Agilent Technologies E2625A, 100 User Manual

User’s Guide
Publication number E2625-97003 March 2002
For Safety information and Regulatory information, see the pages at the end of this manual.
© Copyright Agilent Technologies 1999-2002 All Rights Reserved
Infiniium Option 100 and E2625A Telecommunications Mask Template Kit
In This Book
This Telecommunications Mask Template Kit User’s Guide contains the following information:
Introduction to Option 100 and E2625A Telecommunications Mask Template Kit
Chapter 1 introduces specific product information about the Mask Template Kit. It also discusses the recommended oscilloscope configuration, and shows an overview of the Mask Test Setup dialog box.
Getting Started and Using Masks
C h a p t e r 2 d e s c r i b e s M a s k Te s t i n g a n d s h o w s y o u wa ys t o u s e t h e st a n d a r d m a s k s to verify compliance of your instrument waveforms.
Standard Masks
Chapter 3 lists file names for the standard masks in the Telecommunications Mask Template Kit, along with line impedances and bit rates. The templates for each standard mask are also shown.
ii
Contents
1 Introduction to Option 100 and E2625A Telecommunications
Mask Template Kit
Adding Option 100 to an existing Infiniium Oscilloscope 1-2 Supplied items 1-3 Recommended accessories 1-3 What not to do 1-3 Oscilloscope models 1-4 To access the Mask Test Setup dialog box 1-5 Overview of the Mask Test Setup dialog box 1-6
2 Getting Started and Using Masks
Introduction to Mask Testing 2-2 Make a copy of the Mask Test disk 2-3 To install the Standard Masks 2-4 To calibrate the oscilloscope 2-5 To choose a proper probe for Mask Testing 2-6 To get your waveform on the oscilloscope display 2-7 To enable the Mask Test 2-8 To load a Mask 2-9 To align a Mask to the waveform 2-10 To start and stop the Mask Test 2-11 To understand the Mask Test tab description area 2-12 To run the Mask Test conditionally 2-13 To stop the Mask Test on the first failure 2-14
3Standard Masks
ITU G.703 International Standard Masks 3-4 ANSI T1.102 North American Standard Masks 3-11 FCC Part 68.308 Options A, B, and C Masks 3-17 Pre-Compliance SONET Masks 3-19 IEEE 802.3 Masks 3-21
Contents-1
Contents-2
1

Introduction to Option 100 and E2625A Telecommunications Mask Template Kit

Introduction to Option 100 and E2625A Telecommunications Mask Template Kit
The Infiniium Oscilloscopes Option 100 Telecommunications Mask Template Kit includes the materials necessary for you to verify that your instrument waveforms comply with industry standards. Option 100 provides a quick method for you to certify waveform integrity during early development and later during manufacturing tests.
This manual helps you use the standard masks supplied with Option 100. It shows you how to load standard waveform masks into Infiniium Oscilloscopes and use them to test the compliance of your waveforms. All of the standard masks supplied are shown in Chapter 3.

Adding Option 100 to an existing Infiniium Oscilloscope

The Op tion 100 Telecommunications Mask Template Kit is orderable as E2625A for customers who already have an Infiniium Oscilloscope.
1-2
Introduction to Option 100 and E2625A Telecommunications Mask Template Kit

Supplied items

Supplied items
The Option 100 Telecommunications Mask Template Kit contains the following items:
Mask Template Kit Item
Floppy disk Mask template library floppy disk E2626A Adapter 100/110/120
Adapter cable Bantam (m) to Siemens (f) adapter cable E2623A Adapter 75
Adapter cable Dual-bantam (f) to RJ48C (m) adapter cable E2624A Adapter/coupler RJ48C (f) to RJ48C (f) E2627A Adapter Bantam (m) to BNC (f); for system verification of
Adapter BNC (m) to BNC (f) 50-to-75 ohm adapter; for
Cable BNC (m) to BNC (m) cable; 30 cm long 8120-1838 Storage case Hard-shell case for storage of all kit accessories E2625-45501
Description Separately Orderable
differential impedance adapter
with bantam (f) connector, bantam (m) to Siemens (f) adapter cable
single-ended impedance adapter with
BNC (f) connector
E2621A adapter and Infiniium oscilloscope
system verification of E2622A adapter and Infiniium oscilloscope
Agilent Part Number
E2621A
E2622A
E2628A
E2629A

