6.2 User Interface....................................................................................................................7
6.3 Additional Service Functionalities.......................................................................................7
7 Known Bugs in SOL_1302 ....................................................................................8
8 Important Remarks................................................................................................8
printed in Germany 07 / 2001
DD+DIS154.01EModification Instructions
HOME
1 General
The Upgrade to SOL_1302 is a must Modification!
1.1 Prerequisites
As of Software version SOL_1213 more sensors are checked, error messages may occur, if the
mechanics are not adjusted properly.
Possible Problems after installing the Software
No bugs of the software, but tolerances of the hardware.
Error Reason Cure
24733 Communication problem IOBUS Replace Power Integration
board
"078F" checksum error Read error of the floppy disk Check floppy with "scandisk" or
a similar tool
Check the disk drive A:\ via
user interface.
Replace BER board
1.2 Hardware Requirements
VME – Rack
· 33 MHz Gemini Board
· EEPROM Mon 4.26
· 32 MB MEMORY Board
· BER-Board with Ethernet
· VSCB, VERA Scan-Ctrl-Board, Revision 0 or higher with Serial PROM Version SCBV3026
or higher
· SAB, Scan-Acquisition-Board, Revision 1 or higher
· VERA VME-Adapter Board, Revision 1 or higher
· Floppy (no SCSI)
· 1.2 GB SCSI Hard-Disk
· Stepper Motor Boards GS-IOB-SIN-STEP for download (no EPROM)
Revision 8 Type 5155/100/200 Section 10 / 3
Modification Instructions DD+DIS154.01E
HOME
1.3 Software Requirements
Recommended software configuration:
Either:
Configuration Preview and
One ADC-Solo
Connected to
VIPS
Connected to
Preview & ID-Stations
Connected to
MG/LR3300
Or:
Configuration Preview and
ID-Station
PRID1.1.05 VIPS1.0.05
ID-Station
Processing-
Device
Processing-
Device
Medical-
Gateway (HC)
MEGA3.32
OEM35xxx
Mon 4.03
DCM2.3.3
LTP01804
Medical-
Gateway (HC)
One ADC-Solo
Connected to
VIPS
Connected to
Preview & ID-Stations
For Direct ID / Fast Preview and/or Mammo Cassettes this combination is required:
Configuration Preview and
One ADC-Solo
Connected to
VIPS
Connected to
Preview & ID-Stations
PRID1.1.07 VIPS1.0.09
Processing-
ID-Station
> PRID1.2.09 ³ VIPS1.1.11
Device
MEGA3.32
OEM35xxx
Mon 4.03
DCM2.3.3
LTP01804
Medical-
Gateway (HC)
MEGA3.32
OEM35xxx
Mon 4.03
DCM2.3.3
LTP01804
For the option Fast Preview CCM tool version 1.1.06 is needed.
Section 10 / 4 Type 5155/100/200 Revision 8
DD+DIS154.01E Modification Instructions
HOME
1.4 Order Number:
CM+9.5155.1050.6
1.5 Scope of Delivery
· SOL_1302
4 floppy disks (1 x “HD formatter” floppy (boot floppy) , 3 x application program floppies)
· 1 floppy Language files LNG 1.005 for PRID ³ 1.1.05
· 1 empty floppy disk
· 1 upgrade instruction (DD+DIS154.01E)
Due to wrong usage, from now on the boot floppy is named "HD formatter floppy".
2 Installation of Software (Normal Upgrade-Procedure)
Upgrade to SOL_1302
Do not use the HD formatter (boot floppy)!
· Make a new backup of the machine specific data
· Select
<Service Menu>
· Follow the online instructions.
<Install from floppy>
<Software>
3 Configuration of Direct ID / Fast Preview (Options)
For the configuration of the options Direct ID and Fast Preview see DD+DIS081.01E, which is
part of this upgrade kit.
4 Restore of the Complete System (After a Hard Disk Replacement)
The restore procedure is part of the Service Documentation DD+DIS231.98E (Section 6.5)
Do not forget to restore the test images from CDROM CM+9.5145.1055.0. after the
restore procedure.
Revision 8 Type 5155/100/200 Section 10 / 5
Modification Instructions DD+DIS154.01E
HOME
5 Downgrade Procedure SOL_1302> SOLG1108
If an Erasure-Ctrl-board labeled with F8.5145.6420.X is installed, you must not
execute the downgrade procedure.
The only reason to downgrade is you cannot pass the hardware requirements. It is not
recommended to downgrade from higher software versions to lower ones. You might
run into problems if you do not follow the procedure. Even if everything works fine, you
will lose valuable infocounter information or get nonsense entries in the infocounter
file.
