Preparing the
CT LUCIA 601P / 601PY and CT LUCIA 201P
To remove the lens positioning clip, place two fingers at the top
of the lens holder and use your thumb underneath the clip to
gently release it. Then lift the clip straight up.
Cover the whole lens with a generous amount of OVD
apply a small amount of OVD1 onto the silicon tip of the plunger.
Avoid touching the lens.
1
. In addition,
000000-2166-003
08/2016
2
Check that the lens is in the correct position, central and secure in
the loading chamber.
Close the lid of the loading chamber
the “click-lock” mechanism is engaged.
and listen for the click, to ensure
To advance the lens into the injector tip, gently press the plunger forward
and push the lens into the conical tip of the injector, ensuring the lens
moves smoothly forward. Retract the plunger slightly to separate
the silicone cushion tip from the IOL and ensure the haptics are properly
positioned. The IOL is now ready for implantation.
IMPORTANT:
The lens should be injected straight away. Viscoelastic materials may lose
their lubricity if allowed to stand too long while exposed to air.
Implantation
Press the injector plunger forward and then continuously apply and release pressure. Continue this “pulsing action”, to ensure
the IOL advances smoothly. The IOL will slide smoothly into the eye.
1
We recommend to use ZEISS OVDs and OVD should be at room temperature.
2
For lenses having less than 8 diopters of optical power, it is necessary to push down on the optic body with a forceps or OVD
cannula, during wing closure to ensure that the lens folds in the correct direction. When the “click-lock” mechanism is engaged,
the lens is securely in place.
0297
CT LUCIA 201P
CT LUCIA 601P / 601PY
EN_32_022_0133II
The contents of the brochure may differ from the current status of approval of the product in your country. Please contact our
regional representative for more information. Subject to change in design and scope of delivery and as a result of ongoing
technical development. CT LUCIA is a trademark of Carl Zeiss Meditec.