Agilent Technologies 81200 User Manual

Agilent 81200
The 81200 Data Generator/ Analyzer Platform
Configuration Guide Release 2.1
Simplify your verification and characterization process
The Agilent Technologies 81200 data generator/analyzer platform
a) Using the 81200 as a "proprietary system" means that the 81200 will not be combined with other VXI modules in a standard VXI system. It can how­ever, be controlled by a LAN or a GPIB interface. If a "proprietary sys­tem" is desired, follow steps 1-3, 4a and 5 to configure your system (fig­ure 1).
b) If Agilent 81200 modules are com­bined with other VXI modules then an "open VXI system" (the standard VXI system), is achieved. For this configuration, please follow steps 1-3, 4b and 5 (figure 1).
For information on upgrading an existing Agilent 81200, please refer to page 8.
On each page you'll find tables where you can enter your choice as a reminder for ordering.
STEP 1: Selecting the number of channels required
Figure 2 shows that three different generator front-ends and four differ­ent analyzer front-ends are available. To select the correct front-end, the following should be checked:
• speed
• data format
• levels
• memory depth
Please note that some of the lower speed front-ends have two outputs or two inputs and support a maximum memory depth of 512 Kbit per chan­nel, not 1024 Kbit. SMA cables are not included.
For more details, consult the Agilent
81200 Data Generator/Analyzer Platform, data sheet, p/n 5965-3415E.
STEP 2: Choosing the modules
The generator and analyzer front­ends can be fitted together as follows:
• Generator front-ends can be fitted (any mix) in the Agilent E4831A Clock and Data Generator Module or the Agilent E4841A Data Generator/Analyzer Module.
• Analyzer front-ends only fit into the Agilent E4841A module. Any mix of analyzer and generator front-ends is acceptable, provided that the same sequence is needed for the generator and analyzer.
• At least one clock module is essen­tial; either an Agilent E4805A Central Clock Module or the Agilent E4831 Clock & Data Generator Module. These modules can drive up to eleven or six Agilent E4841A modules respec­tively.
• The sequence of segments and the segment types (pattern, pause, PRBS/PRWS) is the same for all the channels within one module. For example, if you need to set up PRBS and control channels, make sure you run PRBS from one mod­ule and the control signals from another module.
Figure 1: Configuring an Agilent 81200 Data Generator/Analyzer Platform
2
STEP 1 STEP 2
Front-ends
The front-ends determine generator/analyzer function and input/output characteristics.
System Configuration
You are then ready to configure your system using the attached Agilent 81200 Configuration Sheets on page 9.
Modules
A choice of modules provide front-end-housing, clock and sequencing capability, or or combination of these features.
STEP 3
Decision: “Open VXI System”
Integrating the Agilent 81200 modules with other VXI test equipment to a open VXI system or use the Agilent 81200 as “proprietary system”
Propretiary System
STEP 4a
Mainframe(s) & System Configuration
There is a choice of two mainframes. Both include software, PC, hard disk and floppy drive. The larger frame can be augmented with one or two expander frames.
STEP 4b
Requirements for intergration into an “open VXI system”:
VXI controller and slot 0 commander recommendations and additional requirements for operating the Agilent 81200 modules in a open VXI environment.
STEP 5
Choose your user interfaces, support options, hardware and software accessories
3
Product Description Model Number Quantity
Generator front-ends Generator front-ends 660MHz, RZ/NRZ, 3.5 Vpp, variable transition times SMA (f). Agilent E4838A
Generator front-end 330 MHz, RZ/NRZ, variable transition times SMA (f). Agilent E4842A Generator front-end 660 MHz, RZ/NRZ, single channel, 2.5 Vpp SMA (f). Agilent E4843A Dual generator front-end 200 Mbit/s, NRZ, 3.5 VPP SMA (f). Agilent E4846A
Analyzer front-ends Differential analyzer front-end 660 Msa/s, 1 GHz bandwidth, 50SMA (f). Agilent E4837A
Single analyzer front-end 660 Msa/s, 1 GHz bandwidth, 50SMA (f). Agilent E4844A Dual analyzer front-end 330 Msa/s, 1 GHz bandwidth, 50SMA (f). Agilent E4845A High impedance dual analyzer front-end 330 Msa/s, 350 MHz bandwidth, 50SMA (f). Agilent E4847A
Modules 660 MHz data generator/analyzer module – holds four front-ends, any mix Agilent E4841A
660 MHz clock and data generator module – holds two generator front-ends, any mix.
Clocks up to six Agilent E4841A modules. Agilent E4831A
660 MHz central clock module, which clocks up to eleven Agilent E4841A modules Agilent E4805A
and up to two Agilent E4805s.
