The Agilent Technologies 16760A 1500 Mb/s State/800 MHz Timing
logic analyzer offers 64M deep memory with up to 170 channels on a
single time base (5 cards, 34 channels per card). Differential probing
captures input signals as low as 200 mV p-p.
“Getting Started” on
page 13
“Probing and
Selecting the
Sampling Mode” on
page 33
•“Probing and Sampling Mode Selection Steps” on page 15
•“Timing Mode or State Mode Steps” on page 22
•“Eye Scan Mode Steps” on page 26
•“Probing the Device Under Test” on page 35
•“Using the E5378A Single-Ended Probe” on page 35
•“Using the E5379A Differential Probe” on page 37
•“Using the E5380A Mictor-Compatible Probe” on page 39
•“Using the E5382A Single-ended Flying Lead Probe Set” on page 40
•“Using an Analysis Probe” on page 41
•“Choosing the Sampling Mode” on page 43
•“Selecting the Timing Mode (Asynchronous Sampling)” on page 43
•“Selecting the State Mode (Synchronous Sampling)” on page 46
•“In Either Timing Mode or State Mode” on page 53
•“Selecting the Eye Scan Mode” on page 55
•“Formatting Labels for Logic Analyzer Probes” on page 57
“Using the Logic
Analyzer in Timing or
State Mode” on
page 67
•“Setting Up Triggers and Running Measurements” on page 69
•“Using Trigger Functions” on page 70
•“Using Other Trigger Features” on page 75
2
Agilent Technologies 16760A Logic Analyzer
•“Editing the Trigger Sequence (Timing or 200, 400 Mb/s State Only)”
on page 78
•“Editing Advanced Trigger Functions (Timing or 200 Mb/s State Only)”
on page 83
•“Saving/Recalling Trigger Setups” on page 90
•“Running Measurements” on page 91
•“Displaying Captured Data” on page 94
•“Using Symbols” on page 101
•“Printing/Exporting Captured Data” on page 110
•“Cross-Triggering” on page 112
•“Solving Logic Analysis Problems” on page 114
•“Saving and Loading Logic Analyzer Configurations” on page 116
“Using the Logic
Analyzer in Eye Scan
Mode” on page 117
“Reference” on
page 153
“Concepts” on
page 239
•“Setting Up and Running Eye Scan Measurements” on page 119
•“Displaying Captured Eye Scan Data” on page 133
•“Saving and Loading Captured Eye Scan Data” on page 152
•“The Sampling Tab” on page 155
•“The Format Tab” on page 174
•“The Trigger Tab” on page 176
•“The Symbols Tab” on page 193
•“The Eye Scan Tab” on page 204
•“The Calibration Tab” on page 210
•“Error Messages” on page 211
•“Specifications and Characteristics” on page 227
•“Understanding Logic Analyzer Triggering” on page 240
•“Understanding State Mode Sampling Positions” on page 256
•“Understanding Eye Scan Measurements” on page 259
3
Agilent Technologies 16760A Logic Analyzer
See AlsoMain System Help (see the Agilent Technologies 16700A/B-Series Logic
Analysis System help volume)
Glossary (see page 263)
4
Contents
Agilent Technologies 16760A Logic Analyzer
1 Getting Started
Probing and Sampling Mode Selection Steps15
Step 1. Connect logic analyzer to the device under test15
Step 2. Choose the sampling mode16
Step 3. Format labels for the probed signals19
Timing Mode or State Mode Steps22
Step 4. Define the trigger condition22
Step 5. Run the measurement23
Step 6. Display the captured data24
Eye Scan Mode Steps26
Step 4. Select channels for the eye scan measurement26
Step 5. Set the eye scan range and resolution27
Step 6. Run the eye scan measurement28
Step 7. Display the captured eye scan data29
For More Information...31
2 Probing and Selecting the Sampling Mode
Probing the Device Under Test35
Using the E5378A Single-Ended Probe35
Using the E5379A Differential Probe37
Using the E5380A Mictor-Compatible Probe39
Using the E5382A Single-ended Flying Lead Probe Set40
Using an Analysis Probe41
5
Contents
Choosing the Sampling Mode43
Selecting the Timing Mode (Asynchronous Sampling)43
Selecting the State Mode (Synchronous Sampling)46
In Either Timing Mode or State Mode53
Selecting the Eye Scan Mode55
Formatting Labels for Logic Analyzer Probes57
To assign pods to the analyzer57
To set pod threshold voltages58
To set clock threshold voltages59
To assign probe channels to labels60
To import label names and assignments from a netlist62
To import label definitions from an ASCII file63
