The information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright.All
rights are reserved. No part of this document may be photocopied, reproduced, or
translated to another language without the prior written consent of the Agilent
Technologies.
Agilent Technologies Japan, Ltd.
Component Test PGU-Kobe
1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan
The manual’s printing date and part number indicate its current edition. The
printing date changes when a new edition is printed. (Minor corrections and
updates that are incorporated at reprint do not cause the date to change.) The
manual part number changes when extensive technical changes are incorporated.
1990 First Edition (part number: 16048-90000)
November 1998 Second Edition (part number: 16048-90001)
August 1999 Third Edition (part number: 16048-90001)
March 2000 Fourth Edition (part number: 16048-90001)
Safety Summary
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific WARNINGS elsewhere in this manual may impair the
protection provided by the equipment. In addition it violates safety standards of
design, manufacture, and intended use of the instrument.
Agilent Technologies assumes no liability for the customer’s failure to comply
with these requirements.
NOTE16048A comply with INSTALLATION CATEGORY I and POLLUTION
DEGREE 2 in IEC61010-1. 16048A are INDOOR USE product.
2
•Ground The Instrument
To avoid electric shock hazard, the instrument chassis and cabinet must be
connected to a safety earth ground by the supplied power cable with earth
blade.
•DO NOT Operate In An Explosive Atmosphere
Do not operate the instrument in the presence of flammable gasses or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
•Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component
replacement and internal adjustments must be made by qualified maintenance
personnel. Do not replace components with the power cable connected. Under
certain conditions, dangerous voltages may exist even with the power cable
removed. To avoid injuries, always disconnect power and discharge circuits
before touching them.
•DO NOT Service Or Adjust Alone
Do not attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present.
•DO NOT Substitute Parts Or Modify Instrument
Because of the danger of introducing additional hazards, do not install
substitute parts or perform unauthorized modifications to the instrument.
Return the instrument to a Agilent Technologies Sales and Service Office for
service and repair to ensure that safety features are maintained.
•Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous
procedures throughout this manual. Instructions contained in the warnings must
be followed.
WARNINGDangerous voltages, capable of causing death, are presenting this instrument.
Use extreme caution when handling, testing, and adjusting this instrument.
Certification
Agilent Technologies certifies that this product met its published specifications at
the time of shipment from the factory. Agilent Technologies further certifies that
its calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institution’s calibration
facility, or to the calibration facilities of other International Standards Organization
members.
3
Warranty
This Agilent Technologies instrument product is warranted against defects in
material and workmanship for a period corresponding to the individual warranty
periods of its component products. Instruments are warranted for a period of one
year. Fixtures and adapters are warranted for a period of 90 days. During the
warranty period, Agilent Technologies Company will, at its option, either repair or
replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to
Agilent Technologies and Agilent Technologies shall pay shipping charges to
return the product to Buyer. However, Buyer shall pay all shipping charges, duties,
and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by
Agilent Technologies for use with an instrument will execute its programming
instruction when property installed on that instrument. Agilent Technologies does
not warrant that the operation of the instrument, or software, or firmware will be
uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorized modification or misuse, operation outside the environmental
specifications for the product, or improper site preparation or maintenance.
IMPORTANTNo other warranty is expressed or implied. Agilent Technologies specifically
disclaims the implied warranties of merchantability and fitness for a particular
purpose.
Exclusive Remedies
The remedies provided herein are buyer’s sole and exclusive remedies. Agilent
Technologies shall not be liable for any direct, indirect, special, incidental, or
consequential damages, whether based on contract, tort, or any other legal theory.
4
Assistance
Product maintenance agreements and other customer assistance agreements are
available for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service
Office. Addresses are provided at the back of this manual.
Safety Symbol
General definitions of safety symbols used on the instrument or in manuals are
listed below.
Instruction Manual symbol: the product is marked with this symbol when it is
!
necessary for the user to refer to the instrument manual.
WARNINGThis warning sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could
result in injury or death to personnel.
CAUTIONThis Caution sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could result
in damage to or destruction of part or all of the product.
NOTENote denotes important information. It calls attention to a procedure, practice,
condition or the like, which is essential to highlight.
This operating note provides complete information on the 16048A Test Leads. The
16048A is shown pictorially in Figure 1-1, its physical dimensions are given in
Table 1-1, and typical characteristics related to offset error are given in Table 1-2.
To order additional copies of this operating note, use the part number listed on the
rear cover.
