Alaris MedSystem III User manual

MedSystemIII
®
Multi-Channel Infusion Pump with
Advanced Dose Rate Calculation
DIRECTIONS FOR USE
INTRODUCTION
BASIC OPERATION
ADVANCED OPERATION
TABLE OF CONTENTS
INTRODUCTION
ABOUT THE PUMP....................................................................... 1
FEATURES.................................................................................... 2
SYSTEM COMPONENTS ............................................................... 4
OPERATIONAL PRECAUTIONS ...................................................... 6
BASIC OPERATION
PREPARING THE INFUSION .......................................................... 9
PREPARING THE ADMINISTRATION SET ........................................ 9
LOADING THE SET ....................................................................... 9
FRONT PANEL OVERVIEW ............................................................ 10
PROGRAMMING PRIMARY INFUSION........................................... 14
MAKING CHANGES WHILE INFUSING .......................................... 15
KVO STATUS................................................................................ 17
ADVANCED OPERATION
SECONDARY MODE .................................................................... 19
PREPARING THE ADMINISTRATION SET ........................................ 19
PROGRAMMING SECONDARY INFUSION...................................... 20
DOSE RATE CALCULATOR (DRC) USING A SPECIFIC DRUG NAME............... 22
DOSE RATE CALCULATOR PROGRAMMING WITH DRUG? ........................... 24
DEVICE........................................................................................ 26
CONFIG....................................................................................... 28
NOTE .......................................................................................... 29
BATLOG....................................................................................... 29
TROUBLESHOOTING
RESPONDING TO AN ADVISORY, ALARM OR FAULT ...................... 31
ALARM RESPONSE KEYS .............................................................. 31
ADVISORIES ................................................................................ 32
ALARMS...................................................................................... 33
FAULT.......................................................................................... 36
WATCHDOG ................................................................................ 36
OTHER CONDITIONS.................................................................... 36
MAINTENANCE
SPECIFICATIONS .......................................................................... 37
CLEANING................................................................................... 38
INSPECTION REQUIREMENTS ....................................................... 41
SERVICE INFORMATION............................................................... 42
WARRANTY................................................................................. 43
GLOSSARY................................................................................... 44
Alaris MedSystem III Directions For Use v

INTRODUCTION

CHAPTER 1 — INTRODUCTION
The MedSystem III®Multi-Channel Infusion Pump —
• three independent fluid delivery systems in the space of one.
• compact size:
- reduces bedside clutter
- simplifies patient transport
• easy to setup and use, yet provides advanced features.
• accurate delivery of a variety of fluids.
• accommodates assorted container types.
• multiple delivery methods: Intravenous/Intra-arterial/Subcutaneous/Epidural
• Uses administration sets that provide free flow protection.
• Six available Device Types with configurable parameters
(maximum and minimum rates, maximum volumes, baseline and maximum pressures, and air-in-line thresholds) to achieve specific clinical applications:
General Purpose Neonatal Controller Pressure Operating Room General Purpose II Operating Room II
• Displays infusion status for rate, volume remaining and
volume infused.
• Infusions can be programmed to deliver at a specified rate or
over a specified period of time.
• Secondary mode allows fluids and medications to be
delivered at two different rates, sequentially.
• Dose Rate Calculator (DRC) feature performs the volumetric
rate and/or dose rate calculations.
• With DRC activated, displays infusion status for rate, dosing
regimen and drug name.
• Communications Protocol allows clinical monitoring,
instrument configuration and maintenance.
• Field Maintenance software (FMS) available for Biomed to
configure, service and troubleshoot the pump.