Required in all systems that include analyzer front-ends, and for all systems with seven Agilent E4805A
or more Agilent E4841As.
8-line trigger input for TTL signals. When branching on external events Agilent E4805A
(hardware signals) other than VXI-ECL trigger lines or compare errors is required. Opt 002
Deskewprobe, includes Agilent 1144A 880 MHz active probe and a BNC (f) to SMA (m) Agilent E4805A
adapter (part number 1250-1200). Opt 003
Which clock module?
The Agilent E4831A Clock & Data Generator Module can be an eco­nomic way of building pulse or data generators as it includes slots for two front-ends. Please remem­ber that the Agilent E4831A can only be used for generator chan­nels. For applications needing many channels, analyzer channels or more than two looping levels, use the Agilent E4805A Central Clock Module.
STEP 3: Decision - VXI integration or not
If you want to combine the Agilent 81200 modules with other VXI modules to achieve an "open VXI system," then go to STEP 4b.
If you want to use the Agilent 81200 as a “proprietary system,” meaning the Agilent 81200 will not be combined with other VXI modules in one system, then go to STEP 4a.
STEP 4a: Selecting the mainframe
If you’ve made the decision to config­ure a "proprietary system," you are now in the position to select a main­frame. Two mainframes are available, the Agilent E4840A small frame with three free slots and the Agilent E4849B mainframe with 10 free slots. Both mainframes include a built-in PC running Windows NT
4.0®. The user software is also installed. All modules and front-ends ordered with the mainframes come factory installed.
Figure 2: Front-ends and modules
Figure 3: Mainframes and expanders
Generator-only systems.
(Two looping levels)
Agilent E4838A 660 MHz/MRZ
Agilent E4842A 330 MHz RZ/NRZ
(variable transitions)
Agilent E4843A 660 MHz RZ/NRZ
Agilent E4846A Dual 200 Mb/s, NRZ
Front - ends Clock module Data generator/
+
Agilent E4831A Clock & data generator module
+
Generator systems or generator/analyzer systems.
Analyzer front ends
(Five looping levels)
analyzer modules
Up to six Agilent E4841A modules can be added when more generator front-ends are needed.
Agilent E4837A 660 MSa/s, differential
Agilent E4844A 660 MSa/s
Agilent E4845A Dual 330 MSa/s
Agilent E4847A High-impedance dual 330 MSa/s
Up to 8 front-ends
Agilent E480A
3 free slots
3-slot PC
9 free slots
VXI bus
extender
Embedded PC
+
VXI bus
extender
12 free slots
Agilent E4805A Central clock module
Up to 36 front-ends
Agilent E4849B
3-slot PC
10 free slots
(+)
VXI bus
extender
12 free slots
+
Agilent E4841A Data generator/ analyzer module(s)
4
Expander frames
For more module slots, one or two Agilent E4848B expander frames can be connected. To do this, a VXI exten­der module (Agilent E4849B Opt 002) is required. The module occupies one of the otherwise free slots and is con­nected to a corresponding VXI exten­der that is supplied as part of the expander frame.
Note that, when one or two expander frames are fitted, an Agilent E4805A Central Clock Module must be includ­ed in all frames. The Agilent E4831A Clock & Data Generator Module is unsuitable for systems with expander frames.
Personalized system configuration
Now you are in a position to config­ure your system to meet your specific requirements with the Agilent 81200 Configuration Sheets. For configura­tion instructions, please consult the box on page 9. If you do not fill out the configuration sheets, your system will be delivered with the convention­al configuration pre-installed. Now go to STEP 5.
