When developing test sets YSI carefully sees that the tests can be carried out as safely as possible. Some hazards by dangerous substances, however, cannot always be avoided.
Caution
Improper handling of certain reagents can cause damage to your
health.
In any case follow the safety labels on the packing and the safety
instructions of the package insert.
Protective measures specified there have to be followed exactly.
Qualification of the userWe assume that, due to their professional training and experience, the
users are able to correctly understand the safety labels and safety instructions and to appropriately follow the protective measures specified
there.
1.2General information on test sets
Test sets contain special reagents that are added to the test sample according to a certain specification (analysis specification). These reagents react with the test sample. After the reaction the prepared
product is inserted in the cell shaft of the photometer in a cell and the
photometric measurement is carried out.
In addition to the reagents in the test set, further common laboratory
auxiliary reagents can be required such as acids or lyes to adjust a certain pH value.
CategoriesThe test sets can be divided into two categories depending on the way
they are carried out:
Reaction cell tests
They provide highest convenience. All the special reagents required
are prepared in a measuring cell. Normally, a certain quantity of
sample has just to be added (with some reaction cell tests, a dose
of reagent as well).
Reagent tests
They contain all reagents required for measurement. Mostly, the
sample and reagents can be prepared directly in the cell. An empty
cell is required to carry out a reagent test (diameter, see analysis
specification).
5ba76126e0103/2013
Practical partpHotoFlex®Series
Note: Before using the test with your photometer for the first time, determine the reagent blank
value.
YSI model no.:COD2 TC (MR)
Category:KT (reaction cell test)
Cell:16 mm
Measuring range:20 - 1500 mg/l COD
Shake the reaction cell
so that sediment is
Carefully pipette 2.0 ml
of sample into the cell,
Heat the cell in the
thermoreactor for two
Remove the cell from the
thermoreactor and let it
After approx. 10 min
cooling time sway the
COD MR
Program no.
310
*RN310*
Program number
Program number as bar code
1.3Carrying out photometric measurements
1.3.1The analysis specifications
Following the present practical part are the analysis specifications for
all photometric measurements that can be carried out with this photometer.
Program numberIn order to measure, you have to enter on the photometer the program
number quoted in the analysis specification. If you have connected a
bar code reader, you can read in the program number from the analysis
specification with it.
Further specifications YSI model number
Category
Cell to be used. Use suitable cells from the YSI product range only
Measuring ranges and citation forms. After measurement you can
switch between the stated citation forms.
Note
The measuring ranges specified in this Analysis Manual are valid especially for measurements with the pHotoFlex
®
Series photometers and
can be different from other data, e. g. on the package insert.
6ba76126e0103/2013
pHotoFlex®SeriesPractical part
1.3.2Reagent blank values
The evaluation of the photometric measurement always refers to the
comparison value of a sample without the substance to be determined
(reagent blank value). Thus the influence of the basic absorbance of
the reagents on photometric measurement is compensated for.
In practice, measurement of the reagent blank value is carried out with
the same amount of deionized water instead of sample.
Default reagent blank
values
Reagent blank values
without factory
default
For most tests, the reagent blank value is a constant value. It was determined in the factory and stored in the photometer. You can, however, measure the reagent blank value yourself. The default reagent
blank value is then overwritten. The default reagent blank values are
restored when the photometer is reset to default settings.
For some tests, it does not make sense to store a default reagent blank
value in the factory, e.g. if a change of certain test elements during the
storing period cannot be excluded. In this case, a reagent blank value
has to be determined before the first measurement with a new photometer. The photometer informs you if no valid reagent blank value is
available. The measured reagent blank value remains stored in the meter until a new reagent blank value is determined. When the photo-meter is reset to default, all reagent blank values that were not stored in
the factory are erased.
Note
You can increase accuracy if you determine the reagent blank value
with a test of a newly started reagent package and use the reagent
blank value for all tests of this package.
1.3.3Standard adjustment (user calibration)
The pHotoFlex
®
pH and pHotoFlex®STD photometers provide the
standard adjustment function. With this function you can adjust the factory-set calibration curve for some tests in order to optimize the accuracy if necessary. The standard adjustment especially compensates for
lot specific deviations and age-related influences on the reagents.
For which tests the standard adjustment can be used is given in the
L
ISTOFAVAILABLEPHOTOMETRICTESTSETS on page 13.
