GUIDE TO CLEANING AND DISINFECTING ZEBRA HEALTHCARE PRINTERS
Introduction
When using Zebra healthcare printers in healthcare environments, patient safety is of paramount importance. Cleaning and Disinfecting the printer after each patient visit is essential to prevent the spread of infectious agents that cause disease and to keep the product functioning properly. Since repeated use of chemical cleaners can have an adverse eect on traditional amorphous plastics, Zebra has built its healthcare printers with rugged, UV resistant plastics designed and tested to withstand regular cleaning with all major hospital-cleaning agents, including UV­disinfection protocols, for the life of the printer. This document details how to safely clean the printer on a regular basis.
Cleaning is the removal of visible soil (e.g., organic and inorganic material) and biofilm from the product and normally accomplished manually using a dampened micro-denier microfiber wiper (eye wear lens cloth) and a scrubbing motion.
Disinfection describes a process that eliminates many or all pathogenic microorganisms on pre-cleaned surfaces. The ecacy of any disinfectant is diminished when soil is present.
Approved Cleaning Agent Ingredients
The cleaning agents used to disinfect and clean the printer must contain one of the active ingredients listed below:
• Alcohol - Isopropyl and Ethyl (≤80%)
• Ammonium - Quaternary Ammonium Compounds
• Bleach - Sodium Hypochlorite
• Hydrogen Peroxide - Accelerated or Improved
• Sodium Troclosene - Sodium Dichloroisocyanurate
Warning - It may be dangerous to combine some or all of the chemicals listed above. Read all warnings and follow the manufacturers directions provided on the packaging of each cleaning agent.
Approved Cleaning Products
Zebra recommends using pre-saturated wipes to avoid over saturating or “pooling” of cleaning agents on the printer. Never pour or spray cleaning agents directly on any surface of the product. Always use personal protection when using any cleaning agent.
Alcohols:
• Alcohol Prep Pads (PDI)
• CaviWipes™ (Metrex) - approval does not extend to CaviWipes1
• Super Sani-Cloth® (PDI)
• Sani-Cloth® Plus (PDI)
• Viraguard® Towelettes (Veridien)
• Windex® Blue (S.C. Johnson)
Hydrogen Peroxide/Acid:
• Clinical Surface Wipes (Clorox® Healthcare)
• Oxivir® TB Wipes (Diversey)
• Virox® AHP 5 RTU Wipes (Diversey)
Ammonium:
• MadaCide-FDW-Plus (Mada Medical)
• Sani-Cloth
• Super Sani-Cloth® (PDI)
• Formula 409® Glass and Surface Cleaner (Clorox)
Bleach:
• CaviWipes™ Bleach (Metrex)
• Dispatch® Disinfectant Towels (Clorox)
• Sani-Cloth® Bleach Wipes (PDI)
Soap/Saline:
• Saline 10%
• Mild dish soap
Please contact Zebra Product Management for approval before using any unlisted product or cleaning agent.
®
HB (PDI)
Caution - Cleaning agents that contain bleach can cause oxidation (corrosion) of metal components (including connectors on batteries and docking stations) resulting in permanent product damage.
Chemicals Known to Damage Plastics
The following chemicals can potentially damage the plastics on Zebra Healthcare printers and should not come in contact with plastic parts:
• Acetone
• Ammonia solutions
• Aqueous or alcoholic alkaline solutions
• Aromatic and chlorinated hydrocarbons
• Benzene
• Carbolic acid
• Ethanolamine
• Ethers
• Ketones
• TB-lysoform
• Toluene
• Trichloroethylene
Note: Some common hand sanitizers used in the healthcare environment may contain some of the harmful chemicals listed above. Please dry your hands completely and use gloves before handling the printer to prevent inadvertent contamination and damage to the plastics.
This is not a comprehensive list of chemicals known to cause damage to Zebra products. Please contact Zebra Product Management for approval before using any unlisted substance.
Regular Cleaning and Disinfecting
The following steps should be taken by healthcare workers in order to disinfect their product after each patient visit.
1. Cleaning the printer display
a. Remove any dirt particles with a damp cloth. b. Wipe the display window using a pre-moistened
wipe or dampen a soft cloth with an approved cleaning agent, taking care not to saturate the cloth.
c. After cleaning, dry with a soft non-abrasive cloth to
prevent streaking.
d. Do not allow any abrasive material to touch
the display.
e. Do not pour or spray any cleaning agent directly
onto the display window.
f. Do not let liquid pool around or on the display
window.
2. Cleaning the printer housing
a. Use a pre-moistened wipe or dampen a soft cloth
with an approved cleaning agent, taking care not to
saturate the cloth. b. Gently wipe all plastic surfaces thoroughly. c. If necessary, use a cotton-tipped applicator to reach
tight areas. d. Do not wrap the printer in the pre-moistened cloth. e. Do not pour or spray any cleaning agent directly
onto the printer.
3. Cleaning the printer keypad
a. Use a pre-moistened wipe or dampen a soft cloth
with and approved cleaning agent, taking care not
to saturate the cloth. b. Be sure to clean the printer keys, and if necessary
in between the keys, using a cotton-tipped
applicator. c. Do not pour or spray any cleaning agent directly
onto the keypad.
4. Drying
a. After cleaning, dry with a soft non-abrasive cloth
and let air dry before use. b. Always change gloves before handling the printer
to avoid re-contaminating the device.
http://www.zebra.com/locations
NA and Corporate Headquarters
+1 800 423 0442 inquiry4@zebra.com
Zebra and the stylized Zebra head are trademarks of ZIH Corp., registered in many jurisdictions worldwide. All other trademarks are the property of their respective owners. ©2018 ZIH Corp. and/or its affiliates. All rights reserved.
Asia-Pacific Headquarters
+65 6858 0722 contact.apac@zebra.com
EMEA Headquarters
zebra.com/locations contact.emea@zebra.com
Latin America Headquarters
+1 847 955 2283 la.contactme@zebra.com
P1066640-002
Loading...