Pump Based Free Flow Protection .......................................................................................38
Anti-Bolus Function .....................................................................................................38
Spare Parts ..............................................................................................39
Document History ........................................................................................39
Contact Us ...............................................................................................40
Customer Service Information ...........................................................................................40
BDDF00144 Issue 1
2/40
Page 4
Alaris™ GW 800 Volumetric Pumps
Introduction
Introduction
The Alaris™ GW 800 Volumetric Pump (herein after referred to as Pump) is a small lightweight volumetric infusion Pump that provides
accurate and reliable infusions over a range of rates. The ideal Pump for general care and critical care.
This Directions for use can be used with the following Alaris GW 800 Volumetric Pumps:-
• 800TIG2GBD1
• 800TIG2CZD1
• 800TIG2DED1
• 800TIG2ESD1
• 800TIG2EED1
• 800TIG2FID1
• 800TIG2FRD1
• 800TIG2HRD1
• 800TIG2HUD1
• 800TIG2ITD1
• 800TIG2NLD1
• 800TIG2PLD1
• 800TIG2SRD1
• 800TIG2SED1
• 800TIG2TRD1
Intended Purpose
The Alaris GW 800 Volumetric Pump is intended for use by medical staff for the purpose of controlling infusion rate and volume.
Conditions for Use
The Alaris GW 800 Volumetric Pump should only be operated by medical staff competent in the use of automated volumetric pumps
and in the management of infusion therapy. Medical staff should determine the suitability of the device in their care area for its
intended purpose.
The user must be thoroughly familiar with the Pump and have been trained as per training document 0000CF02888.
w
Indications
The Alaris GW 800 Volumetric Pump is indicated for the infusion of fluids, medications, parenteral nutrition, blood and blood products
through clinically acceptable routes of administration; such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of
fluid spaces. The Alaris GW 800 Volumetric Pump is indicated for use on adults and paediatrics.
Contraindications
The Alaris GW 800 Volumetric Pump is contraindicated for enteral therapies.
BDDF00144 Issue 1
3/40
Page 5
Alaris™ GW 800 Volumetric Pumps
About This Manual
About This Manual
The user must be thoroughly familiar with the Pump described in this manual prior to use.
All illustrations used in this manual show typical settings and values which may be used in setting up the functions of the Pump.
These settings and values are for illustrative use only. The complete range of settings and values are detailed in the specifications section.
Keep this Manual for future reference during the Pump’s operational life.
w
Conventions used in this manual
BOLDUsed for Display names, software commands, controls and indicators referenced in this manual, for
'Single quotes'Used to indicate cross-references made to another section of this manual.
ItalicsUsed to refer to other documents or manuals and also used for emphasis.
w
It is important to ensure that you only refer to the most recent version of the Directions for Use and Technical Service
Manual for your BD products. These documents are referenced on bd.com. Paper copies of the Directions For Use can
be obtained free of charge by contacting your local BD representative. An estimated delivery time will be provided
when the order is placed.
example, AC Power Indicator,
Caution: Wherever this symbol is shown an Important note is found. These notes highlight an aspect of
use that is important for the user to be aware of when operating the Pump.
E, ON/OFF button.
BDDF00144 Issue 1
4/40
Page 6
Alaris™ GW 800 Volumetric Pumps
Features of the Alaris GW 800 Volumetric Pump
Features of the Alaris GW 800 Volumetric Pump
Display Indicators
Main Display -
Displays the infusion rate, VTBI, VI
and time remaining for VTBI / Time
infusions. Display will flash when
Pump is operating on battery.
Flow Stop Mechanism -
When activated the mechanism
will stop fluid flow.
Air Sensor
Bevel -
Releases the flow
stop mechanism lever
arm when the door is
closed.
Door Latch -
Press outer latch to
open the Pump door.
Mute Indicator
Alarm Indicator
Keypad
Flow Direction Label
Pressure Sensor
Tubing Guide -
Guide to assist in the loading
of the Infusion set.
Release Lever -
For Rotating cam.
Rotating Cam -
To lock onto the
rectangular bars.
Mains Fuses
Mains Inlet
Potential Equalisation
(PE) connector
RS232 / Nurse call
Connector
Flow Sensor Interface
Folded Pole Clamp
Infra Red
Communications
port (IrDA)
BDDF00144 Issue 1
5/40
Page 7
Controls and Indicators
Controls
SymbolDescription
ON/OFF button - Press once to switch the Pump on. Press and hold down for approximately 3 seconds to switch
the Pump off.
a
Note: Logs are maintained for power down events including when the Pump is powered down or
unexpected power loss.
Alaris™ GW 800 Volumetric Pumps
Controls and Indicators
d
m
b
l
e
hf
Indicators: (when illuminated)
RUN/HOLD button - Press to start the infusion or to put the infusion on hold.
CLEAR/MUTE button - Press to silence alarm or advisories for 1 minute. The audio from the alarm or advisories
will resound after this time. Resets numeric values to zero.
Note: To re-enable the alarm audio press any other button or key to unmute.
PRIME/BOLUS button - Primes the Infusion set. Administers bolus during the infusion.
PRIMARY/SECONDARY button - Switches the Pump between Primary and Secondary infusion modes. (If
enabled).
ENTER button - Scrolls between rate, time, VTBI and total volume infused (VI). Enters values for selected infusion
/ configuration parameters. Confirms the rate during an infusion titration.
CHEVRON keys - Increases or decreases the infusion rate, TIME limit and VTBI. Press and hold to increase the
selection speed. Used to adjust user selectable options
SymbolDescription
S
RATEThe Pump is displaying the infusion rate in millilitres per hour (ml/h).
VTBIThe Pump is displaying the Volume To Be Infused (VTBI) in millilitres (ml).
VIThe Pump is displaying the Volume Infused (VI) in millilitres (ml).
TIMEThe Pump is displaying the infusion time in hours : mins.
MICROThe Pump is operating in the Micro Mode. When not illuminated the Pump is in the Standard Mode.
SECThe Pump is operating in the Secondary Mode. When not illuminated the Pump is in the Primary Mode.
ml/h
AC POWER indicator - When illuminated the Pump is connected to an AC power supply.
(Millilitres / hour) When ml is illuminated the Pump displays the rate, VTBI or VI. When the h is illuminated the
Pump displays the rate or infusion time.
ALARM Indicator - When flashing indicates the Pump is in an alarm condition.
MUTE Indicator - When flashing indicates the Pump audio alarm is muted.
BDDF00144 Issue 1
6/40
Page 8
Symbol Definitions
Labelling Symbols:
SymbolDescription
Consult accompanying documents
Alaris™ GW 800 Volumetric Pumps
Symbol Definitions
x
>
k
IP32
r
s
T
t
+
U
W
Potential Equalisation (PE) Connector
RS232/Nursecall Connector.
Type CF applied part. (Degree of protection against electrical shock)
Protected against direct sprays of water up to 15° from vertical and protected against solid objects greater
than 2.5mm.
Alternating Current
Device complies with the requirements of Council Directive 93/42/EEC as amended by 2007/47/EC.
Date of Manufacture
Manufacturer
Connector for Flow Sensor
Not for Municipal Waste
Fuse rating
V
W
X
Y
,
.
Infusion indicator - Infusing in Standard mode.
Infusion indicator - Infusing in Micro mode.
Infusion indicator - Displays fluid drops detected by the flow sensor when infusing in Standard mode.
Infusion indicator - Displays fluid drops detected by the flow sensor when infusing in Micro mode.
Battery Status indicator - Displays status of battery, battery has greater than 30 minutes run time.
Battery Status indicator - Displays status of battery, battery is low and has approximately 30 minutes or less
run time.
BDDF00144 Issue 1
7/40
Page 9
Operating Precautions
Infusion Sets
• To ensure correct and accurate operation, only use BD single use infusion sets described in the 'Compatible
Infusion Sets' section of this Directions for Use. Use an infusion set with an anti-siphon valve whenever
possible. The anti-siphon valve prevents free flow from occurring if an infusion set is incorrectly loaded or
removed from the Pump.
• It is recommended that Infusion sets are changed in accordance with the Directions for Use.
Carefully read the Directions For Use supplied with the Infusion set prior to use. Use of non-specified Infusion
sets may impair the operation of the Pump and the accuracy of the infusion.
• When combining several apparatus and/or instruments with Infusion sets and other tubing, for example via
a 3-way tap or multiple Infusion, the performance of the Pump may be affected and should be monitored
closely.
• Uncontrolled flow may result if the Infusion set is not properly isolated from the patient i.e. closing a tap in
j
the set or activating an in-line clamp/roller clamp.
• The Infusion set may be fitted with an in-line clamp, which can be used to occlude tubing in case it is
required to stop fluid flow.
• The Alaris GW 800 Volumetric Pump is a positive pressure Pump, which should use Infusion sets fitted with
Luer lock fittings or equivalent locking connectors.
• To infuse from a burette, close the roller clamp above the burette and open the clamp on the vent on top of
the burette.
• Discard Infusion set if the packaging is not intact or the protector cap is detached. Ensure sets are not kinked
as this may occlude the tubing.