Recommended accessories

Other Agilent accessories recommended for use in mask testing include:
Agilent Part Number
E2630A Patch cable Bantam (m) to bantam (m) patch cable E2632A Patch cable Siemens (m) to Siemens (m) patch cable
Accessory Item Description

What not to do

You should not use the Mask Test master disk for daily use. Instead, you should use a copy of the master disk, as described in Chapter 2.
1-3
Introduction to Option 100 and E2625A Telecommunications Mask Template Kit

Oscilloscope models

Oscilloscope models
You can use the Telecommunications Mask Template Kit with these Infiniium Oscilloscopes:
Agilent Model Channels Bandwidth Sample Rate Memory Depth
1
54810 54815 54820 54825 54835
54845
54846
54830
54831
54832
1
1
1
1
1
1
2
2
2
2 500 MHz 1 GSa/s 32k 4 500 MHz 1 GSa/s 32k 2 500 MHz 2 GSa/s 32k 4 500 MHz 2 GSa/s 32k 4 1.5 GHz 2 GSa/s — 4 ch
4 GSa/s — 2 ch
4 1.5 GHz 4 GSa/s — 4 ch
8 GSa/s — 2 ch
4 2.25 GHz 4 GSa/s — 4 ch
8 GSa/s — 2 ch
2 600 MHz 2 GSa/s — 4 ch
4 GSa/s — 2 ch
4 600 MHz 2 GSa/s — 4 ch
4 GSa/s — 2 ch
4 1 GHz 2 GSa/s — 4 ch
4 GSa/s — 2 ch
32k — 4 ch 64k — 2 ch
32k — 4 ch 64k — 2 ch
32k — 4 ch 64k — 2 ch
2M — 4 ch 4M — 2 ch
2M — 4 ch 4M — 2 ch
2M — 4 ch 4M — 2 ch
1
Requires system software version A.03.00 or later.
2
Requires system software version A.02.00 or later.
1-4
Introduction to Option 100 and E2625A Telecommunications Mask Template Kit

To access the Mask Test Setup dialog box

To access the Mask Test Setup dialog box
Select Analyze, then Mask Test... from the oscilloscope main menu, as shown here, to access the Mask Test Setup dialog box:
1-5
Introduction to Option 100 and E2625A Telecommunications Mask Template Kit

Overview of the Mask Test Setup dialog box

Overview of the Mask Test Setup dialog box
You will use the Mask Test Setup dialog box shown here to load and use masks with the Infiniium Oscilloscope. Standard masks are shown in Chapter 3. How to load and use a mask is described in Chapter 2.
Mask Test Setup Dialog Box
Checkmark to enable the Mask Test.
Click to load a Mask.
A Mask Test tab will appear when the Mask Test is enabled.
Check options to invert a mask, use the file setup when aligning a m ask, or enable averaging to reduce noise.
1-6
Click to start and stop the Mask Test.
Click to align or clear a Mask.
Conditionally run the Mask Test and stop on the first failure.
See the Infiniium help system to use Scaling and Automask.
2

Getting Started and Using Masks

Getting Started and Using Masks
Telecommunications (telecom) and data communications (datacom) industries have established standards for equipment manufacturers. One of these standards defines the electrical parameters for waveforms. By applying these standards, both telecom and datacom equipment manufacturers can ensure that their equipment will operate properly together by first certifying waveform integrity.
To verify waveform integrity during the equipment manufacturing phase, Infiniium Oscilloscopes have a feature called Mask Testing. Within seconds, you know whether your equipment’s waveform complies with the industry standard.

