Agilent Technologies 85032B-E 50 User Manual

User’s and Service Guide
Agilent Technologies 85032B/E
50Ω Type-N Calibration Kits
Agilent Part Number: 85032-90020
Printed in USA
Print Date: July 2002
© Copyright 1993, 2000, 2002 Agilent Technologies, Inc. All rights reserved.
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Assistan ce
Product maintenance agreements and other customer assistance agreements are availa ble for Agilent products.
For any assistance, contact Agilent Technologies. Refer to page 5-3.
ii 85032B/E

Contents

1. General Information
Calibration Kit Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 2
Broadband Loads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Opens and Shorts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Calibration Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Installation of the Calibration Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Equipment Required but Not Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Incoming Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Recording the Device Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Clarifying the Sex of a Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
2. Specific atio ns
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Temperature—What to Wa tch Out For . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Mechanical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Pin Depth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Supplemental Electrical Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
3. Use, Maintenance, and Care of the Devices
Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Look for Obvious Defects and Damage First. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
What Causes Connector Wear? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Inspect the Mating Plane Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Inspect Female Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Cleaning Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Gaging Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Connector Gage Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
When to Gage Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Reading the Connector Gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Gaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Gaging Male Type-N Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Gaging Female Type-N Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
How to Make a Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Preliminary Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Final Connection Using a Torque Wrench. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Connecting and Discon nect ing the Two-P iece Female Open (85032B) . . . . . . . . . . . . .3-14
How to Separate a Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
85032B/E iii
Contents
4. Performance Verification
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
How Agile nt Verifies t h e D e v i ces in This Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Limited Recertification (Option 003). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
How Often to Recertify. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Where to Send a Kit for Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
5. Troubleshooting
Troubleshooting Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Returning a Kit or Device to Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
6. Replaceable Pa rts
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
A. Standard Definitions
Standard Class Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Blank Form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Nominal Standard Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Setting the System Impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Blank Form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-6
iv 85032B/E

1 General Information

85032B/E 1-1
General Information

Calibration Kit Overview

Calibration Kit Overview
The Agilent 85032B and 85032E type-N calibration kits are used to calibrate Agilent network analyzers up to 6 GHz for measurements of components with 50type-N connectors.

Kit Contents

The 85032B calibration kit contains the following:
• one male and one female open termination
• one male and one female short termination
• one male and one female 50load
• two type-N-male to 7-mm adapters (included with Option 100)
• two type-N-female to 7-mm adapters (included with Option 100) Refer to T able 6-1 and Figure 6-1 for a complete list of kit contents and their associated
part n u mbers .
The 85032E calibration kit contains t he following:
• one male combination open/short termination
• one male 50load Refer to T able 6-2 and Figure 6-2 for a complete list of kit contents and their associated
part n u mbers .
Broadband Loads
The broadband loads a re inst rument-grade, 50terminat ions that have been optimized for performance up to 6 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.
Opens and Shorts
The opens and shorts are built from parts that are machined to the current state-of the-art precision machining.
The short’s inner conductors have a one-piece construction, common with the shorting plane. This construction provides for extremely repeatabl e connection s .
The female open has a separate-piece inner conductor that is made from a low-dielectric-constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The lengths of these offsets are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.
1-2 85032B/E
General Information
Calibration Kit Overview
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance. The adapters utilize a dual-beaded connector structure to ensure stable, repea table connections. The b eads are des i gned to minimize return loss and are separated far enough so that interaction between the beads is minimized.
The adapters are designed so that their nominal electrical lengths are the same, which allows them to be used in calibration procedures for non-insertable devices.

Calibration Definitions

The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your
network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.
The calibration definitions can be:
• resident within the analyzer
• entered from the front panel
Installation of the Calibration Definitions
The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.
Options
The following options are available for the Agilent 85032B/E. Option 100 (85032B o nly) Option 100 adds the four type-N to 7-mm adapters to the
calibration kit. Option 003 (85032B o nly) This option provides a limited calibration for the devices in
the calibration kit to 3 GHz instead of 6 GHz. This calibration option can be requeste d only from an Agilent service center. It cannot be ordered from the factory.
Option UK6 This option adds a certificate of calibration and the corresponding calibration data for the devices in the calibration kit.
Equipment Required but Not Supplied
Gages, torque and open-end wrenches, and various connector cleaning supplies are not included in the calibration kit but are required to ensure successful operation of the calibration kit. Refer to Table 6-3 on page 6-5 for o rde ring information
85032B/E 1-3
General Information

Incoming Inspection

Incoming Inspection
Verify that the shipment is complete by referring to Figure 6-1 or Figure 6-2. Check for damage. The foam-lined storage case provides protection during shipping. If the case or any device appears damaged, or if the shipment is incomplete, contact
Agilent. See Table 5-1 on page 5-3. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company.
When you send the kit or device to Agilent, in clude a service tag (found near the end of this manual) with the following information:
• your company name and address
• the name of a technical contact person within your company, and the person's complete phone number
• the model number and serial number of the kit
• the part number and serial number of the device
• the type of service required
•a detailed description of the problem
1-4 85032B/E
General Information

Recording the Device Serial Numbers

Recording the Device Serial Numbers
In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in
Table 1-1 for the 85032B and Table 1-2 for the 85032E. Recording the serial numbers will
prevent confusing the devices in this kit with similar devices in other kits.
Table 1-1 Serial Number Record for 85032B
Device Serial Number
Calibration kit Male broadband load Female broadband load Male open Female open Male s hort Female short Type-N-male to 7-mm adapter Type-N-male to 7-mm adapter Type-N-female to 7-mm adapter Type-N-female to 7-mm adapter
_____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________ _____________ __________________
Table 1-2 Serial Number Record for 85032E
Device Serial Number
Calibration kit
_______________________________ Male broadband load Male combination open/short
85032B/E 1-5
_______________________________
_______________________________
General Information

Clarifying the Sex of a Connector

Clarifying the Sex of a Connector
In this manual, the s ex of c a li bra tion de vices a nd ada pt ers ar e r ef erred to in terms of their connector interface. For example, a male open has a male connector.
However, during a measurement calibration, the network analyzer softkey menus label a
type-N calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) on the analyzer’s display refers to the short that is to be connected to the female test port. This will be a male short from the calibration kit.
Conversely, connector gages are referred to in terms of the connector that it measures. For instance, a male connector ga ge has a female connector on the gage so that it can measure male dev ices.

