1998 Agilent Technologies 1998, 2000. All rights reserved.
No part of this manual may be reproduced in any form or by any means
(including electronic storage and retrieval or translation into a foreign
language) without prior agreement and written consent from Agilent
Technologies Inc. as governed by United States and international
copyright laws.
Notice
The material contained in this document is subject to change without
notice. Agilent Technologies makes no warranty of any kind with regard
to this material, including, but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Agilent
Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
Warranty
This Agilent Technologies product has a warranty against defects in
material and workmanship for a period of three years from date of
shipment. During the warranty period, Agilent Technologies will, at its
option, either repair or replace products that prove to be defective. For
warranty service or repair, this product must be returned to a service
facility designated by Agilent Technologies. The Buyer shall pay Agilent
Technologies round-trip travel expenses. For products returned to
Agilent Technologies for warranty service, the Buyer shall prepay
shipping charges to Agilent Technologies and Agilent Technologies shall
pay shipping charges to return the product to the Buyer. However, the
Buyer shall pay all shipping charges, duties and taxes for products
returned to Agilent Technologies from another country.
4
Notice
Agilent Technologies warrants that its software and firmware designated
by Agilent Technologies for use with an instrument will execute its
programming instructions when properly installed on that instrument.
Agilent Technologies does not warrant that the operation of the
instrument software, or firmware, will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by the Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse, operation
outside of the environmental specifications for the product, or improper
site preparation or maintenance. No other warranty is expressed or
implied. Agilent Technologies specifically disclaims the implied
warranties of merchantability and fitness for a particular purpose.
Exclusive Remedies
The remedies supplied are the Buyer's sole and exclusive remedies.
Agilent Technologies shall not be liable for any direct, indirect, special,
incidental, or consequential damages, whether based on contract, tort or
any other legal theory.
Certification
Agilent Technologies certifies that this product met its published
specifications at the time of shipment. Agilent Technologies further
certifies that its calibration measurements are traceable to the United
States Institute of Standards and Technology, to the extent allowed by
the Institute's calibrating facility, and to the calibration facilities of other
International Standards Organization members.
Services and Support
Any adjustment, maintenance, or repair of this product must be
performed by qualified personnel. Contact your customer engineer
through your local Agilent Technologies Service Center. You can find a
list of local service representatives on the Web at:
http://www.agilent.com/Service/English/index.html
5
Safety Summary
Safety Summary
The following general safety precautions must be observed during all
phases of operation of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates
safety standards of design, manufacture, and intended use of the
instrument. Agilent Technologies Inc. assumes no liability for the
customer's failure to comply with these requirements.
General
This product is a Safety Class 1 instrument (provided with a protective
earth terminal). The protective features of this product may be impaired
if it is used in a manner not specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs
as per IEC 60825-1.
Environmental Conditions
This instrument is intended for indoor use in an installation category II,
pollution degree 2 environment. It is designed to operate at a maximum
relative humidity of 95% and at altitudes of up to 2000 meters. Refer to
the specifications tables for the ac mains voltage requirements and
ambient operating temperature range.
Before Applying Power
Verify that the product is set to match the available line voltage, the
correct fuse is installed, and all safety precautions are taken. Note the
instrument's external markings described under Safety Symbols on
page 8.
6
Safety Summary
Ground the Instrument
To minimize shock hazard, the instrument chassis and cover must be
connected to an electrical protective earth ground. The instrument must
be connected to the ac power mains through a grounded power cable,
with the ground wire firmly connected to an electrical ground (safety
ground) at the power outlet. Any interruption of the protective
(grounding) conductor or disconnection of the protective earth terminal
will cause a potential shock hazard that could result in personal injury.
Fuses
Only fuses with the required rated current, voltage, and specified type
(normal blow, time delay, etc.) should be used. Do not use repaired fuses
or short-circuited fuse holders. To do so could cause a shock or fire
hazard.
Do Not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of flammable gases or
fumes.
Do Not Remove the Instrument Cover
Operating personnel must not remove instrument covers. Component
replacement and internal adjustments must be made only by qualified
service personnel.
Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.
7
Safety Summary
Safety Symbols
Caution (refer to accompanying documents)
Protective earth (ground) terminal
In the manuals:
WARNING
CAUTION
The WARNING sign denotes a hazard. It calls attention to a
procedure, practice, or the like, which, if not correctly performed
or adhered to, could result in personal injury. Do not proceed
beyond a WARNING sign until the indicated conditions are fully
understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating
procedure, or the like, which, if not correctly performed or adhered to,
could result in damage to or destruction of part or all of the product. Do
not proceed beyond a CAUTION sign until the indicated conditions are
fully understood and met.
8
About this Book
About this Book
This quick start guide helps you to quickly get familiar with the features
and the user interface of the Agilent 81101A Pulse Generator.
Chapter 1 Introducing the Agilent 81101A Pulse Generator on page
15 gives a general overview of the Agilent 81101A, its features, the user
interface, and the steps required for operating the instrument.
For getting started Chapter 2 Getting Started on page 29 gives the
setup information for some real-world signal examples, which can easily
be used by varying only some parameter values.
Chapter 3 Using the Agilent 81101A on page 43 provides complete
information on the user interface screens.
Installation and maintenance are described in Appendix A Installation
& Maintenance on page 75.
For specifications and information on operating the instrument by means
of remote control, please refer to the Reference Guide, p/n 81101-91021.
9
About this Book
Conventions Used in this Book
This book uses certain conventions to indicate elements of the
Agilent 81101As user interface. The following table shows some
examples:
SoftkeysPress the MODE/TRG softkey to access the Mode/
Trigger screen.
HardkeysPress the MORE key to switch to the alternative
softkey layout.
Alternate KeysPress SHIFT + 0 (ON/OFF) to switch on the output.
The alternate key labelwhich is selected by
pressing the SHIFT keyis given in parentheses.
Screen QuotesMove the entry focus down to P
turn the knob to select
INTERNAL
ULSE-PERIOD
PLL.
and
Entry FocusThe highlight field, that can be moved with the
cursor keys, to change modes, parameters, or
parameter formats.
The purpose of the introduction chapter is to give a general overview of
the Agilent 81101A.
The main features and use models are described in What you can do
with the Agilent 81101A on page 16.
Operating the instrument via the front panel user interface is described in
The Front Panel on page 18 and Operating the Agilent 81101A on
page 20.
Help is Available on page 27 shortly introduces the Agilent 81101As
on-line help system.
Finally, The Rear Panel on page 28 takes a look at the back of the
Agilent 81101A.
15
Introducing the Agilent 81101A Pulse Generator
What you can do with the Agilent 81101A
What you can do with the
Agilent 81101A
This section introduces the basic features and use models of the
Agilent 81101A Pulse Generator.
Basic Features
The Agilent 81101A is a single-channel pulse generator with variable
transition times. It is capable of generating all standard pulses and bursts
of pulses needed to test current logic technologies (for example, TTL,
CMOS, ECL, PECL, LVDS, GTL) and other digital designs up to 50 MHz.
The instrument features two internal oscillators:
a synchronously triggerable internal oscillator
an accurate, stable internal PLL
For even more accuracy, an external frequency reference can be
connected.
Any timing parameter can be varied without glitch or drop out. This
contributes to more accurate and confident characterizations of the
device under test (DUT).
Complete signal setups can be stored locally (9 internal non-volatile
registers available) or on a memory card.
Benchtop Testing
The Agilent 81101A features a graphic display showing all pulse
parameters at a glance. The cursor keys and the modify knob allow fast
and simple operation.
The user interface is designed to minimize the time invested in getting
familiar with the instrument. After familiarization, the instrument
supports quick setups of signals. This leaves you free to concentrate on
the measurement task and testing of the DUT.
16
Introducing the Agilent 81101A Pulse Generator
What you can do with the Agilent 81101A
Automated Testing
The Agilent 81101A features an GP-IB/SCPI-conform command structure
for all features. Using this programming interface, the instrument can be
easily integrated into all phases of test system development such as
planning rack integration and test program generation.
Programs designed for the Agilent 81101A are compatible with all other
models of the Agilent 81100 family.
NOTE
For a command reference list, please refer to the Reference Guide, part
number 81101-91021.
17
Introducing the Agilent 81101A Pulse Generator
The Front Panel
The Front Panel
When used for benchtop testing, the instrument is mainly operated from
the front panel.
Front Panel
Switch
NOTE
Special Function KeysData Entry Keys
SoftkeysMORE KeyInputs and Outputs
front panel switch
The
When the front panel switch is off, the instrument is in standby mode.
The instrument is disconnected from the AC line power only by
disconnecting the power cord.
The four keys below the display are the
keys). The current function of each softkey is indicated in the
corresponding box on the display.
Pressing the MORE key changes the softkey layout.