Mounting the Pump
Alaris™ GW 800 Volumetric Pumps
Operating Precautions
^
%;
$
• The fluid height in the container must not be more than 1 metre above the patients heart.
• Do not mount the Pump in a vertical position with the AC power inlet pointing upwards as this could affect
electrical safety, in the event of a fluid spill over the Pump.
Operating Pressure
• The pumping pressure alarm system is not designed to provide protection against, or detection of, IV
complications which can occur.
Alarm Conditions
• Several alarm conditions detected by this Pump will stop the infusion and generate visual and audible
alarms. Users must perform regular checks to ensure that the infusion is progressing correctly and no alarms
are operating.
• Alarm tone settings are preserved in the case of power loss, however some system faults will result in loss
of alarm settings. The new alarm tone settings will be stored when powering down from tech mode after a
change. The settings will be lost if a cold-start is performed, but should be saved for faults that don't require a
cold start.
Using Collapsable bags, Glass Bottles & Semi Rigid containers
• It is recommended that the air vent be opened on the Alaris GW 800 Volumetric Pump set if using glass
bottles or semi-rigid containers, to reduce the partial vacuum formed as the fluid is infused from the
container. This action will ensure the Pump can maintain volumetric accuracy whilst the container empties.
The action of opening the air vent for semi-rigid containers should take place after the spiking of the
container and priming of the drip chamber.
Steps for the Collapsable bags
Steps for Semi-rigid containers
Follow steps 1 to 3 as shown for
the semi-rigid containers, however
do not open vent as in step 4, but
prime the set as per step 5. Ensure
the bag outlet is fully pierced
before filling the drip chamber.
BDDF00144 Issue 1
2.
Spike the
container
1.
Close the
roller clamp
3.
Fill drip
chamber to
fill line
8/40
Open the air vent
to allow pressure
equalisation - ready
for infusion
5.
Prime the set by
opening / closing the
roller clamp
4.
Page 10
/
MR
&
Alaris™ GW 800 Volumetric Pumps
Operating Precautions
Operating Environment
• When using any infusion Pump in conjunction with other pumps or devices requiring vascular access,
extra care is advised. Adverse delivery of medication or fluids can be caused by the substantial variation in
pressures created within the local vascular system by such pumps. Typical examples of those pumps are used
during dialysis, bypass or cardiac assist applications.
• This Pump is suitable for use in Hospital and clinical environments other than domestic establishments and
those directly connected to the public single phase AC mains power supply network that supplies buildings
used for domestic purposes. (Consult Technical Service Manual, appropriately trained Qualified Service
Personnel or BD for further information).
• This Pump is not intended to be used in the presence of a flammable anaesthetic mixture with air or oxygen
or nitrous oxide.
Electromagnetic Compatibility and Interference
• This Pump is protected against the effects of external interference, including high energy radio frequency
emissions, magnetic fields and electrostatic discharge (for example, as generated by electrosurgical and
cauterising equipment, large motors, portable radios, cellular telephones etc.) and is designed to remain safe
when unreasonable levels of interference are encountered.
• Therapeutic Radiation Equipment: Do not use the Pump in the vicinity of any Therapeutic Radiation
Equipment. Levels of radiation generated by the radiation therapy equipment such as Linear Accelerator, may
severely affect functioning of the Pump. Please consult manufacturer’s recommendations for safe distance
and other precautionary requirements. For further information, please contact your local BD representative.
• Magnetic Resonance Imaging (MRI): The Pump contains ferromagnetic materials which are susceptible
to interference with magnetic field generated by the MRI devices. Therefore, the Pump is not considered
an MRI compatible Pump as such. If use of the Pump within an MRI environment is unavoidable, then BD
highly recommends securing the Pump at a safe distance from the magnetic field outside the identified
Controlled Access Area in order to evade any magnetic interference to the Pump; or MRI image distortion.
This safe distance should be established in accordance with the manufacturers’ recommendations regarding
electromagnetic interference (EMI). For further information, please refer to the product technical service
manual (TSM). Alternatively, contact your local BD representative for further guidance.
• Accessories: Do not use any non-recommended accessory with the Pump. The Pump is tested and compliant
with the relevant EMC claims only with the recommended accessories. Use of any accessory, transducer or
cable other than those specified by BD may result in increased emissions or decreased Pump immunity.
• In some circumstances the Pump may be affected by an electrostatic discharge through air at levels close
to or above 15kv; or by radio frequency radiation close to or above 10v/m. If the Pump is affected by this
external interference the Pump will remain in a safe mode; the Pump will duly stop the infusion and alert
the user by generating a combination of visual and audible alarms. Should any encountered alarm condition
persist even after user intervention, it is recommended to replace that particular Pump and quarantine the
Pump for the attention of appropriately trained Qualified Service Personnel.
• This Pump is a CISPR 11 Group 1 Class B device and uses RF energy only for its internal function in the normal
product offering. Therefore, its RF emissions are very low and are not likely to cause any interference with the
nearby electronic equipment. However, this Pump emits a certain level of electromagnetic radiation which
is within the levels specified by IEC/EN60601-1-2 and IEC/EN60601-2-24. If the Pump interacts with other
equipment, measures should be taken to minimise the effects, for instance by repositioning or relocation.
• For further information on electromagnetic compatibility, please consult Technical Service Manual,
BDTM00005.
BDDF00144 Issue 1
9/40
Page 11
d
m
V
*
Alaris™ GW 800 Volumetric Pumps
Operating Precautions
Earth Conductor
• The Alaris GW 800 Volumetric Pump is a Class I device, therefore must be earthed when connected to an AC
power supply.
• This Pump also has an internal power source.
• When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the
integrity of the external protective conductor on the AC power cable has been compromised, the Pump
should be disconnected from the AC power source and operated utilising the internal battery.
Hazards
• An explosion hazard exists if the Pump is used in the presence of flammable anaesthetics. Exercise care to
locate the Pump away from any such hazardous sources.
• A fire hazard may exist if the Pump is used in the presence of high oxygen concentrations.
• Dangerous Voltage: An electrical shock hazard exists if the Pump’s casing is opened or removed. Refer all
servicing to Qualified Service Personnel.
• Electrostatic discharge (ESD) precautions are required when connecting RS232/Nurse Call. Touching the pins
of the connectors may result in ESD protection failure. It is recommended that all actions must be taken by
appropriately Qualified Service Personnel.
• If this Pump is dropped, subjected to excessive moisture, fluid spillage, humidity or high temperature, or
otherwise suspected to have been damaged, remove it from service for inspection by Qualified Service
Personnel. When transporting or storing the Pump, use original packaging where possible, and adhere to
temperature, humidity and pressure ranges stated in the Specifications section and on the outer packaging.
• Warning: Alaris GW 800 Volumetric Pumps should not be modified or altered in any way, except where
explicitly directed or authorised by BD. Any use of Alaris GW 800 Volumetric Pumps which have been altered
or modified otherwise than in strict application of directions provided by BD, is at your sole risk, and BD
does not provide any warranty for or endorsement on any Alaris GW 800 Volumetric Pump that has been so
modified or altered. BD product warranty shall not apply in the event the Alaris GW 800 Volumetric Pump
has suffered damage or premature wear, or malfunctions or otherwise operates incorrectly, as a result of
unauthorised modification or alteration of the Alaris GW 800 Volumetric Pump.
BDDF00144 Issue 1
10/40
Page 12
Alaris™ GW 800 Volumetric Pumps
Getting Started
Getting Started
Before operating the Pump read this Directions For Use (DFU) manual carefully.
w
Initial Set Up
1. Check that the Pump is complete, undamaged and that the voltage rating specified on the label is compatible with your AC power
supply.
2. Items supplied are :
• Alaris GW 800 Volumetric Pump
• Electronic Instructions For Use Insert
• User Support CD (Directions For Use)
• AC Power Cable (as requested)
• Protective Packaging
3. Connect the Pump to the AC power supply for at least 2½ hours to ensure that the internal battery is charged (verify that the
lit).
The Pump will automatically operate from its internal battery if the Pump is switched on without being connected
w
to the power supply.
Should the Pump fail to perform correctly, replace in its original protective packaging, where possible and contact
Qualified Service Personnel for investigation.
S is
Power Input
The Pump is powered from the AC supply through a standard IEC AC connector. When connected to the AC supply the AC Power
indicator is illuminated.
• To isolate the Pump from AC supply remove the AC connector from the source socket.
w
• The Pump should be positioned to allow access for disconnecting the AC connector.
BDDF00144 Issue 1
11/40
Page 13
Alaris™ GW 800 Volumetric Pumps
Getting Started
Pole Clamp Installation
A pole clamp is fitted to the rear of the Pump and will provide secure fixing to vertical I.V. poles of a diameter between 15 and 40 mm.
Recessed
area
Docking Station/Workstation* or Equipment Rail Installation
1. Pull the folded pole clamp towards you and unscrew the clamp to leave
enough room for the size of the pole.
2. Place Pump around pole and tighten screw until the clamp is secured to the
pole.
Never mount the Pump such that the Infusion stand becomes
w
w
top heavy or unstable.
Ensure pole clamp is folded away and stored within recessed
area at the rear of the Pump before connecting to a Docking
Station/Workstation* or when not in use.
