3600 SERIES USER MANUAL
Programmable DC
Power supply
3600 SERIES
user’s manual
2.3 The operation of the function...................8
2.3.1 Set up voltage..................................9
2.3.2 Set up current...................................10
2.3.3 Enable/disable the output....................10
2.3.4 Store parameters..............................11
2.3.5 Recall parameters.............................12
2.3.6 The function of the Menu...................12
2.3.6.1 Set up the maximum
output voltage..............................13
2.3.6.2 Set up the maximum
output power .............................13
2.3.6.3 Enable/disable the rotary knob..........14
2.3.6.4 Set up the initial out state...............14
2.3.6.5 Save voltage setting.........................14
2.3.6.6 Set up communication speed...........14
2.3.6.7 Set up communication
address (0~31)..............................15
2.3.6.8 Lock the keypad...........................15
2.3.6.9 Clear saved parameters....................15
2.3.6.10 Exit ..............................................15

CONTENTS

Safety Guide...............................................1
Chapter 1 General Introduction ...................4
1.1 General Introduction...............................4
1.2 Specifications.........................................4
1.3 Features ..............................................5
1.4 Dimensions and Structure.......................5
1.4.1 Dimension...........................................5
1.4.2 Structure.............................................6
1.4.2.1 Front view .......................................6
1.4.2.2 Back view........................................7
Chapter 2 Operation....................................8
2.1 General operation .................................8
2.2 Function introduction..............................8
2.2.1. Main functions..................................8
2.2.2 Sub-functions.....................................8
Chapter 3 Installation of PowerMS Software
...................................................16
3.1 System Installation...............................16
3.2 System Start.........................................17
3.3 System Uninstallation...........................18
Chapter 4 The Introduction of main functions
................................................19
4.1 The Definition of the Power Supply..........19
4.2 The COM Port and Address Set.............20
4.3 Start the Communication........................21
4.4 Stop the Communication.......................21
4.5 Select POWER.....................................21
4.6 Select PC to POWER
Control Instructions..............................22
4.7 Set the Voltage Range..........................22
4.8 Set the Max Current..............................22
4.9 Query the Report..................................23
4.10 Explanation of the Interface
Indicating Components.......................24
4.11 The State Bar......................................24
4.12 Power Supply State Indication..............25
4.13 Exit the System..................................25
26
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 3
3600 SERIES USER MANUAL

SAFETY GUIDE

This manual contains the precautions necessary to ensure your personal safety as well as for protection for the products and the connected equipment. These precautions are highlighted with a triangle “W ARNING” symbol in this manual and are marked according to the danger levels as follows:
Danger
It indicates that if appropriate precautions are not taken, serious incidents of personal injuries and deaths or significant damages or losses to the properties will be caused.
Caution
It indicates that if appropriate precautions are not taken,injuries or losses of properties will take place
Note: Reminds you to pay particular attention to the important information related to the products, disposal of products or the specific part of documents.
Warning
Only qualified personnel are allowed to debug and operate this equipment. The qualified personnel are specified as those personnel who carry out commissioning, grounding and apply the volume identification to the circuits, equipment and systems according to the available safety practices and standards.
Note
Only when this product is transported, stored assembled and installed in a proper way and operated and maintained according to the recommendations, can it implement the func­tions properly and reliably.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
1
Page 4
3600 SERIES USER MANUAL
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precau­tions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. The Manufacturer assumes no liability for the customer’s failure to comply with these requirements.