Preventive Maintenance

The best techniques for maintaining the integrity of the devices in this kit include:
• routine visual inspection
• cleaning
• proper g aging
• proper connection techniques All of the above are described in Chapter 3 , “Use, Maintenance, and Care of the Devices.”
Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections , r es ult ing f rom pin dep th v al ues be ing out of the observed limits (see Tab le 2- 2
on page 2-4), or from bad connections, can also damage these devices.
1-6 85032B/E

2 Specifications

85032B/E 2-1
Specifications

Environmental Requirements

Environmental Requirements
Table 2-1 Envi ronmental Requirements
Parameter Limits
Operating temperature Error-corrected temperature range
Storage temperature −40 °C to +75 °C (−40 °F to +167 °F) Altitude
Operation < 4,500 meters (15,000 feet) Storage < 15,000 meters (50,000 feet)
Relative humidity Always non-condensing
Operation 0 to 80% (26 °C maximum dry bu lb) Storage 0 to 95%
a. The temperatu re rang e over which the calibration standards maintain confor ma nce to their
specification s .
b. The allowable network analy z er ambient temp erature drift during measurement calibration
and during measurem ents when t he ne twor k analyzer er ror co rre cti on is t urne d on. Al so, the range over which the network an aly z er m aintains its spe cified performance while correction is turned on.
a
b
+15 °C to +35 °C (+59 °F to +95 °F) ±1 °C of measurement calibration temperature

Temperature—What to W atch Out For

Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range specified in
Table 2-1.
IMPORTANT Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
2-2 85032B/E
Specifications

Mechanical Characteristics

Mechanical Characteristics
Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Agilent Technol ogies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any improper pin depth when the kit leaves the factory.
“Gaging Connectors” on page 3-6 explains how to use gages to determine if the kit devices
have maintained their mechanical i nteg rity. (Refer to Table 2-2 on page 2-4 for typical and observed pin depth limits.)

Pin Depth

Pin depth is the distance the center conductor mating plane differs from being flush with the outer conductor mating plane. Refer to Figure 2-1 . Some coaxial connectors, such as
2.4 mm and 3.5 mm, are designed to have these planes nearly flush. Type-N connectors, however, are designed with a pin depth offset of approximately 5.26 mm (0.207 inch), not permitting these planes to be flush. The male center conductors are recessed by the offset value while the female center conductors compensate by protruding the same amount. This offset necessitates the redefining of pin depth with regard to protrusion and recession.
Protrusion refers to a male type-N connector center conductor having a pin depth value less than 5.26 mm (0.207 inch), or a female type-N connector c enter conductor ha ving a pin depth value greater than 5.26 mm (0.207 inch).
Recession refers to a male type-N connector center conductor having a pin depth value greater than 5.26 mm (0.207 in), or a female type-N connector center conductor having a pin depth value less than 5.26 mm (0.207 inch).
Figure 2-1 Connector Pin Depth
85032B/E 2-3
Specifications
Mechanical Characteristics
NOTE The gages for measuring type-N connectors compensate for the designed
offset of 5.26 mm (0.207 inch), therefore, protrusion and recession readings are in relation to a zero reference plane (as if the inner and outer conductor planes were intended to be flush). Gage readings can be directly compared with the observed values listed in Table 2-2.
The pin depth value of each calibration device in this kit is not specified, but is an important mechanical parameter. The electrical performance of the device depends, to some extent, on its pin depth. The electrical specifications for each device in this kit take
into account the effect of pin depth on the device’s performance. Table 2-2 lists the typical pin depths and measurement uncerta inties, and pr ovid es observed pin depth limits for the devices in the kit. If the pin depth of a device does not measure within the observed pin depth limits, it may be an indication that the device fails to meet electrical specifications. Refer to Figure 2-1 for an illustration of pin depth in type-N connectors.
Table 2-2 Pin Depth Li mit
Device Typical Pin Dept h
Opens 0 to 0.0127 mm
0 to 0.0005 in
Shorts 0 to 0.0127 mm
0 to 0.0005 in
Fixe d Load s 0 to 0.0508 mm
0 to 0.0020 in
a. Approximately +2 sigma to 2 sigma of gage uncer tainty based on studies done at the
factory according to recommended procedur es.
b. Observed pin depth limits are the range of observation limit s seen on the gage reading due
to measurement uncert ainty. The depth could still be within specifications.
Measu rem e n t U n certa in ty
+0.0038 to 0.0038 mm +0.00015 to 0. 00015 in
+0.0038 to 0.0038 mm +0.00015 to 0. 00015 in
+0.0038 to 0.0038 mm +0.00015 to 0. 00015 in
a
Observed Pin Depth Limits
+0.0038 to −0.0165 mm +0.00015 to 0.000 65 in
+0.0038 to −0.0165 mm +0.00015 to 0.000 65 in
+0.0038 to −0.0546 mm +0.00015 to 0.002 15 in
b
2-4 85032B/E
Specifications

Electrical Specifications

Electrical Specifications
The electrical specifications in Table 2-3 apply to the devices in your calibration kit when connected with an Agilent precision interface.
Table 2-3 Electrical Specifications for 50Type-N Devices
Devi ce Specification Frequency (G H z)
Loads Return loss 49 dB (ρ ≤ 0.00355) DC to ≤ 2
Return loss ≥ 46 dB (ρ ≤ 0.00501) > 2 to ≤ 3
Return loss ≥ 40 dB (ρ ≤ 0.01000) > 3 to 6 Female open
Female short
Male open
Male s hort Adapters (type-N to 7-mm) Return loss ≥ 30 dΒ (ρ ≤ 0.03162) DC to 6
a. The specifications for the opens and shorts are given as allowed deviation from the nominal
a
a
a
a
model as de fined i n the standard def i nitions. See Table A-3 on page A-5.
±0.501 ° ±0.484 °/GHz deviati on fro m nominal DC to ≤ 6
±0.490 ° ±0.385 °/GHz deviation from nominal DC to 6
±0.501 ° ±0.234 °/GHz deviation from nominal DC to 6
±0.441 ° ±0.444 °/GHz deviation from nominal DC to 6

Certification

Agilent Technologies cer tif i es tha t this product met it s pu blished sp ec ific a tio ns at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST)
to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Agilent Verifi es the
Devices in This Kit” on page 4-2 for more in fo rmation.

