Agilent Technologies 2100 Installation And Safety Manual

Agilent 2100 Bioanalyzer
Installation and Safety Guide
Notices
CAUTION
WARNING
© Agilent Technologies, Inc.
No part of this manual may be reproduced in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
G2938-90007 Rev B
Edition
Printed in Germany
Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn
This product may be used as a com­ponent of an in vitro diagnostic sys­tem if the system is registered with the appropriate authorities and com­plies with the relevant regulations. Otherwise, it is intended only for gen­eral laboratory use.
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcon­tract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer soft­ware” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or contract clause. Use, duplication or dis­closure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will
receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated condi­tions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indi­cated conditions are fully under­stood and met.
Contents
Contents
1 Safety Information 5
Power Cords 7 Operation 8 Safety Symbols 9 Laser Safety 10 Chemical and Biological Safety 11 Legal Notice 13
2 Site Requirements for the Agilent 2100 Bioanalyzer 15
Power Considerations 16 Bench space 17 Environment 18 Sound Emission 19 Physical Specifications of the Agilent 2100 Bioanalyzer 20
3 Unpacking the Agilent 2100 Bioanalyzer System 21
Damaged Packaging 22
4 Installing the Agilent 2100 Bioanalyzer System 25
Setting up PC and Printer 26 Setting up the Agilent 2100 Bioanalyzer 27
5 Starting the Agilent 2100 Bioanalyzer 31
Turning on the Agilent 2100 Bioanalyzer 32 Connecting the Bioanalyzer via the USB/Serial Cable 33 Starting the Agilent 2100 Bioanalyzer Software 34 Preparing the Assay 35
6 Maintenance of the Agilent 2100 Bioanalyzer 37
7 Spare Parts and Accessories 39
3
Contents
4
1
WARNING

Safety Information

Power Cords 7
Operation 8
Safety Symbols 9
Laser Safety 10
Chemical and Biological Safety 11
Legal Notice 13
The following general safety precautions must be observed during all phases of operation, service, and repair of the Agilent 2100 Bioanalyzer.
All safety instructions should be read and understood before installation, operation and maintenance of the instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture and intended use of the instrument.
Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
A WARNING notice denotes a hazard.
It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death.
Do not proceed beyond a WARNING notice until the indicated conditions are fully
understood and met.
Agilent Technologies
5
1 Safety Information
CAUTION
NOTE
A CAUTION notice denotes a hazard.
It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data.
Do not proceed beyond a CAUTION notice until the indicated conditions are fully
Instrument Safety: This is a Safety Class I instrument (provided with a terminal for protective grounding) and has been manufactured and tested according to international safety standards.
understood and met.
6

Power Cords

WARNING
Safety Information
Power Cords
Different power cords are offered as options with the Agilent 2100 Bioanalyzer. The female end of all power cords is identical. It plugs into the power-input socket at the rear of the instrument. The male end of each power cord is different and designed to match the wall socket of a particular country or region.
Electric Shock
The absence of ground connection and the use of an unspecified power cord can lead to electric shock or short circuit.
Never operate your instrument from a power outlet that has no ground connection.
Never use a power cord other than the Agilent Technologies power cord designed
for your region.
Never use cables other than those supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
1
7
1 Safety Information
WARNING

Operation

Operation
Ensure the proper usage of the equipment
The protection provided by the equipment may be impaired.
The operator of this instrument is advised to use the equipment in a manner as
Before the instrument is switched on, all protective ground terminals, extension cords, auto-transformers, and devices connected to it must be connected to a protective ground socket. Any interruption of the protective grounding will cause a potential shock hazard that could result in serious personal injury. Whenever it is likely that the protection has been impaired, the instrument must be made inoperative and be secured against any intended operation.
To operate the instrument safely:
Do not remove any cover of the Agilent 2100 Bioanalyzer
Avoid any maintenance of the instrument under voltage
Do not let liquid drip into the Agilent 2100 Bioanalyzer. It could cause a
Do not replace components of the instruments (e.g. electrode cartridge or
Do not operate the instrument in the presence of flammable gases or fumes.
Do not install substitute parts or make any unauthorized modification to
specified in this manual.
shock or it could damage the Agilent 2100 Bioanalyzer.
pressure cartridge) with power turned on.
Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
the instrument.
8