Attention

I. Safety
1. Users should operate according to this manual.
2. There is high voltage inside the instrument, please avoid touching it directly.
3. Please read the user manual carefully before you use the instrument to assure your
safety.
4. Ground the Instrument
This product is provided with a protective earth terminal. To minisize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the AC power supply mains through a three-conduc­tor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet.
5.Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified service personnel. Do not split the components when power cable connected.
Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power, discharge circuits and remove external voltage sources before touching components.
Do Not Substitute Parts or Modify Instrument
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a qualified dealer for service and repair to ensure that safety features are maintained.
II. Connecting the power line
1. Inspect the power selection switch on the back panel, to assure if the selected voltage is the same as environment power voltage. If not, please refer to the notice around the power plug.
2. Before connecting the power supply, please be sure that the switch on the front panel should be in the off position.
3600 Series Programmable Power Supply
2
VERSION 2.0 2004.6
Page 5
3600 SERIES USER MANUAL
3. Connect the power cable to the AC in the power supply and the 3 connector plug.
Please be sure that the power supply is grounded.
4. Press the switch on the front panel to switch on the instrument and you can start
to use it.
III. Fuse
1. The fuse is at the position of power input part on the back panel, for the purpose of avoiding failure brought by using wrong voltage. Please pay attention to the follow­ing items when you change the voltage input and the fuse.
2. Before changing the voltage and fuse, please be sure that the AC power is switch­ing off, and be sure that there is no other equipment connected to this instrument.
3. Put a screwdriver into the fuse mounting, press it and the fuse mounting will pop up.
4. Pull out the fuse, change it according to the label beside the power input plug.
Warning: To avoid damaging the instrument, please be sure to change a suitable
type fuse.
IV. Power supply
1. For safety reasons, please be sure not to have any short circuit between plus and minus pole.
2. Output wire should be insulated completely to the connected equipment.
3. Before using the instrument, you should switch on the power supply for 30 minutes.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
3
Page 6
3600 SERIES USER MANUAL

Chapter 1 General Introduction

1.1 General Introduction

The ARRA Y 3600 Series Power Supply is a compact programmable DC power supply. It is equipped with back-light LCD display, number keypad and rotary knob which make it much easier to use. It can be operated at constant voltage mode or constant current mode. Pro­gramming feafures include over-current protection, maximum output voltage and power limitation, and 10 parameters storage. It is an essential instrument for scientific research, education, service and so on institutions.

1.2 Specification

Model 3644A 3645A 3646A Number of outputs 1 1 1
DC outputs
Voltage 0 to 18V 0 to 36V 0 to 72V Current 5A 3A 1.5A
Power(max) 90W 108W 108 W Line Regulation
Voltage 0-3.999V: 0.01% + 3mV 0-3.999V: 0.01% + 3mV 0-3.999V: 0.01% + 3mV
4-18V: 0.02% + 10mV 4-36V: 0.02% + 10mV 4-72V: 0.02% + 10mV
Current 0.02% + 8mA 0.02% + 8mA 0.02% + 8mA
Load Regulation
Voltage 0.02% + 20mV 0.02% + 10mV 0.02% + 10mV Current 0.02% + 15mA 0.02% + 10mA 0.02% + 8mA
Ripple and noise
Voltage <1mVrms <1mVrms <1mVrms Current <5mArms <4mArms <3mArms
Programming accuracy at 25 C+/- 5 C
Voltage 0.1% + 20mV 0.1% + 20mV 0.1% + 20mV Current 0.2% + 20mA 0.2% + 20mA 0.2% + 20mA
Readback accuracy at 25 C +/- 5 C
Voltage 0.2% + 20mV 0-19.999V: 0.2% + 20mV 0-19.999V: 0.2% + 20mV
20-36V: 0.2% + 100mV 20-72V: 0.2% + 100mV
Current 0.2% + 20mA 0.2% + 20mA 0.2% + 20mA
Program resolution
Voltage 0-3.999V: 1mV 0-3.999V: 1mV 0-3.999V : 1mV 4-18V: 10mV 4-36V: 10mV 4-72V: 10mV Current 1mA 1mA 1mA
Readback resolution
Voltage 10mV 0-19.999V: 10mV 0-19.999V: 10mV
20-36V: 100mV 20-72V: 100mV
3600 Series Programmable Power Supply
4
VERSION 2.0 2004.6
Page 7

1.3 Features

3600 SERIES USER MANUAL
Current 10mA 10mA 10mA
Protection Over voltage / Over current / Over power AC Input 110V/220Vac +/- 15%, 47 to 63 Hz Operating environment Weight 6.00kg Dimensions 212.6mmW x 88.1mmH x 250mmD Software PowerMS Power Management System software, ActiveX and
DLL tools support VB, VC++, DELPHI, LABVIEW
Standard Accessories Optional Accessories
* 0 to 35 C for full rated output. At higher temperaturers, the output current is derated 3% per C.
0 to 50 C, 80% RH *
User’s manual, AC cable, handlebars, software CD RS232 adaptor, RS485 adaptor , USB adaptor, Mounting rack
1. LCD display with back light.
2. Number keypad and rotary knob.
3. Maximum output voltage limitation.
4. Maximum output power limitation.
5. 10 parameters storage.
6. Can be series / parallel connected to expand output voltage/ current.
7. Free power management software.