Supplemental Electrical Characteristics

Supplemental electrical characteristics are values which are typically met by a majority of the calibration kit devices tested. These supplemental characteristics are intended to provide information in calibrati o n ki t appl ications by giving typical, but non-warra nted, performance parameters. Table 2-4 lists the typical electrical characteristics of the 50 loads and adapters in the 85032B/E calibration kit.
Table 2-4 Supplemental Electrical Characteristics
Devi ce Specification Frequency (G H z)
Loads Return loss ≥ 23 dΒ (ρ ≤ 0.07079) > 6 to 18 Adapters (type-N to 7-mm) Return loss ≥ 34 dΒ (ρ ≤ 0.01995) DC to 8
Return loss ≥ 28 dΒ (ρ ≤ 0.03981) > 8 to 18
85032B/E 2-5
Specifications
Electrical Specifications
2-6 85032B/E
3 Use, Maintenance, and Care of the
Devices
85032B/E 3-1
Use, Maintenance, and Care of the Devices

Electrostatic Discharge

Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test (DUTs), can also carry an electrostatic charge. To prevent damage to the test set, components, and devices:
always wear a grounded wrist strap having a 1 M resistor in series with it when handling components and devices or when making connections to the test set.
always use a grounded, conductive table mat while making connections.
always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
always ground yourself before you clean, inspect, or make a connection to a static-sensitive dev ice or tes t port. You can, for example, grasp the gr ounded outer s hel l of the test port or cable connector briefly.
always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center
conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Refer to Chapter 6 , “Replaceable Parts ,” for part numbers and in st ru ctions for o rdering ESD protection devices.
Figure 3-1 ESD Protection Setup
3-2 85032B/E
Use, Maintenance, and Care of the Devices

Visual Inspection

Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.
In some cases, magnification is necessary to see damage on a connector; a magnifying device with a magnification of 10× is recommended. However, not all defects that are visible only under magnification wi ll affect the elec trical per formance of the connector. Use the following guidelines when evaluating the integrity of a connector.

Look for Obvious Defects and Damage First

Examine the connectors first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are connected incorrectly.
Connector wear eventually d egra des performance of the dev ice. Calibration d evi ce s sho uld have a long life if their use is on the order of a few times per week. Replace devices with worn connectors.
The test port connectors on the network analyzer test s et may have many connections each day, and are therefore more subject to wear. It is recommended that an adapter be used as
a test port saver to minimize the wear on the test set’s test port connectors.

Inspect the Mating Plane Surfaces

Flat contact between the connectors at a ll point s on their mating plane sur faces is required for a good connection. See Figure 2-1 on page 2-3. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance.
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration.
85032B/E 3-3
Use, Maintenance, and Care of the Devices

Cleaning Connectors

Inspect Female Connectors

Pay special attention to the contact fingers in the female center conductor. These can be bent or broken, and damage to them is not always easy to see. A connector with damaged contact fingers will negatively affect electrical performance and must be replaced.
NOTE Inspection is particularly important when mating nonprecision to precision
devices.
Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
1. Use Compressed Air or Nitrogen
WARNING Always use protective eyewear when using compressed air or
nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kP a (60 ps i) to con trol the vel ocity of th e air stream. High- velocity s treams of compressed air can cause electrostatic effects when directed into a connector. These electrostatic effects can damage the device . Refe r to “Electrostatic Discha r ge” earlier in
this chapter for additional information.
2. Clean the Connector Threads
WARNING Keep isopropyl alcohol away from heat, sparks, and flame. Store in a
tightly closed container. It is extremely flammable. In case of fire, u se alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.
3-4 85032B/E
Use, Maintenance, and Care of the Devices
Cleaning Connectors
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to T abl e 6-3 on page 6-5 for part numbers f or isopropy l alcohol and cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab. b. Clean the connector threads. c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
3. Clean the Mat ing Plane Surfaces a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figu re 2-1 on
page 2-3. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate , then blow the connector dry wit h a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
4. Reinspect Inspect the connector again to make sure that no particles or residue are present.
85032B/E 3-5
Use, Maintenance, and Care of the Devices

Gaging Connectors

Gaging Connectors
The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. (See Table 6-3 on page 6-5 fo r part number information.) They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test acces sories , and the c alibration kit devices .
Do not use the g a g es fo r precise pin depth measurements.

Connector Gage Accuracy

The connector gages are only capable of performing coarse measurements. They do not provide the degree of accuracy necessary to precisely measure the pin depth of the kit devices. This is partially due to the repeatabil ity uncertai nties that a re associated with the
measurement. Only the factory—through special gaging processes and electrical testing— can accurately verify the mechanical characteristics of the devices.
With proper te chnique , however, the gages are useful in detecting gros s pin depth err ors on device connectors. To achieve maximum accuracy, random errors must be reduced by taking the average of at least three measurements having different gage orientations on the connector. Even the resultant average can be in error by as much as ± 0.0001 inch du e to systematic (biasing) errors usually resulting from worn gages and gage masters. The information in Table 2-2 on page 2-4 assumes new gages and gage masters. Therefore, these systematic errors were not included in the uncertainty analysis. As the gages undergo more use, the s ystematic er rors can bec ome more significa nt in the a ccuracy of the measurement.
The measurement uncertainties (see Table 2-2 on page 2-4) are primarily a function of the assembly materials and design, and the unique interaction each device type has with the gage. Therefore, these uncertainties can vary among the different devices. For example, note the difference between the uncertainties of the opens and shorts in Table 2-2.
The observed pin depth limits in Table 2-2 on pag e 2-4 add these uncertainties to the typical factory pin depth values to provide practical limits that can be referenced when using the gages. See “Pin Depth” on page 2-3. Refer to “Kit Contents” on page 1-2 for more information on the design of the calibration devices in this kit.
NOTE When measuring pin depth, the measured value (resultant average of three
or more measurements) is not the true value. Always compare the measured value with the observed pin depth limits in Table 2-2 on page 2-4 to evaluate the condition of device connectors.
3-6 85032B/E
Use, Maintenance, and Care of the Devices
Gaging Connectors

When to Gage Connectors

Gage a connector at the following times:
• Prior to using a device for the first time: record the pin depth measurement so that it can be compared with future r eadings. ( It will serve as a good troub leshooting tool when you suspect damage may have occurred to the device.)
• If either visual inspection or electrical performance suggests that the connector interface may be out of typical range (due to wear or damage, for example).
• If a calibration device is used by someone else or on another system or piece of equipment.
• Initially after every 100 connections, and after that as often as experience indicates.