Prior to each use, check the pole clamp:
• does not show any signs of excessive wear,
• does not show any signs of excessively loose movement in the
extended, mountable position.
If these signs are observed, the Pumps should be taken out of
service for examination by Qualified Service Personnel.
Rectangular bar
Warning: Pump may fall off the Docking Station/Workstation* if not properly mounted which could result in user and/or
patient harm.
It is recommended that infusion bags be located on a hanger directly above the pump with which they are being
w
*Alaris DS Docking Station and Alaris Gateway Workstation.
used. This minimises the potential for confusion of Infusion sets when multiple volumetric pumps are used.
Release lever
Rotating cam
The rotating cam can be fitted to the rectangular bar on the
Docking Station/Workstation* or equipment rails measuring 10mm
by 25mm.
1. Align the rotating cam on the rear of the pump with the
rectangular bar on the Docking Station/Workstation* or the
equipment rail.
2. Push the pump firmly onto the rectangular bar or equipment
rail.
3. Ensure that the pump 'clicks' securely into position onto the rail
or bar.
4. Ensure that the Pump is positioned securely. Verify Pump is
secure by gently pulling the Pump away from the Docking
Station/Workstation* without using the release lever. When the
Pump is securely attached, it should not come off the Docking
Station/Workstation*.
5. To release, push the release lever and pull the pump forwards.
BDDF00144 Issue 1
12/40
Page 14
Loading an Infusion Set
Ensure the appropriate Infusion set for the fluid/drug to be infused has been selected.
w
Follow the instructions supplied with the individual Infusion set.
Use of non-specified Infusion sets may impair the operation of the Pump and the accuracy of the infusion. For Alaris
GW 800 Volumetric Pump Infusion sets refer to 'Infusion Sets' section of the DFU.
Position the IV fluid container to avoid spillage onto the Pump.
Ensure that the tubing is inserted completely into the pumping channel, avoiding any slack.
When using 273-003, 273-003V, 273-303E and 273-303EV Infusion sets, ensure a separation of at least 50cm is
maintained between the Pump and the upper Back Check Valve.
Alaris™ GW 800 Volumetric Pumps
Getting Started
1. Close the in-line clamp on the Infusion set. Press the door latch to open the
tubing cover door.
2. Release the flow stop mechanism by pushing the lever arm up and to the right.
3. Avoiding any slack, insert the infusion set from left to right into the slot
provided, following the flow direction label. Make sure that the infusion set is
pressed firmly past the constriction points and into the slots on either side of
the casing.
4. Re-engage the flow stop mechanism by pushing the lever left and down.
5. Close the tubing cover door. Use the recesses in the cover door to press the
door firmly to ensure the latch is correctly applied. Open in-line clamp on the
Infusion set.
6. Observe fluid chamber and check for no fluid flow.
BDDF00144 Issue 1
13/40
Page 15
Alaris™ GW 800 Volumetric Pumps
Power On/Off
To power up the Pump:
1. Press the a button once and release.
2. Check:
• A high pitch sounder is activated for 3 seconds during this time the main speaker 'beeps' once upon power up sequence.
• All display segments and all indicators are lit.
• If an error occurs during self-test the Pump will alarm.
3. After this self-test the Pump will display last rate setting entered or zero, depending on the configuration.
To power down the Pump:
1. Press and hold a. The Pump will display OFF3-OFF2-OFF1.
2. If the
w
a button is released during the countdown the Pump will not power down and it will return to its previous state.
If the Pump alarms, segments / indicators do not light up correctly or 2 audible sounds are not heard, then the Pump
must be powered down immediately, and Qualified Service Personnel contacted. If transportation to an engineer is
necessary, it is recommended to use the original protective packaging.
Battery Operation
The Pump will operate from the internal battery when AC Power is disconnected.
The following can be observed to confirm that the Pump is running on battery Power:
• A single beep is heard when AC Power is disconnected from the Pump
• The AC Power Indicator is extinguished
• When infusing:
– The ml/h indicator will flash
– Main Display will flash
– Other display indicators, if displaying, will flash then after one minute will extinguish. The display indicators will flash again
after any key is pressed.
To check the battery status see 'User Selectable Options' section.
w
Getting Started
BDDF00144 Issue 1
14/40
Page 16
Priming the Infusion Set
Use an infusion set with an anti-siphon valve whenever possible. The anti-siphon valve prevents free flow from
w
1. Ensure the Pump is switched on and in-line clamp is open.
2. Load the Infusion set (see 'Loading the Infusion Set').
3. Press
4. Press and hold
hospital protocol).
5. Attach the set to the patient or other Infusion set.
6. Start the infusion (see 'Starting the Infusion').
w
occurring if an infusion set is incorrectly loaded or removed from the Pump. Infusion sets with an anti-siphon valve
can only be primed when loaded in the Pump.
When using infusion sets without an anti-siphon valve eg. 273-004, 273-007 and 273-008, the infusion set can be
primed without using the Pump. Use of a flow sensor is recommended when using an infusion set without an antisiphon valve. The flow sensor will cause the Pump to alarm if a significant deviation from the set infusion rate occurs.
b button once E will be displayed.
b button while E is still displayed, prime the Infusion set until there is no visible air in the IV line (according to
Use the prime function to fill the Infusion sets before starting an infusion.
Never connect the Infusion set to the patient during the priming process.
The prime (
After operating the Pump's prime function, the prime function cannot be used again until after the door has been
opened and closed or the Pump has been powered off and then on.
E) volume delivered will not be subtracted from the VTBI, or added to the total volume infused.
Alaris™ GW 800 Volumetric Pumps
Getting Started
Automatic Set Detection
The Pump automatically tests that a compatible BD Infusion set has been loaded correctly (refer to 'Compatible Infusion Sets' section of
this DFU). The test will occur at the start of the first infusion after the Pump is switched on or after the door has been opened; the Pump
will run in reverse for 10 seconds and then forward for 10 seconds, the test takes a maximum of 20 seconds to complete. During this
operation the clinician may observe a blood return that will be more evident if using a small catheter.
If the Pump fails to detect a correct BD Infusion set or detects a possible set misload, then the Pump will alarm and display
refer to 'Alarms and Warnings' section of this DFU.
Please contact your local BD representative if further information or support is required regarding the Automatic Set Detection
operation or the application of this Pump in specific clinical settings, e.g. neonatal.
Q
BDDF00144 Issue 1
15/40
Page 17
Alaris™ GW 800 Volumetric Pumps
Starting the Infusion with a Flow Sensor (Recommended)
Starting the Infusion with a Flow Sensor (Recommended)
The flow sensor automatically monitors the infusion flow rate through the drip chamber. The flow sensor will cause
w
w
Check:
• The Pump is switched on.
• The Infusion set has been primed (refer to 'Priming the Infusion Set' section of this DFU).
• The in-line clamp is open.
• The flow sensor is connected (refer to 'Flow Sensor Operation' section of this DFU).
•
Standard Mode
1. Enter infusion rate using the fh keys.
2. Press
3. Enter VTBI using the
4. Press
5. Press
6. Press
the Pump to alarm if a significant deviation from the set infusion rate occurs. The flow sensor will also be able to
detect empty containers. For this reason, use of a flow sensor is recommended when using an infusion set without
an anti-siphon valve.
When operating the Pump, Users should position themselves at a distance of approximately 0.5 metres from the
display.
U indicates a drop is detected by the flow sensor during infusion.
e button once to confirm the infusion rate.
f h keys or switch VTBI off by pressing the f button until OFF is displayed.
e button to confirm the VTBI.
m to clear VI if required.
d button to start infusing.
Standard Mode with VTBI / Time Infusion on
1. Enter VTBI using the fh keys.
2. Press
3. Enter TIME using the
4. Press
5. Press
6. Press
e button once to confirm the VTBI.
f h keys.
e button to confirm the TIME.
m to clear VI if required.
d button to start infusing.
Micro Mode
1. Enter infusion rate using the fh keys.
2. Press
3. Enter VTBI using the
4. Press
5. Press
6. Press
e button once to confirm the infusion rate.
f h keys or switch VTBI off by pressing the f button until OFF is displayed.
e button to confirm the VTBI.
m to clear VI if required.
d button to start infusing.
Micro Mode with VTBI / Time Infusion on
1. Enter VTBI using the fh keys.
2. Press
3. Enter TIME using the
4. Press
5. Press
6. Press
e button once to confirm the VTBI.
f h keys.
e button to confirm the TIME.
m to clear VI if required.
d button to start infusing.
BDDF00144 Issue 1
16/40
Page 18
Alaris™ GW 800 Volumetric Pumps
Starting the Infusion without a Flow Sensor
Starting the Infusion without a Flow Sensor
When operating the Pump, Users should position themselves at a distance of approximately 0.5 metres from the
w
Check:
• The Pump is switched on.
• The Infusion set has been primed (refer to 'Priming the Infusion Set' section of this DFU).
• The in-line clamp is open.
•
Standard Mode
1. Enter infusion rate using the fh keys.
2. Press
3. Enter VTBI using the
4. Press
5. Press
6. Press
Standard Mode with VTBI / Time Infusion on
display.
indicates infusion without a flow sensor being used.
e button once to confirm the infusion rate.
f h keys.
e button to confirm the VTBI.
m to clear VI if required.
d button to start infusing.