Safety Symbols

Symbol Description
Electrosensitive device.
The apparatus is marked with this symbol when the user should refer to the Installation and Safety Manual in order to prevent risk of harm to the operator and to protect the apparatus against damage.
Indicates dangerous voltages.
Indicates a protected ground terminal.
Laser Radiation. When open avoid exposure to beam.
Safety Information
Safety Symbols
1
9
1 Safety Information
WARNING

Laser Safety

Laser Safety
The Agilent 2100 Bioanalyzer is classified as a "Laser Class 1" product (IEC825-1, CFR1040.10). During normal operation of the Agilent 2100 Bioanalyzer no laser light is accessible to the user.
When the laser is in use, the laser light source emits light at a power that may be harmful to the eyes. To prevent operator exposure to potentially harmful laser light, an interlock mechanism turns off the laser whenever the lid of the Agilent 2100 Bioanalyzer is not closed. Furthermore, the lid can't be completely opened and has a black light absorbing surface in order to prevent any potential laser light reflection.
Harmful laser light
The laser light source emits light at a power that may be harmful to the eyes.
Never look into the beam or direct it towards someone else.
System interlocks should never be disabled.
10

Chemical and Biological Safety

WARNING
WARNING
Toxic and hazardous solvents
The handling of solvents and reagents can hold health risks.
When working with solvents observe appropriate safety procedures (for example,
goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
Pathogenic, toxic, or radioactive samples
Handling and use of pathogenic, toxic, or radioactive samples and of genetically modified organisms holds risks for health and environment.
Ensure that all necessary safety regulations, guidelines, precautions and practices
are adhered to accordingly.
Safety Information
Chemical and Biological Safety
1
Consult the laboratory safety officer for advise on the level of containment required
for the application, and proper decontamination or sterilization procedures to follow if fluids escape from containers.
To operate the instrument safely:
Observe all cautionary information printed on the original solution
containers prior to their use.
Because leaks, spills, or loss of sample may generate aerosols, observe
proper safety precautions.
Agilent 2100 Bioanalyzer covers have not been designed as bioseals for
aerosol or liquid containment.
Handle body fluids with care because they can transmit disease. No known
test offers complete assurance that they are free of micro-organisms. Some of the most virulent – Hepatitis (B and C) and HIV (I-V) viruses, atypical mycobacteria, and certain systemic fungi – further emphasize the need for aerosol protection.
11
1 Safety Information
WARNING
Chemical and Biological Safety
Always follow local state and federal biohazard handling regulation when
Handle all infectious samples according to good laboratory procedures and
Dispose of all waste solutions and products according to appropriate
Harmful chemical and biological substances
Residues of clinical samples may contain chemicals and biological substances that are dangerous and harmful to persons working on the 2100 Bioanalyzer.
The Agilent 2100 Bioanalyzer and any accessories must be decontaminated before
disposing of biohazardous waste material e.g. used Cell LabChips or contaminated pressure adapters.
methods to prevent spread of disease.
environmental health and safety guidelines.
requesting service by an Agilent Technologies Field Service Representative and before returning the 2100 Bioanalyzer to Agilent Technologies for repair or replacement.
12

Legal Notice

Safety Information
Legal Notice
The Agilent 2100 Bioanalyzer is sold for research use only.
By purchasing this instrument, the purchaser is granted the limited right to use only this instrument. Purchase of this instrument does not include any right, express or implied, to use any other patented product, method or process, or to use any other portion or component of any patented system or systems, software, microfluidic devices or reagents, either alone or in conjunction with this product, unless use of such method, process or other portion or component is separately authorized.
1
13
Loading...
+ 29 hidden pages