Introduction to Mask Testing

What is a Mask?
A mask is a template that you load into the oscilloscope used during mask testing. The mask shows the acceptable electrical parameters of a waveform by defining regions of the oscilloscope display where the waveform must not intersect. If the waveform crosses the boundary of a region of the mask, it fails the mask test.
The types of masks that you can use to test waveforms include:
• Industry standard masks
Industry standard masks are included in this Mask Template Kit.
• Automatically generated masks
• User-defined masks
When to use Mask Testing
You should use mask testing whenever your waveform must conform to an industry standard. You can do mask testing during manufacturing to ensure that your design meets the requirements of the telecom or datacom industries.
2-2
Getting Started and Using Masks

Make a copy of the Mask Test disk

How Infiniium does Mask Testing
When you start the mask test, Infiniium creates a database that has unique locations for every pixel in the waveform viewing area. Each location in the database has a 21-bit counter for a maximum count of 2,097,151.
Each time a data point from a channel memory illuminates a pixel in the wav efor m viewi ng a r ea, th e count er for that pi xel i s incre ment e d. Du r ing ma sk testing, any data point that illuminates a pixel in the mask violation region causes a mask test failure.
The oscilloscope’s database continues to build until the oscilloscope stops acquiring data or until mask testing is deactivated.
The oscilloscope’s database is also used for histogram analysis and color grade persistence. Refer to the Infiniium help system for detailed information about these other waveform analysis features.
How to do Mask Testing
The remaining topics in this chapter show you how to do mask testing. Chapter 3 shows the standard masks supplied in the Mask Template Kit.
Make a copy of the Mask Test disk
A disk containing the stand ard masks is shipped with Infiniium Option 100 and E2625A. Before you install the Mask Test templates, you should make a copy of the Mask Test master disk and use the copy of the Mask Test disk for the installation.
2-3
Getting Started and Using Masks

To install the Standard Masks

To install the Standard Masks
Use the copy of the Mask Test disk to install standard masks onto the Infiniium hard drive. The Mask Test disk contains many of the available standard masks that you may need to use during mask testing.
1
Insert the copy of the Mask Test disk in the floppy drive.
2 From the oscilloscope main menu, select Analyze, then Mask Test. 3 In the Mask Test Setup dialog box, select the Enable Mask Test control. 4 In the Mask Test Setup dialog box, click the Load Mask... button.
The Load Mask dialog box will resemble:
This selects the disk where the masks are installed.
This selects the directory where the masks are installed.
5
Click the Install Masks from A: button.
2-4
Getting Started and Using Masks

To calibrate the oscilloscope

To calibrate the oscilloscope
Your Infiniium should be at optimum performance when performing mask testing. If your Infiniium has not been calibrated in the last year or if the Calibration ∆Te m p i s gr e a t e r th a n ± 5 °C , yo u sh o u l d r e c a l i b r a te t h e os c i l loscope before performing a mask test. You can find the Calibration ∆Temp in the Calibration dialog box by clicking on the Utilities menu and selecting Calibration.
1
Make sure Infiniium has warmed up for at least 30 minutes.
2 From the oscilloscope main menu click Utilities, then Calibration. 3 Make sure that the Cal Memory Protect check box is unchecked.
Cal Memory Protect This option protects the oscilloscopes calibration factors from
Details When checked, this option displays a list of anything that has failed
4 Click Start.
Infiniium will display additional instructions and guide you through the step­by-step calibration procedure.
accidentally being changed.
during the cal ibration. You can us e the failure informati on to help repair the oscilloscope.
2-5
Getting Started and Using Masks