Reading the Connector Gage

The gage dial is divided into increments of 0.0001 inch and major divisions of 0.001 inch (see Figure 3-2). For each revolution of the large dial, the smaller dial indicates a change of
0.01 inch. Use the small dial as the indicator of multiples of 0.01 inch. In most connector
measuring applications, this value will be zero. When making a measurement, the gage dial indi cator will travel i n one of tw o directio ns. If
the center conductor is recessed from the zero reference plane, the indicator will move counterclockwise to i ndicate the amount of recession, whi ch is re ad as a ne g ativ e val u e. I f the center conductor protrudes, the indic ator will move clockwi se to indicate the amount of protrusion, which is read as a positive value. Refer to “Pin Depth” on page 2-3 for definitions of protrusion and recession.
Figure 3-2 Reading the Connector Gage
85032B/E 3-7
Use, Maintenance, and Care of the Devices
Gaging Connectors

Gaging Procedures

Gaging Male Type-N Connectors
NOTE Always hold a connector gage by the gage barrel, below the dial indicator.
This gives the best stability, and improves measurement accuracy.
1. Select the proper gage for your connector. (Refer to Table 6-3 fo r gage par t nu mbers).
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-3): a. While holding the gage by the barrel, and without turning the gage or the gage
master, screw the gage master connecting nut onto the male gage, just until you meet resistance. Connect the nut finger tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting
nut to 135 N-cm (12 in-lb). Refer to “Connections” on page 3-12 for more information.
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer
corresponds to the correction value noted on the gage master. Do not adjust the gage
dial to zero, unless the correction value on the gage master is zero. d. Tighten the dial lock screw and remove the gage master. e. Attach and torque the gage master to the gage once again to verify that the setting is
repeatable. Remove the gage master.
4. Gage the device connector (refer to Figure 3 - 3): a. While holding the gage by the barrel, and without turning the gage or the device,
screw the connecting nut of the device being measured onto the gage, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting
nut to 135 N-cm (12 in-lb). Refer to “Connections” on page 3-12 for more information. c. Gently tap the barrel of the gage with your finger to settle the gage reading. d. Read the gage indicator dial. If the needle has moved clockwise , the center conductor
is protruding by an amount indicated by the black
numbers. If the needle ha s moved
counterclockwi se, the center conduct or is recessed by an amount indicated by the red
numbers.
For maximum accuracy, measure the connector a minimum of three times and take
an average of the readings. After each measurement, rotate the gage a quarter-turn
to reduce measurement variati ons that result f rom the gage or t he connector face not
being exactly perpendicular to the center axis. e. Compare the average reading with the o bserved pin d epth limits i n T a ble 2-2 on page
2-4.
3-8 85032B/E
Figure 3-3 Gaging Male Type-N Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85032B/E 3-9
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Female Type-N Connectors
NOTE Always hold a connector gage by the gage barrel, below the dial indicator.
This gives the best stability, and improves measurement accuracy.
1. Select the proper gage for your connector. (Refer to Table 6-3 fo r gage par t nu mbers).
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-4): a. While holding the gage by the barrel, and without turning the gage or the gage
master, screw the gage connecting nut onto the female gage master, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting
nut to 135 N-cm (12 in-lb). Refer to “Connect ions” on pag e 3-12 for more in fo rmation .
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer
corresponds to the correction value noted on the gage master. Do not adjust the gage
dial to zero, unless the correction value on the gage master is zero. d. Tighten the dial lock screw and remove the gage master. e. Attach and torque the gage master to the gage once again to verify that the setting is
repeatable. Remove the gage master.
4. Gage the device connector (refer to Figure 3 - 4): a. While holding the gage by the barrel, and without turning the gage or the device,
screw the gage connecting nut onto the device being measured, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting
nut to 135 N-cm (12 in-lb). Refer to “Connections” on page 3-12 for more information. c. Gently tap the barrel of the gage with your finger to settle the gage reading. d. Read the gage indicator dial. If the needle has moved clockwise , the center conductor
is protruding by an amount indicated by the black
numbers. If the needle ha s moved
counterclockwi se, the center conduct or is recessed by an amount indicated by the red
numbers.
For maximum accuracy, measure the connector a minimum of three times and take
an average of the readings. After each measurement, rotate the gage a quarter-turn
to reduce measurement variati ons that result f rom the gage or t he connector face not
being exactly perpendicular to the center axis. e. Compare the average reading with the o bserved pin d epth limits i n T a ble 2-2 on page
2-4.
3-10 85032B/E
Figure 3-4 Gaging Female Type-N Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85032B/E 3-11
Use, Maintenance, and Care of the Devices

Connections

Connections
Good connections require a skill ed op erator. The most comm o n caus e of me asu rement error is bad connections. The following procedures illustrate how to make good connections.