7ba76126e0103/2013
Practical partpHotoFlex®Series
1.3.4Dosing of sample and reagents
The exact dosing of the individual reagents is a precondition for the
successful carrying out of a test. The test sets contain simple accessories for correct dosing.
Instructions on using
dosing accessories pro-
vided with test sets
Dropping bottles (liquid dosing)
Hold the dropping bottle in an exactly vertical
position with the dropping tip downward and
let the reagent drip out slowly so the correct
drop size forms.
Dosing tops/measurers (powder dosing)
Screw the dosing top on the reagent bottle
instead of the screw cap. For dosing, hold
the reagent bottle in a vertical position and
for each specified dose press the lateral slide
into the measurer up to the stop once. Subsequently, close the reagent bottle with the
original screw cap so the contents cannot become moist.
Microspoon (powder dosing)
Microspoons are integrated in the screw cap
of the reagent bottles. They are available in
several colors for several dosing quantities.
Precise dosing
of liquids
8ba76126e0103/2013
Exact dosing of the sample (and of certain reagents) is especially important. Use suitable laboratory pipettes for this.
pHotoFlex®SeriesPractical part
Use of powder packsSome test sets contain reagents readily dosed as powder in small
packs. Use the powder packs as follows:
Hold the powder pack in a vertical position
as shown in the picture (perforation up)
and hit it slightly against a solid base so
the powder collects in the lower part.
Cut open the pack (tear if necessary)
along a horizontal line, starting at the perforation.
Slightly press the pack from both sides so
that a wide opening forms.
Empty the pack completely. For narrow
vessel openings use a suitable funnel top.
9ba76126e0103/2013
Practical partpHotoFlex®Series
1.4Working with the analysis timer
Many determination procedures include steps that need certain periods
of time to expire. All these periods are stored in the photometer. If the
analysis timer is switched on, a corresponding timer is activated for
each time-critical step according to the proceeding described in the
analysis specification. Only after all timers have expired can the photometric measurement be started.
1.5Sample dilution
Diluting the sample can be required for the following reasons:
The expected concentration of the substance to be determined is
near or above the upper measuring range limit
Other substances in the sample cause the measured values to be
too high or too low due to matrix interferences
With the pHotoFlex
®
Series you can measure diluted samples without
having to multiply the measured value by the dilution factor afterwards.
The measuring range is extended automatically. To do so, enter the dilution number in the photometer before measuring. Admissible are dilution numbers from 1+1 up to 1+99 (volume parts water + volume parts
test sample) in whole number steps.
For diluting, use deionized water and common laboratory dosing equipment with sufficient dosing accuracy (volumetric flasks, pipettes etc.).
Note
Note that the dilution error can increase with an increasing dilution.
Therefore, check whether a different test or method with a suitable
measuring range can be used instead of diluting the sample.
10ba76126e0103/2013
pHotoFlex®SeriesPractical part
1.6Minimizing interfering effects
1.6.1General information
The following factors can affect photometric determination and cause
incorrect measurement results:
Unsuitable pH value of the sample
Turbidity
Interferences due to certain water substances (e.g. complexing
agents often disturb the determination of metals)
Adverse temperature
Improper, especially not representative, sampling
Change of the sample due to too long storing period or unsuitable
storing conditions until measurement
Many tests have package inserts included. Read these package inserts
thoroughly. They inform you of special features to be noted during sampling, preparing the sample and carrying out the test and of possible interferences.
The following chapters provide a detailed description of important influencing factors and practical instructions on remedial actions.
1.6.2Influence of the pH value
The pH value can affect the course of chemical reactions in a photometric determination. For some tests the pH value of a solution has to
be in a certain range. With these tests, the analysis specification informs you of the necessity to check and if necessary adjust the pH value.
Adjusting the pH valueObserve the following points when adjusting the pH range:
Measure the pH value with the aid of a pH meter or pH indicator
Use the acids and lyes specified in the analysis specification
Add the acid or lye drop by drop and measure the pH value after
each drop added. Thus the volume of the sample does not increase
(is not diluted) too much
The volume increase by the drops is negligible if the resulting dilu-
tion is less than 2 %. With a greater dilution, the measurement result
should be converted accordingly. Adding up to five drops per 10 ml
of solution is uncritical as a rule of thumb.
11ba76126e0103/2013
Practical partpHotoFlex®Series
1.6.3Influence of turbidity
With samples that are visibly turbid, measured values can be oscillating
or too high or too low during photometrical determination.