1. Enter VTBI using the fh keys.
2. Press
3. Enter TIME using the
4. Press
5. Press
6. Press
e button once to confirm the VTBI.
f h keys.
e button to confirm the TIME.
m to clear VI if required.
d button to start infusing.
Micro Mode
1. Enter infusion rate using the fh keys.
2. Press
3. Enter VTBI using the
4. Press
5. Press
6. Press
e button once to confirm the infusion rate.
f h keys.
e button to confirm the VTBI.
m to clear VI if required.
d button to start infusing.
Micro Mode with VTBI / Time Infusion on
1. Enter VTBI using the fh keys.
2. Press
3. Enter TIME using the
4. Press
5. Press
6. Press
e button once to confirm the VTBI.
f h keys.
e button to confirm the TIME.
m to clear VI if required.
d button to start infusing.
BDDF00144 Issue 1
17/40
Page 19
Alaris™ GW 800 Volumetric Pumps
Secondary / Piggyback Infusions
Secondary / Piggyback Infusions
Secondary (or piggyback) Infusion mode is only available if configured, refer to Configurable Options section of this DFU.
Secondary Infusion mode is used to administer an intermittent fluid / drug solution e.g. 4 hourly antibiotic infusion using:
• A primary infusion set with an in-line check valve before the Y-Injection site e.g. 273-003 or 273-303E.
• A secondary infusion set e.g. 72213 or 72213N.
Primary fluid container must hang lower (approximately 20cm lower) than the secondary fluid container to allow
w
the secondary infusion to run. Primary infusion will restart on completion of the secondary infusion.
1. Set the primary infusion, but do not start (refer to 'Starting the Infusion' section of this DFU). If Pump is running press
put Pump on hold.
2. Prime Secondary Infusion set, following the instructions supplied.
3. Close the in-line clamp on the secondary set.
4. Connect secondary Infusion set into upper Y-injection site of the primary Infusion set.
5. Lower primary fluid container using extension hook supplied with the secondary Infusion set.
6. Press
7. Enter required rate using the
8. Press
9. Enter VTBI using the
10. Open the in-line clamp on the secondary set.
11. Press
12. Ensure the SEC (Secondary) indicator is lit.
w
l button and S will be displayed.
Rate / VolumeOrVTBI / TIME
hf
keys.Enter VTBI using the
button to scroll to VTBI.Press
e
e button to scroll to TIME.
hf keys.
hf keys.Enter TIME using hf keys.
e
button to scroll further, or press
Note: The infusion rate will automatically revert to the primary infusion rate when the secondary infusion is complete. On completion
of the primary infusion the Pump will continue at Keep Vein Open rate (KVO) rate.
During primary / secondary infusion of 2 drugs into a single lumen line, it is essential to ensure drug / fluid
compatibility by consulting a drug compatibility chart or local pharmacist, prior to infusion.
The secondary set connects to the upper Y-connection on the primary Infusion set.
To set the Secondary Infusion the Pump must be on
Secondary infusion rates above 270ml/h may cause simultaneous flow from secondary and primary fluid sources.
d
button to start the secondary infusion.
G or not running.
d button to
Typical Secondary Infusions:
Secondary Fluid Container
Usually a smaller container e.g.
50ml, 100ml, 200ml or 250ml.
In-line Clamp
e.g. roller clamp.
Secondary Infusion Set
e.g. 72213. Usually shorter
tubing to reach the Y-site on
the primary Infusion set.
Upper Y-Injection Site on
Primary Infusion Set.
IV Pole
2
1
Extension Hook
Normally included with the secondary Infusion
set.
Primary Infusion set must hang lower to allow the
secondary infusion to run and primary infusion to
restart on completion of the secondary.
Primary Fluid Container
Primary Infusion set
e.g. 273-003V with an upper Y-injection site.
In-line Clamp e.g Roberts clamp.
Check Valve
Prevents secondary infusions
from flowing back up the
primary Infusion set instead
of to the patient.
If using a flow sensor it must be
w
on the primary Infusion set.
BDDF00144 Issue 1
18/40
Page 20
Alaris™ GW 800 Volumetric Pumps
Basic Features
Rate Titration
1. Enter the new infusion rate using the fh keys.
2. Press
w
Bolus Infusions
To administer a bolus infusion:
1. Press
2. Press and Hold
Panel Lock
The Panel Lock feature minimises the risk of unintentional changes to the infusion settings, whilst infusing.
e button to confirm the infusion rate.
Note: The rate can be increased or decreased without interrupting the infusion.
If the new rate selected is not confirmed the Pump will revert to the current rate and no change in the infusion rate
will occur.
b button once and C will be displayed.
b button while C is still displayed, release b button after administering the desired bolus volume.
Note: Bolus volume given will be added to the total volume infused (VI) and subtracted from the volume to be infused (VTBI).
Basic Features
p
If Panel Lock is enabled then displays for all (non-operational) button presses.
Panel Lock prevents button operation with the exception of:
• Scrolling between infusion parameters using the
• Muting the alarm using the
• Pausing / resuming the infusion using the
Optimising the Pumps Performance
Pump performance may be optimised by moving a new section of the infusion set in the pumping mechanism. To insert a new section
of tubing:
1. Press
2. Ensure the in-line clamp is closed.
3. Open the Pump door, release the flow stop mechanism and move the Infusion set approximately 15cm along. See 'Loading the
4. Close the door, press
Hold Mode
Press d to pause the infusion. Press d again to resume the infusion.
A call back alarm will activate if the Pump is left on
KVO (Keep Vein Open) Rate
At the end of infusion, the Pump will continue to infuse at a very low rate (refer to 'Specifications' section of this DFU). KVO is used to
keep the patients vein open, in order to prevent blood clots and catheter occlusions.
d to place the infusion on G.
Infusion Set'.
d to restart the infusion.
m button.
e button.
d button.
G for more than 2 minutes.
w
If the KVO rate is greater than the set infusion parameters then the Pump will continue to infuse at the set infusion
rate.
If the KVO rate has been configured to OFF the Pump will stop infusing and generate an alarm.
BDDF00144 Issue 1
19/40
Page 21
Alaris™ GW 800 Volumetric Pumps
Basic Features
Changing the Infusion Set
1. Press d to put the Pump on G.
2. Close in-line clamp and ensure the IV access to the patient is isolated.
3. Disconnect the Infusion set from the patient.
4. Open Pump door and remove Infusion set from the Pump and discard the set and fluid container according to hospital protocol.
5. Place new Infusion set into Pump, see 'Loading the Infusion Set'.
6. Squeeze the drip chamber approximately half full or up to fill line (if the drip chamber is marked) with fluid.
7. Prime the set manually.
8. Restart infusion, see 'Getting Started'.
When changing the Infusion set or the fluid container use aseptic technique according to hospital protocol. It
w
Changing the Fluid Container
1. Press d to put the Pump on G.
2. Remove bag spike on Infusion set from empty / used container. Discard empty / used container according to hospital protocol.
3. Insert spike into new container.
4. Squeeze the drip chamber approximately half full or up to fill line (if the drip chamber is marked) with fluid.
5. Restart infusion, see 'Getting Started'.
w
is recommended that Infusion sets are changed in accordance with the Directions for Use. Carefully read the
Directions For Use supplied with the Infusion set prior to use. The set change interval is 24 hours.
When changing the Infusion set or the fluid container use aseptic technique according to hospital protocol. It
is recommended that Infusion sets are changed in accordance with the Directions for Use. Carefully read the
Directions For Use supplied with the Infusion set prior to use.
SmartSite™ Needle-Free System Instructions
SmartSite Needle-Free Valve is designed to permit safe gravity flow and automated flow, injection and aspiration of fluids without the
use of needles by utilising Luer lock and Luer slip connectors.
Precautions:
w
DIRECTIONS - Use Aseptic Technique
1. Prior to every access, swab top of SmartSite Needle-Free Valve port with 70% Isopropyl alcohol (1-2 seconds) and allow to dry
(approximately 30 seconds).
NOTE: Dry time is dependent on temperature, humidity, ventilation of the area.
2. Prime valve port. If applicable, attach syringe to SmartSite Needle-Free Valve port and aspirate miniscule air bubbles.
3. When used with administration sets always refer to individual set directions for use as change interval may vary according to clinical
application (e.g. infusions of blood, blood products, and lipid emulsions).
NOTE: During use of Needle-Free Valve port, fluid may be observed between the housing and blue piston. This fluid does not enter
NOTE: For product questions or needle-free valve educational materials, contact your BD representative. Consult facility protocols.
Discard if packaging is not intact or protector caps are unattached.
If SmartSite Needle-Free Valve is accessed by a needle in an emergency the valve will be damaged causing leakage.
Replace SmartSite Needle-Free Valve immediately.
SmartSite Needle-Free Valve contraindicated for blunt cannula system.
Do not leave slip Luer syringes unattended.
the fluid path and requires no action.
Consult other organisations that publish guidelines useful in developing facility protocols.