How to Make a Connection

Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table mat. Refer to “Electrostatic Discharge” on page 3-2 for ESD
precautions.
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-4.
4. Use a connector gage to verify that all center conductors are within the observed pin depth values in Table 2-2 on pa ge 2-4. Refer to “Gaging Connectors” on page 3-6.
5. Carefully align the connectors. The male connector center pin must slip concentrically into the contact finger of the female connector.
6. Push the connectors straight together.
CAUTION Do not turn the device body. Only turn the connector nut. Damage to the
center conductor can occur if the device body is twisted.
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables.
Final Connection Using a Torque Wrench
1. Use a torque wrench to make a final connection. Table 3-1 provides information about the torque wrench recommended for use with this calibration kit. A torque wrench is not included in the calibration kit. Refer to Chapter 6 for part numbe r an d ordering inform at ion.
Table 3-1 Torque Wrench Inform ation
Connector Type Torque Setting Torque Tolerance
Type-N 135 N-cm (12 in-lb) ±13.5 N-cm (±1.2 in-lb)
3-12 85032B/E
Use, Maintenance, and Care of the Devices
Connections
Using a torque wrench guarantees that the connection is not too tight, preventing possible connector damage. It also guarantees that all connections are equally tight each time.
2. Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible to do this by hand if one of the connectors is fixed (as on a test port). In all situations, however, it is recommended that you use an open-end wrench to keep the body of the device from turning. Refer to Chapter 6 for part nu mb e r an d o rdering information.
3. Position both wrenches within 90 degrees of each other before applying force. See
Figure 3-5 . Wrenches opposing each other (greater than 90 degrees apart) will cause a
lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together.
Figure 3-5 Wrench Positions
4. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See
Figure 3-6 .
Figure 3-6 Using the Torque Wrench
85032B/E 3-13
Use, Maintenance, and Care of the Devices
Connections
5. Apply downwa rd fo rce perpendicular t o the wrench handle. See Figu re 3- 6 . This applies torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque.
6. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-6. D o not tighten the connection further.
CAUTION You don’t have to fully break the handle of the torque wrench to reach the
specified torque; doing so can cause the handle to kick back and loosen the connection. Any give at all in the handle is sufficient torque.
Do not pivot the wrench handle on your thumb or other fingers , other wis e you apply an unknown amount of torque to the connection when the wrench reaches its break point.
Do not twist the head of the wrenc h r el ative to the outer conductor mating pla ne. If you do, you apply more than the recommended torque.
Connecting and Disconnecting the Two-Piece Female Open (85032B)
The female open standard in the 85032B cali b rati on kit i s composed of two parts: the open body (outer conductor) and the center conductor extender. Refer to Figure 3-7.
Figure 3-7 Connecting the Two-Piece Female Open
To conne ct the female open:
1. Connect the open body to the male test port.
2. Insert the center conductor extender into the hole at the end of the body and push gently until the center conductors mate.
To disconnect the female open:
1. Remove the center conductor extender by pulling gently outwards without twisting, rocking, or bending the extender or the body.
2. Disconnect the body from the test port.
3-14 85032B/E
Use, Maintenance, and Care of the Devices

Handling and Storage

How to Separate a Connection

To a void l ateral (bending) f orce on the connecto r mating plane surf aces, alw ays support the devices and connections.
CAUTION Turn the connector nut, not the device body. Major damage to the center
conductor can occur if the device body is twisted.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connector nut.
3. Complete the separation by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.
Handling and Storage
• Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use.
• Nev er store co n nectors lo o se in a bo x, desk, or bench drawe r. This is the mo st co mmon cause of connector damage during storage.
• Keep connectors clean.
• Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove.
• Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.
85032B/E 3-15
Use, Maintenance, and Care of the Devices
Handling and Storage
3-16 85032B/E

4 Performance Verification

85032B/E 4-1
Performance Verification

Introduction

Introduction
The performance of your calibra tion kit can only be verified b y retur ning the kit to Agi len t Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.

How Agilent Verifies the Devices in This Kit

Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to NIST (National Institute of Standards and Technology). Th e airline and sh o rt characterist ics are develo pe d fro m mechan ical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representa tion. The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) testing of the ca li brat ion de vi ces , Agilent includes the test measurement uncer tai nty a s a guardband to guarantee each device meets the published specification. For recertifications (after sale), no guardband is used and the measured data is compared directly with the specification to determine the pass or fail status. The measurement uncertainty for each device is, however, recorded in the calibration report that accompanies recertified k it s.
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the
institute’s calibration facility. The specifications data provided for the devices in this kit is traceable to NIST through Agilent Technologies.
4-2 85032B/E
Perf ormance Verification

Recertification

Recertification
The following will be provided with a recertified kit:
• a new calibration sticker affixed to the case
• a certificate of calibrat i on
• a calibration report for each device in the kit listing measured values, specifications, and uncertainties
NOTE A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
Agilent Technologies offers a Standard calibration for the recertification of this kit. For more information, contact Agilent Technologies. See Table 5 -1 o n page 5-3.

Limited Recertification (Option 003)

The 50 loads in the calibration kit are specified for use up to 6 GHz. For many applications, the performance above 3 GHz is not utilized.
For a standard recertification, the devices in the calibration kit are tested and calibrated up to 6 GHz. However, a limited recertification can be requested. For this limited recertification, the devices are tested and calibrated up to 3 GHz.
To reques t a limited recer tificati on, make sure the following is clearly written on the order:
Limited Calibration DC–3 GHz (Option 003). All loads that receive a limited calibration are supplied with a limited calibration label
applied to the device.

How Often to Recertify

The suggested initial inter val fo r recertifi cation i s 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establi sh a different recertif ication inter val that reflects the usage and wear of the kit.
NOTE The recertification interva l should begin on the date the kit is first used after
the recertification date.

Where to Send a Kit for Recertification

Contact Agilent Technologies for information on where to send your kit for recertification. See Table 5-1 on pa ge 5-3.
When you return the kit, complete and attach a service tag. Refer to “Returning a Kit or
Device to Agilent” on page 5-3 for details.
85032B/E 4-3
Performance Verification
Recertification
4-4 85032B/E

5 Troubleshooting

85032B/E 5-1
Troubleshooting

T roubleshooting Process

Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.
Figure 5-1 Troubleshooting Flowchart
5-2 85032B/E
Troubleshooting