1.4 Dimensions and Structure

1.4.1 Dimension

Fig 1 Dimension of 3600 Series Programmable DC power supply
3600 Series Programmable Power Supply VERSION 2.0 2004.6
5
Page 8
3600 SERIES USER MANUAL

1.4.2 Structure

1.4.2.1 Front view

Front panel is for users to operate.
Please see the right picture for details.
1. LCD Display
2. Number Keypad
3. Rotary knob
4. Output Terminal Fig 2 Front view of 3600 Series DC power supply
1. LCD display
0.00V 0.00A
0.00W OFF
Fig 3 LCD Display of 3600 Series
power supply
2. Arrangement of the Keyboard
Fig 4 Key board of 3600 Series DC
power supply
In common state, the keypad will execute the prompting functions of the black words. And in special mode, it will change into the functions of the blue color words.
The Left-upper Corner: The set voltage value. (Flashing voltage means the low-voltage.)
The Left-bottom Corner: The output power value. (Flashing power value means the over power.)
The Right-upper Corner: The output current value.
The Right-bottom Corner: The state value.
ON(OFF) : represents the output state of the power supply . PC : represents the operation is controlled by the computer.
0~9: The number keys Store: Save the current setting value Recall: Read the saved setting value Menu: The menu operation key OUT ON/OFF: Enable/disable the output Enter: The confirmation key V-set: The output voltage set I-set: The max output current set : The up moving key : The down moving key V/A: Set the voltage as V Set the current as A mV/ mA : Set the voltage as mV Set the current as mA
3600 Series Programmable Power Supply
6
VERSION 2.0 2004.6
Page 9
3. Rotary knob and function keys
Fig 5 Rotary knob and function keys

1.4.2.2 Back view

3600 SERIES USER MANUAL
Left Operation: The left moving key Right Operation: The right moving key ESC: Can be used to exit from any working state OK: Confirmation key Rotary knob: The rotary dial
Wire rack
fan wrap
AC input switch
Fig 6 Back view
The fuse can be changed easily by using a small screwdriver.
PC connection interface
Power supply plug
Fuse
3600 Series Programmable Power Supply VERSION 2.0 2004.6
7
Page 10
3600 SERIES USER MANUAL

2.1 General operation

1. Connect the power supply with PC
3600 Series power supply

Chapter 2 Operation

3311 adaptor

2.2 Function introduction

2.2.1. Main functions

1. Set up voltage
2. Set up current
3. Enable/disable the output
4. Store parameters
5. Recall parameters

2.2.2 Sub-functions

1. Set up the maximum output voltage
2. Set up the maximum output power
3. Enable/disable the rotary knob
4. Set up initial output state
5. Save voltage setting
6. Set up the communication speed
Fig 2-1 Connect the power supply with PC
7. Set up the communication address
8. Lock the keypad
9. Clear saved parameters

2.3 The operation of the function

There are 5 main functions and 9 sub-functions of this power supply, please refer to the
operation of these functions as follows:
3600 Series Programmable Power Supply
8
VERSION 2.0 2004.6
Page 11