BDDF00144 Issue 1
20/40
Page 22
Alaris™ GW 800 Volumetric Pumps
Clearing Air-In-Line
1. Press d button to silence the air-in-line alarm and put the Pump on hold.
2. Close the in-line clamp.
3. Open the door to view the air bubble.
NOTE: Air-in-line alarms can be activated by both single bubbles and bubbles accumulated over time.
4. Disconnect the Infusion set from the patient and ensure the IV access to the patient is isolated.
5. Close the door.
6. Open the in-line clamp.
7. Press the
8. Press and hold the
9. Close the in-line clamp.
10. Attach the Infusion set to the patient.
11. Open the in-line clamp and restore the IV access to the patient.
12. Press the
b button once and E will be displayed.
b button while E is still displayed until there is no visible air in the IV line (according to hospital protocol).
d button to resume the infusion.
Use aseptic technique according to hospital protocol.
w
Infusion sets without an in-line anti-siphon valve must be clamped before disconnecting from the patient and the
prime (
w
E) procedure will not be necessary to remove the air-in-line, as the air can be removed by gravity.
Basic Features
BDDF00144 Issue 1
21/40
Page 23
User Selectable Options
To set the user options the Pump must be on G or in set-up mode, and the relevant user selectable options are
w
Press
Note: Number of
Check the Battery Status
1. Display shows the status of battery.
•
•
2. Press
Setting the Occlusion Pressure Level
1. When PrES is displayed.
2. Select HI, nor or Lo using the
3. Press
enabled. See also 'Configurable Options' section in this DFU.
The Battery Status option is always enabled.
e button and hold for 2 seconds to enter User Selectable Options.
e button presses are dependent upon the User Select Mode Options which have been enabled. The following
instructions are based upon all options being enabled.
, - battery has greater than 30 minutes run time.
. - battery is low and has approximately 30 minutes or less run time.
e button to return to G or set-up mode or next option.
h f keys to set the occlusion pressure level to High, Normal or Low.
e button to return to G or set-up mode or next option.
Alaris™ GW 800 Volumetric Pumps
User Selectable Options
Setting the Alarm Volume
1. Press e button until tonE is displayed.
2. Select an alarm volume between 1 (Low) and 7 (High) using the
3. Press
e button to return to G or set-up mode, or next option.
h f keys.
Setting a VTBI / Time Infusion
1. Press e button until R is displayed.
2. Select On or OFF using the
3. Press
4. Make sure that the TIME indicator is lit if set to on.
e button to return to G or set-up mode, or next option.
Note: The time increases / decreases in rate dependant units e.g. 10ml @ 99.9ml/h is 6mins, therefore 0:06 is displayed.
h f keys to turn the VTBI / time infusion setting on or off.
Setting to Micro Mode
1. Press e button until T is displayed.
2. Select On or OFF using the
3. Press
4. Make sure that the MICRO indicator is lit if set to on.
e button to return to G or set-up mode, an audible sounder will confirm the status.
h f keys to turn micro mode on or off.
BDDF00144 Issue 1
22/40
Page 24
Alaris™ GW 800 Volumetric Pumps
Configurable Options
The default settings are configurable as displayed in brackets in the table below. Each of the configurable options
w
DescriptionRangeDefaultSetting
Enable VTBI / time infusions (On / OFF)OFF
Maximum priming volume (OFF, 1 to 40 ml)40ml
has a code which must only be altered by Qualified Service Personnel with reference to the technical service manual
(TSM) for this product (Technical Service Manual reference: BDTM00005).
Configurable Options
Clear infusion parameters to zero on
power - up
Maximum VTBI in MICRO Mode (0.1 to 999 ml)999ml
Bolus rate (1 to 999ml/h)400ml/h
Maximum bolus volume (OFF, 1 to 99ml)5ml
Keep vein open rate (OFF, 1.0 to 5.0 ml/h)5.0ml/h
Air in line alarm volume - single bubble (50,100, 250, 500µL)100µl
Enable secondary infusion capability (On / OFF)OFF
Default occlusion pressure on power - up (Lo (250mmHg),Nor (350mmHg),
Alarm volume level (1 - 7)4
Enable Micro mode (On / OFF)OFF
Maximum infusion rate (1 - 999 ml/h)999ml/h
Enable ASCII mode for communications (On / OFF)OFF
Enable Odd parity for communications (On / OFF)OFF
Set Pump address for communications (1 - 250)1
Flow Sensor Connection Mode**(AUTO/On)AUTO
Set - up of current time and date (00:00 to 23:59)
Language selection (EnGL, FrAn, dEut, ItAL, ESPA, SE,
*EnGL - English, FrAn - French, dEut - German, ItAL - Italian, ESPA - Spanish, SE - Swedish, nEd - Dutch.
** If infusion sets without anti-siphon valves are being used, it is recommended to change the flow sensor connection mode setting to On. With
On selected, the Alaris GW 800 Volumetric Pump will not operate unless a flow sensor is connected.
Serial NumberSoftware Version
Configured byDate
Approved byDate
(On / OFF)
(On / OFF)
(On / OFF)
(On / OFF)
BDDF00144 Issue 1
OFF
OFF
OFF
OFF
23/40
Page 25
Alaris™ GW 800 Volumetric Pumps
Alarms
All Alarms are of a high priority and are indicated by a combination of flashing red beacon, an audible alarm and a message on the
display.
1. Check the display for an alarm message and review table below for cause and action. Press
M and K)
2. When the cause of the alarm has been rectified, press the
All alarms will stop the infusion except for F and H, where the infusion status will remain the same state
w
DisplayCauseAction
J
K
L
M
N
O
P
Q
D
F
H
Note: The audio sound pressure level is at least 45 dB depending on configuration of the alarm sound level.
as prior to the alarm.
AIR-IN-LINE
UPSTREAM OCCLUSION
INTERNAL BATTERY DEPLETEDTo silence the alarm connect the Pump to AC power.
DOOR OPEN
Door was opened during an infusion.
SYSTEM FAULTSwitch Pump off.
FLOW ERROR
No drops have been detected during an
infusion (empty container).
Excessive increase or decrease of fluid
flow detected by flow sensor.
Flow sensor attached to secondary
Infusion set.
FLOW SENSOR CONNECTION ERROR
Flow sensor connected / disconnected
during an infusion.
Flow sensor is not connected and the
primary VTBI is OFF.
Too much fluid in drip chamber.
DOWNSTREAM OCCLUSION
A blockage has occurred downstream.
Incorrect infusion set, set incorrectly
loaded or set worn.
Excessive amount of air in line.
Infusion started with upstream tubing
clamped.
273-003 set loaded with upper Y-site too
close to Pump.
Finished pre-set volume to be infused
with KVO switched off.
Low Battery
(At least 30 mins before bAt alarm).
The Pump has been left unattended for
2mins and infusion has not started.
d button to resume the infusion.
See 'Clearing Air-In-Line'.
Remove the occlusion / air and restart the infusion by pressing
the
d button.
Restart operation on AC power to charge the internal battery.
Close the door and restart the infusion.
Remove Pump from service and have the Pump inspected by
Qualified Service Personnel.
Clamp the tubing to stop fluid flow.
Ensure that the infusion set tubing is properly loaded in the
pumping channel following the flow direction label.
Ensure that ample fluid is in the fluid container.
Check for blockage / occlusion in infusion set.
After the tubing is properly inserted, close the Pump door and
resume infusion.
Ensure flow sensor is attached to the primary infusion set.
Restart the infusion with the flow sensor connected /
disconnected, as required.
Connect flow sensor or set a VTBI and re-start the infusion.
Ensure fluid in the drip chamber is not above the fill line.
Remove pressure in the infusion set to prevent a post
occlusion bolus to the patient.
Remove the cause of the blockage.
Restart the infusion.
Remove the infusion set and load the correct or new set (see
'Compatible Infusion sets').
Clear air from set. (Refer to 'Clearing Air-In-Line' section)
Release clamp and restart.
Reload set with Pump at least 30cm from the Y-site.
Set new VTBI.
Connect Pump to an AC power source.
Attend to Pump. Press
m to silence the alarm. (Exceptions are
d to clear alarm.
Alarms
w
Setting the alarm sound pressure level lower than the ambient sound pressure level can impede User recognition of
alarm conditions.
BDDF00144 Issue 1
24/40
Page 26
Alaris™ GW 800 Volumetric Pumps
Advisories
Advisories alert the user and are indicated by an audible alarm, a message on the display or both.
1. Check the display for advisory message. Press
DisplayCauseAction
D
C
E
G
I
Finished pre-set volume to be infused.Pump will infuse at the keep vein open rate, until the
Bolus is being administered.
The Pump is priming the Infusion set.Ensure all air has been primed out of the infusion set,
The Pump is on hold.
Automatic set check.Allow test to complete before operating the Pump
m to silence the alarm.
d button is pressed. Refer to 'KVO Rate' section of
this DFU.
Release
bolus has been administered.
before starting the infusion.
Press
to set-up.
further.
b button to return to infusion once correct
d to return to infusion, or press e to return
Advisories
BDDF00144 Issue 1
25/40
Page 27
Flow Sensor Operation
The flow sensor automatically monitors the infusion flow rate through the drip chamber. The flow sensor will cause
w
Flow Sensor Usage
Infusion set with anti-siphon valve ?Use flow sensor ?