Returning a Kit or Device to Agilent

Returning a K it or Device to Agilent
If your kit or device requir es service , conta ct the Agile nt T ec hnologies offi ce nearest yo u for information on where to send it. See Table 5-1. Include a service tag (located near the end of this manual) on which you provide the following information:
• your company name and address
• a technical contact person within your company, and the person's complete phone number
• the model number and serial number of the kit
• the part number and serial number of each device
• the type of service required
•a detailed description of the problem and how the device was being used when the problem occurred (such as calibration or measurement)
Table 5-1 Contacting Agilent
Online assistance: www.agilent.com/find/assist United States
(tel) 1 800 452 4844
New Zealand
(tel) 0 800 738 378 (fax) (+64) 4 495 8950
Malaysia
(tel) 1 800 828 848 (fax) 1 800 801 664
Taiwan
(tel) 0800-047-866 (fax) (886) 2 25456723
Latin America
(tel) (305) 269 7500 (fax) (305) 269 7599
Japan
(tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Philippines
(tel) (632) 8426802 (tel) (PLDT subscriber only):
1 800 16510170
(fax) (632) 8426809 (fax) (PLDT subscriber only):
1 800 16510288
People’s Republic of China
(tel) (preferred ): 800-810-0189 (tel) (alternate): 10800-650-0021 (fax) 10800-650-0121
Canada
(tel) 1 877 894 4414 (fax) (905) 282-6495
Australia
(tel) 1 800 629 485 (fax) (+6 1) 3 9210 5947
Thailand
(tel) outside Bangk ok :
(088) 226 008
(tel) with in Bangko k :
(662) 661 3999 (fax) (66) 1 661 3714
India
(tel) 1-600-11-2929 (fax) 000-800-650-1101
Europe
(tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
Singapore
(tel) 1 800 375 8100 (fax) (65) 836 0252
Hong Kong
(tel) 800 930 871 (fax) (852) 2506 9233
85032B/E 5-3
Troubleshooting
Returning a Kit or Device to Agilent
5-4 85032B/E

6 Replaceable Parts

85032B/E 6-1
Replaceable Parts

Introduction

Introduction
Table 6-1 lists the replacement part numbers for items included in the 85032B calibration
kit and Figure 6-1 illustrates each of these items.
Table 6-2 lists the replacement part numbers for items included in the 85032E calibration
kit and Figure 6-2 illustrates each of these items.
Table 6-3 lists the replacement part numbers for items recommended or required for
successful operation but not included in the calibration kit. To order a listed part, note the description, the part number, and the quantity desired.
Telephone or send your order to Agilent Technologies. See Table 5-1 on page 5-3.
Table 6-1 Repla cea ble Parts for the 85032B Calibration Kit
Item No. Description Qty Per Kit Agilent Part Number
Calibration Devices (50 Type-N)
1 Male broadband load 1 00909-60009 2 Female broadband load 1 00909-60010 3 Male short 1 85032-60008 4 Female short 1 85032-60009 5 Male open 1 85032-60007 6
7 Type-N-male to 7-mm 2 85054-60009 8 Type-N-female to 7-mm 2 85054-60001
9
10 11
12 Female end cap for type-N as required 1401-0225 13 Male end cap for type-N and 7 mm as required 1401-0214
Female open
Box assembly (incl udes case and foam pad set)
Case (without foam pad set)
Foam pad set
a
Adapters (included with Option 100)
Calibration Kit Storage Case
c
c
Protective End Caps for Connectors
1 85032-60012
b
1 85032-60010 1 85032-80002
1
85032-80003
Miscellaneous Items
15 User’s and service guide 1 85032-90020
a. Includes center conductor extender. b. Refer to “Options” on page 1-3 for description of available op tions.
c. Included in box assembly.
6-2 85032B/E
Figure 6-1 Replaceable Parts for the 85032B Calibration Kit
Replaceable Parts
Introduction
85032B/E 6-3
Replaceable Parts
Introduction
Table 6-2 Replaceable Parts for the 85032E Calibration Kit
Item No. Description Qty Per Kit Agilent Part Number
Calibration Devices (50 Type-N)
1 Male broadband load 1 00909-60009 2 Male combination open/short 1 85032-60011
Calibration Kit Storage Case
3 Case (without foam pad set) 1 9211-1582 4 Foam pad set 1 85023-80005 5 Kit identification label 1 85032-80014
Protective End Caps for Connectors
6
7 U ser’s and service guide 1 85032-90020
Male end ca p for type-N and 7 mm as
required
Miscellaneous Items
1401-0214
Figure 6-2 Replaceable Parts for the 8 5032E Calibration Kit
6-4 85032B/E
Replaceable Parts
Introduction
Table 6-3 Replaceable Parts—Items Not Included in the Calibration Kit
Item No. Descript ion Qty Agilent Part Number
Connector Gages
1 G age set (includes items listed below) 1 85054-60049 2 Female gage 1 85054-60050 3 Female gage master 1 85054-60052 4 Male gage 1 85054-60051 5 6 Centering bead (2 su pp lied with gage set) 1 85054-80028
7 3/4 in, 135 N-cm (12 in-lb) torque wrench 1 8710-1766 8 1/2 in and 9/16 in open-end wrenc h 1 8710-1770 9 3/4 in open-end wrench 1 8720-0011
10 Grounding wrist strap 1 9300-1367 11 5 ft grounding cord for wrist s trap 1 9300-0980 12 2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300-0797
Male gage master
Wrenches
ESD Protec t i ve Device s
a
(Type-N)
1 85054-60053
13 ESD heel strap 1 9300-1308
Connector Cleaning Supplies
14 Isopropyl alcohol 30 ml 8500-5344 15 Foam tipped cleaning swabs 100 9301-1243
a. To ens u re you choose the correct gage, refer to , “Clarifying the Sex of a Connector” on page 1-6.
85032B/E 6-5
Replaceable Parts
Introduction
6-6 85032B/E
A Standard Definitions
85032B/E A-1
Standard Definitions
N 50

Standard Class Assignments

Standard C l ass Assignments
Class assignment organizes calibration standards into a format compatible with the error models used in the measu rement c alibration. A class or group of classe s corr espo nds t o the systematic errors to be removed from the measured network analyzer response. T able A-1 lists the classes of the devices in this calibration kit.
Table A-1 Standard Class Assignments
Calibration Kit Label:
Class A B C D E F G Standard Clas s Labe l
A2
S
11
S
B17 Shorts
11
S
C3 Load
11
S
A28 Opens
22
S
B17 Shorts
22
S
C3 Load
22
Forward Transmi ssi o n 4 Thru Reverse Transmission 4 Thru Forward Match 4 Thru Reverse Mat ch 4 Thru Response 1 7 2 8 4 Response Response and Isolation 1 7 2 8 4 Response & Isolation
8
Opens
A-2 85032B/E
Standard Definitions
Standard Class Assignments