To export label definitions to an ASCII file64
To change the label polarity64
To reorder bits in a label65
To turn labels off or on66
3 Using the Logic Analyzer in Timing or State Mode
Setting Up Triggers and Running Measurements69
Using Trigger Functions70
Using Other Trigger Features75
Editing the Trigger Sequence (Timing or 200, 400 Mb/s State Only)78
Editing Advanced Trigger Functions (Timing or 200 Mb/s State Only)83
Saving/Recalling Trigger Setups90
Running Measurements91
Displaying Captured Data94
To open Waveform or Listing displays94
To use other display tools95
If the captured data doesn't look correct97
If there are filtered data holes in display memory98
To display symbols for data values99
To cancel the display processing of captured data100
6
Contents
Using Symbols101
To load object file symbols102
To adjust symbol values for relocated code103
To create user-defined symbols104
To enter symbolic label values105
To create an ASCII symbol file106
To create a readers.ini file107
Printing/Exporting Captured Data110
Cross-Triggering112
To cross-trigger with another instrument112
Solving Logic Analysis Problems114
To test the logic analyzer hardware114
Saving and Loading Logic Analyzer Configurations116
4 Using the Logic Analyzer in Eye Scan Mode
Setting Up and Running Eye Scan Measurements119
To select channels for the eye scan119
To set the eye scan range and resolution120
To run an eye scan measurement121
To set advanced eye scan options121
To set up qualified eye scan measurements122
To comment on the eye scan settings132
7
Contents
Displaying Captured Eye Scan Data133
To open the Eye Scan display133
To select the channels displayed134
To scale the Eye Scan display135
To set Eye Scan display options136
To make measurements on the eye scan data142
To display information about the eye scan data 149
To comment on the eye scan data151
Analyzer armed from another module contains no "Arm in from IMB"
event212
Branch expression is too complex212
Cannot specify range on label with clock bits that span pod pairs217
Counter value checked as an event, but no increment action specified218
Goto action specifies an undefined level218
Hardware Initialization Failed 218
Maximum of 32 Channels Per Label219
Must assign another pod pair to specify actions for flags219
Must assign Pod 1 on the master card to specify actions for flags 219
No more Edge/Glitch resources available for this pod pair219
No more Pattern resources available for this pod pair220
No Trigger action found in the trace specification221
Slow or Missing Clock221
Timer value checked as an event, but no start action specified222
Trigger function initialization failure 222
Trigger inhibited during timing prestore223
Trigger Specification is too complex 223
Waiting for Trigger225
9
Contents
Specifications and Characteristics227
E5378A Single-Ended Probe Specifications and Characteristics228
E5379A Differential Probe Specifications and Characteristics228
E5380A MICTOR-Compatible Probe Specifications and Characteristics229
1500 Mb/s Sampling Mode Specifications and Characteristics230
1250 Mb/s Sampling Mode Specifications and Characteristics231
800 Mb/s Sampling Mode Specifications and Characteristics232
400 Mb/s Sampling Mode Specifications and Characteristics233
200 Mb/s Sampling Mode Specifications and Characteristics234
Conventional Timing Mode Specifications and Characteristics235
Transitional Timing Mode Specifications and Characteristics235
What is a Specification?236
What is a Characteristic?237
6 Concepts
Understanding Logic Analyzer Triggering240
The Conveyor Belt Analogy240
Summary of Triggering Capabilities 242
Sequence Levels 242
Boolean Expressions245
Branches246
Edges246
Ranges246
Flags247
Occurrence Counters and Global Counters247
Timers248
Storage Qualification249
Strategies for Setting Up Triggers 251
Conclusions255
Understanding State Mode Sampling Positions256
Understanding Eye Scan Measurements259
10
Contents
Glossary
Index
11
Contents
12
1
Getting Started
After you have connected the logic analyzer probes to your device
under test (see “Step 1. Connect logic analyzer to the device under