YESOptional
NORecommended
Model 180 Flow Sensor
the Pump to alarm if a significant deviation from the set infusion rate occurs. The flow sensor will also be able to
detect empty containers. For this reason, use of a flow sensor is recommended when using an infusion set without
an anti-siphon valve.
Alaris™ GW 800 Volumetric Pumps
Flow Sensor Operation
Handles
Pull back handles
Flow sensor
interface
1. Plug the flow sensor into the flow sensor interface located on the top rear part of the Pump.
2. Attach the Model 180 Flow Sensor to the drip chamber of the Infusion set, by pulling back the handles. Refer to the illustration
above.
3. Proceed with load, priming, and set-up instructions as described in section 'Getting Started'.
NOTE: Ensure drip chamber is half full and upright.
Flanges
Drip chamber
Flow
Sensor
Interface
Cover
w
Always attach the flow sensor before starting an infusion.
Avoid using the flow sensor in direct sunlight.
Always ensure lens is clean.
Always replace the flow sensor interface cover when the flow sensor is disconnected.
BDDF00144 Issue 1
26/40
Page 28
Alaris™ GW 800 Volumetric Pumps
Compatible Infusion Sets
Compatible Infusion Sets
The Pump uses standard, single-use, disposable Infusion sets with Luer-lock connectors. The user is responsible for verifying the
suitability of a product used, if it is not recommended by BD.
• BD recommends the use of infusion sets with anti-siphon valves whenever possible. The anti-siphon valve
w
Please note these drawings are not to scale
Standard Sets
273-001VInfusion set with 15µm filter in drip chamber, anti-siphon valve. (230 cm)
273-002VInfusion set with 15µm filter in drip chamber, 1 Y site and anti-siphon valve. (240 cm)
273-003VInfusion set with 15µm filter in drip chamber, 2 Y sites, back check valve and anti-siphon valve. (240 cm)
prevents free flow from occurring if an infusion set is incorrectly loaded or removed from the Pump.
• New sets are continuously being developed for our customers. Please contact your local BD representative for
availability.
• It is recommended that Infusion sets are changed in accordance with the Directions for Use. Carefully read the
Directions For Use supplied with the Infusion set prior to use.
273-004VInfusion set with 15µm filter in drip chamber, roller clamp and Luer back check valve. (220 cm) Suitable for gravity
infusion.
273-005VInfusion set with roller clamp and a back check valve. (220 cm) Suitable for gravity infusion.
273-303EVInfusion set with 15µm filter in drip chamber, two back check valves and two SmartSite Valve Y Port. (295 cm)
273-304VInfusion set with 15µm filter in drip chamber. (270 cm) Suitable for gravity infusion.
Blood Sets
273-007VBlood set with 1 upper Y site, in-line drip chamber with 200µm filter, and Luer back check valve. (290 cm) Suitable
for gravity infusion.
273-008EVBlood set with 1 upper and 1 lower SmartSite Valve Y port, back check valve, in-line drip chamber with 200µm
filter and Luer back check valve. (300 cm) Suitable for gravity infusion.
273-080EVBlood set with 2 spikes, 1 upper and 1 lower SmartSite Valve Y port with anti-siphon valve and in-line drip
chamber with 200µm filter. (255 cm)
BDDF00144 Issue 1
27/40
Page 29
Alaris™ GW 800 Volumetric Pumps
Compatible Infusion Sets
Filter Sets
273-009V1.2µm filter set with anti-siphon valve, with 15µm filter in drip chamber. (245 cm)
273-022V0.2µm filter set with anti-siphon valve, with 15µm filter in drip chamber. (245 cm)
Burette Sets
273-103EVBurette set with 1 SmartSite Valve Y port and anti-siphon valve. (230 cm)
Opaque Sets
273-043VLight Resistant PVC infusion set with anti-siphon valve and Pump segment with 15µm filter in drip chamber. (250
cm)
Low Sorbing Sets
273-053VLow Sorbing PVC infusion set with anti-siphon valve and Pump segment with 15µm filter in drip chamber.
(270 cm)
Secondary Sets
72213-0006Secondary / Piggyback set with 18G needle and hanger. (approx. 84 cm)
72213N-0006Secondary / Piggyback set and extension hook. (approx. 76 cm)
BDDF00144 Issue 1
28/40
Page 30
Alaris™ GW 800 Volumetric Pumps
Compatible Infusion Sets
Filter Extension Sets
C20128Extension set with 1.2µm filter and one Y site. Rotating male Luer lock. (approx. 51 cm)
C20350Extension set with 0.2µm filter and one Y site. Rotating male Luer lock (approx. 51 cm) Low Sorbing (Polyethylene
Lined)
20128E-0006Extension set with 1.2µm filter and one SmartSite Valve Y port. Rotating male Luer lock. (approx. 51 cm)
20350E-0006Extension set with 0.2µm filter and one SmartSite Valve Y port. Rotating male Luer lock (approx. 51 cm) Low
Sorbing (Polyethylene Lined)
Oncology Sets
MFX273-950EOncology set with five SmartSite Valve Y ports. (265 cm)
MFX273-951EOncology set with three SmartSite Valve Y ports. (261 cm)
MFX273-952EAmber Oncology set with five SmartSite Valve Y ports. (265 cm)
MFX273-954EAmber Oncology set with three SmartSite Valve Y ports. (261 cm)
For the following infusion sets carefully read the Directions For Use supplied with the Infusion set prior to use for
w
information on the use of the flow sensor with the infusion sets:
• MFX273-950E
• MFX273-951E
• MFX273-952E
• MFX273-954E
BDDF00144 Issue 1
29/40
Page 31
Associated Products
The Alaris Gateway Workstation
Product SKU80203UNS0y-xx
Supply Voltage115-230VAC, ~50-60Hz
Electrical Rating460VA (Maximum)
Protection Against Electrical ShockClass 1
ClassificationContinuous Operation
Supply to Pump115-230V, ~50-60Hz, 60VA
Alaris™ GW 800 Volumetric Pumps
Associated Products
The Alaris DS Docking Station
y = Connectivity option - 1, 2 or 3
xx = Configuration
Product SKU80283UNS00-xx
Supply Voltage230VAC, ~50-60Hz
Electrical Rating500VA (nominal)
Protection Against Electrical ShockClass 1
ClassificationContinuous Operation
Supply to Pump20VA max 230V 50-60Hz
BDDF00144 Issue 1
30/40
Page 32
Alaris™ GW 800 Volumetric Pumps
Maintenance
Maintenance
Routine Maintenance Procedures
To ensure that this Pump remains in good operating condition, it is important to keep it clean and carry out the routine maintenance
procedures described below. All servicing should only be performed by Qualified Service Personnel with reference to the Technical
Service Manual (TSM).
Circuit diagrams and components parts lists and all other servicing information which will assist the Qualified Service Personnel in
performing repair of the parts designated as repairable are available upon request from BD.
If the Pump is dropped, damaged, subjected to excessive moisture or high temperature, immediately take it out of
w
Interval Routine Maintenance Procedure
As per hospital policyThoroughly clean external surfaces of the Pump before and after prolonged period of storage.
Each usage1. Inspect AC power supply plug and cable for damage.
Before the transfer of the Pump to
a new patient and as required
service for examination by Qualified Service Personnel.
All preventative and corrective maintenance and all such activities shall be performed at a compliant work place
in accordance with the information supplied. BD will not be responsible should any of these actions be performed
outside the instructions or information supplied by BD.
2. Inspect case, keypad and mechanism for damage.
3. Check Start up self-test operation is correct.
4. Check for activation of both the alert indicator and audio function during the Pump start-up.
Clean the Pump by wiping over with a lint-free cloth lightly dampened with warm water and a
standard disinfectant/detergent solution.
Please refer to Technical Service Manual for calibration procedures. The units of measurement used in the calibration
w
w
procedure are standard SI (The International System of Units) units.
Keep this Manual for future reference during the Pump’s operational life.
It is important to ensure that you only refer to the most recent version of the Directions for Use and Technical Service
Manual for your BD products. These documents are referenced on bd.com. Paper copies of the Directions For Use can
be obtained free of charge by contacting your local BD representative. An estimated delivery time will be provided
when the order is placed.
Battery Operation
The internal rechargeable battery allows continued operation when the AC power is unavailable, for example during patient transfer or
AC power failure. The infusion time on battery is rate dependant, see 'Specifications' section of the DFU. From the battery low alarm it
will take about 24 hours to fully recharge when reconnected to the AC power supply, whether the Pump is in use or not. The battery is
automatically charged during AC operation and whenever the Pump is connected to the AC power supply and the AC power indicator is
illuminated.
The battery is maintenance free, sealed Nickel Metal Hydride and requires no routine servicing. However, to achieve optimum operation,
ensure that the battery is fully recharged after full discharge, before storage, and at regular 3 month intervals during storage.
Charge retention will eventually degrade. Where retention is critical the internal battery should be replaced every 3 years.
It is recommended that only Qualified Service Personnel replaces the battery. For further information regarding the replacement of
batteries refer to the Technical Service Manual.
