
communicate
port,
using
Introduction
II Infusion
delivery
intra-arterial,
bags
via a disposable
and
built-in
and
features:
maintenance,
surface-mount
disassembly/reassembly
an
device
or
subcutaneous
of
an
independent
located
above
infusion
for
infusion
low
self-tests
construction
software
no
through
calibration,
TI
software
Field
Maintenance
with
compatible
capable
or
tasks.
The
to
of
lubrication
circuitry
and
customization
the
infusion
System
is a small,
of
controlled
routes.
infusion
below
the
cassette.
Infusions
electronic
help
ensure
the
MedSystem
Software
system through
computer.
and
Each
device
level
of
able
Channels
are
and
software
safe
operation.
III
features
(FMS).
three-channel,
monitored
of
the
three
to
pump
the
instrument.
are
individually
monitored
functions
Infusion
are
available
This
its
RS-232
microprocessor-
infusion
channels
fluid
to
detect
software
communications
via provides
from
syringes,
Each
channel
controlled
occlusions,
are
also
moni-
provides
through
the
allows
intra-
the
use
the
1.1
Objectives
The
purpose
troubleshoot, prepared analysis mended
or
repair
In
light of
and
specialized
it
is
recommended
ule
or
replaceable
minimal It is
suggested
procedures tions
For
tions
for
of
and
under
and
servicing
that a factory
of
the
the
or
no
be
Use
Use,
this
manual
the
presumption
is
to
enable
the
MedSystem
that
of
microprocessor-controlled
authorized
training
instrument.
state-of-the-art
equipment
that
troubleshooting
part
unsoldering that a complete
obtained.
(DFU)
in
and
addition
technology
required
Replaceable
of
circuit
overview
This
can
the
MedSystem
to
this
Technical
qualified
HI
Infusion
the
user
program
used,
in
the
replacement
and
parts
components.
of
be
done
by
If
Service
service
System.
is
experienced
instrumentation.
precede
proprietary
repair
all
be
are
items
features,
consulting
Field
Maintenance
Manual.
personnel
This
and
any
nature
of
individual
performed
that
can
functions,
the
MedSystem
to
maintain,
trained
of
has
in
It
is
disassembly
the
software,
components,
only
to
the
be
replaced
and
operating
II
Software
been
the
mod-
with
MedSystem
HI
Technica!
Service
Manual
1-3
Introduction
1.2
Using
This Throughout to
both
called The
Each principles
them. ter 5 describes
bleshooting,
troubleshooting
This
manual
models.
Manual
includes
the
manual, Where
out.
Technical
chapter
Service
is
clearly
of
operation.
Chapter 4 provides
procedures
and
includes a number
process.
ment.
The
appendices
A-
Technical
B - Required C - Recommended D - Battery E-
Illustrated
E-
List
of
G - Glossary
H-
Safety
I-
Schematics
All
instructions,
each
chapter
tion. It
is
suggested
ning any
type
included
Specifications
Equipment
History
Parts
Abbreviations
of
Terms
Summary
special
to
minimize
that
of
service
information
if
no
information
Manual
defined
on
the
model
number
is
specific
is
specifically
in
the
table
Chapter 3 discusses
both
routine
for
calibrating
of
Chapter 7 is a detailed
are:
and
Supplies
Spare
the
Log
List
Parts
and
notes,
the
need
on
the
Acronyms
and
illustrations
to
refer
Summary,
instrument.
Model
is
of
functional
and
annual
the
to
Appendix
2860
and
Model
specified,
to
one
structured
contents.
the
model,
to
Chapter 2 discusses
tests
maintenance
instrument.
to
be
used
Chapter.6
as
procedure
are
integrated
multiple
H,
pages
be
reviewed
2863
pumps.
information
that
model
facilitate
and
how
number
ease
to
complete
procedures.
discusses
guides during
for
repairing
within
during a single
the
pertains
is
of
use.
the
Chap-
trou-
the
the
instru-
text
of
1.3
1.3.1
1.3.2
_
New
Physical
*..
* +
Instrument
Inspection
Before
affected Check Check
found,
unpacking,
contents. to for
report
Functional
Refer and
to
your
testing
of
Recommended
MedSystem
1H
Technical
Checkout
check
Report
ensure
any
Tests
institution's
that
all
physical
it
to
Customer
incoming
equipment
functional
Service
Manual
accessories
damage
policies
tests
the
shipping
any
shipping
Service.
for
are
container
are
to
the
instrument
specific
before
given
in
for
damage
included
to
in
or
requirements
use.
Chapter 3 of
damage
Customer
the
that
Service.
package.
accessories.
regarding
this
manual.
may
have
If
any
is
inspection
Introduction
1.4
1.5
Customizing
Certain customizing Directions
Points
For
Customer
1-800-482-4822
For
Product
1-800-854-7128
MedSystem
For
of
parts,
disposables,
Service
technical
Support
is
desired,
Use.
Contact
information,
Parameters
II
operating
refer
or
repair
Department
assistance
Department
extension
6003
to
the
service,
parameters MedSystem
telephone:
with
repair,
can
be
I
Directions
or
technical
customized
For
Use
training,
using
telephone:
and
FMS. the
if
FMS
MedSystem
III
Technical
Service
Manual!
1-5
Chapter
2
2.0
Figure
Principles
System
The
MedSystem
wide
shown and
includes
independent system.
monitor
is
controlled
sensors stopped, sage
on
tery
pack
minimized nents
2-1.
Overview
range
of
in
Using
infusion.
cassette
that
or
watchdog
and
the
LCD
or
by
and
low
Block
of
III
Infusion
operating
2-1. A cassette
pumping
the
Based
keypad
interfaces
mechanism,
by a microprocessor.
placement,
the
microprocessor
circuit
the
operator
screen.