The other keys (
and the
operating the instrument (see Operating the Agilent 81101A on
page 20).
special function keys, data entry keys, cursor keys
rotary knob
is used to switch on and off the instrument.
are used to select and modify parameters when
Cursor Keys
softkeys
(software-controlled
Rotary Knob
)
18
Introducing the Agilent 81101A Pulse Generator
The Front Panel
The major inputs and outputs of the instrument are available at the front
panel:
The external input (EXT INPUT) can be used to connect an external
arming source (triggered or gated modes), or to perform pulse
recovery (external width mode). For details, please refer to The
Mode/Trigger Screen on page 44.
The strobe output (STROBE OUT) provides a signal indicating the
duration of a burst.
In burst mode, the STROBE OUT signal marks the start and end of
each burst of pulses generated. The rising edge of the STROBE signal
is synchronized to the start of the first pulse period in a burst, the
falling edge is synchronized to the start of the last pulse period in the
burst. (see The Mode/Trigger Screen on page 44).
You can set the output levels to TTL or ECL (see The Trigger-Level
Screen on page 65)
The trigger signal (TRIGGER OUT) marks the start of each pulse
period (see The Mode/Trigger Screen on page 44). You can set the
output levels to TTL or ECL (see The Trigger-Level Screen on
page 65).
NOTE
NOTE
The OUTPUT provides the signal output, the indicator shows the
current state of the output (on or off).
An input connector for external frequency reference (CLOCK/REF INP.)
is available at the rear panel (see The Rear Panel on page 28). When
the rear panel option UN2 is ordered, all input and output connectors are
available at the rear panel.
For information on the use of the memory card slot, refer to The
Memory Card Screen on page 67.
19
Introducing the Agilent 81101A Pulse Generator
Operating the Agilent 81101A
Operating the Agilent 81101A
This section guides you through the first steps when operating the
Agilent 81101A via the user interface.
NOTE
For information on operating the Agilent 81101A via remote control,
please refer to the Reference Guide, part number 81101-91021.
Switching On the Instrument
After switching on the instrument the display indicates that the
instrument selftest is running. This can take several seconds to complete.
If the selftest fails, you see a flashing E at the bottom of the screen. Press
the HELP key to see a list of the selftest error messages. Use the knob or
the cursor keys to scroll through the list if necessary.
To return to normal operation press HELP again. Note that the selftest
error messages are removed from the error queue after this.
20
Introducing the Agilent 81101A Pulse Generator
Operating the Agilent 81101A
The Basic Screens
The major parameters for pulse generation can be set up in only two
screens.
Mode/Trigger screen
The
and trigger modes with respect to the signal required.
allows you to set the fundamental operating
Press the MODE/
TRG softkey to
access this screen.
CONTINUOUS
PULSES
MODIFY
Single-Pulses at Out1
*PulseStrm
Burst
Pulse-Period: internal Osc
OUTPUTTRG-LEVLIMITSMODE/TRG
In this screen, you can set up the signal to be triggered or continuous,
and to be a pulse stream or a burst (several pulses followed by a pause)
consisting of single or double pulses.
In the lower section you can specify the frequency source for the pulse
period and the trigger source.
21
Introducing the Agilent 81101A Pulse Generator
Operating the Agilent 81101A
Output screen
The
for the signal to be generated.
allows you to specify timing and level parameters
Press the OUTPUT
softkey to access
this screen.
Per 1.000
Delay
Width
LeadE
TraiE
µµµµ
s Normal
0ps
100.0ns
5.00ns
=LeadE
Offset
Amplit
ΩΩΩΩ
50
into 50.0
OFF
1
+0.0mV
1.00V
ΩΩΩΩ
MODIFY
*OFF
ON
OUTPUTTRG-LEVLIMITSMODE/TRG
In the left part of this screen, you can set up the timing for the signal
(period, pulse width, transition times, ...). The right part is for the levels
(normal or inverted signal, offset and amplitude, impedance, ...).
Furthermore, you can switch on and off the output.
Pressing SHIFT + MORE (GRAPH) when the Output screen is displayed,
toggles between the textual display and a graphical representation of the
parameters. Depending on the current cursor position (left or right part),
the timing parameters or the level parameters are visualized.
Per 1.000µµµµs Normal
Delay1 0ps
Width
LeadE
TraiE
100.0ns
5.00ns
=LeadE
Offset
Amplit
50
OUTPUTTRG-LEVLIMITSMODE/TRG
OFF
+0.0mV
1.00V
ΩΩΩΩ
into 50.0
MODIFY
*OFF
ON
ΩΩΩΩ
1
OFF
PERIOD
1
OFF
+11V
1.000
µµµµ
s
OUTPUTTRG-LEVLIMITSMODE/TRG
-11V
OUTPUTTRG-LEVLIMITSMODE/TRG
Output 1
*OFF
ON
22
Introducing the Agilent 81101A Pulse Generator
Operating the Agilent 81101A
Adjusting Parameters
Adjusting parameters within a screen, requires two steps:
selecting the parameter
adjusting its value
Some parameters allow different formats of their values. For example,
the pulse width can be displayed and entered as an absolute value, as
duty cycle (percentage of the period), or as the delay of the trailing edge.
The following sections show the standard procedure for adjusting
parameters, and list some features useful for the advanced user.
Standard Procedure
To experience the standard procedure for adjusting parameters, consider
the following example where the duty cycle is set to 50%.
1
Press the OUTPUT softkey to access the Output screen.
2
Use the cursor keys to select the W
parameter.
IDTH
Selection CursorModify/Enter Area
Per 1.000
Delay
Width
LeadE
TraiE
µµµµ
s Normal OFF
0ps
100.0ns
5.00ns
=LeadE
Offset
Amplit
50
ΩΩΩΩ
into 50.0
1
+0.0mV
1.00V
ΩΩΩΩ
MODIFY
*Width
DutyCycle
TrailDel
OUTPUTTRG-LEVLIMITSMODE/TRG
The available formats for the pulse width parameter are displayed in
the Modify/Enter area.
23
Introducing the Agilent 81101A Pulse Generator
Operating the Agilent 81101A
3
When changing the
parameter format,
the instrument
automatically
recalculates the
value.
Turn the knob to select D
The selected setting is indicated by an *.
4
Move the cursor to the right to select the duty cycle value.
5
Use the data entry keys or the knob to enter the required value: 50.
6
Press the ENTER key to confirm your selection.
UTYCYCLE
.
Per 1.000
Delay1 0.0ps
DtyCyc
LeadE
TraiE
µµµµ
s Normal OFF
50.00%
6.00ns
=LeadE
Offset
Amplit
50
ΩΩΩΩ
into 50.0
+0.0mV
1.00V
MODIFY
50.00
ΩΩΩΩ
%
OUTPUTTRG-LEVLIMITSMODE/TRG
If you need to specify a unit for a parameter, simply press the appropriate
unit key (
With this small example, you have learned the basic steps required for
adjusting parameters. For a complete reference of parameters and
formats available in the individual screens, refer to Chapter 3 Using the
Agilent 81101A on page 43.
NANO, MICRO/MEGA, MILLI/KILO
) instead of the ENTER key.
24
SHIFT
Introducing the Agilent 81101A Pulse Generator
Operating the Agilent 81101A
Advanced Procedures
The following features can be used to make operation more comfortable.
Selecting parameters
Most keys of the front panel have an additional function. The SHIFT
key provides fast access to the additional functions of the data entry
keys and the special function keys.
For example, it is possible to quickly access the pulse width
parameter by pressing SHIFT + 6 (WIDTH). The entry focus will be
positioned on the corresponding editing field.
Changing the step size
SHIFT
SHIFT
NOTE
When modifying a value in the Modify/Enter area, pressing the SHIFT
key followed by cursor left/right moves the cursor to a different digit
in the displayed number.
Thus, you can change the step size for parameter change before you
vary the value either with the up/down cursor keys or the knob.
Overprogramming
Pressing the SHIFT key while turning the knob, it is possible to exceed
specified parameter ranges to utilize the instrument to its limits.
Proper operation of the instrument outside of the specified ranges is not
guaranteed. It is recommended to have the output switched on when
overprogramming to have the internal error check system activated. This
error check system warns you about impossible settings.
25
Introducing the Agilent 81101A Pulse Generator
Operating the Agilent 81101A
Switching the Output On and Off
When you switch the instrument on, the output is switched off to protect
the device under test. The LED indicator next to the output connector
indicates the output state.
ON/OFF
0
MAN
RECALL
STORE
SHIFT
LOCAL
To switch the output on or off
either press SHIFT + 0 (ON/OFF),
or move the cursor to the ON/OFF parameter in the Output screen
and select the appropriate value by turning the knob.
Using the Special Function Keys
The instrument provides the following special function keys:
The MAN key can be used in triggered or gated mode to manually
trigger the instrument if there is no other source available (see The
Mode/Trigger Screen on page 44).
The STORE key can be used to store/recall to/from 1 to 9 individual
settings in the instrument memory.
In the internal memory location 0 there is a default setting stored.
Pressing SHIFT + STORE (RECALL) and selecting 0 resets the instrument to
the default setting.
The SHIFT key provides fast access to additional functions.