Compensating for
turbidity
Depending on the type of sample or substance to be determined, the
influence of turbidity can be compensated for in different ways:
If you are positive that the substance to be determined is exclusively
in the dissolved part, the sample can be filtrated before carrying out
determination. For filtrating, simple common laboratory paper filters
or membrane filters (recommended pore size 0.45 µm) can be used.
If you assume that a considerable part of the substance to be deter-
mined is bound in the solid part of the sample, the substance has to
be brought into an analysable form before carrying out the photometric determination. This is done in a chemical digestion procedure. Suitable digestion reagents can be found in the YSI catalog.
With some procedures, the critical solid substance part is digested
during the determination itself (e. g. COD measurement in aqueous
samples with suspended matter with parts of organic compounds).
Here it is important for an exact determination that the part of suspended matter be representative for the sample. For this the sample
has to be homogenized, e.g. with a disperser.
1.6.4Influence of complexing agents
Complexing agents can disturb the determination of metals by forming
very stable compounds with them. In this form the metals cannot be analyzed. Here, a digestion has to be carried out prior to photometric determination. Suitable digestion reagents can be found in the YSI
catalog.
12ba76126e0103/2013
pHotoFlex®SeriesAnalysis specifications
2Analysis specifications
List of available photometric test sets
NameProg ModelOrder
no.
(5)
Aluminium
Al-1 TP251400Y 0.002 - 0.250 mg/l Al28 mm RTrequired 15
323
Ammonia (free)347MCA-1 TP251419Y
Ammonia
Ammonia
Ammonia
(4)
(4)
(4)
Ammonium vario
Ammonium vario HR
Ammonium vario LR
)
341NH4-1 TP251408Y (0.01 - 0.50 mg/l NH4-N)28 mm RTrequired 17
This programm includes a pH measurement by means of the integrated pH function of the meter. Thus the programm cannot
be used with the pHotoFlex
(5)
Standard adjustment (user calibration) is possible.
Measuring range
(main citation form)
2
2
2
4
4
Cell
4
4
4
4
16 mm KTrequired 53
16 mm KTrequired 54
16 mm KTrequired 55
16 mm KTrequired 56
28 mm RTrequired 57
16 mm RTrequired 58
28 mm RTrequired 59
28 mm RTrequired 60
2
28 mm RTrequired 61
28 mm RTrequired 62
Type
(1)
Blank
value
P.
14ba76126e0103/2013
5:00
Aluminium
Program no.
YSI model no.:Al-1 TP
Category:RT (reagent test)
Cell:28 mm
Measuring range:0.002 - 0.250 mg/l Al
323
*RN323*
Note: Before using this test with your photometer for the first time, determine the reagent blank
value (see notes below).
Pipette 20.0ml sample
into the empty cell.
Dissolve the powder by
shaking.
Notes:
Add the contents of a
VARIO Aluminum
ECR F20 powder pack
and close the cell with
the screw cap.
For reagent blank value
only:
Add 2 drops of VARIO ECR-Masking RGT,
close the cell with the
screw cap and mix.
Dissolve the powder by
shaking.
Allow to react for 5
minutes.
Allow to react for 30
seconds.
Insert the cell in the
photometer cell shaft
and start measurement.
Add the contents of a
VARIO Hexamine F20
powder pack and close
the cell with the screw
cap.
• To avoid errors due to contamination, rinse the equipment with hydrochloric acid solution (approx. 20%), then deionized
water beforehand.
• Adding the VARIO ECR-Masking RGT masking reagent is only required when determining the reagent blank value.
• The sample temperature has to be between 20 and 25 °C.
• We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
ba76126e0103/2013
15
5:00
5:00
Sample temperatureReaction time
(minutes)
Sample temperatureReaction time
(minutes)°C°F°C°F
541 102068 5
1050823732.5
1661625772
Ammonia (free)
Program no.
YSI model no.:MCA-1 TP
Category:RT (reagent test)
Cell:28 mm
Measuring range:0.010 - 0.500 mg/l NH3-N (f)
Perform a zero
adjustment using a
28 mm cell filled with
sample.
347
0.012 - 0.610 mg/l NH3 (f)
Display in mmol/l possible
Pipette 10.0 ml sample
into an empty 28 mm
cell.
*RN347*
Add 1 drop of Free
Ammonia Reagent
solution, close the cell
with the screw cap and
mix.
Allow to react for 5
minutes.
Add the contents of a
Vario Monochlor F
RGT powder pack and
close the cell with the
screw cap.