About the Pump

1
CAUTION: With any
multiple or parallel infusions (i.e., connection of additional infusion systems to the MedSystem III pump and connection to the patient venous system), air infusion, reverse flow, and flow interruption are possible. Please consult the “Overall Problem/Solution Matrix” contained in DIN/VDE Standard 0753 in such cases.
WARNING: The use of
positive displacement infusion devices ported together with gravity flow infusion systems into a common IV site may impede the flow of common “gravity only” systems affecting their performance (hospital personnel must ensure the performance of the common IV site is satisfactory under these circumstances).
WARNING: The pump is
designed to stop fluid flow under alarm conditions other than the Low Battery and KVO. Periodic patient monitoring must be performed to ensure infusion is proceeding as expected.
WARNING: The
MedSystem III is a positive pressure delivery system capable of developing positive fluid pressures to overcome widely varying resistances to flow. It is neither designed nor intended to detect infiltrations and will not alarm under infiltration conditions.
WARNING: Hospital
personnel must ensure the compatibility of the drugs as well as the performance of each pump as part of the overall infusion. Potential hazards include drug interactions, inaccurate delivery rates, inaccurate pressure alarms, and nuisance alarms.
CAUTION: Only systems
that have been qualified to International Standard IEC 60601-1 should be connected to the Communications Receptacle, and the connection should ONLY be performed by qualified personnel. The selection, testing, and use of host computer hardware and software, in conjunction with the MedSystem III pump is strictly the responsibility of the purchaser.
INTRODUCTION
Multi-channel Fluid Delivery System
The instrument combines three independent infusion channels in an unparalleled small size.
Lightweight/portable
The pump with pole clamp weighs just over 5 pounds and is easy to transport.
Unique, rotating pole clamp
The Pump may be attached to a variety of surfaces.
Dose Rate Calculator (DRC)
The pump calculates a volumetric or dose rate based on values entered for patient weight, drug concentration (drug amount and diluent volume) and dosing parameters.
Six Device Types available
General Purpose, Neonatal, Controller Pressure, Operating Room, General Purpose II, and Operating Room II.
Free-flow Protection
The IVAC 28 and 25 Series Administration Sets contain a cassette that provides protection from free-flow conditions. To remove the cassette from the pump, the cassette’s slide clamp is pulled to full extension, occluding the tubing and preventing fluid from flowing.

Features

2
A: Stopped
DRUG? Wt ---- KG= 0.0 LB
A:Conc ------- mg/---- ml A:Dose ------- mcg/kg/min A:Rate ------- ml/h A:VR 1 ml (Vol Rem) A:VI 0 ml DI 0.0 mg Press Select to choose line
Select
Clear
INTRODUCTION
Monitoring System
The instrument continuously monitors pump conditions and alerts with adjustable audio tones and visual messages.
Data Monitoring
The pump can be configured to communicate with a remote computer, such as a centralized patient monitoring nurses station. The COMM receptacle is compatible with RS-232 cabling. A communications manual that describes the programming and hardware involved is available.
Field Maintenance Software (FMS)
The pump can be modified to accommodate specialized clinical applications. The Device Type parameters, occlusion limit, and air-in-line threshold can be reconfigured with the optional FMS software.
Secondary Mode
Allows the user to program two different rates of infusion to run sequentially.
Full Range of Delivery Rates
Rates from 0.1 to 999 milliliters an hour.
Battery Capacity
A fully-charged battery provides 6 to 8 hours of operating time with rates at 125 ml/h per channel.
3
A: Infusing Secondary
A: Secondary Rate 100 ml/h A: Sec VolRem (VR) 100 ml
A: Sec Time(TR) 1 hr A: Sec VolInf(VI) 1 ml
since 12:37p 01 Feb 02
Stop Affects Secondary
Select
Fast
INTRODUCTION

System Components

Slide Clamp
Pressure Dome
Piston
Tubing Collar
Lower Assembly
3 Cassette Holders
Pump Latch Mechanism
Drives the cassette piston to move fluid through the tubing.
Air-in-Line Sensor
Detects bubbles of air during infusion.
Tubing Collar Recess
Holds tubing collar in place.
Front Panel
Instrument Keys
Display Screen
Softkey Pads
Channel Indicator Lights
Green:
• Steady - infusing on AC power
• Flashing - infusing on battery power
Red:
• Slow flashing - Advisory
• Rapid flashing - Alert
Channel Select Keys
Cassette
Portion of administration set, inserts into cassette holder.
4
INTRODUCTION
AC Adapter Power Cord 120V/60 Hz three-pronged
grounded adapter with 4 pin locking connector, standard on Model 2863.
Connector Panel
External Power
External power receptacle con­nects with power cord.
Plug Symbol
Green light on indicates AC power is connected; batteries are charging.
COMM
Communications line receptacle connects with RS-232.
NOT E:
When inserting or removing connectors to the receptacles, avoid excessive force or twisting.
Container Hook
One hook on each side of the instrument.
Rotating Latch
Allows clamp to spin 360° and position at every 90°.
Adjustable Pole Clamp
Jaw with clutch feature, mounts pump to a pole or bedside.
Attaching Pole Clamp
To attach the pole clamp, position the clamp jaw over the mounting surface and turn the knob until the clamp is tightened and the pump feels secure. When the knob is as tight as possible, continued turning will make it click and spin freely without over-tightening.
5
INTRODUCTION
To avoid possible injury to the patient, observe the following precautions:
Epidural Administration
The MedSystem III pump can be used for epidural administration of anesthetic and analgesic drugs. This application is only appropriate when using anesthetic and analgesic drugs labeled for continuous epidural administration and catheters intended specifically for epidural use. Use only an IVAC 28 or 25 series set
wwiitthhoouutt
a ‘Y’ connector or injection port, for epidural infusions. The pump‘s secondary mode must
nnoott
be used when the pump is being used for
epidural administration of anesthetic and analgesic drugs.
- Epidural administration of anesthetic drugs — Use indwelling catheters specifically indicated for short-term (96 hours or less) anesthetic epidural drug delivery.
- Epidural administration of analgesic drugs — Use indwelling catheters specifically indicated for either short-term or long-term analgesic epidural drug delivery.
Administration Sets
• Use only IVAC 28 and 25 Series Administration Sets. The use of other sets will cause improper pump operation.
• Do not use the set if damaged.
• Do not insert a cassette into a channel with a Service status.
• Remove any cassettes from channel(s) requiring service.
• Ensure the cassette is properly installed before starting infusions.
Electromagnetic and Radio Frequencies
Operating the pump near equipment which radiates high-energy electromagnetic and radio frequencies (electrosurgical/cauterizing equipment, portable radios, cellular telephones, etc.) may cause false alarm conditions. If this happens, reposition the pump away from the source of interference; or turn off the pump and manually regulate the flow with the administration set regulating clamp.

