Alaris 8300 EtCO2 Service manual

Page 1
A
L
A
R
M
M
O
O
R
S
T
A
N
D
B
Y
CHANNEL
SELECT
MONITOR
CHANNEL
OFF
EtCO (mm Hg)
2
RR (breaths/min)
Technical Service Manual
Alaris®EtCO2Module, 8300 Series
Alaris®Products
Supports: Guardrails®Suite (v7 or later)
January 2006
Page 2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
This Technical Service Manual is subject to change without notification.

General Contact Information

Customer Advocacy - North America
Clinical and technical feedback.
Phone: 800.854.7128, Ext. 7812
E-Mail: CustomerFeedback@cardinal.com
Technical Support - North America
Maintenance and service information support; troubleshooting.
United States:
Phone:
858.458.6003
800.854.7128, Ext. 6003
Canada:
Phone:
Eastern: 800.908.9918 Western: 800.908.9919
Customer Care - North America
Instrument return, service assistance, and order placement.
United States:
Phone: 800.482.4822
Canada:
Phone: 800.387.8309
Cardinal Health Alaris®Products
10221 Wateridge Circle
San Diego, California 92121 http://www.cardinal.com/alaris
Page 3
Chapter 1 - General Information
1.1 Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 Precaution Definitions
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.3 Specifications
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.4 Operating Features, Controls and Indicators
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-2
1.5 Alarms, Errors, Messages
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.5.1 Silencing Alarms
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Chapter 2 - Checkout and Configuration
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 New Instrument Checkout
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-1
2.3 Configuration Options and Defaults
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.4 Configuration Setup - EtCO
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4.1 Alarm Limits
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4.2 Limit Mode
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.4.3 Waveform Height
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.4.4 Waveform Time Scale
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Chapter 3 - Preventive Maintenance
3.1 Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Regular and Preventive Maintenance Inspections, Calibrations
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.3 Cleaning
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Chapter 4 - Principles of Operation
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 General Information
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

TABLE OF CONTENTS

i
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Page 4
Chapter 5 - Corrective Maintenance
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Disassembly/Reassembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2.1 Latch Assembly and Feet
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.2.2 IUI Connectors
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.2.3 Rear Case
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
5.2.4 Frame Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5.2.5 Logic Board
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5.2.6 Power Board/Oridion Module
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.2.7 IUI Bracket
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5.2.8 Gas Inlet/Outlet Door
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
5.2.9 Luer Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
5.2.10 Display Board
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
5.2.11 Status Indicator Lens
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Chapter 6 - Troubleshooting
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Chapter 7 - Illustrated Part
s Breakdown
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Illustrations
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.3 Parts List
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.4 Ordering Parts
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
ii
TABLE OF CONTENTS
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Page 5
TABLE OF CONTENTS
List of Tables
2-1 Record of Configured Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
5-1 Required Materials, Supplies and Tools
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5-2 Torque Values
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
5-3 Level of Testing Guidelines
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
6-1 Technical Troubleshooting Guide
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6-2 Malfunction Codes
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
7-1 Parts List
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
List of Figures
7-1 Housing Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7-2 Frame/Front Case Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
7-3 IUI Bracket/Frame Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
7-4 Oridion Module/Power Supply Board
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
7-5 Front Case Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
7-6 Rear Case Assembly/Latch Kit
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
7-7 EtCO
2
Cable Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
iii
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Page 6
iv
TABLE OF CONTENTS
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
THIS PAGE
INTENTIONALLY
LEFT BLANK
Page 7
1 GENERAL INFORMATION
Page 8
Page 9
Chapter 1 — GENERAL INFORMATION
1-1Alaris
®
EtCO2Module (8300 Series)
Technical Service Manual
1.1 INTRODUCTION
This manual describes how to service the Alaris
®
EtCO2module 8300 Series (“EtCO
2
Module”). Use this manual in conjunction with the following Alaris®System documents and software:
Alaris
®
PC point-of-care unit 8000 Series (“PC Unit”) / Alaris®Pump module 8100 Series (“Pump Module”) Technical Service Manual
The PC Unit section of the Alaris
®
System
Directions for Use (
DFU)
The EtCO
2
Module section of the Alaris
®
System DFU
Maintenance Software User Manual (version
7 or later)
This manual is intended for personnel experienced in analysis, troubleshooting, and repair of analog/digital microprocessor­based medical equipment.
The EtCO
2
Module is a capnograph
indicated for continuous, noninvasive monitoring of end tidal carbon dioxide (EtCO
2
), fractional inspired carbon dioxide
(FiCO
2
) and respiratory rate (
RR). The
EtCO
2
Module and disposables are
indicated for use with intubated and nonintubated adult, pediatric, and neonatal patients. It is not intended for direct connection to ventilator or breathing systems. Only one (
1) EtCO
2
Module can be
connected to the Alaris
®
System.
The EtCO
2
Module is used with Oridion’s
patented Microstream
®
Disposables/circuits
for sidestream capnography.
To avoid damaging the keypad, do not use sharp objects (pens, pencils, etc.) to activate switches.
CAUTION
Any attempt to service this product by anyone other than an authorized Cardinal Health Service Representative while the product is under warranty may invalidate the warranty.