Shake the cell vigorously
for about 20 seconds to
dissolve solids.
Notes:
Allow to react for 5
minutes.
Insert the cell in the
photometer cell shaft
and start measurement.
When switching to
another program restore
the zero adjustment by
performing a new zero
adjustment using a
28 mm cell filled with
deionized water.
• Collect samples in clean glass bottles and analyze as soon as possible.
• The 5 minutes reaction times refer to a sample temperature of 18-20 °C (64-68 °F).
For other sample temperatures adjust the reaction time according to the follwing table:
Note: Failing to restore
the zero adjustment
will result in all future
measurements being
erroneous!
• We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
ba76126e0103/2013
16
3:00
15:00
Notes:
• The measuring range depends largely on the pH and temperature. On
the basis of the pH and temperature measurement, it is individually calculated and displayed for each determination.
• We recommend to determine a new reagent blank value (deionized
water instead of sample) for each test package started. To determine
the reagent blank value it is not necessary to measure the pH and temperature.
• For further notes please refer to the package insert of the test.
Ammonia
Program no.
YSI model no.:NH4-1 TP
Category:RT (reagent test)
Cell:28 mm
Measuring range:Corresponding to 0.01 - 0.64 mg/l NH4 or 0.01 - 0.50 mg/l NH4-N
341
Measuring ranges for NH3 or NH3-N depending on pH value and temperature,
Example: 0.002 - 0.092 mg/l NH3 at pH 8.5 and 25 °C.
*RN341*
Note: Before using tYSIhe test with your photometer for the first time, determine the reagent
blank value. To determine the reagent blank value it is not
necessary to measure the pH and
temperature.
Step 1: pH and temperature measurement
Measuring mode, pH &
ORP:
Measure pH value and
temperature immediately
after sampling.
Store measured values
with <STO>. If necessary, assign an ID for
easier retrieving.
Switch to the Photometry
measuring mode and
select program no. 341.
When the prompt Assign
pH und temp.
select and accept the
stored values from the
pH and temperature
measurement.
appears,
The meter is ready for
the photometric measurement (step 2).
The measuring range is
shown on the display.
Step 2: photometric measurement
Allow to react for 3
minutes (reaction time).
Pipette 10.0 ml of sample
into the empty cell.
Add the contents of a
VARIO AMMONIA
Salicylate F10 powder
pack and close the cell
with the screw cap.
Allow to react for 15
minutes (reaction time).
Insert the cell in the
photometer cell shaft
and start measurement.
Add the contents of a
VARIO AMMONIA
Cyanurate F10 powder
pack and close the cell
with the screw cap.
Shake the cell vigorously
to dissolve solids.
ba76126e0103/2013
17
approx.
pH 7
20:00
Notes:
• We recommend to determine a new reagent blank value (deionized water
instead of sample) for each test package started.
• If NH
4
-N is present in the sample, the solution becomes green after VARIO
AMMONIA Cyanurate F5 was added.
• If chlorine is present, sodium thiosulfate has to be added to the sample
immediately after sampling. Add 1 drop of a 0.1 mol/l sodium thiosulfate solution
per 0.3 mg/l chlorine to 1 liter sample.
• Iron disturbs the measurement and can be eliminated as follows: Determine the
total iron concentration and prepare an iron standard solution with the determined concentration. Use this solution instead of distilled water to determine the
reagent blank value for ammonium measurement.
Ammonia
Program no.
YSI model no.:NH4-2 TC (LR)
Category:KT (reaction cell test)
Cell:16 mm
Measuring range:Corresponding to 0.03 - 3.20 mg/l NH4 or 0.02 - 2.50 mg/l NH4-N
339
Measuring ranges for NH3 or NH3-N depending on pH value and temperature,
Example: 0.005 - 0.447 mg/l NH3 at pH 8.5 and 25 °C.
*RN339*
Note: Before using the test with your photometer for the first time, determine the reagent blank
value. To determine the reagent blank value it is not
necessary to measure the pH and temper-
ature.
Step 1: pH and temperature measurement
Measuring mode, pH &
ORP:
Measure pH value and
temperature immediately
after sampling.
Store measured values
with <STO>. If necessary, assign an ID for
easier retrieving.
Switch to the Photometry
measuring mode and
select program no. 339.
When the prompt Assign
pH und temp.
select and accept the
stored values from the
pH and temperature
measurement.
appears,
The meter is ready for
the photometric measurement (step 2).