Operational Precautions

Patient Precautions
WARNING:Epidural
administration of drugs other than those indicated for epidural use could result in serious injury to the patient.
WARNING: It is strongly
recommended that the infusion pump source container and administration set used for epidural drug delivery be clearly differentiated from those used for other types of administration.
6
INTRODUCTION
Artifacts
It is normal for intravenous infusion devices to produce non­hazardous currents when infusing electrolytes. These currents vary at a rate proportional to the infusion device flow rate. When the ECG monitoring system is not functioning under optimal conditions, these currents may appear as artifacts, simulating actual ECG readings. To determine if ECG abnormalities are caused by patient condition or the ECG equipment, place the infusion device on hold. If the ECG readings become normal, the ECG equipment requires attention. Proper setup of the ECG equipment should eliminate these artifacts. Reference the appropriate ECG monitoring system documentation for instructions on setup and maintenance.
Dropping/Jarring
If a pump is dropped or severely jarred, it should be immediately taken out of service and inspected by qualified service personnel to ensure proper function prior to reuse.
To ensure proper performance of the pump and to reduce potential injury to the operator, observe the following precautions:
• The power cord must be connected to a properly grounded, 3-wire receptacle (“Hospital Use” or “Hospital Grade”)
• Avoid excessive force or twisting of detachable power cords, when inserting or removing connector terminals.
• Use power cord indoors only.
• Disconnect AC and battery power when performing maintenance.
• Do not use the pump in the presence of flammable anesthetics.
• Do not open the instrument case. The case should only be opened by qualified service personnel using proper grounding techniques.
• Do not stack instruments on top of each other.
User Precautions
WARNING: When the
case is opened, an electrical shock hazard exists which can result in serious injury to persons and instrument component damage.
7
THERE IS
NO
PRINTING
ON THIS
PAGE

BASIC OPERATION

9
BASIC OPERATION
Prepare solution container in accordance with the manufacturers’ instructions.
• A syringe can be used as the container for the IV fluid to be infused. Syringe sizes from 20cc to 60cc of the B-D and Monoject
®
brands can be used.
NOTE:
The IVAC Model 8631A Syringe Holder is available as an accessory that provides a convenient place to hold syringes while they are being used as containers for IV fluid. The Syringe Holder is designed to be easily installed and removed from the top of the pump and to support up to three syringes. Do not use the Syringe Holder as a handle to carry the pump.
Connect the container to the IV set.
Prime the IVAC 28 or 25 Series administration set in accordance with Administration Set Directions for Use.
It is important to prime the set properly to eliminate air bubbles. Ensure the cassette slide clamp is pushed in completely so tubing is not
occluded. Invert the cassette so tubing is up. Slowly open the regulating clamp
and establish fluid flow to fully prime the set. Gently tap the cassette and ‘Y’ sites as necessary to remove all air. Gently massage the pressure dome to ensure no air bubbles are trapped.
Close regulating clamp before inserting and removing the cassette to reduce the risk of free flow.
Press to turn pump on.
With tubing down, use a 45 degree, upward motion to insert cassette into channel.
Push on clear portion of cassette until completely seated. Then push in slide clamp flush with entire cassette.
Pull down gently on tubing collar. Press with thumb to seat tubing collar in recess beneath cassette.
NOTE: Three beeps sound when inserted properly.

