Agilent Technologies 7694 User Manual

Changing the Sample Needle
Agilent 7694 Headspace Sampler
Sample probes and union part numbers are: Nickel needle assembly for G1289 or G1290 301-013-HSP
Silcosteel® needle assembly for G1289 or G1290 301-220-HSP Nickel needle assembly for G1883 232-2790010-EHS Silcosteel needle assembly for G1883 232-2790012-EHS Zero dead volume union for G1290/G1289 325-045-HSP Union elbow for G1883 998-0000053-EHS
The process of changing the sample needle on the 7694A or 7694E Headspace Sampler is as follows.
This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material.
The information contained herein is intended for use by i nform ed i ndividuals who can and must determine its fitness for thei r purpose.
A16176.doc
http:// www.chem.agilent.com
http:// www.chem.agilent.com Page 1
1 of 4444
11
Maintaining the Instrument
Changing the Sampling Needle
Changing the Sampling Needle
1. Remove the two screws on the sides of the sampler cover shown below and lift the cover. Support the cover while working in the sampler area.
Lift front cover
Rear cover
Remove screws
Manifold cover
Figure 16. Removing the sampler cover
2. Loosen the two screws on the manifold cover (it is a stainless steel box) and insulation from the valve compartment.
2
Maintaining the Instrument
Changing the Sampling Needle
Manifold cover and insulation
(Sampler cover
removed for clarity)
Loosen
Figure 17. Removing the manifold cover and insulation
3. The sampling needle is located inside the square aluminum block. It is connected to the sampling valve through a 1/16–inch nickel tube and hexagonal union.
4. Support the hexagonal union with a 7-mm wrench and remove the nut with another 7-mm wrench. Then, loosen and remove the hexagonal union. Lift out the needle.
3
Maintaining the Instrument
Changing the Sampling Needle
Hexagonal union
Sampling needle
G1289/G1290 G1883
Union
Sampling needle
Figure 18. Position of vial sampling needle
5. Replace the sampling needle, tighten the fittings, and perform a leak test before replacing the insulation and cover.