2.3.1 Set up voltage

3600 series power supply provides two methods to set up the output voltage: by the number
keypad and by the rotary knob. Please see the following operation procedure.
Procedure Operation details LCD display
Step 1 Press “V-set” ENTER PASSWORD
Step 2 Enter the password ( Or jump to step 4 if the keypad ENTER PASSWORD is unlocked) Step 3 Press “Enter” button ( it will return to step 2 if your ENTER PASSWORD password is wrong for reentering) **** Step 4 Press “V-set” .Set the voltage voltage to 24.00V by using SET VOLT= 0.000V the number key or the rotary knob. NEW=
Step 5 Press “V/A” or “mV/mA” to confirm the set, SET VOLT= 24.00V it will return to step 4 for reentering if this voltage value NEW= 18 exceed the maximum output voltage. It will exit the setting up voltage operation at any procedure by press ESC button
3600 SERIES USER MANUAL
For example, how to set up the output voltage at 24.3V
1. T o set up by using number keyboard
Step1. Press “V-set” button, Step2. Enter the password by using the number keypad (if the keypad is unlocked, please do step 4), Step3. Press Enter button (if the password is wrong, please do step2 for reentering), Step4. Press “2”, “4”, “.” and “3” button to enter the voltage value, Step5. Press “V/A” button to confirm the voltage value.
2. T o set up by using Rotary knob
(1) If the keypad is unlocked by password, directly rotate the “Rotary knob”, and the voltage will be continually changed from the previews value according the rotation. At the beginning, the cursor will be shown on the last number of the value which is indicated on the LCD, you can move the cursor to the first number, second number etc by using “ ”and “ ” buttons, and then rotate to change each number, and let it stay at 24.3 V, then confirm the value by pressing “V/A” button.
(2) If the keypad is locked by password Step1. Press “V-set” button, Step2. Enter the password by using the number keypad, Step3. Press Enter button (if the password is wrong, please do step2 for reentering), Step4. Rotate the Rotary knob button to change the value, the operation is as same as item(1) Step5. Press “V/A” button to confirm the voltage value.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
9
Page 12
3600 SERIES USER MANUAL

2.3.2 Set up current

3600 Series power supply can be set up for a constant current or a maximum current, please see
the following example.
Conditions: voltage=24V, load R=12Ohm, then V/R=2 A, it represents the power supply pro­vide the load with 2A current. If the current value is set to 2.50A, the power supply works in CV mode, the actual current value should be displayed on the screen as 2.00A. If the load resistor becomes lower, the current will increase. When the current exceed 2.5A, the power supply will change to CC mode, and the output voltage will drop down to maintain the actual current value to 2.5A.
Set up current procedure as follows:
Procedure Operation details LCD display Step 1 Press “I-set” ENTER PASSWORD
Step 2 Enter the password ( Or jump to step 4 if the ENTER PASSWORD keypad is unlocked) Step 3 Press “Enter” button ( it will return to step 2 if your ENTER PASSWORD password is wrong for reentering) **** Step 4 Press “I-set” . Set up a constant current or a maximum SET CURR=0mA current by using number key or the rotary knob. NEW=3 Step 5 Press “V/A” or “mV/mA” to confirm the set. SET CURR=0mA It will return to step 4 for reentering if the current NEW= 15.0 value exceeds the maximum output current. It will exit the setting up current operation at any procedure by press ESC button

2.3.3 Enable/disable the output

By default, the power supply is disabled after power on, users can change the output status by using ON/OFF button. The button is a turnover button. When the original output is ON, press
10
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 13
it then the output will be changed to OFF status, vice versa.

2.3.4 Store parameters

If you often use a voltage and current as 24V and 2A or 12V and 2.3A and etc, you just need to set up the data at the first time and then store the data in the power supply. When you need it, recall it. 3600 series power supply can store up to 10 sets of parameters.
The parameters include 1) Voltage value; 2) Current value; 3) Maximum voltage; 4) Locked/ unlocked keypad; 5) Maximum power ; 6)Baud rate; 7) Communication address.
The store operation always be done after setting up V-set, I-set etc, the operation is as follows:
Procedure The operation Methods LCD display Step 1 Press “Store” button ENTER P ASSWORD
3600 SERIES USER MANUAL
Step 2 Enter the password ( Or jump to step 4 if the keypad ENTER PASSWORD
is unlocked)
Step 3 Press “Enter” button ( it will return to step 2 if your ENTER PASSWORD
password is wrong for reentering) 1234
Step 4 Enter the set value for store number(from 1 to 10) by SAVE 1
using the number key or rotate the rotary knob to change the set value number
Step 5 Press “Store” button to confirm the set value, if the SAVE *
number is out of the range from 1 to 10, it will retune to Step 2 for reenter It will exit the store operation at any procedure by press ESC button
For example, set up the voltage=15V, current=2A, Maximum output voltage=18V, keypad locked, Maximum output power=25W, Baud rate=9600, communication address=05, after done the setup , users can store all the above setup as a set of data, such as the 01 or 02 etc set data.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
11
Page 14
3600 SERIES USER MANUAL