CAUTION
Page 10
1.2 PRECAUTION DEFINITIONS
Refer to the Alaris
®
System DFU
.
1.3 SPECIFICATIONS
Refer to the product-specific (EtCO
2
Module) section of the Alaris®System
DFU.
1.4 OPERATING FEATURES, CONTROLS AND INDICATORS
Refer to the product-specific section of the Alaris
®
System DFU
.
1.5 ALARMS, ERRORS, MESSAGES
The scrolling Channel Message Display bar shows alarms and other messages. Refer to the product-specific (EtCO
2
Module)
section of the Alaris
®
System
DFU for
detailed information.
1.5.1 Silencing Alarms
All alarms can be temporarily silenced by pressing the
SILENCE key on the PC Unit.
If an alarm condition on the EtCO2Module occurs while the audio alarm is silenced, the
only alarm indications will be visual displays and symbols related to the alarm condition.
WARNING
GENERAL INFORMATION
1-2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Page 11
2 CHECKOUT & CONFIGURATION
Page 12
Page 13
Chapter 2 — CHECKOUT AND CONFIGURATION
2-1Alaris
®
EtCO2Module (8300 Series)
Technical Service Manual
2.1 INTRODUCTION
This chapter describes the initial setup and configuration for the EtCO
2
Module.
2.2 NEW INSTRUMENT CHECKOUT
Refer to the EtCO
2
Module Directions for
Use (
DFU
) for instructions regarding unpacking and setting up the instrument for first time use.
When powering up the instrument, verify the instrument beeps and all display
LED
segments flash. This confirms that the instrument has performed its self test and is operating correctly. During operation, the instrument continually performs a self test, and will alarm and display a message if it detects an internal malfunction.
Contact Cardinal Health authorized service personnel if the instrument has physical damage, fails to satisfactorily pass the startup sequence, fails a self test, or continues to alarm.
For new instrument checkout, the minimum checks (described in the Maintenance Software User Manual) are:
Regular Inspection
Functional Tests
Should an instrument be jarred severely or dropped, remove it from use immediately. It should be thoroughly tested and inspected by qualified service personnel to ensure proper function prior to reuse.
CAUTION
Page 14
2.3 CONFIGURATION OPTIONS AND DEFAULTS
NOTES:
Changes to the factory default values are retained after a power cycle.
If Factory Default is Yes, then all the configuration settings are set to their factory default.
If Factory Default is No, then one or more of the configuration settings has been changed. If desired, Factory Default can be selected and set to Yes, which will set all configuration settings to their factory default.
With the Profiles feature enabled, the settings are configured independently for each profile. A hospital-defined, best­practice data set must be uploaded to enable the Profiles feature. Date and Time is a system setting and is the same in all profiles.
System Settings: Refer to the PC Unit section of the Alaris®System
DFU for
information on system settings (including alarm audio profile, battery meter enable/disable, clock setup, key click audio enable, tamper resist enable/disable).
EtCO
2
Module Settings: Refer to the
product-specific (EtCO
2
Module) section of
the Alaris
®
System DFU for information on
EtCO
2
Module settings (including limit
mode, high and low respiratory rate alarm limit, high and low EtCO
2
alarm limit, no
breath alarm setting, and high FiCO
2
alarm
limit).
NOTES:
• Pressing the
EXIT soft key while in a
System Config - Module screen
immediately powers the system down, with no “Powering Down” display.
• Pressing the
EXIT soft key while in a
System Config - EtCO
2
screen returns the
display to the main System Config - Module screen.
• Pressing the
CONFIRM soft key while in a
System Configuration option screen:
accepts existing setting or setting
change
displays next option setting screen (if
applicable) or returns display to System Config - EtCO
2
screen
• Pressing the Point-of-Care Unit’s
CANCEL
key while in a System Configuration option screen:
leaves setting unchangedreturns display to System Config -
EtCO
2
screen
CHECKOUT AND CONFIGURATION
2-2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Page 15
To access System Configuration options:
NOTE: Refer to EtCO2Module DFU for alarm settings. If profiles are enabled, those
settings are set by the profile.
1. Hold OPTIONS key at power up.
2. Press
PAGE DOWN soft key.
3. Press EtCO
2
Module soft key.
2.4.1 Alarm Limits
NOTE: The following examples show adult
alarm limits for illustration. Follow the same steps for neonatal alarm limits.
1. From the System Config - EtCO
2
display, press the Adult alarm limits or Neonatal alarm limits soft key.
2. To change an alarm limit, press the soft key next to that parameter. The selected parameter is highlighted.
CHECKOUT AND CONFIGURATION
2-3
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
2.4 CONFIGURATION SETUP
Page 16
3. Enter a numeric value for the selected alarm limit. Use the numeric keypad or the up or down arrow keys to enter the value. If a valid value appears in the field for three seconds, the display shows the prompt > Press CONFIRM to Apply Changes.
4. Press ENTER on the Point-of-Care Unit to confirm the value. Once you press ENTER, the next limit is highlighted and the display prompts for an entry.
5. To change another alarm limit, press the soft key for that parameter and enter a new value.
6. Once all changes are complete, press the CONFIRM soft key.
2.4.2 Limit Mode
NOTES:
The following examples show adult alarm limits for illustration. Follow the same steps for neonatal alarm limits.