STEP 4b: Integrating Agilent 81200 modules with other VXI test equip­ment
To run the Agilent 81200 modules in a standard VXI test system with other VXI test equipment, it is necessary that a controller with the operating system Windows NT 4.0® is present, on which the Agilent E4873A User Software for the Agilent 81200 can run. Therefore, either an embedded VXI controller, or a Slot 0 command module has to be installed in the mainframe. The command module has to be connected to an external PC, on which the Agilent E4873A User Software is running. For operat­ing the Agilent 81200 platform in a VXI test system, the following list, comprising of VXI controllers, com­mand modules and frames, is recom­mended. Please also refer to figure 4.
VXI Controller:
• Agilent E6235A VXI PC
• National Instruments VXIpc-850
Command modules:
• Agilent E1406A GPIB Slot 0 command module
• Agilent E8491B IEEE-1394 PC link to VXI
• National Instruments MXI-2 Interface
Frames:
• Agilent E8403A VXI Mainframe, C-Size
Please consider the power require­ment of the modules and front-ends when you configure your VXI test sys­tem. For details about power require­ments, consult Agilent 81200 Data Generator / Analyzer Platform, data sheets, p/n 5965-3415E, or download the power requirement calculation table (MS Excel) for the Agilent 81200 from:
www.agilent.com/find/81200_configinfo
It is necessary to order the Agilent E4873A User Software for the Agilent 81200 separately.
Expander frames
For more module slots, one or two Agilent E4848B expander frames can be connected. To do this, a VXI exten­der module Agilent E1482B is required. The module occupies one of the otherwise free slots and is con­nected to a corresponding VXI exten­der that is supplied as part of the expander frame.
Note that, when one or two expander frames are fitted, an Agilent E4805A Central Clock Module must be includ­ed in all frames. The Agilent E4831A Clock & Data Generator Module is unsuitable for systems with expander frames.
Product Description Model Number Quantity
Mainframes Small mainframe – 3 free slots. Agilent E4840A
Mainframe – 10 free slots Agilent E4849B VXI extender module (E1482B, 1 slot) Agilent E4849B Required if 1 or 2 expander frames Opt 002 are needed Expander frame Agilent E4848B
Software Agilent 81200 User Software Agilent E4873A
5 licenses of Agilent 81200 User Software Agilent E4873A
Opt 005
Figure 4: Recommended Standard VXI Interfaces
Open VXI Interfaces
VXI Mainframe with E1406A GPIB Slot 0 Command Module
Any PC running MS Windows NT 4.0®
VXI Mainframe with Agilent E6235A VXI-PC
VXI Mainframe with Agilent E8491A IEEE-1394 PC-link to VXI
VXI Mainframe with National Instruments VXIpc- 850
STEP 5: Choosing the user inter­face, support options, installation options, and test accessories
User Interface
The compact choice is the display and entry panel that mounts onto the front of the mainframe and hinges for optimum viewing/connector access. It can only be used when connected to the frames Agilent E4840A and Agilent E4849B in the “proprietary system.”
Alternatively, a VGA display, PC key­board and mouse provide a larger, higher resolution display and stan­dard-sized keys.
Support, documentation and rackmount options
Please consult the table below. We recommend that a CD-ROM drive is ordered so that user software upgrades can be installed.
Figure 5: User interfaces
Product Description Model Number Quantity
Agilent E4840A/49B Agilent E4840A/49B
User interface options 5" x 4" display and entry panel Opt 001
Interface options 15" VGA monitor Opt 003
17" VGA monitor Opt 004 21" VGA monitor Opt 005 Keyboard, US/English Opt 006 Mouse (two button) Opt 007 CD-ROM SCSI drive – includes SCSI-2 cable and termination Opt 008
Documentation options Additional user manual set Opt OB1
Japanese localization Opt ABJ CD-ROM Service Guide (part number E4849-91022) Opt OBW
Rackmount options Rack flange kit (part number E8400-80923)
(Only for the Agilent E4849B and Agilent E4848B) Opt AX4
Support options One year on-site warranty Opt W01
Five year return repair service Opt W50 Software update notification service (12 per year) Opt +NAO Commercial calibration with test report Agilent E4805A/31A/41A Opt UK6 Five year return calibration with service Agilent E4805A/31A/41A Opt W52
5
The following user interfaces are available as Agilent E4840/49B mainframe options:
Option 001 display and entry panel Includes 5"x 4" flat VGA display, miniature keyboard and touch pad
Option 003/004/005: 15"/17"/21" Ultra VGA displays
Option 006 keyboard (U.S./ English) and option 007 mouse
6
Controlling instruments with the Agilent 81200 (only for "proprietary systems")
The embedded PC of the frames Agilent E4849B and Agilent E4840A includes a GPIB interface so that other GPIB instruments can be con­trolled with the Agilent 81200 plat­form (figure 6). To do this, you need to install suitable controller software such as C or VEE. The GPIB interface also allows the Agilent 81200 to be controlled from another computer, such as an ATE system controller.
