The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequent ial damages in connect ion with the furnishing , performance, or
use of this material.
Certification
Agilent Technologie s certi f ies that this produc t m et i ts p ub lis hed speci fi c ations a t the time
of shipment from the factory. Agilent Technologies further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute’s calibration facility, and to the
calibration facilities of other International Standards Organization members.
Regulatory Inf ormation
The regulatory information is located in Chapter 8 , “Safety and Regulatory Information.”
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “ AS IS,” AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT
PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR
IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR
INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD
AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Assistance
Product maintenance agreements and other customer assistance agreements are availa ble
for Agilent Technologies products. For any assistance, contact your nearest Agilent
Technologies sales or service office. See Table 8-1 for the nearest office.
ii
Safety Note s
Front-Panel Key
SOFTKEY
The following safety notes are used throughout this manual. Familiarize your self with
each of the notes and its meaning before operating this instrument. All pertinent safety
notes for using this product are located in Chapter 8 , “Safety and Regulatory
Information.”
WARNINGWarning denotes a hazard. It calls attention to a procedure which, if
not correctly performed or adhered to, could result in injury or loss
of life. Do not proceed beyond a warning note until the indicated
conditions are fully un derstood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to , would result in damage to or destructi on of
the instrument. Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met.
How to Use This Guide
This guide uses the following conventions:
This represents a key physically located on the
instrument.
This represents a “softkey,” a key whose label is
determined by the instrument’s firmware.
Screen TextThis represents text displayed on the instrument’s screen.
iii
Documentation Map
The Installation an d Quick Start Guide pr ovides procedures for
installing, configuring, and verifying the operation of the analyzer. It
also will help you familiarize yourself with the basic operation of the
analyzer.
The User’s Guide shows how to make measurements, explains
commonly-used features, and tells you how to get the most
perform ance from your ana ly z e r.
The Reference Guide provides re fere n ce inform at i on, su ch as
specifications, menu maps, and key definitions.
The Programm er’s Guide provides general GPIB programming
informat ion, a comm an d reference, and e xample pr og ra ms. T he
Programmer’s Gui de contains a CD-ROM with example programs.
The CD-ROM provides the Installa tio n and Qu ick Start Guide , the
User’s Guid e, the Referen ce Guide , and the Progra mmer’s Guid e in
PDF format for v ie wing or p rin t ing from a PC.
The Service Guide pr ovides information on calibrating,
troubleshooting, and servicing your analyzer . The Serv ic e Gu ide i s not
part of a standard shipment and is available only as Option 0BW, or
by ordering part number 08720-90397. A CD-ROM with the Service Guide in PDF format is included for viewing or printing from a PC.
This chapter contains the following example procedures for making measurements. Mixer
and time domain measurements are covered in Chapter 2 , "Making Mixer Measurements
(Option 089 Only) " and Chapter 3 , “Making Time Domain Measurements.” This chapter
also describes how to use most display, marker, and sequencing functions.
• "Making a Basic Measurement" on page 1-4
• "Measuring Magnitude and Insertion Phase Response" on page 1-7
• "Measuring Electrical Length and Phase Distortion" on page 1-43
— Electr ical Length
— Phase Distortion (deviation from linear phase, group delay)
• Characterizing a Duplexer (ES Analyzers Only)
• "Measuring Amplifiers" on page 1-52
— Measuring Gain Compression
— Measuring Gain Compression and Reverse Isolation Simultaneously
(ES Analyzers Only)
— Making High Power Measurements (ES Analyzers Only)
• "Using the Swept List Mode to Test a Device" on page 1-67
• "Using Limit Lines to Test a Device" on page 1-72
• "Using Test Sequencing to Test a Device" on page 1-114
The following chapters describe how to use more instrument functions (as indicated by
their chapter titles):
• Chapter 4 , "Printing, Plotting, and Saving Measurement Results"
• Chapter 5 , "Optimizing Measurement Results"
• Chapter 6 , "Calibrating for Increased Measurement Accuracy"
1-2
Making Measurements
More Instrument Functions Not Described in This Guide
More Instrument Functions Not Described in This Guide
To learn about instrument functions not covered in this user’s guide, refer to the following
chapters in the reference guide.
“Menu Maps”contains maps of the instrument menu structure.
“Hardkey/Softkey Reference”contains descriptions of all instrument functions.