To choose a proper probe for Mask Testing

To choose a proper probe for Mask Testing
You can use any oscilloscope probe that fits your application. Depending on the type of waveform you are testing, you may need to use one of the probe adapters.
Calibrating the probe adapter
Depending on the requirements of your Telecom or Datacom industrial standard, you may need to use the E2622A 75 probe adapter or the E2621A differential probe adaptor. Before performing a mask test, you should calibrate the probe adapter using the following instructions:
1
In the Channel Setup dialog box, click the Probes... button.
2 In the Probes Setup dialog box, click the Calibration... button. 3 Follow the step-by-step instructions to calibrate the probe.
2-6
Getting Started and Using Masks

To get your waveform on the oscilloscope display

To get your waveform on the oscilloscope display
When doing mask testing, you will try to fit the mask to your waveform. This is how you will verify whether your waveform meets the acceptable electrical parameters defined by the standard mask. By having your waveform displayed on the oscilloscope screen before doing mask testing, when you load and use the mask, you will know immediately if the waveform passes the mask test.
1
Display the waveform you want to test on the oscilloscope screen.
2 Begin mask testing using a standard mask, as described in the following
topics.
2-7
Getting Started and Using Masks

To enable the Mask Test

To enable the Mask Test
1 Click Analyze, then Mask Test to access the Mask Test Setup dialog box. 2 In the Mask Test Setup dialog box, click Enable Mask Test to checkmark
the box.
When enabled, a Mask Test tab appears beneath the waveform viewing area.
Checkmark to enable t he Mask Test.
2-8
Getting Started and Using Masks

To load a Mask

To load a Mask
1 Click Analyze, then Mask Test to access the Mask Test Setup dialog box. 2 Make sure the mask test is enabled. 3 Click the Load Mask... button.
4 In the Load Mask dialog box, click the Mask Test Source field to select an
oscilloscope channel for the mask test. The active channel is the default.
5 Click the Drive: field and select the drive where the mask you want to
load is located. In the
List files of type: field, Mask Files will be displayed
by default.
6 To load a standard mask from the Infiniium hard drive, make sure the
Folder: shows c:\scope\masks. Then view the list of standard masks by
scrolling them in the
7 When you find the standard mask that you want to load, double-click
the file in the
Files: field. The mask file name will be entered and
highlighted in the
Chapter 3 lists the standard mask file names and shows the mask waveforms.
Files: field.
File name: field, and Infiniium will load the mask.
2-9
Getting Started and Using Masks

To align a Mask to the waveform

To align a Mask to the waveform
1 Access the Mask Test Setup dialog box and make sure the mask test is
enabled.
2 Locate the Use File Setup When Aligning option.
When aligning a mask to the waveform, these settings are used:
• Channel range and offset
• Time base range and position
• Trigger mode and level
• Mask test scaling source, position(s), and delay
You can either use the current oscilloscope settings, or you can use the standard settings defined in the mask file.
If you want to use the standard settings defined in the mask file,
3
checkmark the
4 If you want to use the current oscilloscope settings, do not checkmark
this box.
Use File Setup When Aligning option.
When aligning a mask, you can use the current settings for the oscilloscope setup, or use the standard settings in the mask file.
5 Click the Align button to align the mask.
2-10
Getting Started and Using Masks

To start and stop the Mask Test

To start and stop the Mask Test
1 Access the Mask Test Setup dialog box and make sure the mask test is
enabled.
2 At the bottom of the Mask Test Setup dialog box, locate the start and
stop mask test buttons.
• The green triangle starts the mask test.
• The red square stops the mask test.
Starts mask test in the Mask Test Setup dialog box.
3
Click the green triangle to start the mask test.
Stops mask test in the Mask Test Setup dialog box.
Starts/stops mask test from the Mask Test tab.
When you start a mask test, Infiniium also starts the acquisition system.
2-11
Getting Started and Using Masks