If a profile is in use for programming, you cannot change the Limit Mode.
1. To change the Limit Mode, toggle its value from the System Config - EtCO
2
display, or use the Limits mode soft key.
2. To toggle the value: From the System
Config - EtCO
2
display, press the soft
key for the Limit Mode selection. Every key press toggles the value between Adult and Neonatal.
3. To use the soft key: From the System
Config - EtCO
2
display, press the
Limits Mode soft key.
CHECKOUT AND CONFIGURATION
2-4
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
2.4 CONFIGURATION SETUP (Continued)
2.4.1 Alarm Limits (Continued)
Press the Limits Mode soft key to
toggle between the two acceptable values
Page 17
4. Press the soft key to select the Limit Mode Setup display, or press CANCEL to exit the System Config - EtCO
2
display without changing the value.
2.4.3 Waveform Height
1. To change the waveform height, you can toggle its value from the System Config
- EtCO
2
display, or you can use the
Waveform height soft key.
2. To toggle the value: From the System
Config - EtCO
2
display, press the soft
key for the waveform height value.
Every key press toggles the value between 60 and 99 mmHg.
3. To use the soft key: From the System
Config - EtCO
2
display, press the
Waveform height soft key.
4. Press the soft key to select the waveform height, or press CANCEL to exit the Waveform Height display without changing the value.
2.4.4 Waveform Time Scale
1. To change the waveform time scale, you can toggle its value from the System
Config - EtCO
2
display, or you can use
the Waveform time scale soft key.
CHECKOUT AND CONFIGURATION
2-5
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
2.4 CONFIGURATION SETUP (Continued)
2.4.2 Limit Mode (Continued)
Press the
Waveform height
soft key to toggle between the two acceptable values
Press the
Waveform time scale soft key to
toggle between the two acceptable values
Page 18
CHECKOUT AND CONFIGURATION
2-6
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
2. To toggle the value: From the System
Config - EtCO2display, press the soft
key for the Waveform Time Scale value. Every key press toggles the value between 5 and 10 sec.
3. To use the soft key: From the System
Config - EtCO
2
display, press the
Waveform time scale soft key.
4. Press the soft key to select the waveform height, or press CANCEL to exit the Waveform time scale display without changing the value.
2.4 CONFIGURATION SETUP (Continued)
2.4.4 Waveform Time Scale (Continued)
Page 19
CHECKOUT AND CONFIGURATION
2-7
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
EtCO2Alarm Limit, High
EtCO
2
Alarm Limit, Low
FiCO
2
Alarm Limit, High
Limit Mode No Breath Alarm Respiratory Rate Alarm
Limit, High Respiratory Rate Alarm
Limit, Low Waveform Height Waveform Time Scale
Table 2-1. Record of Configured Instruments
#1 #2 #3 #4
Instrument ID/Serial #
NOTE: If profiles are enabled, alarm limits are set by the profile.
Page 20
CHECKOUT AND CONFIGURATION
2-8
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
THIS PAGE
INTENTIONALLY
LEFT BLANK
Page 21
3 PREVENTIVE MAINTENANCE
Page 22
Page 23
Chapter 3 — PREVENTIVE MAINTENANCE
3-1Alaris
®
EtCO2Module (8300 Series)
Technical Service Manual
Failure to perform regular and preventive maintenance inspections may result in improper instrument operation.
WARNING
3.1 INTRODUCTION
Perform regular and preventive maintenance inspections to ensure that the EtCO
2
Module remains in good operating
condition:
Perform regular inspections before each use.
Perform preventive maintenance inspections annually.
These requirements and guidelines are intended to complement the intent of Joint Commission on Accreditation of Healthcare Organizations (
JCAHO) requirements.
3.2 REGULAR AND PREVENTIVE MAINTENANCE INSPECTIONS, CALIBRA TION
Use the Maintenance Software (version 7 or later) to perform calibration and preventive maintenance.
Contact Cardinal Health Technical Support if you need help obtaining or using the Maintenance Software.
3.3 CLEANING
Refer to the PC Unit section of the Alaris
®
System DFU.
Page 24
PREVENTIVE MAINTENANCE
3-2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
THIS PAGE
INTENTIONALLY
LEFT BLANK
Page 25
4 PRINCIPLES OF OPERATION
Page 26
Page 27
Chapter 4 — PRINCIPLES OF OPERATION
4-1Alaris
®
EtCO2Module (8300 Series)
Technical Service Manual
4.1 INTRODUCTION
This chapter describes the principles of operation for the EtCO
2
Module and its
major subsystems. Refer to the PC Unit/Pump Module
Technical Service Manual for Point-of-Care Unit information.
4.2 GENERAL INFORMATION
The EtCO
2
Module uses Oridion’s patented
Microstream
®
non-dispersive infrared (NDIR) spectroscopy to continuously measure the amount of CO
2
during every breath, the
amount of CO
2
present at the end of
exhalation (EtCO
2
) and during inhalation
(FiCO
2
), and the Respiratory Rate. The
EtCO
2
Module is a side stream capnograph.
The Microstream
®
Disposables deliver a sample of the inhaled and exhaled gases from the ventilator disposable or directly from the patient (via an oral/nasal cannula) into the monitor for CO
2
measurement. Moisture and patient secretions are extracted from the sample by the Microstream
®
inline filter while maintaining
the shape of the CO
2
waveform.