When the front panel controls are locked in remote control, pressing
the SHIFT key unlocks the front panel controls.
HELP
The HELP key provides access to the instruments on-line help or in
warning or error state, access to Warning/Error Report screen.
Pressing SHIFT + HELP (AUTOSET) sets the instrument to a valid
setting based on the current period setting.
26
Help
Introducing the Agilent 81101A Pulse Generator
Help is Available
Help is Available
Whenever you are in doubt or the instrument signals warnings or errors,
press the HELP key.
If there are no warnings or errors pending, pressing the HELP key
displays information on the currently selected parameter, the parameter
help. More information is available within the help system:
Parameter Help
The help information gives a short description of the parameter or
setting options and the corresponding SCPI command(s) syntax for
programming the parameter or setting.
If there is more than one screen available (indicated by small arrows),
use the knob or the cursor keys to scroll through the help
information.
To access parameter help from other screens of the help system,
press the ON FIELD softkey.
Warnings and
Errors
Exit Help
Concept Help
Pressing the CONCEPT softkey within the help system displays a short
description of the instrument.
Serial Numbers and Software Revision
Pressing the SERIAL # softkey within the help system displays
information on serial numbers and software revision codes of the
instrument.
If there are warnings or errors pending (indicated by a flushing W or E),
pressing the HELP key displays a list of the current messages. Using the
ERROR QU and WARNING softkeys, you can toggle between both lists. For
more information on warnings and errors, see Warnings and Errors on
page 72.
To exit from the help system, press the HELP key again, or press the EXIT
HELP softkey.
27
Introducing the Agilent 81101A Pulse Generator
The Rear Panel
The Rear Panel
The rear panel always provides two connectors:
The input connector for external frequency reference (CLOCK/REF
INP.). This input can be used if a higher frequency accuracy is
required, or if you need frequency locking.
The GP-IB connector providing the interface for remote control.
The following figure shows the rear panel view with the option UN2.
GP-IB ConnectorExternal Reference
Inputs and Outputs
With option UN2 the major inputs and outputs of the instrument (as
described in The Front Panel on page 18) are available at the rear
panel:
external input (EXT INPUT)
strobe output (STROBE OUT)
trigger signal (TRIGGER OUT)
signal output (OUTPUT)
28
Fuse HolderAC Line Socket
Serial Number
2
2Getting Started
The intention of this chapter is to give the necessary steps to set up
generic signals for first-time users of the Agilent 81101A.
This chapter provides examples for the following types of signals:
“Setting Up a Clock Signal” on page 30
“Setting Up a Pulse Signal” on page 35
“Setting Up a Burst Signal” on page 39
At the end of each example, the required set of device commands is listed
to provide programming examples. For further information on the
commands and a complete command reference please refer to the
Reference Guide, p/n 81101-91021.
The examples are intended to be performed one after the other.
Therefore, the first example provides the most detailed instructions,
while the other examples are described on a higher level.
29
Getting Started
Setting Up a Clock Signal
Setting Up a Clock Signal
Ta sk
Instructions
Set up a continuous clock signal with 25 MHz frequency with PLL
accuracy, a duty cycle of 50 %, 6 ns transition times, a high level of 2.5 V
and low level of 0 V.
Duty Cycle = 50 %
2.5 V
0.0 V
f = 25 MHz
To set the
1
Reset all parameters and modes by pressing SHIFT + STORE (RECALL) + 0.
2
Press the MODE/TRG softkey to enter the Mode/Trigger screen.
Trigger mode CONTINUOUS and operating mode PULSES are
selected by default.
3
Move the entry focus down to P
select
operating mode and trigger mode
ULSE-PERIOD
INTERNAL
PLL.
as required:
and turn the knob to
CONTINUOUS PULSES
Single-Pulses at Out1
Pulse-Period: internal PLL
OUTPUTTRG-LEVLIMITSMODE/TRG
30
MODIFY
intOsc
intPLL
*CLK-IN
Getting Started
Setting Up a Clock Signal
To set the
1
Press the OUTPUT softkey to enter the Output screen.
2
Press SHIFT + 0 (ON/OFF).
This turns on the Output and activates the internal error check to
detect parameter conflicts.
3
Move the entry focus to P
4
Move the entry focus to the right (to the frequency entry field) and
enter a value of 25 MHz by pressing 2 + 5 +
A warning is displayed, indicating a potential conflict in your settings.
You can ignore this, because the conflict will be solved in the next
steps.
5
Move the entry focus down to W
6
Move the entry focus to the right and enter 50 % by typing 5 + 0 +
ENTER.
7
Move the entry focus down to the leading edge entry field and press 6
+
The trailing edge is set to be the same as the leading edge (T
L
Freq 25.00MHz Normal ON
timing parameters
.
NANO
E) by default.
EAD
as required:
and turn the knob to select F
ER
MICRO/MEGA
and select D
IDTH
UTYCYC
REQ
.
le.
RAI
MODIFY
.
E =
Delay1 0.0ps
DtyCyc
LeadE
TraiE
50.00%
6.00ns
=LeadE
Offset
Amplit
ΩΩΩΩ
50
into 50.0
+0.0mV
1.00V
ΩΩΩΩ
OUTPUTTRG-LEVLIMITSMODE/TRG
8
Press SHIFT + MORE (GRAPH) to see a preview of your signal.
6.00
ns
31
Getting Started
Setting Up a Clock Signal
1
ON
LeadEdge1
6.00
ns
OUTPUTTRG-LEVLIMITSMODE/TRG
You can modify the parameter displayed in the Modify/Enter area, and
immediately check how your changes affect the signal. Use the cursor
keys to switch to another parameter.
9
Press SHIFT + MORE (GRAPH) to return to the textual screen.
To set the
1
Move the entry focus to O
select H
2
Move the entry focus to the high level entry field, enter 2.5 and press
the ENTER key.
In the same way, enter 0 for the low level. The impedance is 50Ω
source into 50Ω load by default.
level parameters
IGH-LOW
.
as required:
FFSET/AMPLITUDE
and turn the knob to
NOTE
Freq 25.00MHz Normal ON
Delay1 0.0ps
DtyCyc
LeadE
TraiE
50.00%
6.00ns
=LeadE
High
Low
ΩΩΩΩ
50
into 50.0
+2.50V
+0.00V
ΩΩΩΩ
MODIFY
+2.50
V
OUTPUTTRG-LEVLIMITSMODE/TRG
There is also a graphical screen available for the level parameters.
32
Getting Started
Setting Up a Clock Signal
Here is the signal as displayed on an Agilent 54810A Infinium
oscilloscope. Use the generator’s TRIGGER OUT to trigger the scope.
33
Getting Started
Setting Up a Clock Signal
Programming
Example
If you want to include this clock signal in your GP-IB program, use the
following command lines. The comment lines starting with a # are not
required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlayOFF
# Internal PLL has to be set as period source.
:ARM:SOURceINT2
# Set the frequency to 25 MHz, the duty cycle to 50 % and the
# leading and trailing edge to 6 ns.
:FREQuency25 MHZ
:PULSe:DCYCle50
:PULSe:TRANsition6NS
# Set the high level to 2.5 Volts, the low level to 0.0 Volts.
:VOLTage:HIGH2.5V
:VOLTage:LOW0V
# Enable the output.
:OUTPutON
34
Setting Up a Pulse Signal
Getting Started
Setting Up a Pulse Signal
Ta sk
Instructions
Set up a continuous pulse signal with 50 ns period, a pulse width of 30 ns,
a leading edge of 6 ns, a trailing edge of 10 ns, an amplitude of 3.3 V and
an offset of 1.65 V (high level 3.3 V, low level 0.0 V).
Pulse Width = 30 ns
3.3 V
50 %
0.0 V
Period = 50 ns
To set the
1
Reset all parameters and modes by pressing SHIFT + STORE + 0.
2
Press the MODE/TRG softkey to enter the Mode/Trigger screen.
The default settings can be used for this example. For reference the
Mode/Trigger screen is displayed. There is nothing to change.
CONTINUOUS
operating mode and trigger mode
PULSES
as required:
MODIFY
Single-Pulses at Out1
*PulseStrm
Burst
Pulse-Period: internal Osc
OUTPUTTRG-LEVLIMITSMODE/TRG
35
Getting Started
Setting Up a Pulse Signal
To set the
1
Press the OUTPUT softkey to enter the Output screen.
2
Switch on the output.
3
Enter a period of 50
A warning is shown, which you may ignore, because the parameter
conflict will be solved in the next steps.
4
Enter a pulse width of 30
5
Enter 6
6
For the trailing edge, select A
Per 50.00ns Normal ON
Delay1 0.0ps
Width
LeadE
TraiE
timing parameters
.
NS
for the leading edge.
NS
Offset
30.00ns
6.00ns
Amplit
50
10.0ns
as required:
.
NS
BSOLUTE
+0.0mV
1.00V
ΩΩΩΩ
into 50.0
and enter a value of 10 NS.
