Notices
© Agilent Technologies, Inc. 2012
No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or
translation into a foreign language)
without prior agreement and written
consent from Agilent Technologies, Inc.
as governed by United States and
international copyright laws.
Manual Part Number
G4370-90020
Edition
First edition, January 2013
Printed in China
Agilent Technologies, Inc.
412 Ying Lun Road.
Waigaoqiao Free Trade Zoon
Shanghai 200131 P.R.China
安捷伦科技(上海)有限公司
上海市浦东新区外高桥保税区
英伦路 412 号
联系电话:(800)820 3278
Warranty
The material contained in this document
is provided “as is,” and is subject to
being changed, without notice, in future
editions. Further, to the maximum extent
permitted by applicable law, Agilent
disclaims all warranties, either express
or implied, with regard to this manual
and any information contained herein,
including but not limited to the implied
warranties of merchantability and
fitness for a particular purpose. Agilent
shall not be liable for errors or for
incidental or consequential damages in
connection with the furnishing, use, or
performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty terms
covering the material in this document
that conflict with these terms, the
warranty terms in the separate
agreement shall control。
Safty Notice
A CAUTION notice denotes a hazard. It calls
attention to an operating procedure, practice,
or the like that, if not correctly performed or
adhered to, could result in damage to the
product or loss of important data. Do not
proceed beyond a CAUTION notice until the
indicated conditions are fully understood and
met.
A WARNING notice denotes a hazard. It calls
attention to an operating procedure, practice,
or the like that, if not correctly performed or
adhered to, could result in personal injury or
death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully
understood and met。

Agilent 7667A Mini Thermal
Desorber Site Preparation
Customer Responsibilities 4
Dimensions and Weight 5
Power Consumption 6
Environmental Conditions 7
Gas Selection 7
7667A Connection with GC and GC-MSD 11
Basic Tools and Supplies 15
This section outlines the space and resource requirements for
Agilent 7667A Mini Thermal Desorber(Mini TD) installation.
For a successful and timely installation of the instrument, the
site must meet these requirements before beginning installation.
Necessary supplies (gases, tubing, operating supplies,
consumables, and other usage- dependent items such as columns,
and solvents) must also be available. Refer to the Agilent
website at www.agilent.com/chem for the most up-to-date listing
of GC, GC/MS, and consumables.

Customer Responsibilities
The specifications in this manual outline the necessary space,
electrical outlets, gases, tubing, operating supplies, consumables, and
other usage- dependent items such as columns and solvents required
for the successful installation of instruments and systems.
If Agilent is delivering installation and familiarization services, users
of the instrument should be present throughout these services;
otherwise, they will miss important operational, maintenance, and
safety information.
If Agilent is delivering installation and familiarization services,
delays due to inadequate site preparation could cause loss of
instrument use during the warranty period. In extreme cases, Agilent
Technologies may ask to be reimbursed for the additional time
required to complete the installation. Agilent Technologies provides
service during the warranty period and under maintenance
agreements only if the specified site requirements are met.
4 Site Preparation

Dimensions and Weight
Select the laboratory bench space before the system arrives. Pay
special attention to the total height requirements. Avoid bench space
with overhanging shelves.
The Agilent 7667A Mini Thermal Desorber installs on the left
side/top of the GC. It supports Agilent 7820A, and 5975T LTM
GC/MSD products. See Figure 1 through Figure 2 for example total
dimensions when used as a system with various Agilent products.
Table1 height, width, depth and weight
Figure 1 Example bench space needed for a G4370A - 7820A GC system
Site Preparation 5