1-3
Making Measurements
PRESET : FACTORY
Making a Basic Measurement
Making a Basic Measurement
There are five basic steps when you are making a measurement.
1. Connect the device under test and any required test equipment.
CAUTIONDamage may result to the device under test (DUT) if it is sensitive to the
analyzer’s default output power lev el. To avoid damaging a sensitive DUT, be
sure to lower the output power before connecting the DUT to the analyzer.
2. Choose the measurement parameters.
3. Perform and apply the appropriate error-correction.
4. Measure the device under test (DUT).
5. Output the measurement results.
This example procedure shows you how to measure the transmission response of a
bandpass filt e r.
Step 1. Connect the device under test and any required test
equipment.
Make the connections as shown in Figure 1-1.
Figure 1-1 Basic Measurement Setup
Step 2. Choose the measurement parameters.
Press .
Preset
To set preset the analyzer to the “Factory Preset” conditions, press the
1-4
softkey if it is not selected. Then press .
Preset
Setting the Frequency Range
POWER RANGE MAN
POWER RANGES
Sweep Setup
NUMBER OF POINTS
Trans: FWD S21 (B/R)
TRANSMISSN
AUTOSCALE
SELECT DISK
INTERNAL MEMORY
RETURN
SAVE STATE
Marker Search
SEARCH: MAX
To se t the center frequency to 134 MHz, press:
Making Measurements
Making a Basic Measurement
Center134M/µ
To set the span to 30 MHz, press:
Span30M/µ
NOTEYou could also press the and keys and enter the frequency
StartStop
range limits as start frequency and stop frequency values.
Setting the Source Power
To change the power level to −5 dBm, press:
Power−5x1
NOTEYou could a lso pres s and select
one of the power ranges to keep the power setting within the defined range.
Setting the Measurement
To change the number of measurement data points to 101, press:
To select the transmission measurement, press:
Meas
or on ET models:
To view the data trace, press:
Scale Ref
Step 3. Perform and apply the appropriate error-correction.
Refer to the Chapter 5 , “Optimizing Measurement Results,” for procedures on
correcting measurement errors.
To save the instrument state and error- correction in the analyzer internal memory,
press:
Save/Recall
Step 4. Measure the device under test.
Replace any standard used for error-correction with the device under test.
To measure the insertion loss of the bandpass filter, press:
1-5
Making Measurements
PRINT MONOCHROME
PLOT
Making a Basic Measurement
Step 5. Output the measurement results.
To create a printed copy of the measurement results, press:
(or )
Copy
Refer to Chapter 4 , “Printing, Plotting, and Saving Measurement Results ,” for
procedures on how to set up a printer and define a print, plot, or save results.
1-6
Making Measurements
Trans:FWD S21 (B/R)
TRANSMISSN
AUTO SCALE
Trans:FWD S21 (B/R)
TRANSMISSN
AUTO SCALE
CALIBRATE MENU
RESPONSE
THRU
Measuring Magnitude and Insertion Phase Response
Measuring Magnitud e and Insertion Phase Response
This measurement example shows you how to measure the maximum amplitude of a
surface acoustic wave (SAW) filter and then how to view the measurement data in the
phase format, which provides information about the phase response.
Measuring the Magnitude Response
1. Connect your test device as shown in Figure 1-2.
Figure 1-2 Device Connections for Measuring a Magnitude Response
2. Press and choose the measurement settings. For this example the
You may also want to select settings for the number of data points, averaging, and IF
bandwidth.
3. Remove the device and connect the power cables together (thru) and perfo rm a response
calibration using the following key presses.
Press .
Chan 1Cal
1-7
Making Measurements
AUTO SCALE
Marker Search
SEARCH: MAX
DUAL | QUAD SETUP
DUAL CHAN ON
PHASE
Measuring Magnitude and Insertion Phase Response
If the channels are coupled (the default condition), this calibration is valid for both
channels.
4. Reconnect your test device.
5. To better view the measurement trace, press:
Scale Ref
6. To locat e the maximum amplitude of the device resp onse, as shown in Figure 1-3 , p r ess :
Figure 1-3 Example Magnitude Response Measurement Results
Measuring Insertion Phase Response
7. To view both the magnitude and phase response of the device, as shown in Figure 1-4,
press:
Chan 2
Display
Format
The channel 2 portion of Figure 1-4 shows the insertion phase response of the device under
test. The ana lyz er measures and displays phas e over the range o f −180° to +180°. As phase
changes beyond these values, a sharp 360° transition occurs in the displayed data.