The 50 mL/min sampling flow rate reduces liquid and secretion accumulation, decreasing the risk of obstruction in the sample pathway in humid ICU environments. The small sample size required by Microstream
®
eliminates the need for water traps and prevents excess fluid accumulation.
The EtCO
2
Module draws a gas sample through a microsample cell (15 microliters). This extremely small volume is quickly flushed, allowing for a rise time of
Page 28
PRINCIPLES OF OPERATION
4-2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
approximately 190 ms and accurate CO
2
readings, even at high respiration rates. The Microbeam IR source illuminates the
microsample cell and the reference channel. This proprietary IR light source generates only the specific wavelengths characteristic of the CO
2
absorption spectrum. The IR light that passes through the microsample cell and the IR light that passes through the reference channel are measured by IR detectors.
The microcomputer in the EtCO
2
Module
calculates the CO
2
concentration by
comparing the signals from both channels. No operator intervention is required for
routine moisture or condensate. All Microstream
®
Disposables contain an inline hydrophobic filter to extract condensate and/or patient secretions while maintaining measurement and waveform integrity. For humid conditions within the operating parameters of the EtCO
2
Module
and Microstream
®
Disposables, humidity
has no quantitative effect on the CO
2
concentration, given the small 50 mL/min sample size rate. In high humidity environments or extended monitoring periods (24-72 hours) only Microstream
®
Disposables designed for those instances should be used. In the event of humidity or condensate outside the EtCO
2
Module’s
operating specifications, the EtCO
2
Module will present a “Remove Blocked Disposable” message.
Due to the relatively small sampling size needed for EtCO
2
readings, partial pressure
does not affect the ability of the EtCO
2
Module to measure EtCO2, as long as the 50 mL/min rate can be achieved.
Microstream
®
Disposables are single-use, disposable devices which must be changed with each use. The manufacturer’s sample flow, 50 mL/minute, does not affect the disposable’s life; however, humidity and specific patient conditions may shorten the effective life of the disposables. Microstream
®
Disposables are rated for up to 24 hour and 72 hour use, depending on the specific Microstream
®
Disposable.
The EtCO
2
Module provides readings in compliance with BTPS (body temperature, pressure, saturation) standards. There is no effect on accuracy due to cyclic pressure up to 10 kPa.
NOTE: BTPS (body temperature, pressure, saturation assumed 37°C, 47 mmHg) calculations are made according to:
PCO2= FCO2X (Pb - 47) Where:
FCO2is the fractional concentration of CO
2
in dry gas, and FCO2= % CO2/100.
Pb is the ambient pressure. PCO2is the partial pressure of CO2at
BTPS.
4.2 GENERAL INFORMATION (Continued)
Page 29
5 CORRECTIVE MAINTENANCE
Page 30
Page 31
NOTE: Due to product changes over time, components/ assemblies illustrated in this chapter may differ from the instrument you disassemble. If there are any questions, look for Service Bulletins related to this chapter or contact Cardinal Health Technical Support.
5.1 INTRODUCTION
This chapter describes how to disassemble and reassemble the EtCO
2
Module.
For more efficient repair, read the “Principles of Operation” chapter for information on the mechanical and electrical functions of the EtCO
2
Module.
The surface mount devices on the circuit boards are not field repairable. Return circuit boards to an authorized Cardinal Health Service Center for repair. Attempting circuit board repairs voids all warranties.
Whenever the EtCO2Module is repaired, perform the minimum tests required for new
instrument checkout (use Maintenance Software v
7 or later).
For information on replacement parts, see the “Illustrated Parts Breakdown” chapter.
Following any level of maintenance, perform the applicable tests (refer to “Level of Testing Guidelines” table).
5-1
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Chapter 5 — CORRECTIVE MAINTENANCE
Disconnect the EtCO2Module from the Alaris
®
System before performing maintenance. qualified personnel using proper grounding techniques should open the instrument case.
WARNING
CMOS devices are sensitive to static electrical
charges and may be damaged during repair if the repair activity is not performed in an electrostatic discharge (
ESD) protected environment using
approved ESD protective procedures, including personnel grounding.
CAUTION
Page 32
Follow these procedures in order for the most efficient disassembly of EtCO
2
Module
subassemblies. To reassemble the EtCO
2
Module, reverse the disassembly steps.
NOTE: Perform all maintenance on an antistatic surface, preferably a grounded antistatic mat.
Before adhering gaskets and labels to the instrument, clean the surface with a cotton swab or soft cloth lightly dampened with
70% Isopropyl Alcohol.
5-2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
Disconnect the EtCO2Module from the Alaris
®
System Before attempting
disassembly.
CAUTION
To avoid the risk of electrical hazard or damage to the instrument circuitry, do not spray fluids directly onto the instrument or allow fluids to enter the instrument.
CAUTION
Table 5-1. Required Materials, Supplies and Tools
NOTE: Contact/source information is subject to change.