2.3.5 Recall parameters

In the last paragraph, we know that we can store 10 sets of parameters. Also you can recall any one of them. It means that you needn’t to set up again for the usually requirements and it brings you much conveninent. Users can recall one set of the data from the stored data sets. Which including set up of 1)Voltage value; 2)Current value; 3)Maximum voltage; 4)Locked/ unlocked keypad; 5)Maximum power; 6)Baudrate; 7)Communication address.
The recall operation is as follows:
Procedure The operation Methods LCD display
Step 1 Press “Recall” button CALL 1
Step 2 Enter the password ( Or jump to step 4 if the keyboard ENTER PASSWORD is unlocked) Step 3 Press “Enter” button ( it will return to step 2 if your ENTER PASSWORD password is wrong for reentering) 1234 Step 4 Enter the number of the set data which you want to recall CALL 1 (from 1 to 10) by using the number key or rotate the rotary knob to change the number Step 5 Press “Recall” button to confirm , if the number is out of CALL * the range from 1 to 10, it will retune to Step 2 for reenter It will exit the Recall operation at any procedure by press ESC button
12

2.3.6 The function of the Menu

3600 series power supply provides a Menu operation for some special functions. The opera­ tion and functions are as follows:
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 15
3600 SERIES USER MANUAL
Procedure The operation Methods LCD display Step 1 Press “Menu” button MAX OUT VOLTAGE
KEY SOUND SET
Step 2 Enter the password ( Or jump to step 4 if the keyboard ENTER PASSWORD
is unlocked)
Step 3 Press “Enter” button ( it will return to step 2 if your ENTER PASSWORD
password is wrong for reentering) 1234
Step 4 The LCD display the menu functions one by one, user MAX VOLTAGE SET
can use the UP and DOWN button to change the MAX POWER SET selecting each function, Press “Enter” button to execute ROTARY SW SET the selected function INITIAL OUT SET VOLT. SAVE SET BAUDRATE SET ADDRESS SET KEY LOCK
CLEAR SAVE DATA
EXIT It will exit the Menu operation at any procedure by press ESC button
The menu operation includes MAX VOL TAGE SET, MAX POWER SET, ROTAR Y SW SET,
INITIAL OUT SET, VOLT. SAVE SET, BAUDRATE SET, ADDRESS SET, KEY LOCK and CLEAR SAVE DATA
function. We will describe the details as follows.

2.3.6.1 Set up the maximum output voltage

When you select the MAX VOLTAGE SET function, the LCD will display as:
MAX VOLT. = 24.00 V NEW=
You can set the voltage value by using the number kaypad or rotating the ROTARY knob. Then confirm the value by pressing “Enter” button.

2.3.6.2 Set up the maximum output power When you select the MAX POWER SET function, the LCD will display as:

MAX POWER=108.00W NEW=
You can set the maximum output power value by using the number keypad or totating the ROT AR Y knob. Then confirm the value by pressing “Enter” button.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
13
Page 16
3600 SERIES USER MANUAL

2.3.6.3 Enable/disable the rotary knob

When you select the ROTARY SW SET function, the LCD will display as:
SW ENABLE (Def) SW DISABLE

2.3.6.4 Set up initial output state

When you select the INITIAL OUT SET function, the LCD will display as:
INI. OUT SAVE INI. CLEAR (Def)

2.3.6.5 Save voltage setting When you select VOLT. SAVE SET function, the LCD will display as:

SAVE OUT VOLT DON’T SAVE (Def)
You can select SW ENABLE (Def) to enble the rotary
switch or select SW DISABLE to disble it .
Select INI. OUT SAVE to save the output data before switching off the unit, select INI. CLEAR (Def) to clear the output data before switching off the unit.
Select SAVE OUT VOLT. to save the last output volt­age before switching off the unit, select "DON’T SAVE (Def) to set the last output voltage as zero.
14