1-8
Measuring Magnitude and Insertion Phase Response
Figure 1-4 Example Insertion Phase Response Measurement
Making Measurements
The phase response shown in Figure 1-5 is undersampled; that is, there is more than 180°
phase delay between frequency points. If the ∆Φ≥ 180°, incorrect phase and delay
inform at io n may result . Figu re 1-5 shows an example of phase samples being with
∆Φ less than 180° and greater than 180°.
Figure 1 -5 Phase Samples
Undersampling may arise when measuring devices with long electrical length. To correct
this problem, the frequency span should be reduced, or the number of points increased
until ∆Φ is less than 180° per point. Electrical delay may also be used to compensate for
this effect (as shown in the next example procedure).
1-9
Making Measurements
Using Display Functions
Using Display Functions
This section provides the necessary information for using the display functions. These
functions are very helpful for displaying measurement data so that it will be easy to read.
This section covers the following topics:
• Adding titles to your measurements
• Viewing both primary channels at the same time
• Viewing and customizing four-channel measurements
• Using the memory traces
• Using the memory math functions
• Blanking the analyzer’s display
• Changing the colors of the display
1-10
Titling the Active Channel Display
MORE
TITLE
ERASE TITLE
ENTER
SELECT LETTER
DONE
NEWLINE
FORMFEED
Making Measurements
Using Display Functions
1. Press to access the title menu.
Display
2. Press and enter the title you want for your measurement display.
• If you have a DIN keyboard attached to the analyzer, type the titl e you w ant from the
keyboard. Then press to enter the title into the analyzer. You can enter a
title that has a maximum of 50 characters. (For more information on using a
keyboard with the analyzer, refer to the “Options and Accessories” chapter in the
reference guide.)
• If you do not have a DIN keyboard attached to the analyzer, enter the title from the
analyzer front panel.
a. Turn the front panel knob to move the arrow pointer to the first character of the
title.
b. Press .
c. Repeat the previous two steps to enter the re st of the char acter s in yo ur title. You
can enter a title that has a maximum of 50 characters.
d. Press to complete the title entry.
Figure 1-6 Example of a Display Title
CAUTIONThe and keys are not intended for creatin g display
titles. Those keys are for creating commands to send to peripherals during a
sequence program.
1-11
Making Measurements
DUAL | QUAD SETUP
DUAL CH AN on O FF
SPLIT DISP
SPLIT DISP
Using Display Functions
Viewing Both Primary Measurement Channels
In some cases, you may want to view more than one measured parameter at a time.
Simultaneous gain and phase measurements, for example, are useful in evaluating
stability in negative feedback amplifiers. You can easily make such measurements using
the dual channel display.
1. To see channels 1 and 2 in the same grid, press:
Display
, set to ON, and
to 1X.
Figure 1-7 Example of Viewing Channel 1 and 2 Simultaneously
2. To view the measurements on separate graticules, press: Set to 2X. The
analyzer shows channel 1 on the upper half of the display and channel 2 on the lower
half of the display. The analyzer defaults to measuring S
on channel 1 and S21 on
11
channel 2.
1-12
Figure 1-8 Example Dual Channel with Split Display On
SPLIT DISPLAY 1X
COUPLED CH OFF
COUPLED CH ON off
MARKERS: UNCOUPLED
Making Measurements
Using Display Functions
3. To return to a single-graticule display, press: .
NOTEYou can control the stimulus functions of the two channels independent of
each other by pressing .
Sweep Setup
Dual Channel Mode with Decoupled Stimulus
The stimulus functions of the two channels can be controlled independently using
in the stimulus menu. In addition, the markers can be controlled
independently for each channel using in the marker mode
menu, under the key.
Marker Fctn
NOTEES models onl y: For dual channel, if chann els are unc oupled and you have full
2-port calibrations on both channels, you will not be able to select a
non-ratioed measurement. For example, you can measure S
or B/R, but not
21
input B.
NOTEAuxiliary channels 3 and 4 are permanently coupled by stimulus to primary
channels 1 and 2, respectively. Decoupling the primary channels’ stimulus
from each other does not affect the stimulus coupling between the auxiliary
channels and their primary channels.