Communicating over networks (only for "proprietary systems")
The Agilent 81200's PC also includes a LAN interface. The Agilent 81200 can be controlled from a network computer and for convenient vector transfer, servers may also be accessed (figure 6). The data format used for transferring vectors is a STIL (Standard Test Interface Language) subset (ASCII file with header and footer). Vector files can also be trans­ferred via 3.5" disks or the optional external CD-ROM drive.
Additional test software for automated characterization of your design
If you want to automate your charac­terization measurements and create shmoo plots, jitter measurements and eye diagrams, you can simply install the Agilent E4974A Characterization Software Components. For details, also including the system require­ments, please consult Agilent E4974A
Characterization Software Components, data sheet, p/n 5968-
4259E.
Product Description Model Number Quantity
Agilent 81200 Agilent 81200 Test Fixture Agilent E4839A dedicated Accessories
Pogo cable kit: 4*SMA(m) & 2 Pogo Agilent E4839A
adapter for the Agilent E4839A Opt. 001
Universal DUT Test Board Agilent E4839A
Opt. 002
Adapter Kit: 4*SMA(m) I/O Adapter Agilent 15440A
Test Software Characterisation Software Components
General Accessories Cable Kit: 4*SMA(m) to SMA(m) Agilent 15442A
Cable Kit: 10*SMA(m) to SCI Connector Agilent 15441A SMA coax. cable, 1 m. Agilent 8120-4948 Torque wrench, SMA. Agilent 8710-1582 Adapter Kit: 4* SMA(m) I/O Adapter Agilent 15440A Adapter SMA (m)/BNC (f). Agilent 1250-1200 Adapter right-angle SMA (m-f). Agilent 1250-1249 Adapter right-angle SMA (m-m). Agilent 1250-1397 Adapter tee SMA. Agilent 1250-1698 Pulse adder/splitter, SMA. Agilent 11667B 500 ps transition converter. Agilent 15433B 1 ns transition converter. Agilent 15434B 2 ns transition converter. Agilent 15438B Cable, GPIB. Agilent 10833B VEE 5.0 on CD-ROM. Agilent E2120F BASIC on 3.5" disks. Agilent E2060B
Figure 6: The Agilent 81200 can be integrated into all common system test environments
The supplied GPIB interface let's you use the Agilent 81200's built-in PC as a test system controller
The Agilent 81200 can be integrated into a GPIB test system
The Agilent 81200 can be integrated into LAN's
Agilent 81200 (E4840A or E4849B mainframe) with Windows-NT -based C, HP-Basic or VEE
Other GPIB instruments
GPIB
Any controller or PC with C, C++, HP-BASIC or VEE
LAN
Other GPIB instruments
Agilent 81200 (E4840A or E4849B mainframes)
Server/Hard disk (STIL subset)
Agilent 81200 (E4840A or E4849B mainframes)
VXI system
VXI system
7
DUT Fixturing
For convenient and reliable DUT fix­turing for DUT’s with up to 192 pins, there is the Agilent E4839A Test Fixture available. For details, please consult Agilent E4839A Test Fixture, data sheet, p/n 5968-3580E.
Simulation data link for VHDL and Verilog
BestLink/81200, a simulation data link is available to process and trans­fer simulation data from Verilog and VHDL simulators. For details, please consult the BestLink/81200
Simulation Data Link for the Agilent 81200, product information, p/n
5968-2548E or visit:
www.diagonal.com
Test accessories
Please consult the table on page 6.
STEP 6: Checking your configuration
Front-ends and modules:
What do I need to run generator channels up to 1.32 Gbit/s?
You need two Agilent E4843A or Agilent E4838A generator front-ends, which are EXOR-ed internally added to generate one 1.32 Gbit/s signal. For better signal performance, for signals above 1 Gbit/s, please also consider an external addition, by using an Agilent 11667B (APC-3.5 power split­ter, DC to 26.5 GHz) or an Agilent 11636B (DC to 26.5 GHz power divider, APC-3.5).