Loctite 242 (http://www.loctite.com)
Silicone grease, Dow Corning Molykote
33, or equivalent (http://www.dowcorning.com)
#1 Phillips screwdriver
#2
Phillips screwdriver
Small diagonal cutters
Lint-free cloth (such as, Kimwipes or lint-free tissue)
Page 33
5.2.1 Latch Assembly and Feet
1. Remove two (
2) screws attaching Latch
Assembly to bottom of Rear Case and remove Latch Assembly components.
2. Pull two (2) Feet from underside of module.
NOTE: Pay close attention to Compression Spring location and orientation to ensure proper installation during reassembly.
During Reassembly:
When reinstalling the Latch Assembly, press toward outer edge of the module and screw into position.
Apply a thin layer of Dow Corning Molykote
33 (or equivalent) silicone grease
to Feet.
NOTE: The Feet press-fit into the module.
5-3
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
Feet
Rear Case
Latch Assembly
components
5.2 DISASSEMBLY/REASSEMBLY (Continued)
Compression Spring
Page 34
5-4
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY (Continued)
5.2.2 IUI Connectors
Remove two (
2) screws attaching each IUI
(left and right) to module.
During Reassembly:
Ensure ground clips are still installed on both IUI connectors.
To install IUI Connector Gasket, remove protective backing and adhere to IUI Connector.
IUI Connector, Left
IUI Connector, Right
IUI Connector Seal, Right
IUI Connector Gasket, Right
Ground clip located at each mounting hole on both IUI Connectors
Page 35
5.2.3 Rear Case
1. Remove two (
2) screws (and associated
washers) attaching Rear Case to chassis assembly.
2. Pull Rear Case away from chassis assembly.
During Reassembly:
Ensure that Silicone Tubing gasket in Rear Case is in place and undamaged.
5.2 DISASSEMBLY/REASSEMBLY (Continued)
Rear Case Assembly
Silicone Tubing
gasket
5-5
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
Screws and
Washers (x2)
Page 36
5.2.4 Frame Assembly
1. Remove six (6) screws attaching Frame Assembly to Front Case.
2. Use small diagonal cutters to remove upper snap rivet attaching Power Board to Frame Assembly.
2. Remove the IUI Bracket Strap from the Display Board.
3. Carefully remove Frame Assembly from Front Case Assembly, then place the Frame Assembly to the left of the Front Case, taking care not to damage hoses and wires.
4. Disconnect electrical harness from Oridion Module.
5. Disconnect the inlet and exhaust hoses on the Oridion Module.
During Reassembly:
Fit the top flange of the Frame Assembly
into the square slot on the Status Indicator Lens.
Fit the connectors on the Power Board
and Logic Board into the two (2) connectors on the Display Board.
Ensure that the harness and hoses are
fully connected, and route them so they are not pinched during reinstallation.
Slide the IUI Bracket Strap under the
keypad ground wire on the Display Board.
5-6
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY (Continued)
Frame
Assembly
Exhaust hose
IUI
Bracket
Strap
Snap Rivet
Front Case
Assembly
Electrical
Harness
When using cutters to remove snap rivet, take care not to cut the rivet.
CAUTION
Inlet hose
Screw
Keypad
ground wire
IUI
Bracket
Strap
Keypad ground
wire
Page 37
5-7
5.2 DISASSEMBLY/REASSEMBLY (Continued)
5.2.5 Logic Board
1. Use small diagonal cutters to remove snap rivet attaching Logic Board to Frame Assembly .
2. Remove Logic Board from Frame Assembly.
During Reassembly:
Seat the holes in the Logic Board over the standoffs on the IUI Bracket.
Reinstall the snap rivet and its locking pin to secure the Logic Board to the Frame Assembly.
Snap Rivet
Logic
Board
Standoff on
IUI Bracket
When using cutters to remove snap rivet, take care not to cut the rivet.
CAUTION
Frame
Assembly
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
Page 38
5-8
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2.6 Power Board/Oridion Module
1. Use small diagonal cutters to remove upper (if installed) and lower snap rivet(s) attaching Power Board to Frame Assembly. If installed, remove IUI Bracket Strap.
2. Remove screw holding PCB retainer to Frame Assembly .
3. Remove Power Board from Frame Assembly.
4. Disconnect electrical harness between Power Board and Oridion Module.
5. Separate Power Board and Oridion Module:
Remove three (3) screws from
underside of Power Board.
Remove screw and nut attaching
PCB retainer to Power Board.
During Reassembly:
Ensure that Oridion Module and PCB
retainer are positioned correctly over the Power Board before reinstalling.
Reinstall the IUI Bracket Strap and snap
rivets to secure the Power Board to the Frame Assembly .
5.2 DISASSEMBLY/REASSEMBLY (Continued)
PCB retainer
Power
Board
Oridion
Module
When using cutters to remove snap rivet, take care not to cut the rivet.
CAUTION
Page 39
5-9
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2.7 IUI Bracket
Flex IUI Bracket away from key pins and remove bracket and Ground Strap from Frame Assembly .
During Reassembly:
Position IUI Bracket over key pins on Frame Assembly
5.2 DISASSEMBLY/REASSEMBLY (Continued)
Key pin
Frame
Assembly
IUI Bracket
Ground
Strap
Page 40
5-10
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY (Continued)
5.2.8 Gas Inlet/Outlet Door
NOTE: To hold Torsion Spring in place
during removal, hold Spring Retainer firmly when removing Cotter Pin.