To understand the Mask Test tab description area

To understand the Mask Test tab description area
After you have enabled the mask test in the Mask Test Setup dialog box, you will see a Mask Test tab on the lower left of the oscilloscope screen, beneath the waveform viewing area. The description area below the Mask Test tab shows:
• Mask title
• Amplitude specifics (for ANSI T1.102 North American Standard Masks only)
• Total number of waveforms in the duration of the mask test
• Number of failed waveforms in the mask test
• Regions in the mask
• Waveform violations
Mask Test tab
Description area
When any of the ANSI T1.102 North American Standard Masks are loaded, the Mask Test tab description area will show minimum and maximum amplitude values. These values are the result of testing for specifics of the standard, and indicate the upper or lower limit that the standard allows.
2-12
Getting Started and Using Masks

To run the Mask Test conditionally

To run the Mask Test conditionally
1 Access the Mask Test Setup dialog box and make sure the mask test is
enabled.
2 Click the Run Until selection arrow to view the options.
a If Stop on Failure is disabled, you can stop the mask test based on the
following conditions:
•Forever
The mask test will run continuously. This is the default selection.
•Number of Waveforms
The mask test will run until the number of waveforms specified has been acquired. The number of waveforms is from 1 to 1,000,000,000.
•Time
The m as k test wil l r un until t he am ount of t im e specifie d h as elapse d. The time is in seconds, from 0.1 minute to 1440.0 minutes. The default is 1.0 minute.
b If Stop on Failure is enabled, the first time the waveform crosses the mask
violation region, the mask test will stop. See the next topic.
Click and select how you want the mask test to run.
Checkmark if you want the mask test to stop on the first failure.
3
Select the Run Until option you want to use for the mask test.
2-13
Getting Started and Using Masks

To stop the Mask Test on the first failure

To stop the Mask Test on the first failure
1 Access the Mask Test Setup dialog box and make sure the mask test is
enabled.
2 Click Stop On Failure to checkmark the box. 3 Run the mask test.
When the mask test encounters the first failure, the test will stop and the failure result will be displayed in the Mask Test tab description area.
Analyzing a failure
If your waveform crosses the boundary of the mask violation region, the waveform cross-over areas will be displayed in red. All waveform locations displayed in red indicate that the waveform fails the mask test in these areas. If the waveform just “touches” the boundary, the mask test fails.
Standard masks and the designated violation regions are shown in chapter 3. When you start the mask test with Stop On Failure enabled, if Infiniium
encounters a failure, the oscilloscope acquisition system will be stopped.
2-14
3