DC
power
using
complimentary
duty-cycle
Diagram
System
conditions. A block
provides a sterile
for
monitoring
motor
and
LCD,
on
Rate
and
Various
occlusions,
properly
detect a condition
is
alerted
The
system
from
the
MedSystem
operation.
of
MedSystem
Operation
provides
the
and
by
the
is
powered
metal oxide
Ill
safe
diagram
pathway
infusions.
control
operator
system,
can
Remaining,
sensors
air
in
the
executes
that
red
LEDs,
by either a rechargeable
AC
Adapter.
semiconductor
Infusion
System
and
of
monitor
line. A watchdog
the
is
accurate
the
Each
fluid
overall
for
delivery
channel
delivery
system
contains
and infusion
program
software
inappropriate,
an
audible
the
pumping
the
Power
the
instrument
infusion
circuit
program.
the
tone
and a mes-
consumption
for
a
is
of
an
monitoring
to
mechanism
to
detect
If
the
infusion
NiCd
is
bat-
is
POWER
AC
USER
INTERFACE
Ker
AD
:
-
LEDs
-
AUDIO
AC
ADAPTER
WATCHDOG
MICRO—
PROCESSOR
CORE
AEA
Vext
POWER
SUPPLY
Grup
RECHARGEABLE
BATTERIES
MONITORING
>
-
- »
OCCLUSIONS
+5V
Lee
+13V
|
Bov.
NICO
MOTOR
CONTROL
CASSETTE
LATCH
АВ-М-ВМЕ
Hz
400
MEGHAN
(BACKLIGHT)
FROM
CONTAINER
CASSETTE
το
PATIENT
MedSystem
HI
Technical
Service
Manual
2-3
Principles
The
main
sette
contains
brane
sealed
stroke
section
The
collar
securely
chassis
LD.
optics
ing
different
instrument cally
of
Operation
functional
an
elastomeric
is
exposed
for
by a silicone volume
between
of the
is
seats
in
the
instrument
also
approximately
the
in
sensor.
provide a code
cassette
chassis,
occludes
the
components
valving
piston
the
engages
sleeve
cassette
and
lower To
detect
types.
The
the
when
of
the
membrane
and
for
sensing
in
which
80
microliters.
housing
of
cassette
the
identification
that.can
cassette
slide
link
the
cassette
cassette
between
are
two
pressure.
the
plastic
Air
collar.
the
instrument
placement,
be
used
in
slide
clamp
to
close
is
removed
shown
rigid
The
piston
is
sensed
and
the
optomodule
future
applications
latches
the
pump
from
in
Figure
plastic
pumping
holds
parts.
is
inserted.
in
the
outlet
the
of
the
the
cassette
and
the
instrument.
2-2.
The
The
mem-
chamber
The
piston
tubing
tubing
sensor
cassette.
in
for
identify-
to
the
automati-
cas-
is
the
The
Figure
2-2.
The
2-3.
the
Cassette
physical
The
maim
interface
of
three
boards:
board. backup
MEA
the
to
The
consists
top
board.
the
MEA
battery
MEA.
Main
Functional
configuration
microprocessor
electronics
and
infusion
(1)
power
supply
connect
of
three
The
via
the
core,
assembly
monitoring
the
printed
to
boards:
power
top
board.
of
the
motor
(MEA),
the
MEA
(1)
supply
The
Components
electronics
is
control,
together
systems.
circuit
via the
the
printed
memory
In
(2)
the
board
top
circuit
backup
shown
and
the
auxiliary
as a block
watchdog
with
interface
Model
assembly
board.
(2)
board
assembly
battery
circuits
2860,
(PCBA)
In
the
the
connects
diagram
are
circuits
the
MEA
and
(3)
and
Model
board, and
(PCBA)
in
included
for the
user
consists
the
top
memory
2863,
the
connects
directly
to
in
(3)
the
2-4
MedSystem Ш Technical
Service
Manual
In
both
models,
also
connect
the
main
housing.
communications, connector
module
with
by
contains
an
electroluminescent
an
80V
to
PCBA;
AC
inverter
the
the
drive
MEA.
The
audio
and
empty
however,
three
green
on
modules
The
transducer
the
backlighting
the
and
audio/connector
container
ECD
and
three
red
power
supply
monitoring
PCBA
and
connectors
detectors
LEDs
(ECD)
has
been
and a super-twist
The
PCBA.
Principles
sensors
LCD
mounted
and
keypad
for
external
are
included
discontinued.
backlighting
of
Operation
on
the
chassis
are
mounted
power,
The
dot
RS-232
on
the
display
matrix
is
powered
on
LCD
Figure
2-3.
lag
dE
|
|
WS
lw
«δις
jus
Be пы
|
|
>
E
č
a
<
[2
8
=
o
|
E
o
£
=
|
Block
(Model
|
(BS)
oo]
=
ge
Js
O)
as
|
Diagram
2860
only)
lu
še
Da
a
|
jeg
ce
SE
=5
38
<
of
Electronics
Physical
a y
5
8
8 | [15
5
5%
i |
za
ah <
ge
τας
55
og
=
e
at
E
1
C
Configuration
|
|
|
|
|
4
3
|
En
5
|
35
|
22|
++
EDs
N
38
19
yu
| |
gs
< |
205
SHE
250
そら
8
я
2
|
Ex
Sam
Lo
a
Vbl
S|
(E
|
LIRE
Bui
ЕН
ES
G
a
<
a
EPROM
PCBA
KEYPAD
MEMORY
BACKUP
BATTERY
MODULE
DISPLAY
MedSystem
Ill
Technical
Service
Manual
2-5
Loading...
+ 211 hidden pages