1. From inside Front Case Assembly, straighten Cotter Pin legs and from Spring Retainer.
2. Slowly lift Spring Retainer to relieve compression on Torsion Spring.
3. Remove Torsion Spring.
4. Remove Door from Front Case Assembly.
During Reassembly:
1. Place Torsion Spring on Door shaft with long end facing up.
2. Align Spring Retainer with shaft, push Torsion Spring down, insert Cotter Pin through Spring Retainer, then wind Cotter Pin legs around Door shaft to secure.
3. Wind Torsion Spring 1/4-turn and secure long end of Spring under Spring Retainer.
Door
Status
Indicator
Lens
Front Case
Display
Board
Spring
Retainer
Cotter
Pin
Spring
Retainer
Cotter
Pin
Torsion
Spring
Torsion
Spring
Page 41
5-11
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2.9 Luer Assembly
1. Remove three (
3) screws attaching Luer
Retainer to Front Case.
2. Remove Luer Retainer from Front Case.
3. Separate Luer Assembly from Luer Retainer.
During Reassembly:
Ensure that O-ring is correctly positioned on Luer Assembly, then apply a small bead of water around the O-ring: this helps O-ring to remain in place when installing Luer Assembly into Front Case.
Position Luer Retainer over screw holes in Front Case, then install screws to secure.
5.2 DISASSEMBLY/REASSEMBLY (Continued)
Front Case
O-ring
Luer
Assembly
Luer
Retainer
Page 42
5-12
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2.10 Display Board
1. Remove screw attaching ground wire to Display Board.
2. Cut cable tie, then disconnect Keypad Harness and Backlight harnesses from Display Board.
3. Remove three (3) screws attaching Display Board to Front Case.
4. Remove Display Board from Front Case
During Reassembly:
When reinstalling Display Board, slide LEDs under Status Indicator Lens.
Reinstall screws attaching Display Board to Front Case.
Reinstall Keypad and Backlight Harness, then replace the cable tie.
Reinstall the keypad ground wire.
5.2 DISASSEMBLY/REASSEMBLY (Continued)
Display
Board
Front Case
Status
Indicator
Lens
Keypad Ground
Wire
Keypad
Harness
Page 43
5-13
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2.11 Status Indicator Lens
1. Remove two (2) screws attaching Status Indicator Lens and Gasket to Front Case.
2. Remove Status Indicator Lens.
5.2 DISASSEMBLY/REASSEMBLY (Continued)
Status
Indicator
Lens
Front Case
Gasket
Screws (x2)
Page 44
Table 5-2. Torque Values
Functional Application Item Description Torque Value
FINAL ASSEMBLY
IUI Connectors 6-32 x
7
/16 12 in/lb
Latch Assembly 4-40 x
3
/8
Torx Security 6 in/lb
Rear Case - Internal Frame 6-32 x
7
/
16 12 in/lb
FRONT CASE ASSEMBLY
Display Board 4-40 x
5
/16
6 in/lb
Internal Frame - Front Case 4-40 x
3
/
4 6 in/lb
IUI Bracket Strap and Ground Wire 4-40 x
5
/16 6 in/lb
Status Indicator Lens 4-20 x
5
/16 6 in/lb
Luer Mount 4-20 x
5
/
16 6 in/lb
Power Board/Oridion Module 4-40 x
1
/8 6 in/lb
5-14
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
Page 45
5-15
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
Table 5-3. Level of Testing Guidelines
Display Board NNNN N N N Frame Assembly NNNNNN N N Front Case NNNNNN N N Gas Inlet/Outlet Door NNNNNN N N IUI Bracket NNX NNN N N IUI Connectors NN N Latch Assembly and Feet N Logic Board NNNNNN N Luer Assembly NNNNNN N N Power Board/Oridion Module NNN NN Rear Case NN Status Indicator Lens NNNNNN N N
Miscellaneous: Ð
Instrument dropped NNX NNN N N Instrument software flashed N New instrument checkout NNX NNN N N No fault found
N
(instrument not opened)
No fault found
NNNNNN N N
(instrument opened)
Channel ID/IUI Connector Test
CO
2
Sensor Calibration
Keypad Test
CO
2
Sensor Verification
Alarm Test
Tests to Perform
N
= Required X = If verification fails. Blank = Not Applicable
Repair/Replacement of: Ð
Simultaneous Display Test
Visual/Instrument Inspection
NOTES:
Use the Maintenance Software to perform testing and calibration.
• After performing any calibration procedure, always perform the associated verification test.
Leak Down Test
Page 46
THIS PAGE
INTENTIONALLY
LEFT BLANK
5-16
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
Page 47
6 TROUBLESHOOTING
Page 48
Page 49
Chapter 6 — TROUBLESHOOTING
6-1Alaris
®
EtCO2Module (8300 Series)
Technical Service Manual
6.1 INTRODUCTION
This chapter describes possible technical problems and operating errors that may occur while using the EtCO
2
Module. Refer to this chapter before attempting to service the EtCO
2
Module.
Refer to the PC Unit / Pump Module Technical Service Manual for information on viewing the Malfunction Log.
Use extreme caution in servicing the instrument when connected to AC power. Hazardous voltages are present when AC power is connected, regardless of the setting of the power switch.