Blank Form

The standard class assignments listed in Table A-1 may be changed to meet your specific requirements. Table A-2 is provided to record the modified standard class assignments.
Table A-2 Standard Class Assignments Blank Form
Calibration Kit Label: ______________
Class A B C D E F G Standard Class
Label
S
A
11
S
B
11
S
C
11
S
A
22
S
B
22
S
C
22
Forward Transmissio n Reverse Transmission Forward Match Reverse Match Response Response and Isolation
85032B/E A-3
Standard Definitions

Nominal Standard Definitions

Nominal Standard Definitions
Standard definitions provide the constants needed to mathematically model the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. The nominal values of these constants are theoretically derived from the physical dimensions and material of each cali bration standard , or from actua l measured response. These values are used to determine the measurement uncertainties of the network analyzer. The standard definitions in Table A-3 list typica l c ali brat ion ki t p arameter s used to sp ec ify the mathematical model of each device. This information must be loaded into the network
analyzer to perform valid calibrations. Refer to your network analyzer’s user’s guide for instructions on loading calibration constants.
NOTE The values in the standard definitions table are valid only over the specified
operating temperature range.

Setting the System Impedance

This kit contains only 50 ohm devices. Ensure the system impedance (Z0) is set to 50 ohms. Refer to your network a na ly zer’s user’s guide for i ns tr uctions on setting system im pedance .
A-4 85032B/E
Table A-3 Standard Definitions
Standard Definitions
Nominal Standard Definitions
System Z
Standard
1 2 3
4
5 6 7 8
a
= 50
0
b
F
15
Number
Type
Short Open Load Fixed
Delay/ Thru
Short Open
C0 ×10
119.09 −36.955 26.258 5.5136 0 50 700Μ 0 999
62.140 −143.07 82.920 0.7600 17.411 50 700Μ 0 999
F/Hz
27
C1 ×10
2
F/Hz
36
C2 ×10
Calibration Kit Label:
c
3
F/Hz
45
C3 ×10
Fixed or Sliding
0.093 49.992 700Μ 0 999
17.817 50.209 2.1002
Offset Freq
Delay (ps)
0 50 700Μ 0 999
0 50 700Μ 0 999
0
Z
N 50
(GHz)
Loss (Ω/s)
G
0999
Min
Max
Coax Short (m) Coax Open (m) Coax Broad-
Coax Thru
Coax Short (f) Coax Open (f)
d
Coax or Waveguide
Standard Label
band
a. Ensure system impedance (Z
) of network anal y z er is set to this value .
0
b. Open, short, load, delay/thru, or arbitrary impedance. c. Load or arbitrary impedance only. d. Standard labels that specify sex, (m) or (f), refer to the sex of the analyzer’s test port connector.
85032B/E A-5
Standard Definitions
Nominal Standard Definitions

Blank Form

The standard definitions listed in Table A-3 may be changed to meet your specific requirements. Table A-4 is provided to record the modified standard definitions.
Table A-4 Standard Definitions Blank Form
System Z
Standard
Number
1 2 3 4 5 6 7 8
a
= ______________
0
b
F
15
Type
C0 ×10
Calibrati on K it Label:
____________________
Offset Freq
2
F/Hz
F/Hz
27
36
C1 ×10
C2 ×10
c
3
F/Hz
45
C3 ×10
Fixed or Sliding
Delay (ps)
0
Z
Loss (GΩ/s)
(GHz)
Min
Max
d
Coax or Waveguide
Standard Label
a. Ensu r e syst e m Z
of network analyzer is set to this value .
0
b. Open, short, load, delay/thru, or arbitrary im ped ance. c. Lo ad or arbitrary impedanc e only. d. Standard labels that specify sex, (m) or (f), refer to the sex of the analyzer’s test port connector.
A-6 85032B/E