9
Operation
Product Description
Product Description
The 16048A consists of a direct attachment, 4-terminal pair type fixture which is
equipped with four BNC (m) terminated-coaxial test leads. These test leads are
used to attach user-fabricated test fixtures. DC bias levels of up to ±40V can be
applied to the 16048A. Cable length is 1 meter. The 16048A is shown in Figure
1-1.
Figure 1-1Product Overview
10Chapter 1
Specifications
Table 1-1Specifications of the 16048A
Function:For use with 4 terminal-pair LCR Meters and
INSTALLATION CATEGORY I
POLLUTION DEGREE 2
INDOOR USE
Chapter 111
Operation
Typical Characteristics
Typical Characteristics
Table 1-2Typical Characteristics
Incremental error at freq. ³1 MHz
Parameter reading
Offset value for D
error (%)
5
æö
´2±
èø
-----10
f
±
0.02
æö
´
èø
-----10
2
f
NOTEf is the measurement frequency in megahertz. The incremental errors calculated
from the equation in the table for measurements at frequencies above 1 MHz are
additive.
Figure 1-2
12Chapter 1
Operation
Compensation for Fixture Residual Impedance Error
Compensation for Fixture Residual Impedance Error
The 16048A has inherent stray capacitance, residual inductance, and residual
resistance that affect the accuracy of measured values. To compensate for, or
negate, these residuals to minimize measurement error, the instrument’s
Open/Short compensation procedure should be performed. The procedure is given
in the instrument’s operating manual.
Chapter 113
Operation
Compensation for Fixture Residual Impedance Error
14Chapter 1
2Service
15
Service
Maintenance
Maintenance
An exploded view of the 16048A (for parts identification) is shown in Figure 2-1.
Do not disassemble any further than shown. Maintenance consists principally of
cleaning contacts and replacing worn or damaged parts. Take special care when
cleaning contacts. To order parts, use the Agilent Technologies part numbers listed
in Table 2-1. If a faulty part is located in an assembly that cannot be disassembled,
order the next higher assembly or return the whole device to the nearest Agilent
Technologies Sales/Service Office for repair or replacement.
16Chapter 2
Figure 2-1Parts Identification
Service
Maintenance
Chapter 217
Service
Maintenance
Table 2-1Parts Identification
ReferenceAgilent Part No.Qty.DescriptionNote
1*4SLEEVE
2*4NUT
3*4NUT
4*4WASHER
5*4WASHER
62360-01922SCREW
7*1COVER-BOTTOM
8*4INSULATOR
9*2CONNECTOR-BNC
10*2CONNECTOR-BNC
112200-01031SCREW
122190-02061WASHER
1316047-400001STOPPER
1516048-040001COVER-TOP
161400-07192TIE RAP
17*1GROMMET
18*1CABLE
19*1BUSHING
212360-01151SCREW
223050-00101WASHER
2316021-500211CABLE CRAMP
2416021-500221CABLE CRAMP
25*4BOOT-BNC
26*4SLEEVE
27*4SLEEVE
28*4INSULATOR
291250-00894CONTACT
301250-00524CONNECTOR-BNC
16048-60010
a
1TEST LEAD 1 thru 30
18Chapter 2
Service
Maintenance
ReferenceAgilent Part No.Qty.DescriptionNote
311250-01184CONNECTOR-BNC
FEMALE
322360-01152SCREW1 thru 30
3316032-100211PLATE
342190-00164WASHER
352950-00014NUT
3616032-100221PLATE
3716032-600011BNC BRACKET31thru36
a. Agilent internal-only part number.
*: Not separately replaceable.
Chapter 219
Service
Maintenance
20Chapter 2
REGIONAL SALES AND SUPPORT OFFICES
For more information about Agilent Technologies test and measurement products, applications, services, and
for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one
of the following centers and ask for a test and measurement sales representative. 11/29/99
United States:
Agilent Technologies
Test and Measurement Call Center
P.O.Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 4844
Canada:
Agilent Technologies Canada Inc.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(tel) 1 877 894 4414
Europe:
Agilent Technologies
Test & Measurement
European Marketing Organization
P.O.Box 999
1180 AZ Amstelveen
The Netherlands
(tel) (31 20) 547 9999
Agilent Technologies
24/F, Cityplaza One, 1111 King’s Road,
Taikoo Shing, Hong Kong
(tel) (852)-3197-7777
(fax) (852)-2506-9284
Japan:
Agilent Technologies Japan Ltd.
Call Center
9-1, Takakura-Cho, Hachioji-Shi,
Tokyo 192-8510, Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Latin America:
Agilent Technologies
Latin American Region Headquarters
5200 Blue Lagoon Drive, Suite #950
Miami, Florida 33126
U.S.A.
(tel) (305) 267 4245