2.3.6.6 Set up the communication speed

This function is for monitoring the output data of the power supply by using a computer. When you select BAUDRATE SET function, the LCD will display as:
BAUDRATE=4800 BAUDRATE=9600 BAUDRATE=19200 BAUDRATE=38400
Users can change the communication setup by using UP and DOWN keys or rotating the ROTARY knob, and confirm the value by pressing “Enter” button.
BAUDRA TE 4800 represents BAUDRATE=4800bps
BAUDRA TE 9600 represents BAUDRATE=9600bps
BAUDRA TE 19200 represents BAUDRATE=19200bps
BAUDRA TE 38400 represents BAUDRATE=38400bps
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 17

2.3.6.7 Set up communication address (0~31)

This communication address function is for monitoring multi-power supply system. In the system, one computer can monitor 32 power supplies at the most by the connecting RS232 and 485 bus. So we should give each power supply an address.
When you select ADDRESS SET function, the LCD will display as:
SET ADDRESS =12 NEW=

2.3.6.8 Lock the keypad

After you locked the keypad, you must enter the correct password to unlock it, then you can use the number keys and the ROT A RY knob. This function is for the safety of the using of power supply.
3600 SERIES USER MANUAL
Users can change the communication address by us­ing number kaypad or rotating the ROTARY knob, and confirm the value by pressing “Enter” button. The range of the address is from 0 to 31.
When you select the KEY LOCK function, the LCD will display as:
ENTER P ASSWORD

2.3.6.9 Clear saved parameters

When you select CLEAR SAVE DA TA function, the LCD will display as:
CLEAR *

2.3.6.10 Exit

Users can enter 4 numbers or letters as the password by pressing the number button or by using ROTARY knob to and keys to change the number or ASCII number which will be your password, and confirm the password by pressing “Enter” button.
“*” refers to the serial number of the saved parameters.
Select “Exit” to exit the menu.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
15
Page 18
3600 SERIES USER MANUAL

Chapter 3 System Installation of PowerMS Software

3. 1 System Installation

3.1.1 Put the disk into the CDROM driver. Then select the setup powerMS and the initial diagram as in Fig.
3-1 will be displayed.
3.1.2 Then it will enter the interface as in Fig. 3-2. Press “NEXT” to continue.
Fig 3-1 The Installation Initial Interface
Fig 3-2 The Installation Interface 2
3.1.3 In Fig. 1-3, there is some explanation to some products’ introduction. Read it and press “YES” to
continue, otherwise there will be no way to install.
Fig 3-3 The Installation Interface 3
3.1.4 In Fig. 3-4, click “BROWSE” to select installation directory path. The default directory path is “C:
\Program Files\Array\PowerMS”
16
Fig 3-4 The installation Interface 4
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 19
3600 SERIES USER MANUAL
3.1.5 In Fig. 3-5, users may select the installation type. Generally, select “TYPICAL” and click “NEXT”
to continue.
Fig 3-5 The Installation Interface 5
3.1.6 In Fig. 3-6, enter the file folder’s name and the default name is “POWERMS”. Generally it is not
needed to enter and it is just need to click “NEXT”.
Fig 3-6 The Installation Interface 6
3.1.7 Click “NEXT” and the installation system will enter the files’ copy state. Please wait patiently for
the end of the files’ copy. Then the PowerMS system installation finished.
Fig 3-7 The Installation Interface 7

3.2 System Start

3.2.1 In Fig. 3-8, select the file folder of “Start | Program | Array”. And then click the “PowerMS” in the menu.
Fig 3-8 The System Start Interface
3600 Series Programmable Power Supply VERSION 2.0 2004.6
17
Page 20
3600 SERIES USER MANUAL
3.2.2 Enter the initial interface as shown in Fig. 3-9.
Fig 3-9 The System Start Diagram
3.2.3 Wait for the end of the system initialization and then it will be the main interface as shown in Fig. 3-
10.
Fig 3-10 The PowerMS Main Interface
Explanation:
1. Once the PowerMS system is started, it will be automatically in the minimized state. And this time the icon is in the state bar on the desk. Click the right key of the mouse on the icon, the menu as shown in Fig. 3-11 will be displayed.
Fig 3-11
“Show”: Show the interface. “Hide”: Hide the interface. “Start Communicate”: Start the communication. “Stop Communicate”: Stop the communication. “About PMS”: Show the help contents. “Exit System”: Close the system.
3.3 System Uninstallation
Select “ ”under the program file folder to uninstall it.
18
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 21