MODIFY
10.0
ΩΩΩΩ
ns
OUTPUTTRG-LEVLIMITSMODE/TRG
To set the
1
Enter an offset of 1.65 V.
2
Enter an amplitude of 3.30 V.
level parameters
as required:
Per 50.00ns Normal ON
Delay1 0.0ps
Width
LeadE
TraiE
30.00ns
6.00ns
10.0ns
Offset
Amplit
50
ΩΩΩΩ
into 50.0
OUTPUTTRG-LEVLIMITSMODE/TRG
36
+1.65V
3.30V
ΩΩΩΩ
MODIFY
3.30
V
Getting Started
Setting Up a Pulse Signal
Here is the figure as displayed on the Agilent 54810A Infinium
oscilloscope. Use the generator’s TRIGGER OUT to trigger the scope.
37
Getting Started
Setting Up a Pulse Signal
Programming
Example
If you want to include this pulse signal in your GP-IB program, use the
following command lines. The comment lines starting with a # are not
required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlayOFF
# Pulse stream operating mode is required, but as we start
# from a default status it is not necessary to send a command for
# setting the instrument into pulse stream operating mode.
# Set the period to 50 ns, the pulse width to 20 ns, the leading
# edge to 6 ns and the trailing edge to 10 ns.
:PULSe:PERiod50NS
:PULSe:WIDTh30NS
:PULSe:TRANsition6NS
:PULSe:TRANsition:TRAiling:AUTOOFF
:PULSe:TRANsition:TRAiling10NS
# Set the amplitude to 3.3 Volts, the offset to 1.65 Volts.
:VOLTage3.3V
:VOLTage:OFFSet1.65V
# Enable the output
:OUTPutON
38
Setting Up a Burst Signal
Getting Started
Setting Up a Burst Signal
Ta sk
Instructions
Set up a burst signal with a burst repetition of 2 µs. Each burst consists of
two double-pulses at a period of 500 ns. The pulse width is 100 ns, the
delay between the two pulses of a double-pulse is 200 ns. The levels are
2 Vpp amplitude and 0 V offset.
Double Pulse Delay = 200 ns
1 V
0 V
1 V
Pulse Period = 500 ns
Burst Repetition = 2 µs
To set the
1
Reset all parameters and modes by pressing SHIFT + STORE + 0.
2
Press the MODE/TRG softkey to enter the Mode/Trigger screen.
operating mode and trigger mode
as required:
The internal PLL is used to trigger the startable internal oscillator.
The PLL defines the burst repetition, the oscillator generates the
pulse period. This requires the following settings:
3
Move to CONTINUOUS and turn the knob to select T
4
Move to PULSES and select B
5
Move down and select D
6
Move left and select a burst length of 2.
7
Move down and select T
(burst repetition)
.
OUBLE
RG’D BY
.
URST
(D
OUBLE-PULSES AT OUT
: PLL and set the period to 2.000µs
RIGGERED
1).
.
39
Getting Started
Setting Up a Burst Signal
TRIGGERED BURSTS OF
2-Double-Pulses at Out1
Pulse-Period: internal Osc
Trg'd by: PLL ->
Per 2.000
µµµµ
s
OUTPUTTRG-LEVLIMITSMODE/TRG
To set the
1
Press the OUTPUT softkey to enter the Output menu.
2
Switch on the Output.
3
Enter a pulse period of 500
4
Enter a double-pulse delay (D
To set the
1
Enter an amplitude of 2.00 V.
timing parameters
level parameters
as required:
.
NS
BLDEL
as required:
) of 200NS.
Per 500.0ns Normal ON
MODIFY
2.000
s
µµµµ
MODIFY
DblDel1 200.0ns
Width
LeadE
TraiE
40
100.0ns
5.00ns
=LeadE
OUTPUTTRG-LEVLIMITSMODE/TRG
Offset
Amplit
50
ΩΩΩΩ
into 50.0
+0.0mV
2.00V
ΩΩΩΩ
2.00
V
Getting Started
Setting Up a Burst Signal
The following figure shows the signals as displayed on the
Agilent 54810A Infinium Oscilloscope if you connect OUTPUT to
channel 1, STROBE OUT to channel 2, and use the generator’s
TRIGGER OUT to trigger the scope.
41
Getting Started
Setting Up a Burst Signal
Programming
example
If you want to include this burst signal in your GP-IB program use the
following command lines. The comment lines starting with a # are not
required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay OFF
# Set the instrument to burst mode by selecting a burst count
# of 2. Choose double pulses. Select triggered mode by selecting
# PLL as the trigger source and set the burst repetition to 2 µs.
:TRIGger:COUNt2# Set Burst mode with burst count of 2
:ARM:SOURceINT2# Set Triggered mode with PLL as
# trigger source
:ARM:PERiod2US# Set burst repetition of 2 µs
:PULSe:DOUBleON# second channel generates double pulses
# per period
# Set the double pulse delay to 200 ns and the width to 100 ns.
:PULSe:PERiod500NS
:PULSe:WIDTh100NS
:PULSe:DOUBle:DELay200NS
# Set the output amplitude to 2 Volts and the offset to
# 0.0 Volts.
:VOLTage2V
# Enable the output.
:OUTPutON
42
3
3Using the Agilent 81101A
This chapter provides complete reference information for using the
Agilent 81101A by means of the user interface screens.
Each screen is described in detail. To access the individual screens, use
the softkeys below the screen. There are two softkey layouts:
the default layout
OUTPUTTRG-LEVLIMITSMODE/TRG
the alternative layout displayed after pressing the MORE key
TRG-LEVCONFIGMEMCARDMODE/TRG
At the end of this chapter, Warnings and Errors on page 72 provides
details on the instruments warning and error messaging system.
NOTE
NOTE
For general information on using the Agilent 81101A and information on
the instruments hardkeys, please refer to Chapter 1 Introducing the
Agilent 81101A Pulse Generator on page 15.
For information on using the Agilent 81101A via remote control, please
refer to the Reference Guide, part number 81101-91021.
43
Using the Agilent 81101A
The Mode/Trigger Screen
The Mode/Trigger Screen
This section describes the Mode/Trigger screen, starting with an
overview of the available parameter combinations, followed by detailed
descriptions of each combination.
Overview
To access the Mode/Trigger screen, press the MODE/TRG softkey.
The following figure shows a typical Mode/Trigger screen, where the
individual parameters are indicated. The parameter combinations are
listed in the table on the next page.
➀
TRIGGERED BURSTS OF
2-Double-Pulses at Out1
➃
➄
Pulse-Period: internal Osc
Trg'd by: PLL ->
In this screen, you can set up the signal to be triggered or continuous,
and to be a pulse stream or a burst (several pulses followed by a pause)
consisting of single or double pulses.
44
➁
Per 2.000
OUTPUTTRG-LEVLIMITSMODE/TRG
➅
➂
MODIFY
2.000
µµµµ
s
µµµµ
s
Using the Agilent 81101A
The Mode/Trigger Screen
Furthermore, you can specify the pulse and trigger sources.
The following table provides a list of possible parameter combinations.
The rows refer to the numbers in the figure.
Trigger ModeContinuousTriggeredGatedExt. Width
➀
Pulse ModePulsesBurstPulsesBurstPulsesBurst
➁
Pulse Type
➂
Length
➃
Period Source
➄
Arming Source
➅
Trigger Out
Strobe Out
Single/DoubleSingle/DoubleSingle/Double
265536265536265536
internal Osc.
internal PLL
CLK-IN
not used
on 1st
↑
on last
↓
MAN
key
Ext. Input
marks each pulse period generated
not used
internal Osc.
internal PLL
CLK-IN
MAN
key
Ext. Input
PLL
on1st
↑
on last
↓
internal Osc.
internal PLL
CLK-IN
MAN
key
Ext. Input
not used
↑
↓
on 1st
on last
MAN
key
Ext. Input
not used
The pulse period source can be selected from internal oscillator, internal
PLL, and external CLK-IN:
If you have to trigger after an external event, use the startable
oscillator (
internal Osc.
), which starts on the external signal with a
fixed latency.
If you do not have to trigger but need a high accuracy frequency, then
use the
For even more accuracy, you can use the external CLK-IN.
internal PLL
clock generation circuitry.
In triggered/gated mode the internal PLL can be used as an arming
source. However, the internal PLL cannot be used as pulse and arming
source at the same time.
The following sections explain the individual mode combinations in
more detail. For each combination of trigger mode and pulse mode,
typical timing diagrams are provided, showing single and double pulses.
45
Using the Agilent 81101A
The Mode/Trigger Screen
Continuous Pulses Mode
The following figure shows typical timings for trigger mode
CONTINUOUS and pulse mode PULSES.
Period Source
int Osc, int PLL or Ext Clock
TRIGGER OUT
OUTPUT
Single Pulse
OUTPUT
Double Pulse
Period
Characteristics
NOTE
Pulse periods are generated continuously.
You can select single pulses or double pulses per pulse period.