Dual Channel Mode with Decoupled Channel Power
By decoupling the channel power or port power and using the dual channel mode, you can
simultaneously view two measurements (or two sets of measurements, if both auxiliary
channels are enabled) having different power levels.
1-13
Making Measurements
MEASURE RESTART
AUX CHAN
AUX CHAN
DUAL | QUAD SETUP
DUAL CHAN
AUX CHAN
SPLIT DISP
4X
Using Display Functions
However, there are two configurations that will not sweep continuously.
1. For analyzers with source attenuators, with channel 1 having one attenuation value
and channel 2 set to a different at tenua tion val ue, then continuous sweep is disa bled to
avoid wear on the attenuator. A similar situation where this occurs is when a 2-port cal
is active and the port 1 attenuation value is not equal to the port 2 attenuation value.
Since one attenuator is used for bot h measure ments, this woul d c ause the attenuator to
continuously switch power ranges , so continuous sweep is not all owed. (The exceptio n is
analyzers configured with option 400. Option 400 analyzers have two attenuators and
can allow different attenuation settings on each port.)
2. For ES analyzers with Option 007 or 085 (options using a mechanical transfer switch),
channel 1 is driving one test port and channel 2 is driving the other test port. This
would cause the test port transfer switch to continually cycle. The instrument will not
allow the transfer switch or attenuator to c ontinuously switc h ranges in order to update
these measurements without the direct intervention of the operator.
If one of these conditions exist, the test set hold mode will engage, and the status notation
tsH will appear on the left side of the screen. The hold mode leaves the measurement
function in only one of the two measurements. To update both measurement setups, press
Sweep Setup
. Refer to "Source Attenuator Switch Protection" on
page 7-13.
Viewing Four Measurement Channels
Four m easur ement c hannels c an be viewed simultaneously by enabli ng aux il iar y cha nnels
3 and 4. Although independent of other channels in most variables, channels 3 and 4 are
permanently coupled to channels 1 and 2 respectively by stimulus. That is, if channel 1 is
set for a center frequency of 200 MHz and a span of 50 MHz, channel 3 will have the same
stimulus values.
NOTEChannels 1 and 2 are referred to as primary channels and channels 3 and 4
are referre d t o as au xiliary channels.
Channel 3 or 4 are activated when the Chan 3 or Chan 4 keys are pressed. Alternatively,
you can enable the auxiliary setting to ON. For example, if channel 1 is
active, pressing to ON enables channel 3 and its trace appears on the display.
Channel 4 is similarly enabled and viewed when channel 2 is active.
1. Press to select the type of display of the data. This example uses the log mag
format.
2. If channel 1 is not active, make it active by pressing .
3. Press , set to ON, set to
ON, and set to .
Format
Chan 1
Display
The display will appear as show n in Figure 1-9. Channel 1 is in the upper-left quadrant
of the display, channel 2 is in the upper-right quadrant, and channel 3 is in the lower
half of the display.
1-14
Figure 1-9 Three-Channel Display
Chan 2
AUX CHAN
Making Measurements
Using Display Functions
4. Press Chan 4 (or press , set to ON).
This enables channel 4 and the screen now displays four separate grids as shown in
Figure 1-1 0. Channel 4 is in the lower-right quadrant of the screen.
1-15
Making Measurements
MARKER 1
MARKER 2
MARKER MODE MENU
MARKERS:
Using Display Functions
Figure 1-10 Four-Channel Display
5. Press .
Observe that the amber LED adjacent to the key is lit and the CH4 indicator
Chan 4
Chan 4
on the display has a box around it. This indicates that channel 4 is now active and can
be configured.
6. Press .
Marker
Markers 1 and 2 appear on all four channel traces. Rotating the front panel control
knob moves marker 2 on all four channel traces. Note that the active function, in this
case the marker frequency, is the same color and in the same grid as the active channel
(channel 4).
7. Press .
Observe that the amber LED adjacent to the key is lit. This indicates that
Chan 3
Chan 3
channel 3 is now active and can be configured.
8. Rotate the front panel control knob and notice that marker 2 still moves on all four
channel traces.
9. To independently control the channel markers:
Press , set to UNCOUPLED.
Marker Fctn
Rotate the front panel control knob. Marker 2 moves only on the channel 3 trace.
1-16
Loading...
+ 448 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.