Index

A
adapters Agilent Technologies, contacting alcohol
altitude
B
blank form
broadband loads
C
cal kit
calibration
calibration co nst ants
calibration definitions calibration kit
calibration label calibration re port
calibration sticker cent er con duct or
certificate of calibration certification
characteristics
class assignment class assignments
, 1-3
part numbers
5-3
isopropyl
as cleaning solv ent
, 6-5
, 3-4
, 2-2
standard cla ss assi gnme nts standard definitions
, A-6
, 1-2
contents overview return to Agile nt serial number
bad certificate of constants
limited standards temperature
entering permanently stored
, 1-2
, 1-2
, 5-3
, 1-5
, 5-2
, 1-3, 4-2
, 1-3
permanently stored
, 1-3, 4-3
, 2-5
, 2-2
, 1-3
, 1-3
, 1-3
, 1-3
contents overview return to Agile nt serial number
part number
, 1-2
, 1-2
, 5-3
, 1-5
, 6-5
, 4-2
, 4-3
protrusion recession
, 3-7
, 3-7
, 4-2
specifications mechanical
supplemental standard blank form
, 2-5
, 2-3
, 2-3
, A-2
, A-3
, A-3
cleaning connectors cleaning supplies
ordering
,
part numbers compressed air or nitrogen conductor
mating plane
conductor protrusion recession
connecting the two-piece female
open
, 3-14
connections
cautions in making disconnecting ESD concerns final
, 3-12
how to make preliminary two-piece female open undoing using a torque wrench
connector
cleaning cleaning supplies damage female gage
gaging life
male mating plane sex threads visual inspection wear
connector g age
handling master reading zeroing
connector g age acc u ra cy constants
calibration
contacting Agilent Technologies contents
kit
D
damage
shipment
, 3-15
, 3-3
, 3-4, 3-10
dial
, 3-7
, 3-6
when to do
, 3-3
, 3-8
, 1-6
, 3-4
, 3-3
, 3-8, 3-10
, 3-7
, 3-8, 3-10
entering permanently st ored
5-3
, 1-2
, 1-3
, 6-5
, 6-5
, 2-3
, 2-3
, 2-3
, 3-2, 3-12, 3-15
, 3-15
, 3-12
, 3-12
, 3-12
, 3-4
, 3-7
, 3-5
, 3-8, 3-10
, 1-3
, 1-3
, 1-4
, 3-4
, 3-12
, 6-5
, 3-3
, 3-14
, 3-12
, 3-6
, 1-3
, 3-4
to connectors damaged connectors data, recertification defective connectors definitions
standard deviation from nominal phase device
conductor
mating plane connecting disconnecting handling maintenance part numbers pin depth return to Agile nt storage temperature visual inspection
devices
how Agilent verifies serial numbers
dial
connector gage
dimensions
device
center con ductor
oute r co n ductor
disconnecting the two-piece disconnection
two-piece female open
disconnection s documentation warranty
E
electrical characteristics
supplemental
electrical specifications electrostatic discharge
supplies
part num ber s when making connections
environmental requirements equipment
required
but not supp l i ed
,
supplied
ESD
precautions supplies
part num ber s when making connections
F
female open
, A-4
, 3-12
, 3-15
, 2-3
, 3-15
female open
, 1-3 , 1-2, 6-2, 6-4
, 3-2
, 3-2, 3-4
, 3-14
, 3-3
, 3-3
, 4-2
, 3-3
, 2-5
, 2-3
, 3-15
, 1-6
, 6-5
, 5-3
, 2-2
, 3-3
, 4-2
, 1-5
, 3-7
, 2-3
, 2-3
, 3-14
, 3-14
, 3-15
, ii
, 2-5
, 2-5
, 3-2
, 6-5
, 3-12
, 2-2
, 1-3, 6-2, 6-5
, 6-5
, 3-12
85032B/E Index-1
Index
connecting, 3-14 disconnecting
frequency
specifications
frequency range
G
gage
connector
dial
, 3-7
handling master reading zeroing
gage master
part numbers using
, 3-8, 3-10
gages
, 1-3
gaging
female con nectors male connectors procedures
gaging connectors
when to do
H
handling humidity
I
impedance
system incoming inspection inspection
damage
defects
female con nectors
incoming
mating plane
visual
, 3-3
isopropyl alcohol
as cleaning solvent
K
kit
contents
overview
return to Agilent
serial number
L
label
calibration
part number
limited calibration
, 3-14
, 2-5
, 2-5, 4-3
, 3-8, 3-10
, 3-8, 3-10, 6-5
, 3-7
, 3-8, 3-10
, 6-5
, 3-8
, 3-8
, 3-6
, 3-7
, 3-15
, 2-2
, A-4
, 3-3
, 3-3
, 1-4
, 3-3
, 1-2
, 1-2
, 1-5
, 4-3
, 6-5
, 3-10
, 1-4
, 3-4
, 3-4
, 5-3
, 4-3
limited calibration
requesting
loads
broadband
M
maintenance
preventive maintenance of devices making connections manual
part number mati ng plane
conductor
connector mating plane inspection mating plane surfa ces mechanical characteristics mechanical integrity
N
nitrogen nominal standard d efinitions numbers
serial
record i n g
O
open-end wren ch
part number opens open s a n d short s options orderin g pa r ts
P
part numbers parts
ordering
replaceable
required but not supplied perf orman c e v e rifica ti on
fail permanently stored calibration
pin depth
definition of
effect on electrical
importance of
observed limits
typical values preventive maintenance protrusi on
center conductor
conductor
, 3-5
, 3-4
, 1-5
, 1-2
, 1-3
, 6-2
, 5-2
definitions
, 2-3
specifications
, 1-3
, 4-3 , 1-2
, 3-2
, 1-6
, 1-6
, 3-12
, 6-5
, 2-3
, 3-3
, 3-5
, 2-3
, 2-3
, A-4
, 1-5
, 1-3, 3-15
, 6-5
, 1-2
, 6-2
, 6-2
, 6-2
, 6-2
, 4-2
, 1-3
, 2-3
, 2-4
, 2-4
, 2-4
, 2-4
, 1-6
, 3-7
, 2-3
R
reading connector gage recertification
how often? interval limited what’s included where to send your kit
recession
center conductor conductor
regulations
environmental replaceable parts requirement s
environmental retu rn lo ss
specifications retu rn , ki t or d e v i ce
S
serial numbers
devices
recording service service tag sex
connector shipment
damage
verifying complete shorts
, 1-2
specifications
altitude
certification
electrical
environmental
frequency
humidity
pin depth
return loss
temperature
verifying standard cl ass ass ig nm ents
blank form standard definition s
blank form
nominal standards
calibration
NIST static discharge sticker
calibration storage
altitude
humidity
, 4-3
, 4-3
, 4-3
, 4-3
, 2-3
, 2-5
, 1-5
, 1-5
, 1-5
, 5-2
, 1-4, 4-3, 5-3
, 1-6
, 1-4
, 2-2
, 2-2
, 2-5
, 2-5
, 2-5
, 2-2
, 2-4
, 2-5
, 2-2
, 4-2
, A-3 , A-6
, A-4
, 2-5
, 2-5, 4-3
, 3-2
, 4-3
, 3-15
, 2-2
, 2-2
, 3-7
, 4-3
, 4-3
, 3-7
, 3-5
, 6-2
, 2-2
, 5-3
, 1-4
, 2-2
, A-4, A-6
, A-2
Index-2 85032B/E
Index
temperature, 2-2
supplemental char acteristics supplemental elec trical
characteri stics
supplies
cleaning
system impedance
setting
T
tag
service
temperature
calibration device error -corrected te mp erature
measurement operating range verification and measurement
test data threads
connector
torque wrench
part number
specifications traceability troubleshooting two-piece f em a le open
, 1-3
, A-4
, 1-4, 4-3, 5-3
, 2-2
range
2-2
, 4-2
, 3-4
, 2-5
, 2-2
, 2-2
, 2-2
, 2-2
, 1-3, 3-12
, 6-5
, 3-12
, 4-2
, 5-2
, 3-14
, 2-3
,
V
verification
performance
temperature visual inspection
W
warranty, documentation wrench
open-end
part num ber undoing connections
torque
part num ber
wrenches
part numbers
Z
zeroing
connector gage
zeroing co n nector gag e
3-10
85032B/E Index-3
, 4-2 , 2-2
, 3-3
, ii
, 1-3, 3-15
, 6-5
, 3-15
, 1-3
, 6-5
, 6-5
, 3-8, 3-10
, 3-8,
Loading...