Standard Masks

Standard Masks
The Infiniium Option 100 and the E2625A Telecommunications Mask Template Kit disk contains standard masks used by the telecommunications and data communications industries. These are the masks you can load into the oscilloscope to test your waveform’s compliance against industry standards.
To use any of these masks, follow the procedures in chapter 2.
Standard Masks
The types of standard masks include:
• ITU G.703 International Standard Masks
• ANSI T1.102 North American Standard Masks
• FCC Part 68.308 Options A, B, and C Masks
• Pre-Compliance SONET Masks
• IEEE 802.3 Masks
ITU G.703 International Standard Masks
Mask File Name Line Impedance Bit Rate
1544kb_ITU_DS1.msk 100 2Mb_ITU_120.msk 120 2Mb_ITU_75.msk 75 coax 2.048 Mbps 6312kb_ITU_110_DS2.msk 110 6312kb_ITU_75_DS2.msk 75 coax 6.312 Mbps 8Mb_ITU.msk 75 coax 8.448 Mbps 34Mb_ITU.msk 75 44736kb_ITU_DS3.msk 75 coax 44.736 Mbps 140Mb_ITU_0.msk 75 coax 139.264 Mbps 140Mb_ITU_1.msk 75 140Mb_ITU_1_INV.msk 75 coax 139.264 Mbps 155Mb_ITU_0.msk 75 coax 155.520 Mbps 155Mb_ITU_1.msk 75 155Mb_ITU_1_INV.msk 75 coax 155.520 Mbps
3-2
twisted pair 1.544 Mbps twisted pair 2.048 Mbps
twisted pair 6.312 Mbps
coax 34.368 Mbps
coax 139.264 Mbps
coax 155.520 Mbps
ANSI T1.102 North American Standard Masks
Mask File Name Line Impedance Bit Rate
DS1_ANSI.msk 100 DS1A_ANSI.msk 100 DS1C_ANSI.msk 100 DS2_ANSI.msk 110 DS3_ANSI.msk 75 DS4NA_ANSI.msk 75 DS4NA_ANSI_MAX.msk 75 STS1_ANSI_EYE.msk 75 STS1_ANSI_PULSE.msk 75 STS3_ANSI.msk 75 coax 155.520 Mbps STS3_ANSI_MAX.msk 75
twisted pair 1.544 Mbps twisted pair 2.048 Mbps twisted pair 3.152 Mbps
twisted pair 6.312 Mbps coax 44.736 Mbps coax 139.264 Mbps coax 139.264 Mbps coax 51.840 Mbps coax 51.840 Mbps
coax 155.520 Mbps
FCC Part 68.308 Options A, B, and C Masks
Standard Masks
Mask File Name Line Impedance Bit Rate
FCC_68_OPT_A.msk 100 FCC_68_OPT_B.msk 100 twisted pair 1.544 Mbps FCC_68_OPT_C.msk 100
twisted pair 1.544 Mbps
twisted pair 1.544 Mbps
Pre-Compliance SONET Masks
Mask File Name Bit Rate
OC1_SONET.msk 51.840 Mbps OC3_SONET.msk 155.520 Mbps OC12_SONET.msk 622.080 Mbps
IEEE 802.3 Masks
Mask File Name Bit Rate
10BASE_T_IDL.msk 10 Mbps 10BASE_T_LINK.msk 10 Mbps 100BASE_TX_STP.msk 100 Mbps 100BASE_TX_UTP.msk 100 Mbps
3-3
Standard Masks

ITU G.703 International Standard Masks

ITU G.703 International Standard Masks
The ITU G.703 International Standard Masks resemble the following:
1544kb_ITU_DS1.msk
2Mb_ITU_120.msk
3-4
2Mb_ITU_75.msk
6312kb_ITU_110_DS2.msk
Standard Masks
ITU G.703 International Standard Masks
3-5
Standard Masks
ITU G.703 International Standard Masks
6312kb_ITU_75_DS2.msk
8Mb_ITU.msk
3-6
34Mb_ITU.msk
44736kb_ITU_DS3.msk
Standard Masks
ITU G.703 International Standard Masks
3-7
Standard Masks
ITU G.703 International Standard Masks
140Mb_ITU_0.msk
140Mb_ITU_1.msk
3-8
140Mb_ITU_1_INV.msk
Standard Masks
ITU G.703 International Standard Masks
155Mb_ITU_0.msk
3-9
Standard Masks
ITU G.703 International Standard Masks
155Mb_ITU_1.msk
155Mb_ITU_1_INV.msk
3-10
Standard Masks

ANSI T1.102 North American Standard Masks

ANSI T1.102 North American Standard Masks
The ANSI T1.102 North American Standard Masks resemble the following:
DS1_ANSI.msk
DS1A_ANSI.msk
3-11
Standard Masks
ANSI T1.102 North American Standard Masks
DS1C_ANSI.msk
DS2_ANSI.msk
3-12
DS3_ANSI.msk
DS4NA_ANSI.msk
Standard Masks
ANSI T1.102 North American Standard Masks
3-13
Standard Masks
ANSI T1.102 North American Standard Masks
DS4NA_ANSI_MAX.msk
STS1_ANSI_EYE.msk
3-14
STS1_ANSI_PULSE.msk
STS3_ANSI.msk
Standard Masks
ANSI T1.102 North American Standard Masks
3-15
Standard Masks
ANSI T1.102 North American Standard Masks
STS3_ANSI_MAX.msk
3-16

