No part of this manual may be reproduced
in any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
This product may be used as a component of an in vitro diagnostic system if the system is registered with
the appropriate authorities and complies with the relevant regulations.
Otherwise, it is intended only for general laboratory use.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation
or contract clause. Use, duplication or disclosure of Software is subject to Agilent
Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood and met.
Contents
Contents
1 Safety Information5
Power Cords7
Operation8
Safety Symbols9
Laser Safety10
Chemical and Biological Safety11
Legal Notice13
2 Site Requirements for the Agilent 2100 Bioanalyzer15
Power Considerations16
Bench space17
Environment18
Sound Emission19
Physical Specifications of the Agilent 2100 Bioanalyzer20
3 Unpacking the Agilent 2100 Bioanalyzer System21
Damaged Packaging22
4 Installing the Agilent 2100 Bioanalyzer System25
Setting up PC and Printer26
Setting up the Agilent 2100 Bioanalyzer27
5 Starting the Agilent 2100 Bioanalyzer31
Turning on the Agilent 2100 Bioanalyzer32
Connecting the Bioanalyzer via the USB/Serial Cable33
Starting the Agilent 2100 Bioanalyzer Software34
Preparing the Assay35
6 Maintenance of the Agilent 2100 Bioanalyzer37
7 Spare Parts and Accessories39
3
Contents
4
1
WARNING
Safety Information
Power Cords7
Operation8
Safety Symbols9
Laser Safety10
Chemical and Biological Safety11
Legal Notice13
The following general safety precautions must be observed during all phases of
operation, service, and repair of the Agilent 2100 Bioanalyzer.
All safety instructions should be read and understood before installation,
operation and maintenance of the instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture and intended use of the instrument.
Agilent Technologies assumes no liability for the customer’s failure to comply
with these requirements.
A WARNING notice denotes a hazard.
It calls attention to an operating procedure, practice, or the like that, if not correctly
performed or adhered to, could result in personal injury or death.
➔ Do not proceed beyond a WARNING notice until the indicated conditions are fully
understood and met.
Agilent Technologies
5
1Safety Information
CAUTION
NOTE
A CAUTION notice denotes a hazard.
It calls attention to an operating procedure, practice, or the like that, if not correctly
performed or adhered to, could result in damage to the product or loss of important
data.
➔ Do not proceed beyond a CAUTION notice until the indicated conditions are fully
Instrument Safety: This is a Safety Class I instrument (provided with a terminal for
protective grounding) and has been manufactured and tested according to international
safety standards.
understood and met.
6
Power Cords
WARNING
Safety Information
Power Cords
Different power cords are offered as options with the Agilent 2100
Bioanalyzer. The female end of all power cords is identical. It plugs into the
power-input socket at the rear of the instrument. The male end of each power
cord is different and designed to match the wall socket of a particular country
or region.
Electric Shock
The absence of ground connection and the use of an unspecified power cord can
lead to electric shock or short circuit.
➔ Never operate your instrument from a power outlet that has no ground connection.
➔ Never use a power cord other than the Agilent Technologies power cord designed
for your region.
➔ Never use cables other than those supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
1
7
1Safety Information
WARNING
Operation
Operation
Ensure the proper usage of the equipment
The protection provided by the equipment may be impaired.
➔ The operator of this instrument is advised to use the equipment in a manner as
Before the instrument is switched on, all protective ground terminals,
extension cords, auto-transformers, and devices connected to it must be
connected to a protective ground socket. Any interruption of the protective
grounding will cause a potential shock hazard that could result in serious
personal injury. Whenever it is likely that the protection has been impaired,
the instrument must be made inoperative and be secured against any intended
operation.
To operate the instrument safely:
• Do not remove any cover of the Agilent 2100 Bioanalyzer
• Avoid any maintenance of the instrument under voltage
• Do not let liquid drip into the Agilent 2100 Bioanalyzer. It could cause a
• Do not replace components of the instruments (e.g. electrode cartridge or
• Do not operate the instrument in the presence of flammable gases or fumes.