Any use of battery packs that are not manufactured by BD in the Alaris Volumetric Pump is at your sole risk, and BD does not provide any
warranty for or endorsement on any battery packs that are not manufactured by BD. BD product warranty shall not apply in the event
the Alaris Volumetric Pump has suffered damage or premature wear, or malfunctions or otherwise operates incorrectly, as a result of use
with a battery pack that is not manufactured by BD.
Disposal
Information on Disposal for Users of Waste Electrical & Electronic Equipment
U symbol on the product and/or accompanying documents means that used electrical and electronic products should not be
This
mixed with household waste.
If you wish to discard electrical and electronic equipment, please contact your BD affiliate office or distributor for further information.
Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and
the environment which could otherwise arise from inappropriate waste handling.
Information on Disposal in Countries outside the European Union
Usymbol is only valid in the European Union. The product should be disposed of taking environmental factors into consideration.
This
To ensure no risk or hazard, remove the internal rechargeable battery and the Nickel Metal Hydride battery from the control board and
dispose of as outlined by the local country regulations. All other components can be safely disposed of as per local regulations.
BDDF00144 Issue 1
31/40
Page 33
Alaris™ GW 800 Volumetric Pumps
Cleaning and Storage
Cleaning and Storage
Cleaning the Pump
Before the transfer of the Pump to a new patient and periodically during the use, clean the Pump by wiping over with a lint-free cloth
lightly dampened with warm water and a standard disinfectant / detergent solution.
Do not use the following disinfectant types:
• Disinfectants which are known to be corrosive to metals must not be used, these include:
• NaDcc (such as Presept),
• Hypochlorites (such as Chlorasol),
• Aldehydes (such as Cidex),
• Cationic Surfactants >1% (such as Benzalkonium Chloride).
• Iodine (such as Betadine) will cause surface discoloration.
• Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Recommended cleaners are:
BrandConcentration
Hibiscrub20% (v/v)
Virkon1% (w/v)
The following products were tested and are acceptable for use on the Pump if used in accordance with the specified manufacturer’s
guidelines.
• Warm soapy water• Tristel Fuse sachets
• Mild detergent in water (e.g. Young’s Hospec)• Tristel Trio wipes system
• 70% Isopropyl Alcohol in water• Tuffie 5 wipe
• Chlor-Clean• Virkon Disinfectant
• Hibiscrub• TriGene Advance
• Clinell Universal Wipes
Before cleaning always switch off and disconnect from the AC power supply. Do not allow liquid to enter the casing
w
and avoid excess fluid build up on the Pump.
Do not use aggressive cleaning agents as these may damage the exterior surface of the Pump. Do not steam autoclave,
ethylene oxide sterilise or immerse this Pump in any fluid.
Storing the Pump
If the Pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in a clean, dry
atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in the technical service manual and ensure that the internal
battery is fully charged.
See the Technical Service Manual for further information regarding the charging of the RTC Battery BT1.
w
Cleaning and storing the Infusion set
The Infusion set is a disposable single use item and should be discarded after use according to hospital protocol.
Cleaning the Flow Sensor
Before the transfer of the flow sensor to a new Infusion set and periodically during use, clean the flow sensor by wiping over with a lintfree cloth lightly dampened with warm water and a standard disinfectant / detergent solution. Ensure the connector does not get wet.
Dry flow sensor before use.
To aid cleaning of flow sensors which have been heavily soiled, contaminated or if the handle operation is not free moving, then the flow
sensor may be immersed and soaked in clean soapy water (see
whilst submerged in the water.
After cleaning, the flow sensor should be allowed to dry fully prior to use.
The plug of the flow sensor must not be immersed in water as damage will occur.
A). The inside of the spring mechanism can be cleaned by activating it
w
BDDF00144 Issue 1
32/40
Page 34
Alaris™ GW 800 Volumetric Pumps
Specifications
Specifications
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24.
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.
Electrical Safety
IEC/EN 60601-1 - Typical earth leakage current 40µA.
Potential Equalisation Conductor
The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Pump and the potential
equalisation busbar of the electrical installation. To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Pump to the potential equalisation busbar of the electrical installation.
Dielectric Strength
Proof strength test 1.7kV dc (live and neutral to earth) for 10s
Performance strength test 500V dc (live and neutral to earth)
Proof Strength Test
The proof strength test is applied at the factory. It is not recommended that the proof strength test is reapplied if the Pump is tested
again during service.
Classification
Class I Equipment. Continuous, Portable Equipment, type 4.
AC Power Supply
220 - 240 VAC, 50 - 60Hz, 10VA (nominal).
Protection against fluid ingress
IP32 - Protected against direct sprays of water up to 15° from vertical and protected against solid objects greater than 2.5mm.
Dimensions
137mm (w) x 140mm (h) x 105mm (d). Weight: approx. 1.5kg (excluding power cable).
The maximum volume infused following a single fault condition is 1.0ml.
KVO Infusion Rate
Up to a max. of 5ml/h or the infusion rate if programmed less than the set KVO rate.
Occlusion Pressure
User Selectable: Occlusion Alarm Pressure at 25ml/h - 250mmHg (low), 350mmHg (normal), 500mmHg (high).
Fuse Type
2 X T 63 mA, slow blowing (220 - 240 VAC, nominal).
Air Sensor
Integral Ultrasonic Sensor.
Air in line detector
Configurable 50µl, 100µl, 250µl, 500µl.
Total Time Setting
Up to 99 hours and 59 mins.
Memory Retention
The electronic memory of the Pump will be retained for at least 6 months when not powered up.
Minimum Occlusion Alarm Pressure
100mmHg
Maximum Occlusion Alarm Pressure
1000mmHg
Bolus volume generated at 25.0 ml/h when the minimum occlusion alarm threshold is reached
0.3ml
Bolus volume generated at 25.0 ml/h when the maximum occlusion alarm threshold is reached
0.6ml
Maximum time for activation of occlusion alarm
Maximum time to alarm at 1.0ml/h is <45min (High Pressure)
Maximum time to alarm at 1.0ml/h is <30min (Low Pressure)
Maximum time to alarm at 25ml/h is <5.30min (High Pressure)
Maximum time to alarm at 25ml/h is <2.10min (Low Pressure)
Maximum time to alarm at 999ml/h is <3 secs (High Pressure)
Maximum time to alarm at 999ml/h is <2 secs (Low Pressure)
System Accuracy
Rate Accuracy ± 5% at 25 ml/h under nominal conditions2, tested to IEC60601-2-24 (95% confidence interval / 80% population).
For all conditions the rate accuracy should be adjusted accordingly.
6
w
Bolus Volume Accuracy - ± 10% @ 5ml under nominal conditions2, tested to IEC60601-2-24. Under all conditions3 the bolus volume
accuracy should be de-rated as for rate accuracy.
BDDF00144 Issue 1
34/40
Page 36
Alaris™ GW 800 Volumetric Pumps
Specifications
Occlusion Pressure Accuracy
± 150 mmHg under nominal conditions
± 250 mmHg under all conditions
2
3
Air in Line Accuracy
± 20% or ± 0.025ml5 under nominal conditions
Notes:
1. All accuracy specifications are with a 95% confidence interval / 95% population, unless stated otherwise.
2. Nominal conditions are defined as:
• Set Rate: 125 ml/h (25 ml/h for rate accuracy);
• Disposable Type: 273-001;
• Needle: 18 gauge x 40 mm;
• Solution Type: De-ionized & Degassed Water;
• Temperature: 23° ± 2°C
• Fluid Head Height: 0.3 ± 0.1 m;
• Back Pressure: 0 ± 10 mmHg.
3. All are as normal conditions with the following additions:
• Set Rate: 1 to 999 ml/h;
• Solution Type: All fluids
4
;
• Temperature: +5°C - +40°C
• Fluid Head Height: 0 ± 1.0 m;
4. Tested using Distilled water, 20% lipid, 50% glucose, 0.9% Normal Saline and 5% Alcohol solutions.
5. Whichever is the greater of the air in line limit set.
6. For all conditions the rate accuracy should be adjusted by the following percentages:
• ± 10% over the infusion rate range 1 to 999 ml/h
• Nominal: 0.68 (± 0.36)% over 24 hours of continuous use.
• Nominal: -3.5 (± 1.08)% @ 15°C
• Nominal: -0.9 (± 0.62)% @ 38°C
2
BDDF00144 Issue 1
35/40
Page 37
Alaris™ GW 800 Volumetric Pumps
IrDA, RS232 and Nurse Call Specification
IrDA, RS232 and Nurse Call Specification
RS232 / IrDA Feature
The RS232 / IrDA feature is a standard feature on Alaris GW 800 Volumetric Pump. It allows the Pump to be monitored remotely via a
suitable central monitoring or computer system. It also enables the internal event log of the Pump to be downloaded for technical
support purposes.
The nurse call interface provides a remote backup to the internal audible alarm. It should not be relied upon to
w
Nurse Call Feature
The nurse call interface is for connection to a suitable monitoring device in order to provide remote indication of the Pump entering an
alarm condition.
replace monitoring of the internal alarm.