WARNING
Page 50
TROUBLESHOOTING
6-2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Table 6-1. Technical Troubleshooting Guide
Follow the steps in the order they are listed until the problem/fault is corrected. Before making a final diagnosis, visually inspect the instrument for damage and pay particular attention to the power cord and plug. Following repair/replacement, use Maintenance Software to perform the required tests (see Chapter
5). Perform the Ground Current Leakage and Ground Leakage tests described in the PC Unit /
Pump Module Technical Service Manual to verify that the system is electrically safe.
Problem Remedy
Display Problem 1. Check cable connections.
2. Replace Display.
3. Replace Logic Board.
4. Return to factory.
Key Stuck Alarm 1. Turn instrument off and back on to see if problem clears.
2. Replace Keypad Assembly.
3. Replace Logic Board.
4. Return to factory.
Instrument Malfunction 1. Turn instrument off and back on to see if problem clears.
2. Refer to alarm history, malfunction codes (Table 6-2), and subsystem codes (see PC Unit / Pump Module (8000/8100 Series) Technical Service Manual).
Intermittent Operation 1. Check connections to Logic Board.
2. Replace Logic Board.
3. Return to factory.
Page 51
TROUBLESHOOTING
6-3
Table 6-2. Malfunction Codes
Refer to the PC Unit / Pump Module Technical Service Manual for information on viewing the Malfunction Log.
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Malfunction
Code
Subsystem Explanation Remedy
500 Main Logic Board See NOTE See NOTE 510 Keypad Decoder See NOTE See NOTE 511 Keypad Decoder Comm Error See NOTE See NOTE 530 Keypad Error See NOTE See NOTE 550 IUI Error See NOTE See NOTE 560 Power Supply Board Error See NOTE See NOTE 570 EtCO2Board Error See NOTE See NOTE 571 EtCO2Board Comm Error See NOTE See NOTE
6200 EtCO2Board Error Failed continuous built-in
tests.
Cycle power. If error repeats, replace EtCO
2
Board.
6210 EtCO2Board Error Communications error. Cycle power. If error repeats,
replace EtCO2Board.
6220 EtCO2Board Error Configuration error. Cycle power. If error repeats,
replace EtCO
2
Board.
6230 EtCO2Board Error Received value out of
range.
Replace Microstream Disposable device. If error repeats, replace EtCO
2
Board.
6240 EtCO2Board Error Missing sensor status. Replace Microstream
Disposable device. If error repeats, replace EtCO
2
Board.
NOTE: Refer to Troubleshooting information in the PC Unit / Pump Module Technical Service Manual for more detailed descriptions of codes.
Page 52
TROUBLESHOOTING
6-4
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
THIS PAGE
INTENTIONALLY
LEFT BLANK
Page 53
7 ILLUSTRATED PARTS
BREAKDOWN
Page 54
Page 55
Chapter 7 — ILLUSTRATED PARTS BREAKDOWN
7-1Alaris
®
EtCO2Module (8300 Series)
Technical Service Manual
7.1 INTRODUCTION
The illustrated parts breakdown for the EtCO
2
Module is divided according to its
major assemblies.
7.2 ILLUSTRATIONS
Use the exploded views to identify the parts of each assembly. Item numbers (the numbers in bubbles) in the illustration correspond to the item number in the parts list. Apart that does not have an item number is part of a higher assembly.
NOTE: Due to product changes over time, illustrations in this chapter can differ from the instrument under service. See applicable Service Bulletins or contact Cardinal Health Technical Support for more information.
7.3 PARTS LIST
The parts lists include the following information for replaceable parts :
Item: This number corresponds to the item number in the illustration.
Part Number: Use this number to order the part. If a part number is not provided, that part is not available for sale, is part of a kit or higher assembly, or can only be replaced/repaired by Cardinal Health authorized service personnel.
Description: Description of the part.
QTY: Total number of each item used.
Page 56
Item Part Number Description QTY
15 146980-000 Gasket, Status Indicator 1 16 148140-000 Retainer, Luer 3 17 321245 O-Ring, 0.364 ID x 0.070 W 60 Duro 1
20 10013666
Front Case/Keypad Assembly
(Assembled parts: Front Case, Keypad, items
15, 25, 841 qty. 2)
1
21 148128-102 Door Assembly 1 22 148130-002 Spring, Torsion 1 23 148228-000 Retainer, Spring 1 24 3211246 Pin, Cotter, 1/32 OD x 3/8 L, extended prong 1 25 146978-001 Lens, Status Indicator 1 32 144424-100 LED, Keyboard Backlight 1 35 148112-103 Board Assembly, Display 1 36 321270 Standoff, F/F, 3/16
hex x
3
/8
4-40, AL
3 37 10011768 Module, Oridion 1 39 TC10003682 Board Assembly, Power Supply 1 40 142833-000 Bracket, IUI 1 45 148131-001 Frame 1 46 148227-002 Retainer, PCB 1
7.4 ORDERING PARTS
Parts can be ordered by writing or calling Cardinal Health Customer Service (refer to "General Contract Information" page at beginning of this manual). When requesting a part, provide the following information:
Instrument name and model number; for example, EtCO
2
Module, Model 8300.