• Do not install substitute parts or make any unauthorized modification to
specified in this manual.
shock or it could damage the Agilent 2100 Bioanalyzer.
pressure cartridge) with power turned on.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
the instrument.
8
Safety Symbols
SymbolDescription
Electrosensitive device.
The apparatus is marked with this symbol when the user should refer to the Installation and Safety Manual in order to prevent risk of harm to the operator and to protect the apparatus against damage.
Indicates dangerous voltages.
Indicates a protected ground terminal.
Laser Radiation. When open avoid exposure to beam.
Safety Information
Safety Symbols
1
9
1Safety Information
WARNING
Laser Safety
Laser Safety
The Agilent 2100 Bioanalyzer is classified as a "Laser Class 1" product
(IEC825-1, CFR1040.10). During normal operation of the Agilent 2100
Bioanalyzer no laser light is accessible to the user.
When the laser is in use, the laser light source emits light at a power that may
be harmful to the eyes. To prevent operator exposure to potentially harmful
laser light, an interlock mechanism turns off the laser whenever the lid of the
Agilent 2100 Bioanalyzer is not closed. Furthermore, the lid can't be
completely opened and has a black light absorbing surface in order to prevent
any potential laser light reflection.
Harmful laser light
The laser light source emits light at a power that may be harmful to the eyes.
➔ Never look into the beam or direct it towards someone else.
➔ System interlocks should never be disabled.
10
Chemical and Biological Safety
WARNING
WARNING
Toxic and hazardous solvents
The handling of solvents and reagents can hold health risks.
➔ When working with solvents observe appropriate safety procedures (for example,
goggles, safety gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when toxic or
hazardous solvents are used.
Pathogenic, toxic, or radioactive samples
Handling and use of pathogenic, toxic, or radioactive samples and of genetically
modified organisms holds risks for health and environment.
➔ Ensure that all necessary safety regulations, guidelines, precautions and practices
are adhered to accordingly.
Safety Information
Chemical and Biological Safety
1
➔ Consult the laboratory safety officer for advise on the level of containment required
for the application, and proper decontamination or sterilization procedures to follow
if fluids escape from containers.
To operate the instrument safely:
• Observe all cautionary information printed on the original solution
containers prior to their use.
• Because leaks, spills, or loss of sample may generate aerosols, observe
proper safety precautions.
• Agilent 2100 Bioanalyzer covers have not been designed as bioseals for
aerosol or liquid containment.
• Handle body fluids with care because they can transmit disease. No known
test offers complete assurance that they are free of micro-organisms. Some
of the most virulent – Hepatitis (B and C) and HIV (I-V) viruses, atypical
mycobacteria, and certain systemic fungi – further emphasize the need for
aerosol protection.
11
1Safety Information
WARNING
Chemical and Biological Safety
• Always follow local state and federal biohazard handling regulation when
• Handle all infectious samples according to good laboratory procedures and
• Dispose of all waste solutions and products according to appropriate
Harmful chemical and biological substances
Residues of clinical samples may contain chemicals and biological substances that
are dangerous and harmful to persons working on the 2100 Bioanalyzer.
➔ The Agilent 2100 Bioanalyzer and any accessories must be decontaminated before
disposing of biohazardous waste material e.g. used Cell LabChips or
contaminated pressure adapters.
methods to prevent spread of disease.
environmental health and safety guidelines.
requesting service by an Agilent Technologies Field Service Representative and
before returning the 2100 Bioanalyzer to Agilent Technologies for repair or
replacement.
12
Legal Notice
Safety Information
Legal Notice
The Agilent 2100 Bioanalyzer is sold for research use only.
By purchasing this instrument, the purchaser is granted the limited right to
use only this instrument. Purchase of this instrument does not include any
right, express or implied, to use any other patented product, method or
process, or to use any other portion or component of any patented system or
systems, software, microfluidic devices or reagents, either alone or in
conjunction with this product, unless use of such method, process or other
portion or component is separately authorized.
1
13
Loading...
+ 29 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.