For the pulse period source you can select from
internal Osc (VFO)
internal PLL (higher accuracy)
external CLK-IN (external signal), synchronize to rising or falling
edge
TRIGGER OUT marks each pulse period.
STROBE OUT is not used in continuous pulse mode.
46
Using the Agilent 81101A
The Mode/Trigger Screen
Continuous Burst Mode
The following figure shows typical timings for trigger mode
CONTINUOUS and pulse mode BURST.
Burst of 3
Period
Period Source
int Osc, int PLL or Ext Clock
OUTPUT
Single Pulse
OUTPUT
Double Pulse
TRIGGER OUT
Characteristics
STROBE OUT
1st period
A burst of pulse periods is repeated continuously.
You can select the number of pulse periods per burst in the range of
last period
265536.
You can select single pulses or double pulses per pulse period.
For the pulse period source you can select from
internal Osc (VFO)
internal PLL (higher accuracy)
external CLK-IN, synchronize to rising or falling edge.
TRIGGER OUT marks each pulse period.
STROBE OUT rises at the start of the first pulse period in a burst, and
falls at the start of the last pulse period.
47
Using the Agilent 81101A
The Mode/Trigger Screen
Triggered Pulses Mode
The following figure shows typical timings for trigger mode TRIGGERED
and pulse mode PULSES. The pulses are triggered by the rising edge of
the arming source.
Arming Source
Trg'd by: Ext Input or MAN Key
TRIGGER OUT
OUTPUT
Single Pulses
OUTPUT
Double Pulses
Trg'd by:
Characteristics
NOTE
STROBE OUT
Single pulse periods are triggered by (T
RG'D BY
) an active edge at the
selected arming source:
MAN key on front panel, triggered by press or release or both.
EXT INPUT triggered by rising or falling or both edges.
You can select single pulses or double pulses per pulse period.
TRIGGER OUT marks each pulse period.
STROBE OUT is not used in triggered pulse mode.
The PLL cannot be selected as the arming source. Select continuous
pulse mode with the PLL as period source to achieve the same result.
48
Using the Agilent 81101A
The Mode/Trigger Screen
In contrast to the previous figure, this figure shows a timing diagram
where the pulses are triggered by both rising and falling edges of the
arming source.
Trg'd by:
Arming Source
(Trg'd by:)
Ext INPUT or MAN Key
OUTPUT
Single Pulses
OUTPUT
Double Pulses
TRIGGER OUT
STROBE OUT
Minimum Width
49
Using the Agilent 81101A
The Mode/Trigger Screen
Triggered Burst Mode
The following figures show typical timings for trigger mode TRIGGERED
and pulse mode BURST. The bursts are triggered by the rising edge of the
arming source.
For the first example, the synchronously triggerable internal oscillator is
used to source the period.
Characteristics
Trg'd by:
ARMING SOURCE
(Trg'd by:) EXT INPUT
or MAN Key
or PLL
PERIOD SOURCE
internal Osc.
OUTPUT
Single Pulses
OUTPUT
Double Pulses
TRIGGER OUT
STROBE OUT
A burst of pulse periods is triggered by an active edge at the selected
Period
1st period
Burst of 3
last period
Oscillator starts synchronously
1st period
arming source
MAN key on front panel, triggered by press or release or both.
EXT INPUT triggered by rising or falling or both edges.
PLL (internally triggered bursts)
When using the internal PLL as the arming source, you must specify
the triggering period.
You can select the number of pulse periods per burst in the range of
265536.
You can select single pulses or double pulses per pulse period.
50
Using the Agilent 81101A
The Mode/Trigger Screen
TRIGGER OUT marks each pulse period.
STROBE OUT rises at the start of the first pulse period in a burst and
falls at the start of the last pulse period.
NOTE
For the second example, either the internal PLL or an external CLK-IN
or MAN Key
OUTPUT
Single Pulses
OUTPUT
Double Pulses
Trg'd by:
Period
1st period
PLL/CLK-IN NOT synchronous to EXT INPUT
DELAY = n * Period, 1 < n ≤ 2
Burst of 3
last period
1st period
are used to source the periodboth cannot be triggered synchronously.
ARMING SOURCE
(Trg'd by:) EXT INPUT
PERIOD SOURCE
internal PLL or CLK-IN.
TRIGGER OUT
STROBE OUT
You cannot use the PLL as both pulse period source and arming
(T
RG'DBY
) source at the same time.
51
Using the Agilent 81101A
The Mode/Trigger Screen
Gated Pulses Mode
The following figures show typical timings for trigger mode GATED and
pulse mode PULSES.
For the first example, the synchronously triggerable internal oscillator is
used to source the period.
Characteristics
Gated by:
ARMING SOURCE
(Gated by:) EXT INPUT
or MAN Key
Period
PERIOD SOURCE
internal Osc
OUTPUT
Single Pulses
OUTPUT
Double Pulses
TRIGGER OUT
STROBE OUT
Pulse periods are gated by (enabled by) an active level at the selected
Oscillator starts synchronously
arming source:
MAN key on the front panel, gated while pressed or released or
both.
EXT INPUT (external signal) gated by high, low or both levels.
You can select single pulses or double pulses per pulse period.
TRIGGER OUT marks each pulse period.
STROBE OUT is not used in gated pulses mode.
52
Using the Agilent 81101A
The Mode/Trigger Screen
For the second example, either the internal PLL or an external CLK-IN
are used to source the periodboth cannot be triggered synchronously.
ARMING SOURCE
(Gated by:) EXT INPUT
or MAN Key
PERIOD SOURCE
internal PLL or CLK-IN
OUTPUT
Single Pulses
OUTPUT
Double Pulses
TRIGGER OUT
STROBE OUT
Gated by:
Period
PLL/CLK-IN NOT Synchronous to EXT INPUT
DELAY = n * Period, 1 < n ≤ 2
53
Using the Agilent 81101A
The Mode/Trigger Screen
Gated Burst Mode
The following figures show typical timings for trigger mode GATED and
pulse mode BURST.
For the first example, the synchronously triggerable internal oscillator is
used to source the period.
Characteristics
Gated by:
ARMING SOURCE
(Gated by:) EXT INPUT
or MAN Key
PERIOD SOURCE
internal Osc.
OUTPUT
Single Pulses
OUTPUT
Double Pulses
TRIGGER OUT
STROBE OUT
1st period
Bursts of pulse periods are gated by (enabled by) an active level at
Period
Oscillator starts synchronously
Burst of 3
last period
1st period
last period
the selected arming source:
MAN key on the front panel, gated while pressed or released or
both.
EXT INPUT (external signal) gated by high, low or both levels.
You can select single pulses or double pulses per pulse period.
TRIGGER OUT marks each pulse period.
STROBE OUT rises at the start of the first pulse period in a burst and
falls at the start of the last pulse period.
54
Using the Agilent 81101A
The Mode/Trigger Screen
For the second example, either the internal PLL or an external CLK-IN
are used to source the periodboth cannot be triggered synchronously.
ARMING SOURCE
(Gated by:) EXT INPUT
or MAN Key
PERIOD SOURCE
internal PLL. or CLK-IN
OUTPUT
Single Pulses
OUTPUT
Double Pulses
TRIGGER OUT
STROBE OUT
Gated by:
Period
1st period
PLL/CLK-IN NOT Synchronous to EXT INPUT
DELAY = n * Period, 1 < n
Burst of 3
last period
2
≤
1st period
last period
External Width Mode
In external width mode, the pulse width is determined by an external
signal. You can choose between:
MANK
: Pressing the key generates a leading edge, releasing the key
EY
generates a trailing edge.
EXT-IN: A rising edge at the EXT INPUT generates a leading edge, a
falling edge generates a trailing edge.
The threshold and impedance of the EXT INPUT can be set on the
Trigger Level screen.
The period, delay, and width of the output pulse are not programmable in
this mode, because they are determined by the external signal.
The TRIGGER OUT rises and falls parallel to the OUTPUT signal.
55
Using the Agilent 81101A
The Output Screen
The Output Screen
The Output screen shows both timing and level parameters on one
screen.
To access the Output screen, press the OUTPUT softkey.
Timing Parameters
You can use the left part of the Output screen to view and control the
pulse timing parameters.
NOTE
1.000
Per
Delay
Width
LeadE
TraiE
0ps
100.0ns
5.00ns
=LeadE
µµµµ
s Normal OFF
Offset
Amplit
50
ΩΩΩΩ
+0.0mV
1.00V
into 50.0
1
*Period
Frequency
ΩΩΩΩ
MODIFY
OUTPUTTRG-LEVLIMITSMODE/TRG
The individual timing parameters are described in more detail in the
following. All parameters can be entered in different formats.
With the Agilent 81101A you can vary timing parameters without the risk
of glitches or drop outs, which could cause measurement errors. This
applies to continuous mode with timing values < 100 ms (frequency
< 10 Hz), and consecutive values between one-half and twice the
previous value.