test” on page 15), all measurements will have the following initial steps:
13
Chapter 1: Getting Started
•“Step 2. Choose the sampling mode” on page 16
•“Step 3. Format labels for the probed signals” on page 19
In the timing (asynchronous) or state (synchronous) sampling modes,
measurements will have these steps:
•“Step 4. Define the trigger condition” on page 22
•“Step 5. Run the measurement” on page 23
•“Step 6. Display the captured data” on page 24
In the eye scan sampling mode, measurements will have these steps:
•“Step 4. Select channels for the eye scan measurement” on page 26
•“Step 5. Set the eye scan range and resolution” on page 27
•“Step 6. Run the eye scan measurement” on page 28
•“Step 7. Display the captured eye scan data” on page 29
If you have previously saved a logic analyzer setup to a configuration
file, or if configuration files are included with an analysis probe, you
can load the configuration file to set up the logic analyzer and the
measurement.
Once you have made a logic analyzer measurement, the measurement
can be refined by repeating the measurement set up, run, and display
steps.
Next: “Step 1. Connect logic analyzer to the device under test” on
page 15
14
Chapter 1: Getting Started
Probing and Sampling Mode Selection Steps
Probing and Sampling Mode Selection Steps
You will always take the following steps regardless of the sampling
mode you plan to use.
•“Step 1. Connect logic analyzer to the device under test” on page 15
•“Step 2. Choose the sampling mode” on page 16
•“Step 3. Format labels for the probed signals” on page 19
Step 1. Connect logic analyzer to the device
under test
Before you begin setting up the logic analyzer for a measurement, you
need to physically connect the logic analyzer to your device under test.
There are four probing options available to connect your logic analyzer
to the device under test:
•Single-ended probes with Samtec connectors (see page 35).
•Differential probes with Samtec connectors (see page 37).
•Single-ended probes with MICTOR connectors (see page 39).
•Use an analysis probe to connect to microprocessors and standard buses.
When using an analysis probe, the Setup Assistant guides you through the
connection and setup process for your particular logic analyzer and
analysis probe.
Next: “Step 2. Choose the sampling mode” on page 16
15
Chapter 1: Getting Started
Probing and Sampling Mode Selection Steps
Step 2. Choose the sampling mode
There are three logic analyzer sampling modes to choose from: timing
mode, state mode, and eye scan mode.
In timing mode, the logic analyzer samples asynchronously, based on
an internally-generated sampling clock.
In state mode, the logic analyzer samples synchronously, based on a
sampling clock signal from the device under test. Typically, the signal
used for sampling in state mode is a state machine or microprocessor
clock signal.
In eye scan mode, the logic analyzer samples small windows of time
and voltage on data channels around a clock signal from the device
under test. The resulting eye diagrams let you validate and
characterize the data valid windows of signals latched on the clock.
To choose the sampling mode
1. In the Sampling tab, choose Timing Mode, State Mode, or Eye Scan
Mode.
16
Chapter 1: Getting Started
Probing and Sampling Mode Selection Steps
If you chose Timing Mode
1. Select the timing analyzer conventional/transitional configuration.
In the transitional timing configuration, the logic analyzer can capture a
greater period of execution because only transitions are stored in memory.
2. If you chose the transitional timing configuration, set the sample period.
To capture signal level changes reliably, the sample period should be less
than half (many engineers prefer one-fourth) of the period of the fastest
signal you want to measure.
If you chose State Mode
1. Select the state analyzer speed configuration.
There are trade-offs between high-speed sampling, the number of available
channels, triggering capabilities, and other logic analyzer characteristics.
For example, in the 1250 Mb/s configuration, a periodic clock input signal
is required.