The measuring range is
shown on the display.
Step 2: photometric measurement
Check the pH value of
the sample.
Desired value: approx.
pH 7.
Correct with diluted
sodium hydroxide
solution or hydrochloric
acid as necessary.
Pipette 2.0 ml of sample
into a reaction cell.
Add the contents of a
VARIO AMMONIA
Salicylate F5 powder
pack.
Allow to react for 20
minutes.
Insert the cell in the
photometer cell shaft
and start measurement.
ba76126e0103/2013
Add the contents of a
VARIO AMMONIA
Cyanurate F5 powder
pack and close the cell
with the screw cap.
Shake the cell vigorously
to dissolve solids.
18
approx.
pH 7
20:00
Notes:
• We recommend to determine a new reagent blank value (deionized water
instead of sample) for each test package started.
• If NH
4
-N is present in the sample, the solution becomes green after VARIO
AMMONIA Cyanurate F5 was added.
• If chlorine is present, sodium thiosulfate has to be added to the sample
immediately after sampling. Add 1 drop of a 0.1 mol/l sodium thiosulfate solution
per 0.3 mg/l chlorine to 1 liter sample.
• Iron disturbs the measurement and can be eliminated as follows: Determine the
total iron concentration and prepare an iron standard solution with the
determined concentration. Use this solution instead of distilled water to
determine the reagent blank value for ammonium measur
ement.
Ammonia
Program no.
YSI model no.:NH4-3 TC (HR)
Category:KT (reaction cell test)
Cell:16 mm
Measuring range:Corresponding to 0.5 - 64.4 mg/l NH4 or 0.4 - 50.0 mg/l NH4-N
340
Measuring ranges for NH3 or NH3-N depending on pH value and temperature,
Example: 0.07 - 9.37 mg/l NH3 at pH 8.5 and 25 °C.
*RN340*
Note: Before using the test with your photometer for the first time, determine the reagent blank
value. To determine the reagent blank value it is not
necessary to measure the pH and temper-
ature.
Step 1: pH and temperature measurement
Measuring mode, pH &
ORP:
Measure pH value and
temperature immediately
after sampling.
Store measured values
with <STO>. If necessary, assign an ID for
easier retrieving.
Switch to the Photometry
measuring mode and
select program no. 340.
When the prompt Assign
pH und temp.
select and accept the
stored values from the
pH and temperature
measurement.
appears,
The meter is ready for
the photometric measurement (step 2).
The measuring range is
shown on the display.
Step 2: photometric measurement
Check the pH value of
the sample.
Desired value: approx.
pH 7.
Correct with diluted
sodium hydroxide
solution or hydrochloric
acid as necessary.
Pipette 0.1 ml of sample
into a reaction cell.
Add the contents of a
VARIO AMMONIA
Salicylate F5 powder
pack.
Allow to react for 20
minutes.
Insert the cell in the
photometer cell shaft
and start measurement.
ba76126e0103/2013
Add the contents of a
VARIO AMMONIA
Cyanurate F5 powder
pack and close the cell
with the screw cap.
Shake the cell vigorously
to dissolve solids.
19
3:00
15:00
Ammonium vario
Program no.
YSI model no.:NH4-1 TP
Category:RT (reagent test)
Cell:28 mm
Measuring range:0.01 - 0.50 mg/l NH4-N
324
0.01 - 0.64 mg/l NH
Display in mmol/l possible
*RN324*
4
Note: Before using the test with your photometer for the first time, determine the reagent blank
value.
Pipette 10.0 ml of sample
into the empty cell.
Allow to react for 15
minutes (reaction time).
Notes:
Add the contents of a
VARIO AMMONIA
Salicylate F10 powder
pack and close the cell
with the screw cap.
Insert the cell in the
photometer cell shaft
and start measurement.
Allow to react for 3
minutes (reaction time).
Add the contents of a
VARIO AMMONIA
Cyanurate F10 powder
pack and close the cell
with the screw cap.
Shake the cell vigorously
to dissolve solids.
• We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
• If NH
-N is present in the sample, the solution becomes green after VARIO AMMONIA Cyanurate F10 was added.
4
• If chlorine is present, sodium thiosulfate has to be added to the sample immediately after sampling. Add 1 drop of a 0.1
mol/l sodium thiosulfate solution per 0.3 mg/l chlorine to 1 liter sample.
ba76126e0103/2013
20
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.