56
Using the Agilent 81101A
The Output Screen
When you press SHIFT + MORE (GRAPH) while one of the timing parameters is
selected, you will see a graphical representation of the timing
parameters. The currently selected parameter is displayed in the Modify/
Enter area and is indicated by dashed or bold lines in the graphical
display.
1
OFF
NOTE
Width1
100.0
ns
OUTPUTTRG-LEVLIMITSMODE/TRG
1
OFF
PERIOD
1.000
µµµµ
s
OUTPUTTRG-LEVLIMITSMODE/TRG
Use the left/right cursor keys to toggle between the individual timing
parameters.
Note that in graphics mode you can only adjust the values of each
parameter, not the parameter format. If you want to change the format of
a parameter, for example W
select the parameter name with the cursor.
Pulse Period Parameter
to DTYCYC, you must be in text mode to
IDTH
Set the pulse period as either P
If you select the CLK-IN as the pulse period source on the Mode/Trigger
screen, the pulse period/frequency is determined from the signal applied
to the CLOCK/REF INP. connector:
M
EAS ONCE
The external signal is measured only once. Press ENTER to measure
again.
ERIOD
or F
REQUENCY
.
57
Using the Agilent 81101A
The Output Screen
Pulse Delay Parameter
Delay the leading edge of the pulse within the pulse period. There are
three delay formats available:
M
EAS CONT
The external signal is continuously measured until the instrument
receives a command via GP-IB.
To invoke continuous measurements again, you have to bring the
instrument in local operating mode by pressing SHIFT (LOCAL) and start
continuous measurement again.
D
or DBLDEL (select A
ELAY
D
is the absolute delay from the start of a pulse period to the
ELAY
BSOLUTE
)
start of the leading edge of the pulse.
If double pulses are selected on the Mode/Trigger screen, the D
BLDEL
parameter specifies the delay between the first and second pulse in
each pulse period.
The absolute delay is independent of the pulse period. So the leading
edge does not move relative to the start of the period if you change
the period.
D
% (select % OF P
ELAY
ERIOD
)
The delay from the start of the pulse period to the start of the leading
edge expressed as a percentage of the pulse period. In this format, if
you change the period, the leading edge moves relative to the start of
the period in order to maintain the percentage delay.
P
HASE
(select P
HASE
)
The phase delay in degrees from the start of the pulse period to the
start of the leading edge (360° = 1 pulse period). In this format, if you
change the period, the leading edge moves relative to the start of the
period in order to maintain the phase delay.
58
Using the Agilent 81101A
The Output Screen
Pulse Width Parameter
Set the width of the output pulse. There are three width formats
available:
W
(select W
IDTH
The absolute pulse width measured from the start of the leading edge
to the start of the trailing edge. In this format, the pulse width is
independent of changes in pulse period and delay.
IDTH
)
NOTE
D
TYCYC
(select D
UTYCYCLE
)
The duty cycle is the pulse width measured from the start of the
leading edge to the start of the trailing edge expressed as a
percentage of the period. In this format, if you adjust the period, the
absolute width is adjusted to maintain the duty cycle.
T
RADEL
(select T
RAILDEL
)
The trailing delay is the absolute delay from the start of the pulseperiod to the start of the trailing edge. In this format the trailing edge
remains fixed relative to the start of the pulse period if you adjust the
pulse delay (leading edge delay) or the pulse period.
You cannot have the width format set to D
edge format set to percentage of width (L
EADED
and the leading/trailing-
TYCYC
%/T
E%) at the same
RAIL
time.
Leading Edge Parameter
Set the leading edge transition time of the pulse, measured from 10% to
90% of the pulse amplitude. There are two formats available:
L
E (select A
EAD
BSOLUTE
The absolute transition time measured from 10% to 90% of the pulse
amplitude. In this format the leading edge is independent of the pulse
width.
)
L
E% (select % OF WID)
EAD
The absolute transition time expressed as a percentage of the pulse
width. In this format, if you adjust the pulse width, the transition time
is adjusted to maintain the edge time as a percentage of the width.
59
Using the Agilent 81101A
The Output Screen
NOTE
NOTE
The leading and trailing edges are independently programmable within
certain ranges only, see Transition Timesin the Reference Guide.
You cannot have the width format set to D
edge format set to percentage of width (L
EAD
and the leading/trailing-
TYCYC
E%/T
E%) at the same
RAI
time.
Pulse Trailing Edge Parameter
Set the trailing edge transition-time of the pulse, measured from 10% to
90% of the pulse amplitude. There are three formats available:
=L
E (select =L
EAD
The trailing edge transition time is coupled directly to the leading
edge to maintain a symmetrical pulse.
T
E (select A
RAI
BSOLUTE
The absolute transition time measured from 90% to 10% of the pulse
amplitude. In this format the trailing edge is independent of the
leading edge and the pulse width.
T
E% (select % OF WID)
RAI
The trailing edge transition time expressed as a percentage of the
pulse width. In this format, if you adjust the pulse width, the
transition time is adjusted to maintain the edge time as a percentage
of the width.
EAD
E)
)
NOTE
NOTE
The leading and trailing edges are independently programmable within
certain ranges only, see Transition Timesin the Reference Guide.
You cannot have the width format set to D
edge format set to percentage of width (L
EAD
and the leading/trailing
TYCYC
E%/T
E%) at the same
RAI
time.
60
Using the Agilent 81101A
The Output Screen
Level Parameters
You can use the right part of the Output screen to view and control the
pulse level parameters and to enable or disable the outputs.
Per 1.000
Delay1 0ps
Width
LeadE
TraiE
The individual level parameters are described in more detail in the
following. All parameters can be entered in different formats.
When you press SHIFT + MORE (GRAPH) while one of the level parameters is
selected, you will see a graphical representation of the level parameters.
The currently selected parameter is displayed in the Modify/Enter area
and is indicated in the graphical display.
1
OFF
µµµµ
s Normal
100.0ns
5.00ns
=LeadE
OUTPUTTRG-LEVLIMITSMODE/TRG
+11V
Offset
Amplit
ΩΩΩΩ
50
into 50.0
OFF
+0.0mV
1.00V
ΩΩΩΩ
MODIFY
*OFF
ON
Amplitude1
1.00
-11V
V
OUTPUTTRG-LEVLIMITSMODE/TRG
When output limits are on, the limits are indicated on the Output screen
in graphics mode (see The Limits Screen on page 64).
61
Using the Agilent 81101A
The Output Screen
NOTE
NOTE
NOTE
Note that in graphics mode you can only adjust the values of each
parameter, not the parameter format. If you want to change the format of
a parameter, for example O
FFS-AMPL
to H
IGH-LOW
, you must be in text
mode to select the parameter name with the cursor.
When the output is switched on, the instrument monitors the actual
voltage and current levels at the output. The output is automatically
switched off if voltage levels or power dissipation reach levels which
could damage the output circuits.
Normal/Complement Parameter
Switch the output between normal and complement mode:
N
ORMAL
Pulse leading edge rises from low to high level, trailing edge falls from
high to low level.
C
OMPLMNT
Pulse leading edge falls from high to low level, trailing edge rises from
low to high level.
This parameter is only available in text mode.
Offset/Amplitude, High/Low Level Parameters
Set and display the pulse levels in terms of either offset and amplitude, or
high and low level. You can quickly set TTL or ECL output levels using
the S
TTL and SET ECL formats.
ET
S
TTL
ET
Select high and low level formats and automatically set the levels to
the default TTL levels:
TTL-HI:+2.50 V
TTL-LOW: +0.0 mV
The default levels are set once and can be adjusted afterwards by
moving the entry focus to the values as normal.
H
IGH-LOW
Select high and low level format for the pulse levels.
62
Using the Agilent 81101A
The Output Screen
O
FFS-AMPL
Select offset and amplitude format for the pulse levels. Offset is
measured from 0V to the middle of the pulse amplitude. Pulse
amplitude is the difference between the high and low levels of the
pulse.
S
ECL
ET
Select high and low level format and automatically set the levels to
the default ECL levels:
ECL-HI:850 mV
ECL-LOW:1.80 V
These default levels are set once and can be adjusted afterwards by
moving the entry focus to the value as normal.
Voltage/Current Mode (mV/V/mA/A)
Move the entry focus onto the level units to select between setting the
pulse levels
IN VOLTS
or IN A
MPERES
.
NOTE
NOTE
NOTE
This parameter is only available in text mode.
Output Source Impedance Parameter (50Ω into)
Select an output source impedance of 50 Ω or 1 kΩ.
This parameter is only available in text mode
Load Impedance Parameter (50.0Ω)
Adjust the load impedance value expected at the output to compensate
for non-50Ω loads. The displayed level parameters are then calculated
using this value and therefore represent the levels at a non-50Ω static
load.
This parameter is only available in text mode.
63
Using the Agilent 81101A
The Limits Screen
The Limits Screen
Use the Limits screen to set up voltage and current limits for the pulse
level parameters to prevent accidental damage of the device under test.
To access the Limit screen, press the LIMIT softkey.
