•One G1926A Bar Code Reader (optional, 6890 GC only)–requires INET
integrator or ChemStation
G2613A Injector
Open injector door here
Sample tray quadrant
G2614A Tray
Figure 1The 7683 Automatic Liquid Sampler system
Turret
Compatibility
A 6890 GC without the ALS interface board can use only the 7673 ALS.
A 6890 GC with the ALS interface board can use only the 7683 ALS.
The 6850 GC uses only the 7683 ALS.
2MAR 2000
Page 7
System Overview
Safety
Safety
Before servicing the various components of the GC, observe the following safety
precautions.
Caution
Follow the precautions below when servicing the instrument to avoid data loss
or damage to the instrument.
Printed circuit boards
When storing printed circuit boards (PCBs), always place them in static control
envelopes or enclosures.
Always make sure you are properly grounded (e.g., wearing an ESD strap) before
handling electrostatic sensitive components such as printed circuit boards.
Voltage carrying components
When disconnecting plugs, pull on the plug, not its wires. Pulling on the wires
may cause breakage.
Warnings
The following are potential hazards that can cause personal injury.
Heated components
Parts of the injector may become hot. Turn off the GC oven and inlet/detector
heated zones and allow them sufficient time to cool before servicing those areas.
If you must perform service on components that have not fully cooled, wear
protective gloves.
Voltage carrying components
Whenever possible, disconnect the G2613A Injector and G2614A Tray from the
GC before working on them.
MAR 20003
Page 8
System Overview
Safety
Whenever possible, disconnect the GC from its power source before working on
or near voltage carrying components of the GC.
The following components carry voltage when the GC is plugged in even if the
power switch is off:
•The AC power cord
•The AC power supply
When the power to the GC is turned on, potentially dangerous voltages exist on
these additional components:
•The power transformer
•All electronics boards
•All internal wires and cables connected to these boards
•The injector, tray, and bar code reader cables that connect to the GC
•All internal wires, cables, and boards in the injector, tray, and bar code
reader
The ALS Interface board routes messages between the GC main board and the
components of the automatic liquid sampler system, and transmits power to the
injector(s) and the tray. Errors detected by the injector or the tray are routed to
this board, which then displays the error message on the 6890 GC or at the
ChemStation.
Power from the transformer enters the board at J5. This voltage is partially
rectified through half a diode bridge before being transmitted to the tray and the
injector(s) at approximately 33 VAC. The other half of the diode bridge resides
within the injector and the tray.
The ALS Interface board runs on +5 VDC from the main board through P5.
One fuse is located on this board at F1 to protect the transformer from
overvoltage.
6850 GC
The 6850 GC directly controls and powers the injector and tray. The power
circuitry is identical to that on the ALS interface board.
G2613A Injector
The G2613A Injector completes the rectification of power delivered by the ALS
Interface board (6890 only) or the 6850, and interprets all commands received
into the motions required to make an injection. Sensors and encoders are used
12MAR 2000
Page 17
Diagnostics
Theory of operation
on the syringe carriage, plunger, and turret to locate the position of these parts
and to control injection speed. Any errors detected are reported to the ALS
interface board/GC main board, which displays the error on the GC.
G2614A Tray
The G2614A Tray consists of a platen which holds the sample vial quadrants, a
bracket, a control board, and a robotic arm assembly that manipulates and moves
the sample vials to the injector.
Half-rectified power to the assembly arrives through the cable to the ALS
interface board, along with control signals. The board within the tray completes
rectification of the 33 VAC power input, interprets all commands, and drives the
robotic arm motors and the bar code reader.
Sensors are located within the tray and robotic arm assemblies to allow the tray
board to determine the location of the gripper and to determine whether or not
the gripper is carrying a vial.
MAR 200013
Page 18
Diagnostics
Theory of operation
14MAR 2000
Page 19
5
Replacing the mounting post .............................................16
Replacing the G2614A Tray................................................17
Replacing the G2613A Injector ..........................................19
Replacing the G1926A Bar Code Reader ..........................21
Replacing the ALS Interface board
(6890 Service Manual)...................................................
Making a trial run ................................................................23
Replacement Procedures
Page 20
Chapter 5
Replacement Procedures
Replacing the mounting post
The 6890 GC mounting post locations are shown below.
Back injector
Front post location
Front injector
Back post location
Figure 2Mounting post locations
There is only one post location for the 6850 GC.
16MAR 2000
Page 21
Replacement Procedures
Replacing the G2614A Tray
Replacing the G2614A Tray
1.Unplug the tray cable from the GC.
2.Remove the three injection port cover Torx screws shown in Figure 3 and
remove the old tray.
Tray
Injection port cover
Figure 3Installing the tray
Tray mounting
Captive screw
Remove screws from
cover
screws
(install first)
MAR 200017
Page 22
Replacement Procedures
Replacing the G2614A Tray
CautionDo not move or manipulate the robotic arm or gripper. Moving the arm or gripper
back and forth can cause damage. If necessary, rotate the arm assembly a few
degrees until there is enough clearance.