Chapter 4 The introduction of main functions

4.1 The Definition of the Power Supply

Select the function item and then the interface as shown in Fig. 4-1 will be displayed.
3600 SERIES USER MANUAL
Fig 4-1 The Power Definition Interface
Add: Select “Add” in the function items and input the contents of each item. After the set of the input,
select “Save” to save it.
Delete: Select the POWER record to be deleted in the table and then select “Delete”. Finally select
“Save” and it will be OK.
Modify: Select the POWER record to be modified in the table and then select “Modify” to modify it.
After the modification, select “Save” and it will be OK.
Query: Select “Query” and then wait for the name of the power supply to be queried. Show: Select “Show” and it will show all the records. Print: Select “Print” and it will print all the current records.
Parameter Explanation
Parametrer Explanation Range Remarks Power Name Name Of the Power Supply Must be Input Current Up The Max Current 0~5A/ 0~3A0~1.5A Must be Input Power Up The Max Power 0~90W/0~108W/0~108W Must be Input Voltage Configure The Max Voltage 0~18V/0~36V/0~72V Must be Input Configure ID The ID Number No Consideration
Notes: When selecting the “Add” function item to add POWER, the name and address of the POWER
cannot be repeated. After selecting the inputs, click “OK” and the dialogue frame as shown in Fig. 4-2 will be displayed. In Fig.4-3, select “YES” and the system will close and restart.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
19
Page 22
3600 SERIES USER MANUAL
Fig 4-2 System Prompt the restart

4.2 The COM Port and Lower Machine (Power Supply) Address Set

Login in the identity of “Manager” and then select the quick icon after the system restarts. The dialogue
frame as shown in Fig.4-3 will be displayed.
Fig 4-3 COM Port and address set
In Fig. 4-3, select the COM port from the pull-down table. If the COM port does not exit, the system will prompt the diagram as shown in Fig. 4-4. And click the “OK” button and the page key “ADDRESS” are out of work. And vice versa. (And users must be in the identity of the “Manager” otherwise the “ADDRESS” cannot be used.
Fig 4-4 COM Port Failure Opening frame
Set Default POWER Address: The system will automatically be in the networking state after the start according to the default COM port and the default POWER. It is just needed to enter the address in the “Default POWER” bar.
Set POWER Address: Login in the identity of the “Manager” and select the existing COM port. Then the “ADDRESS” page key will be available.
20
Fig 4-5 POWER Address Set
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 23
3600 SERIES USER MANUAL
In Fig. 4-5, enter the default address (245) of the power supply and then click “READ”. If testing successfully, the “NEW ADDRESS” and “WRITE” functions will be available. If testing failure, then the new address of the power cannot be set and the prompting diagram as shown in Fig. 4-6 will be displayed. This time the communication cable must be checked.
Fig 4-6 Communication Failure
Explanation: For the first installation each POWER must be deployed with but one address in order to communicate correctly. Set the parameter and select “OK” and it will enter into the common communication. The default COM port is COM1 and the default POWER address is 0.

4.3 Start the Communication

After the COM port and ADDRESS set, select the button and the system will start the communication. If the communication is normal, the prompt information as shown in Fig. 4-7 will be displayed. And if the commu­nication is failure, the prompt information as shown in Fig. 4-8 will be displayed.
Fig 4-7 Normal Communication Fig 4-8 Failure Communication

4.4 Stop the Communication

Select the button and the system will stop the communication.

4.5 Select POWER

In Fig. 4-9 , select the POWER name from the listing frame and it will be OK.
Fig 4-9 Selecting the POWER
3600 Series Programmable Power Supply VERSION 2.0 2004.6
21
Page 24
3600 SERIES USER MANUAL

4.6 Select PC to POWER Control Instructions

1.) Methods 1 As in Fig. 4-11, there are four control instructions in total. CLOSE POWER : Close the power output OPEN POWER : Open the power output PC CONTROL : ontrolled by PC POWER SELF : Control by power supply
Fig 4-10-1 Select the Control Instruction
Explanation: The system default control instruction is the PC CONTROL state. And when the system is
closed or the POWER is switched, the system will be automatically set to POWER SELF state.
2.) Methods 2
POWER ON/OFF Set POWER Control Set
Fig 4-10-2 Selecting Control Instruction