For operation in EXOR addition mode you only need two Agilent E4843A respectively Agilent E4838A front­ends per module.
I need generator channels that operate up to 200 Mbit/s and with more than 512 Kbit memory depth per channel. Can I use the Agilent E4846A dual generator front-ends?
No. If you need more than 512 Kbit, you should use the Agilent E4843A/38A (max. 660 Mbit/s) front­ends. The Agilent E4846A dual gener­ator front-ends share the channel's memory depth of 1024 Kbit, this is why they only support 512 Kbit. This is also true for dual analyzer front­ends.
For more details please consult
Agilent 81200 Data Generator/ Analyzer Platform, data sheet, p/n
5965-3415E.
I want to run a PRBS pattern on one chan­nel and data from another channel at the same time. Is that possible?
Yes, but you have to use two mod­ules, one for the PRBS and one for the control signals.
I want to use just the Agilent 81200 analyzer channels. Is this configuration possible?
Yes, but we recommend that one gen­erator channel as a timing reference for the system is used, otherwise the trigger output would have to be used as a timing reference.
Mainframe/system control and accessories:
I'd like to fit a VXI DVM into a spare slot in the Agilent 81200 frame. Will that work?
Yes, if you have configured the Agilent 81200 platform as described in the section STEP 4b as an "open" system.
I test my designs in a remote environment so I won't need monitors and so on con­nected to the Agilent 81200. Will it boot without a monitor and keyboard?
Yes, but you will need a mouse. However, you will need a monitor to perform tasks such as shut down and to install user software upgrades.
Can I fit Agilent 81200 modules into an existing VXI system?
Yes, please refer to STEP 4b. Plug and play drivers for the Agilent 81200 are an integral part of the Agilent E4873A Rev. 2.1 User Software.
Can I combine rack & stack instruments with the Agilent 81200?
Yes. You can use an external con­troller or you can use the Agilent 81200's built-in PC as a controller. A GPIB (IEEE 488.2) interface is already installed on the Agilent 81200 system controller. You will however, need to install controller software such as VEE, Agilent HP-BASIC or C.
I want to integrate the Agilent 81200 into a system. How can I control it and transfer test vectors?
These requirements can be fulfilled via LAN or GPIB. Interfaces for both are installed on the Agilent 81200 Controller. The vector file format is a STIL subset (Standard Test Interface Language; ASCII file with header and footer).
8
Upgrading an existing Agilent 81200 data generator/analyzer platform
The Agilent 81200 Data Generator/ Analyzer Platform can be extended or adapted as needed. Front-ends, mod­ules and expander frames can be added at any time. Orders for individ­ual units are supplied with the instal­lation instructions.
User software upgrades
Revisions and enhancements to the user software will be available from time-to-time on CD-ROM, for which an external CD drive is needed such as the Agilent E4840A/49B Opt 008. The operating system should not be upgraded because the user software is specified for operation on Windows NT Rev 4.0 US-English localization.
Example configurations (for a "proprietary system")
Here are some example configura­tions, providing an overall impression of the scope of the Agilent 81200 Data Generator/Analyzer Platform.
Clock generators, pulse/pattern generators
Requirements
A single or multi-channel generator with:
• independent phase and duty cycle adjustment
• variable slopes
• frequency f and f/n simultaneously
• up to 660 MHz
Example configuration
• Agilent E4840A three-slot mainframe
• Option 001 display and entry panel
• Agilent E4831A clock/data generator
• Two Agilent E4838A front-ends
• SMA cables (Agilent 15442A)
Extensions
If more than two channels are needed, add one or two Agilent E4841A mod­ules with up to four front-ends each.
For data rates below 200 Mbit/s (NRZ) the Agilent E4846A dual-out­put front-end can be used instead of the Agilent E4838A. This can econo­mize on the number of front-ends and modules required.
For data rates up to 660 MHz, the Agilent E4843A front-ends should be used.