2. In the Clock Setup, specify which clock signal edges from the device under
test will be used as the sampling clock.
17
Chapter 1: Getting Started
Probing and Sampling Mode Selection Steps
3. Specify the sampling position. Select the Sampling Positions... button,
then select the Run Eye Finder button to locate the data valid window in
relation to the sampling clock, and automatically set the sampling position
of the logic analyzer.
See Also “To automatically adjust sampling positions” on page 49
In either Timing Mode or State Mode
1. Specify the trigger position.
The trigger is the event in the device under test that you want to capture
data around.
Specify whether you want to look at data after the trigger (Start), before
and after the trigger (Center), before the trigger (End), or use a
percentage of the logic analyzer's memory for data after the trigger (User
Defined).
2. Set the acquisition memory depth.
If you need less data and want measurements to run faster, you can limit
the amount of trace memory that is filled with samples.
If you chose Eye Scan Mode
1. Select the eye scan mode speed configuration.
There are trade-offs between high-speed sampling, the number of available
channels, and other logic analyzer characteristics.
2. In the Clock Setup, specify which clock signal edges from the device under
test will be used as the reference clock for the eye scan measurement.
Next: “Step 3. Format labels for the probed signals” on page 19
18
Chapter 1: Getting Started
Probing and Sampling Mode Selection Steps
Step 3. Format labels for the probed signals
When a logic analyzer probes hundreds of signals in a device under
test, you need to be able to give those channels more meaningful
names than "pod 1, channel 1".
The Format tab is mainly used for assigning bus and signal names
(from the device under test) to logic analyzer channels. These names
are called labels. Labels are also used when setting up triggers and
displaying captured data.
The Format tab also lets you do things like assign pods to the logic
analyzer and specify the logic analyzer threshold voltage.
The Format tab has activity indicators that show whether the signal a
channel is probing is above the threshold voltage (high), below the
threshold voltage (low), or transitioning.
19
Chapter 1: Getting Started
Probing and Sampling Mode Selection Steps
To assign pods to the logic analyzer
1. In the Format tab, select the Pod Assignment button.
2. In the Pod Assignment dialog, drag a pod to the appropriate logic analyzer.
3. Select the Close button.
To specify threshold voltages
The threshold voltage is the voltage level that a signal must cross
before the logic analyzer recognizes a change in logic levels.
1. In the Format tab, select the button under the pod name.
2. In the Pod threshold dialog, select a Standard, External Ref, Differential,
or a User Defined threshold voltage.
3. Select the Close button.
4. Select the Clk Thresh button.
5. In the Clock Thresholds dialog, select the button of the clock whose
threshold you wish to set.
6. In the J, K, etc., threshold dialog, select a Standard, Differential, or a User
20
Chapter 1: Getting Started
Probing and Sampling Mode Selection Steps
Defined threshold voltage.
7. Select the Close button.
8. Select the Close button.
To assign names to logic analyzer channels
1. Select a label button, and either:
•Choose the Rename command, enter the label name, and select the OK
button.
•Or, choose the Insert before or Insert after command, enter the label
name, and select the OK button.
2. In the label row, select the button of the pod that contains the channels
you want to assign.
3. Either choose one of the standard channel assignments--dots (.) mean the
channel is unassigned, asterisks (*) mean the channel is assigned--or
choose Individual.
If you chose Individual:
a. In the "label - pod" dialog, select the channels you want to assign/
unassign.
b. Select the OK button.
Next:
•If you are using the logic analyzer in timing mode or in state mode, go to
“Step 4. Define the trigger condition” on page 22.
•If you are using the logic analyzer in eye scan mode, go to “Step 4. Select
channels for the eye scan measurement” on page 26.
21
Chapter 1: Getting Started
Timing Mode or State Mode Steps
Timing Mode or State Mode Steps
When you have selected the timing or state sampling modes, you need
to perform the following steps.
•“Step 4. Define the trigger condition” on page 22
•“Step 5. Run the measurement” on page 23
•“Step 6. Display the captured data” on page 24
Step 4. Define the trigger condition
The trigger is the event in the device under test that you want to
capture data around.