Prerequisites:
· Backup floppy with machine specific data
· Service - PC
· Interface cable (RS232) Digitizer => Service PC
Downgrade Procedure
· Install the software SOLG1108:
<INSTALL from floppy>
· After installation cancel the upcoming message “To initialize the new software a
reset is necessary“.
<Software>
· Change the attribute of the file ICN.DAT from read only to write access r – x > r – w – x
adcs_D:\> chmod + write ICN.DAT.
· Delete the files ICN.DAT and ICN_1001.DAT:
adcs_D:\> del *.DAT
· Immediately restore the SOL_11xx backup floppy:
<INSTALL from floppy – >
<Machine specific data>
· Activate the reset-switch on ADC-Solo's VME-Rack immediately after restoring the backup.
· End of downgrade procedure
Possible Problems after Downgrading
Error: The machine does not start-up anymore. The display just shows a blinking cursor.
Reason: If the machine accidentally tried to update the infocounters (touch the file
ICN_1001.DAT), a new file has been created. This file format is not readable by
the older software and therefore the system crashes.
Cure: Delete the file ICN_1001.DAT again and reset the machine immediately.
If this procedure does not work at all, the only solution is to start with the HD Formatter (boot)
floppy and load the software. Nevertheless, do not forget to re-install the machine specific data
and the CPF-file.
Activate the CPF with Configure - Change local network ID
Section 10 / 6 Type 5155/100/200 Revision 8
DD+DIS154.01E Modification Instructions
HOME
6 What is New in SOL_1302
6.1 Technical Specifications
· Longer light barrier flags for the scan rolls lift drive are supported. This flag improves the
clamping of the image plate.
· Rescue cycles are performed in case of former errors 247A7 and 247AA.
· Former error code 247A6 is replaced by 247AB and 247AC to show more details.
· Debouncing of the vacuum sensor is improved.
· Alignment of 8’’ x 10’’ cassettes is improved.
· Direct Identification is fully supported.
· Fast Preview is fully supported.
· Scan size 21 cm x 43 cm (asymmetric) is supported.
· Mammo Cassettes are fully supported.
· In Full Leg/Full Spine mode, the scan area has been improved.
It is, however, not possible to work in Full Leg/Full Spine mode together with Direct
Identification.
· I/V-converter signal is now adjusted periodically and tested during start-up.
· During sweep, calibration an additional error might appear due to extended tests during this
calibration. This might lead to the new error code 20C76.
· ADC Solo is able to handle hard disks > 2 GB.
· New acceleration curve for slow scan drive. This modification has an impact on the
throughput.
· Full support for Erasure Control Board F8.5155.6410.1, F8.5155.6410.3, F8.5155.6410.4.
· Correct Size Matching Factor in HT-Formula.
· MT-safe and corrected version of stepper driver and UPI-layer.
· Optimized stepper motor currents.
· The vacuum system tests are now tolerant against the latency of the sensors.
6.2 User Interface
· Name of the Service Report File has changed from d:/SOL_<fn_nr>.rpt to d:/<fn_nr>rpt.rpt
in order to have a unique name for it even after having compressed it for download through
the HTML interface.
6.3 Additional Service Functionalities
· The total amount of cycles is stored now in a list with one entry per month. This list is openended.
Revision 8 Type 5155/100/200 Section 10 / 7
Modification Instructions DD+DIS154.01E
HOME
7 Known Bugs in SOL_1302
· It is possible that the message "<workstation> not ready" appears more frequently due to
network timeouts. In such a case, a retransmission is initiated automatically. There is no
need to reset the digitizer. Even the message on the screen does not need to be confirmed
explicitly. This came along with the introduction of a new network driver that offers a
considerable higher throughput in an ordinary networking environment.
· Direct ID and "Image queue full"
If the digitizer is set to offline mode in the key operator menu and the direct identification
mode is selected, the system hangs after all image buffers (10) are full.
· In heavily loaded networking environments or due to performance restrictions of the
Preview/ID-Station, it is possible that the message "<workstation> not ready" appears more
frequently, if both, fast preview and direct identification are enabled. This will initiate a
retransmission of the image to the workstation, but there is no need for any user
intervention whatsoever.
· Image loss when power-off immediately (within a few seconds) after the scan.
· Cassette not marked “erased” when power-off at a certain position during return of IP into
the cassette.
8 Important Remarks
To avoid any formatting, the ‚format‘ program is delivered no more. 'format' would cause a
complete loss of the test images.
Section 10 / 8 Type 5155/100/200 Revision 8