3.Align the new tray over the three holes in the injection port cover and locate
in place. Tighten the captive screw in the center of the tray mounting arm
to hold the tray in place. Install the two mounting screws into the tray and
tighten until the tray is secure.
4.Install the tray quadrants. Slide the leading edge of each quadrant under the
guides on the tray, then snap the front tab into the slot on the tray. See
Figure 4.
Tray quadrant
Front tab
Guide tab
Slot
Figure 4Install the tray quadrants
18MAR 2000
Page 23
Replacement Procedures
Replacing the G2613A Injector
Replacing the G2613A Injector
NoteDo not mount the G2613A Injector on a mounting post for a 7673 Injector.
Remove the old post and replace it with the new one.
1.Unplug the injector cable from the GC.
2.Lift the injector off of the mounting post.
3.If necessary, install/replace the mounting post on the injection port cover
in the front or back location, as desired.
4.Open the new injector’s door. Slide the packing material and syringe carriage
down, then remove the packing.
5.Mount the new injector. Line up the hole in the base of the injector that is
nearest the cable with the mounting post. Lower the injector about an inch
(2.5 cm) onto the post. See Figure 5.
Alignment pin
Cable
Hole in injector base
Back mounting post
location
Mounting post (front)
Inlet cover
Alignment hole
Figure 5Mounting the injector (6890 GC shown)
MAR 200019
Page 24
Replacement Procedures
Replacing the G2613A Injector
6.Front location: Turn the injector so that the turret is facing the front of the
GC. Lower the injector until the alignment pin in the base enters the
alignment hole in the injection port cover.
Back location: Turn the injector so the turret faces the left side of the GC.
Lower the injector until the alignment pin in the base enters the alignment
hole in the injection port cover.
7.Check your work:
•The injector must be vertical.
•The alignment pin must be properly seated in the alignment hole.
•The injector feet should touch the injection port cover.
If the injector will not sit upright on the GC, check that the plumbing and
cabling under the injection port cover are properly routed in their channels.
20MAR 2000
Page 25
Replacement Procedures
Replacing the G1926A Bar Code Reader
Replacing the G1926A Bar Code Reader
1.Unplug the old bar code reader cable from the tray. See Figure 6.
Connect bar code reader
Figure 6Connecting the bar code reader
2.Remove the bar code reader from the tray.
3.Install the new bar code reader. See Figure 7.
Mounting clips
Bar code reader
Figure 7Mounting location for the bar code reader
MAR 200021
Page 26
Replacement Procedures
Replacing the G1926A Bar Code Reader
Adjusting the bar code reader position
To adjust the bar code reader position:
1.Loosen the bracket mounting screws and the height adjustment screw.
See Figure 8.
Adjust this space
Tray
Mounting screws (2)
GC
Support bracket
Height adjustment screws
Figure 8The bar code reader support bracket
2.Slide the support bracket towards or away from the tray until the top of the
bar code reader is parallel with the tray surface. Tighten the mounting
screws.
3.Adjust the bar code reader’s height. Slightly loosen the height adjustment
screw on the side of the support bracket and raise or lower the bracket until
there is a very small gap—about the thickness of a piece of paper—between
the mounting clips on the bar code reader and the tray. See Figure 8. Tighten
the height adjustment screw.
22MAR 2000
Page 27
Replacement Procedures
Making a trial run
Making a trial run
Use the following trial run to quickly check injector and tray operation.
1.Install an empty syringe in the injector.
2.Place empty bottles in the Solvent A and Waste A turret positions. Place an
empty capped sample vial in the tray 1 position (or the turret sample 1
position, if not using the tray).
3.Set these following parameters on the GC:
ParameterSetting
Injection volume1
# Sample pumps1
Viscosity delay0
# Sample washes1
# Solve A washes1
# Solve B washes0
Slow plungeroff
Pre-dwell time0
Post-dwell time0
Sampling offsetoff (on if cool on-column inlet used)
# Solve A pre-wash1
# Solve B pre-wash0
# Injections/vial1
Samples1–1 (101–101 for turret with ChemStation control)
4.Set the GC oven program to 25°C with a 0°C/min ramp, a hold time of
0.1 minutes, an equilibrium time of 0.3 minutes, and an initial time of
0.3 minutes.
MAR 200023
Page 28
Replacement Procedures
Making a trial run
5.Store, load, then run the sequence.
•If there are no faults, the injector will make one "injection" from the
first vial position.
•If any faults occur, see the 7683 Installation Guide.
24MAR 2000
Page 29
6
Injector, part no. G2613A.......................................................26
Tray, part no. G2614A ............................................................28
Illustrated Parts Breakdown
Page 30
Chapter 6
Illustrated Parts Breakdown
The following pages list the replacement parts for the 7683 Automatic Liquid
Sampler.