Instrument software version. Refer to Alaris
®
System DFU
for directions on how
to view software version.
Part number.
Part description, as provided in parts list.
For labels, specify required language.
ILLUSTRATED PARTS BREAKDOWN
7-2
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Table 7-1. Parts List
NOTE: An “assembly” is a preassembled group of parts. A “kit” is a group of unassembled parts.
Page 57
7-3
ILLUSTRATED PARTS BREAKDOWN
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Item Part Number Description QTY
47 148249-100
Assembly, EtCO2Cable
1 55 TC10003899 Board Assembly, Logic 1 60 146992-001 Strap, Ground, IUI Bracket 1 70
Silicone Grease, Molykote Medium or equivalent As required 75 142794-000 Foot, Rubber 2 80 320763 Tubing, Medium Silicone, 0.030 x 0.065 x 0.18 1
85 10013667 Assembly, Rear Case
(Assembled parts: Rear Case, Nameplate Label, items 75, 80, 603,
620
)
1
90 146456-000 Kit, Latch
(Includes: Latch, Compression Spring, Leaf Spring, Support)
1
120 147077-100 Kit, Right IUI Connector
(Includes: Right Seal, Right IUI Connector)
1
130 147078-100 Kit, Left IUI Connector
(Includes: Left Gasket, Left IUI Connector)
1
135 10013668 Kit, Luer
(Includes: Luer Assembly, item 17)
1
603 125569 Label, Serial Number, Replacement 1 620 148405-000 Label, Patent 1 630 148372-000 Label, Microstream 1 635 10010621 Label, Door 1 645 145820-000 Seal, Instrument 1 820 834031 Cable Tie, 4-in. L 1 825 300348 Screw, Machine, 4-40 x 3/4 PHH PHH 6 830 806112 Rivet, Snap, 0.125 dia., Nylon BLK 3 835 320855 Screw, 6-32 x 7/16 PHH PNH SS 6 840 320908 Washer, Flat, 0.146 ID x 0.270 OD x 0.03 THK, SS 10 841 300923 Screw, Machine, 4-40 x 5/
16 PHL PNH ZNC PLT STL
7
Table 7-1. Parts List (Continued)
Page 58
Table 7-1. Parts List (Continued)
ILLUSTRATED PARTS BREAKDOWN
7-4
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Item Part Number Description QTY
842 303822 Screw, 3LOB, 4-20 x 7/16
PNH PHH ZNC-STL
3
843 300347 Screw, Machine, 4-40 x 5/8
PNH PHH
1 844 300928 Nut, Hex, 4-40 STL PL 1 845 300345 Screw, Machine, 4-40 x 1/4 PHL PNH ZNC PLT STL 4 846 305121 Screw, Machine, PNH 6
10011281
Directions for Use, EtCO
2
1
Not sold by Cardinal Health.
Page 59
7-5
ILLUSTRATED PARTS BREAKDOWN
Figure 7-1. Housing Assembly
130
85
835
635
630
835
120
840
20
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
4 PL
2 PL
2 PL
Page 60
ILLUSTRATED PARTS BREAKDOWN
7-6
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Figure 7-2. Frame/Front Case Assembly
841
825
830
3 PL
6 PL
39
32
55
Page 61
7-7
ILLUSTRATED PARTS BREAKDOWN
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Figure 7-3. IUI Bracket/Frame Assembly
45
60
40
Page 62
3 PL
ILLUSTRATED PARTS BREAKDOWN
7-8
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Figure 7-4. Oridion Module/Power Supply Board
46
39
36
37
8 PL
843
844
840
6 PL
846
Page 63
7-9
ILLUSTRATED PARTS BREAKDOWN
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Figure 7-5. Front Case Assembly
15
25
21
35
16
17
23
24
135
842
3 PL
22
23
24
22
841
3 PL
841
2 PL
Page 64
ILLUSTRATED PARTS BREAKDOWN
7-10
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Figure 7-6. Rear Case Assembly/Latch Kit
85
620
603
Nameplate label
75
90
2 PL
80
1
Labels (items
603 and 620) are provided as separate items and as part of Rear
Case Assembly Kit. Reference Table 7-1, item 85 for a list of parts included in Rear Case Assembly Kit.
3
Before adhering Serial Number Label (item
603):
1. Print instrument model and serial number on label in permanent black ink.
2. Apply clear overlay to completely cover label.
2
Nameplate Label is available only as part of Rear Case Assembly Kit. This label has a regulatory mark, and any level of service that requires its replacement must be performed at the factory.
1
1
2
1
3
Page 65
7-11
ILLUSTRATED PARTS BREAKDOWN
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Figure 7-7. EtCO2Cable Assembly
47
Page 66
THIS PAGE
INTENTIONALLY
LEFT BLANK
ILLUSTRATED PARTS BREAKDOWN
7-12
Alaris®EtCO2Module (8300 Series)
Technical Service Manual
Page 67
Page 68
Alaris®and Guardrails®are registered trademarks of Cardinal Health, Inc. or one of its subsidiaries. Microstream®is a registered trademark of Oridion Medical 1987 Ltd. All other trademarks belong to their respective owners.
Part Number 10011051 ©2005, 2006 Cardinal Health, Inc. or one of its subsidiaries. All rights reserved.
Loading...