Data generators
Requirements
Similar to those of the pulse/pattern generators, additional requirements are:
• deep, segmentable, loopable memory for generating many vectors
• bus, control and clock signals from a single source
Example for 20 channels
• Agilent E4849B mainframe
• Option 001 display and entry
panel
• Agilent E4805A clock module
• 5 Agilent E4841A modules
• 20 Agilent E4838A front-ends
For data rates below 200 Mb/s, the Agilent E4846A dual-output front-end can be used instead of the Agilent E4838A. This way, control channels can be accommodated economically.
As an alternative to the display and entry panel, a full-size monitor, key­board and mouse can be ordered. See figure 4 for details.
Data generator/analyzer
Example for 16 channels
• Agilent E4849B mainframe
• Option 001 display and entry panel
• Agilent E4805A clock module
• 4 Agilent E4841A modules
• 8 Agilent E4843A front-ends
• 8 Agilent E4844A front-ends
With the above configuration, eight data generator channels and eight analyzer channels are provided. Other mixes are feasible, as the Agilent E4841A can accept any mix of front­ends.
Note however, that all front-ends within a module have the same sequence. Thus, when connecting to I/O ports (where analysis must take place at a different time from data generation), the generator front-ends should use different Agilent E4841A modules to the analyzer front-ends. For more than eight channels, the Agilent E4849B mainframe should be ordered, instead of the Agilent E4840A. For I/O applications where there is not enough signal to drive 25(that is the effective resistance of a generator channel connected to an Agilent E4844A or Agilent E4845A analyzer front-end), the Agilent E4847A analyzer front-end with selec­table high-impedance should be used.
Product Description Model Number Quantity
Software Agilent 81200 user software Agilent E4873A
upgrade to latest revision. Opt 001
9
Systems needing more than ten module slots
The Agilent E4849B mainframe can be extended using the Agilent E4848B expander frame. The maximum con­figuration with one expander is:
• Agilent E4849B mainframe
• Option 001 display and entry panel (ormonitor/keyboard/mouse options)
• Option 002 VXI extender module
• Agilent E4805A clock module
• 8 Agilent E4841A modules
• 32 front-ends, any mix
• Agilent E4848B expander frame
• Agilent E4805A clock module
• 11 Agilent E4841A modules
• 44 front-ends, any mix
The maximum number of front-ends for this configuration is 76.
All required connecting cables are supplied with the above products. SMA cables have to be ordered sepa­rately.
Systems needing more than 19 module slots
A second expander frame can be con­nected. The maximum configuration with two expanders is 120 channels.
If in doubt, please contact your local Agilent Technologies representative.
Steps to configure your own per­sonal system:
1) Select the configuration sheets to fill out
• For configuration of the Agilent E4840A Small mainframe, use the configuration sheet on page III
• Configuring the Agilent E4849B mainframe as a standalone system (without the Agilent E4848B Expander frames) use the sheet on page V
• Configuring a system with more than one frame, use the configura­tion sheets on pages VII, IX and XI
2) Select the desired slot for your module
On the 1st row of each sheet you can mark with a cross, what kind of mod­ule you want to have plugged into which slot. For this task, make use of the table on page 3 where you have already entered your choice of mod­ules and front-ends.
3) Select the desired slot for your front-end:
Each module holds up to four front­ends. Make use of the table on page 2 again and write the product number (e.g. E4847A) the kind of front-end you would like to have in the appro­priate slot. Please note:
• Module Agilent E4805A doesn’t hold any front-ends
• Module Agilent E4831A holds only two front-ends (only generator front-ends) in the 3rd and 4th front-end slot
• Module Agilent E4841A holds four front-ends
4) Check the power requirements
Checking for power requirements for each frame with the power require­ment calculation table (MS Excel) is useful, to be on the safe-side. The power requirement spreadsheet can be downloaded from:
www.agilent.com/find/81200_configinfo
5) Fill out the Fax Cover Sheet on the next page, and fax it together with the complet­ed configuration sheets to the Fax number on the cover sheet.
Please make sure that you specify your Agilent order number on the fax cover sheet, otherwise your personal configuration requirement will not be processed.