After you switch on the limits, the pulse level parameters on the OUTPUT
screen cannot be adjusted outside the ranges specified on the LIMITS
screen if the output is switched on. Note that because current and
voltage limits apply, the available ranges of the impedance parameters
are also affected.
When output limits are on, the limits are indicated on the OUTPUT screen
in graphics mode, and the level bar is scaled accordingly:
+1V
1
OFF
MODIFY
*OFF
ON
Amplitude1
500
-1V
mV
64
OUTPUTTRG-LEVLIMITSMODE/TRG
Using the Agilent 81101A
The Trigger-Level Screen
The Trigger-Level Screen
Use the Trigger-Level screen to
set the triggering threshold and input impedance for the EXT INPUT
connector,
set the triggering threshold and input impedance for the
CLOCK/REF INP. connector,
set the output levels for the STROBE OUT and TRIGGER OUT
connectors.
To access the Trigger-Level screen, press the TRG-LEV softkey.
EXT-IN: Threshold
+1.0V 50
CLK-IN: Threshold +1.0V 50
ΩΩΩΩ
ΩΩΩΩ
MODIFY
Set TTL
Set ECL
TRIGGER-OUT: TTL
*Voltage
STROBE-OUT : TTL
OUTPUTTRG-LEVLIMITSMODE/TRG
Threshold and Impedance Parameters
You can individually specify threshold and input impedance of the EXTIN and the CLK-IN connectors.
EXT-IN and CLK-IN Threshold Parameters
Move the entry focus to EXT-IN or CLK-IN to quickly set the triggering
threshold of the external input or the clock input to a TTL- or ECLcompatible level:
65
Using the Agilent 81101A
The Trigger-Level Screen
S
Set the input threshold to +2.5 V.
You can adjust the threshold by moving the entry focus to the value.
S
Set the input threshold to 1.3 V.
You can adjust the threshold by moving the entry focus to the value.
V
Set any threshold level in the range of 10.0 V to +10.0 V. Move the
entry focus to the value to adjust it.
EXT-IN and CLK-IN Impedance Parameters
Toggle the input impedance of the external input connector or the clock
input connector between 50Ω and 10 kΩ.
Trigger and Strobe Level Parameters
TTL
ET
ECL
ET
OLTAGE
You can individually set the output levels of the TRIGGER OUT
connector and the STROBE OUT connector.
Set the output levels (into 50 Ω) to
TTL
High level: +2.50 V
Low level: 0 V
ECL
High level: 0.85 V
Low level: 1.80 V
66
Using the Agilent 81101A
The Memory Card Screen
The Memory Card Screen
Use the Memory Card screen to
store instrument settings to the memory card,
recall instrument settings from the memory card,
delete files from the memory card,
format a memory card.
To access the Memory Card screen, press the MEMCARD softkey. If the
MEMCARD softkey is not displayed, press MORE.
Dir Path \
Filename TESTVI.TXT
MODIFY
*BURSTS
PULSES
Perform Operation
TRG-LEVCONFIGMEMCARDMODE/TRG
Directory Path Parameter
Move the entry focus to DIR P
card, or to view the sub-directories of the current directory (the name of
the current directory is displayed next to D
All sub-directories of the current directory are listed in the Modify/Enter
area.
To change the directory, turn the knob to select the directory name and
press the ENTER key.
Note that when you are in a sub-directory you can return to the parentdirectory by selecting .. from the directory list.
to change the directory on the memory
ATH
IR PATH
).
67
Using the Agilent 81101A
The Memory Card Screen
Filename Parameter
CAUTION
Move the entry focus to the F
ILENAME
parameter to select a file from the
current directory. Use the knob to scroll through the filenames listed in
the Modify/Enter area.
Memory Card Operations
Move the entry focus to P
ERFORM OPERATION
the required operation. Press the ENTER key to perform the operation.
R
EADCARD
Read the DOS file system information from the memory card after
inserting a new card.
R
ECALL
Recall the selected file as the current instrument setting.
S
TORE
Store the current instrument setting to the memory card.
Press ENTER once to start editing the filename for the setting in the
Modify/Enter area. The currently selected filename is used as default.
If you do not modify the filename, the existing file will be overwritten
when you press ENTER.
To modify the filename, move the character cursor with the cursor
keys. Modify a character using the knob. When you have finished,
press ENTER to store the setting.
The filename can be up to 8 characters long.
Press the cursor up or cursor down keys to cancel the store operation
at any time.
Note that the DOS filename suffix .ST0 is added automatically to the
filename when you store the current settings.
and use the knob to select
S
TORE ALL
Store the current instrument setting and the instrument setting
memories 1 to 9 to the memory card. Each setting is stored in a
separate file with the same name but different suffixes (.ST0 for the
current setting, ST1 to ST9 for setting memories 1 to 9).
68
Using the Agilent 81101A
The Memory Card Screen
Press ENTER once to start editing the filename for the setting in the
Modify/Enter area. The currently selected filename is used as default.
CAUTION
CAUTION
If you do not modify the filename, the existing file will be overwritten
when you press ENTER.
To modify the filename, move the character cursor with the cursor
keys. Modify a character using the knob. When you have finished,
press ENTER to store the setting.
The filename can be up to 8 characters long.
Press the cursor up or cursor down keys to cancel the store operation
at any time.
D
ELETE
Delete the selected file from the memory card.
F
ORMAT
Format the memory card.
Formatting a memory card destroys any existing files on the card.
69
Using the Agilent 81101A
The Configuration Screen
The Configuration Screen
Use the Configuration screen to
set the GP-IB address of the Agilent 81101A,
perform a selftest,
select the frequency reference source and the frequency for the PLL.
To access the Configuration screen, press the CONFIG softkey. If the
CONFIG softkey is not displayed, press MORE.
GPIB-Address: 10
Perform Selftest: µP and Signal
MODIFY
Group Params by: TIMING/LEVELS
Pulse-Period: internal PLL
10
PLL-Ref: Internal
Deskew 1: +0ps 2: +0ps
TRG-LEVCONFIGMEMCARDLIMITS
GP-IB Address
Set the GP-IB address for the instrument in the range of 0 to 30. The
default address is 10.
Perform Operation Selftest
Move the entry focus to P
If the selftest fails, a flashing E is displayed. Press HELP to see the list of
error messages.
ERFORM
: S
ELFTEST
and press the ENTER key.
70
Using the Agilent 81101A
The Configuration Screen
PLL Reference
Set the frequency reference source for the PLL:
I
NTERNAL
The internal 5 MHz reference.
CLK-IN
An external reference signal at the CLOCK/REF INP. connector. You
can set the expected frequency of the external reference to 5 MHz or
10 MHz.
71
Using the Agilent 81101A
Warnings and Errors
Warnings and Errors
The Agilent 81101A has two levels of error reporting called warnings
and errors. Checking for errors and warnings is always enabled, unless
you switch it off via the GP-IB using the
:SYSTem:CHEC k
command.
A warning
to a combination of worst case uncertainties at the current settings of
all relevant parameters.
For example, when adjusting the pulse width, all other timing
parameters (leading edge, trailing edge, and pulse period) and their
uncertainties have to be considered in order to check if the width
setting will fits within the pulse period.
Note that the warning limits are therefore not fixed for a particular
parameter, but vary with the settings of the related parameters. It is
also possible that the error and warning limits are the same, that is, a
warning does not occur before the error limit is reached.
If a warning occurs, the settings are still implemented in the hardware
because the worst-case conditions used to evaluate the warning limits
are very unlikely to occur in practice.
A blinking W indicates that one or more warnings have occurred.
Press HELP to view the warning list.
error
An
required parameter settings cannot be implemented in the output
hardware. Multiple errors can occur, but only the first error detected
is displayed.
An error is indicated by a blinking error message at the bottom of the
screen.
is generated when the output signal could be invalid due
is generated when an invalid mode is chosen, or the
NOTE
If you are using the knob to adjust parameters it is normally not possible
to generate warnings or errors. All parameters are automatically limited
to settings that guarantee specified operation.
However, if you press the SHIFT key while turning the knob, you can
adjust a parameter beyond its warning limits.
72
Using the Agilent 81101A
Warnings and Errors
NOTE
NOTE
You can press SHIFT + HELP (AUTOSET) to carry out an autoset. The
instrument resets all parameters, based on the current period setting, to
remove all warning and error conditions.
An Example of Warning and Error Reporting
1
Switch on the instrument and recall the standard settings by pressing
SHIFT + STORE (RECALL) + 0.
The period is now set to 1 µs.
2
Switch on the output by pressing SHIFT + 0 (ON/OFF).
3
On the Output screen, move the parameter cursor to the value of the
W
parameter (100NS).
IDTH
4
Use the knob to make the pulse width as large as possible
(approximately 940
This limit is intended to guarantee that the actual output pulse is
within specifications, for the actual period.
The limit is calculated taking into account a worst case combination
of minimum period from the period setting (1
width from the width setting (940
and trailing edge settings (5.00
Note that if you now try to adjust the leading edge from its current
setting of 5.00 ns with the knob, it cannot be adjusted. This is because the
upper and lower warning limits are currently 5.00 ns. The width is at its
maximum value and width + leading edge + trailing edge <
NS
).