1. In the Trigger tab, and in the Trigger Functions subtab, choose the type of
trigger you want to specify, and select the Replace button.
22
Chapter 1: Getting Started
Timing Mode or State Mode Steps
2. In the Trigger Sequence portion of the Trigger tab, select the buttons to
define the label values and/or other conditions you want to trigger on.
Next: “Step 5. Run the measurement” on page 23
Step 5. Run the measurement
Once the trigger condition has been defined, you can run the
measurement.
1. Select the Run Single button .
When you run a measurement, the Stop button becomes available while
the logic analyzer looks for the trigger condition.
23
Chapter 1: Getting Started
Timing Mode or State Mode Steps
Logic analyzers with deep acquisition memory take a noticeable amount of
time to complete a run; however, messages like "Waiting in level 1" may
indicate you need to stop the measurement and refine the trigger
condition.
When the trigger condition is found, logic analyzer acquisition memory is
filled, the captured data is processed to the display tools, and the Run
Single button becomes available again.
Next: “Step 6. Display the captured data” on page 24
Step 6. Display the captured data
Once you have run a measurement and filled the logic analyzer's
acquisition memory with captured data, you can display it with one of
the display tools.
To open Waveform or Listing displays
Waveform displays are typically used when data is captured with the
timing sampling mode, and Listing displays are used when data is
captured with the state sampling mode.
1. From the Window menu, select your logic analyzer and choose the
Waveform or Listing command.
24
Chapter 1: Getting Started
Timing Mode or State Mode Steps
To add display tools via the Workspace window
1. Select the Workspace button (or from the Window menu, select System
and Workspace).
2. In the Workspace window, scroll down to the Display portion of the tool
icon list.
3. Drag the display tool icon and drop it on the analyzer icon.
4. To open the display tool, select its icon and choose the Display command.
Next: “For More Information...” on page 31
25
Chapter 1: Getting Started
Eye Scan Mode Steps
Eye Scan Mode Steps
When you have selected the eye scan sampling mode, you need to
perform the following steps.
•“Step 4. Select channels for the eye scan measurement” on page 26
•“Step 5. Set the eye scan range and resolution” on page 27
•“Step 6. Run the eye scan measurement” on page 28
•“Step 7. Display the captured eye scan data” on page 29
Step 4. Select channels for the eye scan
measurement
1. In the Eye Scan tab, Labels subtab, select the channels you want in the eye
scan measurement.
26
Chapter 1: Getting Started
Eye Scan Mode Steps
Next: “Step 5. Set the eye scan range and resolution” on page 27
Step 5. Set the eye scan range and resolution
1. In the Eye Scan tab, Scan Settings subtab, select the settings for the eye
scan measurement.
27
Chapter 1: Getting Started
Eye Scan Mode Steps
These settings define the number and size of the time and voltage windows
used in the eye scan.
Measurements using the coarse settings run faster because there are fewer
time and voltage windows to scan, but the resulting eye diagrams have less
resolution.
Measurements using the fine settings result in eye diagrams with better
resolution, but the measurements take longer to run because there are
more time and voltage windows to scan.
Next: “Step 6. Run the eye scan measurement” on page 28
Step 6. Run the eye scan measurement
28
Chapter 1: Getting Started
Eye Scan Mode Steps
Once the eye scan settings have been selected, you can run the
measurement.
1. Select the Run Single button .
The Eye Scan display window opens, and the captured measurement data
begins to appear.
While the eye scan measurement runs, the Stop button becomes available.
The estimated time of the measurement is shown in the status field.
Next: “Step 7. Display the captured eye scan data” on page 29
Step 7. Display the captured eye scan data
29
Chapter 1: Getting Started
Eye Scan Mode Steps
In the Eye Scan display window, use the following subtabs:
•Scale to zoom in or out on the captured data.
•Display to change the display options.
•Measurements to use tools for displaying useful information about the
data.
•Info to display textual information about the captured measurement data.
•Comments to save your comments with the data.
Next: “For More Information...” on page 31
30
Loading...
+ 255 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.