I
Fax Cover Sheet
Agilent 81200 Configuration Guide
To: OFC/BVS Agilent Technologies GmbH,
Herrenberger Straße 130, 71034 Böblingen, Germany
Fax Number: +49 (7031) 464-6532
If you have problems with this fax, please contact: +49 (7031) 464-7674
From:
Name:
Company:
Telephone:
Email:
Sales representative:
Agilent order no. (mandatory):
Comments:
III
Configuration Sheet: E4840A Small Frame
Please choose one of the front­ends: E4837A E4838A E4843A E4844A E4845A E4846A E4847A and write the product number (e.g. E4837A) in the appropriate slot
Front-End 4 Front-End 3 Front-End 2 Front-End 1
0 1 2
E4805A E4831A
E4841A E4831A
E4841A E4831A
Slot
V
Configuration Sheet:
Mainframe E4849B as standalone frame
E4841A
E4805A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
10 11 12
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4805A
E4831A
Please choose one of the front-ends:
E4837A, E4838A, E4843A, E4844A, E4845A, E4846A, E4847 and write the product number (e.g. E4837A)
in the appropriate slot. Use the table on page 3.
81200
Controller
Embedded
Front-End 1 Front-End 2 Front-End 3 Front-End 4
23 4 5 6 7 8 9
1 0
Slot
VII
E4841A
Configuration Sheet:
Mainframe E4849B as Masterframe(if more than one frame is required)
E4805A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
10 11 12
E4841A
E4805A
E4831A
E4805A
Opt 002
Please choose one of the front-ends:
E4837A, E4838A, E4843A, E4844A, E4845A, E4846A, E4847 and write the product number (e.g. E4837A)
in the appropriate slot. Use the table on page 3.
81200
Controller
Embedded
Front-End 1 Front-End 2 Front-End 3 Front-End 4
23 4 5 6 7 8 9
1 0
Slot
IX
E4841A
Configuration Sheet:
E4848B Expander Frame No 1
E4831A
E4841A
E4805A
E4841A
E4805A
E4841A
E4805A
E4841A
E4805A
E4841A
E4805A
E4841A
E4805A
E4841A
E4805A
E4831A
E4831A
E4831A
E4831A
E4831A
E4831A
E4831A
10 11 12
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
Opt 002
Please choose one of the front-ends:
E4837A, E4838A, E4843A, E4844A, E4845A, E4846A, E4847 and write the product number (e.g. E4837A)
in the appropriate slot. Use the table on page 3.
234 56789
1
0
Front-End 1 Front-End 2 Front-End 3 Front-End 4
Slot
XI
1
234 56789
0
10 11 12
Front-End 1 Front-End 2 Front-End 3 Front-End 4
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
E4841A
E4831A
Please choose one of the front-ends:
E4837A, E4838A, E4843A, E4844A, E4845A, E4846A, E4847 and write the product number (e.g. E4837A)
in the appropriate slot. Use the table on page 3.
Slot
Opt 002
E4805A
E4841A
E4841A
E4805A
E4831A
E4841A
E4805A
E4831A
Configuration Sheet:
E4848B Expander Frame No 2
Related Literature
Agilent 81200 Data Generator/Analyzer Platform, brochure
Agilent 81200 Data Generator/Analyzer Platform, data sheet specifications
Agilent E4874A Characterization Software Components, data sheet
Agilent E4839A Test Fixture, data sheet
BestLink/81200 Simulation Data Link for the Agilent 81200 Data Generator/Analyzer Platform, product information
Data transfer between Design, Simulation and the Agilent 81200, product note
Flat Panel Display Link Test, product note
For more information, please visit us at:
www.agilent.com/find/dvt
Pub. Number
5968-4261E
5965-3415E
5968-4259E
5968-3580E
5968-2548E
5967-6276E
5968-8028E
Agilent Technologies Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measure­ment capabilities you paid for and obtain the support you need. Our extensive sup­port resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the produc­tion life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
“Our Promise” means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product informa­tion, including realistic performance spec­ifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measure­ment assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are avail­able.
Your Advantage
“Your Advantage” means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve prob­lems efficiently and gain a competitive edge by contracting with us for calibration, extra- cost upgrades, out-of-warranty repairs, and on-site education and train­ing, as well as design, system integration, project management, and other professional serv­ices. Experienced Agilent engineers and technicians worldwide can help you maxi­mize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
Get assistance with all your test and measurement needs at:
www.agilent.com/find/assist
Or check your local phone book for the Agilent office near you.
Product specifications and descriptions in this document subject to change without notice.
Copyright © 1999, 2000 Agilent Technologies Printed in U.S.A. 5/00 5965-3417E
Loading...