µ
) and maximum
S
) together with minimum leading
NS
).
NS
period.
NOTE
5
Press SHIFT and adjust the pulse width above its warning limit. A
flashing W appears to indicate that a warning condition has occurred.
Note that as long as no errors occur, the output hardware is set up and
attempts to generate the required output.
6
Press HELP to see the warning message:
T
RAILING EDGE
7
Press HELP again to return to the W
1
MAY CUT NEXT PULSE
parameter.
IDTH
73
Using the Agilent 81101A
Warnings and Errors
8
9
10
11
Increase the pulse width further to approximately 980 ns and press
HELP to see the current warnings:
W
1
IDTH
TOO CLOSE TO PERIOD
T
RAILING EDGE
Press HELP again to return to the W
1
MAY CUT NEXT PULSE
parameter.
IDTH
Increase the pulse width further until a flashing error message
appears (approximately 1.10 µs):
OUTPUT 1: W
You have reached the current upper error-limit of the W
IDTH
> P
ERIOD
IDTH
parameter.
The setting is not implemented in the output hardware.
Press SHIFT + HELP to perform an AUTOSET.
74
A
AInstallation & Maintenance
75
Initial Inspection
Initial Inspection
Inspect the shipping container for damage. If the container or cushioning
material is damaged, keep it until the contents of the shipment have been
checked for completeness and the instrument has been verified both
mechanically and electrically.
WARNING
To avoid hazardous electric shock, do not perform electrical tests
when there are signs of shipping damage to any part of the
instruments outer covers or panels.
If the contents are incomplete, or there is mechanical damage, or if the
instrument does not work as expected within its specifications, notify
the nearest Agilent office. (Please request the latest performance tests
from Agilent.) Keep the shipping materials for inspection by the carrier.
The Agilent office will arrange for repair or replacement without
awaiting settlement.
76
Initial Inspection
Standard Deliverables
The Agilent 81101A shipping container contains the following standard
deliverables:
1. The Agilent 81101A Pulse Generator.
2. The English Quick Start Guide and a Reference Guide.
3. A power cord.
1. Agilent 81101A Pulse Generator
2. This Quick Start Guide
and the Reference Guide
3. A Power Cord
77
Initial Inspection
Options and Accessories:
Options
Rear Panel ConnectorsOption UN2
All inputs and outputs are at the rear panel
1 MB SRAM Memory CardOption UFJ
(Agilent part number 0950-3380)
The following figure shows the instruments rear panel with Option UN2
installed.
Handle Accessory
Front Handle KitAgilent part number 5063-9226
Rack Mounting Accessories
Rack Mount KitAgilent part number 5063-9212
Rack Mount and Front
Handle Kit
Rack Slide KitAgilent part number 1494-0059
78
Agilent part number 5063-9219
Quick Start Guide Language Options
French GuideOption ABF
(Agilent part number 81101-91220)
Japanese GuideOption ABJ
(Agilent part number 81101-91520)
Taiwan Chinese GuideOption AB0
(Agilent part number 81101-91620)
Korean GuideOption AB1
(Agilent part number 81101-91720)
Chinese GuideOption AB2
(Agilent part number 81101-91820)
Additional Documentation Options
Service ManualOption 0BW
(Agilent part number 81101-91021)
Initial Inspection
79
Power Requirements
Power Requirements
CAUTION
NOTE
Before applying AC line power to the instrument, ensure that the correct
line fuse is installed in the fuse holder and the correct power cable is
fitted.
When the front panel switch is off, the instrument is in “standby” mode.
The instrument is disconnected from the AC line power only by
disconnecting the power cord. Please make sure that the power cord is
easily identifiable and can quickly be reached by the operator.
The following symbol is close to the ‘Standby’ switch on the front panel
of the instrument.
The instrument can operate from any single-phase AC power source
supplying 100–240 V in the frequency range from 50 to 60 Hz, or 100–
120 V at 400 Hz. The maximum power consumption is 300 VA with all
options installed. When the instrument is switched on the power supply
adapts automatically to the applied AC power (Auto Selection) and
monitors the AC power range during operation.
80
Power Requirements
Replacing the Fuse
1 Remove the power cord.
2
Unscrew the fuse cap at the rear of the instrument beside the powerinlet socket.
Fuse Holder and
Fuse Cap
3
Replace the fuse with the equivalent part:
Agilent
Line VoltageFuse Type
100–240 V~T 10A, 250 V2110-0720
4
Refit the fuse cap.
Part Number
81
Power Cable
Power Cable
In accordance with international safety standards, this instrument is
equipped with a three-wire power cable. The figure below shows the part
numbers of the power cables available. When connected to an
appropriate AC power receptacle, this cable grounds the instrument
cabinet.
E
N
E
L
N
L
E
E
NL
LN
Australia
8120-1369
South Africa
8120-4211
82
Denmark
8120-2956
E
E
LN
NL
Switzerland
Europe
8120-1689
LN
E
Great Britain
8120-1351
LN
E
8120-2104
United States 120V
8120-1378
Japan 120V
8120-4753
Ventilation Requirements
Ventilation Requirements
The instrument is fitted with three cooling fans. Make sure that there is
adequate clearance of 80 mm at the rear and 25 mm at the top and
bottom to ensure adequate air flow. If the air flow is restricted, the
internal operating temperature will be higher, reducing the instrument's
reliability or causing the instrument's thermal-protection circuits to
automatically switch off the instrument.
NOTE
Do not cover the ventilation holes.
Thermal Protection
Overheating Detection
The instrument monitors its internal temperature. If the temperature
exceeds approximately 80°C, the power supply is switched off. The
instrument will switch on again if the temperature falls below
approximately 77°C.
Fan Failure
If a fan is broken or prevented from operating by a blockage, the
temperature will increase. When the temperature exceeds approximately
80°C the overheating detection switches off the instrument for safety
reasons. For reliability it is recommended to send instruments with
broken or defective fans immediately to Agilent Service for repair.
83
Battery
Battery
This instrument contains a lithium battery. Typical life time of the battery
is about 5 years.
NOTE
NOTE
CAUTION
Recycle or dispose used batteries according to local regulations. Or
contact your Agilent representative for information on battery recycling.
The battery is replaceable. Replacement should only be carried out by
qualified service personnel.
There is a danger of explosion if the battery is incorrectly replaced.
The battery must be replaced with the same or equivalent battery type:
Lithium CR2477-N
A
The Agilent replacement part number is:
The battery is protected against charging.
Do not crush, puncture, or incinerate the battery. Do not short the
battery’s external contacts.
type battery.
1420-0557
.
84
Battery Replacement
Battery
NOTE
Disconnect the power cord from AC line voltage to avoid electrical
shock.
1 Remove the bumper which is fitted to the front panel.
2
Remove the strips on the left and right side of the front panel.
3
Remove the front panel which is secured with two screws at each
side.
4
Disconnect the flat cable which connects the front panel to the
chassis.
5
On the right side of the chassis there is a cover secured with one
screw. Remove this cover.
6
Now the battery can be removed from the battery holder. It is
recommended to push the battery on the top, and to pull the battery
on the bottom with one finger for easier removal.
7
Fit the new battery in the holder, take care to not short current the
battery on the chassis.
8
Fit the cover again and secure it with the screw to the chassis.
9
Re-connect the flat cable of the front panel again to the chassis
connector. Make sure that the pin one (red wire) is pointing to the
mid of the chassis.
10
Re-assemble the front cover.
11
Fit the strips to the sides of the front panel, again.
12
Fit the bumper to the front panel.
85
Operating Environment
Operating Environment
Storage Temperature–40 °C to +70 °C
Operating Temperature 0 °C to 55 °C
Humidity95% R.H. (at 40 °C)
AltitudeUp to 2000m
Installation Category II
PollutionDegree 2
WARNING
The instrument is not designed for outdoor use. Do not expose
the instrument to rain or other excessive moisture. Protect the
instrument from humidity and temperature changes which could
cause condensation within the instrument.
Do not operate the instrument in the presence of flammable
gases, fumes or powders. Operation of any electrical instrument
in such an environment constitutes a definite safety hazard.
86
Cleaning Recommendation
Cleaning Recommendation
WARNING
To prevent electrical shock, disconnect the instrument from
mains before cleaning. Use a dry cloth or one slightly dampened
with water to clean external case parts. Do not attempt to clean
internally.
87
Acoustic Noise Emission
Acoustic Noise Emission
Acoustic Noise Emission
For ambient temperatures up to 30°C, under normal operation and at the
typical operator position:
LpA = 52 dBA
Measured in accordance with ISO 7779 / EN 27779.
Geräuschemissionswerte
Bei einer Umgebungstemperatur bis 30°C
LpA = 52 dBA
am Arbeitsplatz, normaler Betrieb.
Angabe ist das Ergebnis einer Typprüfung nach ISO 7779 / EN 27779.