Preparing the Infusion

Loading the Set

Preparing the Administration Set

WARNING: An open
regulating clamp and slide clamp can cause a free­flow condition and may result in serious injury to the patient.
Pressure Dome
Slide Clamp
Tubing Collar
Clear portion of cassette
A
A
A
10
BASIC OPERATION
Status Line
Displays infusion status
(Infusing; Stopped; Standby; KVO; ALARM; FAULT; SERVICE)
for each channel.
NOTE:
S
tatus line in selected
channel is highlighted.
Infusion Rate
Volume Remaining
Volume Infused
Prompt Line
Displays messages that prompt the user to make programming choices and/or take appropriate actions.
Softkeys Prompts
Displays function of specific softkey.
STNDBY – Appears in softkey
information line when
is
pressed during infusion.
Cntrst – (Contrast) Brightens
or dims display.
GP II – When pressed, indi-
cates full name of selected Device Type on the prompt line.
NOT E:
Additional softkey prompts are displayed by pressing More Options.
Softkey Pads (4)

Front Panel Overview

Instrument Control Keys
ON/OFF Recharge Key
Turns the pump on and off.
STANDARD DISPLAY Key
Allows the user to display Standard Display page to view infusion settings for all channels.
MORE OPTIONS Key
Allows the user to display additional softkey functions.
START/STOP Key
Starts or stops infusion on selected channel.
Channel Control Key
Standard Display Page
ON/OFF
STANDARD DISPLAY
MORE
OPTIONS
START
STOP
Stopped
125
VR: 996.2 VI: 12.8
Start affects channel C
STNDBY Cntrst GP
®
ml/h
MedSystem III® DLE
Standby
25
ml/h VR: 138.8 VI: 26.9
A
B
Standby
95
ml/h
VR: 93.2
VI: 16.8
C
BASIC OPERATION
11
Selected channel is indicated by the letter displayed at the beginning of the first five lines.
Status Line
Displays infusion status for se­lected channel.
Infusion Rate
Volume Remaining
Time Remaining
Volume Infused Date/Time
Displays when volume infused was last cleared and infusion began.
Prompt Line
Displays messages that prompt the user to make programming choices and/or take appropriate action.
Soft Key Prompts
Displays function of specific softkey.
Select – Moves highlight bar
through the programmable infusion parameters.
– Increases highlighted value.
– Decreases highlighted value.
Fast
– Increases highlighted
value at greater increments.
Fast
– Decreases highlighted
value at greater increments.
Programming Page
ON/OFF
®
STANDARD DISPLAY
MORE
OPTIONS
MedSystem III® DLE
A: Stopped
A: Primary Rate 25 ml/h A: Pri VolRem (VR) 250 ml
A: Pri Time(TR) 10h 00m A: Pri VolInf (VI) 10 ml
since 12:37p 01 Feb 02
Press Select to choose line
Select Fast
A
B
C
START
STOP
12
BASIC OPERATION
Press .
• Instrument information page is momentarily displayed.
• Continuing to hold down ON/OFF key will keep the information on the display.
• When the ON/OFF key is released, the Standard Display page is displayed.
Press and hold .
• Display disappears.
• Pump is turned off.
Press .
• Standard Display page is displayed.
With the Standard Display page displayed:
Press
,
once.
TotVol, Device, Config, and Note softkeys appear.
Press again.
Batlog and DemoWd softkeys appear.
Press , , or .
• Selected channel programming page is displayed.
With programming page displayed: Press
Select to choose value to change.
• Value is highlighted.
Scroll through values using
, ↓, FAST, or FAST.
and FASTincreases highlighted values in single or multiple
increments.
To turn pump on
To activate additional Standard Display softkey prompts
To turn pump off
To view infusion settings for all active channels
To select channel and display Programming Pages
To program infusion
BASIC OPERATION
13
• ↓and FASTdecreases highlighted values in single or multiple increments.
• Pressing
or ↓changes direction of the FASTor FAST
• Highlight remains flashing until Enter is pressed. If Enter is not
pressed, the entry incomplete advisory will sound.
Press
Enter to accept new value.
• Highlight moves to next programmable value if channel status is
Stopped or Standby.
• If status is
Infusing, highlight remains on selected value.
To recall a previous value after a new value is introduced but not entered, press .
• Recall soft key appears.
Press
Recall.
• Number returns to previous value.
Press .
• Infusion starts or stops immediately, unless the channel’s programming is incomplete, or if an advisory, alarm, or fault condition exists on selected channel.
ALARM is displayed in affected channel status line.
• Alarm condition is displayed on the Standard Display of the affected channel.
Press affected channel , , or .
• Alarm Information page is displayed for that channel.
With the programming page displayed:
Press .
Press
2°Sec to access Secondary page.
Press
CalcOn to access Dose Rate Calculation page.
Press .
Press
CalcOff to discontinue use of the Dose Rate Calculator.
To access alarm information
To activate additional Programming Page softkeys
See the TROUBLESHOOTING section of this manual for more alarm information.
See the ADVANCED OPERATION section of this manual for information on the use of the Dose Rate Calculator function.
Loading...
+ 34 hidden pages