4.7 Set the Voltage Range

There are two methods to set the voltage range: one is by using the rotary button and the other is by using
the keypad (0.004~36.000). If you want to set accurately, please use the keypad.
Fig 4-11 Using the Rotary knob Fig 4-12 Using the Keypad
1) Using the rotary knob: Move the mouse to the icon and then rotate the knob.
2) Using the keypad: Select the “V” button, enter the data and then select the “ENTER” button.

4.8 Set the Max Current

22
There are two methods to set the voltage range: one is by using the rotary knob and the other is by using the
keypad. If you want to set accurately, please use the keypad. In general, you can use the rotary knob.
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 25
Fig 4-13 Fig 4-14
1) Using the rotary knob: Move the mouse to the icon and then rotate the knob.
2) Using the keypad: Select the “A” button, enter the data and then select the “ENTER” button.
Explanation: User must have the authority above the “General” to do set it.

4.9 Query the Report

Select the button and the diagram as shown in Fig. 4-15 will be displayed.
3600 SERIES USER MANUAL
Fig 4-15 Query the Report
1.) Set the Query Conditions: Set the parameters in the “Query” frame.
2.) Query: After setting the conditions, select “SEARCH” button and all the records compatible with the condi tions will be listed.
3.) Set the Report: Select “REPORT’ and it will be OK.
4.) Print the Report: Select “PRINT” and it will be OK.
5.) Query Totally: Select “TOTAL” and it will be OK. The data range must be selected and the other conditions cannot be selected. Its main function is to analyze several POWER so as to list the POWER that overflow the most data. The overflowing data includes the voltage overflowing, the current overflowing and the power overflowing.
6.) Delete the History Record: Select “DELETE” and the diagram as shown in Fig. 4-24 will be displayed. If you confirm to delete, select “YES” and it will be OK.
Fig 4-16 Delete the History Data
3600 Series Programmable Power Supply VERSION 2.0 2004.6
23
Page 26
3600 SERIES USER MANUAL
Explanation: The data range condition must be set.
7.) Close: Select “CLOSE” and return to the upper-interface.

4.10 Explanation of the Interface Indicating Components

1.) Instrument Part
The Voltage Value
Voltage Set Value
(Accuracy rate: 2 digits)
(Accuracy rate: 5 digits)
Common Data: Green Overflowing Data: Red
Fig 4-17 Instrument Indication Common Date and Overflowing Data are both in the normal communication state.
2.) Running Curve: Indicates the data acquired at the nearest 10 ports.
The Voltage Running Diagram The Current Running Diagram
3.) Keypad Explanation
Number Keys: 0-9 “.”: the point key “C”: the cleared key “ ”: the backspace key “V”: the voltage setting key (Unit: V) “A”: the current setting key (Unit: A) “mV”: the voltage setting key (Unit: mV ) “mA”: the current setting key (Unit: mA) “Vmax”: input the max voltage value (36V) “Amax”: input the max current value (3A) “OK”: the confirmation key
Fig 4-18 Keypad Explanation

4.11 The State Bar

Power: p3: represents the current selected POWER.
Panel Part: V: presents the current voltage set state (Unit: V). Max 36: presents that the max set voltage value is 36. 0: the current set value
24
3600 Series Programmable Power Supply VERSION 2.0 2004.6
Page 27
Add: 3: represents the POWER address. V: 36: represents the defined voltage max value. A: 3: represents the defined current max value. W: 108: represents the defined power max value. Sending: represents the communication state.

4.12 Power Supply State Indication

1. Overloading current indication: Blue presents normal. Red presents overloading.
2. Overloading power indication: Blue presents normal. Red presents overloading.
3. Power supply ON/OFF state Blue presents OFF. Red presents ON.
4.Power supply control type: Blue presents CONROL SELF. Red presents PC CONTROL.
3600 SERIES USER MANUAL

4.13 Exit the System

Select the icon to exit the system.
3600 Series Programmable Power Supply VERSION 2.0 2004.6