Refer to the Technical Service Manual for further information regarding the RS232 interface. Since it is possible to
control the Pump using the RS232 interface at some distance from the Pump and hence remote from the patient,
responsibility for the control of the Pump is vested in the software run on the computer control system.
The assessment for the suitability of any software used in the clinical environment to control or receive data from
the Pump lies with the user of the equipment. This software should include detection of the disconnection or other
failure of the RS232 cable. The protocol is detailed in the Technical Service Manual and is for general information
only.
Any connected analogue and digital components are required to meet IEC/EN60950 for data processing and IEC/
EN60601 for medical devices. Anyone connecting additional devices to the signal input or output is a system
configurator and responsible for meeting the requirements of the system standard IEC/EN60601-1-1.
RS232 / Nurse Call Connection Data
Typical Connection Data -
1. Nurse call (Relay) Normally Closed (NC C)
2. Transmit Data (TXD) Output
3. Received Data (RXD) Input
4. Power Input (DSR)
5. Ground (GND)
6. Not used
7. Power Input (CTS)
8. Nurse call (Relay) Normally open (NC O)
9. Nurse call (Relay) Common (NC COM)
BDDF00144 Issue 1
36/40
Page 38
Alaris™ GW 800 Volumetric Pumps
Start-up Graph at 1.0 ml/h (Initial Period)
-0.2
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
0102030405060708090100 110120
Start-up Graph at 25.0 ml/h (Initial Period)
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
45.0
50.0
0102030405060708090100 110120
Trumpet Graph at 1.0 ml/h (Initial Period)
-60
-50
-40
-30
-20
-10
0
10
20
30
40
50
60
05101520253035
Flow Error (%)
Trumpet Graph at 25.0 ml/h (Initial Period)
-15
-10
-5
0
5
10
15
05101520253035
Flow Error (%)
Trumpet Graph at 1.0 ml/h (After 24 Hours)
-80
-70
-60
-50
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
05101520253035
Trumpet Graph at 25.0 ml/h (After 24 Hours)
-15
-10
-5
0
5
10
15
05101520253035
Trumpet and Flow Rate Curves
Trumpet and Flow Rate Curves
In this Pump, as with all infusion systems, the action of the pumping mechanism and variations cause short-term fluctuations in rate
accuracy.
The following curves show typical performance of the system in two ways: 1) the accuracy of fluid delivery over various time periods is
measured (trumpet curves), and 2) the delay in onset of fluid flow when infusion commences (start-up curves).
Trumpet curves are named for their characteristic shape. They display discrete data averaged over particular time periods or 'observation
windows', not continuous data versus operating time. Over long observation windows, short term fluctuations have little effect on
accuracy as represented by the flat part of the curve. As the observation window is reduced, short term fluctuations have greater effects
as represented by the mouth of the trumpet.
Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered. Short
term fluctuations in rate accuracy may have clinical impact depending on the half-life of the drug being infused and the degree of inter
vascular integration, the clinical effect cannot be determined from the trumpet curves alone.
The start-up curves represent continuous flow versus operating time for two hours from the start of the infusion. They exhibit the delay
in onset of delivery due to mechanical compliance and provide a visual representation of uniformity. Trumpet curves are derived from
the second hour of this data. Tests performed per IEC60601-2-24 standard.
Note: The typical flow rate and trumpet curves - Infusion set 273 001
Minimum
Rate Error
Overall Mean Error
= +4.2%
BDDF00144 Issue 1
Flow Rate (%)
Maximum
Rate Error
37/40
Minimum
Rate Error
Overall Mean Error
= +1.1%
Page 39
Alaris™ GW 800 Volumetric Pumps
Technical Description
Technical Description
The following details outline the basic safety checks designed into the Pump to minimise the possibility of under or over infusions.
Power on Self-Tests
The Pump is single fault tolerant, which means the Pump will either stop and alarm, or continue to infuse uninterrupted as a result of a
single fault failure. During the power on self-test sequence the Pump automatically performs system integrity checks and will alarm and
display
M should any of these tests fail. Refer to 'Power On/Off' section of this DFU.
Air-in-Line
Two ultrasonic transducers continuously check for the presence of air in the Infusion set throughout the infusion. This air-in-line feature
operates in two modes:
• Single Bubble Detection - The Pump will alarm and display
volume alarm limit is detected. The alarm limit can be configured to 50, 100, 250 or 500 µL. See also 'Configurable Options' section
of this DFU.
• Air-in-Line Accumulation - This accumulation feature monitors the volume of air that passes through the Infusion set by
accumulating the volume of individual bubbles over a 15 minute window. This feature is particularly useful with infusions for
patients that are highly sensitive to air (i.e. infants, paediatrics) or when infusing products that create significant volumes of small
air bubbles.
Although an individual bubble may not exceed the pre-programmed threshold, the additive volume of bubbles in a
w
15 minute volume may be sufficient to initiate an air-in-line alarm indicated by an Air OCCL message.
J whenever a single air bubble greater than the air in line
Downstream Occlusion Pressure
The Pump includes a pressure sensor to monitor the downstream Infusion pressure. When the IV pressure exceeds the alarm pressure
limit, as a result of, for example kinked IV tubing or blocked cannula, the Pump will alarm and display
To compensate for the variability in Infusion set tubing the Pump performs a relative, baseline pressure measurement. The Pump takes a
reference pressure from the IV line when the infusion commences and alarms at a preset limit above the baseline pressure. The pressure
alarm limits are 250, 350 and 500 mmHg above the baseline pressure, corresponding to the low, normal or high, pressure limits. To avoid
excessively large pressures the Pump is capped at 1000 mmHg.
P.
Upstream Occlusion Pressure
To detect upstream occlusions resulting from, for example, closed clamps or occluded drip chamber filters the Pump continuously
monitors the upstream IV tubing pressure. Upon detection of an occlusion the Pump will alarm and display
the ultrasonic transducers of the air sensor to monitor for upstream occlusions and is therefore unable to differentiate an upstream
occlusion from an air bubble.
J. The Pump uses
Pump Based Free Flow Protection
The Pump is equipped with a flow stop mechanism that is designed to occlude the IV tubing when the Pump door is open and the
tubing remains properly loaded in the Pump. Raising the flow stop mechanism lever arm and pushing it to the right activates this
mechanism. Once activated, the user inserts the IV tubing into the tubing guide channel.
When the Pump door is closed, the bevel integrated into the door releases the lever arm such that it will automatically occlude the
tubing when the door is re-opened. (Refer to 'Features of the Alaris GW 800 Volumetric Pump') The tubing can be removed from the
tubing guide channel by repeating the activation of the lever arm. Once the lever arm is re-activated and the door is opened, the flow
stop mechanism no longer occludes flow in the tubing.
Anti-Bolus Function
The anti-bolus function is designed to reduce the bolus that may occur upon the release of an occlusion following a downstream
occlusion alarm. (Detection of a downstream occlusion is indicated by a
to neutral within 15 seconds by pumping backwards briefly and measuring the Infusion set line pressure through the in-line pressure
detection system. This feature can prevent the fluid bolus to the patient that can occur upon the release of an occlusion, which may be
caused by a downstream clamp.
P alarm.) The Pump returns the Infusion set line pressure
BDDF00144 Issue 1
38/40
Page 40
Alaris™ GW 800 Volumetric Pumps
Spare Parts
Spare Parts
A comprehensive list of spare parts for this Pump is included within the Technical Service Manual.
The Technical Service Manual BDTM00005 is now available in electronic format on the World Wide Web at :- bd.com/int-alaris-technical
A username and password are required to access our manuals. Please contact a local customer services representative to obtain login
details.
Part NumberDescription
1000EL00349Internal Battery Pack
1001FAOPT91AC Power Lead - UK
1001FAOPT92AC Power Lead - European
Document History
IssueDateSoftware Version Description
1July 2017V6r1Initial release
BDDF00144 Issue 1
39/40
Page 41
Alaris™ GW 800 Volumetric Pumps
Contact Us
For full contact information please refer to bd.com.
+27 11 603 2620
Spain+34 902 555 660 info.spain@carefusion.com
Sweden+46 8-7755160bdsweden@bd.com
Switzerland+41 215563301 Info-Switzerland@bd.com
United KingdomFreephone: 0800 917 8776uk-customer-service@carefusion.com
United States of AmericaFreephone: 800 482 4822 CustCareInfusion@carefusion.com
info.alaris.nl@bd.com
NZ-CustomerService@carefusion.com
bdsa_customer_centre@bd.com
Contact Us
BDDF00144 Issue 1
40/40
Page 42
Alaris, BD, BD logo and SmartSite are trademarks or registered
trademarks of Becton, Dickinson and Company or one of its
aliates. All rights reserved. All other trademarks are property of
their respective owners.
This document contains proprietary information of Becton,
Dickinson and Company or one of its aliates, and its receipt or
possession does not convey any rights to reproduce its contents,
or to manufacture or sell any product described. Reproduction,
disclosure, or use other than for the intended purpose without
specic written authorization of Becton, Dickinson and Company
or one of its aliates is strictly forbidden.
BD Switzerland Sàrl,
Route de Crassier 17, Business Park Terre-Bonne,
Batiment A4, 1262 Eysins, Switzerland.
BDDF00144 Issue 1
bd.com