FCC Part 68.308 Options A, B, and C Masks

FCC Part 68.308 Options A, B, and C Masks
FCC_68_OPT_A.msk
Standard Masks
FCC_68_OPT_B.msk
3-17
Standard Masks
FCC Part 68.308 Options A, B, and C Masks
FCC_68_OPT_C.msk
3-18

Pre-Compliance SONET Masks

Pre-Compliance SONET Masks
The Pre-Compliance SONET Masks resemble the following:
OC1_SONET.msk
Standard Masks
OC3_SONET.msk
3-19
Standard Masks
Pre-Compliance SONET Masks
OC12_SONET.msk
3-20

IEEE 802.3 Masks

The IEEE 802.3 Masks resembles the following:
10BASE_T_IDL.msk
Standard Masks
IEEE 802.3 Masks
10BASE_T_LINK.msk
3-21
Standard Masks
IEEE 802.3 Masks
100BASE_TX_STP
100BASE_TX_UTP
3-22
Safety Notices
This apparatus has been designed and tested in accor­dance with IEC Publication 1010, Safety Requirements for Mea­suring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instru­ment (provided with terminal for protective earthing). Before applying power, verify that the correct safety precautions are taken (see the following warn­ings). In addition, note the external markings on the instru­ment that are described under "Safety Symbols."
Warnings
Before turning on the instru­ment, you must connect the pro­tective earth terminal of the instrument to the protective con­ductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two-conductor outlet is not sufficient protec­tion.
Only fuses with the required rated current, voltage, and spec­ified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short­circuited fuseholders. To do so could cause a shock or fire haz­ard.
If you energize this instrument by an auto transformer (for volt­age reduction or mains isola­tion), the common terminal must be connected to the earth termi­nal of the power source.
Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation.
Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another per­son, capable of rendering first aid and resuscitation, is present.
Do not install substitute parts or perform any unauthorized modification to the instrument.
Capacitors inside the instru­ment may retain a charge even if the instrument is disconnected from its source of supply.
Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Do not use the instrument in a manner not specified by the manufacturer.
To clean the instrument
If the instrument requires clean­ing: (1) Remove power from the instrument. (2) Clean the exter­nal surfaces of the instrument with a soft cloth dampened with a mixture of mild detergent and water. (3) Make sure that the instrument is completely dry before reconnecting it to a power source.
Safety Symbols
!
Instruction manual symbol: the product is marked with this sym­bol when it is necessary for you to refer to the instruction man­ual in order to protect against damage to the product..
Hazardous voltage symbol.
Earth terminal symbol: Used to indicate a circuit common con­nected to grounded chassis.
Agilent Technologies Inc. P.O. Box 2197 1900 Garden of the Gods Road Colorado Springs, CO 80901-2197, U.S.A.
Notices
© Agilent Technologies, Inc.
1999-2002
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and interna­tional copyright laws.
Manual Part Number
E2625-97003, March 2002
Print History
E2625-97003, March 2002 E2625-97002, January 2000 E2625-97001, October 2000 E2625-97000, May 1999
Agilent Technologies, Inc. 1601 California Street Palo Alto, CA 94304 USA
Restricted Rights Legend
If software is for use in the per­formance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial com­puter software” as defined in DFAR 252.227-7014 (June 1995), or as a commercial item” as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or dis­closure of Software is subject to Agilent Technologies standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Govern­ment will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Document Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applica­ble law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied war­ranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for inci­dental or consequential damages in connection with the furnishing, use, or per­formance of this document or of any information con­tained herein. Should Agi­lent and the user have a separate written agreement with warranty terms cover­ing the material in this docu­ment that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Trademark Acknowledgements
Windows and MS Windows are U.S. registered trademarks of Microsoft Corporation.
Loading...